Stryker
Beds, Chairs, Stretchers and Stools
Stair-PRO Model 625x Operations and Maintenance Manual Rev E.2 March 2011
Operations and Maintenance Manual
82 Pages
Preview
Page 1
Stair-PRO® Stair Chair Model 6250/6251/6252
Operations/Maintenance Manual
Model 6250
Model 6251
Model 6252 For parts or technical assistance call: 1-800-327-0770 2011/03
6250-001-160 REV E.2
www.stryker.com
Table of Contents Symbols and Definitions ... 5 Symbols... 5 Warning / Caution / Note Definition... 5 Introduction... 6 Specifications... 6 Warranty... 8 Stryker EMS Return Policy ... 9 Return Authorization... 9 Damaged Merchandise... 9 International Warranty Clause... 9 Patent Information... 9 Summary of Safety Precautions... 10 Setup Procedures... 11 Operation Guide ... 12 Operating Guidelines ... 12 Unfolding the Chair... 13 Folding the Chair ... 14 Transferring the Patient to the Stair-PRO® Stair Chair... 15 Using Restraint Straps ... 16 Proper Lifting Techniques... 19 Transporting the Patient on Flat Surfaces ... 20 Transporting the Patient Down Stairs... 21 Transporting the Patient Up Stairs... 24 Operating the Wheel Locks... 25 Adjusting the Wheel Locking Force... 26 Operating the Optional Locking Rear Lift Handles ... 27 Using the Optional Head Support (Model 6252 Only)... 28 Using Additional Assistance ... 29 Cleaning... 30 Washing Procedure... 30 Washing Limitations ... 30 Preventative Maintenance ... 32 Checklist ... 32 Maintenance Record... 33 Training Record... 34 Quick Replacement Part List... 35 Service Information... 36 Vinyl Seat Replacement ... 36 Track Belt Replacement (For Model 6252)... 37 Track Belt Reconditioning (For Model 6252) ... 40 Rear Wheel Replacement... 42 Upper Control Handle Cable Replacement (For Model 6252)... 43
www.stryker.com
6250-001-160 REV E
3
Table of Contents Chair Assembly... 44 6250/6251 Main Frame Assembly ... 49 6252 Main Frame Assembly... 50 Upper Control Handle Assembly - 6252-001-012... 53 6252 Track Frame Assembly, Standard Belt... 54 Internal Latch Assembly... 56 Internal Latch Assembly... 57 Cable and Pin Assembly - 6252-001-016 ... 58 Wheel Assembly - 6060-002-010... 59 6250/6251/6252 Locking Flip-Up Handle Option... 60 6250 Foot Tube Assembly - 6250-001-012... 61 6251/6252 Foot Support Assembly - 6251-001-012... 62 6250/6251/6252 Standard Lower Lift Handle/No Foot Rest... 63 Extension Handle Assembly, Right - 6250-001-014... 65 Extension Handle Assembly, Left - 6250-001-015... 66 6251/6252 Foot Rest Option... 67 6252 Extended Handle/No Foot Rest Option... 68 6252 Extended Handle/Foot Rest Option ... 69 Seat Section Assembly - 6250-001-013 ... 70 Flip-Up Handle Assembly - 6250-001-016... 71 Lock Mechanism Assembly - 6250-001-017... 72 6250/6251/6252 Vinyl Seat Option... 73 6250/6251/6252 ABS Seat Option ... 74 Seat and Back Rest Options ... 75 6252 Head Support Options... 76 Restraint Set, Non-Absorbent/Metal Buckles Option... 77 Restraint Set, Polyester/Metal Buckles Option ... 77 Restraint Set, Polypropylene/Plastic Buckles Option... 78 Optional Vinyl Head Support - 6252-001-140 ... 79 Elastic Oxygen Bottle Holder - 6250-140-000... 80
4
6250-001-160 REV E
www.stryker.com
Symbols and Definitions SYMBOLS Warning/Caution: consult accompanying documentation
Safe Working Load
WARNING / CAUTION / NOTE DEFINITION The words WARNING, CAUTION and NOTE carry special meanings and should be carefully reviewed.
WARNING Alerts the reader about a situation which, if not avoided, could result in death or serious injury. It may also describe potential serious adverse reactions and safety hazards.
CAUTION Alerts the reader of a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property. This includes special care necessary for the safe and effective use of the device and the care necessary to avoid damage to a device that may occur as a result of use or misuse. Provides special information to make maintenance easier or important instructions clearer.
