OLYMPUS
OLYMPUS Video and Image Management Systems
OTV-S7Pro VISERA Pro VIDEO SYSTEM CENTER Instructions Aug 2012
Instructions
286 Pages
Preview
Page 1
INSTRUCTIONS
VISERA Pro VIDEO SYSTEM CENTER
OLYMPUS OTV-S7Pro
Contents
Contents Labels and Symbols ...
1
Important Information - Please Read Before Use...
3
Intended use ...
3
Applicability of endoscopy and endoscopic treatment ...
3
Instruction manual ...
3
User qualifications ...
5
Instrument compatibility ...
6
Repair and modification ...
6
Signal words ...
6
Dangers, warnings and cautions...
7
Cardiac applications...
11
Summary of Equipment Functions ...
12
Chapter 1
Checking the Package Contents...
14
Chapter 2
Nomenclature and Functions ...
15
2.1
Symbols and descriptions ...
15
2.2
Front panel...
16
2.3
Rear panel ...
20
2.4
Keyboard...
23
2.5
Side panels ...
27
2.6
Setup of screen options ...
28
2.7
Monitor ...
29
2.8
Pointer...
32
Inspection ...
33
3.1
Inspection of the power supply ...
34
3.2
Inspection of the examination light...
35
3.3
Video monitor display inspection ...
35
3.4
Inspection of the automatic brightness control function ...
36
3.5
Inspection of the monitor display ...
37
3.6
Inspection of the freeze function ...
38
3.7
Inspection of the release function ...
38
3.8
Inspection of the PinP (Picture in Picture) function...
38
3.9
Inspection of the orientation function ...
38
3.10 Inspection of the special light observation function...
39
3.11 Inspection of the scope switches and foot switches ...
39
Chapter 3
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
i
Contents
3.12 Power OFF...
Chapter 4
Operation...
40
4.1
Operation flow ...
43
4.2
Connection of an endoscope ...
45
4.3
Turning the video system center ON...
49
4.4
Recall of user preset data ...
50
4.5
White balance adjustment...
51
4.6
Patient data ...
56
4.7
Observation of the endoscopic image...
58
4.8
Recording of the endoscopic image...
58
4.9
Termination of the operation ...
59
Functions...
60
Front panel ...
60
Image source buttons... PinP (Picture in Picture) display ... Image enhancement mode (ENH.)... Iris mode... White balance... Brightness adjustment (Exposure) ... STOP button and PC card indicator ... PC card slot and eject button ... RESET button ...
60 62 65 67 68 69 75 76 80
Keyboard...
81
Domepoint ... Clearing characters from the screen (“F1”) ... System setup (“Shift” + “F1”) ... Scope information (“F2”) ... User preset (“Shift” + “F2”) ... Cursor (“F3”)... Patient data (“Shift” + “F3”)... Freeze mode (“F4”) ... Browse (“Shift” + “F4”)... Stopwatch (“F5”)... Auto gain control (AGC) (“F6”) ... Contrast mode (“Shift” + “F6”) ... Image zooming (“F7”)... Color bar (“Shift” + “F7”) ... Image size (“F8”) ... Printer lock (“Shift” + “F8”)... Image enhancement (“F9”)... White balance adjustment (“Shift” + “F9”) ... Color tone adjustment (“COLOR”)... Freeze (“FREEZE”) ... Release (“RELEASE”)... Arrow pointer (“Shift” + arrow keys and domepoint)... Color mode (“Shift” + “Alt” + “1”, “2”, “3”, “4”, “5”) ...
81 82 84 84 85 85 86 86 87 88 89 90 91 93 94 95 96 97 98 99 101 102 104
Chapter 5 5.1
5.2
ii
39
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
Contents
Ending examination (“EXAM END”)...
105
Image recording and playback (PC card) ...
106
Storage level of the PC card... Recording the frozen image on a PC card... PC card menu... Basic operation of the PC card menu ... Formatting of the PC card... Playback images from the PC card ... Deleting images from a PC card... Deleting folder from PC card ... Annotation of images ... Playback image annotation... Playback the images using the personal computer ... Image files and folders...
106 108 110 111 114 115 117 118 119 122 124 125
Image recording and playback (other than PC card) ...
127
Image filing system ... Videocassette recorder (VCR)...
127 129
5.5
Printing images ...
131
Video printer ...
131
5.6
Pre-entry of patient data ...
136
Basic operation in the patient menu ... Entering new patient data ... Displaying patient data ... Editing previously entered patient data... Deleting previously entered patient data ... Clearing all patient data previously entered... Recording patient data into PC card... Loading patient data from PC card ...
136 137 140 141 142 143 144 146
Scope information ...
148
Displaying and entering scope information...
149
Special light observation ...
151
NBI (Narrow Band Imaging)... PDD (Photo Dynamic Diagnosis)...
151 153
5.3
5.4
5.7 5.8
Chapter 6
Fuse Replacement... 155
Chapter 7
Care, Storage and Disposal... 157
7.1
Care ...
157
7.2
Storage ...
158
7.3
Disposal ...
158
Chapter 8
Installation and Connection ... 159
8.1
Installation work flow...
160
8.2
Installation of the equipment ...
161
8.3
Light source ...
164
8.4
Monitor ...
169
8.5
Keyboard...
177
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
iii
Contents
8.6
Videocassette recorder (VCR) ...
178
8.7
Video printer...
180
8.8
Foot switch ...
182
8.9
S6 camera head connecting unit for S7V...
183
8.10 Connection to the AC mains power supply ...
185
Chapter 9
Function Setup... 189
9.1
Turning power ON...
189
9.2
System setup ...
190
Basic operation of the system setup ... System ... Printer... Image filing system... Monitor ... Videocassette recorder ... Saving the system setup ... Summary of settings...
190 193 198 202 204 207 209 210
User preset ...
212
Basic operation of the user preset... Scope switch and foot switch ... Release function... Recording format for PC card... Freeze function... Image enhancement (normal and PDD observation) ... Color mode... Image size ... Iris... Auto gain control (AGC) ... Contrast... Exposure area ... Electronic shutter... Patient data display ... Release index time... Indication of the special light observation... Monitor orientation function ... PinP (Picture in Picture) function... Special light observation ... Image enhancement (NBI observation)... Saving the user preset ... Resetting the user preset data to the factory defaults... Deleting user preset data ... Summary of settings...
212 215 219 220 221 222 224 225 228 230 231 232 233 234 235 237 238 240 243 245 246 247 248 250
9.3
Chapter 10 Troubleshooting ... 252
iv
10.1 Troubleshooting guide ...
252
10.2 Returning the video system center for repair ...
262
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
Contents
Appendix ... 263 System chart ...
263
Transportation, storage, and operation environment ...
267
Specifications ...
267
EMC information ...
273
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
v
Contents
vi
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
Labels and Symbols
Labels and Symbols Safety-related labels and symbols are attached on the locations shown below. If labels or symbols are missing or illegible, contact OLYMPUS.
Rear panel Crossed-out wheeled bin’ symbol CE marking
Fuse rating Caution, refer to instructions.
FUSES T5AL250V
Potential equalization terminal Manufacturer name Electric rating The product name, rated voltage and frequency are shown. VISERA Pro VIDEO SYSTEM CENTER MODEL OLYMPUS OTV-S7Pro INPUT 220-240V 50/60Hz 150VA
Serial number plate
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
1
Labels and Symbols
Back cover of this instruction manual
Manufacturer
Authorized representative in the European Community
2
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
Important Information - Please Read Before Use
Important Information - Please Read Before Use Intended use This video system center has been designed to be used with Olympus camera heads, endoscopes, light sources, monitors, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis, treatment and video observation. Do not use this video system center for any purpose other than its intended use.
Applicability of endoscopy and endoscopic treatment If there is an official standard on the applicability of endoscopy and endoscopic treatment that is defined by the hospital’s administration or other official institutions such as academic societies on endoscopy, follow that standard. Before starting endoscopy and endoscopic treatment, thoroughly evaluate its properties, purposes, effects, and possible risk (their natures, extent and probability). Perform endoscopy and endoscopic treatment only when its potential benefits are greater than its risks. Fully explain to the patient the potential benefits and risks of the endoscopy and endoscopic treatment as well as any examination/treatment methods that can be performed in its place, and perform the endoscopy and endoscopic treatment only after obtaining the consent of the patient. Even after starting the endoscopy and endoscopic treatment, continue to evaluate the potential benefits and risks, and immediately stop the endoscopy/treatment and take proper measures if the risks to the patient become greater than the potential benefits.
