Possis Medical, Inc
AngioJet Rheolytic Thrombectomy System
AngioJet Ultra Console Operations Manual Rev 4
Operations Manual
16 Pages
Preview
Page 1
System Console Operations Manual
CONSOLE OPERATIONS MANUAL
TABLE OF CONTENTS 1.
DEVICE DESCRIPTION... 2 The Console The Thrombectomy Set
2.
INDICATIONS AND USAGE... 3
3.
CONTRAINDICATIONS... 3
4.
WARNINGS and PRECAUTIONS... 3
5.
ADVERSE EVENTS... 4
6.
CONSOLE SPECIFICATIONS... 5
7.
CLINICIAN USE INFORMATION... 6 7.1 Prepare Console... 7 7.2 Load Pump... 8 7.3 Prime the Catheter... 9 7.4 Ultra System Dismantling... 11 7.5 Alarms and Error Messages... 12 7.6 Ultra System Errors... 12 7.7 Maintenance... 12 7.8 Glossary... 13 7.9 Symbols... 14
8.
HOW SUPPLIED... 14
1
CAUTION: Federal (USA) Law restricts this device to sale by or on the order of a physician.
1. DEVICE DESCRIPTION
This manual provides instructions for operating the AngioJet Ultra Thrombectomy System, with specific focus on the AngioJet Ultra Console. The AngioJet Ultra System consists of the following components:
A. The AngioJet Ultra Console (Console)
Control Panel
Drawer Open/Close Button Pump Drawer Roller Pump
Foot Pedal Bracket Main Power Circuit Breaker Switch Power Cord
Foot Pedal
Back View
Front View Figure 1 - AngioJet Ultra Console
The Console is a multiple-use device that controls the Thrombectomy Set. It drives the pump, regulates fluid inflow and outflow, provides the operator with Ultra System set-up prompts, total infused saline volume, and Ultra System malfunction warnings. The Console is activated by pressing a foot pedal.
B. The AngioJet Ultra Thrombectomy Set (Thrombectomy Set - Several models available) The single-use, disposable Thrombectomy Set includes : •
Pump
•
Catheter
•
Saline delivery and waste tubing
•
Collection bag
Pump
Collection bag
Delivery and waste tubing
Catheter
Figure 2 - Thrombectomy Set
2
A bag of sterile, heparinized saline (not included) supplies the pump with saline through the saline delivery tubing. The pump pressurizes the saline. The Thrombectomy Set uses this pressurized, high-velocity saline to create a low-pressure zone at the catheter tip. This allows the catheter to break up and remove thrombus. The waste tubing transports the thrombus debris from the catheter to the collection bag for ultimate disposal.
2. INDICATIONS AND USAGE The Console is intended for use only in conjunction with an AngioJet Ultra Thrombectomy Set. Refer to the individual Thrombectomy Set Information for Use manual for specific clinical applications.
3. CONTRAINDICATIONS Refer to the individual Thrombectomy Set Information for Use manual for specific contraindications.
4. WARNINGS AND • PRECAUTIONS
The AngioJet Ultra System should be used only by operators who have received appropriate training on its installation and use.
•
Use the AngioJet Ultra Console only with an AngioJet Ultra Thrombectomy Set. This Console will not operate with a previous model pump set and catheter.
•
Do not attempt to bypass any of the Console safety features.
•
If the catheter is removed from the patient and/or is inoperative, the waste tubing lumen, guide catheter, and sheath should be flushed with sterile, heparinized solution to avoid thrombus formation and maintain lumen patency. Reprime the catheter by submerging the tip in sterile, heparinized solution and operating it for at least 20 seconds before reintroduction to the patient.
•
Physician discretion with regard to the use of heparin is advised.
•
Refer to the individual AngioJet Ultra Thrombectomy Set Information for Use manual for specific warnings and precautions.
•
Do not move the collection bag during catheter operation as this may cause a collection bag error.
•
Monitor thrombotic debris/fluid flow exiting the catheter through the waste tubing during use. If blood is not visible during console activation, the catheter may be occlusive within the vessel or the outflow lumen may be blocked. Ensure adequate patient anticoagulation to prevent thrombus formation in outflow lumen.
