QIAsymphony SP - AS O-Ring Change - Important Note Dec 2011

Guide

1 Page

December 2011  Important Note Dear QIAsymphony SP and AS user, The QIAsymphony® maintenance instructions currently state that the tightness of the tip-adapter O-rings should be checked weekly. This weekly tightness test is no longer required and we now recommend replacing the O-rings every month. Note: Please use the O-ring tool (see image, below) according to the enclosed quick-guide for replacing the O-rings.  Accessories Product  Contents  Cat. no.  O-ring change tool set  O-ring tool, 2 pegs 32 O-rings, tweezers  9019164  Set; O-rings (32)  32 O-rings  9019168  Best regards, QIAGEN  Trademarks: QIAGEN®, QIAsymphony® (QIAGEN Group).  9243497EN 12/2011  Sample & Assay Technologies