User Guide
394 Pages
Preview
Page 1
L52 Series Transducer
User Guide
Manufactured by SonoSite, Inc. 21919 30th Drive SE Bothell, WA 98021 USA T: 1-888-482-9449 or 1-425-951-1200 F: 1-425-951-1201 SonoSite Ltd Alexander House 40A Wilbury Way Hitchin Herts SG4 OAP UK T: +44-1462-444800 F: +44-1462-444801
Caution:
United States federal law restricts this device to sale by or on the order of a veterinarian.
180PLUS, M-Turbo, MicroMaxx, NanoMaxx, S Series, SonoSite, SonoSite TITAN, TITAN, and the SonoSite logo are registered trademarks or trademarks of SonoSite, Inc. Non-SonoSite product names may be trademarks or registered trademarks of their respective owners. The SonoSite ultrasound system(s) referenced in this document may be covered by one or more of the following U.S. patents: 5722412, 5817024, 5893363, 6135961, 6203498, 6364839, 6371918, 6383139, 6416475, 6447451, 6471651, 6569101, 6648826, 6575908, 6604630, 6817982, 6835177, 6962566, 7169108, 7449640, 7534211, 7549961, 7588541, 7591786, 7604596, 7643040, 7686766, 7694814, D456509, D461895, D509900, D538432, D544962, D558351, D559390, D591423, D592750, D592760, and by the following counterpart foreign patents: AU727381, AU730822, CA2372158, CA2373065, CN ZL 97113678.5, CN ZL 98106133.8, CN ZL 98108973.9, CN ZL 200830007734.8, DE60021552.0, DE60029777.2, DE60034670.6, DE69730563.5, DE6980539.6, DE69831698.3, DE60 2004 23 816.3-08, FR0815793, FR0875203, FR0881492, FR1175713, FR1180970, FR1589878, GB0875203, GB0881492, GB1175713, GB1180970, GB1180971, GB1589878, IT0815793, IT0881492, IT1175713, IT1589878, KR528102, KR532359, NO326202, NO326814, NZ542968, RCD000897368-0001, SP0815793, SP0881492, SP1589878. Patents pending
P07895-05 07/2010 Copyright 2010 by SonoSite, Inc. All rights reserved. Printed in the USA.
ii
English
L52 Series Transducer User Guide
Deutsch
Introduction ...1 Imaging ...1 Measurements and Calculations ...4 Safety ...5 Troubleshooting and Maintenance ... 21
Introduction Español Français
This user guide supplements the following ultrasound system user guides: • NanoMaxx Ultrasound System User Guide • M-Turbo Ultrasound System User Guide • S Series Ultrasound System User Guide • MicroMaxx Ultrasound System User Guide • TITAN Ultrasound System User Guide • SonoSite Ultrasound System User Guide
Italiano
It provides information for using the following transducers: • L52n/10-5 MHz (L52n) transducer on the NanoMaxx™ ultrasound system • L52x/10-5 MHz (L52x) transducer on the M-Turbo® ultrasound system, S Series™ ultrasound system, or MicroMaxx® ultrasound system • L52e/10-5 MHz (L52e) transducer on the MicroMaxx® ultrasound system • L52/10-5 MHz (L52) transducer on the TITAN® high-resolution ultrasound system or SonoSite®180PLUS™ ultrasound system
Português
See the ultrasound system user guide for additional safety information; for instructions on preparation, use, and maintenance of the ultrasound system; and for intended uses for each exam type and imaging mode.
Imaging Transducer, Exam Type, and Imaging Mode The following table describes the transducer, exam type, imaging mode, and optimization that may be available on your system.
1
Transducer, Exam Type, and Imaging Mode (NanoMaxx) Imaging Mode 2D M Mode
CPD
Color
OB
X
X
X
Vascular (Vas)
X
X
X
Musculoskeletal (Msk)
X
X
X
Transducer
Exam Type
L52n
.
