Quick Reference Guide
9 Pages
Preview
Page 1
ESG-150 (Electrosurgical Generator) Quick Reference Guide
CAUTION: This guide is only a summary of the instructions for use of the ESG-150 Electrosurgery Generator. Be sure to reference the instruction manual that was included with your product purchase.
ESG-150 Electrosurgical Generator
Bipolar Graphical User Interface
Neutral
On/Off
Monopolar
ESG-150 User Flexibility
Basic Level
Advanced Level
CONTINUED ON NEXT PAGE
ESG-150 (Electrosurgical Generator) Quick Reference Guide Monopolar Home Screen
Basic Level
Advanced Level Active Socket
Contact Quality Monitor
Active Socket
Contact Quality Monitor
Active Mode
Power Adjustment Effect Level Adjustment
Change Output Socket
System Setup
Change Output Socket
System Setup
Bipolar Home Screen
Basic Level
Advanced Level Active Socket
Active Socket
Active Mode
Power Adjustment Effect Level Adjustment
Change Output Socket
System Setup
Change Output Socket
System Setup
CONTINUED ON NEXT PAGE
ESG-150 (Electrosurgical Generator) Quick Reference Guide ESG-150 System Operating Steps
1 Turn on the ESG-150.
2a Select Basic or Advanced Level Functionality.
1) Select the “Menu” button.
2) Select “User Level”.
4) Select the “OK” button to confirm.
5) Functionality has changed on the home screen.
3) Select “Basic” or “Advanced” Functionality.
CONTINUED ON NEXT PAGE
ESG-150 (Electrosurgical Generator) Quick Reference Guide ESG-150 System Operating Steps
2b Select Mode: Monopolar; Bipolar.
1) Select the mode by tapping on the desired mode located on the bottom of the touch screen.
2c (Optional) Select User Defined Settings.
1) Select the “Menu”.
2) Choose “Select”.
3) Select desired setting.
2) In the menu screen, tap to select the desired mode.
3) Once selected, the home screen will display the change.
3 Mode menu selection.
1) S elect the mode by tapping on color bar on the touch screen.
CONTINUED ON NEXT PAGE
ESG-150 (Electrosurgical Generator) Quick Reference Guide ESG-150 System Operating Steps
4 Changing the Power Level. Select
to adjust power level.
5 Changing the Effect. Select to adjust effect. (Effect is only available in the Advanced Level Function.)
6
1) Apply the neutral electrode to the patient.
2) Connect to the ESG-150.
CONTINUED ON NEXT PAGE
ESG-150 (Electrosurgical Generator) Quick Reference Guide ESG-150 System Operating Steps
7 Monitor Neutral Electrodes: Contact Quality Monitor. If the red neutral electrode lights up and alarm tone sounds: • Is the neutral electrode correctly connected to the device? • Does the neutral electrode have sufficient contact to the skin?
Correct = Green
Error = Red
8 Connect device in appropriate receptacle. Bipolar Devices
Neutral Electrodes
Monopolar Devices
CONTINUED ON NEXT PAGE
ESG-150 (Electrosurgical Generator) Quick Reference Guide Creating User Defined Settings
STEP 1 On the home screen Select Preferred Modes Select Power Level Select Effect Level Select Menu
STEP 2
Settings/Setup Screen • Select “Save”
STEP 3
Save Setting Screen • Enter the setting name • Select “OK”
STEP 4
The setting will appear on the upper right of the touchscreen.
CONTINUED ON NEXT PAGE
ESG-150 (Electrosurgical Generator) Quick Reference Guide ESG-150 Operation Settings Basic Procedures Procedure
Instrument or Technique
Mode
Power
Effect
Snare Polypectomy: Sessile polyp
Close the snare slowly and gently
PulseCut Slow
120
2
Snare Polypectomy: Pedunculated polyp
Close the snare slowly and gently
ForcedCoag
20
4
Endoscopic Mucosal Resection (EMR)
Close the snare slowly and gently
PulseCut Slow
120
2
Hemostasis
Coagrasper
SoftCoag
50
3
Sphincterotomy
Sphincterotome
PulseCut Fast
120
2
Procedure
Instrument or Technique
Mode
Power
Effect
Less extended lesion, bleeding
CD-6C-1
SoftCoag
40
3
Polypoid Lesion (intraluminal mass)
SD-7C-1
PulseCut Slow
50
2
Web-like Stenosis
KD-31C-1
PulseCut Fast ForcedCoag
40 40
2 3
Tissue Ablation
Activate MAPC probe within a 2-5 mm distance from tissue
ForcedArgon
35
N/A
Pulmonary Procedures
The above settings are examples based upon previous general clinical usage of the device. Please note that clinical staff are solely responsible for selection of an appropriate mode and power level, dependent upon the instrument being used and the condition of the tissue being treated. Due to the heterogeneity of instruments and clinical conditions, Olympus cannot accept liability for clinical risks arising from the use of these example settings.
CONTINUED ON NEXT PAGE
ESG-150 (Electrosurgical Generator) Quick Reference Guide ESG-150 Monopolar Modes
The wired foot switch is an accessory and must only be used in conjunction with the compatible ESG-150 to activate the generator. The foot switch consists of: • Yellow cut pedal to activate the selected cutting mode. • Blue coagulation pedal to activate the selected coagulation mode. • Foot switch plug (7-pin connector) to connect the foot switch with the ESG-150. The length of the connecting cable is 4m.
YELLOW PEDAL
BLUE PEDAL
Press to activate the selected CUTTING mode
Press to activate the selected COAGULATION mode
CUT MODES
COAGULATION MODES
PureCut
PulseCut
ForcedCoag
SoftCoag
Continuous cutting
Controlled cutting with an intermittent duty cycle of different duration
Superficial pinpoint coagulation
Deeper coagulation
High Power Cut Support (HPCS) – immediate cutting Fast Spark Monitor (FSM) – optimized cutting power
Versatile coagulation technology