ORTHOFIX
OSCAR Pro User Manual
User Manual
64 Pages
Preview
Page 1
USER MANUAL
Keep this leaflet: you may need to read it again. Discard previous versions Important information - please read prior to use . See also instruction leaflets PQPRO for single use probes and PQRMD for reusable medical devices.
Oscar Pro
™
System
CONTENTS
0. IMPORTANT INFORMATION UPON DELIVERY 0.1. FIRST VERIFICATION UPON DELIVERY 0.2. DEVICE PARTS IDENTIFICATION 0.3. SYMBOL EXPLANATION 0.4. SPECIFICATIONS 0.5. ELECTRICAL SAFETY STATEMENT 0.6. ELECTROMAGNETIC COMPATIBILITY (EMC) STATEMENTS 0.7. SUMMARY OF SAFETY INFORMATION 0.7. 1. Warnings 0.7. 2. Precautions 0.7.3 Risks due to the re-use of “Single Use” Device 0.7.4 Sterile & Non-Sterile Products 1. FIRST POWER ON 2. CLINICAL INFORMATION 2.1. INTENDED PURPOSE 2.2. INDICATIONS FOR USE 2.3. CONTRAINDICATIONS 2.4. INTENDED PATIENTS 2.5. INTENDED USERS 2.6. DISCLAIMER 2.7. POSSIBLE ADVERSE EVENTS 2.8 MRI (Magnetic Resonance Imaging) SAFETY INFORMATION 2.9. EXPECTED CLINICAL BENEFITS AND PERFORMANCE CHARACTERISTICS OF THE DEVICE 2.10. INFORMATION FOR THE PATIENT 2.11 NOTICE ABOUT SERIOUS INCIDENTS 3. PRODUCT DESCRIPTION 3.1. PRINCIPLE OF OPERATION 3.2. SYSTEM OVERVIEW 4. SYSTEM SETUP 4.1. INSTALLATION OF GENERATOR 4.2. ASSEMBLY AND CONNECTION OF HANDSET, PROBE AND ACCESSORIES 4.3. FOOTSWITCHES 4.4. SYSTEM INITIALIZATION AND USE 5. OSCAR PRO™ GENERATOR MENU FUNCTIONS MAP 5.1. STATUS BAR 5.2. ON SCREEN HELP 5.3. SETTINGS
8 8 9 10 11 13 13 16 16 17 18 18 18 21 21 21 21 21 21 21 21 21 21 22 22 22 22 23 26 27 28 31 32 35 36 36 36
3
4
CONTENTS
5.3.1. Language settings 5.3.2. System settings 5.4. SURGICAL FUNCTIONS 5.4.1. Cement Removal 5.4.2. Cementless Prosthesis Removal 5.4.3. Osteotomy 5.5. SELECT CHANNEL & PROBE SCAN 5.5.1. Selecting a channel 5.5.2. Cementless Prosthesis Removal 5.5.3. Enabling the footswitch 5.5.4. Switching channels 5.6. OPERATE 5.6.1. Activating ultrasound 5.6.2. Starting/Stopping irrigation 5.6.3. Switching handset/channel 5.6.4. Probe changes during surgery 5.7. END OF USE 5.7.1. Switching off the generator user interface 5.7.2. Purging the irrigation tube 5.7.3. Turn off the generator 6. SYSTEM & MAINTENANCE AREA 6.1. RESET 7. CONNECTION AREA 7.1. COPYING OR SENDING THE DATA LOG 7.1.1. Copy Data Log onto USB 7.1.2. Sending the data log via Wi-Fi 7.2. ADMIN AREA 8. AUDIBLE AND VISIBLE FEEDBACK SUMMARY 9. INSTRUCTIONS FOR PROCESSING AND REPROCESSING 9.1. WARNINGS 9.2. LIMITATIONS ON REPROCESSING 9.3. POINT OF USE 9.4. CONTAINMENT AND TRANSPORTATION 9.5. CLEANING AND STERILIZATION 9.5.1. Disconnection and Disassembly 9.5.2. Cleaning
36 37 37 37 38 38 38 39 39 39 39 40 40 41 42 42 42 42 42 42 43 43 43 44 44 45 45 46 46 47 47 47 47 47 48 54
CONTENTS
9.5.3. Maintenance, inspection and function testing 9.5.4. Packaging 9.5.5. Sterilization 9.5.6. Deviations from Cleaning and Sterilization Instructions 9.6. STORAGE 10. TROUBLESHOOTING 10.1. EVENTS 10.1.1. Non-blocking events 10.1.2. Blocking events 10.2.1. Touchscreen frozen 11. SERVICE & REPAIR 11.1. CHANGING FUSES 11.2. SERVICE & REPAIR ACTIVITIES 12. DISPOSAL
55 57 57 57 57 58 58 58 58 59 60 60 60 61
FIGURE 1 LOADING PAGE
19
