Smith & Nephew
0.9MM DOUBLE TROCAR TIP MINI KOMPRESSOR
Reprocessing , Cleaning , Sterilization and Inspection of Non-Sterile Family Group 3 Devices
12 Pages
Preview
Page 1
Cleaning, Sterilization and Inspection of Non-sterile Family Group 3 Devices
Table of Contents Product Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Intended Users. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Intended Use/Indications, Contraindications, and Patient Target Group. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Warnings and Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Cleaning and Sterilization Warnings. . . . . . . . . . . . . . . . . . . . . . 5 Adverse Event Reporting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Storage, Handling and Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Preparation/General Guidance for Cleaning and Disinfection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Manual Cleaning Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Automated Cleaning and Disinfection Instructions. . . . . . . . 7 Inspection and Functional Check Prior to Use. . . . . . . . . . . . . 7 Sterilization Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Glossary of Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
English
Product Description
Product Description
Inspection for Completeness
These instructions apply to all reusable medical devices that are sold by ASCENSION◊ Orthopedics, Inc. that are initially supplied non-sterile. These instructions also apply to single use medical devices that are supplied non-sterile and require processing. Non-sterile, single use medical devices (i.e., screws, pins, drills) are commonly placed in containment devices and therefore, require processing prior to use. The medical devices are intended for use in orthopedic surgical procedures according to the Instructions for Use and Surgical Techniques for the associated implant system. Use with devices other than those identified in the Instructions for Use and Surgical Techniques is not recommended.
Upon receipt in the hospital, instrument sets should be inspected for completeness. Inspect for removable screws; detachable handles; and auxiliary exchangeable parts. Many organizing cases have shadow graphs, outlines, catalog numbers, and instrument names or sizes silk-screened or otherwise marked on the case or tray.
Intended Users The Ascension Orthopedics, Inc. devices are intended to be used by orthopedic surgeons with thorough knowledge of the anatomy, physiology, and surgical principles involved. Physicians are strongly encouraged to obtain instruction from experienced clinicians or to observe surgical use of the subject devices. The use environment is a professional healthcare facility.
Intended Use/Indications, Contraindications, and Patient Target Group Intended use/indications, contraindications, and patient target group are based on the implant devices and therefore information related to indications for use and contraindications can be found in the associated implant system Instructions for Use.
EU MDR Class I devices include: • Case bases, lids and inserts that are intended for the housing of non-sterile devices during transport, reprocessing and/or storage • Non-invasive instruments, including handles, impactors, impaction stands, mallets and hammers, sizing gauges, taper disassembly tools, and torque limiters that are intended to be used in orthopedic surgical procedures.
Materials These devices do not contain phthalates, CMR (carcinogenic, mutagenic or toxic to reproduction) or endocrine disruptive substances nor tissues of biological origin above the acceptable concentration of 0.1% weight by weight (w/w) unless indicated on the label.
4
LC-04-0000-0013 Rev. NB
Orthopaedic surgical procedures follow a precise order in which the instruments are used. Also, many instruments have dimensional features which govern bone resections, determine implant sizes, and measure intramedullary canal sizes, depth of drill holes, angles of tube, etc. Therefore, it is very important that all requested sizes of a specific instrument series are available (specific instruments are routinely omitted from instrument sets due to infrequent use unless requested by the user). Contact your Ascension Orthopedics representative if requested instruments have been omitted but are required for surgery.
Warnings and Precautions U.S. Federal law restricts this device to sale by or on the order of a physician. • The devices are shipped non-sterile. Devices may require disassembly before reprocessing as indicated by the tray configuration and/or as indicated in the Instruction for Use and re-assembled prior to use as per the applicable Surgical Technique. Cleaning, disinfection and sterilization is required before each use. • It is the healthcare professional’s responsibility to be familiar with the appropriate surgical techniques prior to use of the device(s). • Read these instructions completely prior to use. The healthcare professional should have a full understanding of the product labeling information including, but not limited to, the following: Instructions for Use (IFU) for the implant, Surgical Techniques and other relevant product materials that have been provided by the manufacturer. • The patient should be warned of surgical risks and made aware of possible adverse effects. It Is recommended that the operating surgeon record all information given to the patient. • Prior to each use, inspect the device to ensure it is functioning properly and not damaged. Do not use a damaged device. Contact your Ascension Orthopedics representative for repair or replacement of damaged items. • For single use devices, after use, the device may be a potential biohazard and should be handled in accordance with accepted medical practice and applicable local and national requirements.
Cleaning, Sterilization and Inspection of Non-sterile Family Group 3 Devices
English
• The instructions provided in this reprocessing manual have been validated by ASCENSION◊ Orthopedics, Inc. and are capable of preparing orthopedic devices for use. It is the responsibility of the hospital to ensure that reprocessing is performed using the appropriate equipment and materials, and that personnel in the reprocessing facility have been adequately trained in order to achieve the desired result. Equipment and processes should be validated and routinely monitored. Any deviation by the processor from these instructions should be properly evaluated for effectiveness to avoid potential adverse consequences. • The useful life of these devices depends on many factors including the method and duration of each use, and the handling between uses. Careful inspection and functional testing of the device before use is the best method of determining the end of serviceable life for the medical device. Do not use devices that show evidence of damage and wear. Evidence of damage and wear on a device may include but is not limited to corrosion (i.e. rust, pitting), discoloration, excessive scratches, flaking, wear and cracks. Improperly functioning devices, devices with unrecognizable markings, missing or removed (buffed off) part numbers, damaged and excessively worn devices should not be used. • Personal Protective Equipment (PPE) should be worn when handling or working with contaminated or potentially contaminated devices. PPE includes: gown, mask, goggles or face shield, and shoe covers. • Universal precautions are standards of infection control practices designed to reduce the risk of transmission of bloodborne infections. Universal precautions should be observed by all hospital personnel that work with contaminated or potentially contaminated devices.
Cleaning and Sterilization Warnings
• Rinse aids can leave residues and is not recommended for use unless validated by the processor and is of proven biocompatibility. • Always observe the cleaning agent concentrations, temperatures, contact times and final rinse instructions specified by the manufacturer. • For ultrasonic cleaning, the enzymatic detergent solution should be changed before it becomes heavily soiled so that effective cleaning is not inhibited. • Do not sonicate powered devices. • It is the responsibility of the user to ensure that the sterilization process is performed using qualified equipment, materials and personnel such that the recommended parameters are achieved. • Steam sterilization is the only method that has been validated for processing by Ascension Orthopedics. Sterrad or hydrogen peroxide-based gas systems have not been validated. • These recommended procedures are intended as a general guide for sterilization of reusable medical devices sold by Ascension Orthopedics Inc. Some devices are labeled with more specific instructions. • It is the responsibility of the user to validate their sterilization equipment to ensure that the recommended parameters are achieved. Steam sterilizers should be validated to local standards, guidance and ISO 17665. • It is the responsibility of the user to ensure that the cleaning process is performed following these procedures to achieve the desired result.
Adverse Event Reporting
• Exercise caution when handling devices with sharp points or cutting edges.
If, during the use of the device or as a result of its use, a serious incident has occurred, please report it to the manufacturer and to your national authority.
Cleaning and Sterilization Warnings
Storage, Handling and Disposal
• Single-use devices must not be reused. Use of a device is defined as contact of the device with blood or bodily fluids from a patient. Reuse may increase risk of breakage, failure, patient infection, patient injury and revision surgery. Refer to the device label to identify single-use devices. • It is the responsibility of the user to ensure that the sterilization process is performed using qualified equipment, materials and personnel such that the recommended parameters are achieved. • All cleaning should be performed in a manner designed to minimize exposure to bloodborne pathogens. Manual cleaning should be done while the instrument is immersed. Cleaning, Sterilization and Inspection of Non-sterile Family Group 3 Devices
• Reusable devices that will be stored between cleaning and sterilization should be dry to prevent microbial contamination that could result from wet devices. Containment devices can be stacked for storage. Store in a dry, dust-free place. • Storage of devices after cleaning and sterilization should be in designated, limited access areas that are well ventilated and provide protection from dust, moisture, insects, vermin and temperature and humidity extremes. • Storage is event related and not time related. • Ascension Orthopedics does not recommend stacking of wrapped containment devices or rigid containers.
LC-04-0000-0013 Rev. NB
5
Preparation/General Guidance for Cleaning and Disinfection
• Reusable devices that have been inspected and have reached the end of their lifetime should be sterilized, if possible, and disposed according to standard hospital procedure and any applicable laws. Single use devices should be disposed of according to institutional procedures, after use. Devices that are unable to be sterilized before disposal should be disposed properly as a potential biohazard.
