OPERATIVE TECHNIQUE
OSCAR ultrasonic arthroplasty revision
CONTENTS
INTRODUCTION
3
THE EQUIPMENT
4
OPERATIVE TECHNIQUE
14
CEMENT REMOVAL
17
UNCEMENTED STEM REMOVAL 21 PROBLEMS AND SOLUTIONS
23
INTRODUCTION
INTRODUCTION The task of removing PMMA bone cement during revision arthroplasty has always posed a challenge to the surgical team. Difficulty in visualisation, ambiguity in discriminating between cement and bone and unpredictable scheduling all add to unnecessary complications and in the worst cases severely compromised bone stock and the need for expensive prostheses and other reconstructive materials. Removing these unpredictable problems requires a device that can be relied upon to restore a planned schedule and outcome. Ultrasound, when applied correctly, can make the cement removal part of the operation predictable, safe and by definition more efficient and quicker. PMMA bone cement responds instantly to a vibrating wave-guide. The putty-like consistency of the cement is then easily removed from the endosteum through novel design of the wave-guides (probes). Cancellous bone is affected by ultrasound and will be removed when subjected to the energy levels deployed during cement removal. Cortical bone will not be affected due to the probe design and the energy levels used. It does not absorb ultrasound as readily as the cement and therefore the surgeon will have awareness of the position of the probe within the bone through both audible and tactile feedback. It is important to refer to the operating techniques discussed later in this manual, because any tool is potentially dangerous if used contrary to the controlled methods described. The design of the probes is critical to safe cement removal; by always transferring the softened cement as the probe moves heat build up is not a major issue, provided irrigation is deployed according to the techniques outlined later. As technology has advanced control of the production of ultrasound, its delivery and monitoring, is now fully under digital circuitry. OSCAR was introduced specifically to facilitate the removal of bone cement during hip revision procedures using ultrasound to soften the cement shell holding the implant in place. Special probes are deployed in a sequence to collect and remove the softened cement from the host bone. The technique reduces manual force to a minimum and practically eliminates the risk of bone fracture and perforation. The technology has become accepted as the standard means of joint revision where cement fixation is involved. Ultrasound and the Removal of Bone Cement Ultrasound is the name given to high frequency vibrations (above 16 kHz) which travel through air, liquid or solid media as pressure and displacement waves. In a bounded system, standing waves may be established, which produce a much greater concentration of acoustic energy, and scope for rapid local heating at absorption sites. Attempts were made
at the Endo Clinic in Hamburg in the early 1970’s to remove bone cement by melting it with ultrasound. Bone cement is remarkable in that it can maintain a temperature gradient of 200o over a distance of 1mm. This means that if molten cement can be removed rapidly, the residual cement hardly increases in temperature, thus minimising damage to adjacent bone. In the initial studies in Hamburg, however, it was found difficult to remove the molten cement. The technology is now available to do this, using a series of specially designed handsets and probes in conjunction with a portable ultrasonic generator (“OSCAR” System for Cemented Arthroplasty Revision). Animal studies have demonstrated that when the tip of the probe is in contact with the endosteal bone surface for a period of 10 seconds, cell death to a depth of only 50 microns occurs. This contrasts with cell death to a depth of 500 microns, which occurs when conventional PMMA is applied to the endosteal surface. Cement removal using this technique is extremely rapid, and can normally be accomplished in 25 minutes or less. The risk of bone perforation or fracture is negligible when OSCAR is used, as the probes are designed to emit a characteristic sound the moment they come into contact with the bone surface. The equipment is simple to operate, and both efficient and predictable in its performance. It has been designed with surgeons, scrub nurses, technicians and hospital engineers in mind and its use permits the accurate scheduling of theatre lists. Cementless Prosthesis Revision Over the last 15 years total joint surgery has moved, in some countries, away from cemented fixation to the use of press-fit prostheses. These are normally either porous or hydroxyapatite coated stems and acetabular shells, which encourage bony in-growth to provide fixation and stability. Inevitably, a percentage of these will require revision and the specific task of separating the implant from the bone will require osteotomes and burrs leading to a complex and potentially damaging procedure. Precedents exist for the use of ultrasound powered osteotomes and adding this function to the OSCAR system was a natural extension of the intended use in order to cater for the changing needs in arthroplasty revision surgery. Ultrasound remains one of the most effective and safe energy forms available for powering surgical instruments. By careful design of the oscillatory system, collateral damage due to local heating can be minimized to levels significantly below that associated with high speed burrs and oscillating saws. Low force cutting coupled with precise control of energy delivery to the operating site, ensure that ultrasonically powered osteotomes will answer surgical needs in growing demand for revision arthroplasty.
