REAL INTELLIGENCE CORI for Knee Arthoplasty Addendum rev A

Addendum

15 Pages

REAL INTELLIGENCE™ CORI™ ADDENDUM CAMERA VISIBILITY NOTES Customers may experience intermittent visibility of the REAL INTELLIGENCE tracker arrays as a result of normal system behavior for maintaining tracker accuracy. If intermittent visibility is occurring more frequently than expected, please see the following troubleshooting tips and notes to achieve more consistent visibility of the tracker arrays: 1. Angle the tracker array slightly up or down away from the direct line of sight from the camera. • Note: Moving the tracker array should be completed within the Camera Orientation Adjustment screen, ideally prior to collecting any registration points. Alternatively, if unable to move the tracker arrays, you may move the leg to angle the tracker array. 2. Position the camera closer to or farther away from the tracker arrays. • Note: On the Camera Orientation Adjustment screen (with camera view also accessible by pressing the tracker icons in the upper righthand corner), the green range represents distances of 1.2 – 2.0 m. It is recommended to position the camera either between 1.2 – 1.4 m or greater than 1.6 m from the tracker arrays. • In positioning of the camera, please be mindful of the sterile field and boundary in the operating room. • Refer to the REAL INTELLIGENCE◊ CORI◊ for Knee Arthroplasty User Manual: “Preparing for the Intraoperative Knee Procedure: Adjusting the Tracking Camera” section for a recommended positioning of tracker arrays within the green range.  3. If steps 1 and 2 do not improve the intermittent visibility, replace the flat markers on the tracker array with intermittent visibility. After replacing flat markers, be sure to re-verify the checkpoints.  Blue Belt Technologies, Inc. 2875 Railroad Street Pittsburgh, PA 15222 USA  0123  www.smith-nephew.com ifu.smith-nephew.com Smith & Nephew Operations B.V. Bloemlaan 2, 2132 NP Hoofddorp Netherlands  +1 763 452 4910 Telephone +1 763 452 4980 Fax CustomerSupport.Robotics@smith-nephew.com  ™Trademark of Smith & Nephew, registered U.S. Patent & Trademark Office. This product may be covered by one or more U.S. patents. See smith-nephew.com/patents for details. ©2024 Smith & Nephew, Inc. All rights reserved. 4/2024  1000264245 REVA
File Type: PDF
File Size: 2.75 MB
File Name: Smith & Nephew - 1000264245 - REAL INTELLIGENCE CORI for Knee Arthoplasty Addendum rev A - 2024-04.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions