Instructions for Use
41 Pages
Preview
Page 1
5-011-178-12(1) 2021-08
LCD Monitor
Instructions for Use Before operating the unit, please read this manual thoroughly and retain it for future reference.
LMD-X3200MD
© 2020 Sony Corporation
Indications for Use/Intended Use Symbols on the product
The LCD Monitor is intended to provide 4K 2D color video displays of surgical and other medical images.
Safety sign Follow the warnings in the instructions for use for parts of the unit on which this symbol appears. NOTE Background color: Blue Symbol: White
Notes This equipment is not intended for diagnostic use. This equipment is for medical professionals. This equipment is intended for use in medical environments, such as clinics, examination rooms, and operating rooms.
Warning for hot surfaces This symbol indicates parts of the unit that can become very hot. Be sure to obey the warning information found in the instructions for use. NOTE Background color: Yellow Triangular band: Black Symbol: Black
Warning To reduce the risk of fire or electric shock, do not expose this equipment to rain or moisture.
Consult the instructions for use Follow the directions in the instructions for use for parts of the unit on which this symbol appears.
To avoid electrical shock, do not open the cabinet. Refer servicing to qualified personnel only.
This symbol indicates the manufacturer, and appears next to the manufacturers name and address.
No modification of this equipment is allowed.
Warning To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
This symbol indicates the EU Importer, and appears next to the EU Importers name and address.
Warning
This symbol indicates the European Community representative, and appears next to the European Community representatives name and address.
This unit has no power switch. To disconnect the main power, unplug the power plug. When installing the unit, incorporate a readily accessible disconnect device in the fixed wiring, or connect the power plug to an easily accessible socket-outlet near the unit. Do not position the medical equipment where it is difficult to unplug the power plug. If a fault should occur during operation of the unit, operate the disconnect device to switch the power supply off, or disconnect the power plug.
This symbol indicates the medical device in the European Community. This symbol indicates the date of manufacture. This symbol indicates the serial number. This symbol indicates the version of the accompanying document. This symbol indicates the Unique Device Identifier (UDI), and appears next to the bar code representation of the Unique Device Identification. This symbol indicates the equipotential terminal which brings the various parts of a system to the same potential.
2
For customers in the U.S.A.
Storage and transport temperature This symbol indicates the acceptable temperature range for storage and transport environments.
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. You are cautioned that any changes or modifications not expressly approved in this manual could void your authority to operate this equipment. All interface cables used to connect peripherals must be shielded in order to comply with the limits for a digital device pursuant to Subpart B of part 15 of FCC Rules.
Storage and transport humidity This symbol indicates the acceptable humidity range for storage and transport environments. Storage and transport pressure This symbol indicates the acceptable atmospheric pressure range for storage and transport environments. This symbol indicates the DC input connector for connecting the AC adaptor (DC input). Signal input This symbol indicates the signal input connector. Signal output This symbol indicates the signal output connector.
If you have any questions about this product, you may call; Sony Customer Information Service Center 1-800222-7669 or http://www.sony.com/
DC output This symbol indicates the DC output connector.
Suppliers Declaration of Conformity Trade Name : SONY Model : LMD-X3200MD Responsible party : Sony Electronics Inc. Address : 16535 Via Esprillo, San Diego, CA 92127 U.S.A. Telephone Number: 858-942-2230 This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Caution Federal law (United States of America) restricts this device to sale by or on the order of a licensed healthcare practitioner.
3
For customers in Canada This unit has been certified according to Standard CAN/CSA-C22.2 No. 60601-1.