FIGURE 2 FIRST POWER ON - LANGUAGE SETTING
19
FIGURE 3 FIRST POWER ON - DATE/TIME SETTING
19
FIGURE 4 SURGICAL FUNCTION MENU
20
FIGURE 5 AWAITING HANDSET CONNECTION
20
FIGURE 6 READY TO SCAN
20
FIGURE 7 INTERCONNECTION DIAGRAM FRONT VIEW
23
FIGURE 8 INTERCONNECTION DIAGRAM REAR VIEW
23
FIGURE 9 HANDSET MAIN PARTS
24
FIGURE 10 OSCAR PRO™ DIAGRAMMATIC VIEW
25
FIGURE 11 OSCAR PRO™ INSTRUMENT TRAY
25
FIGURE 12 HANDSET HORN POSITION
28
FIGURE 13 HANDSET HOLDER FOR PROBE COUPLING
28
FIGURE 14 HANDSET-PROBE COUPLING
28
FIGURE 15 HANDSET HANDLE POSITION
29
FIGURE 16 IRRIGATION NOZZLE ATTACHMENT
29
FIGURE 17 IRRIGATION TUBE ROUTING
29
FIGURE 18 IRRIGATION TUBE POSITIONING
29
FIGURE 19 PLUGGING THE HANDSET CABLE INTO THE HANDSET
30
5
6
CONTENTS
FIGURE 20 CLIPPING IN THE TUBE & HANDSET CABLE
30
FIGURE 21 HANDSET CABLE PLUG
30
FIGURE 22 LINING UP THE CABLE CONNECTOR AND THE RECEPTACLE
30
FIGURE 23 FOOTSWITCH CONNECTION
31
FIGURE 24 LOADING PAGE
32
FIGURE 25 USER MANUAL REFERENCE
32
FIGURE 26 SURGICAL FUNCTION MENU
33
FIGURE 27 SELECT CHANNEL
33
FIGURE 28 PROBE SCAN
33
FIGURE 29 ENABLING THE FOOTSWITCH
34
FIGURE 30 OPERATE
34
FIGURE 31 IRRIGATION
35
FIGURE 32 STATUS BAR
36
FIGURE 33 ON SCREEN HELP EXAMPLE
36
FIGURE 34 SETTINGS - LANGUAGE PAGE
36
FIGURE 35 SETTINGS PAGE
37
FIGURE 36 SURGICAL FUNCTION MENU PAGE
37
FIGURE 37 SELECT CHANNEL PAGE
39
FIGURE 38 PROBE SCAN PAGE
39
FIGURE 39 OPERATE PAGE
40
FIGURE 40 OPERATE PAGE - DETAILS
40
FIGURE 41 POWER OFF ICON
42
FIGURE 42 CONFIRM IRRIGATION LINE PURGE
42
FIGURE 43 MAIN SWITCH OFF
42
FIGURE 44 SYSTEM & MAINTENANCE AREA ICON
43
FIGURE 45 MAINTENANCE AREA PAGE
43
FIGURE 46 CONFIRMING RESET
43
FIGURE 47 ACCESS TO MAINTANCE AREA
43
FIGURE 48 CONNECTION PAGE AREA
44
FIGURE 49 DATA TRANSFER SUCCESSFUL
44
FIGURE 50 WIFI CONNECTIVITY PAGE
45
FIGURE 51 DATA TRANSFER SUCCESSFUL
45
FIGURE 52 ACCESS TO ADMINISTRATOR AREA
45
FIGURE 53 POWER OFF
48
USER MANUAL
FIGURE 54 MESSAGES AT POWER OFF
48
FIGURE 55 MAIN SWITCH OFF
49
FIGURE 56 POWER CABLE DISCONNECTION
49
FIGURE 57 EQUIPOTENTIAL CABLE DISCONNECTION
49
FIGURE 58 HANDSET CABLE DISCONNECTION FROM GENERATOR
50
FIGURE 59 HANDSET CABLE DISCONNECTION FROM HANDSET
50
FIGURE 60 WATER IRRIGATION DISCONNECTION
51
FIGURE 61 HANDSET HANDLE DISCONNECTION
51
FIGURE 62 PROBE DISASSEMBLY
52
FIGURE 63 REMOVAL OF IRRIGATION TUBE FROM PUMP
52
FIGURE 64 DISCONNECTION OF IRRIGATION TUBE FROM GENERATOR
52
FIGURE 65 DISASSEMBLY OF HANDSET HORN ENCLOSURE
53
FIGURE 66 DISASSEMBLY OF HANDSET BUTTON LEVER
53
FIGURE 67 INSTRUMENT TRAY COMPOSITION
56
FIGURE 68 EVENT ICON EXAMPLE
58
FIGURE 69 WARNING ICON EXAMPLE
58
FIGURE 70 ERROR ICON EXAMPLE
58
FIGURE 71 ERROR ICON EXAMPLE
58
FIGURE 72 REPLACEMENT OF MAIN FUSES
60
7
8
USER MANUAL
0. IMPORTANT INFORMATION UPON DELIVERY Section 0 and Section 1 of this Manual contain important information for the identification, handling and first power on of the OSCAR PRO™. Conventions on Warnings and Precautions WARNING
Disclosure of residual risk after all possible mitigations applied by Orthofix.
PRECAUTION
Prescription of risk mitigation to be adopted by user to minimise exposure to hazard.
NOTE
Useful information to allow efficient use of the system.
