Quick Reference Guide
16 Pages
Preview
Page 1
Puritan Bennett™ 980 Ventilator Patient Setup Quick Reference Guide NEW PATIENT COMPLETE SETUP† To set up a new patient: • Open the Wye connector of the breathing circuit (the ventilator has a patient detection system). • Power the ventilator on. • Touch the New Patient button. (see figure 1) • Select gender and height (once selected, use the knob to adjust). (see figure 2) • Select Ventilation type • Select Mode • Select Mandatory or Spontaneous Type • Select Trigger Type • Select Primary Settings (see figure 3) –– Rate –– Tidal Volume –– Pressure –– %O2
(figure 1)
–– PEEP • Press Start when done.
“SAME PATIENT” SETUP†
(figure 2)
To set up using Same Patient: • Open the Wye connector of the breathing circuit (the ventilator has a patient detection system). • Power the ventilator on. • Touch the Same Patient button. (see figure 1) Select Start. (see figure 3)
(figure 3)
QUICK START† This feature uses default values or institutionally configured breath delivery settings to ventilate the patient after predicted body weight (PBW) has been entered. • Open the Wye connector of the breathing circuit (the ventilator has a patient detection system). • Power the ventilator on. • Touch the New Patient button. • Select gender and height. • Touch Quick Start. (see figure 4)
(figure 4)
SHORT SELF TEST (SST) This test checks circuit integrity, calculates circuit compliance and filter resistance, and checks ventilator function.
To run a short self test: • Enter through the first power-up screen. (see figure 1) • Select SST • Select Run all SST. (see figure 5)
(figure 5)
• Choose: –– Circuit type (Adult, Ped, Neo) (see figure 6) –– Humidifier type –– Humidifier volume –– Follow the prompts to completion. • Select Accept. * Adult/Pediatric ventilators have a specific exhalation filter and door. ** Neonate ventilators have a specific exhalation filter and door. †
(figure 6)
Short Self Test (SST) shall be conducted in accordance to instructions in the Operator’s Manual. This guide is provided as a convenience companion document to the Operator’s Manual. It is not intended to replace the Operator’s Manual, which should always be available while using the ventilator. It is important to familiarize yourself with all information in the Operator’s Manual relevant to your institution’s use of the ventilator, including on-screen helps, instructions, warnings and cautions. COVIDIEN, COVIDIEN with logo, Covidien logo and positive results for life are U.S. and internationally registered trademarks of Covidien AG. Other brands are trademarks of a Covidien company. ©2014 Covidien. 13-VE-0089(1)a
6135 Gunbarrel Avenue Boulder, CO 80301 800-635-5267 www.covidien.com/rms
Puritan Bennett™ 980 Ventilator Touch Screen Quick Reference Guide Finding your way around and how to use the touch screen Icons Home Touch this icon to close all open dialogs on the touch screen and return to the main screen. Alarm Touch this icon to display the alarm settings screen. Configure Touch this icon to display the configure screen. Logs Touch this icon to display the logs screen, which contains tabs for Alarms, Settings, Patient Data, Diagnostics, EST/SST status, General Event, and Service logs. 1
2
3
4
5
6
Bezel Keys 1. Display brightness key
6. Inspiratory pause key
2. Display lock key
7. Expiratory pause key
3. Alarm volume key 4. Manual inspiration key 5. Rotary encoder (knob)
8. Alarm reset key 9. Alarm silence key
7
8
9
Elevate O2 Touch this icon to configure and increase the oxygen concentration for two minutes Screen Capture Touch this icon to capture and store the image displayed on the touch screen. Help Drag this icon to the item in question and release. A tooltip will appear describing the item’s function.
How to Use the Touch Screen Action
Used For
How to Use
Swiping
Opening or closing dialogs or panels, moving waveform data, expanding or collapsing tooltips, scrolling lists, or alarm banners
Swipe toward the center of the screen to open dialogs or panels. Swipe toward the side of the screen (or upward if viewing the additional patient data or large-font patient data panels) to close. A downward swipe anywhere in the patient data area opens the additional patient data panel, and another swipe on the additional patient data tab displays the large font patient data panel.
