TABLE OF CONTENTS English 1. Device description... 4 1.1 Intended use... 1.2 Functional description... 1.3 User qualification... Assembly... Operation...
4 4 5 5 5
5. Performing length measurement... 18 5.1 Measuring patients smaller than 130.5 cm... 18 5.2 Measuring patients taller than 130.5 cm... 19
2. Safety information... 5
6. Hygiene Treatment... 20
2.1 Saftey information in this user manual... 5 2.2 Basic safety precautions... 5
6.1 Cleaning... 20 6.2 Disinfecting... 20 6.3 Sterilizing... 20
3. Overview... 7
7. Function Check... 21
3.1 Controls and mounting elements... 7 3.2 Marking on the device and on the rating plate... 8 3.3 Marking on the packaging... 9
8. Troubleshooting... 21
4. Before you really get started… . 10 4.1 Scope of delivery... 4.2 Mounting measuring rod... Mounting options... Preparation... Mounting measuring rod to the wall... 4.3 Verifying position of measuring rod... Checking position... Adjusting position... 4.4 Using the heel stop... Preparing the heel stop... Fitting the heel stop... 4.5 Fitting the measuring flap extension...
10 11 11 12
9. Servicing... 21 10. Technical Data... 22 10.1 General technical data... 22 10.2 Technical data, length measuring... 22 11. Warranty... 23
12 14 14 14 15 15 16 17
•3
1. DEVICE DESCRIPTION 1.1 Intended use The medical measuring rod seca 222 is used primarily in hospitals, general practice and nursing facilities on wards. The measuring rod is for conventional determination of height and for the early detection of growth disorders; it assists the physician supervising treatment in making a diagnosis or deciding on a course of treatment. To make an accurate diagnosis, however, other specific examinations have to be ordered by the physician and their results taken into account, in addition to determining height values.
1.2 Functional description The measuring rod is to stand on the floor and is fitted to a wall with two screws. It is possible to compensate for protruding walls or attachments (such as skirting boards or cable ducts) with a maximum of 5 spacer washers per screw. The measuring rod consists of two telescopic elementes: • Lower telescopic element: Measuring range 1: 6 cm -130.5 cm (3 inch - 51.4 inch) • Upper telescopic element: Measuring range 2: 130.5 cm - 230 cm (51.4 inch - 90 inch) The head slide of the measuring rod is locked at the top of the upper telescopic element to measure persons taller than 130.5 cm (51.4 inch). The head slide needs to be released to measure persons smaller than 130.5 cm (51.4 inch). The measuring flap can be folded down. The heel stop helps the patient to assume the correct postion for measurement. The heel stop ist attached to the measuring rod and supported against the wall.
4•
English
1.3 User qualification Assembly
Devices that are shipped partially assembled may only be mounted by sufficiently qualified persons such as specialist dealers, hospital technicians or seca service technicians.
Operation
The device may only be operated by healthcare professionals.
2. SAFETY INFORMATION 2.1 Saftey information in this user manual The signal words used in this user manual have the following meaning: CAUTION! Used to identify a hazardous situation. If you fail to take note of this information, minor to moderate injuries may result. ATTENTION! Used to identify possible incorrect usage of the device. If you fail to take note of this information, you may damage the device, or the measured results may be incorrect.
2.2 Basic safety precautions • Please observe the information in this user manual. • Keep this user manual in a safe place. This user manual is part of the product and must be readily available. CAUTION! Patient hazard, damage to device
• Use the supplied assembly material only. • Fold down the measuring slide after each measurement. • Have the device serviced regularly as described in the respective section of this document.
Safety information • 5
• Technical changes to the device are not permitted. The unit contains no user-serviceable parts. Make sure that maintenance and repair are only carried out by an authorised seca service partner. You can find your local service partner at www.seca.com or send an email to service@seca.com. • Make sure you only use genuine seca accessories and spare parts. Otherwise the warranty provided by seca will become null and void.
6•
English
3. OVERVIEW 3.1 Controls and mounting elements
3
5
6
4
2
7
1 155.5 cm
8
10
9
11 12
No. 1 2 3 4
Element Lower telescopic element Upper telescopic element Measuring flap extension Measuring flap
Function Measuring range 1:
6 cm - 130.5 cm 3 inch - 51.4 inch Measuring range 2: 130.5 cm - 230 cm 51.4 inch - 90 inch Helps the patient to assume the correct position for measurement • To be positionned on top of the patient‘s head • Can be folded down
Overview • 7
No.
