Contents 1 Documentation 1.1 Associated documentation 1.2 Electronic documentation 2 Required information 2.1 Indication for use 2.2 Operating principle 2.3 Using accessories not supplied by the manufacturer 2.4 Connecting and disconnecting accessories during use 2.5 Repairing or modifying the medical device 2.6 Warranty 2.7 Latest document update 2.8 Date of first CE marking 3 Unpacking the medical device 4 Connect the medical device 4.1 Connecting the medical device to the water system 4.2 Connecting the medical device to the water system 4.3 Installing a water filter 4.4 Connecting the medical device to the electrical network 4.5 Connecting the medical device to the electrical network 5 Installing the medical device 5.1 Fixing the medical device to a non-removable support 5.2 Install cords 5.3 Installing the control pedal 5.4 Connecting the ultrasonic 5.5 Attaching a tip or a file 6 Dispensing a treatment 6.1 Accessory usage conditions 6.2 Preparation for use 7 Medical device description 7.1 Control unit 7.2 Light indicator 7.3 Handpiece 7.3.1 Handpiece cord 7.3.2 Handpiece support 7.4 Adjusts the power 7.5 Setting the irrigation flow 7.6 Air inlets 7.7 Control pedal 7.8 Mains Connector 7.9 Switch 8 Disinfection and sterilising 8.1 Clean and disinfect the medical device 8.2 Cleaning, disinfecting and sterilising accessories 9 Monitoring and routine maintenance 9.1 Cleaning the irrigation system 9.2 Clean the water filter cartridge. 9.3 Corrective Maintenance 9.3.1 Replacing the fuses 10 Identifying incorrect operation 10.1 Not working 10.2 No spray
5 5 5 7 7 7 7 7 7 8 8 8 9 11 11 11 11 11 12 13 13 13 13 13 13 15 15 15 17 17 17 17 17 17 17 17 18 18 18 18 19 19 20 21 21 21 22 22 25 25 25
10.3 The power is not as expected 10.4 Ultrasounds not working 10.5 Water leakage 11 Technical specifications of the medical device 11.1 Identification 11.2 Generator 11.3 Length of cords 11.4 Irrigation 11.5 Footswitch 11.6 Environmental characteristics 11.7 Environmental restrictions 11.8 Main performance characteristics 12 Regulations and standards 12.1 Applicable standards and regulations 12.2 Medical class of the device 12.3 Symbols 12.4 Manufacturer identification 12.5 Branch addresses 12.6 Disposal and recycling 13 Index
25 26 26 27 27 27 27 27 27 27 28 28 29 29 29 29 32 33 34 36
Foreword The medical device that you are about to install and use in your practice is a medical device designed for professional use. It is therefore a key tool with which you will provide treatment within the context of your work. To ensure optimum safety for yourself and your patients, comfort in your daily practice and to benefit fully from the technology of your medical device, please read the documentation provided carefully. Please refer to the cleaning, disinfection and sterilisation instructions for accessories and to the cleaning, disinfection and sterilisation instructions for handpieces for information about the following: l l l l
Preparation of parts for sterilisation Detailed manual and automated instructions Information concerning conditioning for sterilisation Recommendations for the inspection of parts
Please refer to the instructions for the entire range of SATELEC, a company of Acteon group dental ultrasonic generators for information about the following: l l l l l l l
Documentation format Documentation archiving period Warnings concerning user and patient populations Treatment area Medical device usage interactions, contraindications and prohibitions Disposal and recycling of the medical device Manufacturer responsibility
Please refer to the User Manuals, Quick Start Guides and Quick Clean Guides for each medical device for information about the following: l l l l
Unpacking and installing the medical device Using the medical device Monitoring and maintaining the medical device Technical specifications of the medical device
1 Documentation This document contains the following information: l l l l l l l
Indications for use Medical device description Installation of the medical device Medical device use Preparation for cleaning and disinfection of the medical device Monitoring and general maintenance of the medical device Maintenance to be performed by the user
1.1 Associated documentation This document must be used in association with the following documents: Document title
References
General instructions relating to the complete range of dental ultrasonic generators
J00011
Cleaning, disinfection and sterilisation instructions for wrenches
J81001
Cleaning, disinfection and sterilisation instructions for tips
J02001
Cleaning, disinfection and sterilisation instructions for handpieces
J12911
Ultrasonic generator power settings table
J58000
Consulting electronic user instructions
J00007
Newtron® P5 User Manual
J61101
Quick Clean Newtron® P5
J61001
Quick Start Newtron® P5
J61000
SLIM handpiece user manual
J12921
Air and water filter and cartridge replacement and installation instructions
J11330
The Quick Start and Quick Clean documents are summaries created for your approval. The only binding instructions are the user manuals and regulatory documentation associated with the medical device.
