Owners Manual
28 Pages

Preview
Page 1
WARMING CABINETS MODELS SS2201-M(G) SS2203-M(G) SS2205-M(G) SS2207-M(G) SS2209-M(G)
OWNER'S MANUAL (Includes Operation, Maintenance and Replacement Parts) Read this manual before starting to work! This information is necessary for safe and efficient operation of the equipment. TEC-F-0044 REV0 4/12/2013
TABLE OF CONTENTS TITLE
Page 1 PAGE
SECTION 1. SAFETY INFORMATION... 2 1-1. Special User Attention ... 2 1-2. Safety Messages ... 2 SECTION 2. EQUIPMENT SPECIFICATIONS... 4 2-1. Intended Use ... 4 2-2. Installation ... 4 2-3. Environmental Conditions ... 4 2-4. Certification... 4 2-5. Equipment Labels ... 5 2-6. Dimensions... 6 SECTION 3. OPERATION ... 12 3-1. Introduction ... 12 3-2. Power Requirements ... 13 3-3. Start-up Procedure ... 13 3-4. Controls ... 14 3-5. Shutdown Procedure ... 17 SECTION 4. MAINTENANCE ... 18 4-1. Cleaning ... 18 4-2. Sterilization ... 18 4-3. Operator Maintenance ... 18 4-4. Troubleshooting ... 19 4-5. Preventive Maintenance ... 20 4-6. End of Useful Life and Disposal ... 20 SECTION 5. REPLACEMENT PARTS ... 21 5-1. MODEL SS2201-M, SS2201-MG, SS2203-M, SS2203-MG... 21 5-2. MODEL SS2205-M, SS2205-MG... 22 5-3. MODEL SS2207-M, SS2207-MG... 23 5-4. MODEL SS2209-M, SS2209-MG... 24 SECTION 6. REVISION HISTORY... 25 Model Identification: The identification tag is located inside the upper chamber at the center of the backside near the top. Record your Model Number and Serial Number below. For prompt service, please have your model and serial number ready when contacting SKYTRON for assistance. Model # _______________________________ Serial # _______________________________
Distributed by: SKYTRON 5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI 49512 (616) 656-2900 www.skytron.us
Manufactured by: MAC Medical, INC. 820 South Mulberry Street Millstadt, IL 62260 (618) 476-3550 www.macmedical.com
Although current at the time of publication, SKYTRON’S policy of continuous development makes this manual subject to change without notice. M WARMING CABINET OWNER’S REV0
Page 2
SECTION 1. SAFETY INFORMATION
1-1. Special User Attention
1-2. Safety Messages
Prior to use, all personnel that will operate the Warming Cabinet must be instructed in the correct usage and operation.
The following precautions must be reviewed by all personnel before operating the Warming Cabinet.
A routine program must be implemented by the facility for proper usage instructions for all personnel that may operate the Warming Cabinet. All hospital personnel should be aware that sensible care must be exercised to maintain patient safety and keep the Warming Cabinet performing at peak efficiency. The design of the SKYTRON M Warming Cabinet does not utilize internal user serviceable parts or procedures. Service must be performed by SKYTRON authorized service representatives using SKYTRON authorized replacement parts and service techniques.
DANGER Indicates a hazardous situation that, if not avoided, will result in death or serious injury.
WARNING Indicates a hazardous situation that, if not avoided, could result in death or serious injury. Explosion and Fire Hazard: DO NOT use Warming Cabinet in the presence of flammable anesthetics or solvents. Loading the cabinet with any item that could introduce flammable agents into the cabinet atmosphere can cause a fire or explosion. Injury and Burn Hazard: DO NOT use liquids on, or inject in, living tissue unless actual liquid temperature has been measured and is acceptable. The temperature of the Warming Cabinet contents may be hotter than the displayed air temperature. For patient safety, in accordance with optimal medical practice, always verify liquid temperature prior to using. Explosion Hazard: DO NOT exceed 150° F for non-vented closures; (screw caps, crimp seals, plastic pouches, etc.) DO NOT exceed pre-sterile solution manufacturer’s temperature requirements. Injury and Burn Hazard: Disconnect the AC power cord and allow the warmer to cool before performing preventive maintenance.
