OLYMPUS
Electrosurgical Generator
ESG-300 and APU-300 Argon Plasma Unit Electrosurgical Generator Quick Reference Guide
Quick Reference Guide
10 Pages
Preview
Page 1
ESG-300 (Electrosurgical Generator)
Quick Reference Guide
CAUTION: This guide is only a summary of the instructions for use of the ESG-300 Electrosurgery Generator. Be sure to reference the instruction manual that was included with your product purchase.
ESG-300 and APU-300 Display Interface
Electrosurgical Generator (ESG-300)
Bipolar
Graphical User Interface
Monopolar On/Off Neutral
Argon Plasma Unit (APU-300) Purge button
On/Off Argon
Monopolar Home Screen
Active socket
Contact Quality Monitor
Procedure name Setting name
Active Mode
Power adjustment
Effect level adjustment Open user defined settings Change output socket
System setup
CONTINUED ON NEXT PAGE
ESG-300 (Electrosurgical Generator) Quick Reference Guide Bipolar Home Screen
Output Socket Name Active Mode Name
Power adjustment
Effect level adjustment Open user defined settings Change Output Sockets
System setup
Argon Home Screen
Foot Switch Indicator Contact Quality Monitor Output Socket Name
Communication Indicator Setting Name
Argon Gas Flow Rate
Power Level
Type of Connected Argon Probe
Effect Level
Argon Supply Indicator
CONTINUED ON NEXT PAGE
ESG-300 (Electrosurgical Generator) Quick Reference Guide ESG-300 System Operating Steps
1 Turn on the ESG-300 & APU-300.
2a Select Mode: Monopolar; Bipolar; Argon Monopolar
Bipolar
Argon
Select mode by tapping on the desired mode located on the bottom of the touch screen
2b (Optional) Select User Defined Setting or Procedure.
1.) Select “Open”
2.) Select setting or procedure
3.) Verify setting is correct
CONTINUED ON NEXT PAGE
ESG-300 (Electrosurgical Generator) Quick Reference Guide ESG-300 System Operating Steps
3 Mode Menu
1.) Select the mode by tapping on color bar on the touch screen
4 Changing the Power Level
Select to adjust power level
2.) In the menu screen, tap to select the desired mode
3.) Once selected, the home screen will display the change
5 Changing the Effect
Select to adjust effect
6 Argon Plasma Applications
2.) Confirm supply connection indicator displays connection and tank status 1.) Open the argon tank
CONTINUED ON NEXT PAGE
ESG-300 (Electrosurgical Generator) Quick Reference Guide ESG-300 System Operating Steps
7
1.) Apply the neutral electrode to the patient
2.) Connect to the ESG-300
8 Monitor Neutral Electrodes: Contact Quality Monitor If the red neutral electrode lights up & alarm tone sounds: • Is the neutral electrode correctly connected to the device? • Does the neutral electrode have sufficient contact to the skin?
Correct = Green
Error = Red
CONTINUED ON NEXT PAGE
ESG-300 (Electrosurgical Generator) Quick Reference Guide ESG-300 System Operating Steps
9 Connect device in appropriate receptacle
Bipolar Devices Monopolar Devices Neutral Electrodes
Argon Plasma Probes
Creating User Defined Settings
STEP 1 On the home screen • Select Preferred Modes • Select Power Level • Select Effect Level • Select Menu
CONTINUED ON NEXT PAGE
ESG-300 (Electrosurgical Generator) Quick Reference Guide Creating User Defined Settings
OK
STEP 2
Edit Settings/Procedures screen • Select “New” Setting
STEP 3
Save Setting Screen • Enter the setting name • Select “OK”
STEP 4
Edit Settings/Procedures screen • Select “Return” to return to the Home Screen
Creating User Defined Procedures
A user-defined procedure is a compilation of user-defined settings.
