seca 220 Measuring rod for seca column scales
TABLE OF CONTENTS English 1. Device description... 3
7. Function Check... 17
1.1 Intended use... 3 1.2 Functional description... 3 1.3 User qualification... 3 Assembly... 3 Operation... 4
8. Troubleshooting... 17
2. Safety information... 4 2.1 Safety information in these instructions for use... 4 2.2 Basic safety precautions... 4 3. Overview... 5 3.1 Controls and mounting elements... 5 3.2 Marking on the device and on the rating plate... 6 3.3 Marking on the packaging... 7 4. Before you really get started…. . . 8 4.1 Scope of delivery... 8 4.2 Mounting the measuring rod . . . 9 Selecting the mounting position... 9 Attaching spacer to column . . 10 Attaching measuring rod to spacer... 11 4.3 Verifying position of measuring rod... 12 Checking position... 12 Adjusting position... 13 5. Performing length measurement... 14 5.1 Measuring patients smaller than 130.5 cm... 14 5.2 Measuring patients taller than 130.5 cm... 15 6. Hygiene Treatment... 16 6.1 Cleaning... 16 6.2 Disinfecting... 16 6.3 Sterilizing... 16
2•
9. Servicing... 17 10. Technical Data... 18 10.1 General technical data... 18 10.2 Technical data, length measuring... 18 11. Warranty... 19
English
1. DEVICE DESCRIPTION 1.1 Intended use The medical measuring rod seca 220 is used with seca column scales primarily in hospitals, medical practices and in-patient care facilities in accordance with national regulations. In combination with a scale the measuring rod serves as a complete measuring station and is therefore used for conventional determination of weight and height as well as establishment of general state of nutrition; It assists the physician supervising treatment in making a diagnosis or deciding on a course of treatment. To make an accurate diagnosis, however, other specific examinations have to be ordered by the physician and their results taken into account, in addition to determining weight and height values.
1.2 Functional description The measuring rod is compatible to a range of seca column scales. The measuring rod consists of two telescopic elements: • Lower telescopic element: Measuring range 1: 60 cm -130.5 cm • Upper telescopic element: Measuring range 2: 130.5 cm - 200 cm The head slide of the measuring rod is locked at the top of the upper telescopic element to measure persons taller than 130.5 cm. The head slide needs to be released to measure persons smaller than 130.5 cm. The measuring flap on the head slide can be folded down.
1.3 User qualification Assembly
Devices that are shipped partially assembled may only be mounted by sufficiently qualified persons such as specialist dealers, hospital technicians or seca service technicians.
•3
Operation
The device may only be operated by healthcare professionals.
2. SAFETY INFORMATION 2.1 Safety information in these instructions for use CAUTION! Used to identify a hazardous situation. If you fail to take note of this information, minor to moderate injuries may result. ATTENTION! Used to identify possible incorrect usage of the device. If you fail to take note of this information, you may damage the device, or the measured results may be incorrect. NOTE: Includes additional information about use of the device.
2.2 Basic safety precautions • Please observe the information in this user manual. • Keep this user manual in a safe place. This user manual is part of the product and must be readily available. CAUTION! Patient hazard, damage to device
• Use the supplied assembly material only. • Fold down the measuring slide after each measurement. • Have the device serviced regularly as described in the respective section of this document. • Technical changes to the device are not permitted. The unit contains no user-serviceable parts. Make sure that maintenance and repair are only carried out by an authorised seca service partner. You can find your local service partner at www.seca.com or send an email to service@seca.com.
4•
English • Make sure you only use genuine seca accessories and spare parts. Otherwise the warranty provided by seca will become null and void.
3. OVERVIEW 3.1 Controls and mounting elements
4
5
3 2 6 1
155.5 cm
7
11
13 12 10 9
No. 1 2
Element Lower telescopic element Upper telescopic element
3
Measuring flap
8
Function Measuring range 1: 60 cm - 130.5 cm Measuring range 2: 130.5 cm - 200 cm • To be positionned on top of the patient‘s head • Can be folded down Overview • 5
No.
