OLYMPUS
OLYMPUS Scopes Small Intestinal
SIF-Q260 EVIS LUCERA SMALL INTESTINAL VIDEOSCOPE Operation Manual March 2007
Operation Manual
88 Pages
Preview
Page 1
INSTRUCTIONS
EVIS LUCERA SMALL INTESTINAL VIDEOSCOPE
OLYMPUS SIF TYPE Q260
Refer to the endoscope’s companion manual, the “REPROCESSING MANUAL” with your endoscope model listed on the cover, for reprocessing information.
Contents
Contents Symbols...
1
Important Information - Please Read Before Use...
2
Intended use ...
2
Applicability of endoscopy and endoscopic treatment ...
2
Instruction manual...
3
User qualifications ...
3
Instrument compatibility ...
4
Reprocessing before the first use/reprocessing and storage after use ...
4
Spare equipment ...
4
Maintenance management ...
4
Prohibition of improper repair and modification ...
5
Signal words ...
5
Warnings and cautions ...
5
Examples of inappropriate handling...
10
Chapter 1
Checking the Package Contents...
11
Chapter 2
Instrument Nomenclature and Specifications ...
14
2.1
Nomenclature...
14
2.2
Endoscope functions...
16
2.3
Specifications...
18
2.4
Attaching the chain for water-resistant cap (MAJ-1119) ...
21
Preparation and Inspection ...
24
3.1
Preparation of the equipment...
24
3.2
Inspection of the endoscope ...
26
3.3
Preparation and inspection of accessories ...
30
3.4
Attaching accessories to the endoscope ...
33
3.5
Inspection and connection of ancillary equipment ...
36
3.6
Inspection of the endoscopic system ...
39
3.7
Inspection and preparation of the single use splinting tube (ST-SB1) 43
Chapter 3
Chapter 4
Operation ...
47
4.1
Insertion ...
49
4.2
Using EndoTherapy accessories ...
56
4.3
Withdrawal of the endoscope...
62
4.4
Transportation of the endoscope ...
63
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
i
Contents
Chapter 5
ii
Troubleshooting ...
64
5.1
Troubleshooting guide ...
64
5.2
Withdrawal of the endoscope with an abnormality...
68
5.3
Returning the endoscope for repair...
70
Appendix...
71
System chart ...
71
EMC information...
76
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
Symbols
Symbols The meaning of the symbol(s) shown on the package of the components, the back cover of this instruction manual and/or this instrument are as follows:
Refer to instructions.
Caution
Endoscope
TYPE BF applied part
Manufacturer
Authorized representative in the European Community
Lot number
Serial number
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
1
Important Information - Please Read Before Use
Important Information - Please Read Before Use Intended use This instrument has been designed to be used with an Olympus video system center, light source, single use splinting tube, balloon control unit, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps) and other ancillary equipment for endoscopy and endoscopic surgery within the upper digestive tract (including the esophagus, stomach, duodenum, and small intestine) by oral insertion or the lower digestive tract (including the anus, rectum, sigmoid colon, colon, ileocecal valve, and small intestine) by anal insertion. Do not use this instrument for any purpose other than its intended use. Select the endoscope to be used according to the objective of the intended procedure based on the full understanding of the endoscope’s specifications and functionality as described in this instruction manual.
Applicability of endoscopy and endoscopic treatment If there is an official standard on the applicability of endoscopy and endoscopic treatment that is defined by the hospital’s administration or other official institutions such as academic societies on endoscopy, follow that standard. Before starting endoscopy and endoscopic treatment, thoroughly evaluate its properties, purposes, effects, and possible risks (their nature, extent and probability). Perform endoscopy and endoscopic treatment only when its potential benefits are greater than its risks. Fully explain to the patient the potential benefits and risks of the endoscopy and endoscopic treatment as well as any examination/treatment methods that can be performed in its place, and perform the endoscopy and endoscopic treatment only after obtaining the consent of the patient. Even after starting the endoscopy and endoscopic treatment, continue to evaluate the potential benefits and risks, and immediately stop the endoscopy/treatment and take proper measures if the risks to the patient become greater than the potential benefits.
