Forehead Thermometer 乱ᓣԧ⏽䅵 ൟো Model MC-720
INSTRUCTION MANUAL Փ⫼䇈ᯢк
Thank you very much for purchasing the Forehead Thermometer. Be sure to read this Instruction Manual thoroughly before use. Keep this Instruction Manual at hand for future reference. 䴲ᐌᛳ䇶ᙼ䌁фᴀ乱ᓣԧ⏽䅵DŽ Փ⫼ࠡ䇋䆺㒚䯙䇏ᴀ䇈ᯢкDŽ 䇋ֱᄬᴀ䇈ᯢкҹৢ⫼П䳔DŽ
MC-720-AP_A_M.book Page 1 Friday, May 22, 2015 11:01 AM
Contents Thank you for purchasing the OMRON Forehead Thermometer MC-720.
Intended Use:
The OMRON Forehead Thermometer MC-720 offers safe, comfortable and quick temperature measurement from the forehead. It is mainly designed for household use. Important Safety Information ...1 1. Overview ...3 2. Preparation ...4 2.1 Removing the Insulating Tape ...4 2.2 Switching between °C and °F ...4 2.3 Setting the Buzzer ...5 3. Using the Unit ...6 3.1 Taking a Reading ...6 3.2 Using the Memory Function ...10
4. Troubleshooting and Maintenance ... 11 4.1 The Icons and Error Messages ... 11 4.2 Questions and Answers ... 13 4.3 Maintenance ... 14 4.4 Replacing the Battery ... 15 5. Technical Data ... 16 6. Some Useful Information ... 19
Please read this instruction manual thoroughly before using this unit. Please keep for future reference. For specific information about your own temperature, CONSULT YOUR DOCTOR.
MC-720-AP_A_M.book Page 1 Friday, May 22, 2015 11:01 AM
Important Safety Information To assure correct use of the product, basic safety measures should always be followed, including the precautions listed below. Warning: The following situations are potentially hazardous and, if not avoided, could result in death or serious injury. • Conducting self-diagnosis based on the measurement results and/or treatment can be dangerous. Please follow the instructions of your doctor. Self-diagnosis may worsen the symptoms. • A high or prolonged fever requires medical attention, especially for young children. Please contact your doctor. • Keep the unit out of children's reach. • If a child swallows the battery, immediately consult with a doctor. • Do not throw batteries into a fire. The battery may explode. • Remove the battery when the unit will not be used for three months or more. Failure to do so may lead to fluid leakage, heat generation or bursting, resulting in damage to the unit. • Do not use the unit to measure any item with a temperature higher than 80.0°C (176.0°F). General Precautions • Do not apply a strong shock to, drop, step on, or vibrate the main unit. • Do not use a mobile phone near the unit. • The main unit is not waterproof. Be careful when handling this unit so that no liquid (alcohol, water, or hot water) will get into the main unit. When the unit is wet with vapor, wait until it dries or wipe it lightly with a soft dry cloth. • Do not disassemble, repair, or modify the unit. • When the probe lens becomes dirty, wipe it lightly with a soft dry cloth or a cotton swab. Do not wipe the probe lens with tissue paper or a paper towel. For details, refer to section 4.3. • If there is any temperature difference between the places where the unit is stored and where you are going to measure, leave the unit in the room where you are going to use it for more than 30 minutes to allow it to reach room temperature first, then measure.
EN
1
MC-720-AP_A_M.book Page 2 Friday, May 22, 2015 11:01 AM
Important Safety Information • If the forehead is cold, wait until the forehead is warmed up before taking a temperature measurement. The measured result may indicate low when you use an ice bag or an ice pack, or immediately after coming in from outdoor in winter. • Do not use the unit when wet after swimming, bathing, or through sweating after physical exercise, as skin temperature may have fallen. Wipe away moisture and allow 30 minutes for the body to acclimatize to room temperature before measuring. • Do not touch the probe lens with a finger or breathe on it. • Do not attempt measurements when the unit is wet as inaccurate readings may result. • Check the icon on the display before and after the measurement so that the measurement is taken in the appropriate mode. • When taking the temperature of an object with low emissivity such as gold or aluminum, inaccurate readings may result. • When taking the temperature of something with low emissivity (such as milk) in surface measurement mode, slightly lower readings may result. • Do not use the unit in places where strong static electricity or electromagnetic fields are present. Doing so may lead to inaccurate readings and may contribute to instrument failure. • When you inform your doctor of your temperature, make sure you state that you measured the temperature from the forehead.
