Welcome The Lumis™ VPAP ST-A is a bilevel positive airway pressure device.
WARNING • Read this entire guide before using the device. • Use the device according to the intended use provided in this guide. • The advice provided by your prescribing doctor should be followed ahead of the information provided in this guide.
Indications for use Lumis 150 VPAP ST-A The Lumis 150 VPAP ST-A device is indicated to provide non-invasive ventilation for patients weighing more than 13 kg or more than 30 kg in iVAPS mode with respiratory insufficiency or obstructive sleep apnoea (OSA). It is intended for home and hospital use. Lumis 100 VPAP ST-A The Lumis 100 VPAP ST-A device is indicated to provide non-invasive ventilation for patients weighing more than 13 kg with respiratory insufficiency or obstructive sleep apnoea (OSA). It is intended for home and hospital use. The humidifier is intended for single patient use in the home environment and re-use in a hospital/institutional environment. Contraindications Positive airway pressure therapy may be contraindicated in some patients with the following pre-existing conditions: •
severe bullous lung disease
•
pneumothorax or pneumomediastinum
•
pathologically low blood pressure, particularly if associated with intravascular volume depletion
•
dehydration
•
cerebrospinal fluid leak, recent cranial surgery, or trauma.
Adverse effects You should report unusual chest pain, severe headache, or increased breathlessness to your prescribing physician. An acute upper respiratory tract infection may require temporary discontinuation of treatment. The following side effects may arise during the course of therapy with the device: •
drying of the nose, mouth, or throat
•
nosebleed
•
bloating
•
ear or sinus discomfort
•
eye irritation
•
skin rashes.
English
1
At a glance The Lumis includes the following: •
Device
•
HumidAir™ humidifier (if supplied)
•
Air tubing
•
Power supply unit
•
Travel bag
•
SD card (already inserted).
Contact your care provider for a range of accessories available for use with the device including: •
Air tubing (heated and non-heated): ClimateLineAir™, ClimateLineAir Oxy, SlimLine™, Standard
•
HumidAir humidifier
•
Side cover for use without the humidifier
•
Filter: Hypoallergenic filter, standard filter
•
Air10™ DC/DC converter (12V/24V)
•
SD card reader
•
Air10 oximeter adapter
•
Air10 USB adapter
•
Power Station II
•
Air10 tubing elbow
About your device
1
Air outlet
6
HumidAir humidifier
2
Air filter cover
7
Screen
3
Retention clip
8
Adapter cover
4
Power inlet
9
SD card cover
5
Serial number and device number
10 LED alarm indicator
2
About the control panel Start/Stop button
Press to start/stop therapy. Press and hold for three seconds to enter power save mode.
Dial
Turn to navigate the menu and press to select an option. Turn to adjust a selected option and press to save your change.
Home button
Press to return to the Home screen.
Different icons may be displayed on the screen at different times including: Ramp Time
Wireless signal strength (green)
Humidity
Wireless transfer not enabled (grey)
Humidifier warming
No wireless connection
Humidifier cooling
Airplane Mode
Alarm muted
English
3
Setup
CAUTION Do not overfill the humidifier as water may enter the device and air tubing. 1. With the device on a stable level surface, grip the retention clip on the back of the device and pull up to open. Note: The retention clip is shown in the open position. 2. (a) Plug the power connector into the device power inlet then (b) push down the retention clip to secure in place. Connect one end of the power cord into the power supply unit and the other end into the power outlet. 3. Connect the air tubing firmly to the air outlet located on the rear of the device. 4. Open the humidifier and fill it with water up to the maximum water level mark. Do not fill the humidifier with hot water. 5. Close the humidifier and insert it into the side of the device. 6. Connect the free end of the air tubing firmly onto the assembled mask. See the mask user guide for detailed information. Recommended masks are available on www.resmed.com. Note: Ensure the device is placed so that the LED alarm indicator is clearly visible. 4
Starting therapy 1. Fit your mask. 2. Press Start/Stop or breathe normally if SmartStart™ is enabled. You will know that therapy is on when the Monitoring screen is displayed. The pressure bar shows the inspiratory and expiratory pressures in green. The green bar will expand and contract as you breathe in and out.
The screen will go black automatically after a short period of time. You can press Home or the dial to turn it back on. If power is interrupted during therapy, the device will automatically restart therapy when power is restored. The Lumis device has a light sensor that adjusts the screen brightness based on the light in the room.
