Instructions for Use
42 Pages
Preview
Page 1
4-596-096-11(2) 2016-10
LCD Monitor
Instructions for Use Before operating the unit, please read this manual thoroughly and retain it for future reference.
LMD-X310MT LMD-X550MT
© 2016 Sony Corporation
Indications for Use/Intended Use The LCD Monitor is intended to provide 4K, 3D and 2D color video displays of images from endoscopic/ laparoscopic camera systems, surgical microscope and other compatible medical imaging systems. The LCD Monitor is a widescreen, high-definition, medical grade monitor for real-time use during minimally invasive surgical procedures and is suitable for use in hospital operating rooms, surgical centers, clinics, doctors’ offices and similar medical environments.
WARNING To reduce the risk of fire or electric shock, do not expose this apparatus to rain or moisture. To avoid electrical shock, do not open the cabinet. Refer servicing to qualified personnel only. No modification of this equipment is allowed. LMD-X550MT only To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth. Symbols on the products Safety sign (LMD-X310MT only) Follow the warnings in the instructions for use for parts of the unit on which this mark appears. NOTE Background color: Blue Symbol: White Consult the instructions for use Follow the directions in the instructions for use for parts of the unit on which this mark appears. This symbol indicates the manufacturer, and appears next to the manufacturer’s name and address. This symbol indicates the European Community representative, and appears next to the European Community representative’s name and address.
This symbol indicates the date of manufacture. This symbol indicates the equipotential terminal which brings the various parts of a system to the same potential. Storage and transport temperature This symbol indicates the acceptable temperature range for storage and transport environments. Storage and transport humidity This symbol indicates the acceptable humidity range for storage and transport environments. Storage and transport pressure This symbol indicates the acceptable atmospheric pressure range for storage and transport environments. For the customers in the U.S.A. This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. You are cautioned that any changes or modifications not expressly approved in this manual could void your authority to operate this equipment. All interface cables used to connect peripherals must be shielded in order to comply with the limits for a digital device pursuant to Subpart B of part 15 of FCC Rules.
This symbol indicates the serial number.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
This symbol indicates the version of the accompanying document.
For the customers in Canada CAN ICES-3 (A)/NMB-3(A) This unit has been certified according to Standard CAN/ CSA-C22.2 No.60601-1.
2
Important safeguards/notices for use in the medical environments 1. All the equipments connected to this unit shall be certified according to Standard IEC 60601-1, IEC 60950-1, IEC 60065 or other IEC/ISO Standards applicable to the equipments.
4. For this particular equipment, all accessory equipment connected as noted above, must be connected to mains via an additional isolation transformer conforming with the construction requirements of IEC 60601-1 and providing at least Basic Insulation.
2. Furthermore all configurations shall comply with the system standard IEC 60601-1. Everybody who connects additional equipment to the signal input part or signal output part configures a medical system, and is therefore, responsible that the system complies with the requirements of the system standard IEC 60601-1. If in doubt, consult the qualified service personnel.
5. This equipment generates, uses, and can radiate radio frequency energy. If it is not installed and used in accordance with the instruction manual, it may cause interference to other equipment. If this unit causes interference (which can be determined by unplugging the power cord from the unit), try these measures: Relocate the unit with respect to the susceptible equipment. Plug this unit and the susceptible equipment into different branch circuit.
3. The leakage current could increase when connected to other equipment.
Consult your dealer. (According to standard IEC 606011-2 and CISPR11, Class B, Group 1)
Important EMC notices for use in the medical environments • The LMD-X310MT/X550MT needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the instructions for use. • The portable and mobile RF communications equipment such as cellular phones can affect the LMD-X310MT/ X550MT.
Warning The use of accessories and cables other than those specified, with the exception of replacement parts sold by Sony Corporation, may result in increased emissions or decreased immunity of the LMD-X310MT/X550MT. Guidance and manufacturer’s declaration - electromagnetic emissions The LMD-X310MT/X550MT is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD-X310MT/X550MT should assure that it is used in such an environment. Emission test
Compliance
Electromagnetic environment - guidance
Group 1
The LMD-X310MT/X550MT uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 RF emissions Class B CISPR 11 Harmonic emissions IEC 61000-3-2 Voltage fluctuations/flicker emissions
Not applicable (LMD-X310MT)
The LMD-X310MT/X550MT is suitable for use in all establishments, including domestic establishments and those directly connected to the public lowvoltage power supply network that supplies buildings used for domestic purposes.