M WARMING CABINET OWNER’S REV0
Page 3
CAUTION Indicates a hazardous situation that, if not avoided, could result in minor or moderate injury. Explosion or Fire Hazard: The Warming Cabinet is intended for warming non-flammable solutions and blankets made of Muslin, 100% cotton or wool, ONLY. Operating the Warming Cabinet with flammable items could result in explosion or fire. Burn Hazard: DO NOT raise the setpoint temperature to increase the rate of heating. On double or triple chamber Warming Cabinets, do not exceed 30°F (16.5°C) higher temperature than the chamber above. Example: Upper chamber 110° (43°C) lower chamber must not exceed 140° (60°C). This could overheat the contents leading to possible patient burns. Burn Hazard: The recommended temperature limit for warming solutions is 110°F (43°C). Solutions warmed above 111°F (44°C) can cause second degree burns. The recommended limit for blankets is 130°F (54°C). Even slightly higher temperatures have been associated with patient burns. Burn Hazard: Use personal protective equipment when loading the contents into the Warming Cabinet. Avoid contact with hot internal surfaces that could cause burns. Injury and Burn Hazard: Avoid injury by using proper personal protective equipment when unloading the Warming Cabinet. Internal surfaces are hot, glass may shatter when cooled suddenly and solution bags or bottles may burst when picked up.
M WARMING CABINET OWNER’S REV0
The design of the Warming Cabinet allows limited user serviceable parts or procedures. For optimal usage, safety and durability, service must be performed by a SKYTRON authorized service technician using SKYTRON authorized replacement parts and service techniques.
CAUTION CAUTION without the safety alert symbol, indicates a possibility of damage to equipment not related to personal injury. Any parts or assemblies not listed in this section must be serviced or replaced by SKYTRON authorized service personnel only. This is necessary to avoid any possibility of damage to the equipment.
NOTICE Indicates important information not related to personal injury.
Page 4
SECTION 2. EQUIPMENT SPECIFICATIONS
2-1. Intended Use The SKYTRON Warming Cabinet is intended to be used by medical personnel for the purpose of providing heated storage of blankets, sterile water and saline solutions used in the care of patients in surgery, recovery, OB/GYN, ICU, ER and trauma areas in healthcare facilities where all operators are instructed on the usage, limitations and hazards. No other use is authorized or recommended. 2-2. Installation The SKYTRON Set-up Instructions specify the unpacking, installation, and testing of the M Series Warming Cabinet. Prior to starting the installation, review the Set-up Instructions and local electric codes including the Occupational Health and Safety Act for any requirements pertaining to the proper installation of this equipment.
CAUTION Explosion or Fire Hazard: The Warming Cabinet is intended for warming non-flammable solutions and blankets made of Muslin, 100% cotton or wool, ONLY. Operating the Warming Cabinet with flammable items could result in explosion or fire.
Contact your SKYTRON representative for seismic calculations and tie-down hardware if applicable. 2-3. Environmental Conditions DURING TRANSPORT AND STORAGE (IN ORIGINAL PACKAGING MATERIALS) • AMBIENT TEMPERATURE: -40° - 70° C (-40° - 159° F) • RELATIVE AIR MOISTURE: 10% - 100%, INCLUDING CONDENSATION • AIR PRESSURE: 500 hPa - 1060 hPa (14 inHg - 31.3 inHg) DURING USE - FOR DRY LOCATIONS • AMBIENT TEMPERATURE: 15° - 30° C (60° - 85° F) • RELATIVE AIR MOISTURE: 30% - 60% NON CONDENSING • AIR PRESSURE: 700 hPa - 1060 hPa (20.7 inHg - 31.3 inHg) The Warming Cabinet is intended for use in a stable environment, with an ideal ambient temperature of 72° F (22°C) or less. The Warming Cabinet should never be located next to any appliance that may produce heat, such as an autoclave. If the Warming Cabinet must be located next to a heat producing appliance, action must be taken to prevent heat transfer. Contact your SKYTRON representative for assistance. 2-4. Certification
M WARMING CABINET OWNER’S REV0
Page 5 2-5. Equipment Labels
1
3
5
QC Final Inspect OK
2 2 Glass Door
4
BACK SIDE
FRONT
1 OPERATING INSTRUCTIONS:
2 (Part No. W0227-SK)
1. To apply power to the system:
5
• Press the power switch to the ON position
2. To change the setpoint:
• Press SET • Press the arrow keys to alter the setpoint either up or down • Press SET again
3. To silence the alarm:
TEMPERATURE SET POINT FOR BLANKETS FOLLOW BLANKET MANUFACTURERS RECOMMENDATIONS FOR SET POINT. FOR INTRAVENOUS AND IRRIGATION FLUIDS FOLLOW TEMPERATURE GUIDE LINES PRINTED ON CONTAINER OR CONTACT YOUR SUPPLIER FOR TEMPERATURE AND EXPIRATION PERIODS. THIS APPLIANCE NOT APPROVED FOR WARMING BLOOD PRODUCTS.