2
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
Important Information - Please Read Before Use
Instruction manual This instruction manual contains essential information on using this instrument safely and effectively. Before use, thoroughly review this manual and the manuals of all equipment that will be used during the procedure and use the equipment as instructed. Note that the complete instruction manual set for this endoscope consists of this manual and the “REPROCESSING MANUAL” with your endoscope model listed on the cover. The reprocessing manual accompanied the endoscope at shipment. Keep this and all related instruction manuals in a safe, accessible location. If you have any questions or comments about any information in this manual, please contact Olympus.
! Terms used in this manual NBI (Narrow Band Imaging) observation: This is a special light observation using the narrow band observation light. WLI (White Light Imaging) observation: This is an observation using the standard RGB illumination.
User qualifications If there is an official standard on user qualifications to perform endoscopy and endoscopic treatment that is defined by the medical administration or other official institutions, such as academic societies on endoscopy, follow that standard. If there is no official qualification standard, the operator of this instrument must be a physician approved by the medical safety manager of the hospital or person in charge of the department (department of internal medicine, etc.). The physician should be capable of safely performing the planned endoscopy and endoscopic treatment following guidelines set by the academic societies on endoscopy, etc., and considering the difficulty of endoscopy and endoscopic treatment. This manual does not explain or discuss endoscopic procedures.
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
3
Important Information - Please Read Before Use
Instrument compatibility Refer to the “System chart” in the Appendix to confirm that this instrument is compatible with the ancillary equipment being used. Using incompatible equipment can result in patient or operator injury and/or equipment damage. This instrument complies with the EMC standard for medical electrical equipment; edition 2 (IEC 60601-1-2: 2001). However, when connected with an instrument that complies with the EMC standard for medical electrical equipment; edition 1 (IEC 60601-1-2: 1993), the whole system complies with edition 1.
Reprocessing before the first use/reprocessing and storage after use This instrument was not cleaned, disinfected, or sterilized before shipment. Before using this instrument for the first time, reprocess it according to the instructions given in the endoscope’s companion manual, the “REPROCESSING MANUAL” with your endoscope model listed on the cover. After using this instrument, reprocess and store it according to the instructions given in the endoscope’s companion reprocessing manual. Improper and/or incomplete reprocessing or storage can present an infection-control risk, cause equipment damage, or reduce performance.
Spare equipment Be sure to prepare another endoscope to avoid interruption of the examination due to equipment failure or malfunction.
Maintenance management The probability of failure of the endoscope and ancillary equipment increases as the number of procedures performed and/or the total operating hours increase. In addition to the inspection before each procedure, the person in charge of medical equipment maintenance in each hospital should inspect the items specified in this manual periodically. An endoscope with a suspected irregularity should not be used, but should be inspected by following Section 5.1, “Troubleshooting guide” on page 64. If the irregularity is still suspected after inspection, contact Olympus.
4
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
Important Information - Please Read Before Use
Prohibition of improper repair and modification This instrument does not contain any user-serviceable parts. Do not disassemble, modify, or attempt to repair it; patient or operator injury and/or equipment damage can result. Equipment that has been disassembled, repaired, altered, changed, or modified by persons other than Olympus’ own authorized service personnel is excluded from Olympus’ limited warranty and is not warranted by Olympus in any manner.
Signal words The following signal words are used throughout this manual:
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.
Indicates additional helpful information.
Warnings and cautions Follow the warnings and cautions given below when handling this instrument. This information is to be supplemented by the warnings and cautions given in each chapter.
•
After using this instrument, reprocess and store it according to the instructions given in the endoscope’s companion manual, the “REPROCESSING MANUAL” with your endoscope model listed on the cover. Using improperly or incompletely reprocessed or stored instruments may cause patient cross-contamination and/or infection.
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
5
Important Information - Please Read Before Use
6
•
Before endoscopy, remove any metallic objects (watch, glasses, necklace, etc.) from the patient. Performing high-frequency cauterization treatment while the patient is wearing metallic objects may cause burns on the patient in areas around the metallic objects.
•
Do not strike, bend, hit, pull, twist, or drop the endoscope’s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector with excessive force. The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations. It could also cause parts of the endoscope to fall off inside the patient.