Stopping therapy 1. Remove your mask. 2. Press Start/Stop or if SmartStart is enabled, therapy will stop automatically after a few seconds. Note: If Confirm Stop is enabled, a message is displayed asking if you want to stop therapy. Turn the dial to select Yes and then press the dial to stop therapy. Once therapy has stopped, the Sleep Report gives you a summary of your therapy session. Usage hours–Indicates the number of hours of therapy you received last session. Mask Seal–Indicates how well your mask sealed: Good mask seal. Needs adjusting, see Mask Fit.
Humidifier–Indicates if your humidifier is working properly: Humidifier working. Humidifier might be faulty, contact your care provider. If set by your care provider, you will also see: Events per hour–Indicates the number of apnoeas and hypopnoeas experienced per hour. More Info–Turn the dial to scroll down to view more detailed usage data.
English
5
Power save mode Your Lumis device records your therapy data. In order to allow it to transmit the data to your care provider, you should not unplug the device. However, you can put it into power save mode to save electricity. To enter power save mode: Press and hold Start/Stop for three seconds. The screen goes black.
•
To exit power save mode: • Press Start/Stop once. The Home screen is displayed.
My Options Your Lumis device has been set up for your needs by your care provider, but you may find you want to make small adjustments to make your therapy more comfortable. Highlight My Options and press the dial to see your current settings. From here, you can personalise your options.
Ramp Time Designed to make the beginning of therapy more comfortable, Ramp Time is the period during which the pressure increases from a low start pressure to the prescribed treatment pressure. You can set your Ramp Time to Off or between 5 to 45 minutes. To adjust Ramp Time: 1. In My Options, turn the dial to highlight Ramp Time and then press the dial. 2. Turn the dial to adjust the ramp time to your preferred setting and press the dial to save the change.
6
Ramp Down Ramp Down is intended to make stopping therapy more comfortable by reducing your pressure over a fixed 15 minute period. This option will only be available to you via your care provider. To enable Ramp Down: 1. In My Options, turn the dial to highlight Ramp Down and then press the dial. 2. Turn the dial to select On and then press the dial to save the change. To start Ramp Down: 1. Press the Start/Stop button.
Note: If Confirm Stop is enabled, a message is displayed asking if you want to start Ramp Down. Turn the dial to select Yes and then press the dial to start Ramp Down. The Ramp Down icon
and time remaining will be displayed at the bottom left of the screen.
Once Ramp Down is complete, the device will continue to run at low pressure. To stop therapy at any time, press Start/Stop.
Humidity Level The humidifier moistens the air and is designed to make therapy more comfortable. If you are getting a dry nose or mouth, turn up the humidity. If you are getting any moisture in your mask, turn down the humidity. You can set the Humidity Level to Off or between 1 and 8, where 1 is the lowest humidity setting and 8 is the highest humidity setting. To adjust the Humidity Level: 1. In My Options, turn the dial to highlight Humidity Level and then press the dial. 2. Turn the dial to adjust the humidity level and press the dial to save the change.
If you continue to get a dry nose or mouth, or moisture in your mask, consider using ClimateLineAir heated air tubing. ClimateLineAir together with Climate Control delivers more comfortable therapy.
English
7
Mask Fit Mask Fit is designed to help you assess and identify possible air leaks around your mask. To check Mask Fit: 1. Fit the mask as described in the mask user guide. 2. In My Options, turn the dial to highlight Run Mask Fit and then press the dial. The device starts blowing air. 3. Adjust the mask, mask cushion and headgear until you get a Good result.
To stop Mask Fit, press the dial or Start/Stop. If you are unable to get a good mask seal, talk to your care provider.
More options There are some more options on your device which you can personalise. Mask
This option shows your mask type setting. If you use more than one type of mask, adjust this setting when switching between masks.
Run Warm Up
This option allows you to pre-heat the water before starting therapy, so that the air is not cold or dry at the beginning of therapy.
Ramp Down*
This option is intended to make stopping therapy more comfortable by reducing your pressure over a fixed 15 minute period.
Leak Alert*
When Leak Alert is enabled, the device beeps if the mask leaks too much air or if you remove the mask during therapy.
SmartStart*
When SmartStart is enabled, therapy starts automatically when you breathe into your mask. When you remove your mask, it stops automatically after few seconds.
*When enabled by your care provider.