3
4
EQUIPMENT
THE EQUIPMENT The OSCAR 3 system is comprised of a portable generator which has 2 output channels. Both channels 1 and 2 drive standard OSCAR cement removal handsets or OSCAR osteotome handsets. The channels can be activated via an air footswitch or via the handset switches. There is also a large range of cement removal probes, prosthesis removal probes, and a dedicated cleaning system.
The Generator The generator is mains powered and completely portable. It consists of an integrated power supply unit, enabling the generator to be powered from 240Vac mains, a control circuit, 2 power circuits and an LCD. The control circuit controls the 2 power circuits. A block diagram of the electrical system is shown in (Figure 1).
POWER SUPPLY UNIT
POWER CIRCUIT 1
Power supply
or
Drive Sgnal
FREQUENCY SYNTHESIS
Output
POWER CIRCUIT 2
CONTROL CIRCUIT Display Data
or
LCD DISPLAY
Feedback Handset Key Cement Removal Handset Osteotome Handset Figure 1 - OSCAR 3 Generator overview
EQUIPMENT
The two channel outputs running the handsets are microprocessor controlled and are designed to produce ultrasonic energy at the resonant frequency of the attached handset. The frequency range is 27.9 to 28.5 kilohertz. Careful control of the generator output ensures that the energy delivered to the handsets increases automatically in response to the increased load. In effect, as a probe is pushed into bone cement, (or bone for osteotomes) more energy is supplied to the handset - ranging from a few watts to a 150-watt output. In practice, the peak loads do not exceed 130 Watts when piercing solid cement or cutting bone.
Channel 1: Cement removal & Osteotome
LED channel status indicators
Large LCD illuminated display. System status, frequency and power meter displayed
Figure 2 - OSCAR 3 Front panel Channel 2: Cement removal & Osteotome Footswitch Air Nozzles
Once connected to the mains supply, the generator (Figure 2) is activated by the ON/OFF switch at the back on the LCD side of the enclosure. Both of the output channels have a 4 way socket for connection of OSCAR cables, which in turn connect to cement removal or osteotome handsets. Each channel also has an air nozzle socket, positioned below each output socket. The footswitch tubing connects to these air nozzles.
5
6
EQUIPMENT
The Handsets Cement Removal Handset Each handset contains a piezo-ceramic sandwich transducer - the ultrasonic “motor”, and a mechanical focusing device - the “horn” (B). These are enclosed in a water and pressure tight stainless steel casing which incorporates a small hand operated switch (A). This switch when depressed once and held will operate the handset; a click will be felt. The electrical connection is at the rear end of the handset (C). The shroud and end cap (D & E) are removed from the handset prior to autoclaving. A variety of probes may be fitted to the output end of the horn via a screwed coupling, and these permit the surgeon to carry out a range of procedures with maximum efficiency. Each probe consists of a waveguide, which is an integral number of half wave lengths long, its shape and overall length being chosen to suit a particular case. The signal from the generator induces the crystals within the handset to vibrate, producing ultrasonic energy. This energy is focused along the wave guide to the tip of the probe, which will vibrate at approximately 28.3 kHz. The rapid oscillatory movement produces heat when the tip of the probe is in contact with bone cement, causing the cement toliquefy almost immediately.
C
A
E
D
B
Figure 3 - Cement removal handset
Osteotome Handset In order to cut both cortical and cancellous bone the transducer which is used to generate axial vibration in the OSCAR handset has been redesigned so that its output impedance is compatible with bone tissue. This results in reduced displacement amplitude and the ability to sustain a higher distal load. Bone cutting handsets are identified by their gold outer sleeve case and larger size connecting thread at the output end. (See Figure 4 below).