STEP 1
STEP 2
On the Home screen
Edit Settings/Procedures screen
• Select Menu
• Select “New” Procedure
STEP 3
New Procedures screen • Use up/down to highlight the setting • Use arrow to move the setting on the left into the procedure column on the right
OK
STEP 5
Save Procedure Screen • Enter the procedure name • Select “OK”
STEP 6
Edit Settings/Procedures screen • Select “Return” to return to the Home Screen
CONTINUED ON NEXT PAGE
ESG-300 (Electrosurgical Generator) Quick Reference Guide ESG-300 Operation Settings Basic Procedures Procedure
Instrument or Technique
Mode
Power
Effect
Snare Polypectomy: Sessile polyp
Close the snare slowly and gently
PulseCut Slow
120
2
Snare Polypectomy: Pedunculated polyp
Close the snare slowly and gently
ForcedCoag
20
4
Endoscopic Mucosal Resection (EMR)
Close the snare slowly and gently
PulseCut Slow
120
2
Hemostasis
Coagrasper
SoftCoag
50
3
Sphincterotomy
Sphincterotome
PulseCut Fast
120
2
Procedure
Instrument or Technique
Mode
Power
Effect
Marking
Knife (apply gentle pressure)
ForcedCoag
20
2
SoftCoag
50
3
Pre-Cut/Incision
Knife
PulseCut Fast
120
2
BlendCut
30
2
Submucosal Dissection
Knife
PowerCoag
30
2
Hemostasis
Closed Knife
PowerCoag
30
2
Procedure
Instrument or Technique
Mode
Power
Effect
Incision
Knife
PulseCut Fast
120
2
Dissection/Tunneling
Knife
SprayCoag
40
2
Myotomy
Knife
PulseCut Fast
120
2
Hemostasis
Coagrasper
SoftCoag
50
3
Endoscopic Submucosal Dissection (ESD)
Peroral Endoscopic Myotomy (POEM)
The above settings are examples based upon previous general clinical usage of the device. Please note that clinical staff are solely responsible for selection of an appropriate mode and power level, dependent upon the instrument being used and the condition of the tissue being treated. Due to the heterogeneity of instruments and clinical conditions, Olympus cannot accept liability for clinical risks arising from the use of these example settings.
CONTINUED ON NEXT PAGE
ESG-300 (Electrosurgical Generator) Quick Reference Guide ESG-300 Operation Settings Argon Plasma Coagulation (APC) Procedure
Instrument or Technique
Mode
Power
Effect
Diffuse bleeding in the right colon/duodenum
Activate MAPC probe within a 2-5 mm distance from tissue
SmartArgon
120
4
Diffuse bleeding in the remaining colon and rectum
Activate MAPC probe within a 2-5 mm distance from tissue
PulsedArgon Fast
30
N/A
GAVE/radioproctitis
Activate MAPC probe within a 2-5 mm distance from tissue
PulsedArgon Fast
30
N/A
Angiodysplasia
Activate MAPC probe within a 2-5 mm distance from tissue
PulsedArgon Fast
30
N/A
Tumor Reduction
Activate MAPC probe within a 2-5 mm distance from tissue
ForcedArgon
50
N/A
Barrett’s Esophagus
Activate MAPC probe within a 2-5 mm distance from tissue
PulsedArgon Fast
40
N/A
Procedure
Instrument or Technique
Mode
Power
Effect
Less extended lesion, bleeding
CD-6C-1
SoftCoag
40
3
Polypoid Lesion (intraluminal mass)
SD-7C-1
PulseCut Slow
50
2
Web-like Stenosis
KD-31C-1
PulseCut Fast ForcedCoag
40 40
2 3
Tissue Ablation
Activate MAPC probe within a 2-5 mm distance from tissue
ForcedArgon
35
N/A
Pulmonary Procedures
The above settings are examples based upon previous general clinical usage of the device. Please note that clinical staff are solely responsible for selection of an appropriate mode and power level, dependent upon the instrument being used and the condition of the tissue being treated. Due to the heterogeneity of instruments and clinical conditions, Olympus cannot accept liability for clinical risks arising from the use of these example settings.
CONTINUED ON NEXT PAGE
ESG-300 (Electrosurgical Generator) Quick Reference Guide ESG-300 Monopolar Modes
The wired foot switch is an accessory and must only be used in conjunction with the compatible ESG-300 to activate the generator. The foot switch consists of: • Yellow cut pedal to activate the selected cutting mode. • Blue coagulation pedal to activate the selected coagulation mode. • Toggle button to switch between user-defined settings within the procedure. • Foot switch plug (7-pin connector) to connect the foot switch with the ESG-300. The length of the connecting cable is 4m.
YELLOW PEDAL
BLUE PEDAL
Press to activate the selected CUTTING mode
Press to activate the selected COAGULATION mode
CUT MODES
PureCut
PulseCut
Continuous cutting
Controlled cutting with an intermittent duty cycle of different duration
COAGULATION MODES
BlendCut
Cutting of varying tissue structures with increased hemostasis capacities
PowerCoag
Superficial coagulation with increased dissection capability
High Power Cut Support (HPCS) – immediate cutting Fast Spark Monitor (FSM) – optimized cutting power
ForcedCoag
SprayCoag
SoftCoag
Superficial pinpoint coagulation
High peak voltage for superficial coagulation without contact between the HF instrument and the tissue
Deeper coagulation
Versatile coagulation technology
Olympus is a registered trademark of Olympus Corporation, Olympus America Inc., and/or their affiliates.
3500 Corporate Parkway, PO Box 610, Center Valley, PA 18034
I Medical devices listed may not be available for sale in all countries.
For more information, contact your Olympus sales representative, or call 800-848-9024. www.medical.olympusamerica.com ©2018 Olympus America Inc. All rights reserved. Printed in USA OAIET0818QRG27107