Touch and hold
Displaying a tooltip dialog on whatever item is touched; the tooltip appears to glow indicating the touch and hold action
Touch an item and hold for at least 0.5 seconds.
Double touch
Maximizing or minimizing the viewable area of a dialog, control, or waveform, expanding or collapsing tooltips
Double-tapping maximizes the viewable waveform area. Double-tapping again min-imizes the viewable waveform area.
Drag
Gaining information about a particular screen parameter
Drag the help icon, located at the lower right of the touch screen, to the item in question and drop. If a blue glow appears, a tooltip is available.
Sweep
Changing x- and y-axis scales, moving the waveform cursor, moving scroll-bars, scrolling lists; scrolling speed varies depending upon how far outside the list boundary the finger is positioned
Touch the axis and drag to the right to increase the waveform x-axis scale, and to the left to decrease. Use similar motions with the y-axis but drag up and down. To move the cursor (when the waveform is paused), touch the cursor and drag it right or left.
This guide is provided as a convenience companion document to the Operator’s Manual. It is not intended to replace the Operator’s Manual, which should always be available while using the ventilator. It is important to familiarize yourself with all information in the Operator’s Manual relevant to your institution’s use of the ventilator, including on-screen helps, instructions, warnings and cautions. COVIDIEN, COVIDIEN with logo, Covidien logo and positive results for life are U.S. and internationally registered trademarks of Covidien AG. Other brands are trademarks of a Covidien company. ©2014 Covidien. 13-VE-0089(1)b
6135 Gunbarrel Avenue Boulder, CO 80301 800-635-5267 www.covidien.com/rms
Puritan Bennett™ 980 Ventilator Leak Sync Software Quick Reference Guide WHAT’S NEW Puritan Bennett™ Leak Sync software now works with VC+ and VS * (compared to Leak Compensation software on the Puritan Bennett 840 ventilator).
KEY FEATURES The Leak Sync software automatically detects and compensates for leaks in the breathing circuit. This unique software is designed to differentiate between flow due to leaks and flow due to patient respiratory effort and then to adjust quickly.
(figure 1)
LEAK SYNC SOFTWARE • Can be used with invasive and non invasive ventilation • Maximum leak compensation flow: –– Neo 15 L/min –– Ped 40 L/min –– Adult 65 L/min • Is automatically enabled when the ventilator type is NIV or if New patient is selected and the circuit type is Neonate • When the Leak Sync software is active, LS is displayed in the lower left-hand side.
INITIAL SETUP • Select the Vent Setup button (lower left) (see figure 2) • Select More Settings • Select Enabled or Disabled for Leak Sync • Select Accept All • Leak Sync software typically adjusts for a leak present within three (3) breaths
(figure 2)
VALUES DISPLAYED When Leak Sync software is enabled the following values will be displayed in the patient data cells. • VLEAK Inspiratory leak volume, the total volume delivered during inspiration to compensate for the leak. • %LEAK Percent leak, the percentage of total delivered volume during inspiration attributed to the leak over total delivered inspiratory volume • VTL The volume inspired for each breath when Leak Sync is enabled. • LEAK Exhalation leak, the leak rate during exhalation at PEEP (see figure 3)
(figure 3)
KEY ALARMS DSENS displays L/min instead of % when the Leak Sync software is enabled. (see figures 4 and 5) (figure 4)
(figure 5)