Class D (LMD-X550MT) Not applicable (LMD-X310MT) Complies (LMD-X550MT)
IEC 61000-3-3 Warning If the LMD-X310MT/X550MT should be used adjacent to or stacked with other equipment, it should be observed to verify normal operation in the configuration in which it will be used.
3
Guidance and manufacturer’s declaration - electromagnetic immunity The LMD-X310MT/X550MT is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD-X310MT/X550MT should assure that it is used in such an environment. Immunity test
IEC 60601 test level
Compliance level LMD-X550MT
Electrostatic ±6 kV contact discharge (ESD)
±6 kV contact
±6 kV contact
IEC 61000-4-2
±8 kV air
±8 kV air
±8 kV air
Electrical fast transient/burst
±2 kV for power supply lines
IEC 61000-4-4
±1 kV for input/ output lines
±1 kV for input/ output lines
±1 kV for input/ output lines
Surge
±1 kV line(s) to line(s)
Not applicable
IEC 61000-4-5
±2 kV line(s) to earth
±1 kV differential Mains power quality should be that of a typical commercial or hospital mode environment. ±2 kV common mode
Voltage dips, short interruptions and voltage variations on power supply input lines
< 5% UT (> 95% dip in UT) for 0.5 cycle
Not applicable
< 5% UT (> 95% dip in UT) for 0.5 cycle
IEC 61000-4-11
±2 kV for power supply lines
40% UT (60% dip in UT) for 5 cycles
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 25 cycles
70% UT (30% dip in UT) for 25 cycles
< 5% UT (> 95% dip in UT) for 5 sec
< 5% UT (> 95% dip in UT) for 5 sec
Power frequency 3 A/m (50/60 Hz) magnetic field
3 A/m
3 A/m
IEC 61000-4-8 NOTE: UT is the a.c. mains voltage prior to application of the test level.
4
Electromagnetic environment guidance
LMD-X310MT
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the LMD-X310MT/X550MT requires continued operation during power mains interruptions, it is recommended that the LMDX310MT/X550MT be powered from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Guidance and manufacturer’s declaration - electromagnetic immunity The LMD-X310MT/X550MT is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD-X310MT/X550MT should assure that it is used in such an environment. Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance Portable and mobile RF communications equipment should be used no closer to any part of the LMDX310MT/X550MT, including cables, than the recommended separation distance calculated from the equation appliance to the frequency of the transmitter. Recommended separation distance
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
3 Vrms
d = 1.2 √P
3 V/m
d = 1.2 √P 80 MHz to 800 MHz d = 2.3 √P 800 MHz to 2.5 GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. b Interference may occur in the vicinity of equipment marked with following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the LMD-X310MT/X550MT is used exceeds the applicable RF compliance level above, the LMD-X310MT/X550MT should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the LMD-X310MT/X550MT. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
5
Recommended separation distances between portable and mobile RF communications equipment and the LMD-X310MT/X550MT The LMD-X310MT/X550MT is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the LMD-X310MT/X550MT can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the LMD-X310MT/X550MT as recommended below, according to the maximum output power of the communications equipment. Rated maximum output power of transmitter W
Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d = 1.2 √P
80 MHz to 800 MHz d = 1.2 √P
800 MHz to 2.5 GHz d = 2.3 √P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Warning The apparatus shall not be exposed to dripping or splashing. No objects filled with liquids, such as vases, shall be placed on the apparatus. Warning To prevent injury, this apparatus must be securely attached to the floor/wall in accordance with the instructions for use. When fixing the unit, use more than 4 of the supplied screws for the VESA mount or specified screws. Fasten the screws to the hole positions firmly and symmetrically to the upper and lower, left and right positions towards the center of the monitor. Consult with Sony qualified personnel for floor/wall mount installation. Caution When you dispose of the unit or accessories, you must obey the laws in the relative area or country and the regulations in the relative hospital regarding environmental pollution. Caution When installing, the installation space must be secured in consideration of the ventilation and service operation. Leave a space 4 cm (1 5/8 inches) or more behind, 10 cm (4 inches) or more from the left and right sides of, 6 cm
6
(2 3/8 inches) or more from the bottom side of, and 30 cm (11 7/8 inches) or more above the unit. For the customers in the U.S.A. Caution Federal law (United States of America) restricts this device to sale by or on the order of a licensed healthcare practitioner.
