INSTRUCTIONS
Endoscopy CAD system
Labels and Symbols
1
Important Information - Please Read Before Use
3
Chapter 1
Checking the Package Contents
11
Chapter 2
Nomenclature and Functions
13
Chapter 3
Installation and Connection
17
Chapter 4
Function setup
37
Chapter 5
Inspection
55
Chapter 6
Operation
61
Chapter 7
Function
73
Chapter 8
Reprocessing, Storage, and Disposal
87
Troubleshooting
95
OIP-1
Chapter 9
Appendix
105
Contents
Contents Labels and Symbols ... 1 Important Information - Please Read Before Use ... 3 Intended use/Intended purpose ... 3 Contraindications ... 3 Clinical benefit ... 3 Applicability of endoscopy and endoscopic treatment ... 3 Instruction manual ... 4 User qualifications ... 6 Instrument compatibility ... 6 Repair and modification ... 6 Signal words ... 7 Precautions ... 7
Chapter 1 Checking the Package Contents ... 11 1.1
Checking the package contents ... 11
Chapter 2 Nomenclature and Functions ... 13 2.1
Front panel ... 13
2.2
Rear panel ... 14
2.3
Side panels ... 15
2.4
Monitor ... 16
Chapter 3 Installation and Connection ... 17 3.1
Precautions for installation and connection ... 17
3.2
Installation workflow ... 18
3.3
Installation of the equipment ... 19 Installation of the equipment ... 20
3.4
Connection to the AC mains power supply ... 22 When the mobile workstation (WM-NP3) is used ... 23 When no mobile workstation is used ... 24
3.5
Connection of the video system center ... 24 Compatible video system center ... 24 CV-1500 ... 25
3.6
Connection of the monitor ... 26 Compatible monitors ... 26 OEV321UH ... 26 OEV262H ... 28 OEV261H ... 30
OIP-1 INSTRUCTION MANUAL
i
Contents
3.7
Connection of the video recorder ... 32 Compatible video recorders ... 32 IMH-200 ... 33 IMH-20 ... 34 IMH-10 ... 35
Chapter 4 Function setup ... 37 4.1
Setup list ... 37
4.2
Basic operation of the user settings ... 38 Loading the user settings using the the video system center ... 38 Loading the user settings directly from the Endoscopy CAD system ... 40 Editing user settings ... 41 Initializing the user settings ... 44
4.3
User settings (Sub-image display) ... 45
4.4
User settings (Marker/flag) ... 46
4.5
User settings (Auto ON/OFF) ... 48
4.6
Basic operation of the system setup ... 49 Displaying the system setup ... 49 Editing the system setup ... 50 Initializing the system setup ... 51
4.7
System setup ... 52
4.8
Setting custom switches ... 54
Chapter 5 Inspection ... 55 5.1
Precautions of workflow ... 55
5.2
Inspection workflow ... 56
5.3
Inspection of the power supply ... 57
5.4
Monitor display inspection ... 57
5.5
Inspection of the CAD ... 58
5.6
Inspection of power off ... 60
Chapter 6 Operation ... 61 6.1
Precautions for operation ... 61
6.2
Operation workflow ... 62
6.3
Turning ON the Endoscopy CAD system and ancillary equipment ... 63
6.4
Loading the user settings ... 63
6.5
Display methods of the CAD ... 64 Displaying the detection result in normal mode ... 64 Displaying the detection result in target mode ... 65
ii
OIP-1 INSTRUCTION MANUAL
Contents
6.6
Starting the endoscopic examination ... 66 Operation of the CAD in normal mode ... 66 Operation of the CAD in target mode ... 67 Switching the sub-image display in normal mode ... 67 CAD operation chart ... 68 Change settings of the sub image ... 70 Non-display of the CAD detection result ... 71
6.7
Termination of the operation ... 72
6.8
Transportation ... 72
Chapter 7 Function ... 73 7.1
Operation of the setting menu ... 73
7.2
Displaying the OIP-1 information ... 75 OIP-1 information ... 75
7.3
Confirming the message history ... 78 Message history ... 78
7.4
Export/import the setting data ... 82 Inserting the portable memory ... 82 Removing the portable memory ... 83 Export settings ... 83 Import settings ... 85
Chapter 8 Reprocessing, Storage, and Disposal ... 87 8.1
Reprocessing ... 87 Surface disinfectant cleaner ... 88 Signs of degradation from reprocessing ... 89 Preparing equipment for reprocessing ... 89 Reprocessing procedure ... 90
8.2
Storage ... 93
8.3
Disposal ... 93
Chapter 9 Troubleshooting ... 95 9.1
Troubleshooting ... 95
9.2
Troubleshooting guide ... 96
9.3
Returning the Endoscopy CAD system for repair ... 103
Appendix ... 105 Combination equipment ... 105 System chart ... 105
Specifications ... 107 Environment ... 107 Specifications ... 108
OIP-1 INSTRUCTION MANUAL
iii
Contents
Default values table ... 110 User setting ... 110 System setup ... 110
EMC information ... 111 Open Source Software license information of this instrument ... 117
iv
OIP-1 INSTRUCTION MANUAL
Labels and Symbols
Labels and Symbols Safety-related labels and symbols are attached on the locations shown below. If labels or symbols are missing or illegible, contact Olympus.
Rear panel Electric rating The product name, rated voltage, and frequency are shown. ENDOSCOPY CAD SYSTEM OIP-1 POWER INPUT 100 – 240 V
50/60 Hz 275 VA
Medical Device CE marking Potential equalization terminal
Date of Manufacture
Serial number
Manufacturer name
Authorized representative in the European Community Crossed-out wheeled bin’ symbol
OIP-1 INSTRUCTION MANUAL
Fuses information
Indicates to refer to the instruction manual for the cable to be connected. Refer to the instruction manual.
1
Labels and Symbols
Front panel
UDI label A label required by some countries’ regulations regarding identification of medical devices also known as Unique Device Identification (UDI).
Back cover of this instruction manual Symbol
Description Manufacturer
Authorized representative in the European Community
Translation
Importer (into European Union)
2
OIP-1 INSTRUCTION MANUAL
Important Information - Please Read Before Use
Important Information - Please Read Before Use
Intended use/Intended purpose The Endoscopy CAD system processes the electronic signals received from the endoscopy video system center and overlays additional information on the observation monitor. The device directs the user’s attention to areas of interest for further clinical assessment. The device is intended to assist physicians in the detection of mucosal abnormalities such as potential colorectal polyps during colonoscopy. The device is an adjunctive tool, and the user should not rely solely on the device for detection of abnormalities.
Contraindications None known.
Clinical benefit Depends on the connected endoscope.
Applicability of endoscopy and endoscopic treatment If there is an official standard on the applicability of endoscopy and endoscopic treatment that is defined by the healthcare facility’s administration or other official institutions such as academic societies on endoscopy, follow that standard. Before starting endoscopy and endoscopic treatment, thoroughly evaluate its properties, purposes, effects, and possible risks (their natures, extent, and probability). Perform endoscopy and endoscopic treatment only when its potential benefits are greater than its risks. Fully explain to the patient the potential benefits and risks of the endoscopy and endoscopic treatment as well as any examination/treatment methods that can be performed in its place, and perform the endoscopy and endoscopic treatment only after obtaining the consent of the patient. Even after starting the endoscopy and endoscopic treatment, continue to evaluate the potential benefits and risks, and immediately stop the endoscopy/treatment and take proper measures if the risks to the patient become greater than the potential benefits.
OIP-1 INSTRUCTION MANUAL
3
Important Information - Please Read Before Use
Instruction manual This instruction manual contains essential information on using the Endoscopy CAD system safely and effectively. Before use, thoroughly review this manual and the manuals of all equipment that will be used during the procedure and use the equipment as instructed. Keep this and all related instruction manuals in a safe, accessible location. If you have any questions or comments about any information in this manual, contact Olympus.
Terms used in this manual Video recorder: The video recorder is a device that records the video.
Video system center: The video system center is a device that processes the signal from the endoscope and sends the endoscopic image to the monitor.
Healthcare facility grade wall mains outlet: The healthcare facility wall mains outlet is a wall AC mains power outlet socket having the exclusive terminal for grounding.
Portable memory A digital medium for storage of images, etc.
Internal memory This is a space to store the setting data etc. It also includes an area to set additional program data.
Custom switch The custom switch is a general term for the buttons, keys, and switches to which functions can be assigned on video system center: “CUSTOM” key on the keyboard, remote switch on the endoscope, remote switch on the foot switch, and custom button on the touch panel of video system center.
CAD The CAD is Computer Aided Diagnosis/Detection generally. In this equipment, this function displays supplementary information on the monitor which indicates the existence and locations of potential colorectal polyps.
PIP (Picture in Picture) The PIP function displays the image of the endoscopic live image and the image of an external device on the monitor simultaneously, where one image is embedded into the other image.
4
OIP-1 INSTRUCTION MANUAL
Important Information - Please Read Before Use
POP (Picture out Picture) The POP function displays the image of the endoscopic live image and the image of an external device side by side on the monitor simultaneously.
Optical-digital observation This is observation using specific filtered light
NBI (Narrow Band Imaging) observation This is optical-digital observation using blue narrow band light.
RDI (Red Dichromatic Imaging) observation This is optical-digital observation using red narrow band light.
AFI (Auto-Fluorescence Imaging) observation This is optical-digital observation using blue light.
WLI (White Light Imaging) observation This is observation using white light.
SDTV This is a video signal with the same resolution as analog TV broadcasting.
HDTV HDTV is a higher resolution video signal than SDTV.
4K This video signal is a video signal that supports resolution four times that of HDTV and has a color gamut of BT.709.
Color gamut The range of available colors.
Isolation transformer The isolation transformer is a safety device that is used to isolate the noninsulated equipment with potentially high leakage currents to decrease the possibility of an electric shock.
OIP-1 INSTRUCTION MANUAL
5
Important Information - Please Read Before Use
User qualifications If there is an official standard on user qualifications to perform endoscopy and endoscopic treatment that is defined by the medical administration or other official institutions, such as academic societies on endoscopy, follow that standard. If there is no official qualification standard, the operator of this instrument must be a physician approved by the medical safety manager of the healthcare facility or person in charge of the department (department of internal medicine, etc.). The physician should be qualified to operate and perform safely the planned endoscopy and endoscopic treatment following guidelines set by the academic societies on endoscopy, etc., and considering the difficulty of endoscopy and endoscopic treatment. This manual does not explain or discuss endoscopic procedures.
Instrument compatibility Refer to “ System chart” on page 105 to confirm that the Endoscopy CAD system is compatible with the ancillary equipment being used. Using incompatible equipment may result in patient injury or equipment damage and makes it impossible to obtain the expected functionality. This instrument complies with the EMC standard for medical electrical equipment, edition 4 (IEC 60601-1-2: 2014). When connecting to an instrument that complies with a previous edition of the EMC standard for medical electrical equipment edition, the EMC characteristics could be vulnerable.
Repair and modification The Endoscopy CAD system does not contain any user-serviceable parts. Do not disassemble, modify, or attempt to repair it; patient or operator injury, equipment damage and/or the impossibility to obtain the expected functionality can result. Some problems that appear to be malfunctions may be correctable by referring to Chapter 9, “Troubleshooting”. If the problem cannot be resolved using the information in Chapter 9, contact Olympus. This instrument is to be repaired by Olympus technicians only. Olympus is not liable for any injury or damage which occurs as a result of repairs attempted by non-Olympus personnel.
6
OIP-1 INSTRUCTION MANUAL
Important Information - Please Read Before Use
Signal words The following signal words are used throughout this manual:
DANGER
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.
NOTE
Indicates additional helpful information.
Precautions Follow the warnings and cautions given below when handling the Endoscopy CAD system. This information is to be supplemented by the warnings and cautions given in each chapter.
WARNING • Never install and operate the Endoscopy CAD system in locations where: − The concentration of oxygen is high; − Oxidizing agents (such as nitrous oxide (N2O)) are present in the atmosphere; − Flammable gases are present in the atmosphere; − Flammable liquids are near. • Strictly observe the following precautions. Failure to do so may place the patient and medical personnel in danger of an electric shock. − Keep fluids away from all electrical equipment. If fluids are spilled on or into the unit, stop operation of the Endoscopy CAD system immediately and contact Olympus. − Do not prepare, inspect, or use the Endoscopy CAD system with wet hands.
OIP-1 INSTRUCTION MANUAL
7
Important Information - Please Read Before Use
WARNING • Do not use this instrument for any purpose other than its “ Intended use/Intended purpose” on page 3. • Be aware that the recognition processing of this instrument may not detect a lesion and may falsely detect normal tissue as abnormal. Considering this point, use the detection function just as a reference. • Do not use the Endoscopy CAD system in any place where it may be subject to strong electromagnetic radiation (for example, in the vicinity of a microwave therapeutic equipment, MRI, wireless set, short-wave therapeutic equipment, cellular/portable phone, etc.). This may impair the performance of the video system center. • Use the Endoscopy CAD system only under the conditions described in “ Environment” on page 107. Otherwise, improper performance, compromised safety, and/or equipment damage may result. • Be sure that this instrument is not used adjacent to or stacked with other equipment (other than the components of this instrument or system) to avoid electromagnetic interference. Electromagnetic interference may cause malfunction or failure. • Electromagnetic interference may occur on this instrument near equipment marked with the following symbol or other portable and mobile RF (Radio Frequency) communications equipment such as cellular phones. If electromagnetic interference occurs, mitigation measures may be necessary, such as reorienting or relocating this instrument, or shielding the location.
• The Endoscopy CAD system does not contain any user-serviceable parts. Do not disassemble, modify, or attempt to repair it; patient or operator injury, equipment damage and/or the impossibility to obtain the expected functionality can result.
8
OIP-1 INSTRUCTION MANUAL
Important Information - Please Read Before Use
CAUTION • Never insert anything into the ventilation grilles of the Endoscopy CAD system. It can cause an electric shock and/or a fire. • Use only the power cord specified by Olympus. Otherwise, equipment malfunction and/or power cord burns may result. Also, never use it for other equipment. Contact Olympus. • Do not allow a foreign object to penetrate the inside of the AC power inlet, portable memory port and other terminals described in Chapter 2, “Nomenclature and Functions”. The Endoscopy CAD system may malfunction. • Avoid using the Endoscopy CAD system in a dusty environment. This may damage the Endoscopy CAD system. • Do not apply excessive force to the Endoscopy CAD system and/or other instruments connected. Otherwise, damage and/or malfunction can occur.
NOTE • Confirm that there is no dust on the ventilation grills. Clean and vacuum dust from the ventilation grills using a vacuum cleaner, when necessary. Otherwise, the Endoscopy CAD system may break down and get damaged from overheating. • Do not use a pointed or hard object to press the buttons on the front panel. This may damage the buttons.
OIP-1 INSTRUCTION MANUAL
9
Important Information - Please Read Before Use
10
OIP-1 INSTRUCTION MANUAL
1.1 Checking the package contents
Chapter 1
1.1
Checking the Package Contents Ch.1
Checking the package contents
Match all items in the package with the components shown below. Inspect each item for damage. If the Endoscopy CAD system is damaged, a component is missing, or you have any questions, do not use the Endoscopy CAD system; immediately contact Olympus.
Endoscopy CAD system
Endoscopy CAD system (OIP-1)
Accessories
12G-SDI cable 1.5 M (MAJ-2426)
Communication cable 1.8 M (MAJ-1948)
Foot holder (MAJ-1433, 4 pcs. with a pattern)
Instruction manual
OIP-1 INSTRUCTION MANUAL
11
1.1 Checking the package contents
Ch.1
12
OIP-1 INSTRUCTION MANUAL
2.1 Front panel
Chapter 2
2.1
Nomenclature and Functions
Front panel Ch.2 1
2
3
4
5
6
7
8
9
12
No.
Nomenclature
10
11
Description
1
Power switch
Press to turn the Endoscopy CAD system ON or OFF.
2
Power indicator
Lights up when the Endoscopy CAD system is ON.
3
CAD indicator
Lights up when the CAD is ON.
4
CAD button
Switches the CAD between ON and OFF.
5
MODE button 1
Switches the CAD between ON and OFF and selects the display method of the CAD.
6
MODE button 2
Turns the sub-image display on and off and selects the display position when the CAD is ON. The message box position is same as the sub-image position.
7
BACK button
Returns to the previous screen.
8
Arrow button
Selects the item when the menu is displayed.
9
ENTER button
Determines the selected item.
10
MENU button
Turns the home menu display on and off.
11
Portable memory port
Insert the portable memory in this port to record data or to update the program inside of the Endoscopy CAD system.
12
Portable memory access indicator
Lights up green when the portable memory is set; blinks up orange while accessing the portable memory.
OIP-1 INSTRUCTION MANUAL
13
2.2 Rear panel
2.2
Rear panel
Ch.2 1
2
3
4
5
6
No.
Nomenclature
1
SDI IN terminal
Connect the video system center.
2
SDI OUT1 terminal
Connect the following devices.
7
8
Description
• Monitor • Video recorder 3
SDI OUT2 terminal
Connect the following devices. • Monitor • Video recorder
4
Potential equalization terminal
This terminal is connected to a potential equalization terminal of the other equipment connected to the Endoscopy CAD system. The electric potential of their equipment is made equal.
5
1000BASE-T terminal
This terminal cannot be used because this is used exclusively by Olympus service personnel for maintenance.
6
LINK OUT terminal
Connect the equipment having the LINK-IN and LINK-OUT terminals can be daisy-chained.
7
LINK IN terminal
Connect the video system center.
8
AC power inlet
Connects the provided power cord to supply the AC power via this inlet.
14
OIP-1 INSTRUCTION MANUAL
2.3 Side panels
2.3
Side panels
Ch.2 1
No.
Nomenclature
Description
1
Ventilation grilles
Prevent the internal temperature of endoscopy CAD system from rising by ventilating.
OIP-1 INSTRUCTION MANUAL
15