Schiller
Schiller AT1 User Guide rev G Nov 2005
User Guide
198 Pages
Preview
Page 1
AT-1 User Guide - English AT-1Gebrauchsanweisung -Deutsch Mode d'emploi pour l’A rançais l’ATT-1 - FFrançais Article Number a Sept. 1995, b
Jun. 1999, c
2.510171g
Feb. 2000, d Feb. 2001,
f Aug. 2004,
e Jan. 2002
g Nov. 2005
Associated Document Guide to the SCHILLER Interpretation and Measurement Program E/ D/ F
Article Number 2.510179
SCHILLER AG Altgasse 68 CH - 6341 Baar, Switzerland Phone: ++ 41 41 766 42 42 Fax: ++ 41 41 761 08 80 www.schiller.ch
AT-1User Guide - English AT-1 Gebrauchsanweisung -Deutsch Mode d'emploi pour l’A rançais l’ATT-1 - FFrançais Article Number a Sept. 1995, b
Jun. 1999, c
2.510171 g
Feb. 2000, d Feb. 2001,
f Aug. 2004,
e Jan. 2002
g Nov.2005
Associated Document Guide to the SCHILLER Interpretation and Measurement Program E/ D/ F
SCHILLER AG Altgasse 68 CH - 6341 Baar, Switzerland Phone: ++ 41 41 766 42 42 Fax: ++ 41 41 761 08 80 www.schiller.ch
Article Number 2.510179
ENGLISH
AT-1 Operating Instructions English
ENGLISH
i
AT-1 Operating Instructions English
i
Important Information
iv
Safety Information
vii
Short Form Instructions
1
Introduction
3
Modes of Operation Automatic Mode
4 4
a. without external printer connected or printer switched off
4
b. with external printer connected and switched on
4
Modes of Operation Manual Mode
5 5
Automatic Mode
6
a. without external printer connected or printer switched off
6
b. with external printer
7
Manual Mode
8
Location & Power
10
Switching On and Off
11
Keyboard
13
Settings Default Settings Language Filters Baseline Filter
14 17 18 19 19 ii
Important Information
iv
Safety Information
vii
Short Form Instructions
1
Introduction
3
Modes of Operation Automatic Mode
4 4
a. without external printer connected or printer switched off
4
b. with external printer connected and switched on
4
Modes of Operation Manual Mode
5 5
Automatic Mode
6
a. without external printer connected or printer switched off
6
b. with external printer
7
Manual Mode
8
Location & Power
10
Switching On and Off
11
Keyboard
13
Settings Default Settings Language Filters Baseline Filter
14 17 18 19 19 ii
20 21 22 24 24
Automatic Mode (ECG) Settings, Internal Printer
26
Automatic Mode (ECG) Settings, External Printer
30
Automatic Mode (ECG) Settings
34
Service Printout
36
Installing Software Options
38
Patient Cable Connections
39
Settings in Automatic Mode Patient Data Field
41 41
Measurements Heart Rate (HR) Intervals Electrical Axes
42 42 42 43
Care & Maintenance
48
Replacing the Recording Paper
50
Thermal Paper Handling
52
Trouble Shooting
53
Technical Data
ENGLISH
Mains Filter Myogram Filter Defining Lead Sequence & Printout Acoustic QRS Indication Time / Date
55
Mains Filter Myogram Filter Defining Lead Sequence & Printout Acoustic QRS Indication Time / Date
20 21 22 24 24
Automatic Mode (ECG) Settings, Internal Printer
26
Automatic Mode (ECG) Settings, External Printer
30
Automatic Mode (ECG) Settings
34
Service Printout
36
Installing Software Options
38
Patient Cable Connections
39
Settings in Automatic Mode Patient Data Field
41 41
Measurements Heart Rate (HR) Intervals Electrical Axes
42 42 42 43
Care & Maintenance
48
Replacing the Recording Paper
50
Thermal Paper Handling
52
Trouble Shooting
53
Technical Data
55 iii
ENGLISH
iii
Important Information Battery Care and Disposal Instructions Disposal Instructions and Battery Care
° ° °
DO NOT DISPOSE OF THE BATTERY BY FIRE OR INCINERATOR - DANGER OF EXPLOSION DO NOT ATTEMPT TO RECHARGE THE BATTERY - DANGER OF EXPLOSION DO NOT OPEN THE BATTERY CASING - DANGER OF ACID BURN
Only dispose of the battery in official recycling centres or municipally approved areas. Alternatively, used batteries can be returned to SCHILLER AG for disposal.
