OLYMPUS
PSF-1 INTESTINAL VIDEOSCOPE Operation Manual Jan 2020
Operation Manual
152 Pages
Preview
Page 1
OPERATION MANUAL
INSTRUCTIONS
INTESTINAL VIDEOSCOPE
Symbols
1
Important Information - Please Read Before Use
2
Chapter 1
Checking the Package Contents
15
Chapter 2
Instrument Nomenclature and Specifications
19
Chapter 3
Preparation and Inspection
31
Chapter 4
Operation
83
Chapter 5
Troubleshooting
113
OLYMPUS PSF-1
Appendix
Refer to the endoscope’s companion manual, the “REPROCESSING MANUAL” with your endoscope model listed on the cover, for reprocessing information. USA: CAUTION: Federal law restricts this device to sale by or on the order of a physician.
127
Contents
Contents Symbols ... 1 Important Information - Please Read Before Use ... 2 Intended use ... 2 Contraindications ... 2 Applicability of endoscopy and endoscopic treatment ... 3 Instruction manual ... 4 User qualifications ... 4 Instrument compatibility ... 5 Reprocessing before the first use/reprocessing and storage after use ... 5 Spare equipment ... 5 Maintenance management ... 5 Prohibition of improper repair and modification ... 6 Signal words ... 6 Precautions ... 6 Precautions for disappeared or frozen endoscopic image ... 12 Examples of inappropriate handling ... 13
Chapter 1 Checking the Package Contents ... 15 1.1
Checking the package contents list ... 15 Packaged items for the Americas, Europe, Australasia, Middle East, and Africa ... 16 Packaged items for countries other than the Americas, Europe, Australasia, Middle East, and Africa ... 17
Chapter 2 Instrument Nomenclature and Specifications ... 19 2.1
Nomenclature and functions ... 19 Control section, insertion section ... 20 Endoscope connector ... 24
2.2
Specifications ... 26 Environment ... 26 Function list ... 27 Specifications ... 28
Chapter 3 Preparation and Inspection ... 31 3.1
The workflow of preparation and inspection ... 31
3.2
Preparation of the equipment ... 33
3.3
Inspection of the endoscope ... 35 Inspection of the endoscope ... 35 Inspection of the bending mechanisms ... 41
PSF-1 OPERATION MANUAL
i
Contents
3.4
Inspection of accessories ... 45 Inspection of the air/water and suction valves ... 45 Inspection of the biopsy valve (MB-358) ... 46 Inspection of the single use biopsy valve (MAJ-1555) ... 47 Inspection of the auxiliary water inlet cap (MAJ-215) ... 47 Inspection of the auxiliary water tube (MAJ-855) ... 48 Inspection of the single use PowerSpiral tube (DPST-1) ... 49 Inspection of the mouthpiece (for transoral insertion only) ... 49
3.5
Attaching accessories to the endoscope ... 50 Attaching the suction valve ... 50 Attaching the air/water valve ... 52 Attaching the biopsy valve ... 53 Attaching the auxiliary water inlet cap ... 54
3.6
Inspection of ancillary equipment ... 54
3.7
Connection of the endoscope and ancillary equipment ... 55 Connection to the light source ... 55 Connection of the water container ... 56 Connection of the suction tube ... 57 Connection of the auxiliary water tube ... 58
3.8
Inspection of the endoscopic system ... 59 Inspection summary ... 59 Inspection of the endoscopic image ... 60 Inspection of the remote switches ... 61 Inspection of the air-feeding function ... 62 Inspection of the objective lens cleaning function ... 63 Inspection of the suction function ... 65 Inspection of the instrument channel ... 67 Inspection of the auxiliary water feeding function ... 67
3.9
Attachment of the single use PowerSpiral tube (DPST-1) and inspection of the rotation function ...69 Attachment of the single use PowerSpiral tube (DPST-1) ... 70 Inspection of the single use PowerSpiral tube rotation function by the inspection mode ... 75
Chapter 4 Operation ... 83 4.1
Precautions ... 83
4.2
Insertion ... 87 Holding and manipulating the endoscope ... 87 Insertion of the endoscope with the single use PowerSpiral tube (DPST-1) ... 88 Observation of the endoscopic image ... 91 Angulation of the distal end ... 91 Air/water feeding and suction ... 92
ii
PSF-1 OPERATION MANUAL
Contents
4.3
Using EndoTherapy accessories ... 98 Insertion of EndoTherapy accessories into the endoscope ... 100 Operation of EndoTherapy accessories ... 101 Withdrawal of EndoTherapy accessories ... 102 Use of nonflammable gases ... 102 High-frequency cauterization treatment ... 103
4.4
Withdrawal of the endoscope ... 104
4.5
Removal of the single use PowerSpiral tube (DPST-1) ... 107
4.6
Transportation of the endoscope ... 110 Transporting within the hospital ... 110 Transporting outside the hospital ... 111
Chapter 5 Troubleshooting ... 113 5.1
