Stryker
Accessories for Use with Drivers TPS, CORE and RemB Instructions for Use Rev N July 2009
Instructions for Use
85 Pages
Preview
Page 1
NOTE: This portion of the document should not appear
Dimensions:
Color/Material/Finish:
Label Stock:
5 in. x 7 in. Booklet
Black Graphics on White 20# Bond
NA
on produced Labels or IFUs.
Description/Type:
Instructions For Use Stryker Instruments (269) 323-7700 (800) 253-3210
Print Location:
Suppliers/Services:
Print Center
Crimson Life Sciences
DSGN-fm-50366 rev. None Effective: July 1, 2008
Part Number:
Rev.
4100-005-700
N
Page 1 of 1
Accessories For Use With Cordless Drivers REF 4100-000-000, 4101-000-000, 4200-000-000, 4300-000-000
TPS Universal Driver TM
REF 5100-099-000
CORE Universal Driver TM
REF 5400-099-000
RemB Electric Universal Driver TM
REF 6400-099-000
Instructions for Use
ENGLISH ESPAÑOL DEUTSCH FRANÇAIS ITALIANO NEDERLANDS SVENSKA DANSK SUOMI PORTUGUÊS NORSK POLSKI ΕΛΛΗΝΙΚΑ 日本語
中文 한국어
2009/07
4100-005-700 Rev-N
www.stryker.com
EN
4100-005-700 Rev-N
User/Patient Safety* WARNINGS: • Before attaching or removing any accessory, or before passing the handpiece to another person, always put the safety switch in the safety position to prevent inadvertent running of the handpiece. • Use only Stryker-approved components and accessories, unless otherwise specified. DO NOT modify any component or accessory. *If you need more information, contact your Stryker sales representative or call Stryker customer service at 1-800-253-3210. Outside the US, contact your nearest Stryker subsidiary.
Cleaning and Sterilization Recommendations For cleaning and sterilization information, see the appropriate booklet: • 4100-000-000, 4101-000-000, 4200-000-000 or 4300-000-000 Cordless Driver Instructions For Use • TPS Cleaning, Maintenance and Sterilization Recommendations • CORE Cleaning, Maintenance and Sterilization Recommendations • RemB Cleaning, Maintenance and Sterilization Recommendations
Reorder Information DESCRIPTION
REF
DESCRIPTION
REF
Sagittal Saw
4100-400-000
Hudson/Modified Trinkle Drill
4100-135-000
1/4" Keyed Chuck
4100-131-000
Hudson/Modified Trinkle Reamer
4100-235-000
5/32" Drill
4100-132-000
Right Angle Drive - AO Large Reamer
4100-310-000
1/4" Reamer
4100-231-000
Right Angle Drive Modified Trinkle Reamer
4100-335-000
Wire Collet
4100-062-000
Trinkle Drill
4100-160-000
Pin Collet
4100-125-000
Trinkle Reamer
4100-260-000
Pin Collet
4100-126-000
1/4" Keyless Drill Chuck
4100-133-000
1/8" Keyless Drill Chuck
4100-134-000
Hudson is a trademark and/or registered trademark of Hudson Respiratory Care Co. AO is a trademark and/or registered trademark of AO Technology AG.
2
www.stryker.com
4100-005-700 Rev-N
EN
Accessory Information DESCRIPTION
REF
Blades
2296-003-XXX Series 2296-033-XXX Series
Sagittal Saw Attachment (REF 4100-400-000)
5400-003-XXX Series 5400-134-XXX Series
Operating Instructions WARNINGS: • DO NOT attempt to insert or remove any cutting accessory or attachment while the handpiece is operating. Failure to comply may result in operating room staff injury. • DO NOT apply excessive pressure, such as bending or prying, with the blade. Excessive pressure may bend or fracture the blade and cause tissue damage and/or loss of tactile control. Failure to comply may result in patient and/or operating room staff injury. • Before operating the handpiece, gently tug the blade to verify it is secure. Failure to comply may result in patient and/or operating room staff injury. • Before operating the handpiece, gently rotate the blade mount to verify it is seated properly (should not rotate, fully recessed). Failure to comply may result in excessive handpiece vibration/noise and patient and/or operating room staff injury. NOTE: Operate Sagittal Saw Attachment with lower trigger only. Cutting efficiency is lost if both triggers are depressed to operate this attachment in the oscillate mode.
1. Insert the blade into the blade mount retaining gap.
2. Depress the button to allow the blade to fully enter the gap and seat against the post, then release the button.
3. Gently rotate the blade against the post until it snaps into the desired location. If necessary, depress the button again to rotate the blade to the desired location.
4. To index the blade mount, pull out and slightly rotate the mount until it snaps (recesses) into the desired location. The blade mount will index in 45-degree increments or eight possible locations.
