FRONT Stryker CrossFlow Flash Card 01
Stryker Color Palette
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
QUICK START GUIDE
CROSSFLOW integrated Arthroscopy Pump CROSSFLOW Integrated Arthroscopy Pump
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
BACK Stryker CrossFlow Flash Card 01
Stryker Color Palette
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Interface & Hand Control Map
1 Pump Start Up:
CrossFlow Interface & Hand Control Button Map.
User Interface Reference
Hand Control Button Reference
1.
Surgeon Profiles & Joint Settings
8.
Start/Stop
14. Flow/Suction (%)
1.
Hot Swap Function
2.
Set Pressure (mmHg)
9.
Resection Integration Settings Menu
15. Hot Swap Function
2.
Adjust Set Pressure Up/Down
3.
Adjust Set Pressure Up/Down
10. Wash/Clear Menu
16. Hardware Combination Selection
3.
Start/Stop
4.
Wash (use for better visibility)
11. Close Pump Customization Menu
17. Scope/Cannula Combination in Use
4.
Adjust Flow/Suction Up/Down
5.
Pump Settings
5.
Wash (use for better visibility)
6.
Upload Surgeon Profiles
12. Drain (use post procedure to drain fluid from joint)
7.
Footswitch/Hand Control Settings
13. Adjust Flow/Suction Up/Down
NOTE: Identifies which scope/cannula combo is selected for the procedure 18. Actual Pressure
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
FRONT Stryker CrossFlow Flash Card 01
Stryker Color Palette
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
QUICK START GUIDE
CROSSFLOW integrated Arthroscopy Pump CROSSFLOW Integrated Arthroscopy Pump
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
BACK Stryker CrossFlow Flash Card 01
Stryker Color Palette
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Interface & Hand Control Map
1 Pump Start Up:
CrossFlow Interface & Hand Control Button Map.
User Interface Reference
Hand Control Button Reference
1.
Surgeon Profiles & Joint Settings
8.
Start/Stop
14. Flow/Suction (%)
1.
Hot Swap Function
2.
Set Pressure (mmHg)
9.
Resection Integration Settings Menu
15. Hot Swap Function
2.
Adjust Set Pressure Up/Down
3.
Adjust Set Pressure Up/Down
10. Wash/Clear Menu
16. Hardware Combination Selection
3.
Start/Stop
4.
Wash (use for better visibility)
11. Close Pump Customization Menu
17. Scope/Cannula Combination in Use
4.
Adjust Flow/Suction Up/Down
5.
Pump Settings
5.
Wash (use for better visibility)
6.
Upload Surgeon Profiles
12. Drain (use post procedure to drain fluid from joint)
7.
Footswitch/Hand Control Settings
13. Adjust Flow/Suction Up/Down
NOTE: Identifies which scope/cannula combo is selected for the procedure 18. Actual Pressure
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
FRONT Stryker CrossFlow Flash Card 02
Stryker Color Palette
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Start Up
2 Pump Start Up:
How to start up the pump & navigate to the Run Screen.
CROSSFLOW Integrated Arthroscopy Pump
CROSSFLOW Integrated Arthroscopy Pump
NOTE: “OUTFLOW” indicates the Outflow Cassette is properly inserted.
1
2
Turn on Pump: = Power on Pump
Insert Cassettes:
Wait for Insert Cassette Screen to appear, then align the colored side of the cassette facing out toward its matching colored button. Green checkmarks will indicate that the cassette(s) have been properly inserted. Once the Inflow Cassette is inserted, the pump will advance to the User Profile Selection Screen.
The CrossFlow start up screen will appear and a progress bar will indicate that the pump is starting up.
= Inflow Cassette (Blue)
= Outflow Cassette (Red)
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
BACK Stryker CrossFlow Flash Card 02
Stryker Color Palette
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Start Up Pump Start Up: How to start up the pump & navigate to the Run Screen.
3
4
Select User Profile:
Select the name of the surgeon performing
5
Select Joint:
Press the icon that identifies the operative joint:
the case. If no profile is uploaded for that surgeon, press “Standard Settings”.
Shoulder
Knee
Hip
Small Joint
= Advance to Next Screen
Start Pump:
Once tubing is set up and saline is spiked: = Start/Stop Pump
= Return to User Profile Selection Screen and Start New Case = Return to Previous Screen
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
FRONT Stryker CrossFlow Flash Card 02
Stryker Color Palette
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Start Up
2 Pump Start Up:
How to start up the pump & navigate to the Run Screen.