Troubleshooting ... 113
5.2
Troubleshooting guide ... 114 Angulation ... 114 Air/water feeding ... 114 Suction ... 115 Image quality or brightness ... 116 Auxiliary water feeding ... 117 EndoTherapy accessories ... 117 Rotation function ... 117 Other ... 118
5.3
Withdrawal of the endoscope with an irregularity ... 119 Withdrawal when the WLI and NBI endoscopic images appear on the monitor ... 119 Withdrawal when either the WLI or the NBI endoscopic image does not appear on the monitor ... 120 Withdrawal when no endoscopic image appears on the monitor or a frozen image cannot be restored ... 121 Withdrawal when the rotation function does not work ... 122 Withdrawal when limit stop function frequently works ... 123 Retrieval of the dropped single use PowerSpiral tube from the endoscope ... 124
5.4
Returning the endoscope for repair ... 126
PSF-1 OPERATION MANUAL
iii
Contents
Appendix ... 127 Combination equipment ... 127 System chart ... 127 Reprocessing equipment ... 130 Compatible video system center ... 131 Compatible light source ... 131 Compatible EndoTherapy accessories ... 131 Compatible electrosurgical accessories ... 135
Labeling information for irrigation system ... 137 Backflow-prevention valve ... 137 Distal irrigation system and proximal irrigation system ... 137 Consumable device ... 139 Reusable device ... 139
EMC information ... 140
iv
PSF-1 OPERATION MANUAL
Symbols
Symbols The meaning(s) of the symbol(s) shown on the component packaging, the back cover of the instruction manual, and/or the instrument are as follows: Symbol
Description Refer to instructions.
Endoscope
TYPE BF applied part
Single use only
Lot number
Manufacturer
Date of Manufacture
Authorized representative in the European Community
Serial number
Ingress protection rating is 7.
For US Customers only For a Symbols Glossary, visit us: http://www.olympus-global.com/en/common/pdf/symbolsglossary.pdf
PSF-1 OPERATION MANUAL
1
Important Information - Please Read Before Use
Important Information - Please Read Before Use
Intended use This instrument is intended to be used with an Olympus video system center, light source, single use PowerSpiral tube, PowerSpiral control unit, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. The INTESTINAL VIDEOSCOPE PSF-1 is indicated for use within the upper digestive tract (including the esophagus, stomach, duodenum, and small intestine) by oral insertion or the lower digestive tract (including the anus, rectum, sigmoid colon, colon, ileocecal valve, and small intestine) by anal insertion.
Contraindications Patients not suitable for a prolonged endoscopic procedure under deep anesthesia or general endotracheal intubation including: • Medical instability preventing anesthesia • Inability to gain consent • Known perforation • Uncontrolled coagulopathy • Recently placed feeding jejunostomy (e.g. less than two weeks) • Pediatric patients, especially infants and toddlers • Patients having a stent or other instruments implanted in the intestinal tract that may obstruct the passage of the single use PowerSpiral tube. For antegrade approach • Perforated ulcer • Esophageal or gastric varices • Foregut stenosis • Deep mucosal laceration • Suspected or diagnosed eosinophilic esophagitis • Unable to accept biteblock
2
PSF-1 OPERATION MANUAL
Important Information - Please Read Before Use
For retrograde approach • Severe active inflammation of colon • Anal stenosis • Colonic stricture
Applicability of endoscopy and endoscopic treatment If there are official standards on the applicability of endoscopy and endoscopic treatment that are defined by the hospital’s administrators or other official institutions, such as academic societies on endoscopy, follow those standards. Before starting endoscopy and endoscopic treatment, thoroughly evaluate its properties, purposes, effects, and possible risks (their nature, extent, and probability). Perform endoscopy and endoscopic treatment only when its potential benefits are greater than its risks. Fully explain to the patient the potential benefits and risks of the endoscopy and endoscopic treatment as well as any examination/treatment methods that can be performed in its place, and perform the endoscopy and endoscopic treatment only after obtaining the consent of the patient. Even after starting the endoscopy and endoscopic treatment, continue to evaluate the potential benefits and risks, and immediately stop the endoscopy/treatment and take proper measures if the risks to the patient become greater than the potential benefits.