2
MC-720-AP_A_M.book Page 3 Friday, May 22, 2015 11:01 AM
1. Overview Main unit:
Probe lens
Display: Surface temperature
EN
Room temperature Last reading
Probe lens cap Display
Battery icon
ON/MEM button START button
Mute icon Battery compartment
Temperature mode (°F) Memory icon Forehead temperature
Temperature mode (°C)
3
MC-720-AP_A_M.book Page 4 Friday, May 22, 2015 11:01 AM
2. Preparation 2.1
Removing the Insulating Tape When using the unit for the first time, pull the insulating tape out of the battery compartment. The unit is turned on and after 1 minute the room temperature appears on the display.
2.2
Back of the unit
Notes: • The room temperature remains on the display even after the unit turns off. • Place the unit on a level surface, in a place that is at room temperature (away from direct sunlight or air conditioning, etc.).
Switching between °C and °F This unit is set in °C as default.
1. While the room temperature is displayed, press and hold the START button. 2. While holding it down, press and hold the ON/MEM button until °F appears on the display with two beeps.
4
Notes: • To select °C mode again, start from step1. • When the unit is switched between °C and °F, all the readings stored in the memory are deleted.
MC-720-AP_A_M.book Page 5 Friday, May 22, 2015 11:01 AM
2. Preparation
2.3
Setting the Buzzer The buzzer is activated by default, but can be deactivated. Note: The buzzer is only available in forehead measurement mode.
EN
1. Press the ON/MEM button to turn on the unit. 2. Press and hold the ON/MEM button for 3 seconds. The "
" icon flashes on the display.
3. Release the ON/MEM button. The "
" icon remains lit and the buzzer is deactivated.
Notes: • If the ON/MEM button is pressed down for more than 5 seconds after the " flashing, the unit turns off without deactivating the buzzer. • To activate the buzzer again, start from step1.
" icon starts
5
MC-720-AP_A_M.book Page 6 Friday, May 22, 2015 11:01 AM
3. Using the Unit 3.1
Taking a Reading Forehead Measurement Mode This mode displays the oral-equivalent value of the measured forehead temperature. Always make sure the probe lens is clean and undamaged, and the forehead is clean with no sweat, cosmetics, scars, etc. Prior to measurement, ensure that the patient has not taken a bath or exercised within the last 30 minutes, and has remained in a stable environment for at least 5 minutes. It is recommended to perform measurement three times. If the three measurements are different, select the highest temperature.
1. Pull off the probe lens cap to remove it. 2. Press the ON/MEM button. All indicators appear on the display for 1 second. Then the display shown on the far right appears with two beeps.
6
MC-720-AP_A_M.book Page 7 Friday, May 22, 2015 11:01 AM
3. Using the Unit
3. Hold the unit with the probe lens 1 to 3 cm away from the center of the forehead. Notes: • Holding the unit too long in the hand may cause a higher ambient temperature reading of the probe. This could make the body temperature measurement lower than usual. • Keep still during measurement.
EN
4. Press the START button. The measurement is complete in 1 second with a long beep.
7
MC-720-AP_A_M.book Page 8 Friday, May 22, 2015 11:01 AM
3. Using the Unit
5. Move the unit away from the forehead and check the measurement result. The display lights up for 5 seconds. Notes: • If the result is higher than 42.2°C (108.0°F), “Hi” appears on the display. If the result is lower than 34.0°C (93.2°F), “Lo” appears on the display. • You can take another measurement after the display light turns off with two beeps. Make sure the " " icon remains lit. • After the second measurement, the first measurement is displayed in the top right corner, while the second is displayed in the center.
6. Press and hold the ON/MEM button until "OFF" appears on the display to turn off the unit. The unit stores the measurement in memory, then turns off with the room temperature on the display. The unit also turns off automatically if no operation is performed for 1 minute.
8
MC-720-AP_A_M.book Page 9 Friday, May 22, 2015 11:01 AM
3. Using the Unit
Surface Measurement Mode Surface measurement mode shows the actual and unadjusted surface temperature of an object, which is different from the body temperature. This mode can help you determine if the object temperature is suitable for a patient or baby (such as baby milk). The unit is set to forehead measurement mode as default. Follow the steps below to switch to surface measurement mode.
EN
1. Press the ON/MEM button to turn on the unit. 2. Press and hold the ON/MEM button. While holding it down, press the START button. The "
" icon appears on the display.
3. Move the unit close to the object and press the START button.
It measures continuously until the START button is released. Notes: • Move the probe lens as close to the object as possible when using in surface measurement mode (1 cm apart is recommended). Do not apply the probe lens to the object directly. • Surface measurement mode shows the surface result. The surface and internal temperature may be different. Make sure it is safe when measuring an object with extremely high or low temperature. • The display does not light up and the buzzer is off when using this mode. • To switch to forehead measurement mode, go back to step 2 and make sure the icon changes to the icon. • The unit automatically turns off if no operation is performed for 1 minute with the room temperature on the display. To manually turn off the unit, you can press and hold the ON/MEM button until “OFF” appears on the display. • Surface measurement mode is not intended for body temperature measurement or medical use.