1.2 Electronic documentation
The user instructions for your device are available in electronic format on the specified website and not in printed format. If the website is unavailable, try again later. You can also request a free printed copy of the user instructions within seven days via our website, by telephone or in writing. The electronic user instructions are available in PDF format (Portable Document Format). You will need to have a PDF file reader installed to read the electronic user instructions. It is important for you to have read and understood the content of the user instructions relating to the use of your device and its accessories. Never use your device without first reading the user instructions. The device user instructions can be consulted at the following addresses:www.ultradent.com and www.satelec.com. When you receive your device, you are asked to print and/or to download all documents or sections of documents that you may need to refer to in the event of an emergency, if you are unable to connect to the internet or if your electronic display tool is not working (computer, tablet, etc.). We recommend that you visit the website regularly to consult and to download the latest version of your device's user instructions. Users are asked to keep documentation close at hand for reference when necessary. All paper or electronic documentation relating to your medical device must be kept for the device’s entire service life.
User Manual | Newtron® P5 | J61101 | V6 | (13) | 09/2017 | NBACEN030F - Page 5/38
Keep the original documentation for your medical device and its accessories for future reference. When loaning out or selling the medical device, the documentation must be provided with it.
Page 6/38 - User Manual | Newtron® P5 | J61101 | V6 | (13) | 09/2017 | NBACEN030F
2 Required information 2.1 Indication for use This medical device is used with a dental ultrasonic handpiece to which an ultrasonic instrument is attached. It is designed for the treatment of prophylaxis,periodontics, endodontics and preservation and restoration dentistry.
2.2 Operating principle An electrical signal emitted by the medical device is supplied to the ultrasonic handpiece. This is connected to the medical device via a cord. The handpiece comprises a piezoelectric ceramic transducer, which transforms the electrical signal into ultrasonic vibrations. Mechanical vibrations are transmitted to a tip or a dental file attached to the end of the ultrasonic handpiece.
2.3 Using accessories not supplied by the manufacturer The handpiece is designed to operate with SATELEC, a company of Acteon group dental tips and files. The use of tips or files made by other manufacturers will damage the handpiece, break tips and files and void the warranty.
2.4 Connecting and disconnecting accessories during use Do not tighten or loosen the tips when the handpiece is activated.
2.5 Repairing or modifying the medical device Contact the supplier of your device. Using the services of an unapproved repairer could render your device dangerous for you and your patients. Do not repair or modify the device without seeking the prior permission of SATELEC, a company of Acteon group. If the device is modified or repaired, specific checks and tests must be carried out to ensure that the medical device is still safe to use. In the event of doubt, contact an approved dealer or the SATELEC, a company of Acteon group Customer Service team:
www.acteongroup.com [email protected] SATELEC, a company of Acteon group, at the request of technical personnel working for the network of approved dealers, will provide any information required to repair defective parts on which they may perform repairs.
User Manual | Newtron® P5 | J61101 | V6 | (13) | 09/2017 | NBACEN030F - Page 7/38
2.6 Warranty
The screws marked A, B, C, D and E must not be unscrewed by the user under any circumstances as this may void the warranty for the medical device.
2.7 Latest document update 09/2017
2.8 Date of first CE marking 2013
Page 8/38 - User Manual | Newtron® P5 | J61101 | V6 | (13) | 09/2017 | NBACEN030F
3 Unpacking the medical device When you receive your medical device, check for any damage that may have occurred during transportation. If you have received this medical device by mistake, please contact the supplier to arrange for it to be collected. If you have any questions or requirements, contact your supplier. The Newtron® P5 B.LED includes the following items: l
l l l l l l l l l
A Newtron® P5 B.LED unit with non-detachable footswitch cord, a non-detachable handpiece cord, a SLIM handpiece support A disconnectable water supply pipe A Newtron SLIM handpiece, with, depending on configuration, an installed blue LED ring, a white LED ring A box containing disclosing liquid F.L.A.G.™ (depending on selected options) A LED handpiece Quick Start-Clean guide [J12930] A Newtron® P5 B.LED [J61000] Quick Start guide A Newtron® P5 B.LED [J61001] Quick Clean guide An attachment kit Tips and wrenches depending on selected options A power cord
User Manual | Newtron® P5 | J61101 | V6 | (13) | 09/2017 | NBACEN030F - Page 9/38
4 Connect the medical device 4.1 Connecting the medical device to the water system Ask an approved dental installation technician to connect your medical device to the water system. The water supply system pressure may vary throughout the day. The water supply system pressure must be adapted to the values recommended for your medical device. It is very important to make sure that the maximum pressure permitted for the medical device is never reached or exceeded. If in doubt, you are strongly advised to install or arrange for installation of a water pressure limiting system. Before a long period of absence or when the medical device is not in use for a long period of time, turn off the water supply to the device or dental practice.
4.2 Connecting the medical device to the water system The supply pipe connector is used to connect the medical device to the domestic water distribution system. The connector is extended by a pipe to which you must attach a filter.The filter needs to be cleaned and replaced regularly as specified in the chapter Clean the water filter cartridge. page 21. The water supply must comply with the quality criteria compatible with the practice of dental treatments.
4.3 Installing a water filter Water supplied by the system in place may not be of a quality suitable for dental treatments. In this case, you must install a water filter on the medical device's water supply pipe. Procure the relevant kit and the installation and maintenance instructions for water and air filters and cartridges [J11330]. SATELEC, a company of Acteon group accepts no responsibility for damage caused by an installation provided by a service provider that has not been approved by us.
4.4 Connecting the medical device to the electrical network Have your medical device connected to the mains power by an approved dental installation technician. Check that the mains voltage is compatible with that indicated on the medical device or its mains adapter. A different voltage would cause damage to the medical device and could injure the patient and the user. Any variation in the electrical network voltage or electromagnetic field that is non-compliant with the limits in force, could interfere with the medical device's operation. Medical devices equipped with a protective earth must be connected to a supply network equipped with a protective earth. Do not plug the medical device into an extension lead and do not put the mains cord in a cable cover or cable tray.
User Manual | Newtron® P5 | J61101 | V6 | (13) | 09/2017 | NBACEN030F - Page 11/38
4.5 Connecting the medical device to the electrical network 1. Set the medical device's mains switch to "O" OFF position. 2. Connect the mains cord to the control unit's mains connector. 3. Connect the power lead to the mains socket.
Page 12/38 - User Manual | Newtron® P5 | J61101 | V6 | (13) | 09/2017 | NBACEN030F
5 Installing the medical device Place the medical device in the position that is suitable for your activity. The medical device must be placed on a secure and flat surface or a surface with a maximum slope of five degrees. Check that the cords do not hinder the movement or free circulation of anyone. Adjust the position of your medical device to correspond to your angle of vision and the characteristics of your workstation, e.g. lighting or distance between the user and the medical device. Ensure that your medical device is readily accessible. The disconnecting devices - the switch and the power plug - must be easy to access. Do not install your medical device near or on another device.
5.1 Fixing the medical device to a non-removable support After its initial installation, the medical device is not designed to be moved. The medical device must be fixed to ensure that it cannot be removed or moved without the use of a tool.
5.2 Install cords Never rotate the handpiece connector on its cord as this can damage your medical device. Never wrap the handpiece cord around the medical device. Make sure that it is not possible to wheel over or walk on the different cords. The cord attached to its handpiece must be easily accessible. Make sure that the cord is slack during use.
5.3 Installing the control pedal The control pedal must be positioned near the feet of the operator and must be readily accessible.
5.4 Connecting the ultrasonic Check for any traces of humidity on the handpiece connections. If the connections are damp, dry them with the multipurpose syringe. Lubricate the irrigation circuit seal located on the metal shaft on the back of the handpiece with silicone paste. This will prolong its service life and prevent leaks. Do not use spray lubricant on dental instruments.
Connect the handpiece to the socket by aligning the indexing points, without rotating it.
5.5 Attaching a tip or a file A tip or a file vibrates correctly when it is perfectly tightened without being forced beyond its stop point. Tighten the tip with the torque wrench (F81320, F81321, F81322 or F81323) to ensure optimum ultrasonic function. Over-tightening of the tip or file with the open-ended wrench can result in breakage of the tip, file or handpiece. To prevent self-locking of the tip or the file, the latter must be removed and sterilised after each use.