Unit Disposal Instructions Units no longer required can be returned to SCHILLER AG for disposal. Alternatively dispose of the unit in municipally approved recycling centres. This equipment has been tested and found to comply with the limits for a class A digital device, pursuant to both Part 15 of the FCC (Federal Communications Commission) Rules and the radio interference regulations of the Canadian Department of Communications. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with this instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
iv
Important Information Battery Care and Disposal Instructions Disposal Instructions and Battery Care
° ° °
DO NOT DISPOSE OF THE BATTERY BY FIRE OR INCINERATOR - DANGER OF EXPLOSION DO NOT ATTEMPT TO RECHARGE THE BATTERY - DANGER OF EXPLOSION DO NOT OPEN THE BATTERY CASING - DANGER OF ACID BURN
Only dispose of the battery in official recycling centres or municipally approved areas. Alternatively, used batteries can be returned to SCHILLER AG for disposal.
Unit Disposal Instructions Units no longer required can be returned to SCHILLER AG for disposal. Alternatively dispose of the unit in municipally approved recycling centres. This equipment has been tested and found to comply with the limits for a class A digital device, pursuant to both Part 15 of the FCC (Federal Communications Commission) Rules and the radio interference regulations of the Canadian Department of Communications. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with this instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
iv
ENGLISH
Important Information
Power Supply The mains connection is on the rear of the unit. The power supply voltage is set by the factory for 100-115V (nom. 110V) or 220-240V (nom. 230V) working. The setting is indicated by the indented metal strip on the fuse panel. Contact your dealer if the voltage needs to be changed. The mains indicator lamp on the keyboard is always lit when the unit is connected to the mains supply. The unit can either be operated from the mains supply or from the built-in rechargeable battery. Changing a Mains Fuse If it is necessary to change a fuse, always replace with the correct rating i.e. 2x200mAT for 230V, or 2x315mAT for 110V. To change a fuse press the two retaining lugs on the side of the fuse panel (situated below the mains connector on the back panel). Remove the fuse panel and replace the fuse(s). Click back the fuse panel.
v
ENGLISH
Important Information
Power Supply The mains connection is on the rear of the unit. The power supply voltage is set by the factory for 100-115V (nom. 110V) or 220-240V (nom. 230V) working. The setting is indicated by the indented metal strip on the fuse panel. Contact your dealer if the voltage needs to be changed. The mains indicator lamp on the keyboard is always lit when the unit is connected to the mains supply. The unit can either be operated from the mains supply or from the built-in rechargeable battery. Changing a Mains Fuse If it is necessary to change a fuse, always replace with the correct rating i.e. 2x200mAT for 230V, or 2x315mAT for 110V. To change a fuse press the two retaining lugs on the side of the fuse panel (situated below the mains connector on the back panel). Remove the fuse panel and replace the fuse(s). Click back the fuse panel.
v
Important Information Terms of Warranty The SCHILLER AT-1 is warranted against defects in material and manufacture for the duration of one year (as from date of purchase). Excluded from this guarantee is damage caused by an accident or as a result of improper handling. The warranty entitles free replacement of the defective part. Any liability for subsequent damage is excluded. The warranty is void if unauthorized or unqualified persons attempt to make repairs.
PHYSICIAN‘S RESPONSIBILITY
In case of a defect, contact your dealer or the manufacturer. THE AT-1 ELECTROCARDIOGRAPH IS PROVIDED FOR THE EXCLUSIVE USE OF QUALIFIED PHYSICIANS OR PERSONNEL UNDER THEIR DIRECT SUPERVISION. THE NUMERICAL AND GRAPHICAL RESULTS FROM A RECORDING MUST BE EXAMINED WITH RESPECT TO THE PATIENTS OVERALL CLINICAL CONDITION. THE RECORDING PREPARATION QUALITY AND THE GENERAL RECORDED DATA QUALITY, WHICH COULD EFFECT THE REPORT DATA ACCURACY, MUST ALSO BE TAKEN INTO ACCOUNT.
The manufacturer can only be held responsible for the safety, reliability, and performance of the apparatus if: *
assembly operations, extensions, readjustments, modifications, or repairs are carried out by persons authorized by him, and
*
the AT-1 and approved attached equipment are used in accordance with the manufacturers instructions.
IT IS THE PHYSICIANS RESPONSIBILITY TO MAKE THE DIAGNOSIS OR TO OBTAIN EXPERT OPINION ON THE RESULTS, AND TO INSTITUTE CORRECT TREATMENT IF INDICATED.