• •
Refer to individual Thrombectomy Set Instructions for Use manual for specific instructions regarding heparinization of the Thrombectomy Set. 3
5. ADVERSE EVENTS
4
•
The Console contains no user-serviceable parts. Refer service to qualified personnel.
•
Removal of outer covers may result in electrical shock.
•
This device may cause electromagnetic interference with other devices when in use. Do not place Console near sensitive equipment when operating.
•
Equipment not suitable for use in the presence of flammable anesthetic mixture with air or with oxygen or nitrous oxide.
Refer to the individual Thrombectomy Set Information for Use manual for specific observed and/or potential adverse events.
6. CONSOLE SPECIFICATIONS
Model number
5000A
Dimensions DxWxH
25” x 16.5” x 54” (63.5 x 42 x 137 cm)
Weight
135 lbs (57 kg)
Voltage requirements
100/120/220/240
Frequency requirements
50/60 Hz
Power requirements
900VA
Logic power backup outage
60 seconds for conditions of power loss
Equipment class
Class 1
Degree of protection against electrical shock
Defibrillator proof type CF equipment
Enclosure protection against ingress of liquid
IPX1
Foot pedal protection against ingress of liquid
IPX8
Mode of (electrical) operation
Continuous
Transport & storage temperature
– 25C to 57C
Transport & storage humidity
10% to 100%
Transport & storage atmospheric pressure
500hPa to 1060hPa
Operation temperature
10C to 40C
Operation humidity
30% to 75%
Operation atmospheric pressure
700hPa to 1060hPa
Fuse
100/120V, 10A, 250V, 5X20 mm Time lag 220/240V, 6.3A, 250V, 5X20 mm Time lag
NOTE: Replace fuse with the type and rating specified. Failure to do so may result in device damage or risk of fire. The AngioJet Ultra Console is compliant with EN60601-1, UL60601-1, and CAN/CSA-C22.2 No. 601-1-M90.
5
7. CLINICIAN USE INFORMATION
A thorough understanding of the Ultra System components is required for proper operation. Read this manual and the Information for Use supplied with the Thrombectomy Set before attempting to use any of the components of the Ultra System. Improper Ultra System preparation or abnormal component operation will result in error messages. Please refer to Section 7.6 - Alarms and Error Messages of this manual for instructions on what to do if error messages are displayed. Time Display
Counter Reset Button
Icons
Status Panel
Scroll Buttons
Alarm Reset Button
Power Button
Catheter Identification Button
Figure 3 – Control Panel with Labels
Time Display: During catheter priming, the time display counts down to zero. During the procedure, the time display counts up from zero. Icons: Display the progress of set-up and operation steps and also indicate when an alarm or Ultra System error has occurred. -Please Wait. -Install Pump. -Connect Saline. -Prime Catheter. -Ultra System Ready. -Contact Possis Technical Support.
6
Status Panel: Provides instructions, procedure status, and alarm resolution strategies. Power Button: Press to activate the control panel. Alarm Reset Button: Press to clear or override an alarm. Catheter Button: Displays catheter model installed in the AngioJet Ultra Console. Press this button to see a 3-second display of catheter model. Scroll Buttons: These buttons are active only when there is more text to be displayed than can fit on the status panel screen. Press the DOWN button to advance one screen forward; press the UP button to return to the previous screen. Counter Reset Button: Press this button to return the time display and infused volume to zero. NOTE: You cannot use the COUNTER RESET Button to override the time needed for priming the catheter.
7.1 Prepare Console
NOTE: The Ultra System is designed to be interactive. The status panel will provide prompts to guide the technician through set-up as well as provide error resolution when necessary. Preparation of the Ultra System requires the assistance of a sterile and a nonsterile technician. The catheter is used within the defined sterile field, the Console and pump are operated outside the sterile field. The following directions are for the nonsterile technician except where otherwise noted.
1 Plug in the Console and ensure the main power circuit breaker Main Power Circuit Breaker Switch switch is turned ON (Figure 4).