Transducer, Exam Type, and Imaging Mode (M-Turbo or MicroMaxx) Imaging Mode 2D M Mode
CPD
Color
PW
CW
OB
X
X
X
X
-
Vascular (Vas)
X
X
X
X
-
Muscle (Msk)
X
X
X
X
-
Transducer
Exam Type
L52x
Transducer, Exam Type, and Imaging Mode (S Series) Imaging Mode
2
2D M Mode
CPD
Color
PW
OB
X
X
X
X
Vascular (Vas)
X
X
X
X
Muscle (Msk)
X
X
X
X
Transducer
Exam Type
L52x
English
Transducer, Exam Type, and Imaging Mode (MicroMaxx) Imaging Mode 2D M Mode
THI
2D MB
2D S
CPD
Color
PW
TDI PW
CW
L52e
OB
X
-
X
X
X
X
X
-
-
Vascular (Vas)
X
-
X
X
X
X
X
-
-
Muscle (Msk)
X
-
X
X
X
X
X
-
-
Español
Exam Type
Deutsch
Transducer
Transducer, Exam Type, and Imaging Mode (TITAN) Français
Imaging Mode Exam Type
2D
THI
CPD
DCPD
Color
M Mode
PW
CW
L52
OB
X
-
X
-
-
X
-
-
Vascular (Vas)
X
-
X
-
-
X
-
-
Muscle (Msk)
X
-
X
-
-
X
-
-
Italiano
Transducer
Português
Transducer, Exam Type, and Imaging Mode (180PLUS) Imaging Mode Transducer
Exam Type
2D
CPD
L52
OB
res, gen, pen
low, med, high
Vascular (Vas)
res, gen, pen
low, med, high
3
Measurements and Calculations Calculations This table shows the calculations available by exam type for the L52 Series transducer.
Calculations for L52n (NanoMaxx) Exam Type
Calculations
OB
OB
Calculations for L52x (M-Turbo and MicroMaxx) Exam Type
Calculations
Musculoskeletal (Msk)
Percent Reduction Volume
OB
OB
Vascular (Vas)
Percent Reduction Volume
Calculations for L52x (S Series) Exam Type
S Series System
Calculations
OB
S-VetMed
OB
Calculations for L52e/L52 (MicroMaxx and TITAN)
4
Exam Type
Calculations
Muscle (Msk)
Percent Reduction Volume
OB
OB
Vascular (Vas)
Percent Reduction Volume
English
Calculations for L52 (180PLUS) Calculations
OB
OB
Vascular (Vas)
Volume Volume Flow
Deutsch
Exam Type
Español
Safety Guidelines for reducing MI and TI
Table 1: MI Transducer
Français
The following are general guidelines for reducing MI or TI. If multiple parameters are given, the best results may be achieved by minimizing these parameters simultaneously. In some modes, changing these parameters does not affect MI or TI. Changes to other parameters may also result in MI and TI reductions. Please note the MI and TI values on the right side of the screen.
Depth
L52x (M-Turbo, S Series, or MicroMaxx)
↑
L52e (MicroMaxx)
↑
L52 (TITAN)
↑
L52 (180PLUS)
↑
Português
↑
Italiano
L52n (NanoMaxx)
↓ Decrease or lower setting of parameter to reduce MI. ↑ Raise or increase setting of parameter to reduce MI.
5
Table 2: TI (TIS, TIC, TIB) Color Power Doppler Settings
Transducer
PW Settings
PRF
Depth
L52n (NanoMaxx)
-
↑
-
L52x (M-Turbo or MicroMaxx)
↓
↑
Sample Vol ↓
L52x (S Series)
↓
↑
-
L52e (MicroMaxx)
↓
↑
Sample Vol ↑
L52 (TITAN)
↓
↑
-
L52 (180PLUS)
-
-
-
↓ Decrease or lower setting of parameter to reduce TI. ↑ Raise or increase setting of parameter to reduce TI. - Data are not applicable for this transducer/mode.