• Press ALARM RESET
FOR PROPER CIRCULATION ALLOW 1” / 25MM OF SPACE BETWEEN TOP/SIDES OF CABINET AND ITEMS 1” / 25MM OF SPACE
1” / 25MM OF SPACE
DOWNLOADING DATA:
W0227-SK
1. To retrieve warmer data insert flash drive into USB port. 2. The digital display will change from displaying temperature to displaying this code 3. Once the transfer of data is complete the digital display will return to displaying cabinet temperature. 4. Remove the flash drive.
3 QC Final Inspect OK
4 (Part No. W0034) WARNING
PATIENT BURN HAZARD. DO NOT USE LIQUIDS ON-OR INJECT INLIVING TISSUE UNLESS ACTUAL LIQUID TEMPERATURE HAS BEEN MEASURED AND IS ACCEPTABLE. TEMPERATURE OF WARMING CABINET CONTENTS MAY BE HOTTER THAN THE DISPLAYED CHAMBER AIR TEMPERATURE. FOR PATIENT SAFETY, IN ACCORDANCE WITH GOOD MEDICAL PRACTICE, ALWAYS CHECK LIQUID TEMPERATURE PRIOR TO USING.
M WARMING CABINET OWNER’S REV0
1” / 25MM OF SPACE
BURN HAZARD. DO NOT EXCEED 1500 F FOR NON-VENTED CLOSURES (SCREW CAPS, PLASTIC POUCHES, ETC.) DO NOT RAISE SET TEMPERATURE TO INCREASE RATE OF HEATING, ALLOW APPROXIMATELY 6 HOURS FOR SOLUTIONS TO REACH DESIRED TEMPERATURE.
WARNING
Page 6 2-6. Dimensions Exterior and Interior - inches (millimeter) SS2201-M(G)
SS2203-M(G)
SS2205-M(G)
SS2207-M(G)
SS2209-M(G)
Width
30 (762)
30 (762)
30 (762)
30 (762)
30 (762)
Height
24½ (622)
36 (914)
74½ (1892)
74½ (1892)
78¼ (1988)
Depth
26½ (673)
26½ (673)
26½ (673)
26½ (673)
26½ (673)
Upper
Upper
Exterior
Interior Width
25¼ (641)
25¼ (641)
25¼ (641)
25¼ (641)
25¼ (641)
Height
13½ (343)
23¼ (591)
57½ (1461)
13½ (343)
11¼ (286)
Depth
21½ (546)
21½ (546)
21½ (546)
21½ (546)
21½ (546)
Middle Width
25¼ (641)
Height
10¾ (273)
Depth
21½ (546)
Lower
Lower
Width
25¼ (641)
25¼ (641)
Height
34½ (876)
18½ (470)
Depth
21½ (546)
21½ (546)
Capacity - cubic feet (liter) SS2201-M(G) Total Volume 5.27 ft
3
(149 L)
SS2203-M(G)
SS2205-M(G)
SS2207-M(G)
SS2209-M(G)
8.63 ft
21 ft
3
(244 L)
3
(595 L)
17.2 ft
3
(487 L)
14 ft3
(396 L)
5.27 ft3
(149 L)
3.9 ft3
(110 L)
Upper Chamber
N/A
N/A
N/A
Middle Chamber
N/A
N/A
N/A
N/A
3.7 ft3
(105 L)
Lower Chamber
N/A
N/A
N/A
11.93 ft3 (338 L)
6.4 ft3
(181L)
SS2201-M(G)
SS2203-M(G)
SS2205-M(G)
SS2207-M(G)
SS2209-M(G)
175 (79.4)
220 (99.8)
355 (161.5)
430 (195.5)
495 (225)
Weights - pounds (kilograms) Weight
Electrical Information SS2201-M(G)
SS2203-M(G)
SS2205-M(G)
SS2207-M(G)
SS2209-M(G)
Electrical
120VAC, 50/60Hz,1PH
120VAC, 50/60Hz,1PH
120VAC, 50/60Hz,1PH
120VAC, 50/60Hz,1PH
120VAC, 50/60Hz,1PH
Amp Rating
7.5 Amps
7.5 Amps
7.5 Amps
12.0 Amps
13.5 Amps
Power Usage
300 Watts
350 Watts
530 Watts
750 Watts
656 Watts
btu/hr
1024
1195
1810
2557
2238
M WARMING CABINET OWNER’S REV0
Page 7
CONTROL PANEL 30” (762 mm) UPPER CHAMBER
ON
ACTUAL SETPOINT
OFF
HEAT ALARM
SET
DATA
ALARM RESET
SETPOINT UNLOCK LOCK
ADJUSTABLE SHELF
24½” (622 mm)
