OLYMPUS
WM-NP2 MOBILE WORKSTATION Instructions Issue 3 Oct 2011
Instructions
100 Pages
Preview
Page 1
INSTRUCTIONS 取扱説明書 MOBILE WORKSTATION モービルワークステーション
WM-NP2
USA: CAUTION: Federal law restricts this device to sale by or on the order of a physician.
Contents Important Information - Please Read Before Use...
5
Intended use...
5
Instruction manual...
5
Signal words...
5
Handling precautions...
6
Chapter 1
Standard Sets...
9
1.1
Standard sets of components...
10
1.2
Optional accessories...
11
Chapter 2
Symbols...
12
2.1
Top tray...
12
2.2
Separation transformer...
12
Chapter 3
Preparation for Use...
14
3.1
Castors...
14
3.2
Shelf size...
15
3.3
Fixing positions of intermediate shelves...
16
3.4
Fixing position of the drawer unit...
20
3.5
Clip-in accessories...
24
3.6
Scope hanger assembly...
26
3.7
Keyboard assembly...
33
3.8
Equipment installation...
34
3.9
Power connector...
37
3.10 General information...
37
Chapter 4
Inspection...
39
Chapter 5
Operation...
40
5.1
Operation...
40
5.2
Manoeuvring...
41
Chapter 6
Care, Storage and Maintenance...
43
6.1
Care...
43
6.2
Storage...
44
6.3
Maintenance...
45
Chapter 7
Specifications...
46
7.1
WM-NP2 mobile workstation...
46
7.2
Separation transformer...
46
Chapter 8
Spare parts...
48
Chapter 9
Technical Description...
49
MOBILE WORKSTATION WM-NP2
1
2
Chapter 10
Troubleshooting...
51
Chapter 11
End of Life...
52
MOBILE WORKSTATION WM-NP2
Important Information - Please Read Before Use
Important Information - Please Read Before Use Intended use The Olympus WM-NP2 mobile workstation is intended for use in medical facilities under the direction of a trained physician, and has been designed to be used with a range of Olympus equipment to facilitate GI endoscopy, endoscopic ultrasound, respiratory and surgical endoscopic procedures. Do not use the equipment for any purpose other than its intended application. The safety and performance of an endoscopic system depends not only on the endoscope but also on the ancillary equipment used with it. Ensure any ancillary equipment is compatible with the endoscope and other equipment used.
Instruction manual Please read this manual carefully before attempting to use the Olympus WM-NP2 mobile workstation as it contains important information on the proper care, handling and use of the equipment. In addition, read the manuals of any other units (light source, endoscope, etc.) that form part of the system. If you have any questions or comments regarding the information in this manual, please contact Olympus. These instructions should be retained for reference during the life of the product.
Signal words The following signal words are used throughout this manual: WARNING
CAUTION
NOTE
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.
Indicates additional helpful information.
MOBILE WORKSTATION WM-NP2
5
Important Information - Please Read Before Use
Handling precautions Follow the warnings and cautions given below when handling this workstation. This information is to be supplemented by the warnings and cautions given in each chapter. WARNING
• The WM-NP2 mobile workstation should only be used in a medical facility under the direction of a trained physician. • Explosion hazard - never install or use the WM-NP2 mobile workstation within the zone of risk of flammable gases. The workstation is fitted with antistatic castors, none of the medical electrical equipment with which the workstation is intended to be used is ‘AP’ or ‘APG’ rated, therefore never install or use the WM-NP2 mobile workstation within the zone of risk of flammable gases. • To prevent patient shock, the combination of equipment used on the WM-NP2 mobile workstation should never be applied directly to the heart. • After wiping with a piece of moistened gauze, dry the workstation and mounted devices thoroughly before using it again. If it is used while still wet, there is the risk of an electric shock. • Use personal protective equipment to guard against dangerous chemicals and potentially infectious material, otherwise, blood, mucus and other potentially infectious material from the patient could pose an infection risk. During cleaning, wear appropriate protective equipment, such as eye wear, face mask, moisture-resistant clothing and chemicalresistant gloves that fit properly and are long enough so that your skin is not exposed. • To prevent user or patient cross-contamination, do not hang contaminated endoscopes on the scope hanger. • The WM-NP2 should be used in accordance with the following. Otherwise, the workstation will be ‘top heavy’ and may lose stability and topple, causing equipment damage and/or personal injury. Do not attach more than one LCD monitor arm to each side
of the workstation. LCD monitor arms and monitors should not be installed and used on an empty workstation. LCD monitor arms and monitors should be installed after all other equipment is installed on an appropriate place on the workstation. • (USA only): The power cord on this unit is fitted with a ‘hospital grade’ plug. Grounding reliability can only be achieved when this equipment is connected to an equivalent receptacle marked ‘hospital grade’.
