OLYMPUS
OLYMPUS Scopes Gastrointestinal System
GIF-EZ1500 GASTROINTESTINAL VIDEOSCOPE Operation Manual
Operation Manual
128 Pages
Preview
Page 1
OPERATION MANUAL
INSTRUCTIONS
GASTROINTESTINAL VIDEOSCOPE
Symbols
1
Important Information - Please Read Before Use
2
Chapter 1
Checking the Package Contents
13
Chapter 2
Instrument Nomenclature and Specifications
15
Chapter 3
Preparation and Inspection
27
Chapter 4
Operation
65
Chapter 5
Troubleshooting
89
OLYMPUS GIF-EZ1500
Appendix
Refer to the endoscope’s companion manual, the “REPROCESSING MANUAL” with your endoscope model listed on the cover, for reprocessing information.
99
Contents
Contents Symbols ... 1 Important Information - Please Read Before Use ... 2 Intended use/Intended purpose ... 2 Contraindications ... 2 Applicability of endoscopy and endoscopic treatment ... 2 Instruction manual ... 3 User qualifications ... 4 Instrument compatibility ... 4 Reprocessing before the first use/reprocessing and storage after use ... 4 Spare equipment ... 4 Maintenance management ... 5 Prohibition of improper repair and modification ... 5 Signal words ... 5 Precautions ... 6 Precautions for disappeared or frozen endoscopic image ... 10 Precautions for illumination light emitted from the endoscope’s distal end ... 11 Examples of inappropriate handling ... 12
Chapter 1 Checking the Package Contents ... 13 1.1
Checking the package contents list ... 13
Chapter 2 Instrument Nomenclature and Specifications ... 15 2.1
Nomenclature and functions ... 15 Control section, insertion section ... 16 Endoscope connector ... 20
2.2
Specifications ... 22 Environment ... 22 Function list ... 23 Specifications ... 24
Chapter 3 Preparation and Inspection ... 27 3.1
The workflow of preparation and inspection ... 27
3.2
Preparation of the equipment ... 29
3.3
Inspection of the endoscope ... 31 Inspection of the endoscope ... 31 Inspection of the bending mechanisms ... 34
3.4
Inspection of accessories ... 38 Inspection of the air/water and suction valves ... 38 Inspection of the biopsy valve (MB-358) ... 39 Inspection of the single use biopsy valve (MAJ-1555, sold separately) ... 40 Inspection of the auxiliary water inlet cap (MAJ-215) ... 40 Inspection of the auxiliary water tube (MAJ-855) ... 41 Inspection of the mouthpiece (MB-142) ... 41 GIF-EZ1500 OPERATION MANUAL
i
Contents
3.5
Attaching accessories to the endoscope ... 42 Attaching the suction valve ... 42 Attaching the air/water valve ... 44 Attaching the biopsy valve ... 45 Attaching the auxiliary water inlet cap ... 46
3.6
Inspection of ancillary equipment ... 46
3.7
Connection of the endoscope and ancillary equipment ... 46 Connection to the video system center ... 47 Connection of the water container ... 48 Connection of the suction tube ... 49 Connection of the auxiliary water tube ... 50
3.8
Inspection of the endoscopic system ... 51 Inspection summary ... 51 Inspection of the endoscopic image ... 52 Inspection of the focus switching function ... 53 Inspection of the remote switches ... 55 Inspection of the air-feeding function ... 56 Inspection of the objective lens cleaning function ... 57 Inspection of the suction function ... 59 Inspection of the instrument channel ... 61 Inspection of the auxiliary water feeding function ... 62
Chapter 4 Operation ... 65 4.1
Precautions ... 65
4.2
Insertion ... 68 Holding and manipulating the endoscope ... 68 Insertion of the endoscope ... 69 Observation of the endoscopic image ... 70 Focus switching ... 71 Angulation of the distal end ... 72 Air/water feeding and suction ... 73
4.3
Using EndoTherapy accessories ... 79 Insertion of EndoTherapy accessories into the endoscope ... 80 Operation of EndoTherapy accessories ... 82 Withdrawal of EndoTherapy accessories ... 83 Use of nonflammable gases ... 83 High-frequency cauterization treatment ... 84
4.4
Withdrawal of the endoscope ... 85
4.5
Transportation of the endoscope ... 86 Transporting within the healthcare facility ... 86 Transporting outside the healthcare facility ... 87
ii
GIF-EZ1500 OPERATION MANUAL
Contents
Chapter 5 Troubleshooting ... 89 5.1