Return To Table of Contents www.stryker.com
6250-001-160 REV E
5
Introduction INTRODUCTION This manual is designed to assist you with the operation and maintenance of the Model 6250, 6251, and 6252 StairPRO® Stair Chairs. Read it thoroughly before using the equipment or beginning any maintenance on it.
SPECIFICATIONS
6250
Model
6251
6252
Maximum Load 3
500 lb
228 kg
500 lb
228 kg
500 lb
228 kg
Height
36”
91 cm.
36”
91 cm.
37.5”
95.25 cm.
Width
20.5”
52 cm.
20.5”
52 cm.
20.5”
52 cm.
Depth 2
28”
71 cm.
28”
71 cm.
28”
71 cm.
Folded Depth
8”
20 cm.
8”
20 cm.
8”
20 cm.
Weight
20 lb
9 kg
23 lb
10 kg
31.5 lb
14 kg
Dimensions are measured from the outermost edges of the main frame. Specifications are rounded. Conversions are calculated before rounding. 2 Depth dimensions are measured with extendable handles retracted. 3 Maximum load capacity is total weight distributed in accordance to basic human anatomy. Operators must consider the weight of the patient, equipment, and accessories when determining the total load on the product. 1
Stryker reserves the right to change specifications without notice.
Return To Table of Contents 6
6250-001-160 REV E
www.stryker.com
Introduction COMPONENT IDENTIFICATION
6250
Head End Lift Handles
Serial Number Location
6251
Head End Lift Handles
Serial Number Location
Chair Fold Lock Bar
Chair Fold Lock Bar
Wheel Lock
Wheel Lock
Foot End Lift Handles
Upper Control Handle
Front Caster
Foot End Lift Handles
6252
Stair−TREAD Optional Lift Handles Release Lever Upper Control Handle Release Cable Chair Fold Lock Bar
Stair−TREAD Lock Bar
Wheel Lock Lift Handle Release Button
Lift Handle Release Button Front Caster
Foot End Lift Handles
Return To Table of Contents www.stryker.com
6250-001-160 REV E
7
Warranty Stryker EMS, a division of the Stryker Corporation, offers two distinct warranty options in the United States: One (1) year parts and labor. Under this option, Stryker EMS warrants to the original purchaser that its products should be free from manufacturing non-conformances that affect product performance and customer satisfaction for a period of one (1) year after date of delivery. Stryker’s obligation under this warranty is expressly limited to supplying replacement parts and labor for, or replacing, at its option, any product that is, in the sole discretion of Stryker, found to be defective. Two (2) year parts. Under this option, Stryker EMS warrants to the original purchaser that nonexpendable components of its products should be free from manufacturing non-conformances that affect product performance and customer satisfaction for a period of two (2) years after date of delivery. Stryker’s obligation under this warranty is expressly limited to supplying replacement parts for, or replacing, at its option, any product which is, in the sole discretion of Stryker, found to be defective. Expendable components, i.e. mattresses, restraints, I.V. poles, storage nets, storage pouches, Oxygen straps, and other soft goods, have a one (1) year limited warranty with this option. Under either warranty option, Stryker EMS products are designed for a 7 year expected service life under normal use, conditions, and with appropriate periodic maintenance as described in the maintenance manual for each device. Stryker warrants to the original purchaser that the welds on its EMS products will be free from structural defects for the expected 7 year life of the EMS product as long as the original purchaser owns the product. Original purchasers will also obtain a three (3) year limited parts warranty for the X-frame components of the MX-PRO R3 stretcher provided they also purchase X-frame guards at the time of the original purchase and the guards are installed on the MX-PRO before it is put into service. If Stryker requests, products or parts for which an original purchaser makes a warranty claim, the purchaser shall return the product or part prepaid freight to Stryker’s factory. Any improper use or alteration or repair by unauthorized service providers in such a manner as in Stryker’s judgment affects the product materially and adversely, shall void this warranty. Any repair of Stryker products using parts not provided or authorized by Stryker shall void this warranty. No employee or representative of Stryker is authorized to change this warranty in any way. This statement constitutes Stryker EMS’s entire warranty with respect to the aforesaid equipment. STRYKER MAKES NO OTHER WARRANTY OR REPRESENTATION EITHER EXPRESSED OR IMPLIED, EXCEPT AS SET FORTH HEREIN. THERE IS NO WARRANTY OF MERCHANTABILITY AND THERE ARE NO WARRANTIES OF FITNESS FOR ANY PARTICULAR PURPOSE. IN NO EVENT SHALL STRYKER BE LIABLE HEREUNDER FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING FROM OR IN ANY MANNER RELATED TO SALES OR USE OF ANY SUCH EQUIPMENT.
Return To Table of Contents 8
6250-001-160 REV E
www.stryker.com
Warranty STRYKER EMS RETURN POLICY Cots, Stair Chairs, Evacuation Chairs, Cot Fasteners and Aftermarket Accessories may be returned up to 180 days of receipt if they meet the following guidelines:
Prior to 30 Days • • •
30 day money back guarantee in effect Stryker EMS is responsible for all charges Returns will not be approved on modified items
Prior to 90 Days • •
Product must be unused, undamaged and in the original packaging Customer is responsible for a 10% restocking fee
Prior to 180 Days • •
Product must be unused, undamaged and in the original packaging Customer is responsible for a 25% restocking fee
RETURN AUTHORIZATION Merchandise cannot be returned without approval from the Stryker Customer Service Department. An authorization number will be provided which must be printed on the returned merchandise. Stryker reserves the right to charge shipping and restocking fees on returned items. SPECIAL, MODIFIED, OR DISCONTINUED ITEMS NOT SUBJECT TO RETURN.
DAMAGED MERCHANDISE ICC Regulations require that claims for damaged merchandise must be made with the carrier within fifteen (15) days of receipt of merchandise. DO NOT ACCEPT DAMAGED SHIPMENTS UNLESS SUCH DAMAGE IS NOTED ON THE DELIVERY RECEIPT AT THE TIME OF RECEIPT. Upon prompt notification, Stryker will file a freight claim with the appropriate carrier for damages incurred. Claim will be limited in amount to the actual replacement cost. In the event that this information is not received by Stryker within the fifteen (15) day period following the delivery of the merchandise, or the damage was not noted on the delivery receipt at the time of receipt, the customer will be responsible for payment of the original invoice in full. Claims for any short shipment must be made within thirty (30) days of invoice.
INTERNATIONAL WARRANTY CLAUSE This warranty reflects U.S. domestic policy. Warranty outside the U.S. may vary by country. Please contact your local Stryker Medical representative for additional information.
PATENT INFORMATION Stryker products are covered by one or more of the following patents: United States 5,575,026 6,276,010 6,648,343 6,908,133 5,537,700 6,125,485 6,735,794 7,100,224
6,796,757 7,398,571
D527,103 Other Patents Pending
Return To Table of Contents www.stryker.com
6250-001-160 REV E
9
Summary of Safety Precautions The following is a list of safety precautions that must be observed when operating or servicing this unit. The precautions are repeated throughout the manual, where applicable. Carefully read this list before using or servicing the unit.
WARNING • • • • • • • • • • •
• • •
• • •
Do not modify the Stair-PRO® Stair Chair. Modifying the chair can cause unpredictable operation resulting in injury to the patient or operator. Modifying the chair will also void its warranty. Do not allow untrained helpers to assist in the operation of the Stair-PRO® Stair Chair. Untrained technicians/ helpers can cause injury to the patient or themselves. An unlocked chair can fold during use, causing injury to the patient or operator. Always make sure the chair is locked in the unfolded position before use. To avoid injury, always verify the lift handles are locked in place before using them to lift the chair. Always use all restraint straps to secure the patient on the chair. An unrestrained patient may fall from the chair and be injured. Do not push the Model 6252 with the upper control handle in the fully extended position. Pushing the chair with the handle in the fully extended position may cause the chair to tip when obstacles are encountered. Never leave a patient unattended on the chair or injury could result. Hold the chair securely while a patient is on the chair. The Stair-PRO® Stair Chair is not recommended for use with suspected cervical, spinal, or fracture injuries. To avoid injury, transporting the patient on stairs requires a minimum of two operators. If more people are required to safely control the chair, use the chart on page 29 as reference for proper positioning of the helpers. Only use the wheel locks during patient transfer or without a patient on the chair. Tipping could occur if the chair is moved while wheel locks are applied, resulting in injury to the patient or operator and/or damage to the chair. Never use a wheel lock on a chair with excessively worn wheels. Using a wheel lock on a wheel with less than a 5” diameter (Model 6252) or a 6” diameter (Models 6250/6251) could compromise the holding ability of the wheel lock, possibly resulting in injury to the patient or operator and/or damage to the chair or other equipment. To avoid injury, always verify the Stair-TREAD™ system on the Model 6252 is securely locked in place before transporting the patient. Water, ice and debris on the stairs can affect operator footing and proper operation of the Stair-TREAD™ system. To avoid injury, clear the path or consider an alternate route. Condensation, water, ice and/or debris on the Stair-TREAD™ system can cause unpredictable performance, resulting in a sudden change in the weight the operators must support. To avoid injury, and to aid proper operation of the Stair-TREAD™ system, ensure the belts are clean and dry before transporting the patient. Never lubricate the Stair-TREAD™ system. Lubrication on the system can cause inconsistent operation possibly resulting in injury to the patient or operator. To avoid injury to the operators and/or the patient, operators should never attempt to transport patient loads greater than what they can safely lift. Use any appropriate personal safety equipment (goggles, respirator, etc.) to avoid the risk of inhaling contagion. Use of power washing equipment can aerate contamination collected during the use of the cot.
