SIGMA
Sigma Spectrum Infusion Pumps
Sigma Spectrum 35700BAX & 35700ABB Operators Manual Rev H Nov 2012
Operators Manual
131 Pages
Preview
Page 1
spectrum
Operator’s Manual 35700BAX & 35700ABB Generation 2 Operating System Pump Operating Software v6.05 For Use with MDL Editor v6.2.4
41018-6.05/6.2.4 Rev H November 29, 2012
SIGMA 711 Park Avenue Medina, New York 14103 v 800 356 3454 f 585 798 3909 www.sigmapumps.com
Manual 41018- 6.05/6.2.4 Revision H
II
Pump Operating Software v6.05 For Use With MDL Editor v6.2.4
©2012 Copyright SIGMA 711 Park Avenue Medina, New York 14103 v 800 356 3454 f 585 798 3909 www.sigmapumps.com
III
Manual 41018-6.05/6.2.4 Revision H
TABLE OF CONTENTS Introduction and Safety... 1 Intended Device Use...1 Related Documents...1 Regulatory Information...1 Contacting SIGMA Technical Support...1 Conventions...2 Summary of Warnings and Cautions...2
System Components... 15 Spectrum Pump Illustrations...16 Hardware Labeling...17
Battery Compatibility... 18 Setting Up the Pump... 19 Unpack the Pump...19 AC Power Adaptor...20 Cleaning the Power Adaptor... 20 Connecting the Power Adaptor... 20 Removing the Power Adaptor... 21
Charging the Battery...21 Configuring User Options...22 User Options... 22
Operational Overview... 25 Starting a New Infusion Using the Dose Error Reduction System (DERS)...29 Starting a New Infusion using the BASIC Mode...30 Secondary Infusions...31
Preparing the Pump and IV Sets... 32 Loading an IV Set...33 Unloading an IV Set...36 Preparing the Pump for a Secondary Infusion...36
Programming the Pump... 37 Infusion Programming Modes...38 Keys Used to Program and Operate the Pump...39 Activating a Drug Library on a Pump with a Wireless Module...41 Programming the Pump Using the Dose Error Reduction System...42
IV
Pump Operating Software v6.05 For Use With MDL Editor v6.2.4
Programming the Pump Using the BASIC Mode... 46 Setting up a Secondary Infusion... 50 Clearing the Secondary Mode While Returning to the Primary Infusion... 55 Secondary Callback – Assigned to a Secondary Drug in the Drug Library... 55 Programming a Loading Dose... 58 Programming a Bolus... 59 Multi-Step Programming... 59 Cyclic TPN Mode... 61 Titrating... 62 Delayed Run... 65 Weight Confirmation... 65
Pump Operation... 66 Operating Tips... 66 Placing the Pump in STANDBY (Hold)... 67 Operating the Keypad Lock... 68
Alarms... 69 Silencing an Alarm... 70 Clearing an Alarm... 70 Alarm Messages... 70 Preventing Nuisance Alarms... 75 Managing Bolus before Occlusion (Downstream) Release... 75 Battery Warning Levels... 75
Cleaning and Storage... 78 Compatible Cleaners... 78 Cleaning the Pump... 79 Pump Handling, Transport and Storage... 80
Compatible IV Sets... 81 Compatible Baxter IV Sets... 81 Compatible Baxter IV Sets – WARNINGS... 85 Compatible Hospira IV Sets... 87 Compatible Hospira IV Sets – WARNINGS... 89
Accessories... 90 Tandem Carrier... 90
V
Manual 41018-6.05/6.2.4 Revision H
Multi-Pole...91 AC Power Adaptor Protector... 91 Double Rotating Pole Clamp Assembly... 92
APPENDIX A - Specifications... 93 Master Drug Library (MDL) Editor...93 Drug Library Transfer...93 Standard Gravity IV Sets...93 Standards...93
APPENDIX B - Flow Rate Accuracy... 99 Effect of Fluid Container Height...99 Effect of Back Pressure...99 Flow Profile...99
APPENDIX C - Bolus Accuracy... 108 APPENDIX D - Downstream Occlusion... 109 Time to Occlusion...109 Bolus Volume...109
APPENDIX E - Electromagnetic Compatibility... 110 Emissions...110 Immunity – ESD, transient/burst, voltage disparity, magnetic...111 Immunity – Conducted and Radiated...113 Immunity – Separation Distances...115
