Schiller
ARGUS PRO LifeCare with Defib abd Pacemaker User Manual Rev h
User Manual
176 Pages
Preview
Page 1
ARGUS PRO LifeCare
ARGUS PRO LifeCare
Portable Emergency Patient Monitor with Defibrillator and Pacemaker
14 m
in
:5
Portable Emergency Patient Monitor with Defibrillator and Pacemaker
14 4 m
Sp
70 26
O2 CE : 9 01 23
7%
PP
O2 CE : 9 01 23
:7
4
7%
PP
:7
*2.510659*
S C 1/ HIL m LE in N R AG IB A rt P: N o --.2 .1 / -57 -m 02 6 C m
User Guide
Art. no.: 2.510659 rev.: h
*2.510659*
:5
Sp
70 26
S C 1/ HIL m LE in N R AG IB A rt P: N o --.2 .1 / -57 -m 02 6 C m
Art. no.: 2.510659 rev.: h
in
User Guide
Sales and Service Information
Sales and Service Information
The SCHILLER sales and service centre network is world-wide. For the address of your local distributor, contact your nearest SCHILLER subsidiary. In case of difficulty a complete list of all distributors and subsidiaries is provided on our internet site:
The SCHILLER sales and service centre network is world-wide. For the address of your local distributor, contact your nearest SCHILLER subsidiary. In case of difficulty a complete list of all distributors and subsidiaries is provided on our internet site:
www.schiller.ch
www.schiller.ch
Sales information can also be obtained from: [email protected]
Sales information can also be obtained from: [email protected]
Address Headquarters
Address Headquarters
SCHILLER AG Altgasse 68 CH-6341 Baar Switzerland
Article No: 2.510659 rev.: h Issue date: 28.09.11
Tel: + 41 (0) 41 766 42 42 Fax: + 41 (0) 41 761 08 80
SCHILLER AG Altgasse 68 CH-6341 Baar Switzerland
Article No: 2.510659 rev.: h Issue date: 28.09.11
Tel: + 41 (0) 41 766 42 42 Fax: + 41 (0) 41 761 08 80
ARGUS PRO LifeCare
User Guide
Art. no.: 2.510659 rev.: h
Contents 1
Safety Notes ... 9
1.1
Intended Use ... 9
1.1.1 1.1.2 1.1.3 1.1.4
General ... 9 Defibrillator ... 10 etCO2 ... 10 SpO2 ... 10
1.2
Responsibility of the User ... 11
1.3
Organisational Measures ... 11
1.4
Safety Conscious Operation ... 12
1.5
Operation with other Devices ... 13
1.6
Maintenance... 13
1.7
Safety Symbols and Pictograms ... 14
1.7.1 1.7.2 1.7.3 1.7.4
Symbols Used in this Document ... 14 Symbols Used on the Device and Accumulator ... 15 Symbols Used on the Electrode Packaging ... 16 Symbols used on the SpO2 Sensor Packaging ... 16
1.8
Additional terms ... 17
1.8.1 1.8.2
Implied Authorisation... 17 Terms of Warranty ... 17
2
Introduction ... 18
2.1
Overview... 18
2.2
Measured Parameters ... 19
2.3
Parameter Alarm... 19
2.4
Defibrillator ... 20
2.5
Transcutaneous Cardiac Pacemaker... 21
2.6
Memory ... 22
2.6.1 2.6.2 2.6.3
Trend... 22 Event ... 22 Mini SD slot ... 22
2.7
Options ... 23
3
Initial Preparation ... 24
3.1
Inserting/Changing the Accumulator ... 24
3.2
Connecting an External Power Supply... 25
3.2.1
Power Supply and Charging Status ... 25
3.3
Operating the Unit in Low Temperature ... 25
3.4
Switching On and Off ... 26
3.4.1 3.4.2
Switching On ... 26 Switching Off ... 27
3.5
Protection Against Splash Water and Foreign Bodies ... 28
3.6
Connecting an ARGUS PRO Transport ... 29
3.6.1 3.6.2 3.6.3
Connecting ... 29 Indication... 29 Switching off the ARGUS PRO LifeCare Monitoring Section... 30
Page 1
ARGUS PRO LifeCare
4
Operating and Display Elements ...31
4.1
Connectors... 31
4.2