Disposal of Old Electrical & Electronic Equipment (Applicable in Republic of India)
This symbol indicates that this product and its components, consumables, parts or spares thereof shall not be treated as household waste and may not be dropped in garbage bins. Product owners are advised to deposit their product at the nearest collection point for the recycling of electrical and electronic equipment. Your co-operation shall facilitate proper disposal & help prevent potential negative consequences/hazards to the environment and human health, which could otherwise be caused by inappropriate waste disposal including improper handling, accidental breakage, damage and/ or improper recycling of e-waste. The recycling of materials will help to conserve natural resources. For more detailed information about recycling of this product, please contact your local civic office, your household waste disposal service provider or the store where you made the purchase. You may contact our company’s toll free number in India for assistance. Toll Free: 1800-103-7799 Visit: www.sony.co.in for product recycling
For the customers in the U.S.A. SONY LIMITED WARRANTY - Please visit http:// www.sony.com/psa/warranty for important information and complete terms and conditions of Sony’s limited warranty applicable to this product. For the customers in Canada SONY LIMITED WARRANTY - Please visit http:// www.sonybiz.ca/pro/lang/en/ca/article/resourceswarranty-product-registration for important information and complete terms and conditions of Sony’s limited warranty applicable to this product. For the customers in Europe Sony Professional Solutions Europe - Standard Warranty and Exceptions on Standard Warranty. Please visit http://www.pro.sony.eu/warranty for important information and complete terms and conditions. For the customers in Korea SONY LIMITED WARRANTY - Please visit http:// bpeng.sony.co.kr/handler/BPAS-Start for important information and complete terms and conditions of Sony’s limited warranty applicable to this product.
Reduction in the Use of Hazardous Substances in Electrical & Electronic Equipment (Applicable in Republic of India) This product and its components, consumables, parts or spares comply with the hazardous substances restriction of India’s E-Waste (Management) Rules. The maximum allowable concentrations of the restricted substances are 0.1% by weight in homogenous materials for Lead, Mercury, Hexavalent Chromium, Polybrominated Biphenyls (PBB) and Polybrominated Diphenyl Ethers (PBDE), and 0.01% by weight in homogenous materials for Cadmium, except for the exemptions specified in Schedule II of the aforesaid Rules.
7
LMD-X550MT only Warning This unit has no power switch. To disconnect the main power, unplug the power plug. When installing the unit, incorporate a readily accessible disconnect device in the fixed wiring, or connect the power plug to an easily accessible socket-outlet near the unit. Do not position the ME equipment where it is difficult to unplug the power plug. If a fault should occur during operation of the unit, operate the disconnect device to switch the power supply off, or unplug the power plug.
Warning on power connection Use a proper power cord for your local power supply. 1. Use the approved Power Cord (3-core mains lead) / Appliance Connector / Plug with earthing-contacts that conforms to the safety regulations of each country if applicable. 2. Use the Power Cord (3-core mains lead) / Appliance Connector / Plug conforming to the proper ratings (Voltage, Ampere). If you have questions on the use of the above Power Cord / Appliance Connector / Plug, please consult a qualified service personnel.
For the customers in U.S.A. and Canada Warning on power connection for medical use Please use the following power supply cord. With connectors (plug or female) and cord types other than those indicated in this table, use the power supply cord that is approved for use in your area. United States and Canada Plug Type
HOSPITAL GRADE*
Cord Type
Min. Type SJT Min. 18 AWG
Minimum Rating for Plug and Appliance Couplers
10A/125V
Safety Approval
UL Listed and CSA
*Note: Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle marked 'Hospital Only' or 'Hospital Grade'.
8
Caution This unit is heavy. Make sure to unpack and move the unit with two or more people. Firmly grip the bottom of this unit as shown below.
