SLE Limited

Ventilators

SLE4000 and SLE5000 Cleaning , Disinfection and Sterilisation Guide

Cleaning , Disinfection and Sterilisation Guide

2 Pages

How To...  Clean Your SLE4000/5000 Ventilator Preparation of a new ventilator Remove all transit packaging. Inspect the fresh gas port and proximal airway port for any packing material. (Retain packaging for future use). Remove the protective film from the touch screen. Clean, disinfect and sterilise in accordance with the instructions below. Remove the inlet air and O2 gas port caps. (Retain for future use). Allow the unit to acclimatise, i.e. if the ventilator has been stored in a cold or humid environment whilst in transit/ storage.  Cleaning and disinfecting an in-service ventilator  Cleaning, disinfection and sterilisation All cleaning, disinfection and sterilising should be carried out under the direction of the appropriate hospital authority. DO NOT allow moisture to enter the electronic module or its electrical sockets. Electronic malfunction may result. DO NOT steam autoclave the ventilator or otherwise subject it to temperatures above 62°C. DO NOT immerse any part of the ventilator in any liquid, with the exception of the expiratory exhalation block (SLE part No N6622). Sterilisation method  The table below outlines the areas of the ventilator which can be uniquely cleaned, disinfected and sterilised.  The silencer (N2186/01), flow sensor (N5402) and exhalation block (N6622) must be sterilised between use on patients.  Before cleaning or disinfecting the exterior of the ventilator the following tasks should be performed:  The exhalation block must be cleaned as an essential prerequisite to sterilisation.  • The mains cable should be disconnected from the mains supply. • Remove the patient circuit and bacterial filters. Discard any single use items as per appropriate hospital authority guidelines. Reusable items should be processed as per appropriate hospital authority guidelines and the manufacturers’ instructions. • Disconnect the gas supplies from the wall outlets. • Disconnect the Oxygen and Air hoses from the ventilator and cap the inlet ports. • Open the side flap. • Unlock the exhalation block by turning the clamp through 90 degrees until it is horizontal. • Gently pull away the exhalation block and silencer from the gas ports. • Separate the silencer and exhalation block. Refitting the silencer and exhalation block is the reversal of removal. Do not force the exhalation block into place.  For the silencer, flow sensor and exhalation block: Autoclave with pure dry saturated steam at: 134°C (277°F) (Allowable variation of temperature of +3°C) at 220kPa (32psi) with a minimum holding time of 3 minutes Or... 121°C (248°F) (Allowable variation of temperature of +3°C) at 96kPa (14.1psi) with a minimum holding time of 15 minutes. The silencer can be autoclaved up to 20 times. The body of the silencer should be marked after each autoclave cycle with a high temperature, waterproof, permanent marker to indicate number of sterilisation cycles completed. For the flow sensor only: Ethylene Oxide at 55°C (131°F) plus the recommended time for aeration.
File Type: PDF
File Size: 933 KB
File Name: SLE Limited - SLE4000 and SLE5000 Cleaning, Disinfection and Sterilisation Guide.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions