RENASYS TOUCH – Quick Reference (refer to the User Manual for full details on RENASYS TOUCH)
1. Switching the device between Continuous and Intermittent therapy - To switch between Continuous and Intermittent therapy, ensure therapy is paused and press the therapy mode toggle on the home screen
Therapy mode toggle
Continuous therapy home screen
Intermittent therapy home screen
2. Starting therapy (Continuous) Decrease value
Therapy set point
Increase value
Step 1: Ensure that the device is powered on and the Continuous therapy home screen is displayed Status indicator
Step 2: Use the + or – buttons to select prescribed pressure setting Step 3: Press the Start/Pause therapy button to start delivering therapy When therapy is active the status indicator on top of the device illuminates green and the therapy indicator at the top of the screen will rotate orange
Therapy mode toggle
Start/Pause therapy
3. Starting therapy (Intermittent) Step 1: Ensure that the device is powered on and the Intermittent therapy home screen is displayed Step 2: Change the High and Low pressure settings and cycle times by selecting the setting and use the + or – to the prescribed pressure/ time Step 3: Press the Start/Pause therapy button to start delivering therapy When therapy is active the status indicator on top of the device illuminates green and the therapy indicator at the top of the screen will rotate orange
All 4 settings can be adjusted within the following pressure ranges: High Therapy 25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180, 200mmHg High Cycle times: 3, 5, 8, 10 minutes Low Therapy 0, 25, 40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180mmHg Low Cycle times: 2, 3, 5, 8, 10 minutes
Status indicator
4. Checking for a seal Follow on screen seal checking instructions. Once seal is achieved the screen will display “seal achieved” and then display “delivering therapy” Look for “raisin-like” appearance Listen for a “hissing” sound, indicating a leak Feel the dressing, which should be hard to the touch Possible site of leak • Wrinkle or crease • Skin fold or crevice • Where drain exits film (where applicable) Intervention Seal film edges with film dressing If hole found in Transparent Film, patch with adhesive Transparent Film.
5. Locking the screen The Lock function locks the touchscreen and Start/Pause therapy icon – Power icon is not locked. The status indicator continues to illuminate green to indicate the device is still delivering therapy Status indicator
When delivering therapy, the screen will go dark. To view screen, press either Play or Lock pad button. In the event of an alarm, the device will automatically unlock and the Alarm screen will display. The device will automatically lock and enter Sleep mode after 5 minutes without user interaction.
Lock/Unlock
6. Switching the device between Patient and Clinician mode Step 1: Press the yellow settings button to access the settings screen
Step 2: Press the change mode option from the menu
Step 4: Please enter the password (3141) and press the green tick to switch modes
Step 5: Press yellow button on screen twice to return to the home screen
Supporting healthcare professionals for over 150 years Smith & Nephew, Inc. Fort Worth, TX USA
www.smith-nephew.com www.renasys-npwt.com
Customer Care Center 1-800-876-1261 T 727-392-1261 F 727-392-6914
©2016 Smith & Nephew, Inc. All rights reserved. ™Trademark of Smith & Nephew.
Step 3: Select which mode (Clinician/Patient) required
QUESTIONS? CALL NPWT CLINICAL AND PRODUCT SUPPORT LINE, VIA CUSTOMER CARE AT 1-800-876-1261 For detailed product information, including the indications for use, contraindications, precautions and warnings, please consult the product’s Instructions for Use (IFU) prior to use. RNPE-41-0816-UE