OLYMPUS
OLYMPUS DIEGO ELITE System
DIEGO ELITE Handpiece Cleaning Quick-Reference Guide Edition 48xxxx.X.000 Nov 2013
Quick-Reference Guide
2 Pages
Preview
Page 1
HANDPIECE CLEANING QUICK-REFERENCE GUIDE This is designed as a quick-reference guide only. This guide is not a substitute for reading and learning the entire operator’s manual and/or instructions for use (IFU) provided the DIEGO ELITE System. Olympus cannot encourage or recommend any actions that deviate from or are not covered by the IFU.
PRE-CLEANING Wipe the external surfaces of the handpiece, cord, and plug with enzymatic detergent, sterile water, or tap water immediately after use.
MANUAL CLEANING 1
Fully immerse the handpiece, cord, and plug in the cleaning solution. Allow them to soak for ten minutes. Note the manufacturer’s instructions regarding the temperature and dilution factor for the detergent.
2
While immersed, gently scrub the areas shown below with the appropriate brushes (handpiece cleaning kit item EG70338009)
The gear shaft
The interior of the blade retention mechanism
The three-pin receptacle pocket and plug
The handpiece plug
HANDPIECE CLEANING – QUICK-REFERENCE GUIDE
3
Using a nylon-bristle brush, scrub the shelf that contains the Olympus logo and part number.
4
Actuate the blade retention mechanism.
5
Flush the interior of the handpiece with cool distilled, deionized, or potable water. Actuate the retention mechanism while flushing. Ensure water flows through all drainage ports.
6
Use 70% isopropyl alcohol to wipe and flush the handpiece, power cord, and plug. Flush the exterior to remove any remaining residue.
7
Dry completely with a lint-free cloth. Visibly inspect the entire handpiece for any remaining contamination and repeat the cleaning procedure if needed.
AUTOMATED CLEANING
STERILIZATION
Automated cleaning can be performed in an automatic
The handpiece is designed for sterilization by steam autoclave.
washer-disinfector as an alternative to manual cleaning.
Handpieces can be sterilized wrapped or unwrapped.
Place the handpiece and cord, on the washer rack or place them in the approved tray (Item EGMDHPTRAY).
2
Steam sterilization Cycle
Gravity
prevacuum
Temperature
132° C
132° C
The washer should be set to the Vario TD cycle. After
time
15 minutes
4 minutes
automated cleaning, visually inspect the handpiece for
Drying Time
60 minutes
30 minutes
any remaining residue and repeat the cleaning procedure if needed.
Note: To satisfy international, regional, governmental, or healthcare facility requirements, the handpiece may be processed at temperatures up to 137° C without causing damage.
STORAGE To prevent damage, the handpiece must be completely vacuum dried before storage. Specifications, design, and accessories are subject to change without any notice or obligation on the part of the manufacturer.
Postbox 10 49 08, 20034 Hamburg, Germany Wendenstrasse 14–18, 20097 Hamburg, Germany Phone: +49 (0)40 237 730, Fax: +49 (0)40 230 761 www.olympus-europa.com
E048xxxx · X.000 · 11/13 · PR