English
• To ensure proper instrument functioning, verify that all movable parts have been thoroughly cleaned. • Pay special attention to slots, ratchets, joints and box locks, narrow lumens, blind holes, and other areas that are hard to access. • Deionized or distilled water should be used for the final rinse.
• No hazardous substances are contained or come in contact with the device, components, or raw materials. Therefore, the risk of environmental impact upon proper disposal of the subject devices is considered negligible. The subject devices have not been evaluated for long term environmental impact.
Note: Following cleaning, the instruments must be macroscopically clean, i.e. free from visible dirt or deposits. All movable parts, working tips and blades should be inspected with particular care. After cleaning, visually inspect devices under normal lighting for the removal of visible debris. Repeat cleaning if not visibly clean and reinspect the device.
Preparation/General Guidance for Cleaning and Disinfection
Manual Cleaning Instructions
Verify that all instruments required for use are present in the case. For Manual Cleaning, devices should be grouped according to similar metals before subsequent processing in order to prevent galvanic corrosion. In addition, it is not recommended to use chloride containing cleaning solutions since its use has been linked to corrosion of metallic instruments, especially stainless steel. Please also note the following: • Disinfect and clean devices immediately after use in order to avoid device encrustations. • Solutions used for cleaning must always be prepared in accordance with the manufacturer’s instructions. • Never use metal brushes or metal sponges for manual cleaning. • Use a suitably sized non-metallic bristle brush for cleaning lumens, cannulations, blind holes, and cavities, making sure that every part of the inner surface can be properly accessed. • Clean jointed instruments in closed as well as open positions. • Disassemble instruments as far as possible before cleaning. • Be sure to arrange the items so that the water can easily flow out of cannulations, blind holes, and cavities. • For instruments with long or narrow lumens, standard processing should be used only if the disinfectant can flow easily through the lumens and safe rinsing is guaranteed. • The cases/trays used for cleaning must always be loaded correctly to ensure proper cleaning. • After cleaning, check instruments for cleanliness (visible dirt). This especially applies to cannulated instruments or those with blind holes and crevices. 6
LC-04-0000-0013 Rev. NB
The following steps should be completed in sequence. Please note that all instructions provided are validated. Depending upon the detergent selection, minimum processing times and temperature settings may need to be adjusted for optimal processing: 1. Prepare a neutral pH enzymatic detergent as per the manufacturer’s recommendation (e.g. Enzol® prepared at 1 oz. per gallon of lukewarm deionized water). 2. Disassemble instruments to lowest level. 3. Rinse instruments under lukewarm running water to remove all gross soil. Use a soft bristled brush to aid in the brushing. Agitate the instruments under the running water. Agitation includes actuating all movable parts such as opening and closing hinges and moving the instruments around under the running water. Use a clean soft bristled brush and/or pipe cleaner to brush and aid in the rinse for the exterior and interior of instruments. Use a syringe to flush any lumens. 4. Fully immerse each device in the prepared detergent and allow it to soak for a minimum of two minutes. 5. After soaking the devices, scrub them using a soft bristle brush and circular strokes to remove any visible soil. Pay particular attention to all the areas where the soil could be imbedded (i.e. grooves, crevices, lumens, blind holes). Use a syringe to flush lumens and a pipe cleaner to clean lumens and holes. Perform cleaning under the surface of the prepared detergent solution to limit aerosolization of the cleaning fluid and soil, as well as for worker and environmental safety. 6. Rinse devices in lukewarm water for a minimum of one and a half (1.5) minutes to remove any detergent residuals. In accordance with Step 3, agitate the instruments under the running water, being sure to actuate all movable parts, and using a soft bristled brush for internal and exterior device surfaces. Cleaning, Sterilization and Inspection of Non-sterile Family Group 3 Devices
English
Automated Cleaning and Disinfection Instructions
7. Prepare a neutral pH enzymatic detergent (eg. Enzol) in a sonicator, as per the manufacturer’s recommendation using lukewarm water. Fully immerse the devices in the detergent and sonicate for 10 minutes. 8. After sonication, rinse the devices with running lukewarm water (use the highest grade of water available, distilled or deionized water is recommended) for three (3) minutes. Agitate the instruments under the running water, being sure to actuate all movable parts, and using a clean soft bristled brush for internal and exterior device surfaces, and flush all lumens with a syringe. 9. Dry the devices using a clean lint free cloth and visually examine to determine if all adherent visible soil has been removed. 10. Repeat the above cleaning procedure, if visible debris is detected.