•
Never perform angulation control forcibly or abruptly. Never forcefully pull, twist, or rotate the angulated bending section. Patient injury, bleeding, and/or perforation can result. It may also become impossible to straighten the bending section during an examination.
•
Never insert or withdraw the endoscope’s insertion section while the bending section is locked in position. Patient injury, bleeding, and/or perforation can result.
•
Never operate the bending section, feed air or perform suction, insert or withdraw the endoscope’s insertion section, or use EndoTherapy accessories without viewing the endoscopic image. Patient injury, bleeding, and/or perforation can result.
•
Never operate the bending section, feed air or perform suction, insert or withdraw the endoscope’s insertion section, or use EndoTherapy accessories while the image is frozen. Patient injury, bleeding, and/or perforation can result.
•
Never insert or withdraw the insertion section abruptly or with excessive force. Patient injury, bleeding, and/or perforation can result.
•
If it is difficult to insert the endoscope, do not forcibly insert the endoscope; stop the endoscopy. Forcible insertion can result in patient injury, bleeding, and/or perforation.
•
Never insert or withdraw the endoscope’s insertion section or use EndoTherapy accessories while the image is magnified. Patient injury, bleeding, and/or perforation can result (when using the image magnification function of video system center).
•
Do not touch the light guide of the endoscope connector immediately after removing it from the light source because it is extremely hot. Operator or patient burns can result.
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
Important Information - Please Read Before Use
•
When the endoscopic image does not appear on the monitor, the CCD may have been damaged. Turn the video system center OFF immediately. Continued power supply in such a case will cause the distal end to become hot and could cause operator and/or patient burns.
•
Do not rely on the NBI observation modality alone for primary detection of lesions or to make a decision regarding any potential diagnostic or therapeutic intervention.
•
When combining the endoscope with a single use splinting tube (ST-SB1), there is a risk of perforation or bleeding due to entangling the mucous membrane, or of the tube separating from the endoscope and remaining inside the body. Before use, be sure to read the instruction manuals for the single use splinting tube (ST-SB1) and the balloon control unit (OBCU) to fully understand the characteristics.
•
When performing a small intestinal endoscopic procedure, always be sure to observe the condition of the endoscope and the single use splinting tube under X-ray image. Performing the procedure without viewing the X-ray image may result in patient injury, bleeding, and/or perforation.
•
In normal operation, use the balloon control unit to inflate/deflate the balloon. Otherwise, the balloon may be damaged and then patient injury could result. Before use, be sure to read the instruction manual for the balloon control unit to fully understand its characteristics.
•
Patient debris or fluids may regurgitate through the gap between the endoscope’s insertion section and the scope inlet on the grip of the single use splinting tube. To prevent contamination, take proper measures (e.g., spread a moisture-resistant sheet). And always wear personal protective equipment. Otherwise, it can present an infection-control risk.
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
7
Important Information - Please Read Before Use
8
•
If you feel a resistance while operating the single use splinting tube after balloon deflation, check under X-ray image that the balloon has deflated to the intestinal shape. Check also that the air flow tube of the single use splinting tube is not obstructed or crushed. If the air flow tube is not obstructed or crushed and the resistance persists, disconnect the air flow tube from the single use splinting tube. Connect a syringe and a 3-way stopcock to the single use splinting tube and use it to forcefully suction air from the balloon. Do not continue the procedure if resistance is felt. Doing so may result in patient injury, bleeding, and/or perforation.
•
Do not withdraw the distal end of the endoscope into the single use splinting tube during the endoscopic procedures. Otherwise, the single use splinting tube may be crushed and cannot be withdrawn from the patient. If using the endoscope with a distal attachment, the distal attachment may be dropped off inside the patient body.
•
Do not pull the universal cord during an examination. The endoscope connector will be pulled out from the output socket of the light source and the endoscopic image will not be visible.
•
Do not coil the insertion tube or universal cord with a diameter of less than 12 cm. Equipment damage can result.
•
Do not attempt to bend the endoscope’s insertion section with excessive force. Otherwise, the insertion section may be damaged.