PSF-1 OPERATION MANUAL
3
Important Information - Please Read Before Use
Instruction manual This instruction manual contains essential information on using this instrument safely and effectively. Before use, thoroughly review this manual and the manuals for all equipment that will be used during the procedure and use the equipment as instructed. Note that the complete instruction manual set for this endoscope consists of this manual and the “REPROCESSING MANUAL” with your endoscope model listed on the cover. It also accompanied the endoscope at shipment. Keep this and all related instruction manuals in a safe, accessible location. If you have any questions or comments about any information in this manual, contact Olympus.
Terms used in this manual NBI (Narrow Band Imaging) observation: This is optical-digital observation using narrowband light.
WLI (White Light Imaging) observation: This is observation using white light.
Image sensor: The image sensor is a device that converts light into electrical signals.
User qualifications If there are official standards for user qualifications to perform endoscopy and endoscopic treatment that are defined by the hospital’s medical administrators or other official institutions, such as academic societies on endoscopy, follow those standards. If there are no official qualification standards, the operator of this instrument must be a physician approved by the medical safety manager of the hospital or person in charge of the department (department of internal medicine, etc.). The physician should be capable of safely performing the planned endoscopy and endoscopic treatment following guidelines set by the academic societies on endoscopy, etc., and considering the difficulty of endoscopy and endoscopic treatment. This manual does not explain or discuss endoscopic procedures.
4
PSF-1 OPERATION MANUAL
Important Information - Please Read Before Use
Instrument compatibility Refer to “Combination equipment” on page 127 to confirm that this instrument is compatible with the ancillary equipment being used. Using incompatible equipment can result in patient or operator injury and/or equipment damage. This instrument complies with the EMC standard for medical electrical equipment, edition 4 (IEC 60601-1-2: 2014).
Reprocessing before the first use/reprocessing and storage after use This instrument was not reprocessed before shipment. Before using this instrument for the first time, reprocess it according to the instructions given in the endoscope’s companion “REPROCESSING MANUAL” with your endoscope model listed on the cover. After using this instrument, reprocess and store it according to the instructions given in the endoscope’s companion “REPROCESSING MANUAL”. Improper and/or incomplete reprocessing or storage can pose an infection control risk, cause equipment damage, or reduce performance.
Spare equipment Be sure to prepare another endoscope to avoid interruption of the examination due to equipment failure or malfunction.
Maintenance management The probability of failure of the endoscope and ancillary equipment increases as the number of procedures performed and/or the total operating hours increase. In addition to the inspection before each procedure, the person in charge of medical equipment maintenance in each hospital should periodically inspect the items specified in this manual following applicable regulations, guidelines, etc. required of you. An endoscope with an observed irregularity should not be used, but should be inspected by following Section 5.2, “Troubleshooting guide”. If the irregularity is still observed after inspection, contact Olympus. The covering of the endoscope’s rotation part needs to be replaced. A maintenance instruction is displayed by the force gauge. When the instruction is displayed, contact Olympus.
PSF-1 OPERATION MANUAL
5
Important Information - Please Read Before Use
Prohibition of improper repair and modification This instrument does not contain any user-serviceable parts. Do not disassemble, modify, or attempt to repair it; patient or operator injury and/or equipment damage may result. Equipment that has been disassembled, repaired, altered, changed, or modified by persons other than Olympus’ own authorized service personnel is excluded from Olympus’ limited warranty and is not warranted by Olympus in any manner.
Signal words The following signal words are used throughout this manual:
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.
NOTE
Indicates additional helpful information.
Precautions Follow the warnings and cautions given below when handling this endoscope. This information is to be supplemented by the warnings and cautions given in each chapter.
WARNING • Do not use this endoscope for any purpose other than its intended use. Patient or operator injury and/or equipment damage may result. • After using this endoscope, reprocess and store it according to the instructions given in the endoscope’s companion “REPROCESSING MANUAL” with your endoscope model listed on the cover. Using improperly or incompletely reprocessed or stored instruments may cause patient cross-contamination and/or infection. • Before endoscopy, remove any metallic objects (watch, glasses, necklace, etc.) from the patient. Performing high-frequency cauterization treatment while the patient is wearing metallic objects may cause burns on the patient in areas around the metallic objects.