2 Heparinize a bag of sterile, room-temperature saline at a suggested rate of 5000 units per liter of saline and mix contents (a 1.0 liter bag is recommended, but is not included with the Ultra System). Hang the saline bag on the saline bag hook at the top of the Console.
Figure 4
3 Press the POWER button on the control panel.
7
All indicators on the control panel will illuminate. The status panel will display ANGIOJET ULTRA while the Console performs a self-test. The drawer will open, indicating a successful selftest. The status panel will display the next step (Figure 5).
Figure 5
7.2 Load Pump 1 Sterile technician: Remove the catheter and sufficient tubing for ease of use from the sterile package and inspect for damage. Hand the tray with the rest of the Thrombectomy Set to the nonsterile technician for installation into the Console (Figure 6). NOTE: If a clamp is used to keep the catheter in the sterile field, ensure that the clamp does not deform the tubing.
Figure 6
2 Nonsterile technician: Remove the pump (not the piston head) from the tray and insert the pump into the Console. Ensure that the waste tubing aligns with the roller pump (Figure 7). Roller Pump Waste Tubing Delivery Tubing
3 Remove the cap from the Thrombectomy Set bag spike and insert the spike into the saline bag.
Figure 7
4 Push the drawer button to close the Console drawer (Figure 8).
Figure 8
8
The Console will load the pump and process information from a bar code located on the pump. After the Console successfully identifies the catheter, the status panel will display the model of catheter in use (Figure 9). The Console will automatically prime the pump.
5 Place the foot pedal within easy access of the Figure 9
physician. Choose a location that will minimize accidental activation.
7.3 Prime the Catheter The status panel displays PRIME indicating that the initial pump prime was successful (Figure 10). Catheter priming time is determined by the catheter model and is automatically set by the Console.
Figure 10
1 Prime the catheter by completely submerging
Figure 11
the tip in heparinized saline and pressing the foot pedal (Figures 11 & 12).
Figure 12
9
2 Continue priming until the time display reaches zero seconds.
The status panel displays PRIME COMPLETE (Figure 13).
3 Confirm Ultra System set-up is successfully completed by removing foot Figure 13
from foot pedal.
Status panel displays READY and green icon is illuminated (Figure 14).
The AngioJet Ultra System is now ready to use.
Figure 14
During operation, the infused volume will be displayed on the status panel, the green icon remains lit, and the time display will keep track of the total time that the foot pedal is activated.
CAUTION: Do not move the collection bag during catheter operation as this may cause a collection bag error.
10
7.4 Ultra System Dismantling
Follow proper precautions for the handling of infectious waste. Reuse of Thrombectomy Set is prohibited due to risk of contamination by blood waste products. After use, disassemble the components as follows:
1 Push the drawer button to open the drawer. 2 Carefully remove Thrombectomy Set from Console. 3 Unhook the saline supply and collection bag and dispose together with the Thrombectomy Set.
4 Press POWER button to deactivate control panel. Pump drawer will automatically retract when powering down.
5 Unplug the power cord from the wall outlet and coil the cord on the hooks in the power cord and foot pedal recess. Place the foot pedal into the storage bracket at the bottom of the recess.
6 Clean the Console surfaces thoroughly with a standard, mild germicidal cleaning agent. Do not clean with harsh detergent or chemical agents.
NOTE: Check bar code window for saline build-up, clean with long cotton-tipped swab and water, if necessary (Figure 15).
Figure 15
11
7.5 Alarms and Error Messages
Alarms and error messages indicate improper Ultra System preparation or abnormal component operation.
The Console status panel will display alarm resolution messages and prompts (Example shown in Figure 16). In most instances, alarms will either adjust themselves or prompt the user to repeat a step. Follow the prompts displayed on the status panel for alarm resolution. Figure 16 - Example of alarm
REPLACE THROMBECTOMY SET
The Console allows multiple attempts to correct most alarm conditions. If the alarm persists beyond the third attempt, it is most likely that the Thrombectomy Set has a defect and the Console will prompt the user to replace the Thrombectomy Set (Figure 17).
