OLYMPUS
OTV-S300 VISERA ELITE II VIDEO SYSTEM CENTER Instructions
Instructions
486 Pages
Preview
Page 1
INSTRUCTIONS
VISERA ELITE II VIDEO SYSTEM CENTER
OLYMPUS OTV-S300
Labels and Symbols
1
Important Information - Please Read Before Use
3
Summary of Equipment Functions
13
Chapter 1
Checking the Package Contents
17
Chapter 2
Nomenclature and Functions
19
Chapter 3
Installation and Connection
39
Chapter 4
Function setup
71
Chapter 5
Inspection
251
Chapter 6
Operation
271
Chapter 7
Home button functions
283
Chapter 8
Image view function
345
Chapter 9
Reprocessing, Storage, Disposal, and Transportation
361
Chapter 10 Troubleshooting
371
Appendix
387
Contents
Contents Labels and Symbols ... 1 Important Information - Please Read Before Use ... 3 Intended use/Intended purpose ... 3 Contraindications ... 3 Clinical benefit ... 3 Applicability of endoscopy and endoscopic treatment ... 4 Instruction manual ... 4 User qualifications ... 7 Instrument compatibility ... 7 Repair and modification ... 8 Signal words ... 8 Precautions ... 8 About Security ... 12
Summary of Equipment Functions ... 13
Chapter 1 Checking the Package Contents ... 17 1.1
Checking the package contents ... 17
Chapter 2 Nomenclature and Functions ... 19 2.1
Symbols and descriptions ... 19
2.2
Front panel ... 26
2.3
Touch panel ... 28
2.4
Software keyboard ... 33
2.5
Rear panel ... 34
2.6
Side panels ... 35
2.7
Monitor ... 36
Front view ... 26
Home ... 28 Image view ... 30 Setting ... 32
OTV-S300 INSTRUCTION MANUAL
i
Contents
Chapter 3 Installation and Connection ... 39
ii
3.1
Precautions for installation and connection ... 39
3.2
Installation workflow ... 40
3.3
Installation of the equipment ... 41
3.4
Connection to the AC mains power supply ... 44
3.5
Connection of the monitor ... 47
3.6
Connection of the keyboard ... 63
3.7
Connection of the video recorder ... 64
3.8
Connection of the video printer ... 68
3.9
Connection of the foot switch ... 69
Installation on the mobile workstation ... 42 Installation in another location ... 43 When the mobile workstation is used ... 45 When no mobile workstation is used ... 46
Compatible monitors ... 47 LMD-3251ST ... 48 LMD-2451MT ... 50 Setting of the LMD-2451MT ... 52 LMD-X310ST ... 55 Setting of the LMD-X310ST ... 57 OEV262H ... 61 Compatible keyboards ... 63 Compatible video recorders ... 64 IMH-20 ... 64 IMH-10 ... 66 Compatible video printers ... 68 Compatible foot switch ... 69
OTV-S300 INSTRUCTION MANUAL
Contents
Chapter 4 Function setup ... 71 4.1
Setup list ... 71
4.2
Newly registering patient data ... 77
4.3
Editing the system setup ... 81
Patient data setup ... 71 System setup ... 72 User presets ... 74 Security settings ... 75 Maintenance ... 76
Displaying the patient data screen ... 77 Entering patient data ... 78 Editing the patient data ... 79 Deleting patient data ... 80 Loading the patient data ... 80
Using the wizard ... 81 Displaying the language, date/time screen ... 83 Setting the language ... 84 Setting the date ... 86 Setting the time ... 87 Displaying the video output screen ... 89 Setting the output from the DVI OUT terminal ... 90 Setting the output from the 3G/HD-SDI RECORDER OUT terminal ... 91 Setting the recording mode (HD-SDI) from the recorder output terminal ... 93 Setting the image signal (3G-SDI) from the 3G/HD-SDI recorder output terminal ... 94 Setting the output from the DVI OUT (3D/2D) signal ... 96 Setting the 3D image format ... 97 Setting the output from the 3G/HD-SDI OUT terminal, DVI OUT (3D/2D) ... 99 Setting the noise reduction ... 100 Displaying the monitor screen ... 103 Setting the date format ... 104 Enter comments on the monitor ... 105 Setting the brightness of the touch panel ... 107 Displaying the peripherals screen ... 110 Setting the connection of the REMOTE terminal ... 111 Setting the connection of the KEYBOARD terminal ... 112 Displaying the image record ... 114 Setting the ability of recording the endoscopic images ... 115 Setting the definition of the endoscopic images to be recorded ... 116 Setting the recording format of the endoscopic images ... 118 Setting the patient data to be saved along with the endoscopic image ... 120 Setting the auto deletion of the unsent endoscopic images ... 121 Setting the release time (HD) ... 123 Setting the release time (SD) ... 125 Setting the capture time ... 126 Displaying the link settings screen ... 128 Setting the examination start condition ... 129 Setting the examination end condition ... 130 Setting the lamp ignition timing ... 131 Setting the comments to be display on the monitor ... 133 OTV-S300 INSTRUCTION MANUAL