Element
5
Lock
6
Head slide
7
Read-off line 1 (red)
8
Read-off line 2 (red)
9
Spacer washer
10
Fixing hole
11
Spacer washer
12
Heel stop
Function • Locks head slide at top of upper telescopic element for measurement in measuring range 2 • Releases head slide for measurement on lower telescopic element (measuring range 1) • Locked at top of upper telescopic element • To be released for measurement on lower telescopic element (measuring range 1) Reference when head slide is positioned on lower telescopic element to measure persons smaller than 130.5 cm (51.4 inch) • Reference when measuring persons taller than 130.5 cm (51.4 inch) • Results to be read off at upper telescopic element (here: 155.5 cm) • A maximum of 5 per screw to allow the measuring rod to be attached to a wall precisely vertically • Protruding walls or attachments (e.g. skirting boards or cable ducts) can be compensated The fixing screw of the measuring rod is inserted in this hole A maximum of 5 per screw to support the heel stop against the wall Helps the patient to assume the correct postion for measurement
3.2 Marking on the device and on the rating plate Text/Symbol Model xx/xx
Meaning Model number Batch number, engraved on back of device Observe user manual
0123
8•
Device complies with EC standards and directives • 0123: Notified body medical products: TÜV SÜD Product Service
English
3.3 Marking on the packaging Protect from moisture Arrows indicate top of product. Transport and store in an upright position. Fragile Do not throw or drop. Permitted min. and max. temperature for transport and storage Permitted min. and max. moisture for transport and storage Packaging material can be disposed of through recycling programs
Overview • 9
4. BEFORE YOU REALLY GET STARTED… 4.1 Scope of delivery
b
a
c d e
f g h i No.
a b c d e f g h i j k
10 •
Component Measuring rod seca 222 Measuring flap extension Screw M4 x 60 mm Spacer washer Wall plug ø 5 mm Heel stop Spacer washer Plastic nut M5 Plastic screw M5x40 Screw M4 x 60 mm Wall plug ø 5 mm User manual, not shown
j k Pcs. 1 1 2 10 2 1 10 2 2 2 2 1
English
4.2 Mounting measuring rod Mounting options
The measuring rod must be standing on the floor and attached precisely vertically to a wall for meaningful measuring results to be obtained. The spacer washers supplied allow you to compensate for protruding walls and attachments (e.g. skirting boards or cable ducts). The following illustration shows a variety of options for attaching the rod with the number of spacer washers per screw recommended in each case.
Option I
Option II
m min ax = 5 =1
minmax = =1 5 min
=1
3.3
x=
ma
Option IV
Option III
min
m min ax = 5 =1
=1 max
min
cm
=1
max 3.3
x=
ma
=3.3
cm
=5
cm
Before you really get started… • 11
Preparation
Choose an adequately strong, load-bearing wall to mount the device. Standard wall plugs are supplied for installation on solid brick walls. We recommend use of appropriate special wall plugs for other types of walls. To achieve exact measurement results, the floor where the device is installed must be flat and stable. Soft floorings, e. g. wooden planks, give under the weight of the patient and distort the measured result. To mount the measuring rod to a wall, proceed as follows: ø 5mm
1. Mark the drill holes perpendicular to each other exactly at the positions indicated in the illustration above. 2. Drill the hole using a drill bit ø 5 mm. 3. Insert the wall plugs in the drill holes.
Mounting measuring rod to the wall
You need the following fastenings:
2 x Screw M4 x 60
max 10 x Spacer washer
1. If there are protrusions or attachments (e.g. skirting boards or cable ducts) where you are attaching the rod, measure the depth of these. 2. Select the option to suit the attachment location (see “Mounting options” on page 11).
12 •
English ATTENTION! Malfunction due to installation errors
If no spacer washers are used, the head slide will be trapped against the wall. This will make it impossible to measure height. Ensure that at least 1 spacer washer per screw is used. 3. Put the appropriate number of spacer washers ready: – 1 spacer washer = 1.2 cm – Every subsequent spacer washer: + 0.7 cm – 5 spacer washers = 4.0 cm 4. Pull the upper telescopic element out to the 173 cm mark. The holes in the lower telescopic element are exposed.
173 cm
5. Put a screw in the top bore of the measuring rod. 6. Put the appropriate number of spacer washers on the screw (the illustration shows attachment option II). 7. Screw measuring rod to wall hand-tight. 8. Repeat steps 5. to 7. for the bottom bore. 9. Tighten up both screws fully.
minmax = =1 5
Before you really get started… • 13
4.3 Verifying position of measuring rod Checking position
After mounting the measuring rod, you must check its position and correct if necessary. To do so, proceed as follows: 1. Use a spirit level to ensure that the measuring rod is vertical. Requirement: The measuring rod is inclined neither towards nor away from the wall. 2. If the measuring rod is not vertical, correct its position as described in the section below.
Adjusting position
To correct the position of the measuring rod proceed as follows: 1. Loosen the screws of the measuring rod. 2. Correct the position of the measuring rod by adding or removing spacer washers.
max = min =1 5
3. Tighten the screws on the measuring rod. 4. Check the position of the measuring rod again. 5. If necessary, re-adjust the position of the measuring rod.
max =5 min =1
14 •
NOTE: Your seca service partner will also be happy to help. You can find your local seca service partner at www.seca.com or just send an e-mail to service@seca.com.
English
4.4 Using the heel stop The heel stop supplied helps the patient to assume the correct position for measurement. The heel stop is attached to the measuring rod and supported against the wall. Proceed as outlined below to fit the heel stop.