THERE ARE NO EXPRESS OR IMPLIED WARRANTIES WHICH EXTEND BEYOND THE WARRANTIES HEREINABOVE SET FORTH. SCHILLER MAKES NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO THE PRODUCT OR PARTS THEREOF.
FEDERAL LAW IN THE USA RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN
NOTES
About this Handbook The philosophy of SCHILLER is one of continuous improvement. Our aim is to provide the user with the most up-to-date information and the latest technological developments. Your suggestions and comments are welcome on all SCHILLER documentation. Please contact the SCHILLER Technical Documentation Department.
vi
Important Information
Terms of Warranty The SCHILLER AT-1 / AT-1 smartprint is warranted against defects in material and manufacture for the duration of one year (as from date of purchase). Excluded from this guarantee is damage caused by an accident or as a result of improper handling. The warranty entitles free replacement of the defective part. Any liability for subsequent damage is excluded. The warranty is void if unauthorized or unqualified persons attempt to make repairs. In case of a defect, contact your dealer or the manufacturer.
PHYSICIAN‘S RESPONSIBILITY
The manufacturer can only be held responsible for the safety, reliability, and performance of the apparatus if: *
assembly operations, extensions, readjustments, modifications, or repairs are carried out by persons authorized by him, and
*
the AT-1 / AT-1 smartprint and approved attached equipment are used in accordance with the manufacturers instructions.
THE AT-1 / AT-1 SMARTPRINT ELECTROCARDIOGRAPH IS PROVIDED FOR THE EXCLUSIVE USE OF QUALIFIED PHYSICIANS OR PERSONNEL UNDER THEIR DIRECT SUPERVISION. THE NUMERICAL AND GRAPHICAL RESULTS FROM A RECORDING MUST BE EXAMINED WITH RESPECT TO THE PATIENTS OVERALL CLINICAL CONDITION. THE RECORDING PREPARATION QUALITY AND THE GENERAL RECORDED DATA QUALITY, WHICH COULD EFFECT THE REPORT DATA ACCURACY, MUST ALSO BE TAKEN INTO ACCOUNT.
THERE ARE NO EXPRESS OR IMPLIED WARRANTIES WHICH EXTEND BEYOND THE WARRANTIES HEREINABOVE SET FORTH. SCHILLER MAKES NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO THE PRODUCT OR PARTS THEREOF.
IT IS THE PHYSICIANS RESPONSIBILITY TO MAKE THE DIAGNOSIS OR TO OBTAIN EXPERT OPINION ON THE RESULTS, AND TO INSTITUTE CORRECT TREATMENT IF INDICATED.
About this Handbook
FEDERAL LAW IN THE USA RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN
The philosophy of SCHILLER is one of continuous improvement. Our aim is to provide the user with the most up-to-date information and the latest technological developments. Your suggestions and comments are welcome on all SCHILLER documentation. Please contact the SCHILLER Technical Documentation Department.
vi
DO NOT USE SOLVENT CLEANERS. THIS UNIT COMPLIES WITH EMC REGULATIONS FOR MEDICAL PRODUCTS WHICH AFFORDS PROTECTION AGAINST EMISSIONS AND ELECTRICAL INTERFERENCE. HOWEVER SPECIAL CARE MUST BE EXERCISED WHEN THE UNIT IS USED WITH HIGH FREQUENCY EQUIPMENT.
ENGLISH
Safety Information
BEFORE USING THE UNIT, ENSURE THAT AN INTRODUCTION REGARDING THE UNIT FUNCTIONS AND THE SAFETY PRECAUTIONS HAS BEEN PROVIDED BY A PRODUCT REPRESENTATIVE.