Troubleshooting ... 89
5.2
Troubleshooting guide ... 90 Angulation ... 90 Air/water feeding ... 90 Suction ... 91 Image quality or brightness ... 91 Auxiliary water feeding ... 92 EndoTherapy accessories ... 93 Other ... 93
5.3
Withdrawal of the endoscope with an irregularity ... 94 Withdrawal when the WLI, NBI, and RDI endoscopic images appear on the monitor ... 94 Withdrawal when either the WLI, NBI, or RDI endoscopic image does not appear on the monitor ... 95 Withdrawal when no endoscopic image appears on the monitor or a frozen image cannot be restored ... 96 Withdrawal when the focus setting cannot be switched ... 97
5.4
Returning the endoscope for repair ... 98
Appendix ... 99 Combination equipment ... 99 System chart ... 99 Reprocessing equipment ... 102 Compatible video system center ... 103 Compatible light source ... 103 Compatible accessories ... 103 Compatible EndoTherapy accessories ... 104 Compatible electrosurgical accessories ... 108
Labeling information for irrigation system ... 111 Backflow-prevention valve ... 111 Distal irrigation system and proximal irrigation system ... 111 Consumable device ... 113 Reusable device ... 113
EMC information ... 114
GIF-EZ1500 OPERATION MANUAL
iii
Contents
iv
GIF-EZ1500 OPERATION MANUAL
Symbols
Symbols The meaning(s) of the symbol(s) shown on the component packaging, the back cover of the instruction manual, and/or the instrument are as follows: Symbol
Blue
Description
Symbol
Description
Follow instructions for use.
Consult instructions for use.
Endoscope
TYPE BF applied part
Single use only
Not Made with Natural Rubber Latex
Keep away from sunlight
Keep dry
Lot number
Manufacturer
Date of Manufacture
Authorized representative in the European Community
Translation
Importer (into European Union)
Ingress protection rating is 7.
Medical device
GIF-EZ1500 OPERATION MANUAL
1
Important Information - Please Read Before Use
Important Information - Please Read Before Use
Intended use/Intended purpose This instrument is intended to be used with an Olympus video system center, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. The GASTROINTESTINAL VIDEOSCOPE GIF-EZ1500 is indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum).
Contraindications None known.
Applicability of endoscopy and endoscopic treatment If there are official standards on the applicability of endoscopy and endoscopic treatment that are defined by the healthcare facility’s administrators or other official institutions, such as academic societies on endoscopy, follow those standards. Before starting endoscopy and endoscopic treatment, thoroughly evaluate its properties, purposes, effects, and possible risks (their nature, extent, and probability). Perform endoscopy and endoscopic treatment only when its potential benefits are greater than its risks. Fully explain to the patient the potential benefits and risks of the endoscopy and endoscopic treatment as well as any examination/treatment methods that can be performed in its place, and perform the endoscopy and endoscopic treatment only after obtaining the consent of the patient. Even after starting the endoscopy and endoscopic treatment, continue to evaluate the potential benefits and risks, and immediately stop the endoscopy/treatment and take proper measures if the risks to the patient become greater than the potential benefits.
2
GIF-EZ1500 OPERATION MANUAL
Important Information - Please Read Before Use
Instruction manual This instruction manual contains essential information on using this instrument safely and effectively. Before use, thoroughly review this manual and the manuals for all equipment that will be used during the procedure and use the equipment as instructed. Note that the complete instruction manual set for this endoscope consists of this manual and the “REPROCESSING MANUAL” with your endoscope model listed on the cover. It also accompanied the endoscope at shipment. Keep this and all related instruction manuals in a safe, accessible location. If you have any questions or comments about any information in this manual, contact Olympus.
Terms used in this manual RDI (Red Dichromatic Imaging) observation: This is optical-digital observation using red illumination light.
NBI (Narrow Band Imaging) observation: This is optical-digital observation using narrowband light.
WLI (White Light Imaging) observation: This is observation using white light.
Normal focus mode: In this observation mode, the focusing position of the endoscope is optically shifted to a normal position. At this position, a wider focusing depth can be obtained.
Near focus mode: In this observation mode, the focusing position of the endoscope is optically shifted to a near position. At this position, a close-up image can be obtained.
Focus switching: This is an action to switch the focusing position between the normal focus mode and the near focus mode.
Image sensor: The image sensor is a device that converts light into electrical signals.
GIF-EZ1500 OPERATION MANUAL
3
Important Information - Please Read Before Use
User qualifications Endoscopy and endoscopic treatment should be performed by a physician who meets the following requirements. If there is an official standard on user qualifications to perform endoscopy and endoscopic treatment that is defined by the medical administration or other official institutions, such as academic society on endoscopy, endoscopy and endoscopy physicians follow that standard. If there is no official qualification standard, the operator of this instrument must be approved by the medical safety manager of the healthcare facility or person in charge of the department (department of internal medicine, etc.). The physician should be qualified to operate and perform safely the planned endoscopy and endoscopic treatment following guidelines set by the academic societies on endoscopy, etc., and considering the difficulty of endoscopy and endoscopic treatment. This manual does not explain or discuss endoscopic procedures. This instrument should be used by persons trained in how to use it in the cases other than endoscopy and endosopic treatment.
Instrument compatibility Refer to “Combination equipment” on page 99 to confirm that this instrument is compatible with the ancillary equipment being used. Using incompatible equipment can result in patient or operator injury and/or equipment damage. This instrument complies with the EMC standard for medical electrical equipment, edition 4 (IEC 60601-1-2: 2014). When connecting to an instrument that complies with a previous edition of the EMC standard for medical electrical equipment edition, the EMC characteristics could be vulnerable.
Reprocessing before the first use/reprocessing and storage after use This instrument was not reprocessed before shipment. Before using this instrument for the first time, reprocess it according to the instructions given in the endoscope’s companion “REPROCESSING MANUAL” with your endoscope model listed on the cover. After using this instrument, reprocess and store it according to the instructions given in the endoscope’s companion “REPROCESSING MANUAL”. Improper and/or incomplete reprocessing or storage can pose an infection control risk, cause equipment damage, or reduce performance.
Spare equipment Be sure to prepare another endoscope to avoid interruption of the examination due to equipment failure or malfunction.
4
GIF-EZ1500 OPERATION MANUAL
Important Information - Please Read Before Use
Maintenance management The probability of failure of the endoscope and ancillary equipment increases as the number of procedures performed and/or the total operating hours increase. In addition to the inspection before each procedure, the person in charge of medical equipment maintenance in each healthcare facility should periodically inspect the items specified in this manual following applicable regulations, guidelines, etc. required of you. An endoscope with an observed irregularity should not be used, but should be inspected by following Section 5.2, “Troubleshooting guide”. If the irregularity is still observed after inspection, contact Olympus.
Prohibition of improper repair and modification This instrument does not contain any user-serviceable parts. Do not disassemble, modify, or attempt to repair it; patient or operator injury and/or equipment damage may result. Equipment that has been disassembled, repaired, altered, changed, or modified by persons other than Olympus’ own authorized service personnel is excluded from Olympus’ limited warranty and is not warranted by Olympus in any manner.
Signal words The following signal words are used throughout this manual:
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.
NOTE
Indicates additional helpful information.
GIF-EZ1500 OPERATION MANUAL
5
Important Information - Please Read Before Use
Precautions Follow the warnings and cautions given below when handling this endoscope. This information is to be supplemented by the warnings and cautions given in each chapter.