Instruction manual This instruction manual contains essential information on using this video system center safely and effectively. Before use, thoroughly review this manual and the manuals of all equipment which will be used during the procedure and use the equipment as instructed. Keep this and all related instruction manuals in a safe, accessible location. If you have any questions or comments about any information in this manual, please contact Olympus. VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
3
Important Information - Please Read Before Use
Terms used in this manual Light source: The light source provides light and electrical signals to the endoscope. It also provides electrical signals to the video system center. Video printer: The video printer is a device that prints the frozen video image. Wall mains outlet: The wall mains outlet is a wall AC mains power outlet socket having the exclusive terminal for grounding. Isolation transformer: The isolation transformer is a safety device that is used to isolate non-insulated equipment with potentially high leakage currents to decrease the possibility of electric shock. Image sensor (CCD): Image sensor (CCD) is a device that converts light into electrical signals. Automatic brightness control: The automatic brightness control automatically adjusts the intensity of the light emitted from the light source so that the endoscopic image will be maintained at constant brightness even if the distance between the distal end of the endoscope’s insertion tube and the subject changes. Electronic shutter: Electronic shutter is the device that changes the exposure time of the image sensor so that the brightness of an image is maintained on the video monitor. Color adjustment: Color adjustment adjusts the color balance on the video monitor. Iris: The iris function is used to electrically measure the brightness of an endoscopic image to obtain a control signal for the purpose of automatic light adjustment. Freeze: The freeze function creates a stationary view of the moving image. Release: The release function is used to capture and record an endoscopic image. Edge enhancement: Edge enhancement is an image processing technique that electronically sharpens the edges of an image.
4
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
Important Information - Please Read Before Use
Structure enhancement: Structure enhancement is an image processing technique that electronically emphasizes the detailed patterns and edges of an image to increase sharpness. PinP (Picture in Picture): PinP function displays both the image of the endoscopic live image and the image of an external device on the monitor simultaneously. PC card: A digital medium for storage of images, etc. Wash out: Wash out is the inability to see details in the endoscopic image due to excessive brightness. SDTV: Standard Definition Television. It is the format used in standard video systems. HDTV: High Definition Television. This is a format for high resolution video transmission featuring higher definition than the standard SDTV format. Special light observation: This is an observation using filtered light. NBI (Narrow Band Imaging): This is one of the special light observations using the narrow band observation light. PDD (Photo Dynamic Diagnoses) observation This is one of the special light observations using the exciting light.
User qualifications If there is an official standard on user qualifications to perform endoscopy and endoscopic treatment that is defined by the medical administration or other official institutions, such as academic societies on endoscopy, follow that standard. If there is no official qualification standard, the operator of this instrument must be a physician approved by the medical safety manager of the hospital or person in charge of the department. The physician should be capable of safely performing the planned endoscopy and endoscopic treatment following guidelines set by the academic societies on endoscopy, etc., and considering the difficulty of endoscopy and endoscopic treatment. This manual does not explain or discuss endoscopic procedures.
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
5
Important Information - Please Read Before Use
Instrument compatibility Refer to the “System chart” in the Appendix to confirm that this video system center is compatible with the ancillary equipment being used. Using incompatible equipment can result in patient injury or equipment damage and makes it impossible to obtain the expected functionality. This instrument complies with the EMC standard for medical electrical equipment; edition 2 (IEC 60601-1-2: 2001). However, when connecting to an instrument that complies with the EMC standard for medical electrical equipment; edition 1 (IEC 60601-1-2: 1993), the whole system complies with edition 1.
Repair and modification This video system center does not contain any user-serviceable parts. Do not disassemble, modify, or attempt to repair it; patient or operator injury, equipment damage, and/or the impossibility to obtain the expected functionality can result. Some problems that appear to be malfunctions may be correctable by referring to Chapter 10, “Troubleshooting”. If the problem cannot be resolved using the information in Chapter 10, contact Olympus. This instrument is to be repaired by Olympus technicians only.
Signal words The following signal words are used throughout this manual:
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.
Indicates additional helpful information.
6
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
Important Information - Please Read Before Use
Dangers, warnings and cautions Follow the dangers and cautions given below when handling this video system center. This information is to be supplemented by the dangers and cautions given in each chapter.
•
Strictly observe the following precautions. Failure to do so may place the patient and medical personnel in danger of electric shock. When this video system center is used to examine a patient, do not allow metal parts of the endoscope or its accessories to touch metal parts of other system components. Such contact may cause unintended current flow to the patient. Keep fluids away from all electrical equipment. If fluids are spilled on or into the unit, stop operation of the video system center immediately and contact Olympus. Do not prepare, inspect or use this video system center with wet hands.