6
MOBILE WORKSTATION WM-NP2
Important Information - Please Read Before Use
WARNING
• (RoW only): The power cable on this unit is terminated in flying leads, for a suitable plug to be fitted according to the area of intended use. The cores of the cable are coloured in accordance with the following code: BROWN LIVE (‘HOT’) BLUE NEUTRAL GREEN/YELLOW EARTH (GROUND) Only suitably qualified personnel should fit the power plug to this cable.
CAUTION
• Do not exceed the load capacity detailed in Chapter 7. • Do not lean or stand on any part of the mobile workstation or accessory as failure of the product may result. • When assembling equipment onto the mobile workstation, the installation should meet the requirements of EN IEC 60601-1-1:2001, the safety standard for medical electrical systems. • Before commissioning an installation including a separation transformer, ensure its rating will not be exceeded by the equipment connected to it. • When configuring a system, the user should ensure that there is sufficient spare equipment available that in the event of a fault in any part of the system, the endoscopy procedure may be completed or terminated without endangering the patient. • When installing the mobile workstation in the procedure or operating room, ensure that the mains cable has sufficient slack that it cannot be pulled taut as the mobile workstation is moved around the working area. If necessary, relocate the mains connector to a nearer supply socket. • Section 5.2 Manoeuvring gives detailed instructions and precautions which must be read and understood before attempting to move the mobile workstation. • Spray type medical agents, such as lubricants, anaesthetic or alcohol, should be used away from the mobile workstation. • Keep open containers of liquid away from electrical equipment used on the mobile workstation to prevent operator electrical shock and equipment damage. Do not use the system if liquid has entered the equipment. • To minimise risk of injury, it is recommended that heavy items are placed on the mobile workstation by two people working together. Ensure electrical items used on the mobile workstation are positioned centrally on the shelves. • Never place any items on top of a monitor (if fitted) as these may topple when the mobile workstation is moved. • The electrical resistance of the conductive antistatic castors should be tested at regular intervals in accordance with ISO 2878 to assess antistatic effectiveness. • Avoid contact of the castor wheels with oil and avoid a build up of floor cleaning wax and polish as these will impair the antistatic effectiveness of the castors.
MOBILE WORKSTATION WM-NP2
7
Important Information - Please Read Before Use
CAUTION
• Store and use the WM-NP2 mobile workstation within the environmental conditions described in Chapter 7; failure to do so may lead to equipment malfunction or failure. • Before attempting to move the workstation outside the procedure room, ensure that the endoscope and fluid containers are removed. • There are no operator-serviceable components on the WM-NP2 mobile workstation. Refer servicing to qualified service personnel. • Before commencing shelf adjustment, ensure equipment is disconnected and removed from the shelf. • To minimise the risk of injury and/or damage, do not fit accessories other than those described in Section 1.1 to the WM-NP2 mobile workstation. • The maximum storage load of each drawer is 5kg, otherwise stability may be compromised. • Ensure each drawer is locked when moving the workstation. • Take care not to trap fingers when opening/closing or repositioning the drawer unit. • When installing a keyboard on the keyboard tray, ensure the cable passes around the outside of the sliding runner, otherwise the cable may be pinched when operating the sliding tray. • Check all equipment connectors on a regular basis to ensure there are no loose connections.
8
MOBILE WORKSTATION WM-NP2
Standard Sets
Chapter 1 Standard Sets The Olympus WM-NP2 mobile workstation has been designed to accommodate the Olympus video system center and associated ancillary devices as a practical mobile workstation to facilitate GI endoscopy, endoscopic ultrasound, respiratory and surgical endoscopic procedures. Please ensure familiarity with all important information prior to workstation operation. WARNING
CAUTION
Explosion hazard - never install or use the WM-NP2 mobile workstation within the zone of risk of flammable gases. The workstation is fitted with antistatic castors, none of the medical electrical equipment with which the workstation is intended to be used is ‘AP’ or ‘APG’ rated, therefore never install or use the WM-NP2 mobile workstation within the zone of risk of flammable gases. Section 5.2 Manoeuvring gives detailed instructions and precautions which must be read and understood before attempting to move the mobile workstation.