Preparing the heel stop
You need the following mounting elements:
2 x Plastic screw M5 x 40
2 x Plastic nut max 10 x M5 Spacer washer
1. At the foot of the measuring rod, measure the distance from the wall, the skirting board, the cable duct etc.
= ?c
m 2. Put the appropriate number of spacer washers ready (max 5 per screw): – 1 spacer washer = 1.2 cm – Every subsequent spacer washer: + 0.7 cm – 5 spacer washers = 4.0 cm 3. Align the heel stop as shown in the adjacent illustration.
4. Put one plastic screw M5 x 40 in each of the bores in the heel stop. 5. Put the appropriate number of spacer washers (in this case: 2) on the screws. 6. Screw the spacer washers tightly to the heel stop using a plastic nut.
Before you really get started… • 15
7. Shorten the plastic screws so that they finish flush with the spacer washer.
Fitting the heel stop 1. Undo the screws of the measuring rod.
2. Put the measuring rod in the grooves of the heel stop.
3. Screw the measuring rod firmly to the wall again. 4. Check the position of the measuring rod as described in Section “Verifying position of measuring rod” from page 14. NOTE You can use the metal screws and standard wall plugs supplied as an alternative mounting option for the heel stop. This allows you to screw the heel stop to the wall in the same way as the measuring rod itself.
16 •
English
4.5 Fitting the measuring flap extension You have the option of increasing the contact surface of the measuring flap by using the measuring flap extension. This enables you to position the patient even better and achieve a more accurate measuring result. Push the measuring flap extension onto the measuring flap until you hear it engage. Click!
If required, move the measuring flap extension along the measuring flap in a longitudinal direction.
Before you really get started… • 17
5. PERFORMING LENGTH MEASUREMENT CAUTION! Patient hazard
Allways perform a function check as described in section “Function Check” on page 21 before using the device.
5.1 Measuring patients smaller than 130.5 cm 1. Unlock the head slide by pressing on the lock. 2. Unfold the measuring flap so it engages in a horizontal position. 3. Ask the patient to step under the measuring flap: – Back to the measuring rod – Body kept straight – Heels against the heel stop – Head straight: So-called "Frankfurt plane" parallel to the measuring flap 4. Move the head slide on the lower telescopic element until the measuring flap rests on the patient's head. 5. Read off the measured result at the read-off line (read-off line 1) of the headslide (here: 113.5 cm). 6. Ask the patient to step away from the measuring rod. 7. Fold the measuring flap down. 8. Move up the head slide until it engages in the lock.
113.5 cm
18 •
English
5.2 Measuring patients taller than 130.5 cm 1. Unfold the measuring flap so it engages in a horizontal position. 2. Pull the upper telescopic element out until the patient can comfortably stand under the measuring flap. 3. Ask the patient to step under the measuring flap: – Back to the measuring rod – Body kept straight – Heels against the heel stop – Head straight: So-called "Frankfurt plane" parallel to the measuring flap 4. Slide down the upper telescopic element until the measuring flap rests on the patient's head. 5. Read off the measured result at the read-off line (read-off line 2) of the lower telescopic element (here: 141.5 cm). 6. Ask the patient to step away from the measuring rod. 7. Slide the upper telescopic element to the lowest position. 8. Fold the measuring flap down.
141.5 cm
Performing length measurement • 19
6. HYGIENE TREATMENT ATTENTION! Damage to device
Inappropriate detergents may damage the sensitive surfaces of the devices. Do not use caustic or abrasive detergents. Do not use spirits or benzine.
6.1 Cleaning Use a soft cloth dampened with mild soapsuds to clean the surfaces of the device.
6.2 Disinfecting 1. Follow the instructions on the disinfectant. 2. Disinfect the device at regular intervals with a commercially available disinfectant, e.g. with 70% isopropyl alcohol. Please take note of the following terms: • Prior to every measurement with direct skin contact: – Weighing platform – Head slide and measuring flap – Measuring flap extension (if applicable) • After every measurement with direct skin contact: – Weighing platform – Head slide and measuring flap – Measuring flap extension (if applicable) • If required: – Column and display of the scales – Telescopic elements of the length measuring rod – Heel stop (if applicable)
6.3 Sterilizing This device may not be sterilised.
20 •
English
7. FUNCTION CHECK Perform functional check prior to each operation. A complete functional check includes: • Visual inspection for mechanical damage • Visual and functional inspection of all movable paparts • Functional check of the control elements CAUTION! Risk of injury
If you notice any faults or deviations during the functional check, do not use the unit. Have repairs of the unit carried out by seca service or an authorized service partner.
8. TROUBLESHOOTING Problem
Cause/Solution
Measurement does not return plausible results.
Measurement result read off incorrectly - Use the correct edge for the measuring range currently in use (“Performing length measurement” from page 18) - Read off the decimal places for the current measuring range in the correct direction (“Performing length measurement” from page 18) Incorrect position/posture of patient - Correct the position/posture of the patient (“Performing length measurement” from page 18) Measuring rod mounted incorrectly - Mount the measuring rod correctly (“Before you really get started…” from page 10)
9. SERVICING The seca measuring rod does not require maintenance. Should any part of it fail to function, please contact an authorised service partner. You can find your local service partner at www.seca.com or just send an e-mail to service@seca.com. Function Check • 21