This guide is provided as a convenience companion document to the Operator’s Manual. It is not intended to replace the Operator’s Manual, which should always be available while using the ventilator. It is important to familiarize yourself with all information in the Operator’s Manual relevant to your institution’s use of the ventilator, including on-screen helps, instructions, warnings and cautions. COVIDIEN, COVIDIEN with logo, Covidien logo and positive results for life are U.S. and internationally registered trademarks of Covidien AG. Other brands are trademarks of a Covidien company. ©2014 Covidien. 13-VE-0089(1)c
6135 Gunbarrel Avenue Boulder, CO 80301 800-635-5267 www.covidien.com/rms
Puritan Bennett™ 980 Ventilator PAV™*+ Software Quick Reference Guide KEY FEATURES The PAV™*+ software provides ventilatory support proportional to the patient’s inspiratory effort, letting the patient direct the duration and depth of each breath, enabling: • Variable flow • Variable volume • Pressure delivery proportional to patient demand
KEY VARIABLES The PAV™*+ software uses these key variables to determine patient need. (see figure 1) Flow, Volume, Compliance, Resistance and %Support
INTENDED FOR PATIENTS WITH • Intact Respiratory Drive • Predicated body weight (PBW) 25 kg or greater • Tube ID 6.0 or greater
(figure 1)
TO SET UP PAV™*+ SOFTWARE (see figure 2)
• At the ventilator setup screen, enter the patient’s gender and height or predicted body weight (PBW). • Select Invasive vent type. • Select SPONT mode. • Select PAV+ to select Spontaneous type. • Select the desired trigger type (P-TRIG, V-TRIG). • Select tube type. • Select the Tube ID. Initially, a default value is shown based on the PBW entered at ventilator startup. If this Tube ID is not correct for the airway in use, turn the knob to adjust the Tube ID setting.
(figure 2)
• Select Start.
PAV™*+ MONITORING Graphics Screen
(see figure 3) The PAV+ screen will display: • WOBPT, WOBTOT, CPAV, RPAV, PEEPI, RTOT Data Screen displays: • PAV-based lung compliance (CPAV) • PAV-based lung elastance (EPAV) • PAV-based respiratory resistance (RPAV) • PAV-based total airway resistance (RTOT) • Inspired tidal volume (VTI)
(figure 3)
HOW TO CALCULATE PMUS,PEAK AND ADJUST % SUPPORT ON THE BASIS OF PMUS,PEAK A simple algorithm was designed to adjust the %Support during PAV™*+ mode in order to target a reasonable and predefined range of respiratory muscle pressure.1 • A n estimate of Pmus,Peak, was calculated based on this equation (Paw,Peak-PEEP) x [(100-%Support)/%Support] • Paw,Peak (peak airway pressure) values are obtained from the ventilator • Example of the calculation of Pmuspeak: • Pawpeak = 20, PEEP 5, %Support = 70 • Pmuspeak = (20-5) x ((100-70)/70) = 6.43 • The %Support given by PAV™*+ mode was adjusted to maintain Pmus,Peak between 5 and 10 cmH2O1
KEY ALARMS • High circuit pressure (High PPEAK) • PAV startup too long • PAV R&C not assessed • High VTI
NOTE • Not intended for leak environments • Not intended for NIV Refer to the Proportional Assist™ Ventilation (PAV™+) Management Protocol guide for additional information. (10-VE-5543, VE19408)
1. Carteaux G, Mancebo J, Mercat A, et al. Bedside Adjustment of Proportional Assist Ventilation to Target a Predefined Range of Respiratory Effort. Crit Care Med. 2013;41(9):2125-2132. This guide is provided as a convenience companion document to the Operator’s Manual. It is not intended to replace the Operator’s Manual, which should always be available while using the ventilator. It is important to familiarize yourself with all information in the Operator’s Manual relevant to your institution’s use of the ventilator, including on-screen helps, instructions, warnings and cautions. COVIDIEN, COVIDIEN with logo, Covidien logo and positive results for life are U.S. and internationally registered trademarks of Covidien AG. *Proportional Assist and PAV are registered trademarks of The University of Manitoba, Canada. Other brands are trademarks of a Covidien company. ©2014 Covidien. 13-VE-0089(1)d
