CADD®-Solis VIP Ambulatory Infusion System Alarm Help Screens
When certain alarms occur, additional information may be displayed on the screen. The help screens describe what you can do to try to solve the problem that is causing the alarm.
Stopped
2,200 mL Taper Adult TPN
1:15 PM mL m
Infusion Volume
STEPS:
2,000 mL
Taper Up
1. When an alarm occurs, press SILENCE to quiet the alarm.
1 Hrs 0 Min
Taper Down
1 Hrs 0 Min
Infusion Duration
12 Hrs 0 Min
2. If help screens are available for the alarm, “Help” appears above the right soft key. To view the help screens, select HELP.
'Accept Value', or press 'select' to edit
Exit
Accept Value
3. Follow the applicable instructions provided on the help screen. To page through all of the available help screens, press NEXT repeatedly. Press ACKNOWLEDGE at any time to exit Help. 4. “Retry Help” appears when no additional help steps are available. To page through the help screens again, press RETRY HELP. The alarm screen reappears as in Step 2 above. Repeat steps 2 and 3. When the alarm clears, the help screen will no longer be displayed.
The Following Alarms Have Associated Help Screens. Refer to Steps Above to Navigate the Help Screens. CANNOT START PUMP WITHOUT A LATCHED CASSETTE. Alarm
Alarm
Cannot start pump without a latched cassette.
Acknowledge
Help
Cannot start pump without a latched cassette.
Acknowledge
Silence
Alarm - Cassette Not Latched Latch cassette.
Acknowledge
Next
Alarm - Cassette Not Latched Call your home care provider.
Acknowledge
Retry Help
UPSTREAM OCCLUSION. CLEAR OCCLUSION BETWEEN PUMP AND RESERVOIR. Alarm
Alarm
Upstream occlusion. Clear occlusion between pump and reservoir. Stop Pump
Help
Alarm - Upstream Occlusion Reposition bag in pouch
Stop Pump
Next
Upstream occlusion. Clear occlusion between pump and reservoir. Stop Pump
Silence
Alarm - Upstream Occlusion Call your home care provider.
Stop Pump
Retry Help
Alarm - Upstream Occlusion Fully insert spike.
Stop Pump
Next
Alarm - Upstream Occlusion Unkink tubing.
Stop Pump
Next
CADD®-Solis VIP Ambulatory Infusion System Alarm Help Screens
The Following Alarms Have Associated Help Screens. Refer to Steps Above to Navigate the Help Screens. DOWNSTREAM OCCLUSION. CLEAR OCCLUSION BETWEEN PUMP AND PATIENT. Alarm
Alarm
Downstream occlusion. Clear occlusion between pump and patient.
Downstream occlusion. Clear occlusion between pump and patient.
Alarm - Downstream Occlusion Adjust current position.
Silence
Help
Alarm - Downstream Occlusion Open all clamps.
Alarm - Downstream Occlusion Remove any cap.
Alarm - Downstream Occlusion Call your home care provider.
Next
Next
Retry Help
Stop Pump
Stop Pump
Alarm - Downstream Occlusion Unkink tubing.
Next
Next
CANNOT START PUMP WITHOUT A LATCHED AND LOCKED CASSETTE. (PCA ONLY) Alarm
Alarm
Cannot start pump without a latched and locked cassette.
Acknowledge
Silence
Cannot start pump without a latched and locked cassette.
Acknowledge
Alarm - Cassette Not Latched Latch cassette.
Acknowledge
Next
Alarm - Cassette Not Latched Call your home care provider.
Acknowledge
Retry Help
LOCK CASSETTE TO START PUMP. (PCA ONLY) Alarm
Alarm
Lock cassette to start pump.
Acknowledge
Silence
Lock cassette to start pump.
Acknowledge
Help
Alarm - Cassette Not Locked Lock cassette.
Acknowledge
PRODUCT(S) DESCRIBED MAY NOT BE LICENSED OR AVAILABLE FOR SALE IN CANADA AND OTHER COUNTRIES
Next
Alarm - Cassette Not Locked Call your home care provider.
Acknowledge
Retry Help
MPAUCA-1617
Smiths Medical ASD, Inc. 6000 Nathan Lane North Minneapolis, MN, 55442, USA Tel: 1-614-210-7300 Toll-Free USA 1-800-258-5361 www.smiths-medical.com Find your local contact information at: www.smiths-medical.com/customer-support Smiths Medical is part of the global technology business Smiths Group plc. Please see the Instructions for Use/Operator’s Manual for a complete listing of the indications, contraindications, warnings and precautions. CADD and the Smiths Medical design mark are trademarks of Smiths Medical. The symbol ® indicates the trademark is registered in the U.S. Patent and Trademark Office and certain other countries. All other names and marks mentioned are the trademarks or service marks of their respective owners. ©2017 Smiths Medical. All rights reserved. IN193583EN-082017