ACUSON Redwood Supplement to the User and Reference Manuals Sw Ver VA10 , VA20
Supplement
344 Pages
Preview
Page 1
ACUSON Redwood, Product Version 1.0, 2.0 Software Version VA10, VA20 ©2021-2022 Siemens Medical Solutions USA, Inc. All Rights Reserved. Date of first issue: 2021-08 Date of revision: 2022-08 The following trademarks are owned by Siemens Medical Solutions USA, Inc. (hereinafter "Siemens"): ACUSON, ACUSON Redwood All other product names are references to third-party products and are trademarks of their respective companies. Siemens includes references to third-party products in the user documentation for informational purposes only. Siemens does not endorse third-party products referenced in the user documentation. Siemens does not assume responsibility for the performance of third-party products. Siemens reserves the right to change its products and services at any time. In addition, this publication is subject to change without notice. Manufacturer Siemens Medical Solutions USA, Inc. Ultrasound 22010 S.E. 51st Street Issaquah, WA 98029 U.S.A. Phone: +1-888-826-9702 siemens-healthineers.com
2 / 344
Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen Germany Phone: +49 9131 84-0 siemens-healthineers.com
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
Supplement to the User and Reference Manuals (English) The supplement provides updates to your existing user and reference manuals. Refer to the errata in this document and the existing user and reference manuals for a complete set of operating instructions for your ultrasound system.
Retractable Keyboard Option (New content: Page 1-22, Page A-21, Instructions for Use) The retractable keyboard is an ultrasound system option, not a standard feature.
Language-specific alphanumeric keyboard: English, German, French, Spanish, Italian
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
3 / 344
System Overview (Updated content: Page 1-5, Instructions for Use)
Example of the ultrasound system, right front view. 1 Backlit control panel 2 Front handle 3 Keyboard 4 Front basket 5 Storage bin for transducer connectors 6 Lock and swivel lock for the rear wheel 7 Air filter 8 Speaker 9 Rear basket 10 Touch screen with USB ports 11 Rear handle for repositioning the ultrasound system
4 / 344
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
Accessing the Retractable Keyboard (New content: Page 4-4, Instructions for Use) WARNING: Carefully extend or retract the keyboard. A potential pinch point between the control panel and keyboard can create a risk for injury to the user.
Example of the retractable keyboard on the ultrasound system.
To access the keyboard:
To extend the keyboard, gently pull the keyboard from the ultrasound system.
To retract the keyboard, gently push the keyboard toward the ultrasound system.
Special Keys on the Retractable Keyboard Key Name Blank
Description Accesses special characters located on the alphanumeric keyboard. This key displays the keyboard language and is blank on English keyboards.
Text Start
Activates the text cursor on the image screen.
(Space Bar)
Note: Pressing a key on the keyboard also activates the text cursor and inserts the alphanumeric character on the image screen corresponding to the key.
Alt Gr
(Available only on non-English keyboards) Accesses special characters located on the alphanumeric keyboard.
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
5 / 344
Cleaning and Disinfecting Transducers List of Approved Cleaners
Clinell Universal Wipes & Spray
ENZOL
Prolystica 2X Concentrate Enzymatic Presoak and Cleaner
Protex ULTRA Disinfectant Wipes
Transeptic
(Updated content: Page 3-12, Instructions for Use)
18H6
9EC4
9VE4
CW2
CW5
= approved NA = not approved
List of Approved Disinfectants
CIDEX OPA
gigasept FF (new)
Revital Ox RESERT HLD
Sani-Cloth AF3 Germicidal Disposable Wipes
Sani-Cloth HB Germicidal Disposable Wipe
Super Sani-Cloth Germicidal Disposable Wipe
trophon EPR with NanoNebulant1
trophon EPR with Sonex-HL1
trophon2 with NanoNebulant1
trophon2 with Sonex-HL1
18H6
CIDEX
(Updated content: Page 3-13, Instructions for Use)
5C1
9EC4
NA
2
2
2
2
CW2
NA
NA
CW5
NA
NA
= approved NA = not approved 1 Indicates a reprocessing device with the approved disinfectant for use. For information about the reprocessing device, refer to the device manufacturer's instructions for use.