25¼” (641 mm) INSIDE DIMENSION OF CHAMBER
4 RUBBER LEG LEVELERS
POWER CORD, 7 ft (2.1 m) LONG WITH HOSPITAL GRADE PLUG 26½” (673 mm) 24” (609 mm) 6” (152 mm)
18¼” (463 mm)
21½” (546 mm) INSIDE DIMENSION OF CHAMBER
13½” (343 mm)
Figure 2-1. MODEL SS2201-M, SS2201-MG (shown)
M WARMING CABINET OWNER’S REV0
Page 8
CONTROL PANEL 30” (762 mm) UPPER CHAMBER ACTUAL SETPOINT
OFF
HEAT ALARM
ON
SET
DATA
ALARM RESET
SETPOINT UNLOCK LOCK
34” (863 mm) 36” (914 mm)
2” (50 mm)
25¼” (641 mm) INSIDE DIMENSION OF CHAMBER
ADJUSTABLE SHELVES (2)
29¾” (755 mm) 26½” (673 mm)
POWER CORD 7 ft (2.1 m) LONG WITH HOSPITAL GRADE PLUG
24” (609 mm) 6” (152 mm)
28” (711 mm)
21½” (546 mm) INSIDE DIMENSION OF CHAMBER
23¼” (591 mm)
23” (584 mm)
Figure 2-2. MODEL SS2203-M, SS2203-MG (shown)
M WARMING CABINET OWNER’S REV0
Page 9
POWER CORD, 7 ft (2.1 m) LONG WITH HOSPITAL GRADE PLUG CONTROL PANEL
26½” (673 mm)
30” (762 mm) UPPER CHAMBER HEAT ALARM
SETPOINT
ON
ACTUAL
OFF
SET
DATA
ALARM RESET
SETPOINT UNLOCK LOCK
24” (609 mm) 6” (152 mm)
ADJUSTABLE SHELF (1)
69¾” (1771 mm)
21½” (546 mm) INSIDE DIMENSION OF CHAMBER
25¼” (641 mm) INSIDE DIMENSION OF CHAMBER
57½” (1461 mm)
63¾” (1619 mm)
74½” (1892 mm)
ADJUSTABLE SHELF (1)
4¾” (120 mm)
29¾” (755 mm)
28¾” (730 mm)
23” (584 mm)
Figure 2-3. MODEL SS2205-M, SS2205-MG (shown)
M WARMING CABINET OWNER’S REV0
28¾” (730 mm)
Page 10
POWER CORD, 7 ft (2.1 m) LONG WITH HOSPITAL GRADE PLUG CONTROL PANEL
26½” (673 mm)
30” (762 mm) UPPER CHAMBER HEAT ALARM
SET
DATA
ALARM RESET
6” (152 mm)
LOWER CHAMBER OFF ON
ACTUAL SETPOINT
ON
ACTUAL SETPOINT
OFF
24” (609 mm)
HEAT ALARM
SET
SETPOINT UNLOCK
DATA
ALARM RESET
LOCK
SETPOINT UNLOCK LOCK
ADJUSTABLE SHELF (1)
18¼” (463 mm)
13½” (343 mm)
69¾” (1781 mm) 74½” (1892 mm)
25¼” (641 mm) INSIDE DIMENSION BOTH CHAMBERS
45½” (1156 mm)
21½” (546 mm) INSIDE DIMENSION BOTH CHAMBERS
34½” (876 mm)
ADJUSTABLE SHELVES (2)
4¾” (120 mm)
29¾” (755 mm)
23” (584 mm)
Figure 2-4. MODEL SS2207-M, SS2207-MG (shown)
M WARMING CABINET OWNER’S REV0
Page 11
POWER CORD, 7 ft (2.1 m) LONG WITH HOSPITAL GRADE PLUG
CONTROL PANEL
6” (152 mm)
30” (762 mm) UPPER CHAMBER
SET
DATA
LOWER CHAMBER
ALARM RESET
OFF ON
HEAT ALARM
SET
SETPOINT UNLOCK
DATA LOCK
ALARM RESET
24” (609)
LOWER CHAMBER OFF ON
ACTUAL SETPOINT
HEAT ALARM
ACTUAL SETPOINT
ON
ACTUAL SETPOINT
OFF
26½” (673)
HEAT ALARM
SET
SETPOINT UNLOCK
DATA LOCK
ALARM RESET
SETPOINT UNLOCK LOCK
16” (406 mm)
73½” (1867 mm) 78¼” (1987 mm)
25¼” (641 mm) INSIDE DIMENSION ALL CHAMBERS
22” (559 mm) 10¾” (273 mm)
29½” (749 mm) 1 ADJUSTABLE SHELF PER CHAMBER
4¾” (120 mm)
28⅞” (733 mm)
11¼” (285 mm)
21½” (546 mm) INSIDE DIMENSION ALL CHAMBERS
22⅞” (581 mm)
Figure 2-5. MODEL SS2209-M, SS2209-MG (shown)
M WARMING CABINET OWNER’S REV0
18½” (469 mm)
Page 12
SECTION 3. OPERATION