Important safeguards and notices for use in the medical environments 1. All devices connected to the unit must be certified or compliant according to IEC 60601-1, IEC 60950-1, and IEC 60065 standards and other IEC/ISO standards applicable to the devices. 2. Furthermore, the system as a whole must comply with IEC 60601-1 standards. All peripheral devices connected to the signal input/output sections of the unit constitute the medical-use system, and therefore, the user is responsible for ensuring that the system as a whole complies with IEC 60601-1 standards. If in doubt, consult qualified Sony service personnel. 3. Connecting the unit to other devices may increase the leakage current. 4. For all peripheral devices connected to the unit that operate on commercial power supplies and do not comply with IEC 60601-1 standards, incorporate an isolation transformer that complies with IEC 60601-1 standards and connect to the commercial power supply via the transformer. 5. The unit generates, uses, and may radiate radio frequency energy. If it is not installed and used in accordance with the instruction manual, it may cause interference on other devices. If the unit causes interference (which can be determined by disconnecting the power cord from the unit), try the following. Relocate the unit with respect to the affected devices. Connect the unit and the affected devices to different branch circuits. For more information, consult qualified Sony service personnel. (Applicable standard: IEC 60601-1-2)
4
Important EMC notices for use in medical environments The LMD-X3200MD needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the instructions for use. The LMD-X3200MD is intended for use in a professional healthcare facility environment. Portable and mobile RF communications equipment, such as cellular phones, can affect the LMDX3200MD. Warning Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the LMD-X3200MD. Otherwise, degradation of the performance of this equipment could result. If the LMD-X3200MD will be used adjacent to or stacked with other equipment, normal operation of the LMD-X3200MD under such configurations should be verified via observation. The use of accessories and cables other than those specified, with the exception of replacement parts sold by Sony Corporation, may result in increased emissions or decreased immunity of the LMDX3200MD. Guidance and manufacturers declaration electromagnetic emissions The LMD-X3200MD is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD-X3200MD should assure that it is used in such an environment. Emission test
Compliance
Electromagnetic environment guidance
Group 1
The LMD-X3200MD uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 RF emissions CISPR 11 CISPR 32 Harmonic emissions
The LMD-X3200MD is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Class B Class D (AC input)
IEC 61000-3-2 Not applicable (DC input) Voltage fluctuations/ flicker emissions IEC 61000-3-3
Complies (AC input) Not applicable (DC input)
5
Guidance and manufacturers declaration electromagnetic immunity The LMD-X3200MD is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD-X3200MD should assure that it is used in such an environment. Immunity test
IEC 60601 test level
Compliance level AC input
DC input
Electrostatic ±8 kV contact discharge (ESD) ±15 kV air IEC 61000-4-2
±8 kV contact
±8 kV contact
±15 kV air
±15 kV air
Electrical fast ±2 kV for power transient/burst supply lines
±2 kV for power supply lines
IEC 61000-4-4
±1 kV for input/ output lines
±1 kV for input/ output lines
Surge
±1 kV line(s) to line(s)
±1 kV differential Not applicable mode
±2 kV line(s) to earth
±2 kV common mode
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, a relative humidity of at least 30% is recommended. Mains power quality should be that of a typical commercial or hospital environment.
±1 kV for input/ output lines
IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input lines
0% UT 0% UT Not applicable (100% dip in UT) (100% dip in UT) for 0.5/1 cycles a for 0.5/1 cycles a 40% UT (60% dip in UT) for 5 cycles
40% UT (60% dip in UT) for 5 cycles
IEC 61000-4-11
70% UT (30% dip in UT) for 25/30 cycles a (for 0.5 sec)
70% UT (30% dip in UT) for 25/30 cycles a (for 0.5 sec)
0% UT (100% dip in UT) for 250/300 cycles a (for 5 sec)
0% UT (100% dip in UT) for 250/300 cycles a (for 5 sec)
30 A/m
30 A/m
Power frequency (50/60 Hz) magnetic field
Electromagnetic environment guidance
30 A/m
IEC 61000-4-8 NOTE: UT is the a.c. mains voltage prior to application of the test level. a For example, 10/12 means 10 cycles at 50 Hz or 12 cycles at 60 Hz.