•
Do not touch the electrical contacts inside the electrical connector. CCD damage may result.
•
Do not apply shock to the distal end of the insertion section, particularly the objective lens surface at the distal end. Visual abnormalities may result.
•
Do not twist or bend the bending section with your hands. Equipment damage may result.
•
Do not squeeze the bending section forcefully. The covering of the bending section may stretch or break and cause water leaks.
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
Important Information - Please Read Before Use
•
Turn the video system center OFF before connecting or disconnecting the videoscope cable from the electrical connector on the endoscope. Turn the video system center ON or OFF only when the videoscope cable is connected to both the video system center and the electrical connector on the endoscope. Failure to do so can result in equipment damage, including destruction of the CCD.
•
The endoscope’s remote switches cannot be removed from the control section. Pressing, pulling, or twisting them with excessive force can break the switches and/or cause water leaks.
•
If remote switch 1 does not return to the OFF position after being pressed strongly from the side, gently pull the switch upwards to return it to the OFF position.
•
Do not hit or bend the electrical contacts on the endoscope connector. The connection to the light source may be impaired and faulty contact can result.
•
Do not perform angulation control when the distal end of the single use splinting tube is over the endoscope’s bending section. Otherwise, the bending section or the single use splinting tube may be damaged.
•
Electromagnetic interference may occur on this instrument near equipment marked with the following symbol or other portable and mobile RF (Radio Frequency) communications equipment such as cellular phones. If electromagnetic interference occurs, mitigation measures may be necessary, such as reorienting or relocating this instrument, or shielding the location.
•
To check the electromagnetic interference from other equipment (any equipment other than this instrument or the components that constitute this system), the system should be observed to verify its normal operation in the configuration in which it will be used.
•
This endoscope contains a memory chip that stores information about the endoscope and communicates this information to the video system center CV-260SL and CV-260.
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
9
Important Information - Please Read Before Use
•
This endoscope can be inserted far into the small intestine when it is used in combination with a single use splinting tube.
Examples of inappropriate handling Details on clinical endoscopic technique are the responsibility of trained specialists. Patient safety in endoscopic examinations and endoscopic treatment can be ensured through appropriate handling by the physician and the medical facility. Examples of inappropriate handling are described below.
10
•
Over-insufflating the lumen may cause patient pain, injury, bleeding, and/or perforation.
•
Applying suction with the distal end in prolonged contact with the mucosal surface, with higher suction pressure than required, or with prolonged suction time may cause bleeding and/or lesions.
•
The endoscope has not been designed for use in retroflexed observation in parts of the body other than the stomach. Performing retroflexed observation in a narrow lumen may make it impossible to straighten the angle of the bending section and/or withdraw the endoscope from the patient. Retroflexed observation in parts of the body other than the stomach should be performed only when the usefulness of doing so is determined to be greater than the risk that is posed to the patient.
•
Inserting, withdrawing, and using EndoTherapy accessories without a clear endoscopic image may cause patient injury, burns, bleeding and/or perforation.
•
Inserting or withdrawing the endoscope, feeding air, applying suction, or operating the bending section without a clear endoscopic image may cause patient injury, bleeding, and/or perforation.
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
Chapter 1 Checking the Package Contents
Chapter 1 Checking the Package Contents Match all items in the package with the components shown below. Inspect each item for damage. If the instrument is damaged, a component is missing, or you have any questions, do not use the instrument; immediately contact Olympus. This instrument was not disinfected or sterilized before shipment. Before using this instrument for the first time, reprocess it according to the instructions described in the endoscope’s companion manual, the “REPROCESSING MANUAL” with your endoscope model listed on the cover.