Please refer to the following table to check that all items in the set ordered are present. Please contact Olympus if any parts are damaged or missing.
MOBILE WORKSTATION WM-NP2
9
Standard Sets
1.1 Standard sets of components
B
(supplied boxed)
(supplied boxed)
D A (supplied fitted)
C
(supplied fitted)
F
(supplied fitted)
E
(supplied boxed)
A B C D E F G H I J
10
WM-NP2 (incl. Transformer) MAJ-1641 LCD Monitor Arm MAJ-1640 Keyboard Tray MAJ-1638 Scope Pole MAJ-1653 CO2 Cylinder Holder Double MAJ-1643 Drawer Unit Blister pack Toolkit Blister pack Drawer key IEC lead set MAJ-150 adaptor cable (USA/Japan only)
*1 Japan USA Europe UK ROW 1 1 1 1
*2
USA
1 1 1 1 1
Standard set 3 4 Japan USA Japan Europe USA UK ROW 1 1 1 1
5 Japan USA Europe UK ROW 1 1 1
Europe UK 1 1 1
1
1
1
1
1
1
1 1
1 1
1 1
1 1
1 1
MOBILE WORKSTATION WM-NP2
6
1 1 1 1
Standard Sets
1.2 Optional accessories The following products may be used as optional accessories. Code
Description
MAJ-150
Adaptor cable
MAJ-1662
Nurses control panel arm - short
MAJ-1663
EUS Arm mount kit
MAJ-1664
Nurses control panel arm - long
MAJ-1635
EUS connector holder
MAJ-1638
Scope pole kit
MAJ-1639
Single CO2 holder (up to Ø140mm)
MAJ-1640
Keyboard tray
MAJ-1641
LCD monitor arm
MAJ-1642
IV Pole
MAJ-1643
Drawer unit
MAJ-1650
Single CO2 holder (Ø140mm to Ø205)
MAJ-1653
Double CO2 holder
MAJ-1654
Equipotential terminal strip
MAJ-1655
OEV261H Monitor PSU holder
MAJ-1656
Footswitch holder
MAJ-1657
Keyboard arm - side mounted
MAJ-1658
Universal stowage container
MAJ-1659
Equipment support rail
MAJ-1660
Cable retaining hook
MAJ-1659
Equipment support rail
MOBILE WORKSTATION WM-NP2
11
Symbols
Chapter 2 Symbols The following symbols are used on the separation transformer.
2.1 Top tray Consult instructions for use
Refer to instructions
2.2 Separation transformer Warning to observe rules regarding the use of multiple socket-outlets
Potential equalisation terminal
Input
Output
Not exceeding
Maximum ambient temperature
Mains power connected
Breaker
Heavy unit
Remote switch
Non-inherently short circuit proof separation transformer
12
Ta=40°C
Class A/B
Insulation temperature rating of transformer according to IEC 60085:2004
Unit complies with relevant European directives (MAJ-1648 only)
Disposal conditions - see End of Life (Chapter 11) (MAJ-1648 only)
Consult instructions for use
Indicates the unit has been independently assessed for compliance with UL and CSA standards as a Recognized Component of the workstation (MAJ-1647 only).
MOBILE WORKSTATION WM-NP2
Symbols
View of the separation transformer rear panel
NOTE
MAJ-1648 (220-240V) is shown. Other versions are similar.
MOBILE WORKSTATION WM-NP2
13
Preparation for Use
Chapter 3 Preparation for Use 3.1 Castors 1.
The workstation has two diagonally located braked castors. Ensure both brakes are applied prior to use and released before moving the workstation. NOTE
Brakes should always be applied when the workstation is not being manoeuvred.
Brake castor
Brake castor
Figure 3.1 2.
To lock and unlock the castor brake, refer to Figure 3.2.