Defibrillator and Pacemaker Pad Connector... 32
4.3
Control Keys ... 33
4.4
LCD Screen ... 36
4.4.1
Message and Menu Area... 37
4.5
Changing the Parameters Displayed ... 38
4.6
Symbols on the Display ... 40
4.6.1 4.6.2 4.6.3 4.6.4 4.6.5 4.6.6 4.6.7
Accumulator ... 40 QRS Beeper... 40 Memory Card ... 40 NIBP Interval Measurement... 41 Alarm... 41 Defibrillation Electrode Pad Size and NIBP Cuff Size... 41 Bluetooth... 42
4.7
System Visual and Audible Indicators... 43
4.7.1 4.7.2 4.7.3
Power and Connection LED... 43 Monitoring LED ... 43 Audible Indications... 44
4.8
System and Technical Messages... 45
4.9
ARGUS PRO LifeCare Menu Structure ... 47
4.9.1 4.9.2
Menu Navigation ... 47 ARGUS PRO LifeCare Menu Table... 48
5
Alarm Handling ...56
5.1
Parameter Alarm Indications ... 56
5.2
Stopping / Pausing the Audible Alarm ... 57
5.2.1 5.2.2 5.2.3
Pausing the Audible Alarm when an Alarm Limit Breached:... 57 Pausing all Audible Alarms ... 57 Audible Alarm Paused Indicator... 57
5.3
Alarm Limits and Settings ... 58 Alarm Pause Time ... 58
5.4
Alarm Limits ... 59
Art. no.: 2.510659 rev.: h
5.3.1
Page 2
ARGUS PRO LifeCare
6
Defibrillation Overview ...60
6.1
Introduction... 60
6.1.1 6.1.2 6.1.3 6.1.4
Defibrillating Modes ... 60 Shock Level ... 61 Event Storage ... 61 Direct Printout ... 61
6.2
Safety Notes and General Application Guidelines ... 62
6.2.1 6.2.2
Defibrillating Children... 63 Internal Safety Discharge... 63
6.3
Semi AED (automated) Mode Overview... 64
6.3.1 6.3.2 6.3.3 6.3.4 6.3.5 6.3.6
Device Identifies a Shockable Rhythm ... 64 After the First Shock: ... 65 After a Successful Shock: ... 65 After an Unsuccessful Shock ... 65 Device Identifies no Shockable Rhythm ... 65 Language and Display Support for AED Mode ... 66
6.4
Manual Mode Overview ... 67
6.4.1 6.4.2 6.4.3 6.4.4
Manual Mode (Defibrillator)... 68 Cardioversion... 68 Cardioversion References ... 68 Language and Display Support for Manual Mode... 69
7
Defibrillation Procedure ...70
7.1
Diagnostic Procedural Overview... 70
7.2
Applying the Electrode Pads ... 71
7.2.1 7.2.2 7.2.3
Electrode Pads for Adults and Children ... 71 Applying the Electrode Pads... 71 Checking the Electrodes ... 72
7.3
Defibrillating Procedure - AED Mode... 73
7.4
Ending AED Defibrillation Mode / Changing Mode... 77
7.4.1
Exiting Defibrillation Mode ... 77
7.5
Defibrillating Procedure - Manual Mode ... 78
7.6
Ending Manual Defibrillation Mode / Changing Mode... 81
7.6.1
Exiting Defibrillation Mode ... 81
7.7
Defibrillating Procedure - Cardioversion... 82 Ending Manual Defibrillation Mode / Changing Mode... 87
7.8.1
Exiting Defibrillation Mode ... 87
Art. no.: 2.510659 rev.: h
7.8
Page 3
ARGUS PRO LifeCare
Pacemaker Overview ...88
8.1
Modes ... 88
8.1.1 8.1.2 8.1.3
On demand ... 88 Fixed ... 88 Event Storage ... 89
8.2
Safety Notes ... 90
8.3
Guidelines for the Application of External Pacemakers . 90
9
Pacing Procedure ...91
9.1
Preparing the Patient... 92
9.1.1 9.1.2
Apply ECG Electrodes ... 92 Apply Pacemaker Pads... 93
9.2
Procedure ... 94
9.3
Ending / Changing Pacemaker Mode... 98
9.3.1
Exiting Pacemaker Mode ... 98
10
Storing Intervention Data ...99
10.1
AED Defibrillating Mode... 99
10.2
Manual Defibrillating Mode ... 99
10.3
Pacemaker Mode ... 99
10.4
Inserting the Memory Card ... 100
