OLYMPUS
OLYMPUS Scopes Colono and Sigmodo System
CF-FH260AZL EVIS LUCERA COLONOVIDEOSCOPE Reprocessing Manual Feb 2007
Reprocessing Manual
76 Pages
Preview
Page 1
INSTRUCTIONS
EVIS LUCERA GASTROINTESTINAL VIDEOSCOPE
OLYMPUS GIF-FQ260Z EVIS LUCERA COLONOVIDEOSCOPE
OLYMPUS CF-FH260AZL/I
Refer to the endoscope’s companion manual, the “OPERATION MANUAL” whose cover lists the model of your endoscope, for inspection and operation information.
Contents
Contents Symbols...
1
Chapter 1
General Policy ...
2
1.1
Instructions...
2
1.2
Importance of cleaning, disinfection and sterilization...
2
1.3
Signal words ...
3
1.4
Precautions ...
3
1.5
Reprocessing before the first use/reprocessing and storage after use
6
Compatible Reprocessing Methods and Chemical Agents ...
7
2.1
Compatibility summary...
7
2.2
Detergent solution...
9
2.3
Disinfectant solution...
10
2.4
Rinsing water ...
10
2.5
Ethylene oxide gas sterilization...
10
2.6
Steam sterilization (autoclaving) of accessories ...
12
Chapter 2
Chapter 3
Cleaning, Disinfection and Sterilization Procedures ... 13
3.1
Required reprocessing equipment ...
13
3.2
Cleaning, disinfection and sterilization procedures for the endoscope ... 22
3.3
Precleaning ...
23
3.4
Leakage testing...
29
3.5
Manual cleaning...
35
3.6
High-level disinfection ...
49
3.7
Rinsing after high-level disinfection ...
51
3.8
Sterilization ...
55
3.9
Cleaning, disinfection and sterilization procedures for reusable parts and reprocessing equipment... 57
Chapter 4
Cleaning and Disinfection Equipment...
65
Chapter 5
Storage and Disposal...
68
5.1
Storage of the endoscope...
69
5.2
Storage of reusable parts and reprocessing equipment ...
70
5.3
Disposal ...
70
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
i
Contents
ii
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
Symbols
Symbols The meaning(s) of the symbol(s) shown on the back cover of this instruction manual are as follows:
Manufacturer
Authorized representative in the European Community
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
1
Chapter 1 General Policy
Chapter 1 General Policy 1.1
1.2
Instructions •
This manual contains the cleaning, disinfection and sterilization methods recommended by Olympus for the endoscopes listed on the front cover.
•
This instruction manual contains essential information on reprocessing these instruments safely and effectively.
•
Before reprocessing, thoroughly review the manuals of the reprocessing chemicals and all equipment which will be used and reprocess the equipment as instructed.
•
Note that the complete instruction manual set for this endoscope consists of this manual and the “OPERATION MANUAL” whose cover lists the model of your endoscope, both of which accompanied the endoscope at shipment.
•
Keep this manual and all related manuals in a safe and accessible location.
•
If you have any questions or comments about any information in this manual, or if a problem that cannot be solved occurs while reprocessing, contact Olympus.
Importance of cleaning, disinfection and sterilization The medical literature reports incidents of cross contamination resulting from improper cleaning, disinfection or sterilization. It is strongly recommended that all individuals engaged in reprocessing closely observe all instructions given in this manual and the manuals of all ancillary equipment, and have a thorough understanding of the following items:
2
•
Professional health and safety criteria of your hospital
•
Individual cleaning, disinfection and sterilization protocols
•
Structure and handling of endoscopic equipment
•
Handling of pertinent chemicals
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
Chapter 1 General Policy
For the types and conditions of the means of cleaning, disinfection and sterilization to be adopted, please make judgments from your professional viewpoints.
1.3
Signal words The following signal words are used throughout this manual:
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.
Indicates additional helpful information.
1.4
Precautions •
Failure to properly clean and high-level disinfect or sterilize endoscopic equipment after each procedure may compromise patient safety. To minimize the risk of transmitting diseases from one patient to another, after each procedure the endoscope and its ancillary equipment must undergo thorough manual cleaning followed by high-level disinfection or sterilization as described in Chapter 3, “Cleaning, Disinfection and Sterilization Procedures”. Reprocess not only the external surface of the endoscope but also all channels.