3-1. Introduction SKYTRON offers 5 base models of Warming Cabinets to meet each facility’s unique needs: • Model SS2201-M(G) offers a compact size with one warming compartment. See figure 3-1. • Model SS2203-M(G) provides a mid size solution for areas that need more than a countertop version can hold but less than would be needed by a full size unit. • Model SS2205-M(G) offers a full size with one warming compartment. See figure 3-2. • Model SS2207-M(G), a dual chamber solution, provides upper and lower compartments for maintaining different heated goods at different temperature settings. See figure 3-3.
for increased heating efficiency. Doors are made of double panned stainless steel or double paned tempered glass framed with extruded aluminum. Heated air is circulated over the contents by fan(s) in a convection function providing an even heat distribution. Each compartment is designed to hold a quantity of blankets or solutions. Once a set temperature is selected and obtained it will be controlled throughout within +/- 3° F (1.7 °C) of the selected temperature. From a cold start, each compartment's loaded contents will be evenly heated to a setpoint within 8 to 12 hours. In the event of power loss the warmer will resume normal function once power is restored.
• Model SS2209-M(G) offers a unique triple chamber solution for storage of multiple items that need individual temperature setting in one cabinet eliminating the use of multiple units in a given area. See figure 3-4. The Warming Cabinets are designed to provide heated storage of blankets, sterile water and saline solutions used in the care of patients by healthcare facilities. The body of the warming cabinet is double walled stainless steel construction with insulation
Figure 3-1. Model SS2201-MG
Figure 3-2. Model SS2205-MG M WARMING CABINET OWNER’S REV0
Page 13 3-2. Power Requirements
3-3. Start-Up Procedure
The Stainless Warming Cabinet requires a properly grounded 120VAC, 50/60Hz 15Amp electrical power supply. The cabinet is equipped with a 7 foot (2.3 m) long, 14-3 SJT power cord with a 120V (NEMA 15P) hospital grade plug. The power cord is located at the rear of the cabinet in the upper left hand corner. The main power ON/OFF switch/ circuit breaker is located on the front control panel next to the temperature display.
Plug the power cord into a properly grounded 120VAC electrical outlet. Press the ON/OFF Switch, located on the front control panel, to the ON position. For multi-chamber models press the ON/OFF Switch to the ON position for each chamber. The display should illuminate with the current chamber temperature (Actual) and the setpoint temperature.
For multi-chamber cabinets there are individual ON/OFF switches for each chamber. Each switch is clearly labeled UPPER and LOWER for the dual chamber cabinet and UPPER, MIDDLE and LOWER for the triple chamber cabinet.