iii
Contents
iv
4.4
Setting a user preset ... 135
4.5
Setting the security settings ... 211
Displaying the user presets screen ... 135 Register new user preset ... 136 Editing the user presets ... 138 Copy the user presets ... 140 Delete the user presets ... 141 Displaying the switch settings screen ... 143 Setting the custom switch ... 145 Displaying the operation settings screen ... 151 Setting the recording device to save the frozen image ... 153 Setting the pre-freeze scan area ... 155 Displaying the basic image adjustment screen ... 157 Setting the iris mode ... 158 Setting the contrast ... 161 Setting the iris area ... 163 Setting the iris sensitivity ... 165 Setting the image rotation ... 167 Displaying the WLI image adjustment screen ... 169 Setting the brightness mode ... 171 Setting the brightness ... 173 Setting the color tone and chroma ... 175 Setting the color mode ... 177 Setting the NBI cysto-color adjustment ... 179 Setting the AGC ... 180 Setting the AGC MAX gain ... 182 Setting the enhancement mode ... 184 Displaying the display preferences screen ... 187 Setting the endoscopic image size ... 188 Setting the data display on the monitor ... 190 Selecting the data displayed on the monitor ... 192 Setting the data displayed at start of exam ... 195 Setting the text size on the monitor ... 197 Setting the text color on the monitor ... 199 Setting the message and thumbnail layout on the monitor ... 201 Setting the data layout on the monitor ... 203 Setting the data layout on the right side of the monitor ... 205 Setting the aspect ratio ... 207 Setting the index image display time ... 209 Displaying the security settings screen ... 211 Setting the security function ... 212 Setting the security target ... 213 Registering an account ... 214 Editing an account ... 216 Deleting an account ... 217 Changing a password ... 218 Entering an ID and password ... 219
OTV-S300 INSTRUCTION MANUAL
Contents
4.6
Setting the maintenance ... 221
Displaying the maintenance screen ... 221 Displaying the export settings screen ... 222 Export all the settings ... 224 Export the user presets ... 225 Export the patient data ... 226 Export the system setup ... 227 Export the security settings (account) ... 227 Displaying the import settings screen ... 228 Import all the settings ... 230 Import the user presets ... 231 Import the patient data ... 232 Import the system setup ... 233 Import the security settings (Account) ... 233 Displaying the serial numbers ... 234 Displaying the message history ... 235 Displaying the user presets settings ... 238 Displaying the system setup settings ... 239 Displaying the security settings ... 240 Displaying the scope information ... 241 Displaying the test image on the monitor ... 242 Format the portable memory ... 243 Test the portable memory ... 244 Backup the settings and endoscopic images ... 245 Restore settings and endoscopic images ... 247 Format internal memory ... 248 Execute the system reset ... 249
Chapter 5 Inspection ... 251 5.1
Precautions of workflow ... 251
5.2
Inspection workflow ... 251
5.3
Connection of an endoscope ... 253
5.4
Inspection of the power supply ... 258
5.5
Inserting the portable memory (MAJ-1925) ... 259
5.6
Monitor display inspection ... 260
5.7
Inspection of the monitor display ... 260
5.8
Inspection of the examination light ... 261
5.9
Inspection of the brightness adjustment function ... 264
VISERA ELITE series videoscope ... 254 Fiberscope ... 255 Rigid endoscope and camera head ... 256
Inspection of the automatic brightness adjustment ... 265 Inspection of the manual brightness adjustment ... 266
5.10 Inspection of the freeze function ... 266 5.11 Inspection of the release function ... 267 OTV-S300 INSTRUCTION MANUAL
v
Contents