6
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the LMD-X3200MD requires continued operation during power mains interruptions, it is recommended that the LMDX3200MD be powered from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Guidance and manufacturers declaration electromagnetic immunity The LMD-X3200MD is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD-X3200MD should assure that it is used in such an environment. Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment guidance Portable and mobile RF communications equipment should be used no closer to any part of the LMD-X3200MD, including cables, than the recommended separation distance calculated from the equation appliance to the frequency of the transmitter. Recommended separation distance
Conducted RF
3 Vrms 150 kHz to 80 MHz outside ISM bands c
3 Vrms
6 Vrms 150 kHz to 80 MHz in ISM bands c
6 Vrms
Radiated RF
3 V/m
3 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
IEC 61000-4-6
d = 1.2 ƊP
IEC 60601-1-2: 2007
d = 1.2 ƊP 80 MHz to 800 MHz d = 2.3 ƊP 800 MHz to 2.5 GHz IEC 60601-1-2: 2014
d = 2.0 ƊP 80 MHz to 2.7 GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. b Interference may occur in the vicinity of equipment marked with following symbol:
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
7
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the LMD-X3200MD is used exceeds the applicable RF compliance level above, the LMD-X3200MD should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the LMDX3200MD. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. c The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
Recommended separation distances between portable and mobile RF communications equipment and the LMD-X3200MD The LMD-X3200MD is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the LMD-X3200MD can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the LMD-X3200MD as recommended below, according to the maximum output power of the communications equipment. Rated maximum output power of transmitter W
Separation distance according to frequency of transmitter m IEC 60601-1-2 : 2007
IEC 60601-1-2 : 2014
150 kHz to 80 MHz d = 1.2 ƊP
80 MHz to 800 MHz d = 1.2 ƊP
800 MHz to 2.5 GHz d = 2.3 ƊP
150 kHz to 80 MHz d = 1.2 ƊP
80 MHz to 2.7 GHz d = 2.0 ƊP
0.01
0.12
0.12
0.23
0.12
0.20
0.1
0.38
0.38
0.73
0.38
0.63
1
1.2
1.2
2.3
1.2
2.0
10
3.8
3.8
7.3
3.8
6.3
100
12
12
23
12
20
For transmitters rated a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
8
Guidance and manufacturers declaration electromagnetic immunity The LMD-X3200MD is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the LMD-X3200MD. Otherwise, degradation of the performance of this equipment could result. Immunity test
Proximity fields from RF wireless communications equipment
Band a
Service a
Modulation
IEC 60601 test level
Compliance level
380 390 MHz
TETRA 400
Pulse modulation 18 Hz
27 V/m
27 V/m
430 470 MHz
GMRS 460 FRS 460
FM ±5 kHz deviation 1 kHz sine
28 V/m
28 V/m
704 787 MHz
LTE Band 13, 17
Pulse modulation 217 Hz
9 V/m
9 V/m
800 960 MHz
GSM 800/900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5
Pulse modulation 18 Hz
28 V/m
28 V/m
1,700 1,990 MHz
GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1, 3, 4, 25 UMTS
Pulse modulation 217 Hz
28 V/m
28 V/m
2,400 2,570 MHz
Bluetooth WLAN 802.11 b/g/n RFID 2450 LTE Band 7
Pulse modulation 217 Hz
28 V/m
28 V/m
5,100 5,800 MHz
WLAN 802.11 a/n
Pulse modulation 217 Hz
9 V/m
9 V/m
IEC 61000-4-3
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a For some services, only the uplink frequencies are included.
9
Caution
For customers in Europe
When you dispose of the unit or accessories, you must obey the laws in the relative area or country and the regulations in the relative hospital regarding environmental pollution.
This product has been manufactured by or on behalf of Sony Corporation, 1-7-1 Konan Minato-ku Tokyo, 108-0075 Japan. EU Importer: Sony Europe B.V. Inquiries to the EU Importer or related to product compliance in Europe should be sent to the manufacturers authorized representative, Sony Belgium, bijkantoor van Sony Europe B.V., Da Vincilaan 7-D1, 1930 Zaventem, Belgium.
