UHI-3 Cylinder Hose O-Ring Replacement Procedure Instructions Dec 2021
Instructions
2 Pages
Preview
Page 1
INSTRUCTIONS Replacement Procedure for Packing (O-Ring) of the Cylinder hose (PIN) for UHI-3 (MAJ-1080) and Cylinder hose with switch-over valve (Pin-index) (MAJ-1985) If the packing (O-ring) of the aforementioned product is scratched, cracked, or missing replace it with a new O-ring as described in the following procedure. Handle
Packing (O-ring) Yoke
O-ring lock nut
Packing (O-ring) replacement procedure Preparation Order an Olympus O-ring GT8450 for MAJ-1080 or MAJ-1985. Use the following tools*1 to replace the packing. *1 Not offered by Olympus. • Wrench (6 mm) • Tweezers (without sharp tips)
CAUTION • Be sure to use the packing (O-ring) designated by Olympus. Use of a non-designated packing (O-ring) could lead to a gas leak. • Do not use tweezers for attaching a new packing (O-ring). Otherwise, the packing (O-ring) may be damaged, which could lead to a gas leak. • Be sure to use the designated wrench for attaching or detaching the O-ring lock nut. Otherwise, the O-ring lock nut may be damaged.
1 Disconnect the high-pressure hose from the CO2 cylinder and high flow insufflation unit or endoscopic CO2 regulation unit.
CAUTION When disconnecting the high-pressure hose, be sure to close the CO2 cylinder valve and exhaust CO2 remaining in the instrument. Otherwise, a large amount of gas will jet out of the instrument. When disconnecting the high-pressure hose from the CO2 gas inlet and CO2 cylinder, ensure that the CO2 cylinder valve is closed, then release any residual CO2 remaining inside the instrument.
2 Remove the handle from the yoke by turning the handle counterclockwise.
Figure 1
3 Using the wrench (6 mm), remove the O-ring lock nut from the yoke by turning the O-ring lock nut counterclockwise.
Figure 2
–1–
4 Using the tweezers, remove the old packing (O-ring) from the yoke. 5 Inspect the new packing (O-ring) for any damages (e.g.cracks, scratches). If no damage is detected, place a new packing (O-ring) onto the yoke.
6 Using your finger, place the packing (O-ring) and make sure that it is fit in the groove. Groove
Figure 3
7 Using the wrench (6mm), tighten the O-ring lock nut (with a force of about 2 N-m (reference), for example).
Figure 4
NOTE If the O-ring lock nut is damaged or deformed, please purchase a new O-ring lock nut (GT8451) of the model designated by Olympus.
8 Attach the handle on the yoke by turning the handle clockwise.
Figure 5
9 Inspect that the packing (O-ring) has no scratches or cracks. 10 Check the connection of the CO2 cylinder by following the description on the inspection page in the instruction manual of the high flow insufflation unit or endoscopic CO2 regulation unit. Should a gas leak be suspected, disconnect the high-pressure hose and re-attach the packing (O-ring). If a gas leak still occurs, do not use the equipment and contact Olympus.
Manufactured by 2951 Ishikawa-cho, Hachioji-shi, Tokyo 192-8507, Japan
The meaning(s) of the symbol(s) are as follows: Manufacturer
Fax: (042)646-2429 Telephone: (042)642-2111
Distributed by (Premises/Goods delivery) Wendenstrasse 20, 20097 Hamburg, Germany (Letters) Postfach 10 49 08, 20034 Hamburg, Germany Fax: (040)23773-4656 Telephone: (040)23773-0
Authorized representative in the European Community
3500 Corporate Parkway, P.O. Box 610, Center Valley, PA 18034-0610, U.S.A. Fax: (484)896-7128 Telephone: (484)896-5000
Importer (into European Union)
5301 Blue Lagoon Drive, Suite 290 Miami, FL 33126-2097, U.S.A. Fax: (305)261-4421 Telephone: (305)266-2332
KeyMed House, Stock Road, Southend-on-Sea, Essex SS2 5QH, United Kingdom Fax: (01702)465677 Telephone: (01702)616333
A8F, Ping An International Financial Center, No. 1-3, Xinyuan South Road, Chaoyang District, Beijing, 100027 P.R.C. Fax: (86)10-5976-1299 Telephone: (86)10-5819-9000
(Seocho-dong, Majestar City Tower One), 2F & 3F, 12, Seocho-daero 38-gil, Seocho-gu, Seoul, 06655, Republic of Korea Fax: (02)6255-3210 Telephone: 1544-3200
Elektrozavodskaya str. 27 bld.8, 107023 Moscow, Russia Fax: (7)495-926-7072 Telephone: (7)495-926-7077
438B Alexandra Road #03-07/12, Alexandra Technopark Blk B, Singapore 119968 Fax: 6834-2438 Telephone: 6834-0010
3 Acacia Place, Notting Hill, VIC 3168, Australia Fax: (03)9543-1350 Telephone: (03)9265-5400
Date of Issue 2021-12-08
RA0628 03
–2–
©2014 OLYMPUS MEDICAL SYSTEMS CORP. All rights reserved. Printed in Japan 20150113 *0000