Figure 4 - Osteotome handset
EQUIPMENT
The Probes All the probes in the OSCAR system are detachable from the main handsets. There are a multitude of probes available. Each handset type has its own range of unique probes. Each probe is attached to the main handset via a screw threaded connection. Once connected correctly the probe acts as a waveguide, focusing and directing the ultrasound energy generated within the handset to the probe tip or blade, enabling the handset to perform the task at hand, be it removal of bone cement or bone cutting.
Cement Removal The cement removal probes are for use with the Cement Removal Handset. (See Figure 3). The Cement Removal Handset can be run from either Channel I or II of the OSCAR 3 generator. There are many different cement removal probe designs. All probes are made of a titanium alloy, and screw into the distal part of the horn. They are tightened using the supplied 9mm spanner unless otherwise specified. Each probe is designed to perform a different function, grooving, scraping, piercing or cutting. Those probes designed to cut through regions of compact cement have heads incorporating two or more perforations or grooves. Friction between the bone cement and the high velocity tip surface of the ultrasonic probe causes rapid heating at points of contact. A small volume of cement melts, and under the advancing impetus of the probe tip this molten cement is forced through the ports in the head, or the grooves in the body solidifying behind it when the ultrasonic energy is turned off. Smoke is generated during the procedure. The dominant components are methylmethacrylate, benzine and styrene. All concentrations produced are well below OES/MEL values. (OES = occupational exposure standards, MEL = maximum exposure limits). Once broken up in this way, the cement can readily be removed, either by withdrawing the probe with adhering cement or using conventional instruments. Probes designed solely for scraping material from the endosteal surface of the bone are not equipped with perforations in the probe head. A description of individual probe designs follows.
7
8
EQUIPMENT
Consumable Probes Do not allow any consumable probe to come into contact with a metal surface whilst active, as this will cause damage to the probe. Consumable probes are reusable, however they will wear and become damaged after a number of uses. The number of uses varies depending on type and duration of use. As further probes are developed their status will be made clear as to whether they are re-usable (consumable) or single use.
The Groover
OHG2020 (Consumable) The groover (Figure 5) has a flattened spear-shaped head with one hole on each side, and a forward cutting blade at the most lateral aspect of the device. The groover is used primarily to cut longitudinal channels in the proximal cement mantle, in order to weaken the integral cement mass and permit segmental removal with conventional instruments. When the groover comes into contact with bone, a high pitched squeaking noise is heard, and resistance encountered. Note: this squeaking is a reaction between the probe and the bone and not generated from the module. This audible feedback cannot be relied upon if the bone is osteoporotic, dead or if some residual cement is still present to dampen the sound.
Figure 5 - The groover probe
The Axisymmetric Reverse Scraper
OHS2061, OHS2081, OHS2100 (Consumable) The scraper (Figure 6) is spear shaped, with no perforations within the head and is supplied with three tip diameters, 6mm, 8mm and 10mm. Note: the 10mm scraper has a short wave guide. The cutting edge is machined at an angle of 20 degrees to the axis of the probe, and it is used for removing well bonded proximal or distal cement and membrane from the endosteal surface by applying a reverse scraping action.
Figure 6 - Scraper probes
EQUIPMENT
Piercers
OHP2061, OHP2081, OHP2100, OHP2101, OHP2111, OHP2131 (Consumable) The piercer is a round, spear shaped device with 4 perforations in the head and is used for fenestrating the distal plug of cement to provide a clear channel, which can then be enlarged using the long scraper. The piercer is supplied in various tip diameters ranging from 6mm to 13mm (Figure 7). When the piercer strikes cortical bone there is a high pitched audible squeaking noise and resistance is felt. It should be noted that the piercer will de tune when it is in contact with bone, unless it is circumferentially compressed by bone, when de tuning will not occur but resistance will be felt and no smoke will be generated.
Figure 7 - Piercer probes
The Acetabular Probe OHA2030 (Consumable)
The Acetabular probe (Figure 8) is used to assist in the removal of the acetabular cup. It is important to remember that when using OSCAR, none of the activated instruments should be allowed to make contact with metallic components; this is particularly relevant when considering its application to the acetabular cup.