Automated Cleaning and Disinfection Instructions The following steps should be completed in sequence. Please note that all instructions provided have been validated. The disinfection validation protocols were performed in accordance with ISO 15883-1: 2009 Washer-disinfectors, Part 1; General requirements, terms and definitions and tests and ISO 15883-2:2006 Washer-disinfectors - Part 2: Requirements and tests for washer-disinfectors employing thermal disinfection for surgical instruments, anaesthetic equipment, bowls, dishes, receivers, utensils, glassware, etc. Note: steps “1-6” listed below are manual pre-clean steps which need to be followed prior to step “7” 1. Prepare an enzymatic detergent using lukewarm deionized water as per the manufacturer’s recommendation. 2. Fully immerse the devices and allow to soak for a minimum of two (2) minutes. 3. Following the soak time, flush any lumens of the device using a syringe. 4. Rinse the devices under lukewarm running deionized water for a minimum of one (1) minute, while agitating the devices. Agitation includes actuating all movable parts, such as opening and closing hinges and moving the devices around under the running water. 5. Use a clean soft bristled brush and/or pipe cleaner to brush and aid in the rinse for the exterior and interior of device components. Use a syringe to flush any lumens. 6. Place the devices back into the designated locations of the case/tray, and load the case/tray set into an automated washer disinfector (Steris 444 or equivalent). 7. The washer cycle parameters are as follows: Cleaning, Sterilization and Inspection of Non-sterile Family Group 3 Devices
Recirculation Time (Min.) Pre-Wash 02:00
Water Detergent Temperature Cold Water NA
Enzyme Wash
01:00
Hot Water
Wash
02:00
60°C
Rinse
10:00
Hot Water*
Phase
Enzymatic Cleaner Neutral Detergent NA
The washer disinfector cycle parameters are as follows**: Recirculation Time (Min.) Pre-Wash 02:00
Water Detergent Temperature Cold Water NA
Enzyme Wash
01:00
Hot Water
Enzymatic Cleaner
Wash
02:00
60°C
Neutral Detergent
Thermal Rinse
10:00
82°C*
NA
Phase
*Note: The highest grade of water available should be used during the final rinse cycle, distilled or deionized is recommended. **Note: The disinfection parameters provided have been validated for the following systems: Freedom, TSS, TSS 2.5, RSS, and RSS-S. 8. After washing and disinfecting, dry the devices using a clean lint free cloth and visually examine to determine if all adherent visible soil has been removed. 9. Repeat the cleaning and disinfection procedure if visible debris is detected.
Inspection and Functional Check Prior to Use Following reprocessing and prior to use in surgery, all devices should be inspected to ensure proper function. Markings on instruments used for measuring anatomical dimensions must be legible. These may include gauge markings, angles, inner or outer diameters, length or depth markings, and right/left indications. Notify your ASCENSION◊ Orthopedics representative if scales and other markings are not legible. Additional details regarding inspection and functional check specific to device type or features are provided below: • All reusable devices: Visually inspect for damage or wear, including components in their disassembled state prior to re-assembly. Ensure components are re-assembled securely. • Hinged devices: Check for smooth movement of the hinge without excessive “play.” LC-04-0000-0013 Rev. NB
7
English
Sterilization Instructions
• Locking mechanisms: Check for action. • Cutting features: Check edges for distortion and/or large nicks. Edges should be continuous. • Trials: Articular surfaces should be smooth and free of cracks and deep nicks. • Mating parts: Check to make sure that mating parts fit together without complications. Ensure components are re-assembled securely. • Reamer/drill bits: Inspect “chuck” end for burrs and distortion that might hinder insertion into a drill. • Hammering surfaces: Inspect for burrs and large nicks. • Driving devices: Inspect plastic ends for cracks and large nicks. • Metal surfaces: Inspect for corrosion and major deformation.