6135 Gunbarrel Avenue Boulder, CO 80301 800-635-5267 www.covidien.com/rms
Puritan Bennett™ 980 Ventilator Volume Control Plus (VC+) and Volume Support (VS) Quick Reference Guide KEY FEATURES Automatically adjusts inspiratory target pressure from breath to breath in response to changes in the delivered tidal volume. • Variable flow (see figure 1) • Targeted volume • Variable pressure • Breath-by-breath automatic adjustment in delivered pressure based on inspired volume
INITIAL SETUP
(figure 1)
Patient set up: VC+ (see figure 2)
• Touch the Vent Setup button • Touch Invasive • Touch VC+ • Primary setting: –– Set the frequency (except for VS) (see figure 3) –– Set tidal volume, even if it was set in a previous breath type –– Set inspiratory time • Set rise time % • Key alarms –– PPEAK –– High Inspired spontaneous and mandatory tidal volume –– Low circuit pressure
(figure 2)
LIMITATIONS OR CONTRAINDICATIONS • High effort may result in the regulating of pressure down to unacceptable levels
KEY ALARMS • Volume Not Delivered (High Peak Circuit Pressure) • High Inspired Tidal Volume alarm limit • The High Inspired Tidal Volume limit terminates inspiration and commences exhalation during VC+, VS
Low Circuit Pressure WARNING: Because the VC+ pressure control algorithm does not allow the target inspiratory pressure to fall below PEEP + 3 cmH2O, attempting to set the PPEAK alarm limit at or below this level will turn the alarm off.
(figure 3)
This guide is provided as a convenience companion document to the Operator’s Manual. It is not intended to replace the Operator’s Manual, which should always be available while using the ventilator. It is important to familiarize yourself with all information in the Operator’s Manual relevant to your institution’s use of the ventilator, including on-screen helps, instructions, warnings and cautions. COVIDIEN, COVIDIEN with logo, Covidien logo and positive results for life are U.S. and internationally registered trademarks of Covidien AG. Other brands are trademarks of a Covidien company. ©2014 Covidien. 13-VE-0089(1)e
6135 Gunbarrel Avenue Boulder, CO 80301 800-635-5267 www.covidien.com/rms
Puritan Bennett™ 980 Ventilator BiLevel 2.0 Software Quick Reference Guide WHAT IS BILEVEL The Puritan Bennett™ BiLevel 2.0 software is SIMV Pressure Control with enhancements. Breaths are delivered in a manner similar to SIMV mode with PC selected, but providing two levels of pressure. The patient is free to initiate spontaneous breaths at either pressure level during BiLevel mode. (see figure 1)
2 3
4
5
6
7
1 t 5
The two pressure levels are called Low Pressure (PL) and High Pressure (PH). At either pressure level, patients can breathe spontaneously, and spontaneous breaths can be assisted with tube compensation or pressure support. BiLevel monitors mandatory and spontaneous tidal volumes separately. Inspiratory time and expiratory time in BiLevel become Time high (TH) and Time low (TL), respectively. During these inspiratory and expiratory times, PH is maintained during TH and PL is maintained during TL.
APRV APPROACH As TL begins to shorten with the TH:TL ratio extending beyond 4:1, the breathing pattern assumes a distinctly different shape. In the extreme, the exaggerated time at PH and abrupt release to PL would match the pattern patented by John Downs and defined as Airway Pressure Release Ventilation. Pressure support (PSUPP) can be used to assist spontaneous breaths at PL and PH. PSUPP is always set relative to PL. Target pressure = PL + PSUPP.