CAUTION • •
• • •
Improper usage of the Stair-PRO® Stair Chair can cause injury to the patient or operator. Operate the Stair-PRO® Stair Chair only as described in this manual. Improper maintenance can cause injury or damage to the unit. Maintain the Stair-PRO® Stair Chair as described in this manual. Use only Stryker approved parts and maintenance procedures. Using unapproved parts and procedures could cause unpredictable operation and/or injury and will void the product warranty. Casters are not suitable for all surfaces. Caution should be used at all times. Wheel locks are only intended to help prevent the empty chair from rolling while unattended and to aid in patient transfer. The wheel lock may not provide sufficient resistance on all surfaces or under loads. Release the red track release bar before clicking the Stair-TREAD™ system into the locked position. Failure to follow this procedure could result in the track failing to lock. Always verify the Stair-TREAD™ system is locked by trying to fold it before descending stairs.
Return To Table of Contents 10
6250-001-160 REV E
www.stryker.com
Setup Procedures Unpack the cartons and check all items for proper operation. It is important that the Stair-PRO® Stair Chair is working properly before it is put into service. Have a qualified service person use the following list and the operation instructions to check the chair before it is put into service. • • • • • • • • • • • • • • • •
All fasteners secure (reference all assembly drawings). All welds intact, not cracked or broken. No bent or broken tubing or sheet metal. No debris in wheels. All wheels secure and rolling properly. Chair unfolds and locks properly. No rips or cracks in seat or backrest. Patient restraints intact and working properly. Wheel locks operating properly. Foot end lift handles extend and lock properly. Head end lift handles fold and unfold. Front casters secure, rolling and swiveling properly (if equipped). Upper control handle extends and locks in all positions (if equipped). Stair-TREAD™ system unfolds and locks (if equipped). Track belts roll properly (if equipped). Optional accessories intact and operating properly.
The storage compartment on the vehicle in which the Stair-PRO® Stair Chair will be transported must be large enough to accommodate the folded dimensions of the chair (see page 6). When necessary, modify the vehicle to fit the chair. Do not modify the chair.
WARNING Do not modify the Stair-PRO® Stair Chair. Modifying the chair can cause unpredictable operation resulting in injury to the patient or operator. Modifying the chair will also void its warranty.
Return To Table of Contents www.stryker.com
6250-001-160 REV E
11
Operation Guide OPERATING GUIDELINES • • •
• • • •
Use the Stair-PRO® Stair Chair only as described in this manual. Read all labels and instructions on the chair before using the chair. When a patient is on the chair, use a minimum of two operators to manipulate the chair on stairs. If more people are required to safely control the chair, use the chart on page 29 as reference for proper positioning of the helpers. Do not roll the chair, ascend, or descend stairs without advising the patient. Stay with the patient and control the chair at all times. Only use the wheel locks during patient transfer or without a patient on the chair. Always use the restraint straps when a patient is on the chair. Use properly trained helpers when necessary to control the chair and patient.
WARNING • • •
•
Always use all restraint straps to secure the patient on the chair. An unrestrained patient may fall from the chair and be injured. Only use the wheel locks during patient transfer or without a patient on the chair. Tipping could occur if the chair is moved while the wheel locks are applied, resulting in injury to the patient or operator and/or damage to the chair. Condensation, water, ice and/or debris on the Model 6252 Stair-TREAD™ system can cause unpredictable performance, resulting in a sudden change in the weight the operators must support. To avoid injury, and to aid proper operation of the Stair-TREAD™ system, ensure the belts are clean and dry before transporting the patient. To avoid injury to the operators and/or the patient, operators should never attempt to transport patient loads greater than what they can safely lift.