Additional Cleaning Recommendation Pull out and slightly rotate the blade mount to expose the additional surface area around the mount. Using a stiff non-metallic bristle brush, remove debris from the exposed surface area.
www.stryker.com
3
EN
4100-005-700 Rev-N
1/4" Keyless Drill Chuck (REF 4100-133-000)
1/8" Keyless Drill Chuck (REF 4100-134-000)
Handles wires, pins, and drill bits up to 1/4 in. (6.4 mm) diameter.
Handles wires, pins, and drill bits up to 1/8 in. (3.2 mm) diameter.
Operating Instructions Cutting Accessory Insertion WARNING: You must manually close the chuck jaws to secure the accessory. Operating the handpiece to automatically close the jaws on the accessory may result in injury to the user. 1. Open the chuck jaws by holding the attachment with the knurled ring in one hand and turning the collar with the other hand. 2. Insert the cutting accessory and make certain it is centered in the jaws. 3. Close the jaws by holding the collar and turning the knurled ring until tight. 4. Tug the cutting accessory to ensure it is secure. 5. Visually verify that the cutting accessory is centered in the jaws. 6. Install the attachment in the handpiece.
4
www.stryker.com
4100-005-700 Rev-N
EN
1/4" Keyed Chuck (REF 4100-131-000)
1/4" Reamer (REF 4100-231-000)
Handles wires, pins, and drill bits up to 1/4 in. (6.4 mm) diameter.
Handles wires, pins, and drill bits up to 1/4 in. (6.4 mm) diameter with low speed and high torque.
5/32" Drill (REF 4100-132-000)
Chuck style accessories are supplied with a chuck key.
Handles wires, pins, and drill bits up to 5/32 in. (4 mm) diameter.
Wire Collet (REF 4100-062-000) This collet holds K-wires having diameters from 0.028 – 0.071 in. (0.7 – 1.8 mm). Pin Collet (REF 4100-125-000) This collet holds pins having diameters from 0.078 – 0.125 in. (2.0 – 3.2 mm).
To operate the Cordless Driver with the Wire Collet or Pin Collets: 1. Load the wire or pin into the nose of the collet or back cap of the handpiece. 2. Move the safety switch out of the safety position. 3. Squeeze the wire advance control lever against the instrument to hold the wire at the desired length exposed from the distal end of the instrument. 4. Hold the wire or pin against the bone, then depress the trigger to drive the wire or pin. 5. To obtain additional wire length for inserting, release the wire advance control and pull back the instrument. Then, squeeze the wire advance control lever and the trigger to drive more wire. 6. To withdraw threaded pins, squeeze the wire advance control lever and then depress the reverse trigger.
www.stryker.com
5
EN
AO Small Drill (REF 4100-110-000) Accepts Synthes tools such as drill bits, taps and automatic screwdrivers with the appropriate snaplock shank.
AO Large Reamer (REF 4100-210-000) Accepts Synthes tools such as drill bits, automatic screwdrivers, and flexible reamers with the appropriate snap-lock shank. Utilizes low speed and high torque.
4100-005-700 Rev-N
Hudson/Modified Trinkle Drill (REF 4100-135-000) Accepts tools such as drill bits, and automatic screwdrivers with Hudson or modified Trinkle fitting.
Hudson/Modified Trinkle Reamer (REF 4100-235-000) Accepts tools such as drill bits, automatic screwdrivers and flexible reamers with a Hudson or modified Trinkle fitting. Utilizes low speed and high torque.
Right Angle Drive - AO Large Reamer (REF 4100-310-000)
Right Angle Drive Modified Trinkle Reamer (REF 4100-335-000)
Accepts all Synthes tools with the appropriate snap-lock shank.
Accepts tools with modified Trinkle fitting.
Trinkle Drill (REF 4100-160-000)
Trinkle Reamer (REF 4100-260-000)
Accepts tools such as drill bits and automatic screwdrivers with standard Trinkle fitting.
Accepts tools such as drill bits, automatic screwdrivers and flexible reamers with standard Trinkle fitting. Utilizes low speed and high torque.
To retain a bit or tool in accessories with a retaining collar: 1. Slide back the retaining collar. 2. Insert a bit or accessory. 3. Release retaining collar. 4. Twist bit/accessory to check for proper retention before running handpiece.
6
www.stryker.com
Stryker Instruments 4100 E. Milham Kalamazoo, Michigan (USA) 49001 1-269-323-7700 1-800-253-3210
Stryker France ZAC Satolas Green Pusignan Av. de Satolas Green 69881 MEYZIEU Cedex France 2009/07
4100-005-700 Rev-N
www.stryker.com