User Manual | Newtron® P5 | J61101 | V6 | (13) | 09/2017 | NBACEN030F - Page 13/38
The torque wrenches must be replaced every year.
Page 14/38 - User Manual | Newtron® P5 | J61101 | V6 | (13) | 09/2017 | NBACEN030F
6 Dispensing a treatment 6.1 Accessory usage conditions The accessories and the handpiece must be cleaned, disinfected and sterilised prior to each use.
Refer to the cleaning, disinfection and sterilisation instructions for accessories listed in the chapter Associated documentation page 5.
6.2 Preparation for use
To prepare your medical device, follow the steps below: 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11.
Wear safety goggles and protective gloves. Clean the unit with an alcohol disinfectant wipe. Remove the handpiece from its sterilisation bag. Remove the wrench from its sterilisation bag. Remove the tip from its sterilisation bag. Screw the tip onto the handpiece, first manually and finishing with the wrench. Connect the handpiece to the handpiece cord socket. Place the handpiece on its support. . Switch on the medical device. Check the irrigation parameters depending on the tip chosen. After water drainage, check that the spray works correctly.
Your medical device is now ready to use.
User Manual | Newtron® P5 | J61101 | V6 | (13) | 09/2017 | NBACEN030F - Page 15/38
7 Medical device description 7.1 Control unit The control unit incorporates Newtron® technology patented by SATELEC, a company of Acteon group . Newtron® technology emits ultrasonic vibrations in a controlled way. Relayed by SATELEC, a company of Acteon group tips, these vibrations are used to deliver effective treatments and to ensure patient safety. The control unit incorporates an dental ultrasonic generator equipped with a piezoelectric command.
7.2 Light indicator The light indicator is designed to provide information about the status of the medical device. When the light indicator is illuminated, the medical device is on and ready to use. Each colour corresponds to the power level range.
7.3 Handpiece There is a handpiece with SLIM B.LED to white LED connector and a handpiece with SLIM B.LED to blue LED connector. Refer to the Newtron handpiece user manual [J12921] for more information. The handpieces are designed to operate exclusively with SATELEC dental ultrasonic generators.
7.3.1 Handpiece cord The SLIM cord is only compatible with Acteon handpieces with SLIM connector. The cord ensures irrigation circulation and electrical connection between the medical device and the handpiece.
7.3.2 Handpiece support The support holds the handpiece or the cord sleeve. The handpiece support can be fixed to the front face or the right side face of the medical device. To change the position of this support, unscrew the two screws located under the support, position the support over the two holes located on the right side face and insert and tighten the two fastening screws. The silicone supports can be removed by sliding them along the metal rod. They can be sterilised.
7.4 Adjusts the power The ultrasound power must be adjusted in accordance with the tip used and the required treatment. The operating power of the tips must be selected in compliance with the Acteon tips color coding system (CCS tips). Each tip must be used in accordance with the settings defined in the ultrasonic generator irrigation and power settings table [J58000]. The ultrasound power configuration button is used to set the operating power: 1 to 20. Rotating the button causes the colour of the medical device's backlighting to change. l l l l
Green: 1 to 6: very low to low power, used mainly for periodontics. Yellow: 6 to 11: medium power, used mainly for endodontics. Blue: 11 to 16: high power, used mainly for scaling and preservation and restoration dentistry. Orange: 16 to 20: very high power, used mainly for loosening.
The ultrasound power configuration button can be removed by the user to facilitate the cleaning and disinfection of the control unit. The button cannot be sterilised.
7.5 Setting the irrigation flow The medical device must be set to minimum power to adjust the irrigation flow rate. Press the footswitch until a spray appears. Because work habits, feedback and professional training differ from one professional to another, the user must ensure that the irrigation flow is compatible with the procedure to be carried out to prevent burns to the clinical site. Adjust the irrigation flow using the irrigation flow configuration button. The flow rate must be adapted to the tip used and to the required treatment. Tighten to decrease the flow rate and loosen to increase the flow rate. The irrigation flow configuration button is not designed to be removed.
User Manual | Newtron® P5 | J61101 | V6 | (13) | 09/2017 | NBACEN030F - Page 17/38
To avoid losing any device components, do not loosen the configuration button fully so that it comes off. Each tip must be used in accordance with the settings defined in the ultrasonic generator irrigation and power settings table [J58000].
7.6 Air inlets Air inlets ensure correct ventilation of the control unit. Leave them uncovered to allow air to circulate.