Return To Table of Contents 12
6250-001-160 REV E
www.stryker.com
Operation Guide UNFOLDING THE CHAIR
Figure 1a - Unfolding the Chair (6250 Shown) To unfold the chair (refer to Figure 1a above): 1. 2. 3. 4.
Stand behind the chair. Apply the wheel locks (if desired). Pull the backrest and the extension handle apart. The lock mechanism will automatically engage when the chair is completely unfolded. Verify the lock is engaged by pulling up on the seat. If the lock is properly engaged, the chair will not fold.
WARNING An unlocked chair can fold during use, causing injury to the patient or operator. Always make sure the chair is locked in the unfolded position before use.
Return To Table of Contents www.stryker.com
6250-001-160 REV E
13
Operation Guide FOLDING THE CHAIR
Figure 1b - Folding the Chair (6250 Shown) To fold the chair (refer to Figure 1b above): 1.
2. 3. 4. 5.
Apply the wheel locks (if desired). Buckle the restraint straps and fold them neatly to prevent them from interfering with proper folding of the chair. If the chair is equipped with the optional head support strap, secure it behind the chair frame. Stand at the side of the chair. Pull up on the red lock bar at the rear of the chair. Tip the chair forward. Fold the seat up to the backrest until the front legs lock in the clips on the bottom of the seat tube.
Note: On models 6251 and 6252, rotate the front casters so they do not interfere with folding the chair.
WARNING An unlocked chair can fold during use, causing injury to the patient or operator. Always make sure the chair is locked in the unfolded position before use.
Return To Table of Contents 14
6250-001-160 REV E
www.stryker.com
Operation Guide TRANSFERRING THE PATIENT TO THE STAIR-PRO® STAIR CHAIR To transfer the patient to the chair: 1. 2. 3. 4. 5. 6.
Place the chair beside the patient. Apply the wheel locks to prevent the chair from moving. Open the restraint straps. Transfer the patient to the chair using accepted EMS procedures. Use all the restraints to secure the patient on the chair (page 16). Disengage the wheel locks before transporting.
WARNING The Stair-PRO® is not recommended for use with suspected cervical, spinal, or fracture injuries.
Return To Table of Contents www.stryker.com
6250-001-160 REV E
15
Operation Guide USING RESTRAINT STRAPS
Vinyl Seat Top View
Figure 2 Cross-Chest Configuration
Figure 3 Chest and Lap Strap Configuration
ABS Seat Top View
Figure 4 Lap Strap Attachment Points
WARNING Always use all restraint straps to secure the patient on the chair. An unrestrained patient may fall from the chair and be injured. The restraints can be used in two configurations. For the cross-chest configuration (Figure 2): Buckle the chest restraints in an “X” pattern across the patient’s chest. For the chest and lap configuration (Figure 3): Buckle one restraint across the patient’s chest and the other across the patient’s lap. For either configuration: Match up the color coded restraints and buckle the ankle restraint across the patient’s legs. To avoid damage to the buckles and straps, keep the restraint straps buckled when the chair is not being used with a patient. When attaching the restraint straps to the chair, remember the attachment points must provide strong anchorage and proper restraint position while not interfering with equipment and accessories.
Return To Table of Contents 16
6250-001-160 REV E
www.stryker.com
Operation Guide USING RESTRAINT STRAPS - CONTINUED To attach the chest straps in the cross-chest configuration: 1. 2. 3. 4. 5.
Wrap each strap around the chair frame, insert the end through the loop on the end of the strap and pull it tight. Pull the strap across the patient’s chest, lengthening the strap as necessary. Buckle the strap. Pull the loose end of the strap to tighten it securely around the patient. Repeat for the second strap and the lap belt.
Figure 7 - Lengthen strap as necessary
Figure 5 - Insert the end through the loop
Figure 8 - Buckle strap
Figure 6 - Pull the strap tight
Figure 9 - Tighten strap securely Return To Table of Contents
www.stryker.com
6250-001-160 REV E
17
Operation Guide USING RESTRAINT STRAPS - CONTINUED To attach the ankle strap: 1. 2.
Remove the three plastic pieces from the strap. Thread the strap through one of the plastic “D” rings and slide the ring down to the end of the strap up to the white label.
3.
Loop the strap around the front leg of the chair and pull the loose end of the strap through the ring. Pull until the ring rests against the foot rest tube.
4.
Thread the strap through the other “D” ring. Loop the strap around the other front leg of the chair and pull the loose end of the strap through the loop. Pull until the “D” ring rests against the foot rest tube.