Transducer surface temperature rise Table 3 lists the measured surface temperature rise (in °C) from ambient (23°C ± 3°C) of transducers used on the ultrasound system. The temperatures were measured in accordance with EN 60601-2-37 section 42 where controls and settings were positioned to give maximum temperatures.s
L52n on NanoMaxx
L52x on M-Turbo or S Series
L52x on MicroMaxx
L52e on MicroMaxx
L52 on TITAN
L52 on 180PLUS
Table 3: Transducer Surface Temperature Rise IEC 60601-2-37 (Internal Use)
Still air
7.5
8.8
8.2
13.0
9.3
10.8
Simulated use
5.6
5.9
5.6
5.5
2.4
2.4
Test
6
English
Output display Table 4: TI or MI is ≥ 1.0 CPD
PW Doppler
L52n (NanoMaxx)
MI
Yes
Yes
Yes
n/a
TI
No
No
No
n/a
MI
Yes
Yes
Yes
Yes
TI
No
No
No
Yes
MI
Yes
Yes
Yes
-
TI
No
No
No
-
MI
Yes
Yes
Yes
Yes
TI
Yes
No
No
Yes
MI
No
Yes
Yes
Yes
TI
No
No
No
Yes
MI
No
No
No
No
TI
No
No
No
No
L52x (M-Turbo)
L52x (S Series)
L52x (MicroMaxx)
L52e (MicroMaxx)
L52 (TITAN, 180PLUS)*
*
Italiano
Color
Français
2D/ M Mode
Español
Index
Deutsch
Transducer Model
The L52 transducer never exceeds or equals a MI or TI of 1.0 on the TITAN or 180PLUS system.
Português
7
Acoustic output tables (M-Turbo and S Series) Table 5: Transducer Model: L52x/10-5
Operating Mode: 2D TIS
Index Label
M.I.
Associated Acoustic Parameter
Global Maximum Index Value pr.3
(MPa)
Other Information
Non-scan Scan
Aaprt≤1
Aaprt>1
(a)
-
-
#
-
W0
(mW)
min of [W.3(z1),ITA.3(z1)]
(mW)
-
z1
(cm)
-
zbp
(cm)
-
zsp
(cm)
deq(zsp)
(cm)
fc
(MHz)
5.33
PD PRF pr@PIImax
X (cm) Y (cm) (μsec) (Hz) (MPa)
0.15 7222 3.25
deq@Pllmax
(cm)
Dim of Aaprt
Focal Length IPA.3@MImax
Operating Control Conditions
1.01 2.336
TIB Non-scan
TIC
-
(a)
-
#
-
1.8
- # # #
- - -
- - -
- - -
# # #
-
FLx (cm)
#
-
-
#
FLy (cm)
#
-
-
#
2
(W/cm )
Control 1: Exam Type Control 2: Optimization Control 3: Depth Control 4: MB (Multi Beam)
329.1 Any Gen 4.2 cm Off
(a) This index is not required for this operating mode; value is <1. (b) This transducer is not intended for transcranial or neonatal cephalic uses. # No data are reported for this operating condition since the global maximum index value is not reported for the reason listed. (Reference Global Maximum Index Value line.) - Data are not applicable for this transducer/mode.
8
Operating Mode: M Mode TIS
Index Label
M.I.