0.1. FIRST VERIFICATION UPON DELIVERY Conventions on Warnings and Precautions PRECAUTION
In case of visible damage to packaging or device upon delivery, do not use the device and contact the manufacturer for assistance or replacement.
PRECAUTION
After receiving OSCAR PRO™, transport and store the device in its dedicated transport case. Using different container may result in damage, malfunction or failure before or during use.
USER MANUAL
0.2. DEVICE PARTS IDENTIFICATION
O4CABLE
Part #
Code
Description
1
OS4000*
OSCAR PRO GENERATOR
2
O4BEAM
OSCAR PRO HOLDER FOR WATER BAG L459mm
3
PERISTALTIC PUMP, embedded in the generator
4
2000972
STERILE KIT FOR PERISTALTIC PUMP (not for sale in Canada)
4
800001
STERILE KIT FOR PERISTALTIC PUMP US (not for sale in Canada)
5
O4HAND*
OSCAR PRO UNIVERSAL HANDSET
6
PROBE, various models, provided separately
7
O4CABLE
OSCAR PRO UNIVERSAL HANDSET CABLE
8
O4UHH
OSCAR PRO UNIVERSAL HANDSET HANDLE
9
O4FOOT-W*
OSCAR PRO FOOTSWITCH FOR IRRIGATION
10
O4FOOT-U*
OSCAR PRO FOOTSWITCH FOR ULTRASOUND
11
OSCAR PRO MAIN LEAD, various models
12
SPAN7mm
SPANNER 7mm
13
SPAN8/9mm
SPANNER 8/9mm
14
O4HOLD
OSCAR PRO HANDSET HOLDER
15
O4IPL200
OSCAR PRO SLAP HAMMER
16
O4TRAY-02
OSCAR PRO LID TRAY
17
O4TRAY-01
OSCAR PRO BASE TRAY
* Component of the device: OSCARPRO – ULTRASOUND MEDICAL SYSTEM
9
10 USER MANUAL
0.3. SYMBOL EXPLANATION APPENDIX 1
Markings
Symbol
Description Medical Device Catalogue number Batch code Follow instruction for use
0123
CE marking in conformity to applicable European Medical Device Directives/Regulations Date of manufacture
Manufacturer
VOLTAGE: 115-240 V 50-60 Hz
Supply AC Voltage Supply AC Frequency
POWER: 250 VA
Input Power
28kHz 150W
Output Power
T5A H 250V 5x20mm
Fuse Details Supply Power ON Supply Power OFF Disposal. In accordance with the WEEE 2012/19/UE Directive, the product and all its parts (cable, batteries, accessories, etc.) may not be treated as domestic waste. For further information regarding disposal, contact the appropriate department of your local council or the manufacturer service personnel. Warning: counterrotating rollers. Do not operate with open cover. On peristaltic pump embedded in the generator: Flow direction of irrigation liquid.
FOR INTERMITTENT USE ON/OFF
10 | 20 s Output connector Type B Applied Part Equipotential From left to right: Ultrasound - Footswitch connections - Irrigation Wi-Fi connectivity USB port (for mass storage NOT connected to other devices, network or power supply) Caution
USER MANUAL
Stand-by Caution: Federal law (USA) restricts this device to sale by or on the order of a physician. Temperature limitation 90%
Humidity limitation
10 % 1060hPa
Atmospheric Pressure limitation 500hPa
Non-sterile Unique Device Identifier Caution : Consult instructions for use for important cautionary information Consult instructions for use or consult electronic instructions for use
FRAGILE
Fragile Ultrasound Irrigation Locking – Unlocking direction
0.4. SPECIFICATIONS APPENDIX 2
Technical Specification
Symbol
Description
Dimensions:
Generator
377 mm(width) x 180 mm(height) x 415 mm(depth)
Weight:
Generator
9.1 Kg.
Transport case (loaded with generator) and accessories
19.5 Kg
Handset
0.8 Kg
Transport case (loaded with handset, probes, accessories)
11.9Kg
Fuse Type
T5A H 250V 5x20mm
Power Supply input
115 – 240 V
Power Consumption
250 VA
Output – Frequency of operation
28kHz (Channel I and II)
Output - Accuracy of frequency display
1%
11
12 USER MANUAL
Output power
150 W
Volume Adjustment
53 dbA - 88 dBa
Mode of operation
Intermittent Use ON | OFF (Channel I & II) 10s | 20s
Electrical Classification
Generator Class I - generator has a protective earth connection Handset probes: Type B (patient applied parts are at earth potential)
Power Supply Switch
Mean to isolate equipment circuits electrically from the supply mains on all poles simultaneously Transmission and receiving frequencies [bandwidth]
Maximum output
2.4GHz 801.11b [2400~2483.5 MHz]
12.95mdBm
19.72 mW
2.4GHz 801.11g [2400~2483.5 MHz]
12.89mdBm
19.45 mW
2.4GHz 801.11n HT20 [2400~2483.5 MHz]
12.78mdBm
18.97 mW
2.4GHz 801.11n HT40[(2400~2483.5 MHz]
10.28mdBm
10.67 mW
Wireless: Radio-frequency specifications
Handset materials
Handset enclosure: Aluminum alloy EN-AW 7075 conforming to UNI EN 573 Handset transducer: Titanium alloy + Piezoelectric ceramic + Silicone rings
Slap Hammer material
Stainless Steel AISI304 conforming to ASTM F899 and ISO 7153-1
Ingress protection
Generator: IPX0 Handset: autoclavable
Environment for transport and storage
Environment for use
Other Accessories
Temperature: Humidity: Atmospheric pressure:
-40°C to +50°C | -40 °F to +122 °F 10% to 90% 500hPa to 1060hPa
Temperature: Humidity: Atmospheric pressure:
10°C to 30°C | +50 °F to +86 °F 30% to 75% 700hPa to 1060hPa
Transport cases for the generator (GENERATOR FLIGHT CASE) and for the instrument tray (INSTRUMENT FLIGHT CASE). Reference section 9.5 for details of reprocessing.