7
VEN_10251_C
(figure 1 BiLevel Mode) 1. P ressure (y-axis) 2. Low pressure PL 3. High pressure PH 4. S pontaneous breath
5. Synchronized transitions 6. P ressure support 7. T ime-based transitions
Primary Settings Description • Low pressure (PL) • High pressure (PH) • Low pressure time (TL)
• High pressure time (TH) • TH:TL ratio
NOTE: The pressure support level is always referenced to PL
INITIAL SETUP • Enter PBW or gender and height. • Select the BiLevel mode button. After selecting BiLevel mode, the ventilator uses the PC mandatory breath type, which cannot be changed. • Choose the spontaneous type (PS or TC). • Choose trigger type (P-TRIG or V-TRIG). • The default settings for BiLevel mode appear. Select desired ventilator settings to make changes. PH must always be at least 5 cmH2O greater than PL. • Set TL, TH, or the ratio of TH to TL. To select settings that would result in a TH:TL ratio greater than 1:1 or 4:1, you must touch the Continue button to confirm. (see figure 1 and then figure2) • Accept All. (see figure 3) All spontaneous breaths are assisted by a pressure of 1.5 cmH2O. However, to add additional pressure support at PH, PSUPP set must be > (PH -PL).
(figure 1)
APRV SPECIFICS In APRV: • The PH and PL level is set to optimize pulmonary compliance and increase mean airway pressure, which may improve oxygenation. The f setting also impacts the mean airway pressure. • The differences between PH and PL levels, TL and the number of releases per minute (f setting) help manage CO2 and alveolar ventilation. • In APRV the operator can configure the BiLevel padlock settings to allow direct control of TL to assure that changes in the f setting will not inadvertently lengthen the TL period and result in destabilization of end-expiratory alveolar volume. With the TL period locked, changes in set f will change the TH period to accommodate the new f setting while maintaining the set TL period.
(figure 2)
(figure 3)
This guide is provided as a convenience companion document to the Operator’s Manual. It is not intended to replace the Operator’s Manual, which should always be available while using the ventilator. It is important to familiarize yourself with all information in the Operator’s Manual relevant to your institution’s use of the ventilator, including on-screen helps, instructions, warnings and cautions. COVIDIEN, COVIDIEN with logo, Covidien logo and positive results for life are U.S. and internationally registered trademarks of Covidien AG. Other brands are trademarks of a Covidien company. ©2014 Covidien. 13-VE-0089(1)f
6135 Gunbarrel Avenue Boulder, CO 80301 800-635-5267 www.covidien.com/rms
Puritan Bennett™ 980 Ventilator Data Management Quick Reference Guide VITAL PATIENT DATA BANNER Patient data is displayed across the top banner (see figure 1). Up to eight (8) variables are visible at one time with the ability to scroll left- and rightto view an additional six (6) values in the upper right hand corner. Values may be configured to show different patient data values or may be left intentionally blank to reduce the number of visible patient data cells.
To configure: • 1. Select a patient data preset you want to configure (see figure 2) • 2. Double tap on any one of the patient data cells displayed (see figure 3) • 3. Select the new display data value you want to display
(figure 1)
(figure 2)
(figure 3)
LARGE FONT SCREEN A drop-down display is available with cells shown in a larger font for ease of viewing. You may configure which patient data or graphics you would like to display within the large font screen
To display the large-font patient data panel: • Swipe the Vital Patient Data banner tab downward or touch the Vital Patient Data tab. The additional patient data panel appears. • Swipe the Additional Patient Data banner’s tab downward or touch the Additional Patient Data banner’s tab. Patient data appear in a larger font.
(figure 4)
To configure the large-font patient data displayed on the GUI: • Double tap on any one of the displayed parameters to access the menu list of parameters to pick from. (see figure 4) • Select the patient displayed parameter you wish to display. • Swipe the Large Font Patient Data panel tab upward or touch the tab to return to the banner to its normal font size.
(figure 5)
WAVEFORM DISPLAY To configure the waveforms and loops: • Touch the Waveform Layout icon. (see figure 5) • Touch the Graphs tab to reveal the five configurations available. • Touch on the individual waveform configuration you wish to select.