6
PSF-1 OPERATION MANUAL
Important Information - Please Read Before Use
WARNING • Do not strike, hit, or drop the distal end, insertion tube, bending section, control section, universal cord, or endoscope connector of the endoscope. Also, do not bend, pull, or twist the distal end, insertion tube, bending section, control section, universal cord, or endoscope connector of the endoscope with excessive force. The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations. It could also cause parts of the endoscope to fall off inside the patient. • Never perform angulation control forcibly or abruptly. Never forcefully pull, twist, or rotate the angulated bending section. Patient injury, bleeding, and/or perforation may result due to unintended retroflexion of the bending section. It may also become impossible to straighten the bending section during an examination. • Never insert or withdraw the endoscope’s insertion section while the bending section is locked in position. Patient injury, bleeding, and/or perforation may result. • Never operate the bending section, feed air or perform suction, insert or withdraw the endoscope’s insertion section, or use EndoTherapy accessories without viewing the endoscopic image or while the endoscopic image is frozen. Patient injury, bleeding, and/or perforation may result. • Never insert or withdraw the insertion section abruptly or with excessive force. Patient injury, bleeding, and/or perforation may result. • If it is difficult to insert the endoscope, do not forcibly insert the endoscope; stop the endoscopy. Forcible insertion can result in patient injury, bleeding, and/or perforation. • When using the electronic zoom function of the video system center, never insert or withdraw the endoscope’s insertion section or use EndoTherapy accessories while the image is magnified. Patient injury, bleeding, and/or perforation can result. • Do not touch the light guide on the endoscope connector immediately after removing it from the light source because it is extremely hot. Operator or patient burns can result. • If the endoscopic image becomes dimmer during the procedure, it may indicate that blood or mucus is adhering to the light guide lens on the distal end of the endoscope. Rotate the single use PowerSpiral tube backward and withdraw the endoscope from the patient, remove blood or mucus, and confirm that the light guide lens has no irregularities for continued use. If you continue to use the endoscope with an obstructed light guide lens, the temperature at the distal end of the endoscope may rise, which may cause patient injury or operator and/or patient burns. • When the endoscopic image does not appear on the monitor, the image sensor may have been damaged. Turn the video system center OFF immediately. Continued power supply in such a case will cause the distal end of the endoscope to become hot and could cause operator and/or patient burns. PSF-1 OPERATION MANUAL
7
Important Information - Please Read Before Use
WARNING • Do not rely on the NBI observation mode alone for primary detection of lesions or to make a decision regarding any potential diagnostic or therapeutic intervention. • Do not use this endoscope without attaching the single use PowerSpiral tube. When the endoscope is inserted into the body cavity without attaching the single use PowerSpiral tube, patient injury, bleeding, and/or perforation may be caused. • The limit function is not intended to guarantee patient safety. The PowerSpiral control unit has a limit function to stop rotation when the motor current exceeds a certain level, which protects the endoscope’s rotation part. It must be noted that the limit function is not intended to guarantee patient safety. The physician should carefully evaluate the clinical factors so that an excessive rotational load will not be applied to the patient’s body cavity. Be sure to examine the following conditions that change a load on the rotation part during inspection mode. How many level bar indicators are lit. How the reactive forces felt by the right hand holding the insertion section increase. Based on these kinds of information, decide an allowable rotational load. Do not continue the procedure if such allowable rotational load cannot be determined. Continuing to give an excessive load may cause patient injury, bleeding, and/or perforation in the patient’s body cavity. • Do not forcibly advance, retract, and twist the endoscope regardless of rotation while the single use PowerSpiral tube is in the body cavity, which may result in patient injury, bleeding, and/or perforation. • Do not continue to rotate the single use PowerSpiral tube, if the distal end of the endoscope does not move while the single use PowerSpiral tube is rotating. Continuing to rotate the single use PowerSpiral tube in the same location may result in patient injury, bleeding, and/or perforation. • Do not twist the insertion section of the endoscope when the single use PowerSpiral tube is put under heavy rotational load, which may further increase the rotational load. This may result in patient injury, bleeding, and/or perforation. • Before performing endoscopy or treatment, thoroughly evaluate conditions of the patient. Passage of the single use PowerSpiral tube may result in patient injury, bleeding, and/or perforation depending on a patient’s condition. • When inserting the endoscope, thoroughly observe the inside of the body cavity to evaluate whether it is appropriate to continue the endoscopy. Passage of the single use PowerSpiral tube may result in patient injury, bleeding, and/or perforation.