5.12 Inspection of the endoscopic image rotation function ... 267 5.13 Inspection of the custom switches ... 267 5.14 Inspection of the 3D/2D switch function ... 268 5.15 After inspection ... 269
Chapter 6 Operation ... 271 6.1
Precautions for operation ... 271
6.2
Operation workflow ... 275
6.3
Turning ON the video system center and ancillary equipment ... 276
6.4
White balance adjustment ... 276
6.5
Load of user settings ... 276
6.6
Patient data ... 277
6.7
Observation and recording of the endoscopic image ... 279
6.8
Termination of the operation ... 280
Entering directly patient data with the touch panel ... 277 Entering directly patient data with the keyboard ... 278 Loading the patient data ... 278
Chapter 7 Home button functions ... 283
vi
7.1
Function list ... 283
7.2
Using the frequently used functions ... 285
7.3
Using the function for observation or recording ... 293
Loading the user preset ... 285 Igniting the examination lamp ... 287 Extinguishing the examination lamp ... 288 Adjusting the white balance ... 289 Setting the brightness ... 291
Displaying the observation and recording screen ... 293 Changing the 3D/2D mode ... 293 Changing the observation mode ... 295 Freezing the endoscopic image ... 297 Recording the frozen image (Release) ... 298 Recording the live image with the video recorder ... 300 Recording the frozen image only with the video printer (Capture) ... 301 Changing the image enhancement ... 302 Changing the endoscopic image size ... 303 Changing the zoom ratio ... 304 Rotating the endoscopic image ... 306
OTV-S300 INSTRUCTION MANUAL
Contents
7.4
Using the color or brightness functions ... 307
7.5
Using function related to the display ... 321
7.6
Using the prepare functions ... 332
Displaying the color screen ... 307 Adjusting the brightness ... 308 Changing the iris mode ... 309 Changing the brightness mode ... 310 Adjusting the color tone ... 311 Changing the color mode ... 312 Changing the NBI cysto-color adjustment ... 314 Changing the contrast ... 315 Changing the iris area ... 316 Changing the iris sensitivity ... 317 Changing the laser mode ... 319 Changing the AGC ... 320 Displaying the information screen ... 322 Changing the data display ... 322 Entering a comment ... 323 Changing the text size ... 325 Changing the text color ... 326 Displaying the scope information ... 327 Displaying the switch information ... 328 Displaying the arrow pointer ... 329 Changing the Cursor ... 331 Displaying the preparation screen ... 332 Igniting the examination lamp ... 333 Extinguishing the examination lamp ... 334 Adjusting the white balance ... 335 Loading the user preset ... 336 Entering the patient data ... 338 Loading the patient data ... 340 Executing the exam end ... 341 Saving the user preset ... 342 Locking the touch panel ... 343 Resetting the current user preset ... 344
Chapter 8 Image view function ... 345 8.1
Deleting, exporting, and capturing the endoscopic images ... 345
8.2
Confirming the memory information ... 354
Displaying the image view screen ... 345 Deleting the folder with the endoscopic images ... 346 Deleting the endoscopic images ... 346 Printing the endoscopic images with the video printer ... 348 Exporting the folder with the endoscopic images to the portable memory ... 350 Exporting the endoscopic images to the portable memory ... 352 Image files and folders ... 355 Playing back the images using the personal computer ... 358
OTV-S300 INSTRUCTION MANUAL
vii
Contents
Chapter 9 Reprocessing, Storage, Disposal, and Transportation . 361 9.1
Reprocessing ... 361
9.2
Surface disinfectant cleaner ... 363
9.3
Signs of degradation from reprocessing ... 364
9.4
Preparing equipment for reprocess ... 365
9.5
Storage ... 369
9.6
Disposal ... 370
9.7
Transportation ... 370
Equipment needed ... 365 Reprocessing video system center and accessories ... 366
Chapter 10 Troubleshooting ... 371 10.1 Troubleshooting ... 371 10.2 Troubleshooting guide ... 372 10.3 Returning the video system center for repair ... 386
Appendix ... 387 Combination equipment ... 387 System chart ... 387
Specifications ... 390 Transportation, storage, and operation Environment ... 390 Specifications ... 391