Figure 8 - Acetabular probe
9
10
EQUIPMENT
Single Use Probes Single use probes should not be reprocessed under any circumstances.
The Single Use Piercer OHP2080SU (Disposable)
The piercer (Figure 9) loses its cutting edge over time in the same way as the axisymmetric reverse scraper. In response to customer feedback, Orthofix supplies a single use piercer probe. The single use probe has an SU in its batch number etched onto its’ side to distinguish it from the reusable probe. In addition the wording ‘single use’ is etched onto the shaft. This probe also differs from the consumable piercers by being ½ wave-length, and requiring a 7mm spanner for fixation. It is extended using special extension bars, OHE2001SU and OHE2000SU, long and short respectively, and although they have the designation SU at the end of the identification, this does not signify they are single use. There is no wording ‘single use’ on the shaft.
Figure 9 - Single use piercer probe
The Single Use Axisymmetric Reverse Scraper OHS2080SU (Disposable)
Over time the cutting edge of the scraper probe loses its sharpness. Many customers expressed an interest in a cheaper, disposable probe that can be discarded after a single use so as to provide optimum cutting ability at all times. In response to this request Orthofix supplies a single use axisymmetric reverse scraper (Figure 10). Its size and mass are much reduced from the reusable probes, making them cheaper to purchase. The single use probe has an SU in its batch number etched onto its’ side to distinguish it from the reusable probe. In addition the wording ‘single use’ is etched onto the shaft. It is designed to be used with the special extension bars, OHE2001SU and OHE2000SU, long and short respectively, and although they have the designation SU at the end of the identification, this does not signify they are single use. There is no wording ‘single use’ on the shaft.
Figure 10 - Single use reverse scraper probe
EQUIPMENT
The Slim Shafted Piercer (OHP2062SU)
and Scraper
(OHS2062SU) (Disposable) Upper limb revision surgery involves removing cement from the thinner cortices and narrower medullae of the ulna and humerus. Two probes, one piercer and one scraper (Figure 11) each with a 6mm head, are available with 4mm shanks to allow good access. Due to the fragile nature of the slim shank these probes have been designated single use. Their clinical use is similar to the femoral piercers and scrapers, however, in accordance with safety instructions already mentioned, increased frequency of irrigation should be used when removing cement from the humerus and ulna.
Figure 11 - Slim shafted piercer probe / scraper probe
Inertial Probe Loading Instrument (Slaphammer) IPL200 (Consumable) and Extraction Probe EXP2681SU (Disposable)
This extraction probe (Figure 12) is designed to remove larger pieces of cement and the cement plug. However, the deployment of this probe is to be undertaken with care because the technique relies on the probe being embedded in cement, which is allowed to solidify. Instructions for use are located in the Surgical Technique section of this manual.
Figure 12 - Extraction probe Figure 13 - Slaphammer
11
12
EQUIPMENT
Osteotome Probes The osteotome probes are classed as SINGLE USE. They are for use with the osteotome handset (See Figure 4). Osteotome handsets can be run from Channels I & II of the OSCAR 3 generator. After use, the Osteotome probes should be disposed of as per normal operating room procedures, ideally in a suitably sized sharps container.
Flat Serrated Osteotome Probe (Disposable) Figure 14
Flat Non Serrated Osteotome Probe (Disposable) Figure 15
Curved Serrated Osteotome Probe (Disposable) Figure 16
Uncemented Stem Removal Fixation of uncemented stems is achieved by creating a porous coating on the implant or by adding hydroxyapatite to selected portions of the surface, both of which encourage bony ingrowth and generation of new cancellous bone. When revision is necessary, the implant is removed by cutting through what is a substantially cancellous bone interface surrounding the stem. This is done using a variety of flat, narrow osteotomes coupled to the specially designed osteotome handset. The essential feature of the osteotomes is a series of edge serrations forming a saw blade. The osteotomes are introduced beside the prosthesis stem to form a series of extended slots which effectively isolate the implant from the host bone. In some cases it may be necessary to cut into cortical bone when the depth of cancellous bone is limited.