Items to be sterilised (moist heat sterilisation)
Detachable parts
O4HAND OSCAR PRO UNIVERSAL HANDSET O4CABLE OSCAR PRO UNIVERSAL HANDSET CABLE O4UHH OSCAR PRO UNIVERSAL HANDSET HANDLE O4HOLD OSCAR PRO HANDSET HOLDER SPAN7MM SPANNER 7 MM SPAN8/9MM SPANNER 8/9MM O4IPL200 OSCAR PRO SLAP HAMMER O4TRAY OSCAR PRO INSTRUMENT TRAY See section 3.2 for identification. O4HAND-05 O4HAND-04
HANDSET HORN ENCLOSURE HANDSET BUTTON LEVER
OSCAR PRO™ has been designed and built in accordance with ISO 13485 Quality Assurance standard for medical devices and Part 820 of the Title 21 of the Code of Federal Regulations of the USA.
USER MANUAL
0.5. ELECTRICAL SAFETY STATEMENT The equipment complies with EN 60601-1 and IEC 60601-1. The verification of electrical safety according to EN 60601-1 and IEC 60601-1 is performed by Orthofix before delivery and is valid until the date specified in the dedicated label on the back panel of the generator. The generator is provided with a receptacle for connection of a POTENTIAL EQUALIZATION CONDUCTOR conforming to requirements of EN 60601-1 (ref. Figure 8 and symbol in section 0.3). Warning and Precaution WARNING
The electrical safety label is valid if all parts of the system are used, transported and stored according to the instructions of this manual. After the expiry date of electrical safety, Orthofix is not responsible for malfunctioning, damage to the device, environment or other objects, or harm to users or patient, that may be related to the loss of electrical safety.
PRECAUTION
To avoid the risk of electric shock, this equipment must be connected to a grounded power supply. Appropriate system grounding can only be ensured when an approved, hospital-grade receptacle and matching power cord are used. Install plug and receptacles as per local regulations before operating the unit. Power cord, plug and receptacle should be examined to verify that they are in good working condition before connecting the generator. Never pull on the power cord to remove it from the receptacle.
The product is covered by worldwide patents covering all international markets. Orthofix will provide circuit diagrams, component parts lists and descriptions on request. However, Orthofix does not designate any part of the equipment as user-repairable.
0.6. ELECTROMAGNETIC COMPATIBILITY (EMC) STATEMENTS This equipment has been tested and found to comply with the limits for a medical device. However, should interference occur, try the following measures: 1. Turn equipment off and on to confirm the source of the interference. 2. Increase separation between this equipment and other devices. 3. Connect this equipment to a power socket different from that to which the other devices are connected. 4. Consult medical physics department. Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this manual. Portable and mobile RF communications equipment can affect medical electrical equipment. Warning WARNING
Careful consideration should be given to patients who have pacemakers or other active implanted devices. Interference produced can cause pacemaker or other active implantable device to malfunction, enter an unsafe mode, or cause permanent damage to the device. Consult the pacemaker or other active device manufacturer Instruction For Use for specific info.
13
14 USER MANUAL
Table 1 Guidance and manufacturer’s declaration - electromagnetic emissions
OSCAR PRO™ is intended for use in the electromagnetic environment specified below. The customer or the user of OSCAR PRO should assure that it is used in such an environment. Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The OSCAR PRO™ must emit electro-magnetic energy in order to perform its intended function. Nearby electronic equipment may be affected.
RF emissions CISPR 11
Class A
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations /flicker emissions IEC 61000-3-3
Complies
OSCAR PRO™ is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Table 2 Guidance and manufacturer’s declaration - electromagnetic immunity
OSCAR PRO™ is intended for use in the electromagnetic environment specified below. The customer or the user of OSCAR PRO should assure that it is used in such an environment. IMMUNITY test
IEC 60601 test level
Compliance level
Electromagnetic environment guidance
Electrostatic discharge (ESD) IEC 61000-4-2
8kV contact 2kV, ± 4kV, ± 8kV, ± 15kV air
8kV contact 2kV, ± 4kV, ± 8kV, ± 15kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
± 2kV for power supply lines ± 1kV for signal lines
Main power quality should be that of a typical commercial or hospital environment.