This guide is provided as a convenience companion document to the Operator’s Manual. It is not intended to replace the Operator’s Manual, which should always be available while using the ventilator. It is important to familiarize yourself with all information in the Operator’s Manual relevant to your institution’s use of the ventilator, including on-screen helps, instructions, warnings and cautions. COVIDIEN, COVIDIEN with logo, Covidien logo and positive results for life are U.S. and internationally registered trademarks of Covidien AG. Other brands are trademarks of a Covidien company. ©2014 Covidien. 13-VE-0089(1)g
6135 Gunbarrel Avenue Boulder, CO 80301 800-635-5267 www.covidien.com/rms
Puritan Bennett™ 980 Ventilator Proximal Flow System Option Quick Reference Guide FEATURES • The Puritan Bennett™ Proximal Flow system option is used for measuring flows, pressures and tidal volumes of invasively ventilated neonatal patients with predicted body weights (PBW) of 0.3 kg (0.66 lb) to 7.0 kg (15.4 lb) using ET tube sizes from 2.5 mm to 4.0 mm. The NeoMode 2.0 software option must also be installed on the ventilator. • When using the Proximal Flow system option, flow, pressure and volume waveform, along with delivered and exhaled volumes, are derived from Proximal Flow sensor measurements at the patient circuit wye. In addition, proximal flow data are displayed on the waveform plots with a Y appearing in inverse video next to the measurement symbol.
HOW TO KNOW WHEN PROXIMAL FLOW IS ENABLED: • Proximal Flow Sensing is enabled when a “-Y” is after a measured parameter. For example when exhaled tidal volume is being measured by Proximal flow it will change from VTE to VTE-Y in inverse video.
Data Symbol
Description
VTI-Y
Inspired tidal volume (mandatory or spontaneous at patient circuit wye)
VTE-Y
Exhaled tidal volume (at patient circuit wye)
VTE SPONT-Y
Exhaled spontaneous tidal volume (at patient circuit wye)
VTE MAND-Y
Exhaled mandatory tidal volume (at patient circuit wye)
VE TOT-Y
Exhaled total minute volume (at patient circuit wye)
VY
Flow throughout the breath cycle (at patient circuit wye)
VT-
Tidal volume (at patient circuit wye)
PCIRC Y
Pressure throughout the breath cycle (at patient circuit wye)
(table 1) Proximal Flow Option Patient Data Symbols
In the event you need to disable or enable the Proximal Flow system option: • In the constant access icons area, touch the Configure icon. A menu containing tabs appears. (see figure 1) • Touch the Options tab. A screen appears containing the Installed Options and Prox tab. • Touch the Enabled or Disabled button to enable or disable the Prox Flow system option. To connect the Proximal Flow sensor to the ventilator 1. Verify the Proximal Flow sensor, pneumatic lines, and connector are not damaged in any way. 2. O pen the connector panel door and firmly attach the sensor connector to the right-most receptacle in the BDU’s front connector port labeled Prox. Reference Attaching Proximal Flow Sensor to Ventilator, p. F-11.
(figure 1)
To attach the Proximal Flow sensor between the endotracheal tube and patient circuit 1. Connect the larger end of the sensor (marked with “UP” and an arrow) to the endotracheal tube. Reference the figure below. Do not force the connection; when the sensor is oriented correctly, insertion requires little effort.
PURGE An automatic purge is used to clear the lines at preset intervals. A manual purge control is also available. To access this function, see above.
CALIBRATIONS Manual calibration is done during SST. Automatic calibrations are done at preset intervals during ventilation. WARNING: Do not install the Proximal Flow sensor in the patient circuit if the sensor is not also connected to the BDU. Caution: Do not use aerosolized medications with the Proximal Flow sensor. Such medications may damage the sensor.
This guide is provided as a convenience companion document to the Operator’s Manual. It is not intended to replace the Operator’s Manual, which should always be available while using the ventilator. It is important to familiarize yourself with all information in the Operator’s Manual relevant to your institution’s use of the ventilator, including on-screen helps, instructions, warnings and cautions. COVIDIEN, COVIDIEN with logo, Covidien logo and positive results for life are U.S. and internationally registered trademarks of Covidien AG. Other brands are trademarks of a Covidien company. ©2014 Covidien. 13-VE-0089(1)h