Element
4
Lock
5
Head slide
6 7 8
Read-off line 1 (red) Read-off line 2 (red) Spacer
9
Carriage
10
Guide pins
11
Mounting range
12
Fixing hole
13
Guide holes
Function • Locks head slide at top of upper telescopic element for measurement in measuring range 2 • Releases head slide for measurement on lower telescopic element (measuring range 1) • Locked at top of upper telescopic element • To be released for measurement on lower telescopic elment (measuring range 1) For measurements in measuring range 1 For measurement in measuring range 2 Connects the measuring rod to the scale Slides vertically • For mounting the spacer on the scale • For adjusting the measuring rod • Engage in the guide holes of the measuring rod • Ensure that the measuring rod is mounted in a vertical position • 2 mounting positions at the rear for mechanical seca scales • 1 mounting postion on the front for electronic seca scales For details see “Selecting the mounting position” on page 9 The fixing screw of the measuring rod is inserted in this hole The guide pins of the spacer are inserted here
3.2 Marking on the device and on the rating plate Text/Symbol Model xx/xx
Meaning Model number Batch number, engraved on back of device Follow instructions for use
0123
6•
Device conforms to EC directives • 0123: Notified body for medical products: TÜV SÜD Product Service
English
3.3 Marking on the packaging
Protect from moisture Arrows indicate top of product. Transport and store in an upright position. Fragile Do not throw or drop. Permitted min. and max. temperature for transport and storage Permitted min. and max. moisture for transport and storage Packaging material can be disposed of through recycling programs
Overview • 7
4. BEFORE YOU REALLY GET STARTED… 4.1 Scope of delivery
a
b c d
No.
a b c d
8•
Component Measuring rod seca 220 Spacer connecting measuring rod to column of scale Screw M5 x 30 mm Connects spacer to scale Screw M6 x 13 mm (total length: 22 mm) Connects measuring rod to spacer Instructions for use, not shown
Pcs. 1 1 2 1 1
English
4.2 Mounting the measuring rod Selecting the mounting position
To ensure the measured results are correct, the position of the measuring rod varies depending on the model of the scale. The following table gives the mounting position for your seca column scale: • First row: Mounting positions of the spacer and the measuring rod on the scale • Left-hand column: Mounting positions of the measuring rod on the spacer
Position at spacer
Position at scale
A B C
A seca 711 seca 712
seca 700 seca 711-220 kg
-
-
seca 709 seca 710
-
-
-
B
C
Electronic column scales, e.g. seca 704
Before you really get started… • 9
Attaching spacer to column
To mount the spacer on the column of the scale, proceed as follows.
1.
1. Slide the carriage so you can insert a screw in the bottom hole. 2. Line up the spacer with the holes in the column. 2.
3. Insert a screw in the bottom hole of the spacer and tighten securely.
3. 4. Slide the carriage so you can insert a screw in the top hole. 4.
5. Insert a screw in the top hole of the spacer and tighten securely.
5.
6. Slide the carriage into the centre.
6.
10 •
English
Attaching measuring rod to spacer
CAUTION! False measurement results
No proper measurement results will be possible if the mounting position selected is not suitable for your column scale. Make sure to select the correct mounting position for your column scale (see “Selecting the mounting position” on page 9.) 155 cm
1. Pull the upper telescopic element out to the 155 cm mark. The holes in the lower telescopic element are exposed.
2. Adjust the measuring rod as provided for your model of scale (here: position for models seca 709 und seca 710).
3. Position the measuring rod on the spacer at the height provided for your model of scale (here: position for models seca 709 und seca 710). 4. Check that the guide pins of the spacer are inserted in the guide holes of the measuring rod. 5. Screw the measuring rod to the spacer. 6. Check the position of the measuring rod (see “Verifying position of measuring rod” on page 12).
Before you really get started… • 11
4.3 Verifying position of measuring rod Checking position
After mounting the measuring rod, you must check its position and correct if necessary. To do so, proceed as follows: 1. Unfold the measuring flap.
155 cm
2. Pull the upper telescopic element out to the 155 cm-mark. The fixing screw in the lower telescopic element is exposed so that you can correct the position of the measuring rod if necessary. 3. Measure the distance between the underside of the measuring flap and the patient standing area, e.g. using a tape measure. NOTE: The area on which the patient stands for height measurement is the weighing platform or the floor, depending on the model of the scale.