WARNING • Do not use this endoscope for any purpose other than its intended use. Patient or operator injury and/or equipment damage may result. • After using this endoscope, reprocess and store it according to the instructions given in the endoscope’s companion “REPROCESSING MANUAL” with your endoscope model listed on the cover. Using improperly or incompletely reprocessed or stored instruments may cause patient cross-contamination and/or infection. • Before endoscopy, remove any metallic objects (watch, glasses, necklace, etc.) from the patient. Performing high-frequency cauterization treatment while the patient is wearing metallic objects may cause burns on the patient in areas around the metallic objects. • Do not strike, hit, or drop the distal end, insertion tube, bending section, control section, universal cord, or endoscope connector of the endoscope. Also, do not bend, pull, or twist the distal end, insertion tube, bending section, control section, universal cord, or endoscope connector of the endoscope with excessive force. The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations. It could also cause parts of the endoscope to fall off inside the patient. • Never perform angulation control forcibly or abruptly. Never forcefully pull, twist, or rotate the angulated bending section. Patient injury, bleeding, and/or perforation may result due to unintended retroflexion of the bending section. It may also become impossible to straighten the bending section during an examination. • Never insert or withdraw the endoscope’s insertion section while the bending section is locked in position. Patient injury, bleeding, and/or perforation may result. • Never operate the bending section, feed air or perform suction, insert or withdraw the endoscope’s insertion section, or use EndoTherapy accessories without viewing the endoscopic image or while the endoscopic image is frozen. Patient injury, bleeding, and/or perforation may result. • Never insert or withdraw the insertion section abruptly or with excessive force. Patient injury, bleeding, and/or perforation may result. • If it is difficult to insert the endoscope, do not forcibly insert the endoscope; stop the endoscopy. Forcible insertion can result in patient injury, bleeding, and/or perforation.
6
GIF-EZ1500 OPERATION MANUAL
Important Information - Please Read Before Use
WARNING • When using the electronic zoom function of the video system center, never insert or withdraw the endoscope’s insertion section or use EndoTherapy accessories while the image is magnified. Patient injury, bleeding, and/or perforation can result. • Never insert or withdraw the endoscope’s insertion section while the focus setting is set to the near focus mode. Patient injury, bleeding, and/or perforation can result. • Do not touch the light guide on the endoscope connector immediately after removing it from the output socket of the video system center because it is extremely hot. Operator or patient burns can result. • If the endoscopic image becomes dimmer during the procedure, it may indicate that blood or mucus is adhering to the light guide lens on the distal end of the endoscope. Immediately withdraw the endoscope from the patient, remove blood or mucus, and confirm that the light guide lens has no irregularities to use it again. If you continue to use the endoscope with its obstructed light guide lens, the temperature at the distal end of the endoscope may rise, which may cause patient injury or operator and/or patient burns. • When the endoscopic image does not appear on the monitor, the image sensor may have been damaged. Turn the video system center OFF immediately. Continued power supply in such a case will cause the distal end of the endoscope to become hot and could cause operator and/or patient burns. • Do not rely on the NBI or RDI observation mode alone for primary detection of lesions or to make a decision regarding any potential diagnostic or therapeutic intervention. • During endoscopic treatment, keep the insertion section and the bending section as straight as possible. If there is a loop or a bend on the insertion section or the bending section, the operation cannot be performed as intended, and patient injury, bleeding, and/or perforation can result.
CAUTION • Do not pull the universal cord during an examination. The endoscope connector will be pulled out from the output socket of the video system center and the endoscopic image will disappear. • Do not coil the insertion tube or universal cord with a diameter of less than 12 cm. Equipment damage may result. • Do not attempt to bend or twist the endoscope’s insertion section with excessive force. The insertion section may be damaged. • Do not apply shock to the distal end of the endoscope, in particular the objective lens surface at the distal end of the endoscope. An abnormal endoscopic image may result.