Figure 3-3. Model SS2207-MG M WARMING CABINET OWNER’S REV0
Figure 3-4. Model SS2209-MG
Page 14 3-4. Controls The controls are located on the upper panel (figure 3-5) on the front of the cabinet. For multi-chamber cabinets there are individual controls for each chamber. Each control is clearly labeled UPPER and LOWER for the dual chamber cabinet and UPPER, MIDDLE and LOWER for the triple chamber cabinet.
Figure 3-5. Model SS2207-M(G) dual chamber control panel.
1
ON
ACTUAL SETPOINT
OFF
UPPER CHAMBER
2
HEAT ALARM 3
SET
DATA 8
7
6
ALARM RESET
9
4 5 SETPOINT UNLOCK LOCK
Figure 3-6. Control Components 1. ON/OFF Switch/Circuit Breaker – provides power to the warming compartment and control.
5. Up ▲– used to adjust the setpoint of the chamber and silence the audible overheat alarm.
2. HEAT light – visual indicator that the heating system is active
6. Down ▼– used to adjust the setpoint of the chamber.
3. ALARM light – visual indicator of an overheat condition.
7. SET Button – press to use the Up/Down arrows to change the setpoint temperature and press again to lock the new value.
4. Display – shows the current chamber temperature (Actual) and the setpoint temperature in Fahrenheit (F) or Celsius (C). The display also provides Loss of Power and the Overheat (alarm). The overheat alarm is an audible and visual display ‘HI’.
8. DATA port – USB port used to retrieve temperature values for a set period of time. 9. Key switch – used to lock out any changes to the control.
M WARMING CABINET OWNER’S REV0
Page 15 Turning the warmer on.
WARNING Explosion or Fire Hazard: DO NOT use Warming Cabinet in the presence of flammable anesthetics or solvents. Loading the cabinet with any item that could introduce flammable agents into the cabinet atmosphere can cause a fire or explosion. 1. Make sure the power cord is plugged into a properly grounded 120VAC electrical outlet. 2. Press the control panel switch to the ON position. The display will show the actual chamber temperature. Setpoint temperature range.
CAUTION Burn Hazard: DO NOT raise the setpoint temperature to increase the rate of heating. On double or triple chamber cabinets, do not exceed 30° F (16.7 °C) higher temperature than the chamber above. Example: Upper chamber 110° F (43 °C) lower chamber must not exceed 140° F (60 °C). This could overheat the contents leading to possible patient burns. Temperature settings for fluids vary depending on use or manufacturer. Always refer to AORN guidelines and the fluid manufacturer’s recommendations for proper temperature settings.
WARNING Explosion Hazard: DO NOT exceed 150° F for non-vented closures; (screw caps, crimp seals, plastic pouches, etc.) DO NOT exceed pre-sterile solution manufacturer’s temperature requirements. Single chamber Warming Cabinet models: SS2201, SS2203 and SS2205 setpoint temperature range: • 90° F (32 °C) to 160° F (71 °C)
Double chamber Warming Cabinet model SS2207 setpoint range: • UPPER: 90° F (32 °C) to 160° F (71 °C) • LOWER: 90° F (32 °C) to 160° F (71 °C) Triple chamber Warming Cabinet model SS2209 Setpoint range: • UPPER: 90° F (32 °C) to 160° F (71 °C) • MIDDLE: 90° F (32 °C) to 160° F (71 °C) • LOWER: 90° F (32 °C) to 160° F (71 °C) All loads should be allowed time to stabilize at the Setpoint temperature: Solutions - Approximately 8 to 12 hours Blankets - Approximately 6 to 8 hours Selecting the temperature setpoint.
CAUTION Burn Hazard: The recommended temperature limit for warming solutions is 110° F (43 °C). Solutions warmed above 111°F (44 °C) can cause second degree burns. The recommended limit for blankets is 130° F (54 °C). Even slightly higher temperatures have been associated with patient burns. 1. Press the SET button on the temperature control. The setpoint temperature will change and the last digit will flash. 2. Use the Up arrow button to raise the setpoint temperature or use the Down arrow button to lower the setpoint to the temperature desired. If the setpoint is changed more than 10° below the actual temperature, the HI temperature alarm will activate. Press the ALARM RESET (▲ up arrow) button until it quits. Turn off the warming cabinet chamber and wait for the contents to cool adequately. 3. To prevent adjustment to the setpoint temperature, the control may be locked by use of the Key switch. Adjusting the Shelf
M WARMING CABINET OWNER’S REV0
Page 16 1. Turn the power OFF to the heating chamber where the adjustment is needed. 2. Allow the chamber to cool, then unload the contents. 3. Remove the shelf and determine its new position. 4. Remove the four (4 per shelf) shelf supports by compressing the shelf support and twisting the bottom outward toward the center of the chamber. Refer to figure 3-7.