Table of Contents Precaution ... 10 On Safety ...10 On Installation ...10 Precautions for using this unit safely ...10 Precautions for connecting this unit with other medical devices ...10 On simultaneous use with an electrosurgical knife, etc. ...10 Recommendation to Use more than One Unit ...10 LCD image display ...10 About the LCD Display Panel ...11 About the Screen Protect Panel ...11 On a Long Period of Use ...11 On Burn-in ...11 Handling the 3D Glasses ...11 On Fan Error ...11 On Temperature Error ...12 On Moisture Condensation ...12 On Cleaning ...12 On Repacking ...12 Disposal of the Unit ...12
LMD-X310MT ... 32 LMD-X550MT ... 33
Dimensions ... 41 LMD-X310MT ... 41 LMD-X550MT ... 41 LMD-X310MT/X550MT ... 42
3D Viewing Angle (vertical) ... 42
Features ... 13 Location and Function of Parts and Controls ... 15 Front Panel... 15 Input Signals and Adjustable/Setting Items ... 16 Rear/Bottom Panel ... 17
Connecting the AC Power Cord ... 19 LMD-X310MT ...19 LMD-X550MT ...20
Removing the Connector Cover ... 21 Configuration Example: 4K Surgical Endoscope ... 22 Initial Setting ... 22 Using the Menu ... 24 Adjustment Using the Menus ... 25 Items ...25 Adjusting and changing the settings ...26 Color Tone Adjustment menu ...26 Screen Control menu ...27 3D Setting menu ...27 PIP / POP menu ...28 Input Configuration menu ...29 System Configuration menu ...29 Initial Setup menu ...30 Preset menu ...31
Troubleshooting ... 31 Error Messages ... 32 Specifications ... 32
The terms HDMI and HDMI High-Definition Multimedia Interface, and the HDMI Logo are trademarks or registered trademarks of HDMI Licensing LLC in the United States and other countries.
Table of Contents
9
Precaution
Precautions for connecting this unit with other medical devices Before you utilize this device and/or connect this
On Safety LMD-X310MT is a DC powered device. Use with the
supplied AC adaptor (AC-300MD). LMD-X550MT is an AC powered device. Operate the unit on 100-240 V AC only. The nameplate indicating operating voltage, etc. is
located on the AC adaptor. Should any solid object or liquid fall into the cabinet,
unplug the unit and have it checked by qualified personnel before operating it any further. Unplug the unit from the wall outlet if it is not to be used for several days or more. To disconnect the AC power cord, pull it out by grasping the plug. Never pull the cord itself. The socket-outlet shall be installed near the equipment and shall be easily accessible.
On Installation
device to any other medical device, please be aware of and abide by the following precautions: (a) Before actually using this device for medical practice, please check and confirm that you do not experience any discomfort in your use that could be disruptive or impeditive in conducting your intended activity or medical practice. (b) If you experience or are likely to experience such discomfort, please refrain from using this device. (c) Generally, discomfort (such as eye strain, fatigue, nausea, or motion sickness) can be provoked by such factors as quick movements or shakiness of video picture, focal position of video pictures, distance between objects and image capturing modules, user’s point of gaze in video pictures, other varying conditions of video pictures to be input to this device, and individual user’s health conditions. Before you utilize this unit, check if the image of the connected medical device is displayed properly on the screen of this unit.
Prevent internal heat build-up allowing adequate air
circulation. Do not place the unit on surfaces (rugs, blankets, etc.) or near materials (curtains, draperies) that may block the ventilation holes. Do not install the unit near heat sources such as radiators or air ducts, or in a place subject to direct sunlight, excessive dust, mechanical vibration or shock. Do not place the monitor near equipment which generates magnetism, such as a transformer or high voltage power lines.
On simultaneous use with an electrosurgical knife, etc. If this unit is used together with an electrosurgical knife, etc., the picture may be disturbed, warped or otherwise abnormal as a result of strong radio waves or voltages from the device. This is not a malfunction. When you use this unit simultaneously with a device from which strong radio waves or voltages are emitted, confirm the effect of this before using such devices, and install this unit in a way that minimizes the effect of radio wave interference.
Precautions for using this unit safely Some people may experience discomfort (such as eye
strain, fatigue, or nausea) while watching video images. Sony recommends that all viewers take regular breaks while watching video images. The length and frequency of necessary breaks will vary from person to person. You must decide what works best. If you experience any discomfort, you should stop watching the video images until the discomfort ends; consult a doctor if you believe necessary. Avoid watching the display in environments where your head may shake, or while you are walking or performing exercise, because there is a higher possibility that you experience discomfort.
10
Precaution
Recommendation to Use more than One Unit As problems can occasionally occur for the monitor, when the monitor is used for safety control of personnel, assets or stable picture, or for emergencies, we strongly recommend you use more than one unit or prepare a spare unit.
LCD image display Due the physical characteristics of LCD panels, there may be a decrease in brightness or change in color
temperature over a long period of use. These problems are not a malfunction. In addition, these occurrences will not affect recorded data.