Default settings list ... 397 EMC information ... 401 License information of Open Source Software ... 407 GNU GPL ... 407 Wind River® VxWorks® ... 409 End User License Agreement of Open Source Software modules ... 409 Wind River® Diab Compiler ... 441 End User License Agreement of Open Source Software modules ... 441 Wind River® VxWorks® GNU C/C++ Compiler ... 452 End User License Agreement of Open Source Software modules ... 452 Others ... 455 newlib ... 455
Index ... 473
viii
OTV-S300 INSTRUCTION MANUAL
Labels and Symbols
Labels and Symbols Safety-related labels and symbols are attached on the locations shown below. If labels or symbols are missing or illegible, contact Olympus.
Rear panel Blue
Indicates to refer to the instruction manual for the cable to be connected. Refer to the instruction manual.
Authorized representative in the European Community
CE marking
Manufacturer Potential equalization terminal
Electric rating The product name, rated voltage, and frequency are shown.
Serial number
Date of manufacture
OTV-S300 INSTRUCTION MANUAL
MEDICAL DEVICE VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300 POWER INPUT: 100–240 V 50/60 Hz 400 VA
1
Labels and Symbols
Front panel
CAUTION Dry the video connector before insertion. ATTENTION Vaillez séchez le connecteur vidéo avant de l’insérer.
UDI label A label required by some countries’ regulations regarding identification of medical device also known as Unique Device Identification (UDI).
Back cover of this instruction manual Symbol
Description Manufacturer
Authorized representative in the European Community
Importer (into European Union)
Translation
2
OTV-S300 INSTRUCTION MANUAL
Important Information - Please Read Before Use
Important Information - Please Read Before Use
Intended use/Intended purpose This video system center is intended to be used with OLYMPUS camera heads, endoscopes, monitors, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis, treatment, and video observation.
Contraindications The use of medical devices not specifically designed for cardiac applications may cause ventricular fibrillation or seriously affect the cardiac function of the patient. As specified by the international standard IEC 60601-1, any applied part used for observation or treatment of the heart or areas near the heart must meet “TYPE CF applied part” requirements for low electrical leakage current. When using endoscopes for endoscopic cardiac applications, the applied part requirements include all devices directly connected to the endoscope, such as the light guide cable, camera head and telescope holder. Each of these devices must individually meet the “TYPE CF applied part” requirements for leakage current limits if they are to be used for cardiac applications. Other combinations of equipment may cause ventricular fibrillation or seriously affect the cardiac function of the patient.
Clinical benefit Provides a detailed view of anatomical structures which led to an improvement of safety and efficacy of the surgical procedure.
OTV-S300 INSTRUCTION MANUAL
3
Important Information - Please Read Before Use
Applicability of endoscopy and endoscopic treatment If there is an official standard on the applicability of endoscopy and endoscopic treatment that is defined by the healthcare facility’s administration or other official institutions such as academic societies on endoscopy, follow that standard. Before starting endoscopy and endoscopic treatment, thoroughly evaluate its properties, purposes, effects, and possible risks (their natures, extent, and probability). Perform endoscopy and endoscopic treatment only when its potential benefits are greater than its risks. Fully explain to the patient the potential benefits and risks of the endoscopy and endoscopic treatment as well as any examination/treatment methods that can be performed in its place, and perform the endoscopy and endoscopic treatment only after obtaining the consent of the patient. Even after starting the endoscopy and endoscopic treatment, continue to evaluate the potential benefits and risks, and immediately stop the endoscopy/treatment and take proper measures if the risks to the patient become greater than the potential benefits.