Shelf Support
(4 per shelf )
Figure 3-7. Shelf Support
6. Count the mounting locations at each corner of the chamber to be sure that the shelf will be level and install the remaining supports. 7. Reinstall the shelf with the notches on the bottom of the shelf aligned with the shelf supports. 8. Pull outward on shelf to insure it is locked properly on the supports.
CAUTION Burn Hazard: Use personal protective equipment when loading the contents into the Warming Cabinet. Avoid contact with hot internal surfaces that could cause burns. SKYTRON recommends the use of personal protective equipment while loading and unloading of contents. Load only sterile water, saline solutions, or blankets made of Muslin, 100% cotton or wool, do not warm synthetic blend fabrics or items containing plastic, rubber, or metal snaps, studs, hooks etc. Check for proper placement of the shelf on shelf supports prior to loading. The blanket shelf is perforated to facilitate even heating of blankets placed on the shelf and must be used to hold blankets. Contents should be loaded to allow for a minimum of one inch from the top, back and sides of the compartment. Minimum spacing of one quarter inch between fluid containers is recommended for evenly distributed heating. Refer to figure 3-8. UPPER CHAMBER OFF
1” (25.4 mm)
ON
ACTUAL SETPOINT
5. Install the shelf support into the new location by inserting the top tab into the upper slot of the new position, compress the shelf support and rotate it downward until the bottom tab snaps into the lower slot.
Loading the Warming Cabinet
HEAT ALARM
SET
DATA
ALARM RESET
SETPOINT UNLOCK LOCK
1” (25.4 mm) 1/4” (6.4 mm) 1” (25.4 mm)
Figure 3-8. Content Spacing For efficiency purposes bagged solutions should be placed on shelves, stacking of fluid bags increases the heating time required to achieve set temperature.
M WARMING CABINET OWNER’S REV0
Page 17 Blankets must be folded and stacked to allow a one inch minimum space from the sides, back and top of the compartment or the shelf above. Also, the stacked blankets are not to protrude past the front edge of the shelf. In case of power failure the unit will resume normal operation when power is restored. Follow the fluid manufacturer’s guidelines for unused solutions that have cooled or been removed from heated storage. Unloading of the cabinet
CAUTION
Use personal protective equipment while loading and unloading the contents of the Warming Cabinet. Rotate warmed contents on a first in first out basis, failure to do so may present cold or discolored contents.
WARNING Injury and Burn Hazard: DO NOT use liquids on, or inject in, living tissue unless actual liquid temperature has been measured and is acceptable. The temperature of the Warming Cabinet contents may be hotter than the displayed temperature. For patient safety, in accordance with optimal medical practice, always verify liquid temperature prior to using.
M WARMING CABINET OWNER’S REV0
Retrieval of Recorded Temperatures All SKYTRON M Series warmers are equipped with a temperature monitoring program that stores actual chamber temperatures in 30 minute intervals. The information recorded may be down loaded to a USB flash drive at any time. To retrieve the stored data: 1. Insert the SKYTRON flash drive (W0600-S) or equivalent into the USB port marked DATA, located on the front face of the control panel. 2. The digital display will change from displaying the actual temperature to displaying the code shown in figure 3-9.
ACTUAL
Injury and Burn Hazard: Avoid injury by using proper personal protective equipment when unloading the Warming Cabinet. Internal surfaces are hot, glass may shatter when cooled suddenly and solution bags or bottles may burst when picked up.
3-5. Shutdown Procedure
Figure 3-9. Data Download Code 3. Once the transfer is complete, the digital display will return to normal operation. 4. Remove the flash drive. The information may be accessed on any Windows compatible PC that has an Excel program. Turning off the Warming Cabinet Move the ON/OFF Switch/circuit breaker to the OFF position for the compartment to shut down.