Art. no.: 2.510659 rev.: h
8
Page 4
ARGUS PRO LifeCare
11
Monitoring ...101
11.1
ECG ... 101
11.1.1 11.1.2 11.1.3 11.1.4 11.1.5
Connecting the ECG Patient Cable ... 101 Electrodes Identification and Colour Code IEC/AHA ... 105 ECG Measurement Messages... 105 Viewing the ECG... 106 ECG Settings ... 107
11.2
Non-Invasive Blood Pressure... 108
11.2.1 11.2.2 11.2.3 11.2.4 11.2.5 11.2.6
Procedure ... 108 Taking an Extra Measurement... 108 Stopping a Measurement... 108 NIBP Measurement Messages ... 108 NIBP Display... 109 NIBP Settings... 110
11.3
SpO2 and Peripheral Pulse ... 111
11.3.1 11.3.2 11.3.3 11.3.4
SpO2 Display ... 112 SpO2 Measurement Messages ... 112 SpO2 Settings... 112 Placing the Sensor... 112
11.4
Invasive Blood Pressure ... 113
11.4.1 11.4.2 11.4.3
Invasive Pressure Measurement Messages ... 113 Preparing IBP Measurement... 114 Pressure Calibration ... 114
11.5
CO2 Mainstream Measurement... 115
11.5.1 11.5.2 11.5.3 11.5.4 11.5.5 11.5.6 11.5.7 11.5.8 11.5.9
11.6 11.6.1 11.6.2 11.6.3 11.6.4 11.6.5
Safety notes ... 115 Preparing the IRMA sensor... 116 Initial operation of the IRMA sensor... 116 Placement of IRMA sensor ... 117 Zero reference calibration for the IRMA CO2 sensor... 117 O2 Compensation ... 118 Sensor LED indications... 119 Oxygen Cell Replacement ... 119 CO2 Measurement Messages ... 119
CO2 Sidestream Measurement ... 120
ISA gas analyser (sidestream measurement)... 120 Operation of the ISA gas analyser ... 121 Sensor LED indications... 121 Respiration rate messages ... 121 Zero adjustment of the CO2 sidestream sensor ... 121
Temperature ... 122
11.7.1
Temperature Measurement Messages ... 122
Art. no.: 2.510659 rev.: h
11.7
Page 5
ARGUS PRO LifeCare
12
Bluetooth Operation ...123
12.1
Settings - Defining the Pairing Host ... 123
12.1.1 12.1.2
Ortivus Monitor... 123 Enabling the SCHILLER OP1 printer ... 124
12.2
Establishing a Connection... 125
12.2.1 12.2.2 12.2.3
Establishing a connection with an Ortivus Monitor ... 125 Establishing a connection with an OP-1 printer ... 125 Bluetooth icons ... 125
13
SCHILLER OP-1 Printer ...126
13.1
Printer Control Elements ... 126
13.1.1
Printer with Blue tooth... 126
13.2
Connecting the Printer ... 127
13.2.1 13.2.2
USB Cable ... 127 Bluetooth Wireless Connection... 127
13.3
Power Supply ... 128
13.3.1 13.3.2
Exchanging and Charging the Accumulator... 128 External Power Supply... 128
13.4
Changing the Paper... 129
13.5
Mounting the Printer with the ARGUS PRO LifeCare... 130
13.6
Obtaining a Printout ... 131
13.6.1 13.6.2 13.6.3 13.6.4 13.6.5 13.6.6 13.6.7 13.6.8 13.6.9
Auto Printout During Defibrillation... 131 Printout of Defined Data... 132 Print Online (Print Screen) ... 132 Print Endless... 133 Print Resting ECG... 134 Print Event ... 135 Stopping the Print ... 136 Printing the Trend Data... 137 Print Two Selected Pages of Trend Data... 138
13.7
Printer Maintenance ... 139
13.7.1 Cleaning the Thermal Print Head... 139 13.7.2 Print Quality and Printer Alignment Functional Check ... 139 13.7.3 Printer Accumulator Check ... 139
Page 6
SCHILLER Charging Unit CS-1 ...140
14.1
Accumulator Charging Options ... 140
14.2
Inserting an Accumulator ... 140
14.3
The Control Panel ... 141
14.4
Input and Output Supplies... 142
14.5
Accumulator Calibration ... 143