CAUTIONS SWITCH THE UNIT OFF BEFORE CLEANING AND DISCONNECT FROM THE MAINS. THE DEVICE MUST ONLY BE OPERATED USING BATTERY POWER IF THE EARTH CONNECTION IS SUSPECT OR IF THE MAINS LEAD IS DAMAGED OR SUSPECTED OF BEING DAMAGED. USE ONLY ACCESSORIES AND OTHER PARTS RECOMMENDED OR SUPPLIED BY SCHILLER AG. USE OF OTHER THAN RECOMMENDED OR SUPPLIED PARTS MAY RESULT IN INJURY INACCURATE INFORMATION AND / OR DAMAGE TO THE UNIT. IT MUST BE ENSURED THAT NEITHER THE PATIENT NOR THE ELECTRODES (INCLUDING THE NEUTRAL ELECTRODE) COME INTO CONTACT WITH OTHER PERSONS OR CONDUCTING OBJECTS (EVEN IF THESE ARE EARTHED). THERE IS NO DANGER WHEN USING THE ECG UNIT FOR A PACEMAKER PATIENT OR WITH SIMULTANEOUS USE OF OTHER ELECTRICAL STIMULATION EQUIPMENT. HOWEVER, THE STIMULATION UNITS SHOULD ONLY BE USED AT A SUFFICIENT DISTANCE FROM THE ELECTRODES. IN CASE OF DOUBT, THE PATIENT SHOULD BE DISCONNECTED FROM THE RECORDER. IF SEVERAL UNITS ARE COUPLED THERE IS A DANGER OF SUMMATION OF LEAKAGE CURRENTS. WHEN OPERATING SEVERAL DEVICES FOR MEDICAL AND NON-MEDICAL APPLICATION DO NOT USE ANY EXTENSION CABLES OR DISTRIBUTION BOXES FOR THE CONNECTION. EARTH MUST BE CONNECTED WHEN THE EXTERNAL PRINTER IS USED.
vii
NOTES DO NOT USE SOLVENT CLEANERS. THIS UNIT COMPLIES WITH EMC REGULATIONS FOR MEDICAL PRODUCTS WHICH AFFORDS PROTECTION AGAINST EMISSIONS AND ELECTRICAL INTERFERENCE. HOWEVER SPECIAL CARE MUST BE EXERCISED WHEN THE UNIT IS USED WITH HIGH FREQUENCY EQUIPMENT. BEFORE USING THE UNIT, ENSURE THAT AN INTRODUCTION REGARDING THE UNIT FUNCTIONS AND THE SAFETY PRECAUTIONS HAS BEEN PROVIDED BY A PRODUCT REPRESENTATIVE.
CAUTIONS SWITCH THE UNIT OFF BEFORE CLEANING AND DISCONNECT FROM THE MAINS. THE DEVICE MUST ONLY BE OPERATED USING BATTERY POWER IF THE EARTH CONNECTION IS SUSPECT OR IF THE MAINS LEAD IS DAMAGED OR SUSPECTED OF BEING DAMAGED. USE ONLY ACCESSORIES AND OTHER PARTS RECOMMENDED OR SUPPLIED BY SCHILLER AG. USE OF OTHER THAN RECOMMENDED OR SUPPLIED PARTS MAY RESULT IN INJURY INACCURATE INFORMATION AND / OR DAMAGE TO THE UNIT. IT MUST BE ENSURED THAT NEITHER THE PATIENT NOR THE ELECTRODES (INCLUDING THE NEUTRAL ELECTRODE) COME INTO CONTACT WITH OTHER PERSONS OR CONDUCTING OBJECTS (EVEN IF THESE ARE EARTHED). THERE IS NO DANGER WHEN USING THE ECG UNIT FOR A PACEMAKER PATIENT OR WITH SIMULTANEOUS USE OF OTHER ELECTRICAL STIMULATION EQUIPMENT. HOWEVER, THE STIMULATION UNITS SHOULD ONLY BE USED AT A SUFFICIENT DISTANCE FROM THE ELECTRODES. IN CASE OF DOUBT, THE PATIENT SHOULD BE DISCONNECTED FROM THE RECORDER. IF SEVERAL UNITS ARE COUPLED THERE IS A DANGER OF SUMMATION OF LEAKAGE CURRENTS. WHEN OPERATING SEVERAL DEVICES FOR MEDICAL AND NON-MEDICAL APPLICATION DO NOT USE ANY EXTENSION CABLES OR DISTRIBUTION BOXES FOR THE CONNECTION. EARTH MUST BE CONNECTED WHEN THE EXTERNAL PRINTER IS USED.
vii
ENGLISH
Safety Information
Safety Information
WARNINGS TO PREVENT ELECTRIC SHOCK DO NOT DISASSEMBLE THE UNIT. NO SERVICEABLE PARTS INSIDE. REFER SERVICING TO QUALIFIED PERSONNEL ONLY. DO NOT USE THIS UNIT IN AREAS WHERE THERE IS ANY DANGER OF EXPLOSION OR THE PRESENCE OF FLAMMABLE GASES SUCH AS ANAESTHETIC AGENTS. THIS PRODUCT IS NOT DESIGNED FOR STERILE USE. THIS PRODUCT IS NOT DESIGNED FOR OUTDOOR USE. DO NOT, UNDER ANY CIRCUMSTANCES, IMMERSE THE UNIT OR CABLE ASSEMBLIES IN LIQUID. DO NOT USE HIGH TEMPERATURE STERILISATION PROCESSES (SUCH AS AUTOCLAVING). DO NOT USE E-BEAM OR GAMMA RADIATION STERILISATION. THIS UNIT IS CF CLASSIFIED ACCORDING TO IEC EN 60601-1. THIS MEANS THAT THE PATIENT CONNECTION IS FULLY ISOLATED AND DEFIBRILLATION PROTECTED. SCHILLER CAN ONLY GUARANTEE PROTECTION AGAINST DEFIBRILLATION VOLTAGE HOWEVER, WHEN THE ORIGINAL SCHILLER PATIENT CABLE IS USED. DO NOT TOUCH THE CASING DURING DEFIBRILLATION. IF THE PATIENT CABLE SHOULD BECOME DEFECTIVE AFTER DEFIBRILLATION, LEAD OFF WILL BE DISPLAYED AND AN ACOUSTIC ALARM GIVEN.
viii
Safety Information WARNINGS TO PREVENT ELECTRIC SHOCK DO NOT DISASSEMBLE THE UNIT. NO SERVICEABLE PARTS INSIDE. REFER SERVICING TO QUALIFIED PERSONNEL ONLY. DO NOT USE THIS UNIT IN AREAS WHERE THERE IS ANY DANGER OF EXPLOSION OR THE PRESENCE OF FLAMMABLE GASES SUCH AS ANAESTHETIC AGENTS. THIS PRODUCT IS NOT DESIGNED FOR STERILE USE. THIS PRODUCT IS NOT DESIGNED FOR OUTDOOR USE. DO NOT, UNDER ANY CIRCUMSTANCES, IMMERSE THE UNIT OR CABLE ASSEMBLIES IN LIQUID. DO NOT USE HIGH TEMPERATURE STERILISATION PROCESSES (SUCH AS AUTOCLAVING). DO NOT USE E-BEAM OR GAMMA RADIATION STERILISATION. THIS UNIT IS CF CLASSIFIED ACCORDING TO IEC EN 60601-1. THIS MEANS THAT THE PATIENT CONNECTION IS FULLY ISOLATED AND DEFIBRILLATION PROTECTED. SCHILLER CAN ONLY GUARANTEE PROTECTION AGAINST DEFIBRILLATION VOLTAGE HOWEVER, WHEN THE ORIGINAL SCHILLER PATIENT CABLE IS USED. DO NOT TOUCH THE CASING DURING DEFIBRILLATION. IF THE PATIENT CABLE SHOULD BECOME DEFECTIVE AFTER DEFIBRILLATION, LEAD OFF WILL BE DISPLAYED AND AN ACOUSTIC ALARM GIVEN.
viii
AT-1
Manual ECG Recording (internal printer only) Prepare skin, hook up patient.
Automatic ECG Recording with external printer connected and switched on: Prepare skin, hook up patient.
Switch unit on, press ON
ENGLISH
Shor orm Instructions Shortt FForm
Press MAN START Change lead group with Press STOP to stop the printout.
Switch unit on, press ON Press AUTO to record and print externally
Electrode hook-up check
with external printer NOT connected or switched off: To record and print on internal printer: Prepare skin, hook up patient.
electrode dc offsets. Best results are obtained when the electrode voltage readings (right column) are between +50mV and -50mV. Filter On/Off
For AUTO Format 1 press For Auto Format 2 press
for print-out of
Press
Press and then
Press COPY for additional copies.
to switch the (Myogram) filter On / Off.
System Configuration to print system
Press settings.
AT-1 User Guide Art. No. 2.510171 g
AT-1
1
Manual ECG Recording (internal printer only) Prepare skin, hook up patient.
Automatic ECG Recording with external printer connected and switched on: Prepare skin, hook up patient.
Switch unit on, press ON
ENGLISH
Shor orm Instructions Shortt FForm
Press MAN START Change lead group with Press STOP to stop the printout.
Switch unit on, press ON Press AUTO to record and print externally
Electrode hook-up check
with external printer NOT connected or switched off: To record and print on internal printer: Prepare skin, hook up patient.
electrode dc offsets. Best results are obtained when the electrode voltage readings (right column) are between +50mV and -50mV. Filter On/Off
For AUTO Format 1 press For Auto Format 2 press
for print-out of
Press
Press and then
Press COPY for additional copies.
to switch the (Myogram) filter On / Off.
System Configuration Press
to print system
settings. AT-1 User Guide Art. No. 2.510171 g
1
Introduction
2
Introduction
2
NOTE:
ENGLISH
Introduction The only difference between A tprint is that A ATT-1 and A ATT-1 smar smartprint ATT-1 tprint has a connector for an external ink-jet printer smartprint printer,, which smar makes it possible to generate printouts in A4 size.