For use only with the following serial numbers of 5C1 transducers: ≥2019XXXX.
2
6 / 344
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
Updates to Ophthalmic Use (Updated content: Page 1-6, Instructions for Use) WARNING: The ultrasound system is not intended for ophthalmic use.
Updates to Workflow Protocols Editing Protocol Views (Updated content: Page 6-5, Instructions for Use) To
Do This
Delete the current view during protocol
Note: You can delete the view without any acquired images. 1. Press Delete View. The ultrasound system displays a message to confirm the deletion. 2. Tap OK. The ultrasound system deletes the current view. The measurement results are not deleted from the report.
Delete a completed view
Note: You can delete the view without any acquired images. 1. Tap the name of the completed view. 2. Press Delete View. The ultrasound system displays a message to confirm the deletion. 3. Tap OK. The ultrasound system removes the selected view. The measurements are not deleted from the report.
Speed of Sound Optimizing 2D Images (Updated content: Page 8-7, Instructions for Use) To
Do This
Adjust the speed of sound
1. Tap Speed of Sound. 2. Select a setting. A higher setting is for fatty tissue and a lower setting is for less fatty or soft tissue.
Speed of Sound (Updated content: Page A-10, Instructions for Use)
Optimizes the 2D image by adjusting the speed of sound
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
7 / 344
Clinical Measurement Range and Accuracy (Updated content: Page A-24, Instructions for Use) (Applicable for the ACUSON Redwood 1.0 ultrasound system) Note: The range of accuracy for the heart rate detected by the physio module is 30 bpm to 300 bpm.
Clinical Measurements: Range and Accuracy (New content: Page A-27, Instructions for Use) (Applicable for the ACUSON Redwood 2.0 ultrasound system) Direct Measurement
Range
Accuracy
30 bpm - 300 bpm
Less than ±2 beats per minute or ±1%, whichever is greater.
Physio Heart Rate
8 / 344
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
Updates to Configuration Settings General (New content: Pages 1-7 to 1-8, System Reference) Selection
Description
Keyboard (Restart Required)
Provides language characters for the touch screen keyboard English German French Italian Spanish
S-Video S-Video Format
--Selects the signal standard format for separate video connection, for example, Television Standards Committee (NTSC) or Phase Alternating Line (PAL).
Protocol Selections (New content: Page 1-23, System Reference) (Available only for the ACUSON Redwood 2.0 ultrasound system) Selection
Description
Append Paused Images
Designates the location in the list of protocol views for inserting an image acquired and saved when the protocol is paused. Within the Protocol inserts the image at the current location of the paused protocol. At the End of the Protocol inserts the image at the end of the list of protocol views.
Configuring a Local Application Entity (Updated content: Page 6-4, System Reference) The following table replaces the selections in step 3 of the existing procedure on page 6-4 in your manual. Selection
Required Setting
Alias
System-defined "computer name" or user-defined name for the export host
Secure Study Transfer
When enabled, encrypts patient data for a secure DICOM transfer.
AE Title (Secure [SCP + SCU])
AE Title used by DICOM Storage Class Providers (SCPs) to identify the ultrasound system for encrypted DICOM communication
AE Title (Unsecure [SCP + SCU])
AE Title used by DICOM Storage Class Providers (SCPs) to identify the ultrasound system for unencrypted DICOM communication
Client Authentication
When enabled, authenticates the client's certificate for encrypted DICOM communication.
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
9 / 344
Errata to the User and Reference Manuals (English) The errata provides corrected information for the user and reference manuals. Refer to the supplement in this document and the existing user and reference manuals for a complete set of operating instructions for your ultrasound system.