GIF-EZ1500 OPERATION MANUAL
7
Important Information - Please Read Before Use
CAUTION • Do not use the distal end of the endoscope close to locations exposed to strong magnetism. The focus switching function of the endoscope may malfunction and an endoscopic image can not be obtained. • Do not twist or bend the bending section with your hands. Equipment damage may result. • Do not squeeze the bending section forcefully. The covering of the bending section may stretch or break and cause water leakage. • Turn the video system center ON only when the endoscope connector is connected to the output socket of the video system center. In particular, confirm that the video system center is OFF or the “Disconnect scope” button is tapped before connecting or disconnecting the endoscope connector. Failure to do so can result in equipment damage, including destruction of the image sensor. • The endoscope’s remote switches cannot be removed from the control section. Pressing, pulling, or twisting them with excessive force can break the switches and/or cause water leakage. • If remote switch 1 does not return to the OFF position after being pressed strongly from the side, gently pull the switch upwards to return it to the OFF position. • Do not hit or bend the electrical contacts on the endoscope connector. The connection to the video system center may be impaired and a faulty contact can result. • If endoscope’s suction is insufficient, select another suction system without using the endoscope and use it according to the directions given in its instruction manual. Otherwise, a proper endoscopic image may not appear on the monitor. • This endoscope supports radio communication function (receive frequency: 13.56 MHz) that identifies endoscopes. Electromagnetic interference may occur on this endoscope near equipment marked with the following symbol or other portable and mobile RF (Radio Frequency) communications equipment, such as cellular phones. If electromagnetic interference occurs, mitigation measures may be necessary, such as moving the RF communications equipment away, reorienting or relocating this endoscope, or shielding the location.
• Be sure that this endoscope is not used adjacent to or stacked with other equipment (other than the components of this endoscope or system) to avoid electromagnetic interference.
8
GIF-EZ1500 OPERATION MANUAL
Important Information - Please Read Before Use
NOTE • This endoscope contains a memory chip that stores information about the endoscope and communicates this information to the video system center CV-1500. • In case of any serious incident happens to user or patient while operating the device, the incident should be reported to Olympus and the competent authority of the respective member state or country.
GIF-EZ1500 OPERATION MANUAL
9
Important Information - Please Read Before Use
Precautions for disappeared or frozen endoscopic image WARNING • If the endoscopic image disappears unexpectedly or the frozen image cannot be restored during an examination, immediately stop using the endoscope and withdraw it from the patient as described in Section 5.3, “Withdrawal of the endoscope with an irregularity”. Inserting or withdrawing the endoscope, using EndoTherapy accessories, performing suction, feeding air, or performing angulation control under these conditions could result in patient injury, bleeding, and/or perforation. • Follow the precautions given below. Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination. − Connect the endoscope connector to the video system center completely by pushing the endoscope connector until it clicks. Otherwise, a faulty contact can result. − Do not bend, hit, pull, or twist the insertion section, bending section, control section, universal cord, and endoscope connector. The endoscope may be damaged, and water leakage and/or breakage of internal parts like the cable can result. − Before connecting the endoscope connector to the video system center, confirm that the endoscope connector, including the electrical contacts, is completely dry and clean. If the endoscope is used with the electrical contacts wet and/or dirty, the endoscope and video system center may malfunction. − If air bubbles emerge from the endoscope continuously during the leakage test, do not use the endoscope. Water may enter the endoscope and cause a short circuit. This may result in image sensor damage.
CAUTION • Turn the video system center ON only when the endoscope connector is connected to the video system center. In particular, confirm that the video system center is OFF or the “Disconnect scope” button is tapped before connecting or disconnecting the endoscope connector. Failure to do so can result in equipment damage, including destruction of the image sensor. • Do not hit or bend the electrical contacts on the endoscope connector. The connection to the video system center may be impaired and a faulty contact can result.
10
GIF-EZ1500 OPERATION MANUAL
Important Information - Please Read Before Use
Precautions for illumination light emitted from the endoscope’s distal end WARNING Do not contact patient’s body cavity with distal end of endoscope for a long time during endoscopic observation or treatment, or irradiate certain place. Burn may result due to contact with light energy or heat caused by irradiation.