8
Working with alarms The device is fitted with an alarm feature that monitors your therapy and alerts you to changes that may affect your treatment. When power is connected to the device, the yellow LED alarm indicator will flash and the alarm will sound to confirm that the alarm is working. When an alarm is activated, the yellow LED alarm indicator will flash, the alarm will sound and a message will appear on the screen.
Muting activated alarms To mute the alarm: 1. Press the dial. A list of activated alarms will appear and the flashing alarm will be displayed in the top right corner of the screen. mute icon The alarm will be muted for 2 minutes. 2. To return to the previous screen, highlight OK and press the dial.
Once the condition that activated the alarm is corrected, the alarm sound and flashing icon will stop. If the condition that activated the alarm remains after 2 minutes, the alarm will re-occur. See the Alarms troubleshooting section for help with managing common alarm conditions.
Multiple alarms If multiple alarms are activated at the same time, the most recent alarm message will be displayed on the screen and any other activated alarms will be shown in the Alarms list.
English
9
Viewing the alarms To view the alarm list: 1. From the Monitoring screen, turn the dial clockwise until the last Monitoring screen is displayed. 2. To view the alarm details, highlight the alarm and press the dial.
Caring for your device It is important that you regularly clean your Lumis device to make sure you receive optimal therapy. The following sections will help you with disassembling, cleaning, checking and reassembling your device.
Disassembling
1. Hold the humidifier at the top and bottom, press it gently and pull it away from the device. 2. Open the humidifier and discard any remaining water. 3. Hold the cuff of the air tubing and gently pull it away from the device. Grip the retention clip and pull up to release the power cord. 4. Hold both the cuff of the air tubing and the swivel of the mask, then gently pull apart.
10
Cleaning You should clean the device weekly as described. Refer to the mask user guide for detailed instructions on cleaning your mask. 1. Wash the humidifier and air tubing in warm water using mild detergent. 2. Rinse the humidifier and air tubing thoroughly and allow to dry out of direct sunlight and/or heat. 3. Wipe the exterior of the device with a damp cloth and mild detergent. Notes: •
Empty the humidifier daily and wipe it thoroughly with a clean, disposable cloth. Allow to dry out of direct sunlight and/or heat.
•
The humidifier may be washed in a dishwasher on the delicate or glassware cycle (top shelf only). It should not be washed at temperatures higher than 65ºC.
•
Do not wash the air tubing in a dishwasher or washing machine.
Checking You should regularly check the humidifier, air tubing and the air filter for any damage. 1. Check the humidifier: • Replace it if it is leaking or has become cracked, cloudy or pitted. • Replace it if the seal is cracked or torn. • Remove any white powder deposits using a solution of one part household vinegar to 10 parts water. 2. Check the air tubing and replace it if there are any holes, tears or cracks. 3. Check the air filter and replace it at least every six months. Replace more often if there are any holes or blockages by dirt or dust. To replace the air filter:
1. Open the air filter cover and remove the old air filter. The air filter is not washable or reusable. 2. Place a new air filter onto the air filter cover and then close it. Make sure the air filter is fitted at all times to prevent water and dust from entering the device.
Reassembling When the humidifier and air tubing are dry, you can reassemble the parts. 1. Connect the air tubing firmly to the air outlet located on the rear of the device. 2. Open the humidifier and fill it with room temperature water up to the maximum water level mark. English
11
3. Close the humidifier and insert it into the side of the device. 4. Connect the free end of the air tubing firmly onto the assembled mask.
Therapy data Your Lumis device records your therapy data for you and your care provider so they can view and make changes to your therapy if required. The data is recorded and then transferred to your care provider wirelessly or via an SD card.
Data transmission Your Lumis device has the capability of wireless communication so that your therapy data can be transmitted to your care provider to improve the quality of your treatment. This is an optional feature that will only be available if you choose to benefit from it. It also allows your care provider to update your therapy settings in a more timely manner or upgrade your device software to ensure you receive the best therapy possible. The data is usually transmitted after therapy has stopped. In order to make sure that your data is transferred, leave your device connected to the mains power at all times and make sure that it is not in Airplane Mode. Notes: •
Therapy data might not be transmitted if you use it outside of the country or region of purchase.
•
Devices with wireless communication might not be available in all regions.
SD card An alternative way for your therapy data to be transferred to your care provider is via the SD card. Your care provider may ask you to send the SD card by mail or to bring it in. When instructed by your care provider, remove the SD card. Do not remove the SD card from the device when the SD light is flashing. To remove the SD card:
1. Open the SD card cover. 2. Push in the SD card to release it. Remove the SD card from the device. Place the SD card in the protective folder and send it back to your care provider. For more information on the SD card refer to the SD card protective folder provided with your device. Note: The SD card should not be used for any other purpose.