The ECG unit is a 3-channel ECG recorder with all (12) ECG signals simultaneously processed to provide instant ECG recordings. Three automatic recording modes - two for the internal printer and one for the external printer - can be individually preset to enable one button ECG recording of preferred print formats. Individual lamps are provided to give power, paper error, filter, lead group and lead off indications. In addition, any detected disturbance (i.e. loose electrode or end of paper), gives an audible alarm and the corresponding indicator lamp flashes. The ECG unit includes the following features: • Low weight and compact dimensions • Printout from integrated quality thermal printer - OR - from external A4 inkjet printer • Built-in rechargeable battery for mains-independent use - 2 hours normal use on one battery charge • Simple one key operation for main functions • Automatic or manual recording modes • Selectable printing formats - external and internal printer • Choice of 10 languages (AT-1) and 9 languages (AT-1 smartprint) for printing. • Interpretation program option (including measurements) for children and adults
AT-1 User Guide Art. No. 2.510171 g
3
NOTE:
ENGLISH
Introduction The only difference between A tprint is that A ATT-1 and A ATT-1 smar smartprint ATT-1 tprint has a connector for an external ink-jet printer smartprint printer,, which smar makes it possible to generate printouts in A4 size.
The ECG unit is a 3-channel ECG recorder with all (12) ECG signals simultaneously processed to provide instant ECG recordings. Three automatic recording modes - two for the internal printer and one for the external printer - can be individually preset to enable one button ECG recording of preferred print formats. Individual lamps are provided to give power, paper error, filter, lead group and lead off indications. In addition, any detected disturbance (i.e. loose electrode or end of paper), gives an audible alarm and the corresponding indicator lamp flashes. The ECG unit includes the following features: • Low weight and compact dimensions • Printout from integrated quality thermal printer - OR - from external A4 inkjet printer • Built-in rechargeable battery for mains-independent use - 2 hours normal use on one battery charge • Simple one key operation for main functions • Automatic or manual recording modes • Selectable printing formats - external and internal printer • Choice of 10 languages (AT-1) and 9 languages (AT-1 smartprint) for printing. • Interpretation program option (including measurements) for children and adults AT-1 User Guide Art. No. 2.510171 g
3
Modes of Operation Automatic Mode a.
without external printer connected or printer switched off
Automatic Mode provides a printout giving 10 seconds of ECG recording of all 12 leads with a choice of 2 different formats (only for S version). The following can be programmed freely for each of the 2 formats before recording: • Lead Format • Chart Speed • With the optional interpretation program installed it is also possible to select the measurement table, average cycles with optional markings and interpretation statements for the printout. For further information see paragraph `Automatic Mode (ECG) Settings` on page 26. b. with external printer connected and switched on Only one Automatic Mode format possible. Other details as above. For further information see paragraph `Automatic Mode (ECG) Settings` on page 30.
4
Modes of Operation Automatic Mode a. without external printer connected or printer switched off Automatic Mode provides a printout giving 10 seconds of ECG recording of all 12 leads with a choice of 2 different formats (only for S version). The following can be programmed freely for each of the 2 formats before recording: • Lead Format • Chart Speed • With the optional interpretation program installed it is also possible to select the measurement table, average cycles with optional markings and interpretation statements for the printout. For further information see paragraph `Automatic Mode (ECG) Settings` on page 26. b. with external printer connected and switched on Only one Automatic Mode format possible. Other details as above. For further information see paragraph `Automatic Mode (ECG) Settings` on page 30.
4
ENGLISH
Modes of Operation Manual Mode Manual Mode provides a real-time printout of 3 selected leads and is only possible on the internal thermal printer. The following can be freely selected before or during recording: • Lead Group • Chart Speed • Sensitivity • Myogram Filter For further information see paragraph `Manual Mode` on page 8.
AT-1 User Guide Art. No. 2.510171 g
5
ENGLISH
Modes of Operation Manual Mode Manual Mode provides a real-time printout of 3 selected leads and is only possible on the internal thermal printer. The following can be freely selected before or during recording: • Lead Group • Chart Speed • Sensitivity • Myogram Filter For further information see paragraph `Manual Mode` on page 8.