Indications for Use Statement (Updated content: Page 1-7, Instructions for Use) (Available only for the ACUSON Redwood 2.0 ultrasound system) Product
Indications for Use Statement
ACUSON Redwood Ultrasound System
The ACUSON Redwood ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Fetal, Abdominal, Pediatric, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus, and other pelvic structures), Adult, Pediatric and Neonatal Cardiac, Pelvic, Neonatal Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides the ability to measure anatomical structures: fetal, abdominal, pediatric, small organ, neonatal cephalic, cardiac (adult, pediatric and neonatal), transesophageal, transrectal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculoskeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
10 / 344
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
Intended Users (Updated content: Page 1-9, 1-10, Instructions for Use) (Available only for the ACUSON Redwood 2.0 ultrasound system) User
Interaction with Ultrasound Equipment
Sonographer
Acquires diagnostic views of anatomy, blood flow, and related pathology Performs measurements and analysis of the acquired images Prepares exam data for review and interpretation by a qualified physician
Physician
Radiologist
Ranges from novices (for example, students) to advanced practitioners with certification in multiple subspecialties Educated in anatomy, physiology, patient care, and identification of pathology in ultrasound images Many sonographers have a Bachelor's degree; some have advanced degrees in related health care subjects
Performs ultrasound exams
Medical doctor
Interprets exam data
Trained in ultrasound imaging techniques
Writes and assembles exam findings in a report
Skilled in interpreting ultrasound exam data
Performs ultrasound exams
Medical doctor
Performs image-guided access and biopsy procedures
Expert in diagnostic imaging, including computer tomography, magnetic resonance imaging, X-ray, ultrasound, and nuclear medicine
Interprets exam data Writes and assembles exam findings in a report
System Administrator and Customer Service Engineer
Expected Experience and Other Characteristics
Configures the ultrasound system for use in a networked environment
Advanced training in imaging physics with typically two to six years of post-doctoral training in the field of radiology; some specialize in diagnostic ultrasound A system administrator is an individual within your organization who is designated to set up system parameters to connect the ultrasound system or workstation to a picture archiving and communication system (PACS). Customer service engineers are Siemens Healthineers representatives who configure the ultrasound system during initial installation, support troubleshooting activities, and repair the system.
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
11 / 344
Supplying Power to the Ultrasound System (Updated content: Page 4-13, Instructions for Use) To power off the ultrasound system: 1. 2. 3. 4. 5.
Press the partial power on/off control. To restart the system, click Restart. To power off the system, click Shutdown. To cancel the shutdown, click Cancel. To completely disconnect the ultrasound system from the power supply (supply mains), shut down and then unplug the ultrasound system from the power supply.
Activating Hibernate (Updated content: Page 4-14, Instructions for Use) Hibernate decreases the time to power on and off the ultrasound system. When hibernate is activated, the ultrasound system pauses the data transfer of images in progress. A delayed transfer resumes when the ultrasound system is powered on and the network connection is reestablished. To activate hibernate: Note: The ultrasound system should be shut down or restarted after using the hibernation function for 20 times, or after being in hibernation for 24 hours or longer.
1. 2. 3. 4.
Briefly press the partial power on/off control. Click Hibernate. Transport the ultrasound system to another location, if necessary. To exit hibernate and power on the ultrasound system, press the partial power on/off control.
12 / 344
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
Summary Table for Acoustic Output (Updated content only for the summary table: Page D-4, Instructions for Use. Appendix D provides the required acoustic output data for all transducers.)
2D with Color and PW Doppler
2D with M-mode
X
X
X
X
Contrast
2D with PW Doppler
X
Elasticity
PW Doppler
X
CW Doppler
2D with Color
9VE4
M-mode with Color
Transducer Model
B-mode (2D)
Operating Mode
Exam (Updated content: Page 1-17, System Reference) Selection
Required Setting
(parameters)
Optimizes the selected parameters when using TEQ technology for Doppler. Gain optimizes TEQ gain offset for the current imaging mode. Dynamic Range adjusts the dynamic range. Scale/Baseline adjusts the scale factor of the Pulsed Repetition Frequency (PRF) and the position of the spectral or color baseline. Auto Invert reverses the vertical orientation of the Doppler spectrum on the baseline, if required.