About the LCD Display Panel The LCD panel fitted to this unit is manufactured with
high precision technology, giving a functioning pixel ratio of at least 99.99%. Thus a very small proportion of pixels may be “stuck”, either always off (black), always on (red, green, or blue), or flashing. In addition, over a long period of use, because of the physical characteristics of the liquid crystal display, such “stuck” pixels may appear spontaneously. These problems are not a malfunction. Do not leave the LCD screen facing the sun as it can damage the LCD screen. Take care when you place the unit by a window. Do not push or scratch the LCD screen. Do not place a heavy object on the LCD screen. This may cause the screen to lose uniformity. If the unit is used in a cold place, a residual image may appear on the screen. This is not a malfunction. When the monitor becomes warm, the screen returns to normal. The screen and the cabinet become warm during operation. This is not a malfunction.
To prevent any of the above issues, we recommend reducing brightness slightly, and to turn off the power whenever the unit is not in use.
On Burn-in For LCD panel, permanent burn-in may occur if still images are displayed in the same position on the screen continuously, or repeatedly over extended periods. Images that may cause burn-in Masked images with aspect ratios other than 17:9 for the LMD-X310MT and 16:9 for the LMD-X550MT Color bars or images that remain static for a long time Character or message displays that indicate settings or the operating state To reduce the risk of burn-in Turn off the character displays Press the MENU button to turn off the character displays. To turn off the character displays of the connected equipment, operate the connected equipment accordingly. For details, refer to the operation manual of the connected equipment. Turn off the power when not in use Turn off the power if the monitor is not to be used for a prolonged period of time.
Handling the 3D Glasses About the Screen Protect Panel The screen protect panel is made of toughened glass, but there is a possibility that it may crack. Handle with care. Avoid strong impact, such as dropping from a high place. Do not damage the panel with a sharp object. The glass may crack due to the damage.
On a Long Period of Use Due to the characteristics of LCD panel, displaying static images for extended periods, or using the unit repeatedly in a high temperature/high humidity environments may cause image smearing, burn-in, areas of which brightness is permanently changed, lines, or a decrease in overall brightness. In particular, continued display of an image smaller than the monitor screen, such as in a different aspect ratio, may shorten the life of the unit. Avoid displaying a still image for an extended period, or using the unit repeatedly in a high temperature/high humidity environment such an airtight room, or around the outlet of an air conditioner.
For the method and cautions of handling the supplied
3D Eye Shield Kit (CFV-E30SK), see the instructions for use of CFV-E30SK. Do not touch the lens surface of the 3D glasses. Do not leave the 3D glasses in ambient hightemperature, such as near heating equipment or inside a car. Do not put extra pressure on the 3D glasses to avoid them becoming misshapen. Make sure hard accessories or buckles not to touch the lens surface of 3D glasses while holding or transporting. Avoid wearing the 3D glasses when they are aging, breaking or being damaged. Tiny scratches on the lens surface may interfere with your viewing enjoyment. Laying down or looking away from the screen will lessen the 3D effect or shifts image colors.
On Fan Error The fan for cooling the unit is built in. When the fan error indication appears on the screen, turn off the power and contact an authorized Sony dealer.
Precaution
11
On Temperature Error When this unit is used in a high temperature environment and the internal temperature rises, an error is displayed on the screen. When the error is displayed, contact an authorized Sony dealer.
On Moisture Condensation If the unit is suddenly taken from a cold to a warm location, or if ambient temperature suddenly rises, moisture may form on the outer surface of the unit and/ or inside of the unit. This is known as condensation. If condensation occurs, turn off the unit and wait until the condensation clears before operating the unit. Operating the unit while condensation is present may damage the unit.