12
Travelling You can take your Lumis device with you wherever you go. Just keep the following in mind: •
Use the travel bag provided to prevent damage to the device.
•
Empty the humidifier and pack it separately in the travel bag.
•
Make sure you have the appropriate power cord for the region you are travelling to. For information on purchasing, contact your care provider.
•
If you are using an external battery, you should turn off the humidifier in order to maximise the life of your battery. Do this by turning the Humidity Level to Off.
Travelling by plane Your Lumis device may be taken on board as carry-on luggage. Medical devices do not count toward your carry-on luggage limit. You can use your Lumis device on a plane as it meets the Federal Aviation Administration (FAA) requirements. Air travel compliance letters can be downloaded and printed from www.resmed.com. When using the device on a plane: •
Make sure the humidifier is completely empty and inserted into your device. The device will not work without the humidifier inserted.
•
Turn on Airplane Mode. To turn on Airplane Mode: 1. In My Options, turn the dial to highlight Airplane Mode and then press the dial. 2. Turn the dial to select On and then press the dial to save the change. The Airplane Mode icon is displayed at the top right of the screen.
CAUTION Do not use the device with water in the humidifier on a plane due to the risk of inhalation of water during turbulence.
Troubleshooting If you have any problems, have a look at the following troubleshooting topics. If you are not able to fix the problem, contact your care provider or ResMed. Do not try to open the device.
General troubleshooting Problem/possible cause
Solution
Air is leaking from around my mask Mask may be fitted incorrectly.
Make sure your mask is fitted correctly. See your mask user guide for fitting instructions or use the Mask Fit function to check your mask fit and seal.
English
13
Problem/possible cause
Solution
I am getting a dry or blocked nose Humidity level may be set too low.
Adjust the Humidity Level. If you have ClimateLineAir heated air tubing, see the ClimateLineAir user guide.
I am getting droplets of water on my nose, in the mask and air tubing Humidity level may be set too high.
Adjust the Humidity Level. If you have ClimateLineAir heated air tubing, see the ClimateLineAir user guide.
My mouth is very dry and uncomfortable Air may be escaping through your mouth.
Increase the Humidity Level. You may need a chin strap to keep your mouth closed or a full face mask.
Air pressure in my mask seems too high (it feels like I am getting too much air) Ramp may be turned off.
Use the Ramp Time option.
Air pressure in my mask seems too low (it feels like I am not getting enough air) Ramp may be in progress Ramp Down may be in progress
Wait for air pressure to build up or turn Ramp Time off.
. .
Press Start/Stop to stop therapy then press Start/Stop to restart and continue therapy.
Non-vented mask is used.
Only use a vented mask.
Mask vents might be blocked.
Check if you have sufficient venting. Unblock mask vents if necessary.
Expiratory pressure (EPAP) may be set too low.
Talk to your care provider about your settings.
My screen is black Backlight on the screen may have turned off. It turns off automatically after a short period of time.
Press Home or the dial to turn it back on.
Power may not be connected.
Connect the power supply and make sure the plug is fully inserted. Note: the retention clip should be in the open position when inserting the plug. For instructions refer to the Setup section.
I have stopped therapy, but the device is still blowing air Device is cooling down.
Device blows a small amount of air in order to avoid condensation in the air tubing. It will stop automatically after 20 minutes.
My humidifier is leaking Humidifier may not be assembled correctly. 14
Check for damage and reassemble the humidifier correctly.
Problem/possible cause
Solution
Humidifier may be damaged or cracked.
Contact your care provider for a replacement.
My therapy data has not been sent to my care provider Power may not be connected.
Connect the power supply and make sure the plug is fully inserted. Note: the retention clip should be in the open position when inserting the plug. For instructions refer to the Setup section.
Wireless coverage may be poor.
Make sure that the device is placed where there is coverage (ie, on your bedside table, not in a drawer or on the floor). The Wireless signal strength icon indicates good coverage when all bars are displayed, and poor coverage when fewer bars are displayed.
is displayed on the The No wireless connection icon top right of the screen. no wireless network available.
Make sure that the device is placed where there is coverage (ie, on your bedside table, not in a drawer or on the floor). If instructed to do so, send the SD card to your care provider. The SD card also contains your therapy data.
Device may be in Airplane Mode.