Sterilization Instructions Sterilization validations were completed for our instrument sets. The validation protocols were performed in accordance with ANSI/AAMI ST79:2017 Steam Sterilization and Sterility Assurance in Health Care, ANSI/AAMI ST77: 2013/ (R)2018 Containment Devices for Reusable Medical Device Sterilization, and ISO 17665-1:2006 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices. All testing was done using the overkill approach with Geobacillus stearothermophilus spores. The results confirmed 106 Sterility Assurance Level (SAL) for the sets when using the recommended cycles. In accordance with our validation results, the following cycles are recommended for wrapped goods utilizing FDA-cleared sterilization wrap:
• Powered devices: Verify that power is supplied when the device is turned on and ceases when the device is turned off. • Measuring devices (i.e. gauges, calipers): Check for legible measuring markings.
8
LC-04-0000-0013 Rev. NB
Cleaning, Sterilization and Inspection of Non-sterile Family Group 3 Devices
English
Sterilization Instructions
Recommended Sterilization Parameters Device Set System
Pre-vacuum
Gravity
Pre-vacuum2
Pre-vacuum2
132°C/4 Min.
132°C/15 Min. 134°C/3 Min. 134°C/18 Min. Dry Time Requirements (minutes)
ConstructTM Ankle Fusion Plating System1 CMC®/NUGRIP◊ (Metal tray only) FIRST CHOICE◊ DRUJ (PUH & MUH1) TITAN HRA System* Integra® Freedom Wrist
40
40
30
30
30
30
30
30
MOVEMENT Great Toe System
30
35
30
30
Proximal Humeral Fracture Plate System
30
30
30
30
PYROTITAN HRA
30
30
30
30
30
30
30
30
30
30
30
30
TITAN Modular Shoulder System Glenoid Tray
30
35
30
30
TITAN Modular Shoulder System Humeral Tray I
30
40
30
30
TITAN Modular Shoulder System Humeral Tray II
30
40
30
30
◊
1 1
◊
TITAN Coated Retractors TITAN FIN-LOCK Glenoid Tray ◊
1
TITAN Retractors
30
40
30
30
TITAN Reverse Shoulder System
30
30
30
30
TITAN TSS Long Stem1/Revision Instruments
30
45
30
30
Titanium Bone Wedge 1
30
30
30
30
Total Foot System 2 TITAN Modular Shoulder System, 2.5 Glenoid Tray TITAN Modular Shoulder System, 2.5 Humeral Tray I TITAN Modular Shoulder System, 2.5 Humeral Tray II TITAN Reverse Shoulder System - S TITAN Modular Shoulder System, 2.5 Monoblock Sounder Instrumentation PANTA◊ 2 Nail Arthrodesis System, Targeting and Compression Tray 1 PANTA 2 Nail Arthrodesis System, Nail Prep Tray 1
60
60
65
65
40
55
35
35
40
50
35
35
40
60
35
35
30
30
30
30
30
30
30
30
30
30
30
30
30
30
30
30
30
30
30
30
1
PANTA 2 Nail Arthrodesis System, Screw Case
1
* Dry time data for the shared cells is based upon the highlighted sterilization challenge set for the device set family grouping (#3). 1
This system does not have a registered CE mark.
2
134°C setting is for use Outside the USA only.
Cleaning, Sterilization and Inspection of Non-sterile Family Group 3 Devices
LC-04-0000-0013 Rev. NB
9
English
Glossary of Symbols
Glossary of Symbols Manufacturer
CE mark
Medical Device
Date of manufacture
Non-sterile
Authorized representative in the European community
Batch code
Do not re-use
Catalog number
Consult instructions for use
Quantity
Prescription use only
Ascension Orthopedics, Inc. 11101 Metric Blvd Austin, TX 78758 USA 12
Integra LifeSciences Services Integra LifeSciences Corporation Immeuble Séquoïa 2 97 allée Alexandre Borodine Parc Technologique de la Porte des Alpes 69800 Saint Priest – France Phone: +33 (0) 4 37 47 59 10
LC-04-0000-0013 Rev. NB
Stainless steel
MATL SS
Made in the United States of America
Trademark of Smith+Nephew. All trademarks acknowledged.
◊
www.smith-nephew.com ifu.smith-nephew.com +1 800 654 2873 Telephone +1 800 980 7742 Fax
©2021 Smith & Nephew, Inc. All rights reserved. 2021-12
Cleaning, Sterilization and Inspection of Non-sterile Family Group 3 Devices