Electrical fast transient/burst IEC 61000-4-4
± 2kV for power supply lines ± 1kV for signal lines ± 1kV line to line ± 2kV line to earth
± 1kV differential mode ± 2kV common mode
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
<5% UT (>95% dip in UT) for 0.5 cycle 40% UT (60% dip in UT) for 5 cycles 70% UT (30% dip in UT) for 25 cycles <5% UT (>95% dip in UT) for 5s
<5% UT (>95% dip in UT) for 0.5 cycle 40% UT (60% dip in UT) for 5 cycles 70% UT (30% dip in UT) for 25 cycles <5% UT (>95% dip in UT) for 5s
Main power quality should be that of a typical commercial or hospital environment. If the user of OSCAR PRO™ requires continued operation during power main interruptions, it is recommended that OSCAR PRO™ be powered from an uninterruptible power supply.
Power frequency (50/60Hz) magnetic field IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at levels characteristic of a typical commercial or hospital environment
Surge IEC 61000-4-5
NOTE UT is the a.c. main voltage prior to application of the test level
Main power quality should be that of a typical commercial or hospital environment.
USER MANUAL
Table 3 List of Cables Item
Length
Type
Handset Cable
3.0 m
Shielded 4-conductor
Power cord
3.0 m
Unshielded 3-conductor
Precaution PRECAUTION
The use of accessories, handsets and cables other than those provided with OSCAR PRO™ may result in increased emissions and reduced immunity of the device. Such issues may result in user and/or patient harm. Use only Orthofix branded equipment and accessories.
Table 4 Guidance and manufacturer’s declaration - electromagnetic immunity
OSCAR PRO™ is intended for use in the electromagnetic environment specified below. The customer or the user of OSCAR PRO should assure that it is used in such an environment. IMMUNITY test
IEC 60601 test level
Compliance level
Conducted RF IEC 61000-4-6
3Vrms 150kHz to 80MHz
3Vrms
Radiated RF IEC 61000-4-3
3V/m 80MHz to 2.5GHz
3V/m
Electromagnetic environment guidance Portable and mobile RF communications equipment should be used no closer to any part of OSCAR PRO™, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance d = 1.2√P d = 1.2√P, 80MHz to 800MHz d = 2.3√P, 800MHz to 2.3GHz where P is the maximum power output rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol.
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which OSCAR PRO™ is used exceeds the applicable RF compliance level above, OSCAR PRO should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary,such as re-orienting or relocating OSCAR PRO. b Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m. a
15
16 USER MANUAL
Table 5 Recommended separation distances between portable and mobile RF communications equipment and OSCAR
OSCAR PRO™ is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of OSCAR PRO™ can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment and OSCAR PRO™ as recommended below, according to the maximum output power of the communications equipment. Rated maximum output power of transmitter
Separation distance according to frequency of transmitter 150kHz to 80MHz d = 1.2√P
80MHz to 800MHz d = 1.2√P
800MHz to 2.5GHz d = 2.3√P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80MHz and 800MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
0.7. SUMMARY OF SAFETY INFORMATION Please read this section of the manual carefully. It contains a summary of all precaution, warning and caution statements contained in the manual. However, the user is advised to read the entire manual and operate the device only in accordance with all of the instructions contained herein. Servicing of this device should only be performed by qualified technicians authorised by Orthofix. There are no service controls accessible to the user.
0.7.1. Warnings CONTEXT Preparation
Use Use Use
INFORMATION The electrical safety label is valid if all parts of the system are used, transported and stored according to the instructions of this manual. After the expiry date of electrical safety, Orthofix is not responsible for malfunctioning, damage to the device, environment or other objects, or harm to users or patient, that may be related to the loss of electrical safety. Careful consideration should be given to patients who have pacemakers or other active implanted devices. Interference produced can cause pacemaker or other active implantable device to malfunction, enter an unsafe mode, or cause permanent damage to the device. Consult the pacemaker or other active device manufacturer Instruction For Use for specific info. Portable and mobile RF communication equipment can affect medical electrical equipment. If RF equipment is necessary, monitor the OSCAR PRO™ for appropriate function during the procedure. During use, warning messages may be displayed on the screen in relation to a specific event or machine status. Refer to section 10 of this manual for the meaning of each message.
Use
In case of excessive load, the OSCAR PRO™ warns the user with specific luminous and acoustic signals. See section 8 and section 10.
Reprocessing
Single use devices MUST NOT BE REUSED, as they are not designed to perform as intended after the first usage. The used probes MUST BE discarded after the surgical procedure. Do not attempt to reuse or re-sterilise any single-use items.
Reprocessing
Personnel working with contaminated medical devices must follow safety precautions as per the healthcare facility’s procedures.
Reprocessing
Take care when handling handsets with probes attached, some edges are serrated and sharp.
Reprocessing
DO NOT USE detergents and disinfectants with fluoride, chloride, bromide, iodide or hydroxyl ions. Their use will damage the devices and may cause device failure during use.
Reprocessing
Contact with saline solutions should be kept to a minimum.
Reprocessing
The use of ultrasonic washers is not permitted for cleaning the ultrasonic generator, the handsets, the main lead and the footswitches with their pneumatic tubes.
