User Guide Supplement
44 Pages
Preview
Page 1
S Series Ultrasound System User Guide Errata P12397-02 Patient report If you transfer or export an exam from the S Series™ ultrasound system, the following measurements do not appear in the exported patient report. If you need to record them, save the report as an image (press Save) before transferring or exporting, or write them manually into the patient record. • GYN exam: Follicle • Cardiac exam: Automatically calculated CO values (CO, SV, CI, SI, VTI, Vmax, Vmean, PGmax, PGmean, LVOT D, LVOT area)
Language selection If you need to change the Language selection on the Presets setup page, make sure to first archive, transfer, or export all saved patient information you want to keep. The system may not correctly maintain the saved patient information after the Language selection is changed.
Video Mode selection If you change the Video Mode selection (NTSC or PAL) on the Connectivity setup page, restart the system. Changing the Video Mode selection can disable the USB ports until the system is restarted.
Acoustic output tables The following tables provide updated acoustic output information.
Table 1: Transducer Model: HFL38x
Operating Mode: PW Doppler TIS
Index Label
Associated Acoustic Parameter
Global Maximum Index Value pr.3
Other Information
1.0
(MPa)
Non-scan Scan
Aaprt≤1
Aaprt>1
-
1.2
-
-
46.55
(mW)
min of [W.3(z1),ITA.3(z1)]
(mW)
-
z1
(cm)
-
zbp
(cm)
-
zsp
(cm)
deq(zsp)
(cm)
fc
(MHz)
PD PRF pr@PIImax
X (cm) Y (cm) (μsec) (Hz) (MPa)
deq@Pllmax
(cm)
Focal Length IPA.3@MImax
Non-scan
TIC
2.2
(b)
46.55
#
2.37
W0
Dim of Aaprt
Operating Control Conditions
MI
TIB
0.9
1.1 0.33
5.32
-
5.33
-
5.33
#
- -
1.04
-
1.04
#
0.4
-
0.4
#
1.29 1008 2.404 0.46
FLx (cm)
-
3.72
-
#
FLy (cm)
-
2.5
-
#
(W/cm ) 2
323.35
Control 1: Exam Type
Bre/Vas SmP
Vas/Ven
Vas/Ven
Control 2: Sample Volume Control 3: PRF Control 4: Sample Volume Position
1 mm
12 mm
12 mm
1008
10417
10417
Zone 2
Zone 7
Zone 7
(a) This index is not required for this operating mode; value is <1. (b) This transducer is not intended for transcranial or neonatal cephalic uses. # No data are reported for this operating condition since the global maximum index value is not reported for the reason listed. (Reference Global Maximum Index Value line.) - Data are not applicable for this transducer/mode.
2
Table 2: Transducer Model: L38xi/10-5
Operating Mode: M Mode TIS
Index Label
Associated Acoustic Parameter
Global Maximum Index Value pr.3
Other Information
1.5
(MPa)
Non-scan Scan
Aaprt≤1
Aaprt>1
-
(a)
-
-
#
(mW)
min of [W.3(z1),ITA.3(z1)]
(mW)
-
z1
(cm)
-
zbp
(cm)
-
zsp
(cm)
deq(zsp)
(cm)
fc
(MHz)
PD PRF pr@PIImax
X (cm) Y (cm) (μsec) (Hz) (MPa)
deq@Pllmax
(cm)
Focal Length IPA.3@MImax Control 1: Exam Type Control 2: Optimization Control 3: Depth
Non-scan
TIC
1.2
(b)
37.1
#
3.54
W0
Dim of Aaprt
Operating Control Conditions
MI
TIB
1.0
0.9 0.49
5.76
-
#
-
5.20
#
-
#
-
1.86
#
-
#
-
0.40
#
0.146 1600 4.32 0.49
FLx (cm)
-
#
-
#
FLy (cm)
-
#
-
#
(W/cm ) 2
776 Any
Any
Gen
Pen
4.7 cm
7.3 - 9.0 cm
(a) This index is not required for this operating mode; value is <1. (b) This transducer is not intended for transcranial or neonatal cephalic uses. # No data are reported for this operating condition since the global maximum index value is not reported for the reason listed. (Reference Global Maximum Index Value line.) - Data are not applicable for this transducer/mode.
3
SonoSite, Inc.
SonoSite Ltd
21919 30th Drive SE Bothell, WA 98021 USA T: 1-888-482-9449 or 1-425-951-1200 F: 1-425-951-1201
Alexander House 40A Wilbury Way Hitchin, Herts SG4 OAP UK T: +44-1462-444800 F: +44-1462-444801
07/2010 Copyright 2010 by SonoSite, Inc. All rights reserved.
0086
P12397-02
*P12397-02* 44