Warning on power connections Use a proper power cord for your local power supply. 1. Use the approved Power Cord (3-core mains lead) / Appliance Connector / Plug with earthing-contacts that conforms to the safety regulations of each country if applicable. 2. Use the Power Cord (3-core mains lead) / Appliance Connector / Plug conforming to the proper ratings (Voltage, Ampere).
Warning The apparatus shall not be exposed to dripping or splashing. No objects filled with liquids, such as vases, shall be placed on the apparatus.
Warning Make sure the surface is wide enough so that this apparatuss width and depth dont exceed the surfaces edges. If not, this apparatus may lean or fall over and cause an injury.
If you have questions on the use of the above Power Cord / Appliance Connector / Plug, please consult a qualified service personnel.
Warning To prevent injury, if mounting the unit using a mounting arm, wall fixture, or other mounting device prepared by the customer, mount the unit securely as described in the instruction manual provided with the mounting device. Check beforehand that the mounting device used has sufficient strength to support the added weight of the unit. Check yearly that the mounting device is securely attached.
Warning on power connections for medical use Customers in the U.S.A. and Canada should use the following type of power cord. Customers in other countries or regions should use the power cord prescribed by their country or region.
Caution
U.S.A. and Canada Plug type
HOSPITAL GRADE*
Cord type
Min. Type SJT
When installing, ensure the following space around the periphery of the unit, taking ventilation and servicing into consideration. Rear side: 10 cm (4 in.) or more Left/Right sides: 10 cm (4 in.) or more Bottom side: 5 cm (2 in.) or more Top side: 30 cm (11 7/8 in.) or more
Min. 18 AWG Minimum rating for plug and appliance couplers
10 A / 125 V
Safety approval
UL Listed and CSA
Consult with Sony qualified personnel for the following types of installation location. Wall mount Ceiling mount Mounting arm
* Note: Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle marked Hospital Only or Hospital Grade.
10
Disposal of Old Electrical & Electronic Equipment (Applicable in Republic of India)
Warning If the power indicator is blinking in green, do not touch the unit. The unit may be hot due to a cooling malfunction caused by a fan error or blocked ventilation holes, and touching the unit may result in burns. For assistance, contact Sony qualified personnel.
This symbol and its variations indicate that this product and its components, consumables, parts or spares thereof shall not be treated as household waste and may not be dropped in garbage bins. Product owners are advised to deposit their product at the nearest collection point for the recycling of electrical and electronic equipment. Your co-operation shall facilitate proper disposal & help prevent potential negative consequences/hazards to the environment and human health, which could otherwise be caused by inappropriate waste disposal including improper handling, accidental breakage, damage and/ or improper recycling of e-waste. The recycling of materials will help to conserve natural resources. For more detailed information about recycling of this product, please contact your local civic office, your household waste disposal service provider or the store where you made the purchase. You may contact our companys toll free number in India for assistance. Toll Free: 1800-103-7799 Visit: www.sony.co.in for product recycling
Caution Do not use the device in a MR (Magnetic Resonance) environment. It may cause a malfunction, fire, and unwanted movement.
Caution This monitor should only be used with the specified monitor stand. For information on suitable stands, refer to Specifications. Installation of the monitor on any other stands may result in instability, possibly leading to injury. This equipment is not suitable for use in locations where children are likely to be present.
Reduction in the Use of Hazardous Substances in Electrical & Electronic Equipment (Applicable in Republic of India) This product and its components, consumables, parts or spares comply with the hazardous substances restriction of Indias E-Waste (Management) Rules. The maximum allowable concentrations of the restricted substances are 0.1% by weight in homogenous materials for Lead, Mercury, Hexavalent Chromium, Polybrominated Biphenyls (PBB) and Polybrominated Diphenyl Ethers (PBDE), and 0.01% by weight in homogenous materials for Cadmium, except for the exemptions specified in Schedule II of the aforesaid Rules.
11
For the customers in the U.S.A. SONY LIMITED WARRANTY - Please visit http://www.sony.com/psa/warranty for important information and complete terms and conditions of Sonys limited warranty applicable to this product. For the customers in Canada SONY LIMITED WARRANTY - Please visit http://www.sonybiz.ca/pro/lang/en/ca/ article/resources-warranty for important information and complete terms and conditions of Sonys limited warranty applicable to this product. For the customers in Europe Sony Professional Solutions Europe - Standard Warranty and Exceptions on Standard Warranty. Please visit https://pro.sony/en_GB/supportservices/warranty/support-professionalsolutions-europe-standard-product-warranty for important information and complete terms and conditions. For the customers in Korea SONY LIMITED WARRANTY - Please visit http://bpeng.sony.co.kr/handler/BPAS-Start for important information and complete terms and conditions of Sonys limited warranty applicable to this product.
12
Specifications ... 34 Dimensions ...40 Licenses ...41
Table of Contents Precaution ... 14 On Safety ... 14 On Installation ... 14 Precautions for using this unit safely ... 14 Precautions for connecting this unit with other medical devices ... 14 To prolong the life of the unit ... 14 Cautions for RESPONSIBLE ORGANIZATION when connecting this equipment to ITNETWORK ... 14 On simultaneous use with an electrosurgical knife, etc. ...15 Recommendation to Use more than One Unit ...15 LCD image display ...15 About the LCD Display Panel ...15 On a Long Period of Use ...15 On Burn-in ...15 On Fan Error ... 16 On Temperature Error ... 16 On Moisture Condensation ... 16 On Security ... 16 On Cleaning ... 16 On Repacking ...17
Features ...17 Location and Function of Parts and Controls ... 18 Front Panel ... 18 Input Signals and Adjustable/Setting Items... 19 Rear Panel ... 20
Preparation ... 23 Connecting ... 23 Turning on the Monitor/Switching Input Settings ... 24
Initial Setting ... 25 Using the Menu ... 26 Adjustment Using the Menus ... 27 Items ... 27 Adjusting and Changing the Settings ... 28 Color Tone Adjustment menu ... 28 Screen Control menu ... 29 PIP / POP menu ... 29 Input/Output Configuration menu ... 29 System Configuration menu ...30 Initial Setup menu ... 32 Preset menu ... 33
The terms HDMI and HDMI High-Definition Multimedia Interface, and the HDMI Logo are trademarks or registered trademarks of HDMI Licensing Administrator, Inc. in the United States and other countries.
Troubleshooting ... 33 Error Messages ... 34
13
is a higher possibility that you experience discomfort.
Precaution
Precautions for connecting this unit with other medical devices
On Safety Operate the unit only with a power source as
Before you utilize this device and/or connect
specified in the “Specifications” section. The nameplate indicating operating voltage, etc. is located on the rear panel of monitor and the AC adaptor. Should any solid object or liquid fall into the cabinet, unplug the unit and have it checked by qualified personnel before operating it any further. Unplug the unit from the wall outlet if it is not to be used for several days or more. To disconnect the AC power cord, pull it out by grasping the plug. Never pull the cord itself. The socket-outlet shall be installed near the equipment and shall be easily accessible.
this device to any other medical device, please be aware of and abide by the following precautions: (a) Before actually using this device for medical practice, please check and confirm that you do not experience any discomfort in your use that could be disruptive or impeditive in conducting your intended activity or medical practice. (b) If you experience or are likely to experience such discomfort, please refrain from using this device. (c) Generally, discomfort (such as eye strain, fatigue, nausea, or motion sickness) can be provoked by such factors as quick movements or shakiness of video picture, focal position of video pictures, distance between objects and image capturing modules, user’s point of gaze in video pictures, other varying conditions of video pictures to be input to this device, and individual user’s health conditions. Before you utilize this unit, check if the image of the connected medical device is displayed properly on the screen of this unit.