On Cleaning For the method and cautions of handling the supplied 3D Eye Shield Kit (CFV-E30SK), see the instructions for use of CFV-E30SK. Before cleaning Be sure to disconnect the AC power cord from the AC outlet. On cleaning the monitor and 3D glasses A material that withstands disinfection is used for the front protection plate of the medical use LCD monitor. The protection plate surface is specially treated to reduce light reflection, as are the 3D glasses. When solvents such as benzene or thinner, or acid, alkaline or abrasive detergent, or chemical cleaning cloth are used for the protection plate surface/monitor surface/3D glasses, the performance of the monitor/3D glasses may be impaired or the finish of the surface may be damaged. Take care with respect to the following: Clean the protection plate surface/monitor surface/3D glasses with a 50 to 70 v/v% concentration of isopropyl alcohol or a 76.9 to 81.4 v/v% concentration of ethanol using a swab method. Wipe the protection plate surface gently (wipe using less than 1 N force). Stubborn stains may be removed with a soft cloth such as a cleaning cloth lightly dampened with mild detergent solution using a swab method and then clean using the above chemical solution. Never use solvents such as benzene or thinner, or acid, alkaline or abrasive detergent, or chemical cleaning cloth for cleaning or disinfection, as they will damage the protection plate surface/monitor surface/3D glasses. Do not use unnecessary force to rub the protection plate surface/monitor surface/3D glasses with a
12
Precaution
stained cloth. The protection plate surface/monitor surface/3D glasses may be scratched. Do not keep the protection plate surface/monitor surface/3D glasses in contact with a rubber or vinyl resin product for a long period of time. The finish of the surface may deteriorate.
On Repacking Do not throw away the carton and packing materials. They make an ideal container which to transport the unit. If you have any questions about this unit, contact your authorized Sony dealer.
Disposal of the Unit Do not dispose of the unit with general waste. Do not include the monitor with household waste.
About this manual The instructions in this manual are for the following models: LMD-X310MT LMD-X550MT The illustration of LMD-X310MT is used for the explanations. Any differences in specifications are clearly indicated in the text.
Features Compliance with medical safety standards in U.S.A., Canada and Europe
The blur can be compensated for by using the A.I.M.E. Structure Function at the same time. Users can select the degree of the structure enhancement from levels S1 (minimum) to S4 (maximum) for this purpose according to their preference.
IEC 60601-1 and product safety standards in the U.S.A., Canada and Europe have been obtained for this monitor. The monitor is designed for use in the medical treatment field, with the sheet switch, screen protect panel, etc.
A.I.M.E. is a registered trademark of Sony Corporation. A.I.M.E. is an optional function for enhancing the color or structure of the displayed image. Like the other settings of the monitor, users must adjust or deactivate A.I.M.E. when appropriate.
High brightness/high-resolution 4K panel
Screen protect panel
A 4K high-resolution panel and high brightness/ultrawide field of view technology enables you to use the monitor under various lighting conditions and in numerous ways (installing on the wall, using several monitors to view an image, etc.). Because a color filter with wide-color reproduction and LCD materials with high response speed are used, the motion picture of the video signal is displayed more clearly.
A.I.M.E. (Advanced Image Multiple Enhancer) function
OptiContrast panel The “OptiContrast” design that fills up the air layer between the LCD panel and screen protect panel with bonding resin is employed. By eliminating the air layer, this panel prevents diffused reflection and decrease in contrast. It also prevents condensation due to rapid temperature change.
There are two modes in the A.I.M.E. function – the “Structure Enhancement mode” and “Color Enhancement mode.” Users can select four levels for the Structure Enhancement mode and eight levels for the Color Enhancement mode, depending on their preference. By utilizing the A.I.M.E. function, users can expect the following effects on displayed images.
Control panel
Color Enhancement Function
This unit is a high definition and high performance multi-format LCD monitor compatible with the micropolarizer 3D. Exclusive 3D glasses optimized for this unit are supplied considering long-time work. You can use it for both 2D and 3D with the 2D/3D selection function.
This function clarifies color tone differences between objects. Structure Enhancement Function
This function improves recognition of the outline of objects. Visibility becomes clear making it easy to see objects. The A.I.M.E. Color Enhancement Function emphasizes the contrast of R, G, B signals automatically but does not allow users to select a specific color. Users can only select the degree of contrast, which increases from levels C1 (minimum) to C8 (maximum).
Assigns functions frequently used during an operation to buttons on the front surface of the monitor. The panel provides an user interface superior in operability through navigation by luminescent colors and status of the buttons.
3D
Flat surface for better maintenance The design allows the user to easily wipe liquids and gel off the LCD panel and control buttons – ensuring a high standard of disinfection and cleanliness.
The A.I.M.E. Structure Enhancement Function emphasizes the outline and edges of objects. Users can select the degree of the structure enhancement level, which increases from levels S1 (minimum) to S4 (maximum). The image may seem blurred by applying the A.I.M.E. Color Enhancement Function.
Features
13
Grip that is easy to hold The easy-to-grip ergonomic design enables the user to make tiny adjustments simply and quickly.