APPENDIX F - Low / Very Low Battery Tutorial... 116 Low Battery Tutorial...116 Very Low Battery Tutorial...117
APPENDIX G - Icons... 118 Power Icons...118 Wireless Icons...120 General Icons...121
APPENDIX H - SPECTRUM DEFAULT SETTINGS... 122 Factory Settings for Pump Software...122
VI
Manual 41018 - 6.05/6.2.4 Revision H
INTRODUCTION AND SAFETY Intended Device Use The Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of intravenous fluids. These fluids may include pharmaceutical drugs, blood, blood products, and mixtures of required patient therapy. The intended routes of administration consist of the following clinically acceptable routes: intravenous, arterial, subcutaneous, intrathecal, epidural or irrigation of fluid space. The Spectrum is intended to be used in conjunction with legally marketed and compatible intravenous administration sets and medications provided by the user. The Spectrum Infusion Pump with Master Drug Library are suitable for many user facility applications such as, but not limited to, hospitals, outpatient care areas, home care, and ambulatory care services. The Spectrum Infusion Pump with Master Drug Library is intended to reduce operator interaction through guided programming, thereby helping to reduce errors associated with complex device programming. Parameter programming requires trained health care professional confirmation of limits and drug therapy to physician directive.
Related Documents The following documents also pertain to the SIGMA Spectrum with Generation 2 Operating System: ■ ■
Service Manual P/N 41019 MDL User Manual P/N 41020
Regulatory Information Conforms to UL STD 60601-1 Certified to CAN/CSA STD C22.2 NO 601.1-M90
Contacting SIGMA Technical Support Contact SIGMA for all service information at: Telephone:1-800-356-3454 E-mail: [email protected]
1
Introduction and Safety
Pump Operating Software v6.05 For Use With MDL Editor v6.2.4
Conventions
WARNING:
CAUTION:
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate personal injury or property damage. This word is used to also alert against unsafe practices.
Summary of Warnings and Cautions
WARNING:
The SIGMA Spectrum with Master Drug Library has not been tested or approved for use in motor vehicles or aircraft.
WARNING:
Operation is Limited to Trained and Tested Operators SIGMA Spectrum operation is strictly limited to trained operators whose competency in safe Spectrum operation and in safe IV therapy practices has been tested and proven. Pump owners have sole responsibility for operator training and testing even when SIGMA personnel assist in training processes.
WARNING:
Confirm Safe Operation at Start and Thereafter Confirm safe, accurate pump operation at start and periodically thereafter by: - Ensuring that IV sets or container vents are properly functioning, tubing clamps are in the proper positions and tubing is free from kinks or signs of collapse outside the pump. - Observing the drip chamber to verify that there is no flow from the fluid container when the pump is stopped. - Confirming the drip rate approximates the pump’s flow rate during RUN operation. - Confirming pump settings are as intended. - Confirming correct: patient, route, dose rate, dose mode, time and drug/concentration. - Periodically check battery status and replace if necessary. - With IV therapy, vital signs and IV access sites are monitored per facility’s standard practice of care.