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
13 / 344
Актуализирана информация за настройките за конфигуриране General (Ново съдържание: Страници от 1-7 до 1-8, System Reference) Selection
Description
Keyboard (Restart Required)
Provides language characters for the touch screen keyboard English German French Italian Spanish
S-Video S-Video Format
--Selects the signal standard format for separate video connection, for example, Television Standards Committee (NTSC) or Phase Alternating Line (PAL).
Protocol Selections (Ново съдържание: стр. 1-23, System Reference) (Предлага се само за ултразвуковата система ACUSON Redwood 2.0) Selection
Description
Append Paused Images
Designates the location in the list of protocol views for inserting an image acquired and saved when the protocol is paused. Within the Protocol inserts the image at the current location of the paused protocol. At the End of the Protocol inserts the image at the end of the list of protocol views.
Configuring a Local Application Entity (Актуализирано съдържание: cтр. 6-4, System Reference) Следващата таблица заменя изборите в стъпка 3 на съществуващата процедура, описани на стр. 6-4 във вашето ръководство. Selection
Required Setting
Alias
System-defined "computer name" or user-defined name for the export host
Secure Study Transfer
When enabled, encrypts patient data for a secure DICOM transfer.
AE Title (Secure [SCP + SCU])
AE Title used by DICOM Storage Class Providers (SCPs) to identify the ultrasound system for encrypted DICOM communication
AE Title (Unsecure [SCP + SCU])
AE Title used by DICOM Storage Class Providers (SCPs) to identify the ultrasound system for unencrypted DICOM communication
Client Authentication
64 / 344
When enabled, authenticates the client's certificate for encrypted DICOM communication.
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
Обобщена таблица за изходната акустична мощност (Актуализирано съдържание само за обобщената таблица: cтр. D-4, Инструкции за употреба. Приложение D предоставя данни за необходимата изходна акустична мощност за всички трансдюсери.)
Цветен 2D и PW Doppler
2D с М-режим
X
X
X
X
Контраст
2D с PW Doppler
X
Еластичност
PW Doppler
X
CW Doppler
2D режим, цветен
9VE4
M-режим, цветен
Модел на трансюдера
B-режим (2D)
Работен режим
Exam (Актуализирано съдържание: cтр. 1-17, System Reference) Selection
Required Setting
(parameters)
Optimizes the selected parameters when using TEQ technology for Doppler. Gain optimizes TEQ gain offset for the current imaging mode. Dynamic Range adjusts the dynamic range. Scale/Baseline adjusts the scale factor of the Pulsed Repetition Frequency (PRF) and the position of the spectral or color baseline. Auto Invert reverses the vertical orientation of the Doppler spectrum on the baseline, if required.
68 / 344
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
Ažuriranja konfiguracijskih postavki General (Novi sadržaj: stranice od 1-7 do 1-8, System Reference) Selection
Description
Keyboard (Restart Required)
Provides language characters for the touch screen keyboard English German French Italian Spanish
S-Video S-Video Format
--Selects the signal standard format for separate video connection, for example, Television Standards Committee (NTSC) or Phase Alternating Line (PAL).
Protocol Selections (Novi sadržaj: stranica 1-23, System Reference) (Dostupno samo za ultrazvučni sustav ACUSON Redwood 2.0) Selection
Description
Append Paused Images
Designates the location in the list of protocol views for inserting an image acquired and saved when the protocol is paused. Within the Protocol inserts the image at the current location of the paused protocol. At the End of the Protocol inserts the image at the end of the list of protocol views.