Art. no.: 2.510659 rev.: h
14
ARGUS PRO LifeCare
User Guide
15
Maintenance ... 144
15.1
Maintenance Intervals ... 144
15.1.1 Maintenance Interval Table... 144
15.2
Check of Life-saving Functions ... 146
15.2.1 Visual Inspection ... 146 15.2.2 Defibrillator test ... 146
15.3
Accumulator... 147
15.3.1 Charging and Calibrating the Accumulator ... 147 15.3.2 Internal Backup Battery... 147 15.3.3 Accumulator Disposal ... 147
15.4
Unit Disposal... 148
15.5
Cleaning ... 148
15.5.1 Detergents... 148
15.6
Disinfection ... 149
15.6.1 Disinfectant ... 149 15.6.2 Cleaning and disinfecting the device, cable and sensors ... 150 15.6.3 Cleaning and disinfecting the device, cables and sensors... 150
15.7
Inspection and Check List Tables... 151
15.7.1 15.7.2 15.7.3 15.7.4 15.7.5 15.7.6
Check of Life Saving functions - Every Day ... 151 Every Week... 152 Following Every Intervention or Every Six Months ... 153 Every 12 Months ... 153 Every 24 Months ... 153 Lifed Item Replacement Every 3 - 5 years ... 154
15.8
Trouble Shooting ... 155
15.8.1 Defibrillating and General... 155 15.8.2 Parameter Measurement ... 156
15.9
Disposal of the Unit ... 157
Art. no.: 2.510659 rev.: h
15.10 Disposal of Accumulators ... 157
Page 7
ARGUS PRO LifeCare
16
Technical Data ...158
16.1
System Data ... 158
16.2
Measured Values ... 160
16.2.1 ECG ... 160 16.2.2 Temperature ... 160 16.2.3 NIBP - Non-invasive Blood Pressure ... 160 16.2.4 IBP - Invasive Blood Pressure ... 161 16.2.5 SpO2 - Pulsoximetry ... 161 16.2.6 etCO2 - Capnography... 162
Defibrillating... 163
16.4
Pacemaker... 165
16.5
Options ... 166
16.6
Literature ... 167
16.7
Glossary ... 167
16.8
Electromagnetic Radiation ... 168
17
Accessories and Disposables ...169
17.1
Accessories... 169
18
Administrator Settings ...171
18.1
Password for Menu Options ... 171
19
Index ...173
Art. no.: 2.510659 rev.: h
16.3
Page 8
ARGUS PRO LifeCare
User Guide
Safety Notes Intended Use
1 1.1
1 Safety Notes 1.1
Intended Use
1.1.1
General V The ARGUS PRO LifeCare is for the monitoring of vital parameters such as: – ECG and Heartrate – SpO2 and pulse rate (PP) – – – –
Non invasive blood pressure: systolic, diastolic and mean Invasive blood pressure: systolic, diastolic and mean Temperature CO2
V The system incorporates an external defibrillator and external pacemaker. V The ARGUS PRO LifeCare is designed to meet the specific needs of ground and air rescue services as well as in-house and inter-hospital transportation. V Do not use the ARGUS PRO LifeCare in or near magnetic resonance imaging equipment. V The ARGUS PRO LifeCare is powered from an internal rechargable accumulator. V There is no danger when using the device for a patient with a pacemaker fitted V The device is only intended for single patient use. V Only operate the device in accordance with the specified technical data. V Do not use this unit in areas where there is any danger of explosion or in the presence of flammable gases such as anaesthetic agents. V V
The Defibrillation part of the unit is BF classified. The ECG, Temperature, IBP, NIBP and SPO2 part of the unit is CF classified. It is defibrillation protected only when the SCHILLER original patient cable is used. However, as a safety precaution when possible, remove ECG electrodes before defibrillation.
V This product is not designed for direct cardiac application.
Art. no.: 2.510659 rev.: h
V The ARGUS PRO LifeCare is an emergency device and must be ready for operation at any time and in all situations. Ensure that the device is always equipped with a sufficiently charged accumulator, and keep a spare accumulator on hand.
Page 9
1 1.1
Safety Notes Intended Use
ARGUS PRO LifeCare
1.1.2
Defibrillator V
The defibrillator of the ARGUS PRO LifeCare is used in the treatment of ventricular fibrillation (VF) and ventricular tachycardia (VT). It has both automated and manual modes.
V
The device may be used with the appropriate electrodes on either adults or children.
V
The device may only be used on a suspected cardiac arrest victim exhibiting the following symptoms:
– Not responsive – Not breathing – No pulse V The device must not be used if the patient: – is responsive – is breathing – has a pulse V Always follow the indicated specifications when operating the device.
1.1.3
etCO2 V
1.1.4
The IRMA sensor is intended for the continuous non invasive monitoring of expired CO2 (measured at the CO2 sensor). The IRMA sensor is intended for use with adult, paediatric and neonatal patients.
SpO2 The Masimo SET® and the Nellcor OXIMAX® SpO2 sensors are intended for the continuous non-invasive monitoring of functional oxygen saturation of arterial haemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor). The sensors are intended for use with Adult, Paediatric, and Neonatal populations.
Art. no.: 2.510659 rev.: h
V
Page 10
ARGUS PRO LifeCare
User Guide
1.2
Safety Notes Responsibility of the User
1 1.2
Responsibility of the User V This ARGUS PRO LifeCare is provided for the exclusive use of qualified physicians or trained personnel under their direct supervision. V The numerical and graphical results as well as any interpretation suggested by the device must be examined with respect to the patient's overall clinical condition and the quality of the recorded data. V The responsibilities of the personnel for the operation and maintenance of the device must be specified. V Ensure that personnel have read and understood these operating instructions. In particular this section safety notes must be read and understood. V Damaged or missing components must be replaced immediately with original spare parts and accessories. V It is the owner's responsibility that the valid regulations for safety and prevention of accidents are observed. V Store the device where it is not accessible to children.
1.3
Organisational Measures V Before using the unit, ensure that an introduction regarding the unit functions and the safety precautions has been provided by a medical product representative. V Keep these operating instructions in an accessible place for reference when required. Make sure that they are always complete and legible. V Observe the operating instructions and maintenance instructions.
Art. no.: 2.510659 rev.: h
V In addition to this user guide, legal and other binding regulations for the prevention of accidents and for environment protection must be observed.
Page 11
1 1.4
Safety Notes Safety Conscious Operation
1.4
ARGUS PRO LifeCare
Safety Conscious Operation V
Ensure that the personnel have read and understood these operating instructions and in particular this chapter “Safety Notes".
V
Position the device so that there is no possibility of it falling on the patient or floor.
V
Do not touch the unit casing during defibrillation.