Turn off Airplane Mode, see Travelling by plane.
Data transfer is not enabled for your device.
Talk to your care provider about your settings.
My screen and buttons are flashing but there is no alarm sound or message Software upgrade is in progress.
Software upgrade takes approximately 10 minutes to complete.
Displays message: Read only card, please remove, unlock and re-insert SD card SD card switch may be in the lock (read-only) position.
Move the switch on the SD Card from the lock position to the unlock position and then re-insert it.
Alarms troubleshooting Problem/possible cause
Solution
Display disappears and an alarm is activated Power failure.
Remove your mask until power is restored.
Power cord is disconnected or mains power has been turned off during therapy.
Ensure the power cord is connected and the mains power switch (if available) is on.
Displays message: High leak detected, check your water tub, tub seal or side cover Humidifier may not be inserted properly.
Make sure the humidifier is correctly inserted.
Humidifier seal may not be inserted properly.
Open the humidifier and make sure that the seal is correctly inserted.
Displays message: High leak detected, connect your tubing Air tubing may not be connected properly.
Make sure the air tubing is firmly connected at both ends.
English
15
Problem/possible cause
Solution
Mask may be fitted incorrectly.
Make sure your mask is fitted correctly. See your mask user guide for fitting instructions or use the Mask Fit function to check your mask fit and seal.
Displays messsage: Tubing blocked, check your tubing Air tubing may be blocked.
Check the air tubing and remove any blockages. Press the dial to clear the message and then press Start/Stop to restart the device.
Displays message: Leak detected, check your system setup and all connections Mask may be fitted incorrectly.
Make sure your mask is fitted correctly. See your mask user guide for fitting instructions or use the Mask Fit function to check your mask fit and seal.
Displays message: Low MV detected Minute ventilation level has dropped below the pre-set alarm level.
Contact your care provider.
Displays message: Apnoea detected The device detects an apnoea that has exceeded the pre-set alarm.
Breath normally to disable the alarm. If the problem persists contact your care provider.
Displays message: Low Sp02 detected SpO2 has dropped below the pre-set alarm level.
Check the attachment of the sensor. If the problem persists contact your care provider.
Displays message: No SpO2 data, check your oxi sensor attachment to module/finger Oximeter sensor is not attached properly.
Ensure that the oximeter sensor is attached properly to the module and to your finger.
Oximeter sensor may be faulty.
If the message appears repeatedly, the oximeter sensor might be faulty. Replace the oximeter.
Displays message: Non-vented mask detected, use vented mask or unblock your mask vents Non-vented mask is used.
Only use a vented mask.
Mask vents might be blocked.
Check if you have sufficient venting. Unblock mask vents if necessary.
Expiratory pressure (EPAP) may be set too low.
Talk to your care provider about your settings.
Displays message: System fault, refer to user guide, Error 004 Device may have been left in a hot environment.
Allow to cool before re-use. Disconnect the power supply and then reconnect it to restart the device.
Air filter may be blocked.
Check the air filter and replace it if there are any blockages. Disconnect the power supply and then reconnect it to restart the device.
16
Problem/possible cause
Solution
Air tubing may be blocked.
Check the air tubing and remove any blockages. Press the dial to clear the message and then press Start/Stop to restart the device.
There may be water in the air tubing.
Empty the water from the air tubing. Disconnect the power supply and then reconnect it to restart the device.
Displays message: System fault, refer to user guide, Error 022 Power cord may not be correctly inserted into the device.
Remove the power cord from the device and then re-insert it. Ensure that the power cord is fuly inserted into the device. Note: the retention clip should be in the open position when inserting the plug. For instructions refer to the Setup section. If the problem continues, contact your local ResMed dealer or ResMed office. Do not open the device.
All other error messages, for example, System fault, refer to user guide, Error 0XX An unrecoverable error has occurred on the device.
Contact your care provider. Do not open the device.
Reassembling parts Some parts of your device are designed to easily come off in order to avoid damage to the parts or the device. You can easily reassemble them as described below. To insert the humidifier seal:
1. Place the seal into the lid. 2. Press down along all edges of the seal until it is firmly in place. To reassemble the humidifier lid:
English
17
1. Insert one side of the lid into the pivot hole of the base. 2. Slide the other side down the ridge until it clicks into place.