16 15 14
Figure 14 - Flat Figure 15 - Flat non - serrated osteotome probe Figure 16 - Curved serrated osteotome probe
Hoe Probe
6mm OHH2062SU (Disposable) 8mm OHH2081SU (Disposable) This probe (Figure 17) is designed to remove small pieces of cement during upper extremity revisions. Similar to the reverse scraper, the hoe probe is used in a back scraping motion to remove cement in narrow canals.
Figure 17 - Hoe probes
EQUIPMENT
The Cables
(For use with Cement Removal and Osteotome Handsets) Silicone rubber cables with screened conduction wires are supplied with the equipment to connect the handsets to the generator. Each cable contains 2 separate ground leads for safety reasons. This means in effect, that the patient will be grounded when in contact with the device and this may cause hospital diathermy equipment to alarm and cease functioning, as this apparatus will not work when the patient is grounded. (Action: turn off the diathermy machine while OSCAR is being used) The sterile cable (Figure 18) is connected to the front of the generator, into channel I or II, by aligning the red dot on the end of the cable with the red dot on the generator input sockets. Once aligned the cable is then inserted until a small click is heard. The cable must be held by the boot when connecting. The same procedure is used to connect the cable to the handset. Please note the cables are interchangeable, it does not matter which end goes into the handset or generator. To remove the cable the grooved metal sleeve of the connector must be pulled backwards.
Figure 18 - Handset cable
Footswitch A double air footswitch (Figure 19) can be used on either of the output channels on the OSCAR 3 generator. Channels I & II utilise cement removal and osteotome handsets, which both have handset mounted switches, but can also be operated from the attached footswitch.
Figure 19 - Footswitch
13
14
OPERATIVE TECHNIQUE
OPERATIVE TECHNIQUE Setting Up Taking the required handsets (see Figure 3 & Figure 4), attach the required probes to the handset horn as shown in (Figure 20) seen opposite. Attach the required probe to the handset horn using suitable sterilised spanners. One on the handset horn and one on the probe. Do not use a single spanner to attach the probe as this may damage the handset.
Figure 20 - Attaching a probe to the handset
Connect the handset to either channel on the front panel of the generator. To connect the handset cable, line up the red dots on the cable connector and generator/handset connector and push. The connector will click into place (to disconnect pull on the front grooved part of the connector). If footswitch activation is required then connect the air tubes (note colour coding) from the footswitch to the relevant air nozzle on the front panel to channel I or II. (See Figure 21).
Handset connector fitted
Handset connector fitted Figure 21 - Channel I set up
First ensure that the device is plugged into mains voltage and that the switch on the back of the machine is in the ‘ON’ position.
Channel LED Indicator - Blue
When the generator is first switched on, the LCD will display SELECT A CHANNEL In order to select the required channel, simply depress the required handset switch or footswitch once on cement removal or osteotome handsets. Once selected, the generator will sound a beep and the appropriate channel LED will illuminate. The LCD will display which type of handset has been selected. (See Figure 22).
Handset type displayed On LCD Handset cable connected Figure 22 - LED - Channel ready
OPERATIVE TECHNIQUE
To activate the selected channel, press and hold either the handset switch or footswitch. When activated the relevant channel LED will turn yellow and the sounder will activate. The handset will tune to its resonant frequency. Only one channel can be active at any one time. To change channels, connect appropriate handset and depress the hand/foot switch once. The generator will now tune the handset to its resonant frequency every time the handset is operated. When the handset is activated the LED for that channel will turn yellow and the sounder will beep on / off as long as the handset is active. The frequency is displayed and the power bar graph indicates power delivery to handset. A countdown timer will commence from the moment the handset activates. (Maximum ON time is 30 seconds). When the countdown timer approaches zero, the sounder pitch will rise.
Channel I Activated Channel LED illuminates yellow When handset active
Figure 23 - LED - Channel active
The handsets will feel silky when light pressure is applied to the attached probe. It is most important not to squeeze the probe, as this will cause a friction burn to the tissues. It is equally important not to allow the probe to come into firm contact with the patient’s skin or muscle during operation. The shroud allows the surgeon to hold the handset with both hands in a comfortable and controllable manner. The patient’s tissue should be protected using a dry swab. As pressure is exerted on the handset the power output will increase and this is displayed via the power bargraph. If too much pressure is exerted, the cutting performance will decrease and an alarm will sound. This will occur when the power bargraph is on full deflection to the right. As the pressure on the handset relaxes the alarm will cease and cutting performance will improve.