155 cm
155 cm
155 cm
Requirement: The measured distance (reference dimension) between the underside of the measuring flap and the patient standing area is 155 cm.
4. Should the value measured deviate from the reference dimension, note down the deviation and correct the vertical position of the measuring rod as described in the following section. 12 •
English
Adjusting position
To correct the position of the measuring rod proceed as follows: 1. Slightly loosen the screw on the measuring rod.
2. Adjust the measuring rod together with the carriage of the spacer according to the measured deviation from the reference dimension. 3. Tighten the screw on the measuring rod. 4. Check the distance between the measuring flap and patient standing area once again. 5. If necessary, re-adjust the position of the measuring rod. NOTE: • If you are unable to set the specified reference dimension, check that you have selected the correct mounting position for your seca column scale (see “Selecting the mounting position” on page 9.). • Your seca service partner will also be happy to help. You can find your local seca service partner at www.seca.com or just send an email to service@seca.com.
Before you really get started… • 13
5. PERFORMING LENGTH MEASUREMENT CAUTION! Patient hazard
Allways perform a function check as described in section “Function Check” on page 17 before using the device.
5.1 Measuring patients smaller than 130.5 cm 1. Unlock the head slide by pressing on the lock. 2. Unfold the measuring flap so it engages in a horizontal position. 3. Ask the patient to step under the measuring flap: – Back to the measuring rod – Body kept straight – Head straight: So-called "Frankfurt plane" parallel to the measuring flap 4. Move the head slide on the lower telescopic element until the measuring flap rests on the patient's head. 5. Read off the measured result at the read-off line (read-off line 1) of the headslide (here: 113.5 cm). 6. Ask the patient to step away from the measuring rod. 7. Fold the measuring flap down. 8. Move up the head slide until it engages in the lock.
113.5 cm
14 •
English
5.2 Measuring patients taller than 130.5 cm 1. Unfold the measuring flap so it engages in a horizontal position. 2. Pull the upper telescopic element out until the patient can comfortably stand under the measuring flap. 3. Ask the patient to step under the measuring flap: – Back to the measuring rod – Body kept straight – Head straight: So-called "Frankfurt plane" parallel to the measuring flap 4. Slide down the upper telescopic element until the measuring flap rests on the patient's head. 5. Read off the measured result at the read-off line (read-off line 2) of the lower telescopic element (here: 141.5 cm). 6. Ask the patient to step away from the measuring rod. 7. Slide the upper telescopic element to the lowest position. 8. Fold the measuring flap down.
141.5 cm
Performing length measurement • 15
6. HYGIENE TREATMENT ATTENTION! Damage to device
Inappropriate detergents may damage the sensitive surfaces of the devices. Do not use caustic or abrasive detergents. Do not use spirits or benzine.
6.1 Cleaning Use a soft cloth dampened with mild soapsuds to clean the surfaces of the device.
6.2 Disinfecting 1. Follow the instructions on the disinfectant. 2. Disinfect the device at regular intervals with a commercially available disinfectant, e.g. with 70% isopropyl alcohol. Please take note of the following terms: • Prior to every measurement with direct skin contact: – Weighing platform – Head slide and measuring flap – Measuring flap extension (if applicable) • After every measurement with direct skin contact: – Weighing platform – Head slide and measuring flap – Measuring flap extension (if applicable) • If required: – Column and display of the scales – Telescopic elements of the length measuring rod – Heel stop (if applicable)
6.3 Sterilizing This device may not be sterilised.
16 •
English
7. FUNCTION CHECK Perform functional check prior to each operation. A complete functional check includes: • Visual inspection for mechanical damage • Visual and functional inspection of all movable paparts • Functional check of the control elements CAUTION! Risk of injury
If you notice any faults or deviations during the functional check, do not use the unit. Have repairs of the unit carried out by seca service or an authorized service partner.
8. TROUBLESHOOTING Problem
Cause/Solution
Measurement does not return plausible results.