9
MC-720-AP_A_M.book Page 10 Friday, May 22, 2015 11:01 AM
3. Using the Unit
3.2
Using the Memory Function This unit automatically stores the last 25 measurement results (excluding Hi/Lo results). Notes: • If the memory is full, the unit will delete the oldest reading. • When the unit is switched between °C and °F, all the readings stored in the memory are deleted.
1. Press the ON/MEM button to turn on the unit. 2. Press the ON/MEM button again. The memory number appears on the display.
3. Release the ON/MEM button. The most recent result appears on the display. Press the ON/MEM button repeatedly to view the older results.
4. Press and hold the ON/MEM button until "OFF" appears on the display to turn off the unit.
10
The unit turns off with the room temperature on the display.
MC-720-AP_A_M.book Page 11 Friday, May 22, 2015 11:01 AM
4. Troubleshooting and Maintenance 4.1
The Icons and Error Messages
Error Display
Cause
Remedy
Battery is low.
Replace the battery. (Refer to section 4.4)
Taking measurement before unit stabilizes.
Wait until the “ ” icon stops flashing.
The unit is showing a rapid ambient temperature change.
Allow the unit to rest in a room for at least 30 minutes at room temperature: 10°C–40°C (50°F–104°F).
The ambient temperature is not within the range between 10°C–40°C (50°F–104°F).
Allow the unit to rest in a room for at least 30 minutes at room temperature: 10°C–40°C (50°F–104°F).
Errors 5–9 indicate a malfunction of the unit.
Remove the battery, wait for 1 minute, reinsert the battery, and turn on the unit again. If the message reappears, please contact the OMRON representative.
EN
11
MC-720-AP_A_M.book Page 12 Friday, May 22, 2015 11:01 AM
4. Troubleshooting and Maintenance Error Display
12
Cause
Remedy
• Forehead measurement mode: The temperature taken is higher than +42.2°C (108°F). • Surface measurement mode: The temperature taken is higher than +80°C (176°F).
Use the unit within its measurable range. If a malfunction still exists, please contact an OMRON representative.
• Forehead measurement mode: The temperature taken is lower than +34°C (93.2°F). • Surface measurement mode: The temperature taken is lower than -22°C (-7.6°F).
Use the unit within its measurable range. If a malfunction still exists, please contact an OMRON representative.
The unit cannot be powered on to ready stage.
Change to a new battery. (Refer to section 4.4)
MC-720-AP_A_M.book Page 13 Friday, May 22, 2015 11:01 AM
4. Troubleshooting and Maintenance
4.2
Questions and Answers 1. The temperature indicated is rather high.
• The unit has been stored in a cold room. When the unit has been stored in a cold room, the measured temperature may be rather high if you measure the temperature shortly after bringing the unit to a room that is at normal temperature. Measure temperature after leaving the unit for 30 minutes in the room where it will be used. It is advisable to store the unit in the room where you are going to measure temperature. • The probe lens has been exposed to direct sunlight. • Measurement was taken shortly after the subject cried. Crying can increase facial temperature.
EN
2. The temperature indicated is rather low.
• The body is sweaty. Do not use the unit when wet after swimming, bathing, or through sweating after physical exercise, as skin temperature may have fallen. Wipe away moisture and allow 30 minutes for the body to acclimatize to room temperature before measuring. • The probe lens is dirty. • Measurement was taken in surface temperature mode. Use forehead measurement mode. • Measurement was taken shortly after the subject was outdoors in cold weather. Wait until the forehead has warmed up. • Measurement was taken at a location other than the forehead. Be sure to measure exactly at the forehead, not surrounding areas such as the hairline.
13
MC-720-AP_A_M.book Page 14 Friday, May 22, 2015 11:01 AM
4. Troubleshooting and Maintenance 3. There are variances in the measurement values. • Measurements were not taken at the same location. • The probe lens is dirty.
4. The temperature measured on this unit is not equivalent to oral-temperature. • Forehead thermometers may not always calculate the exact oral-equivalent value. It is recommended to take regular measurements on subjects such as family members, etc., who are in good physical condition, to know their normal temperature.