Page 18
SECTION 4. MAINTENANCE
4-1. Cleaning Clean all exterior and interior surfaces of the warming cabinet as needed with a damp towel and a mild non-abrasive detergent. Rub surfaces in a back and forth motion in the direction of the grain, do not wipe in a circular motion or perpendicular to the grain. 4-2. Sterilization
WARNING Explosion Hazard: DO NOT exceed 150° F for non-vented closures; (screw caps, crimp seals, plastic pouches, etc.) Do not exceed pre-sterile solution manufacturer’s temperature requirements. Always follow current AORN Journal Guidelines to ensure proper cleaning and disinfection procedures. 4-3. Operator Maintenance
CAUTION Injury Hazard: The design of the Warming Cabinet allows limited user serviceable parts or procedures. For optimal usage, safety and durability of the product, service must be performed by a SKYTRON authorized service technician using SKYTRON authorized replacement parts and service techniques.
It is recommended to perform preventive maintenance on the SKYTRON M Series Warming Cabinets every six months or when any service is performed.
WARNING Injury and Burn Hazard: Disconnect the AC power cord and allow the warmer to cool before performing preventive maintenance. Every six months the following must be completed: 1. Turn off the power and unplug the warmer from the power source. 2. Use a damp cloth and mild non abrasive detergent wipe all surfaces in a back and forth motion in the direction of the grain, do not wipe in a circular motion or perpendicular to the grain. 3. Clean all exterior surfaces 4. Remove all contents and clean all shelving, all interior chamber walls, floor and ceiling starting at the top and working your way down 5. Restore power to the warmer and allow temperatures to stabilize prior to loading solutions and blankets.
Users are responsible for the thorough inspection of the equipment prior to and after each use. Should any problems or deficiencies arise the results must be reported to the facilities maintenance personnel. The safety of personnel and patients relies on the proper and routine maintenance of this equipment.
M WARMING CABINET OWNER’S REV0
Page 19 4-4. Troubleshooting The following alert messages and operating conditions (figure 4-1) will occur when the warming cabinet is operating outside of acceptable conditions. Figure 4-1. TROUBLESHOOTING GUIDE DESCRIPTION
ACTION REQUIRED
HI with audible alarm
Cabinet temperature is 10° F (5.5 °C) higher than setpoint – silence the alarm by pressing ALARM RESET (Up ▲ arrow) and open door(s) to allow chamber to cool.
LLLL
Input temperature is lower than input range – check temperature probe and connections. See below for additional troubleshooting
HHHH with audible alarm
Input temperature is higher than input range – check temperature probe and connections. See below for additional troubleshooting
OPEn with audible alarm JIC continuous or flashing
Temperature probe is in fault – replace and calibrate. Control failure – replace and calibrate.
Unit will not power up
1. 2. 3. 4. 5. 6.
Chamber does not heat
1. Is the power turned on? 2. Is temperature set above chamber ambient temperature? 3. Is the circulation fan operational? Open the door and press in on the door switch 4. Is there voltage on the output terminal of the controller? 5. Is the thermal overload “OPEN” ? 6. Is the door(s) closed? 7. Contact your SKYTRON authorized service representative
Over temperature alarm activated
HI is
Check outlet for power Check if warmer is plugged in Check if the ON/OFF switch/circuit breaker is turned on Check the fuse on the incoming supply Check for power at the junction box in the control panel Contact your SKYTRON authorized service representative
1. 2. 3. 4.
Is the circulating fan operational? Are the contents loaded properly? Has the chamber set temperature been lowered Temperature of the lower chamber cannot be in excess of +30°F (+16 °C) above the upper chamber 5. Contact your SKYTRON authorized service representative
Overheat Alarm (HI) Condition When the chamber temperature exceeds the setpoint by 10° F (5 °C) or the set temperature is lowered by more than 10° F (5 °C), the display will read HI and the audible alarm will sound. To silence the audible alarm press the Alarm Reset button until it quits.
M WARMING CABINET OWNER’S REV0
Turn off the Warming Cabinet chamber, wait for the contents to cool adequately then unload the and reload the contents following the loading guidelines previously outlined. Turn on the chamber and monitor performance. If the chamber continues to overheat into an alarm (HI) condition turn off the chamber and contact your SKYTRON authorized service personnel.