•
Never install and operate the video system center in locations where: The concentration of oxygen is high; Oxidizing agents (such as nitrous oxide (N2O)) are present in the atmosphere; Flammable gases are present in the atmosphere; Flammable liquids are near. Otherwise, explosion or fire may result because this video system center is not explosion-proof.
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
7
Important Information - Please Read Before Use
•
In case of instrument failure or malfunction, always keep another video system center in the room ready for use.
•
Never insert anything into the ventilation grills of the video system center. It can cause an electric shock and/or fire.
•
Although the illumination light emitted from the endoscope’s distal end is required for endoscopic observation and treatment, it may also cause alteration of living tissues such as protein denaturation of liver tissue and perforation of the intestines by inappropriate use. Observe the following warnings on the illumination. Always set the minimum required brightness. The brightness of the image on a video monitor may differ from the actual brightness at the distal end of an endoscope. Especially in combination with endoscopes using an electronic shutter function, pay attention to the brightness level setting of the light source. When this instrument is used with a light source compatible with automatic brightness control function, be sure to use this function. The automatic brightness control function can keep the illumination light properly. Refer to the instruction manual of the light source for details. Do not continue observation in the proximity to tissue or keep the distal end of the endoscope in contact with living tissue for a long time. It may cause patient burns. When discontinuing the use of the endoscope, be sure to turn the light source OFF.
8
•
This product may interfere with other medical electronic equipment used in combination with it. Before use, refer to the Appendix to confirm the compatibility of this instrument with all equipment to be used.
•
Do not use this product in any place where it may be subject to strong electromagnetic radiation (for example, in the vicinity of a microwave therapeutic device, MRI, wireless set, short-wave therapeutic device, cellular/portable phone, etc.). This may impair the performance of the product.
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
Important Information - Please Read Before Use
•
If the endoscopic image dims during use, blood, mucus or debris may adhere to the light guide on the distal end of the endoscope. Carefully withdraw the endoscope from the patient and remove the blood or mucus in order to obtain optimum illumination and to ensure the safety of the examination. If you continue to use the endoscope in such a condition, the distal end temperature may rise and cause mucosal burns. It may also cause patient and/or operator injury.
•
Do not rely on the special light observation method alone for primary detection of lesions or for a decision regarding any potential diagnostic or therapeutic intervention.
•
For reasons described below, do not rely on the NBI imaging modality alone for primary detection of lesions or to make a decision regarding any potential diagnostic or therapeutic intervention. It has not been demonstrated to aid in differentiating establishing the presence or absence of dysplasia or neoplastic changes within mucosa or mucosal lesions.
•
To display endoscopic images, connect the output terminal of the video system center directly to the monitor. Do not make the connection via any ancillary equipment. Images may disappear during observation depending on the condition of the ancillary equipment.
•
Do not use a pointed or hard object to press the buttons on the front panel and/or keyboard. This may damage the buttons.
•
Do not touch the electrical contacts inside the video system center’s connectors.
•
Do not apply excessive force to this video system center and/or other instruments connected. Otherwise, damage and/or malfunction can occur.
•
Do not connect or disconnect the endoscope connector while this video system center is turned ON. Connecting or disconnecting the endoscope while this video system center is ON may destroy the CCD. Turn the video system center OFF before connecting or disconnecting the endoscope.
•
Clean and vacuum dust the ventilation grills using a vacuum cleaner, when necessary. Otherwise, the video system center may break down and gets damaged from overheating.
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
9
Important Information - Please Read Before Use
•
Be sure that this instrument is not used adjacent to or stacked with other equipment (other than the components of this instrument or system) to avoid electromagnetic interference.
•
Electromagnetic interference may occur to this instrument when it is placed near equipment marked with the following symbol or other portable and mobile RF communications equipment such as cellular phones. If radio interference occurs, mitigation measures may be necessary, such as reorienting or relocating this instrument or shielding the location.
As defined by the international safety standard (IEC 60601-1), medical electrical equipment is classified into the following types: TYPE CF applied part (the instrument can safely be applied to any part of the body, including the heart), and TYPE B/BF applied part (the instrument can safely be applied to any organ except the heart). The part of the body that an endoscope or electrosurgical accessory can safely be applied to depends on the classification of the equipment to which the instruments are connected. Before beginning the procedure, check the current leakage classification type of each instrument to be used for the procedure. Classification types are clearly specified in the instruments’ instruction manuals. Symbol
Classification
TYPE CF applied part
TYPE BF applied part
TYPE B applied part
10
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
Important Information - Please Read Before Use
Cardiac applications •
Use only the devices listed in the “System chart” in the Appendix for endoscopic observation or treatment of the heart or areas near the heart. Other combinations of equipment may cause ventricular fibrillation or seriously affect the cardiac function of the patient.
•
For cardiac applications, never support the endoscope with a metal surgical arm which is not electrically isolated from the ground. If not isolated, the endoscope will be connected to the ground through the surgical arm and bed, and will conduct unexpected leakage current which may seriously affect the cardiac function of the patient.
•
The use of medical devices not specifically designed for cardiac applications may cause ventricular fibrillation or seriously affect the cardiac function of the patient. As specified by the international standard IEC 60601-1, any applied part used for observation or treatment of the heart or areas near the heart must meet “TYPE CF applied part” requirements for low electrical leakage current. When using endoscopes for endoscopic cardiac applications, the applied part requirements include all devices directly connected to the endoscope, such as the light guide cable, camera head and telescope holder. Each of these devices must individually meet the “TYPE CF applied part” requirements for leakage current limits if they are to be used for cardiac applications.
•
The Olympus light guide cables and camera heads listed in the “System chart” in the Appendix (TYPE CF applied part) which are suitable for cardiac applications bear a mark.
•
The Olympus surgical holder for telescope (SH-1) has an electrically isolated arm structure which isolates the endoscope from the ground. This design makes the SH-1 suitable for cardiac applications.
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
11
Summary of Equipment Functions
Summary of Equipment Functions This instrument is a system controller of the endoscopic image observation system that displays, records, and prints the endoscopic images. Some of the functions of this instrument described below are enabled only when the required equipment is connected to this instrument. For more details, refer to the instruction manuals for this instrument and the other instruments connected.
Displaying the endoscopic images on the monitor •
The endoscopic image and other external images can be displayed on the same monitor at the same time (PinP function). “PinP (Picture in Picture) display” on page 62
•
Either a standard-definition (SDTV) monitor or high-definition (HDTV) monitor can be used.
Special light observations Endoscopic observation using filtered light or exciting light is available. Section 5.8, “Special light observation” on page 151
Adjusting the endoscopic images Images can be adjusted to enable clear and convenient observation.
12
•
Adjustment of the image color “Color tone adjustment (“COLOR”)” on page 98
•
Adjustment of the image brightness “Brightness adjustment (Exposure)” on page 69
•
Changing the iris mode “Iris mode” on page 67
•
Changing the contrast mode “Contrast mode (“Shift” + “F6”)” on page 90
•
Enhancement of edge lines and patterns of the images “Image enhancement mode (ENH.)” on page 65
•
Changing the image size “Image size (“F8”)” on page 94
•
Enlargement of the images “Image zooming (“F7”)” on page 91
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
Summary of Equipment Functions
Entering patient data •
The patient data such as name, sex, etc. can be entered and displayed on the monitor with the endoscopic live image. Section 4.6, “Patient data” on page 56 and Section 5.6, “Pre-entry of patient data” on page 136
•
Up to 40 sets of patient data can be stored on the PC card. These patient data can be copied to the other OTV-S7Pro. “Recording patient data into PC card” on page 144
Customizing the operations Up to 20 scope switch settings and other functions such as iris mode, image enhancement, etc., can be stored. Section 9.3, “User preset” on page 212
Recording images •
The endoscopic image can be recorded on the PC card. Section 5.3, “Image recording and playback (PC card)” on page 106
•
The endoscopic image can be recorded on the image-recording device connected to this instrument, and the recorded images can be played back. Section 5.4, “Image recording and playback (other than PC card)” on page 127
•
The endoscopic image can be printed from the printer connected to this instrument. Section 5.5, “Printing images” on page 131
Operation of ancillary equipment •
Videocassette recorder “Videocassette recorder (VCR)” on page 129
•
Video printer Section 5.5, “Printing images” on page 131
•
Image filing system “Image filing system” on page 127
VISERA Pro VIDEO SYSTEM CENTER OTV-S7Pro
13