On Installation Prevent internal heat build-up allowing
adequate air circulation. Do not place the unit on surfaces (rugs, blankets, etc.) or near materials (curtains, draperies) that may block the ventilation holes. Do not install the unit near heat sources such as radiators or air ducts, or in a place subject to direct sunlight, excessive dust, mechanical vibration or shock. Do not place the monitor near equipment which generates magnetism, such as a transformer or high voltage power lines.
To prolong the life of the unit Turn off the power to preserve the performance when not in use for a prolonged time.
Cautions for RESPONSIBLE ORGANIZATION when connecting this equipment to ITNETWORK
Precautions for using this unit safely Some people may experience discomfort (such
as eye strain, fatigue, or nausea) while watching video images. Sony recommends that all viewers take regular breaks while watching video images. The length and frequency of necessary breaks will vary from person to person. You must decide what works best. If you experience any discomfort, you should stop watching the video images until the discomfort ends; consult a doctor if you believe necessary. Avoid watching the display in environments where your head may shake, or while you are walking or performing exercise, because there
– connection of the PEMS to an IT-NETWORK that includes other equipment could result in previously unidentified RISKS to PATIENTS, OPERATORS or third parties; – the RESPONSIBLE ORGANIZATION should identify, analyze, evaluate and control these RISKS; – subsequent changes to the IT-NETWORK could introduce new RISKS and require additional analysis; and – changes to the IT-NETWORK include:
14
the physical characteristics of the liquid crystal display, such “stuck” pixels may appear spontaneously. These problems are not a malfunction. Do not leave the LCD screen facing the sun as it can damage the LCD screen. Take care when you place the unit by a window. Do not push or scratch the LCD screen. Do not place a heavy object on the LCD screen. This may cause the screen to lose uniformity. If the unit is used in a cold place, a residual image may appear on the screen. This is not a malfunction. When the monitor becomes warm, the screen returns to normal. The screen and the cabinet become warm during operation. This is not a malfunction.
changes in the IT-NETWORK configuration; connection of additional items to the IT-
NETWORK; disconnecting items from the IT-NETWORK; update of equipment connected to the IT-
NETWORK; and upgrade of equipment connected to the IT-
NETWORK.
On simultaneous use with an electrosurgical knife, etc. If this unit is used together with an electrosurgical knife, etc., the picture may be disturbed, warped or otherwise abnormal as a result of strong radio waves or voltages from the device. This is not a malfunction. When you use this unit simultaneously with a device from which strong radio waves or voltages are emitted, confirm the effect of this before using such devices, and install this unit in a way that minimizes the effect of radio wave interference.
On a Long Period of Use Due to the characteristics of LCD panel, displaying static images for extended periods, or using the unit repeatedly in a high temperature/ high humidity environments may cause image smearing, burn-in, areas of which brightness is permanently changed, lines, or a decrease in overall brightness.
Recommendation to Use more than One Unit
In particular, continued display of an image smaller than the monitor screen, such as in a different aspect ratio, may shorten the life of the unit. Avoid displaying a still image for an extended period, or using the unit repeatedly in a high temperature/high humidity environment such an airtight room, or around the outlet of an air conditioner.
As problems can occasionally occur for the monitor, when the monitor is used for safety control of personnel, assets or stable picture, or for emergencies, we strongly recommend you use more than one unit or prepare a spare unit.
LCD image display
To prevent any of the above issues, we recommend reducing brightness slightly, and to turn off the power whenever the unit is not in use.
Due the physical characteristics of LCD panels, there may be a decrease in brightness or change in color temperature over a long period of use. These problems are not a malfunction. In addition, these occurrences will not affect recorded data.
On Burn-in For LCD panel, permanent burn-in may occur if still images are displayed in the same position on the screen continuously, or repeatedly over extended periods.
