TEE Transducer Care WARNINGS Cautions
Be sure to use the appropriate Personal Protective Equipment (PPE) whenever necessary. DO NOT skip any steps or abbreviate the cleaning process in any way. DO NOT allow bodily fluids to dry on the transducer. Always clean the transducer immediately after use. DO NOT bend the endoscopic shaft smaller than a 20 cm (8 inch) curve. DO NOT use unapproved cleaning agents such as alcohol or bleach (for example: Sani-Cloth) as these can damage the
transducer and void the warranty. For more information about approved cleaners and disinfectants, see www.sonosite.com/ support/cleaners-disinfectants.
DO NOT spray or immerse the controls, the cord, or the connector with any fluid. Be careful to prevent moisture from getting on the connector.
DO NOT soak the transducer longer than recommended by the chemical manufacturer. Follow the manufacturer’s recommendations for concentration, temperature, and duration.
Refer to the following diagram when caring for your transducer Cannot be submerged
Connector
Cable
Controller
Can be submerged 90 cm
Endoscopic shaft Scan head
Preventing damage STEP 1 ALWAYS use a bite guard. Remove any dentures before placing the bite guard in the patient’s mouth.
Cleaning
For a list of approved cleaners and disinfectants, refer to the cleaners and disinfection tool available at www.sonosite.com/support/cleaners-disinfectants.
STEP 2 DO NOT
STEP 4 During the
STEP 5 Rinse the shaft for at least 3 min in a large volume of clean, lukewarm water (for example, 8 L). Inspect the transducer for any remaining biological material.
STEP 3 Place up to
B Wipe the controller and cable with a clean, non-linting cloth or single-use endoscopic sponge moistened with cleaner. Remove cleaner by wiping again with a fresh cloth or sponge moistened with clean water.
STEP 6 Dry the
allow bodily fluids to dry on the transducer. Remove all visible biological material by wiping the entire transducer, except the connector, with a cloth or wipe moistened with water.
90 cm of the endoscopic shaft in an approved cleaner. Make sure the controller cannot fall into the solution. CHECK the expiration date of the cleaner, and follow ALL chemical manufacturer’s instructions for soak time and temperature.
soak: A Scrub the shaft for at least 3 min using a soft brush or single-use endoscopic sponge moistened with cleaner.
transducer with a clean, non-linting towel. Examine the transducer and cable for any cracks. DO NOT use the transducer if any damage is found.
1
Testing for electrical leakage STEP 7 Set up the
equipment. A Insert the transducer into the transducer adapter.
A
For complete instructions on using the leakage tester, refer to the ULT2000 Series User Manual.
STEP 8
C
B
B Insert the fork adapter securely in between the control wheels on the controller and do not allow it to swivel.
D
Make sure the tester displays Ready for Testing, press FULL TEST. Repeat the test two more times. If any of the results are Fail, do not use the transducer and contact FUJIFILM SonoSite for repair.
C Connect the splitter to the transducer adapter, the fork adapter, and the leakage tester. D Connect the conductivity probe to the leakage tester and insert into 0.9% saline solution or Cidezyme/Enzol.
Disinfecting STEP 9 A Place up to 90 cm of the shaft in an approved disinfectant. Make sure the controller cannot fall into the solution. CHECK the expiration date of the disinfectant, and follow ALL chemical manufacturer’s instructions for soak time and temperature.
B Wipe the controller and cable with a sterile, non-linting cloth or single-use endoscopic sponge moistened with disinfectant. Remove cleaner by wiping again with a fresh sterile cloth or sponge moistened with sterile water.
STEP 10 Rinse the shaft for at least 1 minute in a large volume of sterile, deionized water (for example, 8 L). Repeat 3 separate times. Discard the rinse water after each rinse.
STEP 11 Dry the transducer with a
sterile, non-linting towel or medical-grade air. Examine the transducer and cable for any cracks. DO NOT use the transducer if any damage is found.
Transporting and storing STEP 12 Transport and store the transducer in accordance with your organization’s policies.
For complete instructions on cleaning, disinfecting, transporting, and storing the TEE transducer, refer to the TEE Transducer User Guide.
SonoSite is a registered trademark of FUJIFILM Sonosite, Inc. in various jurisdictions. FUJIFILM is a registered trademark of FUJIFILM Corporation. Value From Innovation is a trademark of FUJIFILM Holdings America Corporation. All other trademarks are the property of their respective owners.
Copyright © 2019 FUJIFILM SonoSite, Inc. All rights reserved.
P05497-09
*P05497-09* 2
Manufacturer: FUJIFILM SonoSite, Inc. 21919 30th Drive SE Bothell, WA 9802, USA T: 1-888-482-9449 or 1-425-951-1200 F: 1-425-951-1201
EC Authorized Representative FUJIFILM SonoSite B.V. Joop Geesinkweg 140 1114 AB Amsterdam, The Netherlands
Australia Sponsor FUJIFILM SonoSite Australasia Pty Ltd 114 Old Pittwater Road Brookvale, NSW, 2100 Australia