5.
6.
Attach the final “male” end clip on the strap.
Figure 10-18 - Attaching the leg strap Return To Table of Contents 18
6250-001-160 REV E
www.stryker.com
Operation Guide USING RESTRAINT STRAPS - CONTINUED •
•
• • •
To lengthen the restraint, grasp the buckle, turn it at an angle to the webbing and pull it out. A hemmed tab at the end of the webbing prevents the buckle from coming off the strap. To shorten the restraint, grasp the hemmed tab and pull the webbing back through the buckle until the required tightness is achieved. When the chair is put into service, open the restraints and place them at either side of the chair until the patient is positioned on the seat. Lengthen the restraint, buckle it around the patient and shorten it until the required tightness to properly secure the patient is achieved. To open the restraint, press the tabs on the side of the buckle to release the buckle and pull the tang out of the receiver. To close the restraint, push the tang into the receiver until a ”click” is heard. Whenever a restraint is buckled on a patient, the attendant should verify the tang is locked and the extra webbing is not tangled in the chair or hanging loose. Inspection of the restraints should be done at least once a month (more frequently if used heavily). Inspection should include checking for a bent or broken receiver or tang, torn or frayed webbing, etc. Any restraint showing wear or not operating properly must be replaced immediately.
PROPER LIFTING TECHNIQUES When lifting the Stair-PRO® Stair Chair and patient, remember these five basic guidelines: • • • • •
Keep your hands close to your body. Keep your back straight. Coordinate your movements with your partner and lift with your legs. Avoid twisting. Always operate the Stair-PRO® Stair Chair as described in this manual.
Return To Table of Contents www.stryker.com
6250-001-160 REV E
19
Operation Guide TRANSPORTING THE PATIENT ON FLAT SURFACES
Figure 19 - Transporting the Patient on a Model 6250
Figure 20 - Transporting the Patient on a Model 6252
CAUTION Casters are not suitable for all surfaces. Caution should be used at all times. Model 6250 To roll the Model 6250 Stair-PRO® over flat surfaces, use either the locking rear lift handles (if equipped), or the backrest tube to tip the chair back. Once the chair is tipped back, use either the head end lift handles or the backrest tube to roll and guide the chair. Lift the chair over and around obstructions with the head end and foot end lift handles. Model 6251 To roll the Model 6251 Stair-PRO® over flat surfaces, push and guide the chair from the rear of the patient, using either the head end lift handles or the backrest tube. Lift the chair over and around obstructions with the head end and foot end lift handles. Model 6252 To roll the Model 6252 Stair-PRO® over flat surfaces, follow the same method as above for the Model 6251. In addition to the head end lift handles, the extendable upper control handle can be used in any position to roll and guide the chair. Extend the upper control handle by pulling the red release cable with one hand, and pulling up on the control handle with the other. Release the cable to lock the handle in either the intermediate or fully extended position. Lift the chair over and around obstructions with the head end and foot end lift handles.
WARNING Do not push the Model 6252 with the upper control handle in the fully extended position. Pushing the chair with the handle in the fully extended position may cause the chair to tip when obstacles are encountered.
Return To Table of Contents 20
6250-001-160 REV E
www.stryker.com
Operation Guide TRANSPORTING THE PATIENT DOWN STAIRS
Figure 21 - Foot end lift handles
Figure 22 - Head end lift handles
WARNING To avoid injury, transporting the patient on stairs requires a minimum of two operators. If more people are required to safely control the chair, use the chart on page 29 as reference for proper positioning of the helpers. To avoid injury, always verify the lift handles are locked in place before using them to lift the chair. Models 6250 and 6251 1. Foot end operator - descend in front of the Stair-PRO® Stair Chair. 2. Head end operator - approach the stairs squarely. 3. Foot end operator - push each red foot end lift handle button and pull out each handle until it stops. Release the button and verify the handle is locked securely in position. 4. Head end operator - unfold the head end lift handles and face down the stairs. 5. Foot end operator - either face backward for improved patient monitoring or forward for an improved view of the stairs and easier maneuvering around obstacles. Note: Although the front operator can face either direction while lifting and carrying, any applicable protocols for carrying chairs should be followed.
Figure 23 and 24 - Transporting down stairs 6.
Both operators - simultaneously lift the chair, using the head and foot end lift handles and following the proper lifting techniques (see page 19). Carry the chair slowly down the stairs, avoiding any obstructions. Return To Table of Contents
www.stryker.com
6250-001-160 REV E
21