2
Manual 41018 - 6.05/6.2.4 Revision H
- The Spectrum Generation 2 Operating System is not intended to replace clinician patient observation. - When using the pump periodic patient monitoring must be performed to ensure that the infusion is proceeding as intended. - The pump was not designed nor is it intended to detect infiltrations or extravasations. Never operate the Spectrum unless all of the above safe operations are being practiced. WARNING:
Prevent Inaccuracy The following can cause flow rate inaccuracies and must be avoided: - Incompatible brand IV sets and compatible brand IV sets with unusually large or small diameters or unusually stiff materials. - Operating temperatures outside of 60-90°F for Standard Battery and 60-80°F for Wireless Battery Module. - Using IV sets longer than is recommended in the Specifications section of this manual. - Using a dropped, damaged, dirty or wet pump. - Pressurizing IV bags. - Positioning IV containers more than 2 feet above or 1 foot below the pump. - Non-vented IV sets with rigid non-vented containers. - Vents on sets or burettes left in the closed position when they should be open. - Using Microdrip or Minidrip chambers for flow rate settings greater than 200 mL/hr. Doing so may influence flow rate accuracy and cause nuisance air-in-line or upstream occlusion alarms. - Laying the IV container flat for long periods of time. Doing so may influence flow rate accuracy and cause air-in-line or upstream occlusion alarms. - Exceeding 500 mL/hr flow rate settings when using sets with backcheck valves. Doing so may influence flow rate accuracy or cause nuisance upstream air or upstream occlusion alarms. Flow rates above 300 mL/hr may cause fluid to be siphoned from the primary container during piggyback operation (see Secondary Infusion). Not applicable with non-DEHP tubing because 250 mL/hr is the maximum flow rate per warning statement.
WARNING:
3
Upstream occlusions caused by improperly vented glass bottles or burettes may not be detected because of the very slow-building vacuums resulting from these situations.
Introduction and Safety
Pump Operating Software v6.05 For Use With MDL Editor v6.2.4
WARNING:
Follow Epidural Precautions Epidural administration of drugs other than those indicated for epidural use can result in serious patient injury. - When administering epidural analgesics, use only catheters specifically labeled for epidural analgesia drug delivery. - To help prevent accidental infusion of non-epidural drugs, DO NOT USE epidural administration sets that contain injection sites. - Label the administration container and IV set “EPIDURAL USE ONLY”. - Clearly identify infusion pumps used for epidural administration. - Use Keypad Lock to guard against unauthorized pump access by a patient or family member.
WARNING:
Follow Neonatal and Pediatric Precautions - Use 60 drop/1 mL IV sets. - Configure the pump with appropriate flow rate, VTBI (Volume To Be Infused), patient weight and occlusion alarm limits. - Prior to connecting to patient, prime IV set, then close roller clamp, load IV set, open slide clamp and roller clamp (if equipped) to avoid possible bolus (0.2 mL) that would result around a door opening/set loading event. - If the pump door is opened with an IV set connected to a patient. Bolusing at door closing must be avoided. Before closing the door, clamp the set below the lower Y injection site. Connect a syringe to the lower Y injection site, close the door, open the slide clamp, collect a 0.085 mL bolus in the syringe and unclamp the set below the Y injection site.
WARNING:
Do Not Allow Uncontrolled Gravity Flow When loading a primed IV set, ensure, before pump manipulation, that the roller clamp below the pump is in the closed position. To open the pump door, the IV set’s slide clamp must first be closed (thus providing “set-based anti-free flow” protection). Do not open the slide clamp when the door is open or during and after IV set unloading. This can cause dangerous, uncontrolled free flow to occur. During IV container changes, always close the set’s roller clamp. When the set is in the pump and the door is closed, the slide clamp can safely be opened. If gravity flow is to be used, the pump door will be open or the set will be outside the pump. Verify gravity flow is maintained at the intended rate whenever the pump door is open and when the set is outside of the pump.
WARNING:
A fluid bolus will occur (maximum of 0.1 mL) when the slide clamp is removed and the administration set is loaded.
4
Manual 41018 - 6.05/6.2.4 Revision H
WARNING:
Proper Disposal Required To dispose of this device or the associated administration sets, adhere to local, state, federal and/or other governing regulations.
WARNING:
Use the Specified Manufacturer’s IV Set Type The label is located on the top of the pump and indicates the specific type of IV tubing that the pump has been calibrated for. The use of other manufacturer’s brands or type tubing could produce pump inaccuracies that could be unsafe for patients.
WARNING:
Risk of Infection Do not use an administration set for longer than the manufacturer’s labeled set change interval to reduce risk of infection.