8
PSF-1 OPERATION MANUAL
Important Information - Please Read Before Use
WARNING • When performing endoscopy in a patient with adhesions and it cannot be confirmed the endoscopy will be performed safely, the procedure must be stopped. However, when its potential benefits are greater than its risk, carefully insert or withdraw by rotating the endoscope through a part of adhesion. Patient and/or adhesion injury, bleeding, and/or perforation may result. • The safety of this endoscopy has not been established in patients with the following conditions. Perform endoscopy and endoscopic treatment in these patients only when its potential benefits are greater than its risks. Known stricturing diseases such as Crohn’s disease Any prior abdominal or pelvic surgery including altered anatomy Pregnancy Radiation enteritis History of dysphagia or known esophageal swallowing disorders Mild to moderate inflammation of colon • If it is assumed that the single use PowerSpiral tube may not be extracted from the patient’s body cavity when the rotation mechanism does not function, then do not use the endoscope for the procedure. Passage of the single use PowerSpiral tube may result in patient injury, bleeding, and/or perforation. • Carefully determine the applicability of a patient with a history of surgery in the gastrointestinal tract. Passage of the single use PowerSpiral tube may result in patient injury, bleeding, and/or perforation. • During endoscopic treatment, keep the insertion section and the bending section as straight as possible. If there is a loop or a bend on the insertion section or the bending section, the operation cannot be performed as intended, and patient injury, bleeding, and/or perforation can result. • Some adverse events due to endoscopic examinations and treatments such as mucosal damage, perforation, mucosal inflammation, bleeding, infection, fever, pain (such as abdominal pain and sore throat), abdominal distention, discomfort (such as abdominal discomfort and swallowing discomfort), digestive dysfunction (such as indigestion, vomiting and nausea), dysphagia (such as odynophagia), respiratory system disorder (such as mediastinal emphysema, respiratory instability, hypoxemia, aspiration pneumonia and cough), hiccup, cardiovascular problem (such as bradycardia, tachycardia, hypotension and hypertension), pancreatitis, hyperamylasemia, hyperlipasemia, intestinal obstruction, intestinal necrosis, intussusception, and parotitis have been reported. Not only during but also after an endoscopic examination with this product, observe the patient and check these adverse events to prevent danger. Take appropriate measures if necessary.
PSF-1 OPERATION MANUAL
9
Important Information - Please Read Before Use
CAUTION • Do not pull the universal cord during an examination. The endoscope connector will be pulled out from the output socket of the light source and the endoscopic image will disappear. • Do not coil the insertion tube or universal cord with a diameter of less than 12 cm. Equipment damage may result. • Do not attempt to bend or twist the endoscope’s insertion section with excessive force. The insertion section may be damaged. • Do not apply shock to the distal end of the endoscope, in particular the objective lens surface at the distal end of the endoscope. An abnormal endoscopic image may result. • Do not twist or bend the bending section with your hands. Equipment damage may result. • Do not squeeze the bending section and the rotation part forcefully. The covering of the bending section and the rotation part may stretch or break and cause water leakage. • Turn the video system center ON only when the endoscope connector is connected to the light source. In particular, confirm that the video system center is OFF before connecting or disconnecting the endoscope connector. Failure to do so can result in equipment damage, including destruction of the image sensor. • The endoscope’s remote switches cannot be removed from the control section. Pressing, pulling, or twisting them with excessive force can break the switches and/or cause water leakage. • If remote switch 1 does not return to the OFF position after being pressed strongly from the side, gently pull the switch upwards to return it to the OFF position. • Do not hit or bend the electrical contacts on the endoscope connector. The connection to the light source may be impaired and a faulty contact can result. • If endoscope’s suction is insufficient, select another suction system without using the endoscope and use it according to the directions given in its instruction manual. Otherwise, a proper endoscopic image may not appear on the monitor. • Do not continue to rotate the single use PowerSpiral tube when the force gauge indicator or the force gauge indicator in endoscopic images shows a high rotational load. The motor case and rotation part may overheat. The temperature of the rotation part may reach 48C (118F), resulting in burns on the patient and/or the operator.
10
PSF-1 OPERATION MANUAL
Important Information - Please Read Before Use
CAUTION • Electromagnetic interference may occur on this endoscope near equipment marked with the following symbol or other portable and mobile RF (Radio Frequency) communications equipment, such as cellular phones. If electromagnetic interference occurs, mitigation measures may be necessary, such as reorienting or relocating this endoscope, or shielding the location.
