Bougie Tube Set Updated Guide
1 Page
Preview
Page 1
28th June 2017
To whom it may concern;
Sterilisation of Olympus Bougie Tube Set (WA33027A) with an autoclave Please follow the instructions in the IFU for WA33027A. This product is compatible with the reprocessing methods listed in the table. For details of the procedures, refer to the Olympus Endoscopy System Guide. + compatible − incompatible
When using an Autoclave to steam sterilise the equipment following cleaning please sterilise according to the Australian standards and your hospital protocol. As a guide the following settings should be utilised: – Temperature: 134°C – Steriliser pressure: 203kPA – Sterilisation time = 3.5 - 4 minutes – Dry time = as per hospital/machine protocol
Regards
Melanie O’Farrell Product Manager – Urology & Gynaecology