WARNING:
Upstream Occlusion Alarm Suspension - Upstream Occlusion Alarm Suspension feature should not be used when delivering critical drugs where the risk of flow stoppage due to an undetected upstream occlusion outweighs that of flow interruption due to nuisance alarms where no upstream occlusion is present. - Upstream Occlusion Alarm Suspension feature should not be used for drugs delivered in RIGID containers since the flow restriction caused by lack of proper container venting may be difficult to recognize when troubleshooting an alarm condition. - Upstream Occlusion Alarm Suspension feature should only be used after the operator visually observes positive line flow.
WARNING:
Baxter IV Sets These warnings apply to sets listed in this document by specific reference. 1. Minidrip sets should not be used for flow rate settings greater than 200 mL/hr. Doing so may influence flow rate accuracy and cause air-inline alarms or upstream occlusion alarms. 2. When using sets with backcheck valves flow rate settings should not exceed 500 mL/hr. Doing so may influence flow rate accuracy or cause upstream air-in-line or upstream occlusion alarms. Piggyback flow rates above 300 mL/hr may cause fluid to siphon from the primary container. Not applicable with non-DEHP tubing because 250 mL/hr is the maximum flow rate per warning statement. 3. When using sets with rigid polyethylene lined tubing, as is often used in nitroglycerine sets, flow rate settings should not exceed 500 mL/hr. Doing so may influence flow rate accuracy. 4. Partially occluded filters can cause air-in-line, upstream occlusion or downstream alarms and can influence negative flow rate accuracy.
5
Introduction and Safety
Pump Operating Software v6.05 For Use With MDL Editor v6.2.4
5. Burettes with closed vents or shutoff valves will cause upstream occlusions that may not be detected by the infusion pump. Rigid unvented containers used with unvented sets or vented sets with vent closed, will cause upstream occlusions that may not be detected by the infusion pump. 6. Closed ball valve operation may not be detected by the pump. 7. Rigid polyethylene lined tubing, often used in nitroglycerine sets, may produce as much as 10 PSI downstream occlusion pressure above the lower limit of the pump specification. 8. Some sets contain two or more slide clamps. Only the slide clamp on the pumping section or on the section with the main roller clamp should be used for pumping operation and clamp detection. Other slide clamps associated with the set need to be observed and controlled by the user. 9. Blood sets with both clamps closed above the blood filter will cause upstream occlusions that may not be detected by the pump. 10. Sets containing manifolds may produce longer times to detection of downstream occlusion. 11. When using the compatible, non-DEHP IV administration sets in Spectrum pumps, the following performance limitations must be observed: 11.1 Flow rate accuracy will range ±10% from the expected volume, when evaluated for over a one-hour period, and not the ±5% specified in Spectrum Operator Manuals for Baxter “S” type DEHP IV sets. 11.2 Flow rate range and IV set usage duration for Baxter nonDEHP IV administration sets is limited to: ■10 – 125 mL/hr with IV tubing use of not greater than 36
hours ■126 – 250 mL/hr with IV tubing use of not greater than 4
hours 11.3 The Baxter non-DEHP administration sets should not be used with the SIGMA Spectrum Pump for medications and therapies requiring infusion flow rates and durations outside of ranges specified above. 11.4 Healthcare professionals should evaluate medications, prescribed therapies and patient populations prior to utilizing the Spectrum Infusion System with non-DEHP IV tubing. See the Specification Section for Downstream Occlusion times and bolus release information.
6
Manual 41018 - 6.05/6.2.4 Revision H
WARNING:
Hospira IV Sets These warnings apply to sets listed in this document by specific reference. 1. Microdrip chambers should not be used for flow rate settings greater than 200 mL/hr. Doing so may influence flow rate accuracy and cause air-in-line or upstream occlusion alarms. 2. When using sets with backcheck valves, flow rate settings should not exceed 500 mL/hr. Doing so may influence flow rate accuracy or cause nuisance upstream air or upstream occlusion alarms. Flow rates above 300 mL/hr may cause fluid to be siphoned from the primary container during piggyback operation (see Secondary Infusion). 3. Partially occluded filters can cause air-in-line upstream occlusion or downstream alarms and influence flow rate accuracy. 4. Burettes with closed vents or shutoff valves will cause upstream occlusions that may not be detected by the infusion pump. 5. Yellow Key Slide Clamp sets are only compatible with Spectrum software of 4.02.06 or higher, keyed for correct direction of flow. 6. Sets having a length that is greater than 48 inches from the exit of the pump to the patient connection end may have an increased downstream occlusion pressure, time to occlusion and bolus at occlusion release. For rates of less than 100 mL/hr, the pump should be set to the LOW downstream pressure setting. 7. Some sets contain two or more slide clamps. Only the slide clamp on the pumping section or on the section with the main roller clamp should be used for pumping operation and clamp detection. Other slide clamps associated with the set need to be observed and controlled by the user. 8. Blood sets with both clamps closed above the blood filter will cause upstream occlusion conditions that may not be detected by the pump. See the Specification Section for Downstream Occlusion times and bolus release information.