About the LCD Display Panel The LCD panel fitted to this unit is
manufactured with high precision technology, giving a functioning pixel ratio of at least 99.99%. Thus a very small proportion of pixels may be “stuck”, either always off (black), always on (red, green, or blue), or flashing. In addition, over a long period of use, because of
Images that may cause burn-in Masked images with aspect ratios other than 16:9
15
SPECIFICATIONS, OR SECURITY PROBLEMS OF ANY KIND. Depending on the operating environment, unauthorized third parties on the network may be able to access the unit. When connecting the unit to the network, be sure to confirm that the network is protected securely. This unit is equipped with a maintenance function performed via a network. Maintenance may be performed with your consent.
Color bars or images that remain static for a
long time Character or message displays that indicate settings or the operating state To reduce the risk of burn-in Turn off the character displays
Press the MENU button to turn off the character displays. To turn off the character displays of the connected equipment, operate the connected equipment accordingly. For details, refer to the operation manual of the connected equipment. Turn off the power when not in use Turn off the power if the monitor is not to be used for a prolonged period of time.
On Cleaning Before cleaning Be sure to disconnect the AC power cord from the AC outlet.
On Fan Error
On cleaning the monitor A material that withstands disinfection is used for the front protection plate of the medical use LCD monitor. The protection plate surface is specially treated to reduce reflection of light. When solvents such as benzene or thinner, or acid, alkaline or abrasive detergent, or chemical cleaning cloth are used for the protection plate surface/monitor surface, the performance of the monitor may be impaired or the finish of the surface may be damaged. Take care with respect to the following: Clean the protection plate surface/monitor surface with a 50 to 70 v/v% concentration of isopropyl alcohol or a 76.9 to 81.4 v/v% concentration of ethanol using a swab method. Wipe the protection plate surface gently (wipe using less than 1 N force). Stubborn stains may be removed with a soft cloth such as a cleaning cloth lightly dampened with mild detergent solution using a swab method and then clean using the above chemical solution. Never use solvents such as benzene or thinner, or acid, alkaline or abrasive detergent, or chemical cleaning cloth for cleaning or disinfection, as they will damage the protection plate surface/monitor surface. Do not use unnecessary force to rub the protection plate surface/monitor surface with a stained cloth. The protection plate surface/ monitor surface may be scratched. Do not keep the protection plate surface/ monitor surface in contact with a rubber or vinyl resin product for a long period of time.
The fan for cooling the unit is built in. When the fan error indication appears on the screen, turn off the power and contact an authorized Sony dealer.
On Temperature Error When this unit is used in a high temperature environment and the internal temperature rises, a temperature error is displayed on the screen. When the temperature error is displayed, contact an authorized Sony dealer.
On Moisture Condensation If the unit is suddenly taken from a cold to a warm location, or if ambient temperature suddenly rises, moisture may form on the outer surface of the unit and/or inside of the unit. This is known as condensation. If condensation occurs, turn off the unit and wait until the condensation clears before operating the unit. Operating the unit while condensation is present may damage the unit.
On Security SONY WILL NOT BE LIABLE FOR DAMAGES OF
ANY KIND RESULTING FROM A FAILURE TO IMPLEMENT PROPER SECURITY MEASURES ON TRANSMISSION DEVICES, UNAVOIDABLE DATA LEAKS RESULTING FROM TRANSMISSION
16
The finish of the surface may deteriorate or the coating may come off.
Features
On Repacking
This monitor displays color video images that are output from medical imaging systems on the LCD (liquid crystal display) panel. Liquid crystal and color filters are laid on the front of flat light source (backlight) on the LCD panel. And then, the LCD panel displays images by controlling the aperture of the liquid crystal according to input signals.
Do not throw away the carton and packing materials. They make an ideal container which to transport the unit. If you have any questions about this unit, contact your authorized Sony dealer.
Compliance with medical safety standards in U.S.A., Canada and Europe IEC 60601-1 and product safety standards in the U.S.A., Canada and Europe have been obtained for this monitor. The monitor is designed for use in the medical treatment field, with the sheet switch, screen protect panel, etc.