Configuring a Local Application Entity (Ažurirani sadržaj: stranica 6-4, System Reference) Odabiri u 3. koraku postojećeg postupka na stranici 6-4 u priručniku zamijenjeni su sljedećom tablicom. Selection
Required Setting
Alias
System-defined "computer name" or user-defined name for the export host
Secure Study Transfer
When enabled, encrypts patient data for a secure DICOM transfer.
AE Title (Secure [SCP + SCU])
AE Title used by DICOM Storage Class Providers (SCPs) to identify the ultrasound system for encrypted DICOM communication
AE Title (Unsecure [SCP + SCU])
AE Title used by DICOM Storage Class Providers (SCPs) to identify the ultrasound system for unencrypted DICOM communication
Client Authentication
When enabled, authenticates the client's certificate for encrypted DICOM communication.
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
75 / 344
Sažeti tablični prikaz akustičnih izlaza (Ažurirani sadržaj samo za sažeti tablični prikaz: stranica D-4, Upute za uporabu. U Dodatku D navedeni su potrebni podaci o akustičnim izlazima za sve pretvornike.)
2D s bojom i Dopplerom pulsirajućeg vala
2D s M načinom rada
X
X
X
X
Kontrast
2D s Dopplerom pulsirajućeg vala
X
Elastičnost
Doppler pulsirajućeg vala
X
Doppler kontinuiranog vala
2D s bojom
9VE4
M način rada s bojom
Model pretvornika
B način rada (2D)
Način rada
Exam (Ažurirani sadržaj: stranica 1-17, System Reference) Selection
Required Setting
(parameters)
Optimizes the selected parameters when using TEQ technology for Doppler. Gain optimizes TEQ gain offset for the current imaging mode. Dynamic Range adjusts the dynamic range. Scale/Baseline adjusts the scale factor of the Pulsed Repetition Frequency (PRF) and the position of the spectral or color baseline. Auto Invert reverses the vertical orientation of the Doppler spectrum on the baseline, if required.
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
79 / 344
Aktualizace týkající se nastavení konfigurace General (Nový obsah: strany 1-7 až 1-8, System Reference) Selection
Description
Keyboard (Restart Required)
Provides language characters for the touch screen keyboard English German French Italian Spanish
S-Video S-Video Format
--Selects the signal standard format for separate video connection, for example, Television Standards Committee (NTSC) or Phase Alternating Line (PAL).
Protocol Selections (Nový obsah: strana 1-23, System Reference) (K dispozici pouze pro ultrazvukový systém ACUSON Redwood 2.0) Selection
Description
Append Paused Images
Designates the location in the list of protocol views for inserting an image acquired and saved when the protocol is paused. Within the Protocol inserts the image at the current location of the paused protocol. At the End of the Protocol inserts the image at the end of the list of protocol views.
Configuring a Local Application Entity (Aktualizovaný obsah: strana 6-4, System Reference) Následující tabulka nahrazuje výběry v kroku 3 stávajícího postupu na straně 6-4 příručky. Selection
Required Setting
Alias
System-defined "computer name" or user-defined name for the export host
Secure Study Transfer
When enabled, encrypts patient data for a secure DICOM transfer.
AE Title (Secure [SCP + SCU])
AE Title used by DICOM Storage Class Providers (SCPs) to identify the ultrasound system for encrypted DICOM communication
AE Title (Unsecure [SCP + SCU])
AE Title used by DICOM Storage Class Providers (SCPs) to identify the ultrasound system for unencrypted DICOM communication
Client Authentication
86 / 344
When enabled, authenticates the client's certificate for encrypted DICOM communication.
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
Opdateringer til konfigurationsindstillinger General (Nyt indhold: Side 1-7 til 1-8, System Reference) Selection
Description
Keyboard (Restart Required)
Provides language characters for the touch screen keyboard English German French Italian Spanish
S-Video S-Video Format
--Selects the signal standard format for separate video connection, for example, Television Standards Committee (NTSC) or Phase Alternating Line (PAL).