Measurement result read off incorrectly - Use the correct edge for the measuring range currently in use (“Performing length measurement” from page 14) - Read off the decimal places for the current measuring range in the correct direction (“Performing length measurement” from page 14) Incorrect position/posture of patient - Correct the position/posture of the patient (“Performing length measurement” from page 14) Measuring rod mounted incorrectly on column scale - Mount the measuring rod correctly (“Before you really get started…” from page 8)
9. SERVICING The seca measuring rod does not require maintenance. Should any part of it fail to function, please contact an authorised service partner. You can find your local service partner at www.seca.com or just send an e-mail to Function Check • 17
service@seca.com.
10.TECHNICAL DATA 10.1 General technical data General technical data Dimensions • Depth • Width • Height (rectracted) • Height (extended)
17 mm 40 mm 980 mm 1480 mm
Weight
approx. 0.7 kg
Temperatur range • Operation • Storage • Transport
+10 °C to +40 °C (50 °F to 104 °F) -10 °C to +65 °C (14 °F to 149 °F) -10 °C to +65 °C (14 °F to 149 °F)
Air pressure • Operation • Storage • Transport
700 - 1060 hPa 700 - 1060 hPa 700 - 1060 hPa
Humidity • Operation • Storage • Transport
30% - 80% no condensation 0% - 95% no condensation 0% - 95% no condensation
Medical device in accordance with Directive 93/42/EEC
Class I with measuring function
10.2 Technical data, length measuring Technical data, length measuring Measuring range 1
60 cm - 130.5 cm
Measuring range 2
130.5 cm - 200 cm
Graduation Accuracy
18 •
1 mm better than ± 5 mm
English
11.WARRANTY We offer a two-year warranty from the date of delivery for defects attributable to faulty material or poor workmanship. This excludes all moveable parts such as (rechargeable) batteries, cables, power supply units, etc. Defects which are covered by the warranty shall be rectified free of charge for customers on production of the sales receipt. No further claims can be accepted. The costs of shipment in both directions shall be borne by the customer where the device is not located at the customer's premises. In the event of any damage during shipment warranty claims can only be asserted where the complete original packaging was used for shipment and the scales were secured inside in the same manner as in the original packaging. You should therefore keep all packaging. The warranty shall become null and void where the device is opened by persons not expressly authorised to do so by seca. We ask customers based abroad to contact their local sales agent directly in the case of warranty claims.
Warranty • 19
Das Längenmessgerät The equipment for measuring length Appareil de mesure à fonctionnement Strumento per la misura della statura Equipo para medición longitudinal Højdemåleren Längdmätinstrumentet Lengdemåleren Pituuden mittauslaite De lengtemeter O aparelho de medição da altura Η συσκευή μέτρησης Přístroj k měření délek Pikkusemõõteseade A távolságmérő Ilgio matuoklis Iekārta garuma mērīšanai Urządzenie do pomiaru długości Naprava za merjenje dolžine Dĺžkové meradlo Boy ölçüm cihazı
220
Chargennr.: siehe Rückseite Messstab Batch no.: see back of measuring rod N° de lot : voir au dos de la toise de mesure N° di lotto: v. parte posteriore dell’asta graduata N° de lote: véase lado trasero de la varilla Charge nr.: se på bagsiden af længdemåleapparat Partinr.: se mätstavens baksida Lot nr.: se bakside målestav. Valmistuserän numero: ks. mittasauvan takaosaa Chargenummer: zie achterzijde meetlat Nº de carga: vide retaguarda da vareta de medição Αρ. σειράς: Βλέπε στην πίσω πλευρά του μετρητή Číslo šarže: viz zadní strana - měřidlo Partiinumber: vt mõõtelati tagaküljelt Gyártási tételszám: lásd a távolságmérő hátlapját Partijos Nr.: žr. skirtuką galinėje pusėje Sērijas Nr.: skat. mērlatas aizmugurē Nr partii: patrz na odwrocie listwy pomiarowej Štev. šarže: oglejte si zadnjo stran merilne palice Číslo šarže: pozri zadnú stranu meracej laty Şarj numarası: ölçüm çubuğunun arka kısmına bakin