Associated Acoustic Parameter
-
(a)
-
-
#
(mW)
min of [W.3(z1),ITA.3(z1)]
(mW)
-
z1
(cm)
-
zbp
(cm)
-
zsp
(cm)
deq(zsp)
(cm)
fc
(MHz)
(cm)
Focal Length IPA.3@MImax
(b)
#
#
#
1.8
# 5.33
- - -
# # #
- - -
# # #
# # #
0.15 1600 3.25 #
FLx (cm)
-
#
-
#
FLy (cm)
-
#
-
#
2
(W/cm )
329.1 Any Gen 4.2 cm Off
Português
Control 1: Exam Type Control 2: Optimization Control 3: Depth Control 4: MB (Multi Beam)
Italiano
deq@Pllmax
(a)
Français
PD PRF pr@PIImax
X (cm) Y (cm) (μsec) (Hz) (MPa)
TIC
Español
Other Information
1.01 2.336
Aaprt>1
W0
Dim of Aaprt
Operating Control Conditions
(MPa)
Aaprt≤1
Non-scan
Deutsch
Global Maximum Index Value pr.3
TIB
Non-scan Scan
English
Table 6: Transducer Model: L52x/10-5 (M-Turbo only)
(a) This index is not required for this operating mode; value is <1. (b) This transducer is not intended for transcranial or neonatal cephalic uses. # No data are reported for this operating condition since the global maximum index value is not reported for the reason listed. (Reference Global Maximum Index Value line.) - Data are not applicable for this transducer/mode.
9
Table 7: Transducer Model: L52x/10-5
Operating Mode: Color/CPD TIS
Index Label
Associated Acoustic Parameter
Global Maximum Index Value pr.3
Other Information
(MPa)
1.33 2.807
Non-scan Scan
Aaprt≤1
Aaprt>1
(a)
-
-
#
-
W0
(mW)
min of [W.3(z1),ITA.3(z1)]
(mW)
-
z1
(cm)
-
Zbp
(cm)
-
Zsp
(cm)
deq(Zsp)
(cm)
Fc
(MHz)
PD PRF pr@PIImax
X (cm) Y (cm) (μsec) (Hz) (MPa)
deq@Pllmax
(cm)
Dim of Aaprt
Focal Length IPA.3@MImax
Operating Control Conditions
M.I.
TIB
1.8
Non-scan
TIC
-
(b)
-
#
- -
4.37
#
-
-
-
#
# #
- -
- -
- -
# #
0.61 5378 3.628 -
FLx (cm)
#
-
-
#
FLy (cm)
#
-
-
#
2
(W/cm )
411.1
Control 1: Mode Control 2: Exam type Control 3: Optimization/Depth Control 4: PRF Control 5: Color Box Position/Size
CPD Vas Med/4.2 3472 bottom/ short/ wide
(a) This index is not required for this operating mode; value is <1. (b) This transducer is not intended for transcranial or neonatal cephalic uses. # No data are reported for this operating condition since the global maximum index value is not reported for the reason listed. (Reference Global Maximum Index Value line.) - Data are not applicable for this transducer/mode.
10
Operating Mode: PW Doppler TIS
Index Label
M.I.
-
1.32
-
-
(MPa)
W0
(mW)
min of [W.3(z1),ITA.3(z1)]
(mW)
63.6
z1
(cm)
1.9
Zbp
(cm)
1.9
Zsp
(cm)
deq(Zsp)
(cm)
Fc
(MHz)
2.1
(cm)
Focal Length
4.36
(W/cm )
- - -
- - -
4.36
4.36
#
2.29
1.476
#
0.55
0.55
#
0.43
- -
FLy (cm) IPA.3@MImax
1.5
1.38 1008 3.30
FLx (cm) 2
#
Italiano
deq@Pllmax
62.81
Français
(μsec) (Hz) (MPa)
(b)
0.45
X (cm)
PD PRF pr@PIImax
2.54
2.443
Y (cm)
Other Information
-
pr.3
Dim of Aaprt
- -
9.13
#
3.4
#
288.97
Control 1: Exam Type Control 2: PRF
Any 1008 Hz
Any Any
Control 3: SV Size Control 4: SV Position
1 mm Zone 4
3 mm Zone 7
Any 1302 or 1563 Hz 1 mm Zone 5
Português
Operating Control Conditions
Aaprt>1
TIC
Español
Associated Acoustic Parameter
1.17
Aaprt≤1
Non-scan
Deutsch
Global Maximum Index Value
TIB
Non-scan Scan
English
Table 8: Transducer Model: L52x/10-5 (M-Turbo only)
(a) This index is not required for this operating mode; value is <1. (b) This transducer is not intended for transcranial or neonatal cephalic uses. # No data are reported for this operating condition since the global maximum index value is not reported for the reason listed. (Reference Global Maximum Index Value line.) - Data are not applicable for this transducer/mode.