Simple connector notation All the connectors face downwards, allowing for easy and organized cable connection.
Cover designed for attachment
14
Features
Location and Function of Parts and Controls Front Panel
Power indicator When the power is turned on, the power indicator lights in green. When the protection function is activated and the display brightness is reduced due to the rise in temperature, it flashes in amber. CONTROL button Displays or clears the operating buttons on the front panel. Selects the items depending on the menu types. OSD menu operation buttons MENU button Press to display the on-screen menu. Press again to clear the menu. /// buttons Press to select the items and setting values. 2D/3D select button Switches between 2D and 3D displays. CUSTOM button
Turns on or off the assigned function. You can adjust the assigned function by pressing the / buttons. (Refer to the custom buttons of the system configuration menu on page 30 and of the default setting on page 31.) The following functions are assigned in the default setting. CUSTOM 1: Brightness CUSTOM 2: Contrast A.I.M.E. button Sets the A.I.M.E. 1) function to Off and On. When the A.I.M.E. function is set to On, A.I.M.E. Structure and A.I.M.E. Color can be adjusted. 1) A.I.M.E.: Enables better image reproduction and display.
2 or 3 screen display setting buttons PIP/POP: For displaying the 2 or 3 screen display or switching the 2 or 3 screen display mode. SWAP: For swapping between the main display and the sub display. Input select buttons PORT A: Displays the input signal assigned to PORT A. When the input signal of PORT A is
Location and Function of Parts and Controls
15
displayed, a menu which allows you to select the input signal to be assigned to PORT A is displayed. PORT B: Displays the input signal assigned to PORT B. When the input signal of PORT B is displayed, a menu which allows you to select the input signal to be assigned to PORT B is displayed.
(ON/STANDBY) switch Press the side to turn the monitor on. Press the side to switch the unit to the standby state, and the indicator light turns off.
Input Signals and Adjustable/Setting Items Input signal Item
SDI 1
SDI 2
Video
PC
× × 4) × 4)
1) 2) 3) 3)
RGB Range
×
×
Color Space
1) 2) 3) 3)
×
×
2) 3) 3)
2) 3) 3)
Phase Chroma Brightness Contrast A.I.M.E. Gradation Extension Color Temperature Mono Sharpness H Sharpness V
4K Scan Size HD Scan Size SD Scan Size Zoom Flip Pattern SD Aspect Interface Mode 2D/3D Select 3D Signal Format 3D Disparity L/R Priority
×
×
Reflected on the screen only when the 4K signal is input. Reflected on the screen only when the HD signal is input. Reflected on the screen only when the SD signal is input. The setting value can be changed but is not applied to the screen when the PC signal is input.
Location and Function of Parts and Controls
× 4) ×
: Can be adjusted/set × : Cannot be adjusted/set 1) 2) 3) 4)
HDMI PC
Gamma
16
DVI-D Video
× 4) × 4) × 4)
×
× 4) ×
Rear/Bottom Panel
LMD-X310MT
LMD-X550MT
HDMI cable holder Secures the HDMI cable (Ø7 mm or less). Close
Cable
HDMI input connector Input connector for HDMI signals. Use HighSpeed HDMI Cable with Cable Type Logo. (Recommend Sony Cable.)
Signals input to the input connector are output to the output connector. 1) Serial digital component signals are SMPTE ST 259 compliant (SD)/SMPTE ST 292-1 compliant (HD)/ SMPTE ST 424 compliant (3G). 2) Transmission distance 3G-SDI: 70 m (approx. 230 ft) max. (When using 5C-FB coaxial cables (Fujikura. Inc.) or equivalent.) HD-SDI: 100 m (approx. 328 ft) max. (When using 5C-FB coaxial cables (Fujikura. Inc.) or equivalent.) SD-SDI: 200 m (approx. 656 ft) max. (When using 5C-2V coaxial cables (Fujikura. Inc.) or equivalent.)