Instruction manual This instruction manual contains essential information on using the video system center safely and effectively. Before use, thoroughly review this manual and the manuals of all equipment that will be used during the procedure and use the equipment as instructed. Keep this and all related instruction manuals in a safe, accessible location. If you have any questions or comments about any information in this manual, contact Olympus.
Terms used in this manual Camera head: The camera head is a device that converts endoscopic images taken by a fiberscope or rigid endoscope into electrical signals.
Scope information: Inherent information about the camera head or videoscope.
Video printer: The video printer is a device that prints the frozen video image.
Video recorder: The video recorder is a device that records the video.
Hospital grade wall mains outlet: The hospital grade wall mains outlet is a wall AC mains power outlet socket having the exclusive terminal for grounding.
4
OTV-S300 INSTRUCTION MANUAL
Important Information - Please Read Before Use
Isolation transformer: The isolation transformer is a safety device that is used to isolate the noninsulated equipment with potentially high leakage currents to decrease the possibility of an electric shock.
Automatic brightness adjustment: The automatic brightness adjustment automatically adjusts the intensity of the light emitted from the light source so that the endoscopic image will be maintained at constant brightness even if the distance between the distal end of the endoscope’s insertion tube and the subject changes.
Manual brightness adjustment: The manual brightness adjustment function is used to adjust manually the intensity of the light emitted from the light source.
Image sensor: The image sensor is a device that converts light into electrical signals.
Electric zoom: The electric zoom function changes the zoom ratio of the image.
Iris: The iris function is used to electrically measure the brightness of an endoscopic image.
Color adjustment: Color adjustment adjusts the color balance and the chroma on the monitor.
Cysto-color adjustment: Cysto-color adjustment adjusts the color when observing the inside of the bladder.
Freeze: The freeze function makes a frozen image of a live image.
Prefreeze: The prefreeze function is used to display the image with the least blur when being frozen.
Release: The release function is used to record an endoscopic image.
Capture: Stores the image in the ancillary equipment.
Index image: The index image is displayed on the monitor when executing the release function. OTV-S300 INSTRUCTION MANUAL
5
Important Information - Please Read Before Use
Contrast: This is the ratio of brightness between the brightest and darkest areas in an image.
Edge enhancement: Edge enhancement is an image processing technique that electrically sharpens the edges of an image to increase sharpness.
Structure enhancement: Structure enhancement is an image processing technique that electrically emphasizes the detailed patterns and edges of an image to increase sharpness.
AGC (Automatic Gain Control): The AGC is used to increase the brightness of an endoscopic image electrically when the brightness of the image is dark because the distance between the endoscope’s distal end and the object is too long.
Image rotation: The image rotation function is a function that rotates the endoscopic image by 180 degrees.
Portable memory: A digital medium for storage of images, etc.
Internal memory: This is space to store the frozen images, patient data, and user settings data.
Wash-out: Wash out is the inability to see details in the endoscopic image due to excessive brightness.
SDTV: Standard Definition Television. It is the format used in standard video systems.
HDTV: High Definition Television. This is a format for high resolution video transmission featuring higher definition than the standard SDTV format.
3D SIDE BY SIDE: This is a mode that transmits and displays the right and left images generated from two image sensors of the 3D videoscope by horizontally compressing them and arranging them side by side on the monitor.
6
OTV-S300 INSTRUCTION MANUAL
Important Information - Please Read Before Use
3D LINE BY LINE: This is a mode that transmits and displays the right and left images generated from two image sensors of the 3D videoscope by arranging the lines from the left image and from the right image alternately.
WLI (White Light Imaging) observation: This is observation using white light.
Optical-digital observation: This is observation using specific filtered light.
