Reprocessing Flexible Endoscope Accessories Manual Cleaning and High Level Disinfection for the MAJ-891 WARNING This workflow is a SUMMARY of the reprocessing steps
necessary to properly reprocess your MAJ-891. Following disconnection and disassembly of the MAJ-891 utilize the IFU for how to reprocess the MAJ-891.
WARNING
Be sure to disassemble the forceps/irrigation plug before it is cleaned, disinfected or sterilized. Otherwise, the forceps/irrigation plug may not be properly cleaned/disinfected/sterilized.
Assembled MAJ-891 (Forceps/Irrigation Plug) Precleaning
1
Disconnect from the endoscope
Completely immerse in detergent solution
2
Disassembled MAJ-891 (Forceps/Irrigation Plug) Manual Cleaning
3
Completely Disassemble
4
Completely immerse in detergent solution
5
Note: to pass the brush through the biopsy valve, press it between your fingers.
Thoroughly wipe all external surfaces and brush all internal surfaces of each part
Thoroughly rinse after manual cleaning.
Manual Disinfection
6
Completely immerse all parts in the disinfectant solution
7
Fill a 30 ml syringe with disinfectant solution and flush the forceps/irrigation plugby covering the stopcock’s holes with your fingers and fill the channel with disinfectant solution
8
Open the slit of the biopsy valve in disinfectant solution
9
Cover the basin and soak
30 ml x3
Rinsing
10
Drying
Immerse all parts in sterile water. Flush the forceps/irrigation plug utilizing a 30ml syringe.
30 ml x3
11
Wipe and dry the external surfaces with sterile, lint-free cloth
12
Use sterile cotton swabs to dry the internal surfaces of each part
Note: when non-sterile water is used after disinfection, immerse all parts in 70% ethyl or isopropyl alcohol and gently agitate them.
Olympus is a registered trademark of Olympus Corporation, Olympus America Inc., and/or their affiliates.
|
Medical devices listed may not be available for sale in all countries.
For more information, contact your Olympus sales representative, or call 800-848-9024. www.medical.olympusamerica.com 3500 Corporate Parkway, PO Box 610, Center Valley, PA 18034
©2021 Olympus America Inc. All rights reserved. Printed in the USA. OAIENT0621WC41474
Reprocessing Flexible Endoscope Accessories Manual Cleaning Prior to Sterilization for the MAJ-891
WARNING This workflow is a SUMMARY of the reprocessing steps
necessary to properly reprocess your MAJ-891. Following disconnection and disassembly of the MAJ-891 utilize the IFU for how to reprocess the MAJ-891.
WARNING
Be sure to disassemble the forceps/irrigation plug before it is cleaned, disinfected or sterilized. Otherwise, the forceps/irrigation plug may not be properly cleaned/disinfected/sterilized.
Assembled MAJ-891 (Forceps/Irrigation Plug) Precleaning
1
Disconnect from the endoscope
2
Completely immerse in detergent solution
Disassembled MAJ-891 (Forceps/Irrigation Plug)
Manual Cleaning
3
Completely Disassemble
Drying
Manual Cleaning
5
Thoroughly wipe all external surfaces and brush all internal surfaces of each part.
Completely immerse in detergent solution
4
Note: to pass the brush through the biopsy valve, press it between your fingers. Thoroughly rinse after manual cleaning.
6
Use a lint-free cloth to thoroughly wipe and dry the external surfaces of each part.
7
Use sterile cotton swabs to dry the internal surfaces of each part.
Note: Prior to sterilization, the instrument must be thoroughly cleaned and dried. All parts must be sealed in packages separately in packaging appropriate for the sterilization method chosen.
Olympus is a registered trademark of Olympus Corporation, Olympus America Inc., and/or their affiliates.
|
Medical devices listed may not be available for sale in all countries.
For more information, contact your Olympus sales representative, or call 800-848-9024. www.medical.olympusamerica.com 3500 Corporate Parkway, PO Box 610, Center Valley, PA 18034
©2021 Olympus America Inc. All rights reserved. Printed in the USA. OAIENT0621WC41474