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
11
Chapter 1 Checking the Package Contents
Endoscope
Channel cleaning brush (BW-9Y)
Water-resistant cap (MH-553)
Injection tube (MH-946)
Channel-opening cleaning brush (MH-507)
AW channel cleaning adapter (MH-948)
Suction valve (MH-443, 2 pcs)
Air/water valve (MH-438, 2 pcs) Channel plug (MH-944)
Biopsy valve (MB-358, 10 pcs)
Suction cleaning adapter (MH-856)
12
Mouthpiece (MB-142, 2 pcs)
Operation manual
Chain for water-resistant cap (MAJ-1119)
Reprocessing manual
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
Chapter 1 Checking the Package Contents
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
13
Chapter 2 Instrument Nomenclature and Specifications
Chapter 2 Instrument Nomenclature and Specifications 2.1
Nomenclature ! SIF-Q260
Universal cord
1. Suction connector 5. Electrical connector
2. S-cord connector mount
Air pipe
3. Air supply connector 3. Water supply connector Product name (model) and serial number Light guide Contact pins 4. Endoscope connector
14
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
Chapter 2 Instrument Nomenclature and Specifications
9. Air/water valve (MH-438) 8. Suction valve (MH-443) 7. UP/DOWN angulation lock Suction cylinder 6. UP/DOWN angulation control knob
16. RIGHT/LEFT angulation control knob Control section
Air/water cylinder
15. RIGHT/LEFT angulation lock
Grip section 14. Color code
Biopsy valve (MB-358)
Boot
10. Instrument channel Instrument channel port
11. Insertion section limit mark Insertion section
13. Remote switches 1 to 4
Distal end Insertion tube
12. Bending section
Top view
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
15
Chapter 2 Instrument Nomenclature and Specifications
2.2
Endoscope functions 1. Suction connector Connects the endoscope to the suction tube of the suction pump. 2. S-cord connector mount Connects the endoscope with the Olympus electrosurgical unit via the S-cord. The S-cord conducts leakage current from the endoscope to the electrosurgical unit. To connect the S-cord, refer to the instruction manual for the electrosurgical unit. Connect the fitting of the chain for water-resistant cap to this mount as required (see Section 2.4 on page 21). 3. Water supply connector and air supply connector Connects the endoscope to the water container via the water container tube in order to supply water to the distal end of the endoscope. 4. Endoscope connector Connects the endoscope to the output socket of the light source and transmits light from the light source to the endoscope. 5. Electrical connector Connects the endoscope to the video system center via the videoscope cable. The endoscope contains a memory chip that stores information about the endoscope and communicates this information to the video system center CV-260SL and CV-260. For more details, refer to the instruction manual of the CV-260SL or CV-260. 6. UP/DOWN angulation control knob When this knob is turned in the “ U” direction, the bending section moves UP; when the knob is turned in the “D ” direction, the bending section moves DOWN. 7. UP/DOWN angulation lock Moving this lock in the “F ” direction frees angulation. Moving the lock in the opposite direction locks the bending section at any desired position. 8. Suction valve (MH-443) This valve is depressed to activate suction. The valve is used to remove any fluid, debris, flatus, or air from the patient. 9. Air/water valve (MH-438) The hole in this valve is covered to insufflate air and the valve is depressed to feed water for lens washing. It also can be used to feed air for removing any fluid or debris adhering to the objective lens. 10. Instrument channel The instrument channel functions as: ! Channel for the insertion of EndoTherapy accessories ! Suction channel ! Fluid feed channel (from a syringe via the biopsy valve)
16
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
Chapter 2 Instrument Nomenclature and Specifications
11. Insertion section limit mark This mark shows the maximum point to which the endoscope may be inserted into the patient’s body. 12. Bending section This section moves the distal end of the endoscope when the UP/DOWN and RIGHT/LEFT angulation control knobs are operated. 13. Remote switches 1 to 4 The functions of remote switches 1 to 4 can be selected on the video system center. Refer to the instruction manual for the video system center when setting these functions. 14. Color code (yellow) The endoscope can be used with EndoTherapy accessories that have the same color code. For more information on combining the endoscope with particular EndoTherapy accessories, refer to the “System chart” in the Appendix and the instruction manuals of the accessories. 15. RIGHT/LEFT angulation lock Turning this lock in the “F ” direction frees angulation. Turning the lock in the opposite direction locks the bending section at any desired position. 16. RIGHT/LEFT angulation control knob When this knob is turned in the “R ” direction, the bending section moves RIGHT; when the knob is turned in the “ L” direction, the bending section moves LEFT.
EVIS LUCERA SIF TYPE Q260 OPERATION MANUAL
17