High brightness/high-resolution 4K panel A 4K high-resolution (3840 × 2160) panel and wide field of view technology enables you to use the monitor under various lighting conditions and in numerous ways (installing on the wall, using several monitors to view an image, etc.).
A.I.M.E. function 1) This function sharpens images by adjusting color, contrast, and visibility of the dark area. You can select the mode best suited for the resolution of the equipment used. 1) A.I.M.E. is a registered trademark of Sony Corporation.
Control panel Assigns functions frequently used during an operation to buttons on the front surface of the monitor. The panel provides an user interface superior in operability through navigation by luminescent colors and status of the buttons.
Flat surface for better maintenance The design allows the user to easily wipe liquids and gel off the LCD panel and control buttons – ensuring a high standard of disinfection and cleanliness.
17
Location and Function of Parts and Controls Front Panel
Power indicator Indicator
Operating State
Green
Power on
Flashes in Green
Power on with image displayed (reduced backlight due to high temperature)
Orange
Standby
Flashes in Orange
No image displayed (remote standby)
When you press PORT A or PORT B when it is lit in white, an image from the input connector assigned to the PORT A/B is displayed. Multi-image display setting buttons PIP/POP: For displaying the multi-image display or switching the multi-image display mode. SWAP: For swapping between the main display and the sub display.
(On)/ (Standby) switch Press the side to turn the monitor on. Press the side to switch the unit to standby state.
A.I.M.E. button Press to select “Off,” “On,” or “Check Mode” for the A.I.M.E. setting. Also, you can adjust A.I.M.E. Structure, A.I.M.E. Color, and A.I.M.E. Shadow.
Input select buttons PORT A/PORT B: Each input connector can be assigned for PORT A/B. 12G-SDI is assigned for PORT A/B in the default setting. When you press PORT A or PORT B when it is lit in green, a menu for selecting the input connector assigned to the PORT A/B is displayed.
CUSTOM buttons Turns on or off the assigned function. You can adjust the assigned function by pressing the / buttons. (Refer to the Custom Button of the System Configuration menu on page 32 and of the Initial Setup menu on page 32.)
18
Press again to hide the menu. /// buttons Press to select the items and setting values.
The following functions are assigned in the default setting. CUSTOM 1: Brightness CUSTOM 2: Contrast CUSTOM 3: Flip Pattern
CONTROL button Displays or hides the operation buttons on the front panel. Selects the items depending on the menu types.
OSD menu operation buttons MENU button Press to display the on-screen menu.
Input Signals and Adjustable/Setting Items Input signal Item Gamma Phase Chroma Brightness Contrast A.I.M.E. A.I.M.E. Structure A.I.M.E. Color A.I.M.E. Shadow Color Temperature Gain R/G/B Offset Bias R/G/B Offset Mono Sharpness H Sharpness V RGB Range Color Space 4K Scan Size HD Scan Size SD Scan Size 4K Zoom Flip Pattern SD Aspect
HDMI
Display Port
SDI
DVI-D
Video
PC
Video
PC
12G-SDI
3G-SDI
Video
PC
1)
3) 2) 2) 1)
1)
3) 2) 2) 1)
× ×
× ×
Image Division
×
×
HDMI Signal Format
×
×
×
×
×
2) 2)
×
×
×
1)
1)
1)
× ×
× ×
× ×
×
× ×
: Can be adjusted/set × : Cannot be adjusted/set 1) SD aspect is reflected to the screen only when the SD signal is input. 2) The setting value can be changed but is not applied to the screen when the PC signal is input. 3) The setting value can be changed but is not applied to the screen.
19
Rear Panel To use the connectors on the rear panel, remove the cable cover. For details on the cable cover, see page 23.
Cable cover (M) This cover is specifically for the 12G-SDI input/ output connector. To connect/disconnect the cable of the 12G-SDI input/output connector without removing the cable cover (L), pull the cover open with your finger hooked in the concaved part near the on the cover.
Cable cover (L) When connecting or disconnecting the cable of the connector on the rear panel, remove this cover.
20