The Display The LCD will display the handset frequency, the time left for activation in second and a power bargraph, illustrating loading of the handset. (As the mechanical load on the handset / probe is increased, the power delivered to the handset increases). With no load applied, ie, with the handset running in air, the bargraph should display one or two illuminated rectangles. This is not an instrument to measure the power. It is only for indication for the user.
Countdown timer Handset frequency
Power indicator Figure 24 - LED - Channel ready
15
16
OPERATIVE TECHNIQUE
Each channel LED indicator has the following meaning:
• Blue indicator illuminates when channel is selected. • Yellow indicator illuminates when handset is active. HANDSET NOT DETECTED
POOR FEEDBACK RELEASE ALL SWITCHES
ERROR NO LOCK
OVERTEMP
FREQUENCY TOO LOW
ACTIVE TOO LONG
CEMENT RELEASE MODE
The handset type cannot be read by the generator. Return the handset for service. This means too much pressure has been exerted on the handset as the handset is switched on. The channel will automatically reset after 4 secs. If this message repeatedly occurs even when running the handset in air, it is likely that the probe has fatigued and should be replaced. If the message remains then the handset has failed. This means the generator could not find a handset resonance in the correct frequency range. The generator will automatically reset. This means the power devices within the channel power circuit have become too hot and to avoid damage the channel has shut down. The devices may overheat through extended periods of heavy loading. Allow the channel power circuit to cool for several minutes. The generator will reset the channel. If the ‘overtemp’ message continues even after 10minutes have passed, and repeated resets do not clear the message, then the power devices may have failed on that channel output. When the handset is running it will naturally warm up and its resonant frequency will drop. After prolonged periods of use the frequency may drift out of the usable range. Allow the handset to cool. The countdown timer has timed out…the maximum time the handset can be run continuously is 30 seconds. This is to guard against the power devices overheating. This is most likely to occur when a piercer is being used. If the tip of the piercer is pushed too deeply into the cement and allowed to remain in situ, then the cement behind the tip is liable to solidify and trap the probe. If this happens, the load on the probe may be excessive preventing normal resonance because of the mass of cement now attached to the probe. If this occurs the generator will automatically enter a ‘stuck in cement’ mode and display CEMENT RELEASE MODE
Energise the handset, but do not apply any pressure on the handset for the first 2 seconds of use (this gives the channel a better chance to find a resonant feature). The generator will perform repeated scans of the handset, sounding a double beep as it does so. Whilst this occurs, gently remove the probe from the cement. When free from cement the channel mode will return to normal operation. This procedure may lead to an OVER-TEMPERATURE message on the LCD. Allow the channel to cool for several minutes. When the channel resets attempt removal again. This process may take a couple of attempts, depending how deep the probe is stuck in the cement. Should an individual channel fail for any reason, the remaining channels will remain operational. When convenient return the system for servicing to Orthofix Srl.
OPERATIVE TECHNIQUE
Probe Use CEMENT REMOVAL The following description relates to hip surgery but is equally applicable to other forms of revision Arthroplasty. It is essential that the patient be placed on his/her side to facilitate cement removal. Following removal of the femoral component it is important to have adequate access to the cement. To this end it is worthwhile removing bone from the upper portion of the greater trochanter to allow straight access to the cement within the femoral canal.
Removal of the Proximal Cement Mantle The proximal cement in the upper third of the shaft is usually loose, with a membrane between it and the bone. This cement can be removed simply, using the groover to cut longitudinal channels through it. When the cutting edge of the groover is placed in contact with the cement, the friction generated by the ultrasonic beam will heat up the cement, which softens and extrudes through the holes in the side flanges (Figure 25). The groover is then pushed gently down the femoral canal in contact with bone to remove a trough of cement. This trough can be extended within the femoral canal as far as the groover can go. It is usual to make 3 longitudinal troughs within the cement at approximately 120-degree intervals. The amount of force required is similar to that needed push a knife through hard butter. Once the 3 longitudinal grooves have been made, it is necessary to make circumferential grooves at approximately 1.5 to 2 cm intervals down the cement mantle. This can be done using either the groover or a back scraper. The fragments of bone cement can then be folded into the medulla and removed with forceps. The proximal cement is removed incrementally, using this technique, down to the level of the cement plug.