Reprocessing
The ultrasonic generator, main lead and footswitches with their pneumatic tubes must not be immersed in cleaning solution.
Reprocessing
Sterilisation of the ultrasonic generator, main lead and footswitches with their pneumatic tubes is not permitted.
USER MANUAL
Reprocessing
During washing, avoid contact between devices because movement could cause damage to devices and the washing efficacy could be reduced.
Reprocessing
Do not include additional devices or instruments in the sterilisation tray. Sterility is not guaranteed if the sterilisation tray is overloaded.
Maintenance
Use only USB mass storage devices with no connection with other devices and no connection to power supply. Failing to comply with this restriction may expose the hospital network to security violation and compromise the integrity of the device.
Maintenance
A data log copied to a USB device can be also transmitted via Wi-Fi, but not vice versa.
0.7.2. Precautions CONTEXT
INFORMATION
Preparation
In case of visible damage to packaging or devices upon delivery, do not use the device and contact the manufacturer for assistance or replacement.
Preparation Preparation
Preparation
Preparation
Preparation
Preparation Preparation Preparation
Preparation
Preparation
After receiving OSCAR PRO™, transport and store the device in its dedicated transport case. Using different containers may result in damage, malfunction or failure before or during use. Handle handsets, probes and cables with care, as inappropriate treatment may affect the safety and performance of the unit. Before use it is recommended that the cables are inspected for damaged insulation. Damaged cables should be replaced before further use of the system. The OSCAR PRO™ generates high voltages within the generator itself and the connected handset. To avoid injury, the generator should be never be operated before ensuring that its cover is properly closed and has not been tampered with. Do not attempt to remove or disassemble the cover. There are no userserviceable parts inside the generator. All servicing should only be performed by an authorised Orthofix representative. No modification of this equipment is required nor allowed. In conditions of incorrect use or insufficient maintenance, electrical medical devices may present a residual risk of unexpected shutdown, electric shock or fire. Follow use instructions and ordinary maintenance instructions to prevent occurrence of those issues. This manual shall be kept in close proximity to the system for easy referral when needed. Adequate air circulation is needed to cool electronic components inside of the generator. Do not block the cooling fan at the generator rear or the air vents on the generator bottom. Do not place the generator on a towel, foam or other soft surface since the material may block the air vents. Blocking these vents may cause the generator to overheat and malfunction or create a shock hazard. A clear drape can be used to protect the generator front panel but do not cover the pump housing or other generator sections. This device is considered medical electrical equipment. Medical electrical equipment needs special precautions regarding electromagnetic compatibility (EMC) and needs to be installed and put into service according to the EMC information provided in this operator's manual. OSCAR PRO™ must not be used adjacent to or stacked with other electrical equipment. If adjacent or stacked use is necessary, the generator should be observed to verify normal operation. Connecting the generator to a power outlet with inadequate voltage or frequency may cause the unit to malfunction or create a shock or fire hazard. In case of fuse replacement, ensure that the correct fuses are used (refer to section 11). To avoid the risk of electric shock, this equipment must be connected to a grounded power supply. Appropriate system grounding can only be ensured when an approved, hospital-grade receptacle and matching power cord are used. Install plug and receptacles as per local regulations before operating the unit. Power cord, plug and receptacle should be examined to verify that they are in good working condition before connecting the generator. Never pull on the power cord to remove it from the receptacle. The use of accessories, transducers and cables other than those specified may result in increased emissions and reduced immunity of the device. Such issues may result in user and/or patient harm. Use only Orthofix branded equipment and accessories.
Preparation
Allow the devices to reach room temperature before use.
Preparation
When inserting the irrigation line in the pump, make sure that the intended flow direction matches the indication arrow on the pump cover.
Use
Improper connection of the handset cable may present a shock hazard. Confirm that the handset connector is dry prior to plugging it in.
Use
Ensure that the software status bar, positioned in the right corner of the header, is continuously blinking as feedback that the Graphic User Interface is operational.
Use
When enabling the irrigation function, verify that no spillage is occurring along the entire line from the water bag to the handset.
Use Use
Do not lift or drag the handset by the handset cable, as this may damage both the handset and the cable. Never pull on the handset cable to remove it from the connector. The use of accessories which are not OSCAR PRO™ original components may result in decreased system performance or failure of the device. Such issues may result in user and/or patient harm. Use only Orthofix branded equipment and accessories.
Use
DO NOT use this equipment in the presence of flammable anaesthetic mixture with air or with oxygen or nitrous oxide to prevent fire ignition.
Use
OSCAR PRO™ must not be used adjacent to or stacked with other electrical equipment that may interfere with its normal functionality or decrease its immunity. Minimum distances to other electrical medical devices or electrical appliances are stated in the EMC information provided in this operator's manual.
17
18 USER MANUAL
Use
In case of high loads and prolonged use, in exceptional cases beyond the recommended 10s duty cycle, the handset may accumulate heat up to 45°C. Overheating may affect the user and might reduce the handset transducer's life. For this reason, audible and visible alarms inform the user when the system is approaching this condition. Proceed by changing the handset in use or by reducing the active time of the duty cycle in favour of the cooling time, until the temperature limit returns within the acceptable range.
