INSTRUCTIONS
COLPOSCOPE OCS-500 DIGITAL CAMERA ADAPTER
OCS5-DCAD2
Contents
Contents Important Information – Please Read Before Use ...
1
Intended use ...
1
Instruction manual ...
1
User qualifications ...
1
Instrument compatibility ...
2
Repair and modification ...
2
Signal words ...
2
Warnings and cautions...
3
Chapter 1
Checking the Package Contents...
4
Chapter 2
Instrument Nomenclature ...
5
Chapter 3
Installation and Connection ...
6
3.1
Connecting the camera mount...
7
3.2
Connecting the digital camera adapter to the zoom microscope body (OCS5-ZB) ...
7
3.3
Connecting the digital camera to the adapter ...
9
3.4
Connecting the AC adapter...
9
3.5
Adjust the balance of the balance arm...
11
Preparation and Inspection ...
12
Inspecting the connections ...
12
Operation ...
13
5.1
Setting up the digital camera ...
14
5.2
Photography...
15
Care, Storage and Disposal...
17
6.1
Care ...
17
6.2
Storage ...
18
6.3
Disposal ...
18
Troubleshooting ...
19
7.1
Troubleshooting guide ...
19
7.2
Returning the digital camera adapter for repair ...
20
Chapter 4 4.1
Chapter 5
Chapter 6
Chapter 7
DIGITAL CAMERA ADAPTER OCS5-DCAD2
i
Contents
ii
Appendix...
21
System chart ...
21
Operating environment...
22
Specifications ...
22
DIGITAL CAMERA ADAPTER OCS5-DCAD2
Important Information – Please Read Before Use
Important Information – Please Read Before Use Intended use This instrument has been designed to be used with the Olympus colposcope OCS-500. Do not use the instrument for any purposes other than its intended use.
Instruction manual This instruction manual contains essential information on using this instrument safely and effectively. Before use, thoroughly review this manual and the manuals of all equipment which will be used during the procedure and use the instruments as instructed. Keep this and all related instruction manuals in a safe, accessible location. If you have any questions of comments about any information in this manual, please contact Olympus.
Terms used in this manual Colposcope (OCS-500) The colposcope to which this digital camera adapter can be connected. Digital camera (E-1, E-300) The digital camera to which this digital camera adapter can be connected.
User qualifications The operator of this instrument must be a physician or medical personnel under the supervision of a physician and must have received sufficient training in clinical colposcopic technique. This manual, therefore, does not explain or discuss clinical colposcopic procedure.
DIGITAL CAMERA ADAPTER OCS5-DCAD2
1
Important Information – Please Read Before Use
Instrument compatibility Refer to the “System chart” in the Appendix to confirm that this instrument is compatible with the ancillary equipment being used. Using incompatible equipment can result in patient injury and/or equipment damage.
Repair and modification This instrument does not contain any user-serviceable parts. Do not disassemble, modify or attempt to repair it. Otherwise, patient or operator injury and/or equipment damage can result. Some problems that appear to be malfunctions may be correctable by referring to Chapter 7, “Troubleshooting”. If the problem cannot be resolved by referring to Chapter 7, contact Olympus.
Signal words The following signal words are used throughout this manual:
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.
Indicates additional helpful information.
2
DIGITAL CAMERA ADAPTER OCS5-DCAD2
Important Information – Please Read Before Use
Warnings and cautions Follow the warnings and cautions given below when handling this instrument. This information is to be supplemented by the warnings and cautions given in each chapter.
Handle the equipment with care and be especially careful not to strike the adapter's body. Otherwise, lens components may be damaged, making photography impossible.
DIGITAL CAMERA ADAPTER OCS5-DCAD2
3
Chapter 1 Checking the Package Contents
Chapter 1 Checking the Package Contents Match all items in the package with the components shown below. Inspect each item for damage. If the instrument is damaged, a component is missing or you have any questions, do not use the instrument; immediately contact Olympus.
Main unit
Camera mount
4
DIGITAL CAMERA ADAPTER OCS5-DCAD2
Instruction manual
Chapter 2 Instrument Nomenclature
Chapter 2 Instrument Nomenclature
Mount Attach the digital camera adapter to the microscope body here.
Camera mount
Camera clamping knob The knob is used to clamp the digital camera. It is only used when a camera is connected.
DIGITAL CAMERA ADAPTER OCS5-DCAD2
5
Chapter 3 Installation and Connection
Chapter 3 Installation and Connection Follow the precautions given below. Otherwise, the patient or the operator can be subjected to an electric shock. For information on attaching the digital camera, refer to its instruction manual. − Do not splash or spill water, or other liquid onto the equipment. − Do not prepare, inspect, or operate the equipment with wet hands.
If the digital camera is not securely connected to its adapter, both the camera and adapter may fall off, leading to patient or operator injury and/or damage to the equipment. Prepare the instrument and compatible equipment (shown in the “System chart” in the Appendix) before each use. Refer to the respective instruction manuals for each equipment. Install and connect all equipment as described in this chapter.