Figure 3.2
14
MOBILE WORKSTATION WM-NP2
Preparation for Use
3.2 Shelf size Refer to Figure 3.3 for the size of equipment that can be mounted on the product's shelves.
Top shelf
Dimensions in mm
470 325 409
439
Middle shelf
512
473
Base shelf
470
479
617
480
Figure 3.3 MOBILE WORKSTATION WM-NP2
15
Preparation for Use
3.3 Fixing positions of intermediate shelves The product columns contain holes for the intermediate shelves. You can reposition the shelves to suit the size of the equipment being mounted.
On delivery The product is delivered with the shelves fitted in the positions shown in Figure 3.4.
275mm
35mm
215mm
35mm
Shelf fixing holes 25mm apart
292mm
Drawer unit not shown
Figure 3.4 16
MOBILE WORKSTATION WM-NP2
Preparation for Use
Repositioning of shelves CAUTION
Before commencing shelf adjustment, ensure equipment is disconnected and removed from the shelf.
Use the following steps to reposition the intermediate shelves.
1.
Remove any equipment that is on the shelf.
2.
Position the workstation on a flat surface and apply the brakes on two diagonally opposite castors.
3.
Remove the two cable ducts that are fitted on the rear of the intermediate shelf and the cable tie that is fitted at the same height as the shelf. You can remove the cable tie by inserting a flat bladed screwdriver behind the stud in the middle of the cable tie and pulling it towards you, as shown in Figure 3.5. Cable duct
Stud
Cable tie
Figure 3.5 4.
Check that nothing has been placed on the shelf and, referring to Figure 3.6, loosen the two M8 screws that are securing the shelf about half a turn using the supplied 6mm Allen key – do not remove.
MOBILE WORKSTATION WM-NP2
17
Preparation for Use
M8 screw
Spring plunger
Figure 3.6 5.
Take a firm hold of the shelf and, referring to Figure 3.6, unlock the two spring plungers fitted on each side. Pull the spring plunger head towards you, twist 90° and release. The spring plunger remains extended and is unlocked. NOTE
6.
Move the shelf to the required height. Pull each spring plunger head, twist 90° and release. The spring plunger head will spring back to its original position when aligned with the column holes. CAUTION
7.
18
The spring plungers cannot be removed from the shelves.
After moving the shelf, check that the four spring plunger heads are not protruding. If any spring plunger head is protruding, the shelf may tilt and get damaged due to insufficient engagement with the column holes.
Check that the shelf is not tilting and firmly tighten the two M8 screws with the Allen key.
MOBILE WORKSTATION WM-NP2
Preparation for Use
8.
Fit the cable tie and cable ducts that were removed in step 3. Position the cable duct against the shelf, align the cable tie (with stud removed) with the shelf hole, and press the stud all the way through the cable tie into the hole to secure the cable duct.
Figure 3.7
3.4 Fixing position of the drawer unit If you purchase the set containing a drawer unit, it is delivered with the drawer unit already fitted. Use the following steps to reposition the drawer unit.
1.
Position the workstation on a flat surface and apply both castor brakes.
2.
Empty the contents of the drawer.
3.
From the rear of the workstation, remove the four M8 screws, M8 crinkle washers and M8 plain washers securing the drawer (Figure 3.8).
MOBILE WORKSTATION WM-NP2
19
Preparation for Use
Drawer unit
M8 screw M8 crinkle washer M8 plain washer
Figure 3.8 4.
Unlock the drawer. From the front of the workstation, take a firm hold of the drawer unit (see Figure 3.9) and pull forwards, one side of the drawer unit will disengage. CAUTION
Take care not to trap fingers when removing the drawer unit.
Figure 3.9
20
MOBILE WORKSTATION WM-NP2
Preparation for Use
5.
Referring to Figure 3.10, locate the spring plunger on the rear of the drawer unit, pull down and twist 90° to lock “open”. CAUTION
NOTE
Please take care not to slide the drawer unit forwards with the spring plunger unlocked, as the shelf will fall.
The spring plungers cannot be removed from the shelves.
Figure 3.10 6.
Referring to Figure 3.11, from the front of the workstation, take a firm hold of the drawer unit and prepare to support its weight (6kg), pull the drawer unit forwards where it will completely disengage and be free of the shelf.
Figure 3.11 MOBILE WORKSTATION WM-NP2
21