SonoSite
T8-3 Transducer Care
22 Pages
Preview
Page 1
T8-3 Transducer Care WARNINGS Be sure to use the appropriate Personal Protective Equipment (PPE) whenever necessary. DO NOT skip any steps or abbreviate the cleaning process in any way. Cautions DO NOT allow bodily fluids to dry on the transducer. Always clean the transducer immediately after use. DO NOT bend the endoscopic shaft smaller than a 20 cm (8 inch) curve. DO NOT drop the transducer. If it is dropped, perform an electrical safety test and temperature calibration prior to use. DO NOT use unapproved cleaning agents because they can damage the transducer and void the warranty. For more information about approved cleaners and disinfectants, see www.sonosite.com/support/cleaners-disinfectants.
DO NOT spray or immerse the controller, the cable, or the connector with any fluid. Be careful to prevent moisture from getting on the connector.
DO NOT soak the transducer longer than recommended by the chemical manufacturer. Follow the manufacturer’s recommendations for concentration, temperature, and duration.
Clean and disinfect transducer prior to use if it is not housed in a protective covering.
Refer to the following diagram when caring for your transducer Cannot be submerged
Connector
Cable
Controller
Can be submerged 90 cm
Endoscopic shaft Scan head
Preventing damage
Pre-cleaning
STEP 1 ALWAYS
STEP 2 DO NOT allow biological material to dry on the
use a bite guard. Remove any dentures before placing the bite guard in the patient’s mouth.
Cleaning STEP 3 Rinse the
endoscopic shaft and 1 minute scan head with clean, room temperature running water for a minimum of 1 minute. DO NOT rinse the cable or controller.
STEP 4 Soak up to
90 cm of the shaft and scan head in an approved cleaner. Make sure the controller cannot fall into the solution. CHECK the expiration date of the cleaner, and follow ALL chemical manufacturer’s instructions for soak time and temperature.
transducer. Immediately after use: a Use a clean cloth moistened with water to wipe the cable and controller. DO NOT wipe the connector. b Use a second clean cloth moistened with water to wipe the shaft and scan head. INSPECT the transducer for any sign of damage. DO NOT use the transducer if any damage is found.
STEP 5 During the soak: a Wipe cable and controller with a wipe approved for use on the cable and controller. DO NOT wipe the connector. CHECK the expiration date of the wipe. b Wipe shaft and scan head using a clean, soft cloth or sponge moistened with cleaner.
STEP 6 Rinse the
shaft and scan head with 3 minutes room- temperature running water for a minimum of 3 minutes. DO NOT rinse the cable or controller. INSPECT for any remaining biological material. Repeat cleaning if any is found.
STEP 7 Dry the
transducer with a clean, non-linting towel. INSPECT for any sign of damage. DO NOT use the transducer if any damage is found.
For a list of approved cleaners and disinfectants, refer to the cleaners and disinfection tool available at www.sonosite.com/support/cleaners-disinfectants. 1
Testing for electrical leakage STEP 8 Set up the equipment: a Immerse the endoscopic shaft up to the 90 cm mark. Secure the controller so it cannot fall into the test solution.
(b)
b Insert the transducer connector into the transducer adapter.
For complete instructions on using the leakage tester, refer to the ULT2000 Series User Manual. (c)
(d)
STEP 9 Run the leakage test: a Power on the ULT-2020 leakage tester. b Press MODE repeatedly until Device Configuration appears.
(a) (e)
c Confirm that the leakage tester is set to the FUJIFILM Sonosite T8-3 transducer. If it isn’t, change the setting as required.
c Connect the transducer adapter to the leakage tester.
d Confirm that the endoscopic shaft, the scan head, and the conductivity probe contacts are all fully submerged in the leakage test fluid.
d Connect the conductivity probe to the leakage tester. e Immerse the conductivity probe contacts into an approved leakage testing fluid. DO NOT allow the probe to touch the transducer.
e Press FULL TEST to start the test. Ensure the shaft, scan head, and probe contacts remain submerged during testing. If the result is Fail, DO NOT use the transducer. Contact FUJIFILM Sonosite for repair.
Disinfecting STEP 10 Disinfect the transducer:
a Wipe the controller and cable with either a Sani-Cloth or a clean, nonlinting cloth moistened with disinfectant approved for use on cable and controller. DO NOT wipe the connector. CHECK the expiration date of the wipe, and follow ALL manufacturer’s instructions for contact time.
b If in accordance with facility policies, remove disinfectant after specified soak time by wiping with a clean, non-linting cloth or sponge moistened with water. c Soak up to 90 cm of the endoscopic shaft and scan head in disinfectant approved for immersing the shaft. Make sure the controller cannot fall into the solution. CHECK the expiration date of the disinfectant, and follow ALL chemical manufacturer’s instructions for soak time and temperature.
STEP 11 Rinse the shaft and scan head
with room-temperature running water for a minimum of 3 minutes. DO NOT rinse the cable or controller.
3 minutes
STEP 12 Dry the transducer with a clean, non-linting cloth or medical-grade air. INSPECT the transducer for any sign of damage. DO NOT use the transducer if any damage is found.
Transporting and storing STEP 13 Transport and store the
For complete instructions on cleaning, disinfecting, transporting, and storing the T8-3 transducer, refer to the T8-3 Transducer (P29465) User Guide.
transducer in accordance with your organization’s policies.
SONOSITE is a registered trademark of FUJIFILM SonoSite, Inc. in various jurisdictions. FUJIFILM is a registered trademark of FUJIFILM Corporation in various jurisdictions. Value from Innovation is a trademark of FUJIFILM Holdings America Corporation. All other trademarks are the property of their respective owners. Copyright © 2021 FUJIFILM SonoSite, Inc. All rights reserved.