V
Electric shock hazard. The energy supplied to the patient can be conducted through the patient to other persons, possibly resulting in a fatal electric shock. For this reason:
– Do not touch the patient, the electrodes or other conductive objects while defibrillation is in progress. – Do not defibrillate the patient in a puddle of water or on other conductive surfaces. – Switch off the device when it is no longer needed. V To ensure patient's safety neither the electrodes, including the neutral electrode, nor the patient, or persons touching the patient, must come into contact with conducting objects even if these are earthed. If the patient has an implanted pacemaker, be sure not to position the electrode directly on top of it.
V
Immediately report any changes that impair safety (including operating behaviour) to the person responsible.
V
Only connect original SCHILLER adhesive electrodes to the device.
V
Prior to switching on the device, perform a visual inspection to ensure that there is no damage to the device casing or the electrode connector port.
V
Operating a device with a defective casing or defective cables constitutes a danger to the patient or the user. Therefore: – Immediately replace a damaged unit, or damaged cables and connections.
Art. no.: 2.510659 rev.: h
V
Page 12
ARGUS PRO LifeCare
User Guide
1.5
Safety Notes Operation with other Devices
1 1.5
Operation with other Devices V Use only accessories and other parts recommended or supplied by SCHILLER. Use of other than recommended or supplied parts may result in injury, inaccurate information and/or damage to the unit. V Ancillary equipment connected to the analogue and/or digital interfaces must be certified according to the respective IEC standards (e.g. IEC/EN 60950 for data processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all configurations shall comply with the valid version of the system standard IEC/ EN 60601-1-1. Everybody who connects additional equipment to the signal input part or signal output part configures a medical system, and is therefore responsible that the system complies with the requirements of the valid version of the system standard IEC/EN 60601-1-1. If in doubt, consult the technical service department or your local representative. V When the unit is connected to the ARGUS PRO Transport, any other equipment used with the patient must use the same common earth. V Precautions must be observed when using high frequency devices. Use the original SCHILLER patient cables to avoid possible signal interference during data acquisition. V There is no danger when using the unit simultaneously with other electrical stimulation equipment. However, the stimulation units should only be used at a sufficient distance from the electrodes and transducers. If in doubt, the patient should be disconnected from the monitor. V If the ECG patient cable should become defective after defibrillation, an electrode becomes displaced, or an electrode resistance is too high, an acoustic alarm given. V Portable communication devices, HF radios and devices labelled with the symbol (non-ionic electromagnetic radiation) can affect the operation of this device (see Disposal of the Unit, page 157).
1.6
Maintenance V Danger of electric shock. Do not open the device. No serviceable parts inside. Servicing may only be carried out by a qualified technician authorised by SCHILLER. V Before cleaning, switch off the device and remove the accumulator.
Art. no.: 2.510659 rev.: h
V Do not use high temperature sterilisation processes (such as autoclaving). Do not use E-beam or gamma radiation sterilisation. V Do not use solvent or abrasive cleaners on either the unit or cable assemblies. V Do not under any circumstances, immerse the unit or cable assemblies in liquid
Page 13
1 1.7
Safety Notes Safety Symbols and Pictograms
ARGUS PRO LifeCare
1.7
Safety Symbols and Pictograms
1.7.1
Symbols Used in this Document The safety level is classified according ANSI Z535.4. The following overview shows the safety symbols and pictograms used in this manual.
For a direct danger which could lead to severe personal injury or to death.
For a possibly dangerous situation, which could lead to serious bodily injury or to death.
For a possibly dangerous situation which could lead to personal injury. This symbol is also used to indicate possible damage to property.
For general safety notes as listed in this chapter.
Note For possibly dangerous situations, which could lead to damages to property or system failure. Important or helpful user information.
Art. no.: 2.510659 rev.: h
Reference to other guidelines.
Page 14
ARGUS PRO LifeCare
User Guide
1.7.2
Safety Notes Safety Symbols and Pictograms
1 1.7
Symbols Used on the Device and Accumulator The unit/component can be recycled.
Notified body of the CE certification (TÜV P.S.).
Attention: Consult accompanying documents.
Bluetooth general symbol.