WARNING:
The Spectrum Pump is not designed to be MRI-compatible nor is it intended to be used in this manner. Strong magnetic fields (those beyond the level tested) may cause the device to operate improperly. Do not expose the SIGMA Spectrum to strong magnetic fields such as is common with MRI equipment. Doing so may cause injury to the patient and/or damage to the equipment.
WARNING:
Low Battery Do not use battery operation or transport a patient when pump is in a low battery state.
7
Introduction and Safety
Pump Operating Software v6.05 For Use With MDL Editor v6.2.4
WARNING:
Confirm Master Drug Library (MDL) - Master Drug Library Administrators (MDLAs) should confirm the appropriate drug library is installed when deploying the Master Drug Library to pumps. - Users should confirm the appropriate drug library is installed on the deployed pumps.
WARNING:
Hyperbaric Chamber The SIGMA Spectrum Infusion System has not been tested or evaluated for use in a hyperbaric chamber.
WARNING:
IV Set Loading - Load tubing directly from the slide clamp to the top of the tubing channel. Confirm the tubing from the IV container enters the back of the slide clamp and exits the front of the slide clamp prior to loading the tubing section into the pump channel. - Improper or reverse IV set loading will result in a no flow condition to the patient as well as possible back flow of blood from the IV set into the IV tubing and/or occlusion/air-in-line alarms. - Follow the direction of flow diagram and screen prompts to load IV set tubing correctly.
WARNING:
The Spectrum Pump is not designed to be exposed to linear accelerator radiation nor is it intended to be used in this manner. Exposure to radiation of this type may cause the device to operate improperly. Do not expose the SIGMA Spectrum to linear accelerator radiation. Doing so may cause injury to the patient and/or damage to the equipment.
WARNING:
Priming Prior to connecting to patient, prime IV set, removing all air, then close roller clamp, load IV set, open slide clamp and roller clamp.
WARNING:
Only use the AC power adaptor specified for this equipment. Using other power adaptors may cause personal injury or damage to equipment.
WARNING:
The use of accessories or cables other than those specified by SIGMA may result in increased Emissions or decreased Immunity of this medical device.
8
Manual 41018 - 6.05/6.2.4 Revision H
WARNING:
Ensure Secure Mounting of Pump During Use and Transport. During use and transport, securely mount IV pumps to pole by centering the pole in the clamp and turning the mounting knob clockwise. To maintain IV pole stability never exceed 210 cm (83 in) from floor to IV pole top and limit bag volume at this extended height to < 1 liter (1000 cc).
WARNING:
Adjacent or Stacked Use The pump should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the pump should be observed to verify normal operation in the configuration in which it will be used.
NOTE:
The Spectrum Infusion System has been tested to operate normally when used stacked or adjacent to other Spectrum Infusion Systems.
WARNING:
Unauthorized View or Access Always guard the keypad lock code from unauthorized view or access. Uncontrolled access by a patient or family member may cause injury to the patient.
WARNING:
Battery Handling - Do not short circuit battery terminals. - Do not disassemble or modify battery packs. - Do not dispose of batteries or battery pack in fire.
WARNING:
Do Not Reuse Tubing Do not reload pumped-on tubing (segment used in pumping channel) into the pumping channel. Doing so will cause alarms and adversely affect flow rate accuracy.
WARNING:
Pump Storage Remove the Battery Module from the pump when storing the pump for extended periods.
WARNING:
Specifications for Downstream Occlusion detection times and bolus volume, after release of occlusion, are based on specific test conditions. The analytical related conditions are: - A distance of 48” from the point of the downstream occlusion to the SIGMA Spectrum’s Downstream Occlusion sensor (approximately the distance from the IV administration set’s exit from the pumping channel to the point of occlusion). - The 48” test administration set contained one “y”-site (no filters, or other components).
9
Introduction and Safety
Pump Operating Software v6.05 For Use With MDL Editor v6.2.4
- Testing was at the nominal room temperature (72°F ±2°F). Time to Downstream Occlusion and Bolus Volume release will generally increase under the following conditions: longer distances to the occlusion point, additional fluid volumetric area (from filters or other components within the IV set length) and hotter room temperatures. WARNING:
Siphoning Close the clamp on the secondary line or remove the secondary container administration to prevent the secondary medication from flowing when the Primary mode is intended.
WARNING:
CAUTION:
Failure to prime/remove all air bubbles from backcheck valve in a primary set may cause the valve to malfunction, resulting in secondary fluid flow back up into the primary container.
Recyclable Battery Pack. Dispose of Properly The SIGMA Spectrum Infusion System contains a lithium-ion rechargeable battery pack. It should not be disposed of in trash or in fire. It is a recyclable product and should be disposed of properly. Return battery to SIGMA for disposal if an authorized disposal center cannot be found.
CAUTION:
Pump Orientation Always operate the pump vertically on the IV pole or bed rail, with the slide clamp keyhole at the top of the pump.
CAUTION:
Use Stable IV Poles Mount pumps on IV poles that securely hold the pump.
CAUTION:
Maintain Battery Charge To maintain battery charge, keep the pump’s AC Power Adaptor plugged into a powered outlet whenever possible, including when the pump is not in use.
10
Manual 41018 - 6.05/6.2.4 Revision H
CAUTION:
Keypad Usage -
Only program the device with the pad or tip of a finger.
-
Do not use sharp objects to depress keys, such as the tip of a pen or the edge of an ID badge. Doing so may damage the pump keypad making the keys inoperable.
-
If keypad malfunctions, discontinue use immediately and sequester pump pending inspection.
Correct Keypad Usage.
Incorrect Keypad Usage.
CAUTION:
The Power Adaptor with Protector is compatible with the cleaners mentioned in this Operator’s manual. For the methods of cleaning and compatible cleaners that can be used on the AC Power Adaptor with Protector. See “Cleaning and Storage” on page 78.
CAUTION:
Service Personnel Must be Trained by SIGMA Servicing the SIGMA Spectrum Infusion System is restricted to qualified, SIGMA trained, service personnel who employ SIGMA authorized parts and procedures. Use of other parts and servicing procedures is prohibited.
CAUTION:
Perform Preventative Maintenance Annually Pumps should be tested for proper performance annually and whenever damage from drops, fluid intrusion and other causes is suspected. See SIGMA Spectrum Service Manual for complete information.
11
Introduction and Safety
Pump Operating Software v6.05 For Use With MDL Editor v6.2.4
CAUTION:
CAUTION:
Cleaning the Pump and Pump Accessories -
Always wear gloves when cleaning pump and pump accessories.Only use SIGMA specified compatible cleaning fluids.
-
Do not allow fluid to seep inside pump (especially through front case door latch holes or rear case speaker holes) or severe damage may occur.
-
Do not spray solutions directly on the pump and pump accessories.
-
Do not autoclave or ETO (ethylene oxide) to sterilize pumps or pump accessories.
-
Do not apply cleaners directly to battery packs exposed terminals.
-
Do not immerse any part of the pump or battery.
-
Do not use phenol-based cleaners/disinfectants. Phenols degrade plastics and membrane switches.
-
Do not use abrasive cleaners.
-
Do not use rigid cleaning instruments.
-
Always use a lint-free, foam tipped swab to clean the tube channel.
-
Always dispose of all cleaning materials per federal, state and local regulations for infectious waste disposal.