7.7 Control pedal The ON/OFF type control pedal is used by the practitioner to operate the medical device. Pressing the footswitch automatically activates the handpiece ultrasounds, and the irrigation function if it is not deactivated. The footswitch equipped with its cord cannot be disconnected. Its weight and antislip pad ensure good stability. The light function remains active for approx. 9 seconds after the pedal is released.
7.8 Mains Connector The mains connector with its earthing pin is used to connect the device to the electrical network via a disconnectable mains cord.
7.9 Switch The mains switch is used to switch on (position I) or to stop (position O) the medical device.
Page 18/38 - User Manual | Newtron® P5 | J61101 | V6 | (13) | 09/2017 | NBACEN030F
8 Disinfection and sterilising The instructions relating to cleaning, disinfection and sterilisation protocols for accessories provided by SATELEC, a company of Acteon group have been approved for each medical device and accessory. The applicable guides are listed in the chapter Associated documentation page 5. They can be downloaded at the following address: www.satelec.com/documents.
In all cases, the local regulations in force relating to the cleaning, disinfection and sterilisation protocols for accessories take precedence over the information provided by SATELEC, a company of Acteon group.
8.1 Clean and disinfect the medical device The medical device's control unit must be cleaned and disinfected daily. The handpiece must be cleaned, disinfected and sterilised after each use. Do not immerse the handpiece. The medical device must be in OFF or O stop position during cleaning and disinfecting procedures. Refer to the instructions in the chapter Cleaning the irrigation system page 21. Use alcohol disinfectant wipes. Avoid using cleaning and disinfection products that contain flammable agents. Otherwise, ensure that the product has completely evaporated or that there is no fuel left on the medical device and its accessories before switching it on. Do not use an abrasive product to clean the medical device. Never apply sprays directly to the medical device to clean it. Always spray the product onto a wipe, then clean the medical device.
User Manual | Newtron® P5 | J61101 | V6 | (13) | 09/2017 | NBACEN030F - Page 19/38
To prepare for cleaning, remove the various parts of the Newtron® P5 B.LED as shown here.
8.2 Cleaning, disinfecting and sterilising accessories Refer to the cleaning, disinfection and sterilisation instructions for accessories listed in the chapter Associated documentation page 5.
Page 20/38 - User Manual | Newtron® P5 | J61101 | V6 | (13) | 09/2017 | NBACEN030F
9 Monitoring and routine maintenance The only preventive maintenance the medical device requires is: l l l
Monitoring of accessories Routine cleaning, disinfection and sterilisation Cleaning
The medical device requires cleaning of the water filter cartridge. Check the cleanliness of the air inlets on the control unit to prevent any heating. Check the cleanliness of the handpiece nosepiece. It must be clean, smooth and corrosion-free. The handpiece must screw easily and firmly inside it. Check the condition of the handpiece rear seals, which must not be distended, torn or broken. Before and after use, check the medical device and its accessories entirely for any problems. This is necessary to detect any electrical isolation fault or damage. If necessary, replace damaged parts.
9.1 Cleaning the irrigation system Operate the device at minimum power, at maximum irrigation flow rate for two minutes. When the irrigation system has been cleaned, perform the following operations: 1. Disconnect the handpiece and refer to the handpiece cleaning, disinfection and sterilisation instructions [J12911]. 2. Clean and disinfect the medical device as indicated in the chapter Clean and disinfect the medical device page 19. 3. Follow the instructions for cleaning, disinfecting and sterilising SATELEC, a company of Acteon group accessories [J81001] and [J02001].
9.2 Clean the water filter cartridge. The water filter cartridge must be cleaned regularly. The water filter must be replaced every six months. To clean the water filter cartridge, follow the steps below: 1. 2. 3. 4. 5.
Stop the medical device (position O). Unplug the electrical supply. Shut off the water supply. Disconnect the pipe from the device. Unscrew the two filter sections.
User Manual | Newtron® P5 | J61101 | V6 | (13) | 09/2017 | NBACEN030F - Page 21/38
6. Check that there is no leakage and that the spray works correctly. A damaged or blocked cartridge must be replaced.
9.3 Corrective Maintenance In the event of faulty operation, the following corrective maintenance actions may be performed by the user.
9.3.1 Replacing the fuses The medical device is protected by two fuses in the mains connector. To replace the fuses, perform the following operations: 1. Stop the medical device (position O). 2. Disconnect the mains cord from the electrical network.