BF symbol. The device's signal input is defibrillation-proof. CF symbol. The device's signal input is defibrillation-proof. CF symbol. This unit is classified safe for internal and external use. However, It is only defibrillation protected when used with the original SCHILLER patient cable.
Connector port for defibrillation and pacemaker electrode pads.
Symbol for the recognition of electrical and electronic equipment. Do not dispose the accumulator or the unit in household waste. Equipment/components and accessories no longer required must be disposed of in a municipally approved collection point or recycling centre. Alternatively, you can return the equipment to your supplier or SCHILLER for disposal. Improper disposal can harm the environment and human health.
Art. no.: 2.510659 rev.: h
Do not incinerate, cut or crush accumulator.
Page 15
1 1.7
Safety Notes Safety Symbols and Pictograms
ARGUS PRO LifeCare
Rechargable accumulator.
Do not short-circuit the accumulator. Accumulator temperature Operating:
-10...+40 °C
Storage:
-20...+60 °C
Expiry date of the accumulator.
1.7.3
Symbols Used on the Electrode Packaging Open electrode packaging.
Remove protective cover.
Designed for single use only; do not re-use.
Do not bend packaging.
Electrode storage temperature.
Expiry date of the electrodes.
1.7.4
Symbols used on the SpO2 Sensor Packaging Only to be used on patients greater than 30 Kgs.
Art. no.: 2.510659 rev.: h
Sensor designed for children.
Page 16
ARGUS PRO LifeCare
User Guide
1.8
Additional terms
1.8.1
Implied Authorisation
Safety Notes Additional terms
1 1.8
Possession or purchase of this device does not convey any express or implied license to use the device with replacement parts which would alone, or in combination with this device, fall within the scope of one or more patents relating to this device.
1.8.2
Terms of Warranty Your SCHILLER ARGUS PRO LifeCare is warranted against defects in material and manufacture for the duration of one year (as from date of purchase). Excluded from this guarantee is damage caused by an accident or as a result of improper handling. The warranty entitles free replacement of the defective part. Any liability for subsequent damage is excluded. The warranty is void if unauthorised or unqualified persons attempt to make repairs. In case of a defect, send the apparatus to your dealer or directly to the manufacturer. The manufacturer can only be held responsible for the safety, reliability, and performance of the apparatus if: • assembly operations, extensions, readjustments, modifications, or repairs are carried out by persons authorised by him, and • the ARGUS PRO LifeCare and approved attached equipment is used in accordance with the manufacturer's instructions.
Art. no.: 2.510659 rev.: h
There are no express or implied warranties which extend beyond the warranties hereinabove set forth. SCHILLER makes no warranty of merchantability or fitness for a particular purpose with respect to the product or parts thereof.
Page 17
2 2.1
Introduction Overview
ARGUS PRO LifeCare
2 Introduction 2.1
Overview Connector Panel
Mini SD Card Slot Carrying Handle and shock frame (end lugs for mounting) Accessory Bag Printer Connector (USB)
Speaker for system and voice indications
Defibrillator and Pacemaker pad connector
LCD Screen
Control and Function Keys
Microphone
ARGUS PRO LifeCare with Optional Shock Frame and Bag The ARGUS PRO LifeCare is a vital signs patient monitor with integrated defibrillator. It is designed to meet the specific needs of ground and air rescue services and inhouse and inter-hospital transportation. Data is viewed on an LCD and optionally printed on an external thermal printer.
The unit is powered from a removable long-life accumulator that is charged via an independent charger. A fully charged accumulator provides >250 shocks at maximum energy, or 4 hours of monitoring (dependent on parameter settings and NIBP measurement interval). An internal backup battery is also provided to ensure uninterrupted monitoring backup when the main accumulator is not present (e.g. being exchanged). This battery provides approximately two minutes of independent patient monitoring. Additional monitoring options with diagnostic capability can be provided when the ARGUS PRO LifeCare is connected to the SCHILLER ARGUS PRO Transport or the SCHILLER ARGUS PRO Panel.
Page 18
Art. no.: 2.510659 rev.: h
High and low alarms can be set if required for all measured parameters.