Distal-end flushing adaptor Guide
2 Pages

Preview
Page 1
How to use the distal-end flushing adaptor MAJ-2319 for EVIS EXERA III Duodenovideoscope TJF-Q190V WARNING
This quick reference guide is a summary of the steps necessary to properly use your MAJ-2319. Follow the detailed steps instructed in the latest REPROCESSING MANUAL for TJF-Q190V
Attaching & detaching MAJ-2319 to the distal-end of the endoscope*
MAJ-2319
Attaching Green flushing port
1
White flushing port Opening
The instrument channel outlet of the distal end is opened opened. (the elevator control lever is in the opposite direction of the “U” direction). direction)
2 Shaft Clip
Forceps elevator Instrument channel outlet
White cover
Hold the MAJ-2319. Then attach the clip of the green cover to the shaft of the white cover.
Green cover
3
4
5
Put the distal-end of the endoscope inside the green cover with the back side surface viewable.
Shut the white cover and the green cover with a snap.
Gently pull the endoscope to ensure that the distal-end stays within the MAJ-2319 and does not come out.
Detaching
Distal end
Open the white cover and the green cover to detach MAJ-2319 from the endoscope. Then separate the white cover and the green cover. * All process to be managed with all instruments immersed in the appropriate liquid
Manually cleaning process with MAJ-2319
1
2
Detergent solution White flushing port
3
Water White flushing port
White flushing port Distal end
1
Flush the distal end with detergent solution
Detergent solution
180 ml
2
Remove detergent solution from the distal end
Water
90 ml
3
Remove detergent solution from the distal-end flushing adaptor
Air
30 ml
20m
l
Procedure
(from each port)
l
Total flush volume
10m
Flush content
30m
l
Syringe
(Repeat the same procedure from the green flushing port in the all procedure
)
While immersing a 30 ml syringe completely in the appropriate liquid, slowly fill it with the appropriate liquid, attach it to the white flushing port, and forcefully flush the distal-end with 30 ml of the liquid. Then with immersing the syringe attached to the port in the appropriate liquid completely, repeat pumping several times (until the volume reaches the specified number) Cover the white cover and the green cover of MAJ-2319 in clean lint-free cloths to prevent splashing from the openings of the covers. Fill the syringe with air, attach it to the white flushing port, and flush the adaptor with 30 ml of air.
Manually disinfecting process with MAJ-2319
1 Disinfectant solution
White flushing port
2
3
4
Disinfectant solution
Water White flushing port White flushing port
White flushing port
180 ml
2
Remove disinfectant solution from the distal-end flushing adaptor
Air
30 ml
3
Flush the distal-end flushing adaptor with water
Water
30 ml
4
Remove rinse water from the distal-end flushing adaptor
Air
30 ml
l
Disinfectant solution
20m
1
Flush the distal end with disinfectant solution
Procedure
(from each port)
l
Total flush volume
10m
Flush content
30m
l
Distal end
(Repeat the same procedure from the green flushing port in the all procedure
)
While immersing a 30 ml syringe completely in the appropriate liquid, slowly fill it with the appropriate liquid, attach it to the white flushing port, and forcefully flush the distal-end with 30 ml of the liquid. Then with immersing the syringe attached to the port in the appropriate liquid completely, repeat pumping several times (until the volume reaches the specified number) Cover the white cover and the green cover of MAJ-2319 in clean lint-free cloths to prevent splashing from the openings of the covers. Fill the syringe with air, attach it to the white flushing port, and flush the adaptor with 30 ml of air.
Alcohol flushing process with MAJ-2319
1
2
White flushing port
(from each port)
20m
l
Procedure
l
Total flush volume
10m
Flush content
30m
l
White flushing port
(Repeat the same procedure from the green flushing port in the all procedure
)
1
Flush the distal-end flushing adaptor with alcohol
Alcohol
30 ml
Cover the white cover and the green cover of MAJ-2319 in sterile lint-free cloths to prevent splashing from the openings of the covers. Fill the syringe with the alcohol, attach it to the white flushing port, and forcefully flush the adaptor with 30 ml of the alcohol.
2
Remove alcohol from the distal-end flushing adaptor
Air
30 ml
Fill the syringe with air, attach it to the white flushing port, and flush the adaptor with 30 ml of air. Confirm that no alcohol exits from both the openings of the white cover and the green cover of the adaptor during the flush.
For complete product details see Instructions for Use.