11
Acoustic output tables (NanoMaxx) Table 9: Transducer Model: L52n/10-5
Operating Mode: 2D TIS
Index Label
Associated Acoustic Parameter
Global Maximum Index Value pr.3
(MPa)
1.0 2.34
Non-scan Scan
Aaprt≤1
Aaprt>1
(a)
-
-
#
-
W0
(mW)
min of [W.3(z1),ITA.3(z1)]
(mW)
-
z1
(cm)
-
zbp
(cm)
-
zsp
(cm)
deq(zsp)
(cm)
fc
(MHz)
Non-scan
TIC
-
(b)
-
#
-
1.8
- 5.33
# # #
- - -
- - -
- - -
# # #
deq@Pllmax
X (cm) Y (cm) (μsec) 0.15 (Hz) 7707 (MPa) 3.25 (cm)
Focal Length
FLx (cm)
#
-
-
#
FLy (cm)
#
-
-
#
Dim of Aaprt PD PRF pr@PIImax
Other Information
M.I.
TIB
-
Operating Control Conditions
IPA.3@MImax
(W/cm2) 329.1 Control 1: Exam Type Any Control 2: Optimization Gen Control 3: Depth 4.2 cm Control 4: MB (Multi Beam) Off or On
(a) This index is not required for this operating mode; value is <1. (b) This transducer is not intended for transcranial or neonatal cephalic uses. # No data are reported for this operating condition since the global maximum index value is not reported for the reason listed. (Reference Global Maximum Index Value line.) - Data are not applicable for this transducer/mode.
12
Operating Mode: M Mode TIS
Index Label
Aaprt>1
-
(a)
-
-
#
min of [W.3(z1),ITA.3(z1)]
(mW)
-
z1
(cm)
-
zbp
(cm)
-
zsp
(cm)
deq(zsp)
(cm)
fc
(MHz)
Non-scan
TIC
(a)
(b)
#
# Español
Associated Acoustic Parameter
Aaprt≤1
(mW)
#
1.8
# 5.33
# # #
- - -
# # #
# # #
deq@Pllmax Focal Length
FLx (cm)
-
#
-
#
FLy (cm)
-
#
-
#
2
(W/cm ) 329.1 Any Gen 4.2 cm Off or On
Português
Control 1: Exam Type Control 2: Optimization Control 3: Depth Control 4: MB (Multi Beam)
#
Italiano
IPA.3@MImax
Français
- - -
X (cm) Y (cm) (μsec) 0.15 (Hz) 1600 (MPa) 3.25 (cm)
PD PRF pr@PIImax
Other Information
1.0 2.34
Non-scan Scan
W0
Dim of Aaprt
Operating Control Conditions
(MPa)
TIB
Deutsch
Global Maximum Index Value pr.3
M.I.
English
Table 10: Transducer Model: L52n/10-5
(a) This index is not required for this operating mode; value is <1. (b) This transducer is not intended for transcranial or neonatal cephalic uses. # No data are reported for this operating condition since the global maximum index value is not reported for the reason listed. (Reference Global Maximum Index Value line.) - Data are not applicable for this transducer/mode.
13
Table 11: Transducer Model: L52n/10-5
Operating Mode: Color/CPD TIs
Index Label
M.I.
Associated Acoustic Parameter
Global Maximum Index Value pr.3
Other Information
1.2 2.35
Non-scan Scan
A aprt≤1
Aaprt>1
(a)
-
-
#
-
W0
(mW)
min of [W.3(z1),ITA.3(z1)]
(mW)
-
z1
(cm)
-
Zbp
(cm)
-
Zsp
(cm)
deq(Zsp)
(cm)
Fc
(MHz)
1.8
Non-scan
TIc
-
(b)
-
#
- -
4.37
#
-
-
-
#
X
(cm)
#
-
-
-
#
Y
#
-
-
-
#
PD PRF pr@PIImax
(cm) (μsec) (Hz) (MPa)
deq@Pllmax
(cm)
Dim of Aaprt
Focal Length
0.60 7097 3.08 -
FLx (cm) FLy (cm)
IPA.3@MImax Control 1: Mode Operating Control Conditions
(MPa)
TIb
Control 2: Exam type Control 3: Optimization Control 4: depth Control 5: Color Box
2
(W/cm )
#
-
-
#
#
-
-
#
308.5 CPD or Color Msk Res 5.4 Default
(a) This index is not required for this operating mode; value is <1. (b) This transducer is not intended for transcranial or neonatal cephalic uses. # No data are reported for this operating condition since the global maximum index value is not reported for the reason listed. (Reference Global Maximum Index Value line.) - Data are not applicable for this transducer/mode.
14
English
Acoustic output tables (MicroMaxx) Table 12: Transducer Model: L52x/10-5
Operating Mode: 2D TIS M.I.
(MPa)
W0
(mW)
min of [W.3(z1),ITA.3(z1)]
(mW)
z1
(cm)
Zbp
(cm)
Zsp
(cm)
deq(Zsp)
(cm)
Fc
(MHz)
#
-
-
#
- - - - -
#
- - - 1.9 5.42
#
- - -
- - -
# #
0.146 8394 3.19
-
FLx (cm)
#
FLy (cm)
#
- -
- -
# #
325.3
Português
(W/cm2)
(b)
2.23
#
(cm)
-
Italiano
Other Information
-
Y (cm)
deq@Pllmax
Control 1: Exam Type Control 2: Optimization Control 3: Depth Control 4: MB
-
#
(μsec) (Hz) (MPa)
IPA.3@MImax
(a)
X (cm)
PD PRF pr@PIImax Focal Length
Aaprt>1
TIC
Français
pr.3
Dim of Aaprt
Operating Control Conditions
1.0
Aaprt≤1
Non-scan
Español
Associated Acoustic Parameter
Global Maximum Index Value
Non-scan Scan
Deutsch
Index Label
TIB
OB Gen 2.5 - 3.9 On or Off
(a) This index is not required for this operating mode; value is <1. (b) This transducer is not intended for transcranial or neonatal cephalic uses. # No data are reported for this operating condition since the global maximum index value is not reported for the reason listed. (Reference Global Maximum Index Value line.) - Data are not applicable for this transducer/mode.
15
Table 13: Transducer Model: L52x/10-5
Operating Mode: M-Mode M.I.
Index Label
Associated Acoustic Parameter
Global Maximum Index Value
1.0
pr.3
(MPa)
W0
(mW)
min of [W.3(z1),ITA.3(z1)]
(mW)
z1
(cm)
Zbp
(cm)
Zsp
(cm)
deq(Zsp)
(cm)
Fc
(MHz)
Dim of Aaprt
Other Information
deq@Pllmax
(cm)
Focal Length
Operating Control Conditions
Control 1: Exam Type Control 2: Optimization Control 3: Depth
(W/cm )
Aaprt>1
-
(a)
-
-
#
Non-scan
TIC
1.2
(b)
58.3
#
1.7 0.659
5.42
- - -
# # #
- - -
4.35
#
2.71
#
0.55
#
0.146 1600 3.19 0.641
- -
FLy (cm) IPA.3@MImax
A aprt≤1
1.9
FLx (cm) 2
Non-scan
- - -
Y (cm) (μsec) (Hz) (MPa)
Scan
TIB
2.23
X (cm)
PD PRF pr@PIImax
TIS
# #
- -
# #
325.3 OB Gen 2.5 - 3.9
Any Pen 15
(a) This index is not required for this operating mode; value is <1. (b) This transducer is not intended for transcranial or neonatal cephalic uses. # No data are reported for this operating condition since the global maximum index value is not reported for the reason listed. (Reference Global Maximum Index Value line.) - Data are not applicable for this transducer/mode.