NOTE Olympus has conducted animal testing to simulate general endoscopic operation. Results of this testing confirmed that all thermal damage was limited to the submucosa tissue. These test results were obtained under the condition where the distal end of the endoscope was in physical contact to the mucous membrane using maximum light intensity for time duration*1 indicated in the table. Light intensity
Time duration*1
Maximum light intensity of Near focus mode
40 sec.
Maximum light intensity of Normal focus mode
15 sec.
*1 Defined by Olympus as the worst case under the general endoscopic operation
GIF-EZ1500 OPERATION MANUAL
11
Important Information - Please Read Before Use
Examples of inappropriate handling Details on clinical endoscopic technique are the responsibility of trained specialists. Patient safety in endoscopic examinations and endoscopic treatment can be ensured through appropriate handling by the physician and the medical facility. Examples of inappropriate handling are described below.
WARNING • Over-insufflating the lumen may cause patient pain, injury, bleeding, and/or perforation. • Applying suction with the distal end of the endoscope in prolonged contact with the mucosal surface, with higher suction pressure than required, or with prolonged suction time may cause bleeding and/or lesions. • The endoscope has not been designed for use in retroflexed observation in parts of the body other than the stomach. Performing retroflexed observation in a narrow lumen may make it impossible to straighten the angle of the bending section and/or withdraw the endoscope from the patient. Retroflexed observation in parts of the body other than the stomach should be performed only when its usefulness is determined to be greater than the risk that is posed to the patient. Also, do not operate the endoscope forcibly in retroflexed observation. • Inserting, withdrawing, and using EndoTherapy accessories without a clear endoscopic image may cause patient injury, burns, bleeding, and/or perforation. • Inserting or withdrawing the endoscope, feeding air, applying suction, or operating the bending section without a clear endoscopic image may cause patient injury, bleeding, and/or perforation. • For reasons described below, do not rely on the NBI*1 or RDI*2 observation mode alone for primary detection of lesions or to make a decision regarding any potential diagnostic or therapeutic intervention. − NBI or RDI has not been demonstrated to increase the yield or sensitivity of finding any specific mucosal lesion, including colonic polyps or Barrett’s esophagus. *1 Narrow Band Imaging. For more details, refer to the instruction manual for the video system center CV-1500. *2 Red Dichromatic Imaging. For more details, refer to the instruction manual for the video system center CV-1500.
12
GIF-EZ1500 OPERATION MANUAL
1.1 Checking the package contents list
Chapter 1
1.1
Checking the Package Contents Ch.1
Checking the package contents list
Match all items in the package with the components shown below. Inspect each item for damage. If the endoscope is damaged, a component is missing, or you have any questions, do not use the items; immediately contact Olympus.
Endoscope
GIF-EZ1500
The following accessories are attached to the endoscope. • Suction valve (MH-443) • Air/water valve (MH-438) • Biopsy valve (MB-358) • Auxiliary water inlet cap (MAJ-215)
GIF-EZ1500 OPERATION MANUAL
13
1.1 Checking the package contents list
Items
Ch.1
Single use combination cleaning brush (BW-412T, 3 pcs)
Injection tube (MH-946)
Channel plug (MH-944)
AW channel cleaning adapter (MH-948)
Suction cleaning adapter (MH-856)
Mouthpiece (MB-142, 2 pcs)
Auxiliary water tube (MAJ-855)
ETO cap (MB-156)
Operation manual
Reprocessing manual
NOTE Single use biopsy valve (MAJ-1555) is sold separately.
14
GIF-EZ1500 OPERATION MANUAL
2.1 Nomenclature and functions
Chapter 2
Instrument Nomenclature and Specifications
The instrument nomenclature, functions, and specifications are described in this chapter.
2.1
Nomenclature and functions
GIF-EZ1500 OPERATION MANUAL
Ch.2
15