Radio Frequency Interference The Spectrum Pump meets the electromagnetic compatibility (EMC) requirements as specified in the International Electrotechnical Commission’s (IEC) 60601-1-2 (2001-09) standard for emissions and immunity. It is good practice to keep the pump separated away from other equipment, such as hand-held transmitters, cellular phones and electrosurgical equipment that may generate strong radio frequency interference (RFI). Refer to the EMC Immunity Section, Separation Distance, in this manual for recommended minimum distance.
CAUTION:
Confirm Audio Operation When pressing the ON key and all other keys, confirm that an audio beep is heard. If sound cannot be heard, discontinue use of the pump and return to SIGMA for service.
CAUTION:
Confirm Display Operation Regularly observe the pump’s display. Discontinue use of the pump and return to SIGMA for service if display abnormalities are observed.
CAUTION:
Electric Shock Hazard Refer servicing to qualified service personnel at your institution or return to SIGMA.
12
Manual 41018 - 6.05/6.2.4 Revision H
CAUTION:
Accuracy Refer to trumpet curves for flow rate accuracy as a function of short infusion durations. See “APPENDIX B - Flow Rate Accuracy” on page 99. The upstream occlusion detector may not detect partially occluded tubing. Always check to ensure the IV set’s clamp is not closed above the Spectrum Pump and respond appropriately to all primary and secondary check flow prompts. Small bore catheters or needles may cause excessive back pressure at high flow rates. Size the catheters according to expected flow rate and fluid viscosity.
CAUTION:
Follow Physicians Orders Federal (USA) law restricts this device to sale or use by, on the order of, or under the supervision of, a physician or other licensed health care practitioner.
CAUTION:
NOTE:
13
This equipment is not suitable for use in the presence of a Flammable Anesthetic Mixture with Air or with Oxygen or Nitrous Oxide. This statement applies to oxygen enriched environments, such as oxygen tents. It is not meant to apply to patients on breathing tubes. Refer to IEC60601-2-24.
CAUTION:
Securely mount IV pumps to pole by turning the mounting knob clockwise. To maintain IV pole stability never exceed 83” (210 cm) from floor to IV pole top and limit bag volume at this extended height to < 1 liter (1000 cc).
CAUTION:
Always route IV set tubing and AC Power Adaptor cabling to prevent patient hazard or entanglement. Identify the individual IV set lines when multiple pumps and routes of administration are practiced.
CAUTION:
The use of BASIC programming should be restricted and actively monitored by a hospital’s Quality Assurance, Risk, Pharmacy and Nursing departments.
Introduction and Safety
Pump Operating Software v6.05 For Use With MDL Editor v6.2.4
CAUTION:
ECG Artifacts Related to the Use of the Spectrum Pump Peristaltic infusion pumps may produce what is known as piezoelectric artifact on ECG monitors and similar types of monitoring instruments. The Spectrum Pump may produce this effect when the infusion pump is running at rates in the higher ranges of operation, this may be in the frequency range tracked by the ECG monitor. The appearance of the artifact may be affected by set up and/or connection of electrodes, leads, or equipment. See the ECG monitoring system documentation for recommendations on proper set up including electrode connections, site preparation, monitor system set up and electrode placement.
CAUTION:
Static Sensitive Equipment Wherever possible, eliminate any electro-static producing materials or conditions (dry, low humidity, synthetic materials such as blankets, carpeting, etc.)
CAUTION:
Avoid Bright Overhead Light Bright Light sources (>100 watt bulb) directly above the pumps keyhole (load point #1) and within 30.5 cm (1 ft) of the top of the pump may cause alarms or continuous system errors.
CAUTION:
-
Increase the distance between the pump and the light source.
-
Move the pump to an adjacent location rather than directly below the light source.
Avoid Overheating When operating the pump, keep out of bright sunlight or direct heat sources to prevent overheating.
CAUTION:
Sequester Pumps Pending Inspection Devices that are believed to have malfunctioned and/or were involved in an adverse event should be immediately removed from service and quarantined pending their inspection and/ or return to SIGMA for inspection and service.
14