Figure 17 - Example
7.6 Ultra System Errors
A system failure will result in a system error (wrench icon lights red, shown in Figure 18). Turn off the power, then restart before contacting Possis Medical Technical Service for further instructions. It is helpful to make a note of each system error title and number to report to Possis Technical Service.
Figure 18 - Example
7.7 Maintenance
12
Refer to your Limited Warranty and Disclaimer and/or Certificate of Extended Warranty, if applicable, for information on servicing the AngioJet Ultra Console. Possis Medical recommends annual inspection and calibration. There are no user serviceable parts inside the Console. The Console cabinet panels should only be opened by trained service personnel.
7.8 Glossary Alarm: A recoverable or nonrecoverable fault occurring when one of the Ultra System safety sensors responds to abnormal operation of a component. Waste Tubing: The tubing which is inserted in the roller pump and transports the extracted thrombus from the effluent tubing to the collection bag. Collection Bag: The bag which collects the extracted thrombus being removed by the Catheter. The bag hangs on the Console Control Panel: The operator interface on the upper front of the Console. Error: A nonrecoverable fault which has occurred because of Thrombectomy Set or Console malfunction. Saline Delivery Tubing: The tubing which transports saline from the saline bag to the pump. Load Pump: An indicator instructing the operator to install the pump into the Console. Over Pressure Alarm: A recoverable condition occurring when an abnormally high pressure has been detected. Replacement of the Thrombectomy Set may be necessary. Thrombectomy Set: The disposable component of the Ultra System which delivers pressurized saline to the catheter and removes debris from the catheter. It consists of the saline bag spike, saline delivery tubing, pump, catheter, waste tubing, and collection bag. Self Test: An operation performed by the Console to examine the fidelity and the state of its circuit paths and sensors. An appropriate indicator will be illuminated if any abnormality is detected. Total Infused Volume: The numerical display on the control panel which displays the total volume of saline infused in cc. Under Pressure Alarm: A recoverable condition occurring when an abnormally low pressure has been detected.
13
7.9 Symbols
8. HOW SUPPLIED
No.
Symbol Description (one or more of these symbols may appear on the Console)
1
ON (power: connection to mains)
2
OFF (power: disconnection to mains)
3
Standby
4
Eject (Open/close drawer)
5
Attention, see Operations Manual
6
Defibrillator proof type CF equipment
7
IPX 1
Protection against dripping water
8
IPX 8
Protection against the effects of permanent immersion in water
9
Alternating current
10
Equipotentiality
11
Fuse
12
Foot Pedal
13
Electrical and electronic equipment should not be disposed of in the normal waste stream
NON-STERILE, MULTIPLE USE CONTENTS:
One (1) Console One (1) Console Operations Manual One (1) Foot pedal with cord One (1) Power cord
14
Possis Technical Support 1-888-848-7677
All information, drawings, and ideas herein constitute a confidential disclosure of POSSIS MEDICAL INC. and are its exclusive property, and must not, without written authorization, be divulged, reproduced, copied or appropriated in whole or in part for any purpose.
Warranty:
Refer to your Limited Warranty and Disclaimer and/or Certificate of Extended Warranty, if applicable.
AngioJet, is a registered trademark of Possis Medical, Inc. The AngioJet Ultra System includes several catheter models, which are variously marketed for thrombectomy of AV access conduits, peripheral arteries and veins, coronary arteries, and/or saphenous vein grafts. See Information For Use manuals for specific and complete prescribing information. ©2006 Possis Medical, Inc.
Medical Equipment With respect to Electric Shock, Fire, and Mechanical Hazards In accordance with UL 60601-1, CAN/CSA C22.2 No. 601.1, IEC 60601-1, and EN 60601-1
Manufactured and Distributed by: Possis Medical, Inc. 9055 Evergreen Boulevard NW Minneapolis, MN 55433-8003 USA tel +1.763.780.4555 toll free +1.800.810.7677 fax +1.763.780.2227 Authorized European Representative: CEpartner4U BV Esdoornlaan 13 3951 DB Maarn The Netherlands tel +31 343 442 524 mobile +31 6 516 536 26 fax +31 343 442 162
105603-001 04