6
DIGITAL CAMERA ADAPTER OCS5-DCAD2
Chapter 3 Installation and Connection
3.1
Connecting the camera mount Align the red mark on the digital camera with that on the camera mount. Place the camera in close contact with the camera mount and rotate the camera mount clockwise until it clicks to secure it (see Figure 3.1). Red mark
Digital camera
Camera mount
Figure 3.1
3.2
Connecting the digital camera adapter to the zoom microscope body (OCS5-ZB) The following explanations show how to connect the digital camera adapter to the left side of the zoom microscope body. If you are using the TV camera adapter, the eyepiece shaped TV mount and the video adapter, the digital camera adapter may be connected to the right side of the zoom microscope body (for details, refer to the TV camera adapter instruction manual). In this case, connect the digital camera adapter to the right side of the zoom microscope body as described above.
DIGITAL CAMERA ADAPTER OCS5-DCAD2
7
Chapter 3 Installation and Connection
1. Loosen the camera adapter clamping knob on the left side of the zoom microscope body by turning it counterclockwise, then slide up the cap on the left side to remove it (see Figure 3.2). Keep the detached cap in a secure place.
Zoom microscope body
Cap Camera adapter clamping knob
Camera adapter clamping knob
Left side Right side
Figure 3.2
If you loosen the camera adapter clamping knob too much, it may fall off.
2. Align the digital camera adapter mount with the camera adapter mount on the zoom microscope body. Slide the mount down until it stops. Then turn the clamping knob clockwise to secure the camera adapter mount to the zoom microscope body (see Figure 3.3). Camera adapter clamping knob Mount
Digital camera adapter
Figure 3.3
8
DIGITAL CAMERA ADAPTER OCS5-DCAD2
Camera adapter clamping knob
Chapter 3 Installation and Connection
3.3
Connecting the digital camera to the adapter Insert the camera mount connected to the digital camera into the digital camera adapter until it stops (see Figure 3.4). Then turn the camera clamping knob clockwise to secure the connection. Make sure that the digital camera is level.
Camera clamping knob
Figure 3.4
3.4
Connecting the AC adapter •
Do not use an AC adapter other than the provided one. Otherwise, a malfunction of the camera or power supply or electric shock to the operator may result.
•
Connect the AC adapter to an isolation transformer before connecting it to this product. Otherwise, an electric shock may result.
•
The power consumption of the peripheral equipment should not exceed the isolation transformer's capacity. It the power consumption exceeds the capacity of one isolation transformer, use a second isolation transformer.
When the AC adapter is used, be careful where to leave the main part of the AC adapter. Otherwise, the main part of the AC adapter may fall off and will be damaged when operating the arm.
DIGITAL CAMERA ADAPTER OCS5-DCAD2
9
Chapter 3 Installation and Connection
The battery can be used instead of the AC adapter. For details, refer to the digital camera instruction manual.
1. Connect the AC adapter's connection cord plug to the digital camera as described in its instruction manual (see Figure 3.5).
Connection cord plug
Balance arm
Figure 3.5
2. Connect the AC adapter's power plug to the isolation transformer. 3. Secure the AC adapter cord using the cable hooks under the colposcope’s horizontal arm and the balance arm (see Figure 3.6). Horizontal arm Balance arm
Cable hooks Bottom view
Figure 3.6
10
DIGITAL CAMERA ADAPTER OCS5-DCAD2
Chapter 3 Installation and Connection
3.5
Adjust the balance of the balance arm Adjust the balance of the balance arm as described in its instruction manual.
DIGITAL CAMERA ADAPTER OCS5-DCAD2
11
Chapter 4 Preparation and Inspection
Chapter 4 Preparation and Inspection Before each case, prepare and inspect this instrument as instructed below. Inspect other equipment to be used with this instrument as instructed in their respective instruction manuals. Should the slightest irregularity be suspected, do not use the instrument and see Chapter 7, “Troubleshooting”. If the irregularity is still suspected after consulting Chapter 7, contact Olympus. Damage or irregularity may compromise patient or operator safety and may result in more-severe equipment damage. Prepare this instrument and other compatible equipment to be used with this instrument (shown in the “System chart” in the Appendix) for the particular case. Refer to the respective instruction manuals for each item.
4.1
Inspecting the connections If the digital camera is not securely connected to the digital camera adapter, both the camera and adapter may fall off, leading to patient or operator injury and/or equipment damage. If you find a connection that is not secure, repeat the connecting procedures as described in Chapter 3, “Installation and Connection”.
1. Confirm that the digital camera is securely connected to the camera mount. 2. Confirm that the camera mount is securely connected to the digital camera adapter.
3. Confirm that the digital camera adapter is securely connected to the camera adapter mount of the zoom microscope body.
12
DIGITAL CAMERA ADAPTER OCS5-DCAD2
Chapter 5 Operation
Chapter 5 Operation The operator of this instrument must be a physician or medical personnel under the supervision of a physician and must have received sufficient training in clinical colposcopic technique. This manual, therefore, does not explain or discuss clinical colposcopic procedures. It only describes basic operation and precautions related to the operation of this instrument.