Protocol Selections (Nyt indhold: Side 1-23, System Reference) (Kun tilgængelig for ultralydssystemet ACUSON Redwood 2.0) Selection
Description
Append Paused Images
Designates the location in the list of protocol views for inserting an image acquired and saved when the protocol is paused. Within the Protocol inserts the image at the current location of the paused protocol. At the End of the Protocol inserts the image at the end of the list of protocol views.
Configuring a Local Application Entity (Opdateret indhold: Side 6-4, System Reference) Tabellen nedenfor erstatter valgene i trin 3 i den nuværende procedure på side 6-4 i vejledningen. Selection
Required Setting
Alias
System-defined "computer name" or user-defined name for the export host
Secure Study Transfer
When enabled, encrypts patient data for a secure DICOM transfer.
AE Title (Secure [SCP + SCU])
AE Title used by DICOM Storage Class Providers (SCPs) to identify the ultrasound system for encrypted DICOM communication
AE Title (Unsecure [SCP + SCU])
AE Title used by DICOM Storage Class Providers (SCPs) to identify the ultrasound system for unencrypted DICOM communication
Client Authentication
When enabled, authenticates the client's certificate for encrypted DICOM communication.
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
97 / 344
Oversigtstabel for akustisk output (Opdateret indhold, der kun gælder for oversigtstabellen: Side D-4, Brugsanvisning. Appendiks D viser data for akustisk output for alle transducere). 2D med farve og PW Doppler
2D med M-modus
X
X
X
X
Kontrast
2D med PW Doppler
X
Elasticitet
PW Doppler
X
CW Doppler
2D med farve
9VE4
M-modus med farve
Transducermodel
B-modus (2D)
Betjeningsmodus
Exam (Opdateret indhold: Side 1-17, System Reference) Selection
Required Setting
(parameters)
Optimizes the selected parameters when using TEQ technology for Doppler. Gain optimizes TEQ gain offset for the current imaging mode. Dynamic Range adjusts the dynamic range. Scale/Baseline adjusts the scale factor of the Pulsed Repetition Frequency (PRF) and the position of the spectral or color baseline. Auto Invert reverses the vertical orientation of the Doppler spectrum on the baseline, if required.
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals
101 / 344
Updates voor configuratie-instellingen General (Nieuwe inhoud: pagina 1-7 tot 1-8, System Reference) Selection
Description
Keyboard (Restart Required)
Provides language characters for the touch screen keyboard English German French Italian Spanish
S-Video S-Video Format
--Selects the signal standard format for separate video connection, for example, Television Standards Committee (NTSC) or Phase Alternating Line (PAL).
Protocol Selections (Nieuwe inhoud: pagina 1-23, System Reference) (Alleen beschikbaar voor het ACUSON Redwood 2.0 ultrasone systeem) Selection
Description
Append Paused Images
Designates the location in the list of protocol views for inserting an image acquired and saved when the protocol is paused. Within the Protocol inserts the image at the current location of the paused protocol. At the End of the Protocol inserts the image at the end of the list of protocol views.
Configuring a Local Application Entity (Bijgewerkte inhoud: pagina 6-4, System Reference) De volgende tabel vervangt de selecties in stap 3 van de huidige procedure op pagina 6-4 in uw handleiding. Selection
Required Setting
Alias
System-defined "computer name" or user-defined name for the export host
Secure Study Transfer
When enabled, encrypts patient data for a secure DICOM transfer.
AE Title (Secure [SCP + SCU])
AE Title used by DICOM Storage Class Providers (SCPs) to identify the ultrasound system for encrypted DICOM communication
AE Title (Unsecure [SCP + SCU])
AE Title used by DICOM Storage Class Providers (SCPs) to identify the ultrasound system for unencrypted DICOM communication
Client Authentication
108 / 344
When enabled, authenticates the client's certificate for encrypted DICOM communication.
Supplement to the User and Reference Manuals Errata to the User and Reference Manuals