16
Operating Mode: Color/CPD TIS
Index Label
M.I.
Associated Acoustic Parameter
W0
(mW)
min of [W.3(z1),ITA.3(z1)]
(mW)
z1
(cm)
Zbp
(cm)
Zsp
(cm)
deq(Zsp)
(cm)
Fc
(MHz)
#
-
1.4 4.35
#
FLy (cm)
#
Control 4: PRF Control 5: Color Box Position/Size
- - -
- - -
# # #
- -
- -
# #
377.1 Color Any Low/ 2.5 - 3.9 <718 Any
Português
Control 1: Mode Control 2: Exam Type Control 3: Optimization/Depth
- - - - -
- #
(W/cm )
#
0.607 4169 3.33
FLx (cm) 2
- - - -
#
(cm)
(b)
2.70
Y (cm)
deq@Pllmax
-
Italiano
Other Information
-
#
(μsec) (Hz) (MPa)
IPA.3@MImax
-
X (cm)
PD PRF pr@PIImax Focal Length
(a)
Français
(MPa)
Aaprt>1
TIC
Español
pr.3
Dim of Aaprt
Operating Control Conditions
1.3
Aaprt≤1
Non-scan
Deutsch
Global Maximum Index Value
TIB
Non-scan Scan
English
Table 14: Transducer Model: L52x/10-5
(a) This index is not required for this operating mode; value is <1. (b) This transducer is not intended for transcranial or neonatal cephalic uses. # No data are reported for this operating condition since the global maximum index value is not reported for the reason listed. (Reference Global Maximum Index Value line.) - Data are not applicable for this transducer/mode.
17
Table 15: Transducer Model: L52x/10-5
Operating Mode: PW Doppler TIS
Index Label
Associated Acoustic Parameter
Global Maximum Index Value pr.3
Other Information
(MPa)
1.2 2.48
Non-scan Scan
Aaprt≤1
Aaprt>1
-
1.4
-
-
68.5
W0
(mW)
min of [W.3(z1),ITA.3(z1)]
(mW)
-
z1
(cm)
-
Zbp
(cm)
-
Zsp
(cm)
deq(Zsp)
(cm)
Fc
(MHz)
PD PRF pr@PIImax
X (cm) Y (cm) (μsec) (Hz) (MPa)
deq@Pllmax
(cm)
Dim of Aaprt
Focal Length IPA.3@MImax
Operating Control Conditions
M.I.
TIB
2.3
Non-scan
TIC
2.2
(b)
37.5
#
2.4 0.19
4.36
-
4.35
-
4.36
#
- -
2.05 0.55
- -
0.90 0.55
# #
1.39 1008 3.505 0.18
FLx (cm)
-
8.32
-
#
FLy (cm)
-
3.5
-
#
2
(W/cm )
284.3
Control 1: Exam Type Any Control 2: Sample Volume 1 mm Control 3: PRF 1008 Hz Control 4: Sample Volume Position Zone 3
Any 2 mm Any Zone 7
Any 12 mm 10417 Hz Zone 3
(a) This index is not required for this operating mode; value is <1. (b) This transducer is not intended for transcranial or neonatal cephalic uses. # No data are reported for this operating condition since the global maximum index value is not reported for the reason listed. (Reference Global Maximum Index Value line.) - Data are not applicable for this transducer/mode.
18