NBI (Narrow Band Imaging) observation: This is optical-digital observation using narrow band light.
Custom switch: The custom switch is a general term for the buttons, keys, and switches to which functions can be assigned: “CUSTOM” key on the keyboard, remote switch on the endoscope, remote switch on the foot switch, and custom button on the touch panel.
User qualifications This instrument should be used by persons trained in the use of this instrument. If there is an official standard on user qualifications to perform endoscopy and endoscopic treatment that is defined by the medical administration or other official institutions, such as academic society on endoscopy and endoscopy physicians, follow that standard. If there is no official qualification standard, the operator of this instrument must be a physician approved by the medical safety manager of the healthcare facility or person in charge of the department (department of internal medicine, etc.). The physician should be qualified to operate and safety perform the planned endoscopy and endoscopic treatment following guidelines set by the academic societies on endoscopy, etc., and considering the difficulty of endoscopy and endoscopic treatment. This manual does not explain or discuss endoscopic procedures.
Instrument compatibility Refer to “ System chart” on page 387 to confirm that the video system center is compatible with the ancillary equipment being used. Using incompatible equipment may result in patient injury or equipment damage and makes it impossible to obtain the expected functionality. This instrument complies with EMC standard for medical electrical equipment, edition 2 (IEC 60601-1-2: 2001) and edition 3 (IEC 60601-1-2: 2007).
OTV-S300 INSTRUCTION MANUAL
7
Important Information - Please Read Before Use
Repair and modification The video system center does not contain any user-serviceable parts. Do not disassemble, modify, or attempt to repair it; patient or operator injury, equipment damage and/or the impossibility to obtain the expected functionality may result. Some problems that appear to be malfunctions may be correctable by referring to Chapter 10, “Troubleshooting”. If the problem cannot be resolved using the information in Chapter 10, contact Olympus. This instrument is to be repaired by Olympus technicians only. Olympus is not liable for any injury or damage that occurs as a result of repairs attempted by non-Olympus personnel.
Signal words The following signal words are used throughout this manual:
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.
NOTE
Indicates additional helpful information.
Precautions Follow the warnings and cautions given below when handling the video system center. This information is to be supplemented by the warnings and cautions given in each chapter.
WARNING • Do not use this instrument for any purpose other than its “ Intended use/Intended purpose” on page 3. • Strictly observe the following precautions. Failure to do so may place the patient and medical personnel in danger of an electric shock. When the video system center is used to examine a patient, do not allow metal parts of the endoscope or its accessories to touch metal parts of other system components. Such contact may cause unintended current flow to the patient. Keep fluids away from all electrical equipment. If fluids are spilled on or into the unit, stop operation of the video system center immediately and contact Olympus. Do not prepare, inspect, or use the video system center with wet hands.
8
OTV-S300 INSTRUCTION MANUAL
Important Information - Please Read Before Use
WARNING • Never install and operate the video system center in locations where: The concentration of oxygen is high; Oxidizing agents (such as nitrous oxide (N2O)) are present in the atmosphere; Flammable gases are present in the atmosphere; Flammable liquids are near. Otherwise, an explosion or a fire may result because the video system center is not explosion-proof. • In case of video system center failure or malfunction, always keep another video system center ready for use. • Never insert anything into the ventilation grills of the video system center. It can cause an electric shock and/or a fire. • Do not look directly into the distal end of the endoscope, the distal end of the light guide cable, or the output socket of the video system center when they are emitting light. The intense light may cause eye injury. • Do not touch the distal end of the light guide cable or the output socket of the light source immediately after disconnecting it from the video system center because they are extremely hot. Operator or patient injury can result. • Although the illumination light emitted from the endoscope’s distal end is required for endoscopic observation and treatment, it may also cause alteration of living tissues such as protein denaturation of biological tissue by inappropriate use. Observe the following warnings on the illumination. Always set the minimum required brightness. The brightness of the image on a video monitor may differ from the actual brightness at the distal end of an endoscope. Be sure to use the automatic brightness control function. This function can keep the illumination light properly. Do not continue observation in proximity to tissue or keep the distal end of the endoscope in contact with living tissue for a long time. When discontinuing the use of the endoscope, extinguish the examination lamp by tapping the lamp button. • This product may interfere with other medical electronic equipment used in combination with it. Before use, refer to “ System chart” on page 387 to confirm the compatibility of the video system center with all equipment to be used. • Do not use the video system center in any place where it may be subject to strong electromagnetic radiation (for example, in the vicinity of a microwave therapeutic equipment, MRI, wireless set, short-wave therapeutic equipment, cellular/portable phone, etc.). This may impair the performance of the video system center.
OTV-S300 INSTRUCTION MANUAL
9
Important Information - Please Read Before Use
WARNING • When turning ON the video system center, never allow the distal end of an endoscope or a light guide cable to come in contact with the patient and/or other flammable materials, such as operating room drapes. Patient injury and/or a fire can result. • If the endoscopic image dims during use, blood, mucus, or debris may have adhered to the light guide cable on the distal end of the endoscope. Carefully withdraw the endoscope from the patient and remove the blood or mucus to obtain optimum illumination and to ensure the safety of the examination. If you continue to use the endoscope in such a condition, the distal end temperature may rise and cause mucosal burns. It may also cause patient and/or operator injury. • Use only the power cord specified by Olympus. Otherwise, equipment malfunction and/or power cord burns may result. Also, never use it for other equipment. Contact Olympus. • Do not rely on the optical-digital observation method alone for primary detection of lesions or for a decision regarding any potential diagnostic or therapeutic intervention. • Do not use the video system center when the live image cannot be observed. Otherwise, patient injury may occur. • To display endoscopic images, connect the output terminal of the video system center directly to the monitor. Do not make the connection via any ancillary equipment. Images may disappear during observation depending on the condition of the ancillary equipment. • The SDI signal is transmitted by daisy-chaining (chaining in a row) several devices. If one of the devices is turned OFF, the SDI signal is not transmitted to the rest of them. • Electromagnetic interference may occur to the video system center when it is placed near equipment marked with the following symbol or other portable and mobile RF communications equipment such as cellular phones. If radio interference occurs, mitigation measures may be necessary, such as reorienting or relocating the video system center or shielding the location.
10
OTV-S300 INSTRUCTION MANUAL
Important Information - Please Read Before Use
CAUTION • Use the video system center only under the conditions described in “Transportation, storage, and operation Environment” on page 390 and “Specifications” on page 391. Otherwise, improper performance, compromised safety, and/or equipment damage may result. • Do not allow a foreign object to penetrate the inside of the video connector socket, output socket, and portable memory port. The video system center may malfunction. • Do not use a pointed or hard object to press the buttons on the front panel, touch panel, and/or keyboard. This may damage the buttons. • Do not touch the electrical contacts inside the video system center’s connectors by hands. Equipment damage and/or malfunction may result. • Do not apply excessive force to the video system center and/or other instruments connected. Otherwise, damage and/or malfunction can occur. • Do not connect or disconnect the endoscope or camera head while the video system center is turned ON. Connecting or disconnecting the endoscope while the video system center is ON may destroy the image sensor. Turn the video system center OFF before connecting or disconnecting the endoscope. • Do not leave the examination lamp ON when an endoscope is connected to the video system center. The examination light reaches the maximum intensity and the endoscope’s distal end becomes hot. In addition, smoke may also be produced if the debris attached to the distal end is heated. • Avoid using the video system center in a dusty environment. This may damage the video system center. • Confirm that there is no dust on the ventilation grills. Clean and vacuum dust from the ventilation grills using a vacuum cleaner, when necessary. Otherwise, the video system center may break down and get damaged from overheating. • Be sure that the video system center is not used adjacent to or stacked with other equipment (other than the components of the video system center or system) to avoid electromagnetic interference.
NOTE In case any serious incident happens to User or Patient while operating the device, the incident should be reported to Olympus and the competent authority of the respective member state or country.
OTV-S300 INSTRUCTION MANUAL
11