Use Use
Use Use Use Use
DO NOT operate pump with pump cover in raised position. Rollers might pinch loose clothing or fingers, resulting in personal injuries. During use, the probe for cement removal may reach a temperature up to 120°C, depending on the use. Irrigation, either manual or assisted by the OSCAR PRO™ irrigation function, must be used whenever temperature control is deemed necessary to limit overheating effects such as bone necrosis around the cemented area. See section 5.6 of this manual and the operative technique for specific indications. During use, the probe for bone cutting may reach a temperature above 47°C. Irrigation, either manual or assisted by the OSCAR PRO™ irrigation function, must be used to limit overheating effects such as necrosis of bone or other surrounding tissues. See section 5.6 of this manual and the operative technique for specific indications. To ensure effective cooling, the irrigation flow shall be directed on the tip of the probe by adjusting the direction of the dedicated irrigation nozzle. Follow the appropriate sequence of steps to shut down the generator: first, select the power off option from the menu function and then, once the procedure is complete, including pump cleaning if necessary, the user can proceed with the switch off by pressing the ON|OFF switch positioned in the rear panel of the generator. During bone cutting, bone necrosis may result if the probe is not moved relative to the bone. A continuous probe motion is recommended in order to minimise contact duration with the ultrasonic tip and minimise heat build-up. In this instance a more frequent intermittent use must be applied.
Use Use Reprocessing
Extra care should be taken when operating the energised probe in the vicinity of nerves. Do not operate OSCAR PRO™ Probes on the surgical site simultaneously with laser equipment or high frequency surgical equipment. The combined effect may lead to device damage and patient injury. All reusable handset parts and accessories must be properly decontaminated, cleaned and sterilised before each use as per the instructions contained in this manual. Failure to do so may lead to infections, which can result in serious harm or death.
0.7.3 Risks due to the re-use of “Single Use” Device The “SINGLE USE” non implantable device of Orthofix is identified through symbol “ ” reported on the label or indicated in the “Instructions For Use” supplied with the products. The re-use of a “SINGLE USE” non implantable device cannot guarantee the original mechanical and functional performances, compromising the effectiveness of the products and introducing health risks for the patients.
0.7.4 Sterile & Non-Sterile Products Orthofix devices are provided as STERILE or NON-STERILE and they are labeled as such. In the case of STERILE products, product integrity, sterility and performance are assured only if the packaging is undamaged. Do not use if packaging is compromised, unintentionally opened or if a component is believed to be faulty, suspect or damaged. The products supplied NON-STERILE require cleaning, disinfection and sterilization prior to use according to procedures reported in the following instructions.
1. FIRST POWER ON The tasks in this section allow you to: • Configure the system with your language preference. • Set the local date/time. • Verify the basic integrity of the generator interface, handset cables and handsets before installation in the operating theatre. Precautions PRECAUTION
Allow the devices to reach room temperature before use.
PRECAUTION
In conditions of incorrect use or insufficient maintenance, electrical medical devices may present a residual risk of unexpected shutdown, electric shock or fire. Follow use instructions and ordinary maintenance instructions to prevent occurrence of those issues. This manual shall be kept in close proximity to the system for easy referral when needed.
1. Position the OSCAR PRO™ Generator on a flat surface. Precautions PRECAUTION
Adequate air circulation is needed to cool electronic components inside of the generator. Do not block the cooling fan at the generator rear or the air vents on the generator bottom. Do not place the generator on a towel, foam or other soft surface since the material may block the air vents. Blocking these vents may cause the generator to overheat and malfunction or create a shock hazard. A clear drape can be used to protect the generator front panel but do not cover the pump housing or other generator sections.
USER MANUAL
2. Connect the generator to the power supply using the power cord provided. Precautions PRECAUTION PRECAUTION
PRECAUTION
OSCAR PRO™ must not be used adjacent to or stacked with other electrical equipment. If adjacent or stacked use is necessary, the generator should be observed to verify normal operation. Connecting the generator to a power outlet with inadequate voltage or frequency may cause the unit to malfunction or create a shock or fire hazard. In case of fuse replacement, ensure that the correct fuses are used (refer to section 11). The OSCAR PRO™ generates high voltages within the generator itself and the connected handset. To avoid injury, the generator should never be operated before ensuring that its cover is properly closed and not tampered with. Do not attempt to remove or disassemble the cover. There are no user-serviceable parts inside the generator. All services should only be performed by an authorized Orthofix representative. No modification of this equipment is required nor allowed.
3. Using the main power switch located on the back of the generator, turn the system ON (position “I”). 4. A loading page appears on the touchscreen showing the progress of the system start-up.
Figure 1 LOADING PAGE 5. Select your preferred language from the ones listed. The system interface, messages, errors and help section will be presented in the selected language until any system reset. The reset procedure will return the system to the default configuration. Reset procedure described in section 6.
Figure 2 FIRST POWER ON - LANGUAGE SETTING 6. Now, the user can adjust the date and time settings. Setup local date/time using the [+] and [-] symbols. The date/time settings will remain in use until the next system reset. The reset procedure will revert back to the system’s default configuration.