General warnings and cautions WARNING • Make sure that you arrange the air tubing so that it will not twist around the head or neck. • Make sure the power cord and plug are in good condition and the equipment is not damaged. • Keep the power cord away from hot surfaces. • If you notice any unexplained changes in the performance of the device, if it is making unusual sounds, if the device or the power supply are dropped or mishandled, or if the enclosure is broken, discontinue use and contact your care provider or your ResMed Service Centre. • Do not open or modify the device. There are no user serviceable parts inside. Repairs and servicing should only be performed by an authorised ResMed service agent. • Beware of electrocution. Do not immerse the device, power supply or power cord in water. If liquids are spilled into or onto the device, unplug the device and let the parts dry. Always unplug the device before cleaning and make sure that all parts are dry before plugging it back in. • Supplemental oxygen must not be used while smoking or in the presence of an open flame. • Always make sure that the device is turned on and airflow generated before the oxygen supply is turned on. Always turn the oxygen supply off before the device is turned off, so that unused oxygen does not accumulate within the device enclosure and create a risk of fire. • Do not perform any maintenance tasks while the device is in operation. • The device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the device should be observed to verify normal operation in the configuration in which it will be used. • The use of accessories other than those specified for the device is not recommended. They may result in increased emissions or decreased immunity of the device. • Regularly check the antibacterial filter for signs of moisture or other contaminants, particularly during nebulisation or humidification. Failure to do so could result in increased breathing system resistance.
CAUTION • Use only ResMed parts and accessories with the device. Non-ResMed parts may reduce the effectiveness of the treatment and/or damage the device. • Use only vented masks recommended by ResMed or by the prescribing doctor with this device. Fitting the mask without the device blowing air can result in rebreathing of exhaled air. Make sure that the mask vent holes are kept clear and unblocked to maintain the flow of the fresh air into the mask. • Be careful not to place the device where it can be bumped or where someone is likely to trip over the power cord. • Blocking the air tubing and/or air inlet of the device while in operation could lead to overheating of the device. • Keep the area around the device dry, clean and clear of anything (eg, clothes or bedding) that could block the air inlet or cover the power supply unit. • Do not place the device on its side as water might get into the device. • Incorrect system setup may result in incorrect mask pressure reading. Ensure the system is correctly set up. 18
• Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturising or antibacterial soaps or scented oils to clean the device, the humidifier or air tubing. These solutions may cause damage or affect the humidifier performance and reduce the life of the products. • If you use the humidifier, always place the device on a level surface lower than your head to prevent the mask and air tubing from filling with water. • Leave the humidifier to cool for ten minutes before handling to allow the water to cool and to make sure that the humidifier is not too hot to touch. • Make sure that the humidifier is empty before transporting the device. Note: The device is not intended to be operated by persons (including children) with reduced physical, sensory or mental capabilities without adequate supervision by a person responsible for the patient's safety.
Technical specifications Units are expressed in cm H2O and hPa. 1 cm H2O is equal to 0.98 hPa. 90W power supply unit AC input range: DC output: Typical power consumption: Peak power consumption: Environmental conditions Operating temperature:
Operating humidity: Operating altitude: Storage and transport temperature: Storage and transport humidity:
100–240V, 50–60Hz 1.0–1.5A, Class II 115V, 400Hz 1.5A, Class II (nominal for aircraft use) 24V 3.75A 53W (57VA) 104W (108VA) +5°C to +35°C Note: The air flow for breathing produced by this therapy device can be higher than the temperature of the room. Under extreme ambient temperature conditions (40°C) the device remains safe. 10 to 95% relative humidity, non-condensing Sea level to 2,591 m; air pressure range 1013 hPa to 738 hPa -20°C to +60°C 5 to 95% relative humidity, non-condensing
Electromagnetic compatibility The Lumis complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC 60601-1-2:2014, for residential, commercial and light industry environments. It is recommended that mobile communication devices are kept at least 1 m away from the device. Information regarding the electromagnetic emissions and immunity of this ResMed device can be found on www.resmed.com, on the Products page under Service and Support. EN 60601-1:2006 classification Class II (double insulation), Type BF, Ingress protection IP22. Sensors Pressure sensors: Flow sensor:
Internally located at device outlet, analogue gauge pressure type, -5 to +45 cm H2O (-5 to +45 hPa) Internally located at device inlet, digital mass flow type, -70 to +180 L/min
Maximum single fault steady pressure Device will shut down in the presence of a single fault if the steady state pressure exceeds: 30 cm H2O (30 hPa) for more than 6 sec or 40 cm H2O (40 hPa) for more than 1 sec.
English
19