AT-1 User Guide Art. No. 2.510171 g
5
Automatic Mode a.
without external printer connected or printer switched off
In automatic mode mode, a full 12-lead ECG is printed in one of two predefined formats with selectable sensitivity. These formats (see Settings) are selected by the user to suit his specific needs and requirements. When the AUTO GAIN key is pressed before recording in automatic mode, the unit detects very large waveform amplitudes and sets the sensitivity for the extremity and/or precordial leads accordingly to reduce the overlapping of traces. To start the automatic ECG recording in Format 1, press the AUTO START key:
To start the automatic recording in Format 2, press the ALT key followed by the AUTO START key: + Depending upon which option is included, the printout gives the following: • ECG recording of all leads in either Standard or Cabrera format according to selection • Sensitivity
6
• Heart Rate
Automatic Mode a.
without external printer connected or printer switched off
In automatic mode mode, a full 12-lead ECG is printed in one of two predefined formats with selectable sensitivity. These formats (see Settings) are selected by the user to suit his specific needs and requirements. When the AUTO GAIN key is pressed before recording in automatic mode, the unit detects very large waveform amplitudes and sets the sensitivity for the extremity and/or precordial leads accordingly to reduce the overlapping of traces. To start the automatic ECG recording in Format 1, press the AUTO START key:
To start the automatic recording in Format 2, press the ALT key followed by the AUTO START key: + Depending upon which option is included, the printout gives the following: • ECG recording of all leads in either Standard or Cabrera format according to selection • Sensitivity
6
• Heart Rate
• • • • • • •
ENGLISH
Automatic Mode Speed Filter Settings Interpretation statements (with option C) Average Cycles (with options M, C) Intervals (with options M, C) Axes (with options M, C) Detailed Measurement Table (with options M, C)
To obtain an extra printout of the ECG recording in Format 1, simply press the COPY key: To obtain an extra printout of the second format, press the ALT key followed by the COPY key: Note:
The Auto mode settings for the two formats are detailed in the paragraph entitled ' Automatic Mode (ECG) Settings` on page 26.
b. with external printer With the external printer connected and switched on, only one Auto Mode format (selectable) is available. ART kkey: ey: To star startt the automatic ECG recording in Automatic Mode, press the AUTO ST START
For further information see paragraph `Automatic Mode (ECG) Settings` on page 30. AT-1 User Guide Art. No. 2.510171 g
7
• • • • • • •
ENGLISH
Automatic Mode Speed Filter Settings Interpretation statements (with option C) Average Cycles (with options M, C) Intervals (with options M, C) Axes (with options M, C) Detailed Measurement Table (with options M, C)
To obtain an extra printout of the ECG recording in Format 1, simply press the COPY key: To obtain an extra printout of the second format, press the ALT key followed by the COPY key: Note:
The Auto mode settings for the two formats are detailed in the paragraph entitled ' Automatic Mode (ECG) Settings` on page 26.
b. with external printer With the external printer connected and switched on, only one Auto Mode format (selectable) is available. To star ART kkey: ey: startt the automatic ECG recording in Automatic Mode, press the AUTO ST START
For further information see paragraph `Automatic Mode (ECG) Settings` on page 30. AT-1 User Guide Art. No. 2.510171 g
7
Manual Mode Manual mode provides a real-time ECG printout of 3 selected leads with full control of parameter selection. This is only possible on the internal thermal printer. To start the manual recording of a real-time ECG, press the MAN START Printout key
To stop the manual recording (printout) press the STOP key
The printout provides you with the following: • Three (selected) leads with lead identification. • On the lower edge, the chart speed and filter settings (if on). • At the top, the heart rate as current average of 4 beats, trace sensitivity. The following can be freely chosen during or before the recording: Lead Group
by means of the LEAD FORWARD and LEAD BACKWARD key
8
Manual Mode Manual mode provides a real-time ECG printout of 3 selected leads with full control of parameter selection. This is only possible on the internal thermal printer. To start the manual recording of a real-time ECG, press the MAN START Printout key
To stop the manual recording (printout) press the STOP key
The printout provides you with the following: • Three (selected) leads with lead identification. • On the lower edge, the chart speed and filter settings (if on). • At the top, the heart rate as current average of 4 beats, trace sensitivity. The following can be freely chosen during or before the recording: Lead Group
by means of the LEAD FORWARD and LEAD BACKWARD key
8
ENGLISH
Manual Mode The following lead groups are selectable: •
I, II, III
aVR, aVL, aVF
(Cabrera: aVL, I, -aVR / II, aVF, III) •
V1, V2, V3
/
V4, V5, V6
Chart Speed
Select speed 5, 25 or 50 mm/s by means of the SPEED keys:
Sensitivity
Select 5, 10 or 20 mm/mV by means of the SENSITIVITY keys:
Myogram Filter
Switch the filter ON or OFF with the FILTER key: 25 Hz or 35 Hz is displayed on the bottom line of the printout when the filter is switched on. To re-centre the ECG traces, press the 1mV key
Recentering
AT-1 User Guide Art. No. 2.510171 g
9
ENGLISH
Manual Mode The following lead groups are selectable: •
I, II, III
aVR, aVL, aVF
(Cabrera: aVL, I, -aVR / II, aVF, III) •
V1, V2, V3
/
V4, V5, V6
Chart Speed
Select speed 5, 25 or 50 mm/s by means of the SPEED keys:
Sensitivity
Select 5, 10 or 20 mm/mV by means of the SENSITIVITY keys:
Myogram Filter
Switch the filter ON or OFF with the FILTER key: 25 Hz or 35 Hz is displayed on the bottom line of the printout when the filter is switched on.