Figure 3 FIRST POWER ON - DATE/TIME SETTING
19
20 USER MANUAL
7. Touch the Continue button to complete the configuration. The Surgical Functions menu appears. CHECK1. Verify that the system’s clock displays the correct time and date on the bar at the bottom of the page. CHECK2. Verify that the dashes in the top right corner are lit up and they are changing brightness in sequence.
Figure 4 SURGICAL FUNCTION MENU 8. After touching one of the options in the Surgical Functions menu, the generator surges for Handset connection.
Figure 5 AWAITING HANDSET CONNECTION 9. Position the Instrument Tray on a table. Precautions PRECAUTION PRECAUTION
Handle handsets, probes and cables with care, as inappropriate treatment may affect the safety and performance of the unit. Before use it is recommended that the cables are inspected for damaged insulation. Damaged cables should be replaced before further use of the system. Do not lift or drag the handset by the handset cable, as this may damage both the handset and the cable. Never pull on the handset cable to remove it from the connector.
10. Connect a Handset Cable to a Handset and connect the Handset Cable to the socket beside the “I” symbol on the front panel of the Generator. CHECK3. After completing the connection, the box “CHANNEL 1” on the screen should turn from faded to light blue. NOTE: Do not press the activation button on the handset as indicated on the screen, as no scan is possible without a probe connected to the horn. If pressed accidentally, continue to follow the sequence after the scan ends. 11. Unplug the Handset Cable from CHANNEL 1 and repeat step 10 with a different Handset Cable and Handset on CHANNEL 2 (symbol “II” on the front panel of the Generator).
Figure 6 READY TO SCAN
12. Unplug the Handset Cable from CHANNEL 2. 13. Touch the Home icon to return to the Surgical Functions menu. 14. Touch the On-Off icon to switch the interface off. 15. Turn the main power switch off (position “O”). 16. In the event that one or more of CHECK 1, CHECK 2 or CHECK3 fail, contact an Orthofix agent for assistance. The first power ON procedure ends with the above steps. In order to test additional functionalities and/or verify additional hardware components (probe connection and scan, footswitches connection and activation, peristaltic pump activation) in a functional setting, proceed with section 4. This section provides the steps for connecting and putting the complete system into operation. Details on each OSCAR PRO™ Generator software menu are provided in section 5.
USER MANUAL
2. CLINICAL INFORMATION 2.1. INTENDED PURPOSE Ultrasound system for the cutting and removal of bone and acrylic bone cement.
2.2. INDICATIONS FOR USE The OSCAR PRO™ is indicated for orthopaedic surgery on the appendicular skeleton. Specific indicated procedures include: • Osteotomy (ETO and TTO procedure). • Cemented prosthesis revision. • Cementless prosthesis revision.
2.3. CONTRAINDICATIONS DO NOT USE the OSCAR PRO™ if a surgical candidate exhibits or is predisposed to any of the following contraindications: • Orthopaedic application on the axial skeleton (such as artificial disc replacements or spinal arthroplasty). • Patients with general medical conditions not suitable for surgery.
2.4. INTENDED PATIENTS OSCAR PRO™ is ndicated for adults and adolescent (greater than 12 years) patients.
2.5. INTENDED USERS The product is intended for use by Healthcare Professionals (HCP) only. Such HCPs must have full awareness of the appropriate orthopaedic procedures and be familiar with the devices, instruments and surgical procedures.
2.6. DISCLAIMER The HCP is fully responsible for the selection of the appropriate treatment and relevant device for the patient (including post-operative care).
2.7. POSSIBLE ADVERSE EVENTS A successful result is not achieved in every surgical case. Additional complications may develop at any time due to improper use, medical reasons or device failure that require surgical re-intervention. Preoperative and operative procedures including knowledge of surgical techniques and proper selection of the device are important considerations in the successful utilization of the device by the HCP. Possible adverse events resulting from the usage of the OSCAR PRO™ are: • Pain and stiffness at the surgery site. • Necrosis of soft tissues and/or cortical bone due to overheating. • Iatrogenic fractures. • Breaking, loosening, bending or reduced performance of the probes. • Prolonging of surgery and possible extra anesthesia. • Intrinsic adverse events associated with anesthesia and surgery. • Burns to patient or intended user due to accidental contact with probe.
2.8 MRI (Magnetic Resonance Imaging) SAFETY INFORMATION OSCAR PRO™ System is not meant to be used in a MR environment. The use in a MR environment can lead to system malfunction.
2.9. EXPECTED CLINICAL BENEFITS AND PERFORMANCE CHARACTERISTICS OF THE DEVICE EXPECTED CLINICAL BENEFITS • EFFICIENT: Specific probes are utilized in sequence to soften and remove cement from the bone. The technique reduces manual force to the minimum, resulting in low risk of bone fracture and perforation. • SAFE: All probes are designed to preserve the bone quality and soft tissue. The system is designed with specific audible and visual feedback to provide the user a safe and efficient operating environment for their patients. • The surgeon can intuitively guide the probe and reduce the likelihood of unnecessary fracture or perforation of the patient’s bone, whilst removing and clearing bone cement from the intramedullary canal.
21