3G/HD/SD-SDI input/output connector (BNC type) SDI 1 (4K/HD) connector (input) Input connector for serial digital component signals 1) 2). connector (output)
Location and Function of Parts and Controls
17
For the HD/SD quad view setting, each connector is compatible with the four division-images shown below.
connector (input) Input DVI Rev.1.0 applicable digital signal. connector (output) Active through output connector of the DVI digital signal. Note
Signals are output when the power is on. No signal is output when the power is off. Signals protected with HDCP 1) cannot be output. 3G/HD/SD-SDI input/output connector (BNC type) SDI 2 (HD) connector (input) Input connector for serial digital component signals 1) 2). connector (output) Signals input to the input connector are output to the output connector. 1) Serial digital component signals are SMPTE ST 259 compliant (SD)/SMPTE ST 292-1 compliant (HD)/ SMPTE ST 424 compliant (3G). 2) Transmission distance 3G-SDI: 70 m (approx. 230 ft) max. (When using 5C-FB coaxial cables (Fujikura. Inc.) or equivalent.) HD-SDI: 100 m (approx. 328 ft) max. (When using 5C-FB coaxial cables (Fujikura. Inc.) or equivalent.) SD-SDI: 200 m (approx. 656 ft) max. (When using 5C-2V coaxial cables (Fujikura. Inc.) or equivalent.)
SERIAL REMOTE connector (RJ-45) Connect to the network by using a 10BASE-T/100BASETX LAN cable (non-shielded type of 5 or more category, optional). The monitor can be operated according to control commands sent from connected external equipment. Consult your dealer in order to utilize the function.
1) HDCP (High-bandwidth Digital Content Protection) is a copyright protection technology using encryption technology of digital video signals.
SERIAL REMOTE RS-232C connector (D-sub 9-pin, female) The monitor can be operated according to control commands sent from connected external equipment.
(DC 26 V input) connector (LMD-X310MT only) Connects the DC connector of the supplied AC adaptor.
Connect the DC connector to the unit, then connect the power cord of the AC adaptor. To remove the DC connector, remove the power cord of the AC adaptor, then remove the DC connector.
For the DC power supply, make sure to use the supplied AC adaptor, AC-300MD. If another power supply is used, there is a risk of fire or electric shock. (equipotential) terminal Connects the equipotential plug. (AC input) connector (LMD-X550MT only) Connects the supplied AC power cord.
For safety, do not connect the connector to peripheral
device wiring that might have excessive voltage. Follow the instructions for use for this port. The connection speed may be affected by the network system. This unit does not guarantee the communication speed or quality of 10BASE-T/ 100BASE-TX. 12V 1.7A output connector Outputs 12 volts. (A connector for an option for future use) 5V 1.6A output connector Outputs 5 volts. (A connector for an option for future use) DVI-D input/output connector
18
Location and Function of Parts and Controls
Do not come into contact with the terminals of the rear panel connectors and patients at the same time. Doing so may result in a generation of voltage that can be harmful to patients if the unit is malfunctioning. Always disconnect the power cord before connecting and disconnecting connectors.
Using this unit for medical purposes This equipment’s connectors are not isolated. Do not connect any device other than one which conforms to IEC 60601-1.
When an information technology device or AV device that uses an alternating current is connected, current leakage may result in an electric shock to the patient or operator. If use of such a device is unavoidable, isolate its power supply by connecting an isolation transformer, or by connecting an isolator between the connecting cables. After implementing these measures, confirm that the reduced risk now conforms to IEC 60601-1.
Connecting the AC Power Cord LMD-X310MT
1
Make sure that the standby state).
2
Insert the DC connector to the DC 26 V input connector of the rear panel until it locks.
3
Plug the AC power cord into the AC IN connector of the AC adaptor and attach the AC power plug holder to the AC power cord.
switch is set to
(in a
AC IN connector
AC power cord
AC power plug holder
4
Insert the AC power plug holder until the fixing lever locks.
Connecting the AC Power Cord
19
5
Connect the plug of the AC power cord to an AC outlet.
To remove the AC power cord Press the switch to the side and switch the unit to the standby state, then remove the AC power cord from the AC power plug holder by holding the fixing levers of the holder from both sides to release the lock. Then, remove the DC connector from the unit.
LMD-X550MT
1
Make sure that the standby state).
2
Insert the AC power cord to the AC input connector of the rear panel and attach the AC power plug holder to the AC power cord.
switch is set to
(in a
AC input connector
AC power cord
AC power plug holder
3
20
Insert the AC power plug holder until the fixing lever locks.
Connecting the AC Power Cord
4
Connect the plug of the AC power cord to an AC outlet.
To remove the AC power cord Press the switch to the side and switch the unit to the standby state, then remove the AC power cord from the AC power plug holder by holding the fixing levers of the holder from both sides to release the lock.