Recentering
To re-centre the ECG traces, press the 1mV key
AT-1 User Guide Art. No. 2.510171 g
9
Location & PPower ower Location Do not keep or operate the apparatus in a wet, moist, or dusty environment. Also, avoid exposure to direct sunlight or heat from other sources. Do not allow the unit to come into contact with acidic vapours or liquids, as such contact may cause irreparable damage. The unit should not be placed near X-ray or diathermy units, large transformers or motors. The unit must be placed on a flat surface and must not be operated in areas where there is any danger of explosion.
Power Supply The mains connection is on the rear of the unit. The mains indicator lamp on the keyboard is always lit when the unit is connected to the mains supply. The unit can either be operated from the mains supply or from the built-in rechargeable battery. The power source is indicated by the respective indicator lamp. When battery capacity is limited, the battery symbol flashes on and off. To recharge the battery, connect the apparatus to the mains supply by means of the supplied power cable. A totally discharged battery needs less than 15 hours to be fully recharged (60% in less than 3 hours, 90% in less than 7 hours). A fully charged battery gives approximately 2 hours of normal use. The unit can remain connected to the mains supply without any danger of damage to either the battery or the unit.
10
Location & PPower ower Location Do not keep or operate the apparatus in a wet, moist, or dusty environment. Also, avoid exposure to direct sunlight or heat from other sources. Do not allow the unit to come into contact with acidic vapours or liquids, as such contact may cause irreparable damage. The unit should not be placed near X-ray or diathermy units, large transformers or motors. The unit must be placed on a flat surface and must not be operated in areas where there is any danger of explosion.
Power Supply The mains connection is on the rear of the unit. The mains indicator lamp on the keyboard is always lit when the unit is connected to the mains supply. The unit can either be operated from the mains supply or from the built-in rechargeable battery. The power source is indicated by the respective indicator lamp. When battery capacity is limited, the battery symbol flashes on and off. To recharge the battery, connect the apparatus to the mains supply by means of the supplied power cable. A totally discharged battery needs less than 15 hours to be fully recharged (60% in less than 3 hours, 90% in less than 7 hours). A fully charged battery gives approximately 2 hours of normal use. The unit can remain connected to the mains supply without any danger of damage to either the battery or the unit.
10
ENGLISH
Switching On and Off The AT-1 is switched on with the green ON key and off by means of the red OFF key The unit is automatically switched off after 5 minutes (30 seconds if battery capacity is limited) if no key is pressed and the patient cable is not connected.
Potential Equalisation If the AT-1 is used in conjunction with other patient connected equipment, we recommend that the potential equalisation stud on the rear of the unit is connected to the building common ground with the yellow/green ground cable (Part-no. 2.310005). When working from an emergency vehicle, the vehicle common ground can be used.
WARNING ! THE UNIT MUST BE GROUNDED WHEN THE EXTERNAL PRINTER IS CONNECTED.
AT-1 User Guide Art. No. 2.510171 g
11
ENGLISH
Switching On and Off The AT-1 is switched on with the green ON key and off by means of the red OFF key The unit is automatically switched off after 5 minutes (30 seconds if battery capacity is limited) if no key is pressed and the patient cable is not connected.
Potential Equalisation If the AT-1 is used in conjunction with other patient connected equipment, we recommend that the potential equalisation stud on the rear of the unit is connected to the building common ground with the yellow/green ground cable (Part-no. 2.310005). When working from an emergency vehicle, the vehicle common ground can be used.
WARNING ! THE UNIT MUST BE GROUNDED WHEN THE EXTERNAL PRINTER IS CONNECTED.
AT-1 User Guide Art. No. 2.510171 g
11