4.3
Maintenance • Check the unit if it is dropped. If unsure, please contact your local OMRON representative to have your unit checked. • When the probe lens becomes dirty, wipe it lightly with a soft dry cloth or a cotton swab. Do not wipe the probe lens with tissue paper or a paper towel. • Clean the probe lens carefully to avoid damaging it. - Use a 70% alcohol swab or soft cotton moistened with 70% alcohol to clean the probe lens. - Allow the probe lens to fully dry for at least 1 minute. • Do not store the unit in the following types of places. Doing so may damage the unit. - Wet locations. - Locations with high heat and humidity or those that are exposed to direct sunlight. Areas close to heating equipment, dusty locations, or environments where there are high salt concentrations in the air. - Locations where the unit will be subjected to leaning over, falling, shock, or vibration. - Pharmaceutical storage areas or locations where corrosive gases are present.
14
MC-720-AP_A_M.book Page 15 Friday, May 22, 2015 11:01 AM
4. Troubleshooting and Maintenance
4.4
Replacing the Battery Battery: CR2032 lithium button battery Note: To protect the environment, discard the used batteries in accordance with the local regulations regarding waste disposal procedure.
EN
1. Release the screw and remove the battery cover. 2. Remove the battery with a pointed object. Note: Do not use a metal tool as it may damage the surrounding parts.
3. Insert the new battery under the metal hook on the right
Metal hook
side and press the left side of the battery down until it clicks. Note: Replace the new battery with the plus (+) side on the top.
4. Close the battery cover and tighten the screw.
15
MC-720-AP_A_M.book Page 16 Friday, May 22, 2015 11:01 AM
5. Technical Data Product Description: Model: Sensor: Temperature Display: Accuracy:
Measurement Range:
Power Supply: Power Consumption: Battery Life: Operating Temperature and Humidity:
16
Forehead Thermometer MC-720 Thermopile 4-digit, °F display in 0.1 degree increments 3-digit, °C display in 0.1 degree increments Forehead measurement mode ±0.2°C (±0.4°F) within 35.0°C to 42.0°C (95.0°F to 107.6°F), ±0.3°C (±0.5°F) for other range Surface measurement mode ±0.3°C (±0.5°F) within 22.0°C to 42.2°C (71.6°F to 108.0°F). Outside this range, it is ±2°C (±3.6°F) or ±4% (percentage of actual temperature), whichever is greater. Room temperature mode ±2°C (±3.6°F) within 10°C to 40°C (50°F to 104°F). Out of this range, the measurement result cannot meet the aforementioned accuracy. Forehead measurement mode 34.0°C (93.2°F) to 42.2°C (108.0°F) Surface measurement mode -22.0°C (-7.6°F) to 80.0°C (176.0°F) Room temperature mode 10°C (50°F) to 40°C (104°F) 3.0V DC, 1 CR2032 lithium button battery 0.018 W With a new battery, approx. 2,500 measurements 10°C (50°F) to 40°C (104°F), 15 to 85%RH
MC-720-AP_A_M.book Page 17 Friday, May 22, 2015 11:01 AM
5. Technical Data Storage Temperature/ Humidity/Air Pressure: Protection against electric shock: Applied Part:
-20°C (-4°F) to 60°C (140°F), 10 to 95%RH, 200 to 1060 hPa Internally powered ME equipment
EN = type BF
Weight: Outer Dimensions: Package Contents:
Approx. 90 g (with battery installed) 45(w) × 155(h) × 39(d) mm Main unit, test battery (CR2032 installed), probe lens cap, instruction manual.
Notes:
• The specification may be changed without prior notice. • This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co. Ltd., Japan. • This unit fulfills the provisions of the EC directive 93/42/EEC (Medical Device Directive) and the European Standard EN ISO 80601-2-56:2012: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.
17
MC-720-AP_A_M.book Page 18 Friday, May 22, 2015 11:01 AM
5. Technical Data Important information regarding Electro Magnetic Compatibility (EMC) With the increased number of electronic devices such as PC's and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation. Medical devices should also not interfere with other devices. In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe product situations, the EN60601-1-2 standard has been implemented. This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices. This medical device manufactured by OMRON HEALTHCARE conforms to this EN60601-1-2:2007 standard for both immunity and emissions. Nevertheless, special precautions need to be observed: • Do not use mobile (cellular) telephones and other devices, which generate strong electrical or electromagnetic fields, near the medical device. This may result in incorrect operation of the unit and create a potentially unsafe situation. Recommendation is to keep a minimum distance of 7 m. Verify correct operation of the device in case the distance is shorter. Further documentation in accordance with EN60601-1-2:2007 is available at OMRON HEALTHCARE EUROPE at the address mentioned in this instruction manual. Documentation is also available at www.omron-healthcare.com. Correct Disposal of This Product (Waste Electrical & Electronic Equipment) This marking shown on the product or its literature, indicates that it should not be disposed with other household wastes at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources. Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can take this item for environmentally safe recycling. Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial wastes for disposal. Disposal of used batteries should be carried out in accordance with the national regulations for the disposal of batteries.
18