• Be sure that this endoscope is not used adjacent to or stacked with other equipment (other than the components of this endoscope or system) to avoid electromagnetic interference.
NOTE • This endoscope contains a memory chip that stores information about the endoscope and communicates this information to the video system center CV-190 PLUS or CV-190. • When the motor case is close to the receiver dish of UPD-3, the shape of the insertion section may be displayed improperly.
PSF-1 OPERATION MANUAL
11
Important Information - Please Read Before Use
Precautions for disappeared or frozen endoscopic image WARNING • If the endoscopic image disappears unexpectedly or the frozen image cannot be restored during an examination, immediately stop using the endoscope and withdraw it from the patient as described in Section 5.3, “Withdrawal of the endoscope with an irregularity”. Inserting or withdrawing the endoscope, using EndoTherapy accessories, performing suction, feeding air, or performing angulation control under these conditions could result in patient injury, bleeding, and/or perforation. • Follow the precautions given below. Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination. Connect the endoscope connector to the light source completely by pushing the endoscope connector until it clicks. Otherwise, a faulty contact can result. Do not bend, hit, pull, or twist the insertion section, bending section, control section, universal cord, and endoscope connector. The endoscope may be damaged, and water leakage and/or breakage of internal parts like the cable can result. Before connecting the endoscope connector to the light source, confirm that the endoscope connector, including the electrical contacts, is completely dry and clean. If the endoscope is used with the electrical contacts wet and/or dirty, the endoscope and light source may malfunction. If air bubbles emerge from the endoscope continuously during the leakage test, do not use the endoscope. Water may enter the endoscope and cause a short circuit. This may result in image sensor damage.
CAUTION • Turn the video system center ON only when the endoscope connector is connected to the light source. In particular, confirm that the video system center is OFF before connecting or disconnecting the endoscope connector. Failure to do so can result in equipment damage, including destruction of the image sensor. • Do not hit or bend the electrical contacts on the endoscope connector. The connection to the light source may be impaired and a faulty contact can result.
12
PSF-1 OPERATION MANUAL
Important Information - Please Read Before Use
Examples of inappropriate handling Details on clinical endoscopic technique are the responsibility of trained specialists. Patient safety in endoscopic examinations and endoscopic treatment can be ensured through appropriate handling by the physician and the medical facility. Examples of inappropriate handling are described below.
WARNING • Over-insufflating the lumen may cause patient pain, injury, bleeding, and/or perforation. • Applying suction with the distal end of the endoscope in prolonged contact with the mucosal surface, with higher suction pressure than required, or with prolonged suction time may cause bleeding and/or lesions. • The endoscope has not been designed for use in retroflexed observation in parts of the body other than the stomach. Performing retroflexed observation in a narrow lumen may make it impossible to straighten the angle of the bending section and/or withdraw the endoscope from the patient. Retroflexed observation in parts of the body other than the stomach should be performed only when its usefulness is determined to be greater than the risk that is posed to the patient. Also, do not operate the endoscope forcibly in retroflexed observation. • Inserting, withdrawing, and using EndoTherapy accessories without a clear endoscopic image may cause patient injury, burns, bleeding, and/or perforation. • Rotating the single use PowerSpiral tube in the direction the endoscope is inserted, inserting or withdrawing the endoscope, feeding air, applying suction, or operating the bending section without a clear endoscopic image may cause patient injury, bleeding, and/or perforation. • When keeping rotating the endoscope forward and the endoscopic image is not clear by positioning the endoscope’s field of view improperly, patient injury, bleeding, and/or perforation may result. • For reasons described below, do not rely on the NBI*1 observation mode alone for primary detection of lesions or to make a decision regarding any potential diagnostic or therapeutic intervention. NBI has not been demonstrated to increase the yield or sensitivity of finding any specific mucosal lesion, including colonic polyps or Barrett’s esophagus. *1
Narrow Band Imaging. For more details, refer to the instruction manual for the video system center CV-190 PLUS or CV-190.
PSF-1 OPERATION MANUAL
13
Important Information - Please Read Before Use
14
PSF-1 OPERATION MANUAL
1.1 Checking the package contents list
Chapter 1
1.1
Checking the Package Contents Ch.1
Checking the package contents list
Match all items in the package with the components shown below. Inspect each item for damage. If the endoscope is damaged, a component is missing, or you have any questions, do not use the items; immediately contact Olympus.
Endoscope
PSF-1
PSF-1 OPERATION MANUAL
15