Figure 25 - Removal of proximal cement mantle
17
18
OPERATIVE TECHNIQUE
Removal of Well Bonded Cement and Residual Membrane Any residual well bonded cement which remains after removal of the proximal cement mantle, and the fibrous membrane which is usually adherent to the endosteal surface, may be removed using the axisymmetric reverse scraper (back scraper). The back scraper cuts by focusing ultrasonic energy on to the retroverted cutting edge at the back of its spear shaped tip. The scraper can also be used to remove the proximal cement if the mantle is too thin for the groover. The cutting edge is put into contact with the cement or the membrane and withdrawn with a moderate amount of force for removing cement, or just gentle pressure for removing the membrane. The membrane will peel off in strips leaving a raw cancellous surface which is ideal for bone grafting should this be required.
Figure 26
Removal of the Cement Plug Removal of the cement plug is performed using the piercer. The piercer focuses ultrasound to the very tip of its spear shaped head causing the cement to liquefy at the point of contact. The liquid cement then flows backwards through the perforations in the head and solidifies behind it. The piercer is gently pushed into the cement plug, and after advancing it for between 1.5 and 2 cm the switch on the handset is released and the handset is removed 1 or 2 seconds later. This short period of time allows the cement to solidify behind the flange for most efficient removal. (Figures 26 & 27). If the piercer is pushed with too much force, the energy at the tip will reduce and the system will work less efficiently. The cement removed using the piercer can be simply wiped from the waveguide using a wet swab, after each application. Should the tip of the piercer come in contact with bone, the waveguide will de tune slightly and produce an audible high pitched squeaking noise. Resistance will be felt and the volume of smoke generated will diminish.
Figure 27
OPERATIVE TECHNIQUE
Removal of the Cement from the Distal Femur Once the plug has been perforated, it is necessary to remove the cement that is in contact with the distal femur and this can be done either with the groover or, if the femur is too narrow, with a long back scraper. (Figure 28a - 28b) There is sometimes a polythene cement restrictor distal to the cement plug. The ultrasonic device will cut through polythene, but it does so at a slower rate than when cutting through bone cement.
Back Scraper
In these circumstances it is necessary to wait for a few seconds before advancing the probe, as polyethylene absorbs more ultrasonic energy than polymethylmethacrylate.
Cement
Figure 28a
Use of Extraction Probe For rapid removal of the cement plug and under certain conditions, the OSCAR system incorporates an extraction probe and slap hammer. The extraction probe, like the piercer probes, is designed to penetrate the cement in a forward motion. The surgeon sinks the probe into the plug up to a reasonable depth and then rotates the probe and handset 90 degrees. By removing the power and holding the probe insitu in the canal for 30 seconds, the cement re-hardens around the probe. The handset can then be detached and the slap hammer connected. Do not place excessive force on the slap hammer when in use. X-rays will give good indication of when the extraction probe can be used. Ensure that there is good bone stock, delamination of the cement, an even tapered canal, and the cement lies above any isthmus. Incorrect positioning of the probe can easily lead to bone fracture, or other damage within the canal.
Figure 28b
19
20
OPERATIVE TECHNIQUE
Removal of Cement from the Acetabular Cup It is important to remember that when using OSCAR, none of the activated instruments should be allowed to make contact with metallic components; this is particularly relevant when considering its application to the acetabular cup. By reference to the x-rays, identify and locate all metallic parts associated with the prosthesis. The acetabular probe can be used to remove cement posterior to the cup by making a series of curved slots extending from the exposed circular profile towards the centre of the cemented hemispherical surface (Figure 29). Four or five slots should be sufficient to loosen the acetabular component, which may then be removed by careful use of a cement chisel.
Bone
Cement
Polyethylene Cup
Figure 29: Removal of cement from acetabular cup