Quick set-up guide
Remove handset from carton
Open external pouch
Prepare internal pouch for transfer
Transfer inner pouch to sterile field
c a b
Remove device contents
Prepare device. Transfer saline tube, evacuation tube and pump from sterile field
a. Handset pump cartridge b. Saline spike line c. Evacuation line
Remove pump cartridge from clam shell
Turn right to 3 o’clock
Remove saline tube spike cap
Safely hold handset
a Turn console power ON
Insert pump cartridge into console
b
c
Spike saline (3L) bag. Ensure saline reaches handset before engaging foot pedal
Attach evacuation tube to vented canister
Increase power setting to 10 for priming
Depress foot pedal to prime
Reduce power to desired procedure setting
Ready
VERSAJET™ II Hydrosurgery System troubleshooting tips
For live tech support, Mon-Fri 8-5pm EST: 1-800-343-5717 option 2
Pump cartridge difficult to remove
First option, turn console off. Try to remove
Second option, turn to unlocked position and gently push cartridge into console a little further. Quickly remove hand allowing cartridge to spring back before trying to remove
Third option, clamp saline line and remove spike from bag. Hold line and adjust power setting to 10, then open the line. Press foot pedal until water jet stops streaming in operating window. Try to remove
Failure conditions
Press and hold up/down keys simultaneously. F1 means power supply failure has occurred. Replace console
To replace console, email [email protected] and [email protected] the following items: 1) Completed VERSAJET II complaint form 2) No charge PO# 3) Console serial number
F2 means an internal motor failure has occurred. Replace console
F3 means an over-heated failure has occurred. Allow console to cool 10 minutes leaving bottom and back ventilation clear
A return authorization number (RA#) and a decontamination form will be sent to you. Upon receiving your replacement or loaner console, please use its shipper box to return the inoperable console. Please attach the completed decontamination form onto the shipper box and send to: Smith & Nephew 76 S. Meridian Ave. Dock #1/Complaint Oklahoma City, OK 73107
Handset doesn’t prime
Verify pinch clamps are fully open
Check for kinks, obstructions or leaks in the inflow line
Set power level to 10. Press foot switch and pinch the clamp. Drag it down to purge all air from the feed line
User tips
Keep handset moving tangentially at swift and steady pace to prevent grooves in the woundbed
Invert handset away from wound when procedure is finished, then remove foot from footswitch to prevent residual saline from trickling back into the clean wound bed
Smith & Nephew, Inc. USA 970 Lake Carillon Drive Suite 110 St. Petersburg, FL 33716
Customer Care Center 1-800-876-1261 T 727-392-1261 F 727-392-6914
Canada 2280 Argentia Road Mississauga Ontario L5N 6H8
Customer Action Centre T 1-800-463-7439 F 1-800-671-9140
www.smith-nephew.com www.myversajet.com
©2014 Smith & Nephew, Inc., All rights reserved. ™Trademark of Smith & Nephew. Certain trademarks registered in the US Patent & Trademark Office. RA 700 VJPE-04-0713-NAE