•
ALL channels of the endoscope, including the auxiliary water channel where existing, MUST be cleaned and high-level disinfected or sterilized during EVERY reprocessing cycle, even if the channels were not used during the previous patient procedure. Otherwise, insufficient cleaning and disinfection or sterilization of the endoscope may pose an infection-control risk to the patient and/or operators performing the next procedure with the endoscope.
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
3
Chapter 1 General Policy
4
•
If the endoscope is not cleaned meticulously, effective disinfection or sterilization may not be possible. Clean the endoscope and accessories thoroughly before disinfection or sterilization to remove microorganisms or organic material that could reduce the efficacy of disinfection or sterilization.
•
Olympus only confirms validation of endoscope reprocessors it recommends. When using an endoscope reprocessor that is not recommended by Olympus, the manufacturer of the endoscope reprocessor is responsible for validation of the reprocessor with the endoscope models listed in its instruction manual.
•
Before using an endoscope reprocessor, confirm that it is capable of reprocessing the endoscope including all channels. If there are channels and/or other parts of the endoscope that cannot be cleaned and/or high-level disinfected by the endoscope reprocessor, manual cleaning and high-level disinfection or sterilization must be performed as described in Chapter 3, “Cleaning, Disinfection and Sterilization Procedures” after using the endoscope reprocessor. Otherwise, insufficient cleaning and disinfection or sterilization of the endoscope may pose an infection-control risk to the patient and/or operators performing the next procedure with the endoscope. If you are uncertain as to the ability of your endoscope reprocessor to clean and high-level disinfect the endoscope including all channels, contact the endoscope reprocessor’s manufacturer for specific instructions and/or information on connectors.
•
Patient debris and reprocessing chemicals are hazardous. Wear personal protective equipment to guard against dangerous chemicals and potentially infectious material. During cleaning and disinfection or sterilization, wear appropriate personal protective equipment, such as eye wear, face mask, moisture-resistant clothing and chemical-resistant gloves that fit properly and are long enough so that your skin is not exposed. Always remove contaminated personal protective equipment before leaving the reprocessing area.
•
Thoroughly rinse off the disinfectant solution. Rinse the external surface of the endoscope, all channels and the cleaning equipment thoroughly with water to remove any disinfectant solution residue.
•
The disinfection/sterilization room must be adequately ventilated. Adequate ventilation protects against the buildup of toxic chemical fumes.
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
Chapter 1 General Policy
•
Store alcohol in an air-tight container. Alcohol stored in an open container is a fire hazard and will result in a loss of efficacy due to evaporation.
•
Be sure to perform a leakage test on the endoscope prior to manual cleaning, and do not use the endoscope if a leak is detected. Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism or other malfunctions.
•
Prior to each procedure, confirm that the endoscope has undergone proper cleaning, disinfection and sterilization. If it is determined that the endoscope has not been properly reprocessed, reprocess it again following the instructions given in this manual.
•
With the cleaning, disinfection and sterilization methods stated in this instruction manual, prions, which are considered to be the pathogenic substance of the Creutzfeldt-Jakob disease (CJD) cannot be destroyed or inactivated. When using this instrument on a patient with CJD or variant Creutzfeldt-Jakob disease (vCJD), be sure to use this product for such patient only and/or immediately dispose of this product after use in an appropriate manner. For methods to handle CJD, please follow the respective guidelines in your country.
•
This instrument is not durable, or does not have sufficient durability against the respective methods stated in the guidelines of each country for destroying or inactivating prions. For information on the durability against each method, please contact Olympus. If cleaning, disinfection and sterilization methods not stated in this instruction manual are performed, Olympus cannot guarantee the effectiveness, safety and durability of this instrument. Make sure to confirm that there is no abnormality before use, and use under responsibility of a physician. Do not use if any abnormality is found.
•
When aerating or irrigating the endoscope channels, the air or water pressure must not exceed 0.5 MPa (5 bar). Higher pressures may cause damage to the endoscope.
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
5
Chapter 1 General Policy
Model
•
When reprocessing an endoscope that has the flexibility adjustment mechanism, make sure that the insertion tube is set to its most-flexible condition (indicated by the “z” mark on the flexibility adjustment ring). If the endoscope is cleaned and disinfected or sterilized while the insertion tube is too rigid, the endoscope may be damaged (for CF-FH260AZL/I only).
•
When reprocessing an endoscope, confirm that the water-resistant cap is securely attached to the endoscope connector and the zoom connector (for CF-FH260AZL/I only) before immersion in reprocessing fluids. If the water-resistant cap is not securely attached, water, detergent solution and/or disinfectant solution could enter the endoscope and damage the equipment. The compatible water-resistant caps are shown in Table 1.1. Water-resistant cap
GIF-FQ260Z
MH-553
CF-FH260AZL/I
MH-553 and MAJ-583
Table 1.1
1.5
Reprocessing before the first use/reprocessing and storage after use This instrument was not cleaned, disinfected or sterilized before shipment. Before using this instrument for the first time, reprocess it according to the instructions given in this manual. After using this instrument, reprocess and store it according to the instructions given in this manual. Improper and/or incomplete reprocessing or storage can present an infection control risk, cause equipment damage or reduce performance.
6
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
Chapter 2 Compatible Reprocessing Methods and Chemical Agents
Chapter 2 Compatible Reprocessing Methods and Chemical Agents 2.1
Compatibility summary Olympus endoscopic equipment is compatible with several methods of reprocessing. Certain components and accessories, however, are not compatible with some methods, which can cause equipment damage. For appropriate reprocessing methods, refer to Table 2.1 below, the recommendations of your infection control committee and all national and local hospital guidelines and policies.
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
7
Chapter 2 Compatible Reprocessing Methods and Chemical Agents
Steam sterilization (autoclaving) Ethylene oxide gas sterilization ACECIDE disinfectant solution ∗3 (use OER-A) 2 – 3.5% glutaraldehyde 70% ethyl or isopropyl alcohol Detergent solution Ultrasonic cleaning Endoscope
∗1
Water-resistant cap (MH-553) Chain for Water-resistant cap (MAJ-1119) Water-resistant cap (MAJ-583)
∗2
∗2
Channel cleaning brush (BW-20T) Channel-opening cleaning brush (MH-507) Air/water valve (MH-438) Suction valve (MH-443) Channel plug (MH-944) Injection tube (MH-946) AW channel cleaning adapter (MH-948) Suction cleaning adapter (MH-856) Mouthpiece (MB-142) Auxiliary water tube (MAJ-855) Biopsy valve (MB-358) applicable
not applicable
Table 2.1
8
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
Chapter 2 Compatible Reprocessing Methods and Chemical Agents
∗1 The endoscope is only compatible with ultrasonic cleaning as performed in an endoscope reprocessor such as OER and OER-A (OER and OER-A are not available in some areas). ∗2 The water-resistant cap and the chain for water-resistant cap can only be ultrasonically cleaned if connected to the endoscope that is being cleaned in an endoscope reprocessor with an ultrasonic cleaning phase. ∗3 ACECIDE disinfectant solution is exclusively for OER-A.
Alcohol is not a sterilant or high-level disinfectant.
2.2
•
Endo-therapy accessories which are marked by the words “AUTOCLAVE” or “AUTOCLAVABLE”, or accessories with a green model reference label are compatible with autoclaving.
•
The endoscope is compatible with some endoscope reprocessors distributed by Olympus. Refer to the respective instruction manual for details on operation. For any other details, please contact your local Olympus organization.
Detergent solution Use a medical-grade, low-foaming, neutral pH detergent or enzymatic detergent and follow the manufacturer’s dilution and temperature recommendations. Contact Olympus for the names of specific brands that have been tested for compatibility with the endoscope. Do not reuse detergent solutions.
Excessive detergent foaming can prevent fluid from adequately contacting the interior of the channels.
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
9
Chapter 2 Compatible Reprocessing Methods and Chemical Agents
2.3
Disinfectant solution In general, 2.0 – 3.5% glutaraldehyde solutions, when used according to the manufacturer’s instructions for achieving high-level disinfection, are compatible with Olympus endoscopes. Contact Olympus for the names of specific brands that have been tested for compatibility with the endoscope. If the disinfectant solution is reused, routinely check its efficacy with a test strip according to the manufacturer’s recommendations. Do not use solutions beyond their expiration date.
2.4
Rinsing water Once removed from disinfectant solution, the instrument must be thoroughly rinsed with sterile water to remove any disinfectant residue. If sterile water is not available, clean potable tap water or water which has been processed (e.g. filtered) to improve its microbiological quality may be used. When non-sterile water is used after disinfection, wipe the endoscope and flush the channels with 70% ethyl or isopropyl alcohol, then air-dry all internal channels to inhibit the growth of residual bacteria. Do not reuse rinsing water.
2.5
Ethylene oxide gas sterilization The instruments and other accessories listed as compatible with ethylene oxide gas sterilization in Table 2.1 on page 8 can be sterilized by ethylene oxide gas and aerated within the parameters given in Table 2.2. When performing ethylene oxide gas sterilization, follow the cleaning, disinfection and sterilization protocols of your hospital and the instruction manuals of the sterilization equipment.
10
•
Before sterilization, the instrument must be thoroughly cleaned and dried. Residual moisture inhibits sterilization.
•
The results of sterilization depend on various factors such as how the sterilized instrument was packed or the positioning, method of placing and loading of the instrument in the sterilization device. Please verify the sterilization effects by using biological or chemical indicators. Also follow the guidelines for sterilization issued by medical administrative authorities, public organizations or the infection management sections at each medical facility, as well as the instruction manual of the sterilization device.
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
Chapter 2 Compatible Reprocessing Methods and Chemical Agents
•
All instruments must be properly aerated following ethylene oxide gas sterilization to remove toxic ethylene oxide residuals.
•
Disconnect the water-resistant caps from the endoscope connector and the zoom connector (for CF-FH260AZL/I only) before ethylene oxide gas sterilization. If the water-resistant caps are attached during ethylene oxide gas sterilization, the air inside the endoscope will expand and rupture the covering of the bending section and/or damage the angulation mechanism.
•
Exceeding the recommended parameters may cause equipment damage (see Table 2.2).
Ethylene oxide gas exposure parameters Process Ethylene oxide gas sterilization
Aeration
Parameters Temperature
57°C
Pressure
0.1 – 0.17 MPa
Humidity
55%
Exposure time
1.75 hours
Ethylene oxide gas concentration
0.6 – 0.7 mg/cm3 (600 – 700 mg/l)
Minimum aeration parameters
12 hours in an aeration chamber at 50 – 57°C or 7 days at room temperature
Table 2.2
Gas mixture 20% Ethylene oxide gas/80% CO2
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
11
Chapter 2 Compatible Reprocessing Methods and Chemical Agents
2.6
Steam sterilization (autoclaving) of accessories The accessories listed as compatible with steam sterilization (autoclaving) in Table 2.1 on page 8 can be sterilized by steam within the parameters given in Table 2.3. When steam sterilizing, follow the cleaning, disinfection and sterilization protocols of your hospital and the sterilization equipment manufacturer’s instructions.
The results of sterilization depend on various factors such as how the sterilized instrument was packed or the positioning, method of placing and loading of the instrument in the sterilization device. Please verify the sterilization effects by using biological or chemical indicators. Also follow the guidelines for sterilization issued by medical administrative authorities, public organizations or the infection management sections at each medical facility, as well as the instruction manual of the sterilization device.
•
Do not steam sterilize the endoscope. Steam sterilization (autoclaving) will severely damage the endoscope.
•
Exceeding the recommended parameters may cause equipment damage (see Table 2.3).
Process Prevacuum
Table 2.3
12
Parameters Temperature
132 – 134°C
Exposure time
5 minutes
Steam sterilization (autoclaving) exposure parameters
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
Chapter 3 Cleaning, Disinfection and Sterilization Procedures ALL channels of the endoscope, including the auxiliary water channel where existing, MUST be cleaned and high-level disinfected or sterilized during EVERY reprocessing cycle, even if the channels were not used during the previous patient procedure. Otherwise, insufficient cleaning and disinfection or sterilization of the endoscope may pose an infection-control risk to the patient and/or operators performing the next procedure with the endoscope.
3.1
•
Do not coil the endoscope’s insertion tube or universal cord with a diameter of less than 12 cm. The endoscope can be damaged if coiled too tightly.
•
For proper reprocessing results, do not coil the insertion tube or the universal cord with a diameter of less than 40 cm. If the diameter is less than 40 cm, it will be difficult to insert the channel cleaning brush (BW-20T).
Required reprocessing equipment Preparation of the equipment Prior to cleaning, disinfection or sterilization, prepare the equipment shown in Figure 3.1.
Use basins which are at least 40 cm by 40 cm in size and deep enough to allow the endoscope to be completely immersed.
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
13
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
Water-resistant cap (MH-553)
Channel cleaning brush (BW-20T) Injection tube (MH-946)
Channel-opening cleaning brush (MH-507)
Suction pump (KV-4/5, SSU-2) (sold separately)
AW channel cleaning adapter (MH-948)
Channel plug (MH-944)
Leakage tester (MB-155) (sold separately)
Light source (CLV-260SL/260NBI/260) (sold separately)
Suction cleaning adapter (MH-856)
Auxiliary water tube (MAJ-855) (for CF-FH260AZL/I only)
Maintenance unit (MU-1) (sold separately)
Water-resistant cap (MAJ-583) (for CF-FH260AZL/I only)
• Personal protective equipment
• Soft brush
• 500 cm3 (500 ml) containers
• Clean lint-free cloths
• Large basins with tight fitting lids∗1
• Sterile lint-free cloths
• Small basins with tight fitting lids (size: 25 (W) × 10 (H) × 25 (D) cm or more)
• Sterile cotton swabs
• Clean water
• 30 cm3 (30 ml) syringe
• Detergent solution • Disinfectant solution • Sterile water • 70% ethyl or isopropyl alcohol
∗1 Use basins which are at least 40 cm by 40 cm in size and deep enough to allow the endoscope to be completely immersed. Figure 3.1
14
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
Reprocessing equipment parts and functions For inspection of other equipment than that mentioned below, refer to the instruction manual for the equipment being used.
Water-resistant cap (MH-553) The water-resistant cap is attached to the electrical connector on the endoscope to protect the connector from water penetration during reprocessing. For leakage testing, the venting connector on the water-resistant cap must be connected to the leakage tester (see Figure 3.2).
Groove Venting connector Seal
Figure 3.2
Water-resistant cap (MAJ-583, for CF-FH260AZL/I only) The water-resistant cap (MAJ-583) is attached to the zoom connector of the endoscope to protect the connector from water penetration during reprocessing. The water-resistant cap is connected to the endoscope by a chain (see Figure 3.3).
Chain
Seal
Screw
Figure 3.3
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
15
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
Channel plug (MH-944) The channel plug is used to plug the openings of the instrument channel port, air/water and suction cylinders during cleaning (see Figure 3.4).
Biopsy valve cap
Cylinder plug
Figure 3.4
Injection tube (MH-946) The injection tube is used to inject detergent solution, disinfectant solution, water and alcohol into the air/water channel, instrument channel and suction channel and to flush air through the channels to expel fluids (see Figure 3.5). Suction channel tube Notice card Suction channel port
Connector plug
Air pipe port Air/water channel port Filter tube
Air/water channel tube
Suction port (including the filter mesh)
Figure 3.5
16
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
Channel cleaning brush (BW-20T) The channel cleaning brush is used to brush the inside of the instrument and suction channels and the interior and/or openings of the suction valve, air/water valve, AW channel cleaning adapter and biopsy valve (see Figure 3.6).
Metal tip
Shaft
Brush head
Figure 3.6
Channel-opening cleaning brush (MH-507) The channel-opening cleaning brush is used to brush the suction cylinder and the instrument channel port (see Figure 3.7).
Brush head
Brush handle
Shaft
Figure 3.7
EVIS LUCERA GIF-FQ260Z, CF-FH260AZL/I REPROCESSING MANUAL
17