•
Wear personal protective equipment to guard against dangerous chemicals and potentially infectious material. During the procedure, wear appropriate personal protective equipment, such as eye wear, face mask, moisture-resistant clothing and chemical-resistant gloves that fit properly and are long enough so that your skin is not exposed.
•
Anytime you suspect an abnormality with an instrument function, stop the examination immediately. Using a colposcope that is not functioning properly may cause patient injury.
DIGITAL CAMERA ADAPTER OCS5-DCAD2
13
Chapter 5 Operation
5.1
Setting up the digital camera Set the menu of the digital camera by referring to its instruction manual. Set the optimum white balance for the shooting conditions by pointing the zoom microscope body at a white object like a sheet of white paper. Refer to the instruction manual of the digital camera for details.
•
If the white balance is not properly adjusted, the color reproduction of the photomicrograph image is poor.
Menu item
Recommended setting
Mode dial
A (Aperture priority shooting).
Exposure correction
0 The exposure may not match certain specimens. Correct the exposure in this case.
Metering mode
Digital ESP metering
Flash
(Flash off)
ISO speed
400
White balance
One touch white balance NOTE: With certain specimens, also try other modes because there may be more suitable modes.
14
Drive mode
To prevent camera shake due to shutter release, set the drive mode using the remote control or the remote cable release.
Remote control
Triggers the shutter button right away.
SLEEP
OFF
•
When using the remote cable for the digital camera (E-1), refer to the instruction manual of the digital camera (E-1).
•
When using the remote cable for the digital camera (E-300), connecting the optional power battery holder (HLD-3) to the digital camera is need. Refer to the instruction manual of the power battery holder (HLD-3) for details.
DIGITAL CAMERA ADAPTER OCS5-DCAD2
Chapter 5 Operation
5.2
Photography 1. Confirm that the reticle in the eyepiece (right side) is clear. If it is not clear, adjust the diopter as described in the instruction manual for the colposcope (OCS-500). When the field of view is in focus, the digital image will also be in focus (see Figure 5.1).
Figure 5.1
If you need more precise focusing, set the magnification at the maximum level by operating the zoom magnification change knob, adjust the focus, then reset the magnification to the desired value.
2. Maximize the light intensity as described in the instruction manual for the light source. Set the composition and check that the operational part is not activated. Then press the button on the remote control switch of the digital camera or the remote cable release of the digital camera (see Figure 5.2). Remote control switch
Remote cable release
Figure 5.2
DIGITAL CAMERA ADAPTER OCS5-DCAD2
15
Chapter 5 Operation
•
We recommend using the remote control switch or the remote cable release to take pictures. Operating the release button of the digital camera manually may move the camera, which will frequently cause the picture to be blurred, especially at high levels of magnification.
•
When using the remote cable for the digital camera (E-1), refer to the instruction manual of the digital camera (E-1).
•
When using the remote cable for the digital camera (E-300), the optional power battery holder (HLD-3) must be connected to the digital camera. Refer to the instruction manual of the power battery holder (HLD-3) for details.
•
The photographic area is the portion of the observation area inside the square as shown in Figure 5.3. The areas containing the hatched lines are not part of the photographic area. Observation area Photographic area 100%
Figure 5.3
16
DIGITAL CAMERA ADAPTER OCS5-DCAD2
Chapter 6 Care, Storage and Disposal
Chapter 6 Care, Storage and Disposal 6.1
Care •
After cleaning the digital camera adapter, dry it thoroughly before use. If it is used while wet, there is a risk of an electric shock.
•
Patient debris and reprocessing chemicals are hazardous. Wear personal protective equipment to guard against dangerous chemicals and potentially infectious material. During cleaning, wear appropriate personal protective equipment, eye wear, face mask, moisture-resistant clothing and chemical-resistant gloves that fit properly and are long enough so that your skin is not exposed. Always remove contaminated protective clothing before leaving the reprocessing area.
•
Do not attempt to autoclave or gas sterilize the digital camera adapter. Either of these methods may cause equipment damage.
•
Do not wipe the surface of the digital camera adapter with an abrasive material. Otherwise, the surface may be damaged.
If the adapter becomes soiled, perform the following cleaning procedures immediately after use. If cleaning is delayed, organic debris will solidify and it may be difficult to effectively clean the adapter. The adapter should also be cleaned routinely.
1. Should the adapter become soiled with blood or other potentially infectious patient material, first wipe off all gross debris using a soft, lint-free moistened cloth and clean the external surfaces with detergent solution. Then wipe the adapter with a clean, lint-free cloth moistened with 70% ethyl or isopropyl alcohol.
2. Remove any dirt from the colposcope’s objective or lenses using a piece of gauze or lint-free cloth moistened with 70% ethyl or isopropyl alcohol.
3. Make sure that no liquid remains on the adapter or the colposcope. After wiping them with a soft, lint-free moistened cloth, be sure to dry them thoroughly.