CROSSFLOW Integrated Arthroscopy Pump
CROSSFLOW Integrated Arthroscopy Pump
NOTE: “OUTFLOW” indicates the Outflow Cassette is properly inserted.
1
2
Turn on Pump: = Power on Pump
Insert Cassettes:
Wait for Insert Cassette Screen to appear, then align the colored side of the cassette facing out toward its matching colored button. Green checkmarks will indicate that the cassette(s) have been properly inserted. Once the Inflow Cassette is inserted, the pump will advance to the User Profile Selection Screen.
The CrossFlow start up screen will appear and a progress bar will indicate that the pump is starting up.
= Inflow Cassette (Blue)
= Outflow Cassette (Red)
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
BACK Stryker CrossFlow Flash Card 02
Stryker Color Palette
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Start Up Pump Start Up: How to start up the pump & navigate to the Run Screen.
3
4
Select User Profile:
Select the name of the surgeon performing
5
Select Joint:
Press the icon that identifies the operative joint:
the case. If no profile is uploaded for that surgeon, press “Standard Settings”.
Shoulder
Knee
Hip
Small Joint
= Advance to Next Screen
Start Pump:
Once tubing is set up and saline is spiked: = Start/Stop Pump
= Return to User Profile Selection Screen and Start New Case = Return to Previous Screen
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
FRONT Stryker CrossFlow Flash Card 03
Stryker Color Palette
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
General Settings
3 Settings Navigation:
How to access and adjust general settings for the CrossFlow pump.
1
2
Access Settings Menu:
3
Adjust General Settings:
Change Pressure Display: = Actual Pressure Display Settings Screen
= Pump Customization Menu = General Settings Screen
Change how pressure displays on the run screen by pressing the desired icon.
= Adjust Volume, Brightness, Language & Perform Software Upgrades
NOTE: This new pressure display will now appear on the Run Screen of the pump.
4
NOTE: Icon will display as ORANGE when selected.
Return to Run Screen: = Exit Menu General settings are now customized.
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
www.stryker.com
= Pump Customization Menu
BACK Stryker CrossFlow Flash Card 03
Stryker Color Palette
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Pressure & Suction/Flow
4 Pressure & Suction/Flow Settings:
How to access & adjust Set Pressure and Suction/Flow settings.
1
2
Start the Pump:
= Set Pressure Up
= Start/Stop Pump
Settings in Detail:
= Set Pressure Down
Will indicate the console is pumping fluid.
Set Pressure
Suction (INFLOW/OUTFLOW MODE)
Flow (INFLOW-ONLY MODE)
Set Pressure is the pump’s target pressure in the joint.
Suction refers to the flow rate the outflow pump will achieve through the outflow cannula, shaver, and RF tube.
Flow is the maximum flow rate the inflow pump will reach while attempting to achieve the Set Pressure.
NOTE: Suction will adjust in increments of 10%
NOTE: Flow will adjust in increments of 10%
NOTE: On the CrossFlow console, pressure is adjusted in increments of 5mmHg
3
Adjust Pressure:
NOTE: Actual Pressure indicates the pressure in the joint.
Adjust Suction/Flow: = Suction/Flow Rate Up = Suction/Flow Rate Down
NOTE: When used in Inflow-only, fluid usage will be read as “Flow”. When used in Inflow/Outflow mode it will be read as “Suction”.
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
FRONT Stryker CrossFlow Flash Card 03
Stryker Color Palette
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
General Settings
3 Settings Navigation:
How to access and adjust general settings for the CrossFlow pump.
1
2
Access Settings Menu:
3
Adjust General Settings:
Change Pressure Display: = Actual Pressure Display Settings Screen
= Pump Customization Menu = General Settings Screen
Change how pressure displays on the run screen by pressing the desired icon.
= Adjust Volume, Brightness, Language & Perform Software Upgrades
NOTE: This new pressure display will now appear on the Run Screen of the pump.
4
NOTE: Icon will display as ORANGE when selected.
Return to Run Screen: = Exit Menu General settings are now customized.
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
www.stryker.com
= Pump Customization Menu
BACK Stryker CrossFlow Flash Card 03
Stryker Color Palette
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Pressure & Suction/Flow
4 Pressure & Suction/Flow Settings:
How to access & adjust Set Pressure and Suction/Flow settings.
1
2
Start the Pump:
= Set Pressure Up
= Start/Stop Pump
Settings in Detail:
= Set Pressure Down
Will indicate the console is pumping fluid.
Set Pressure
Suction (INFLOW/OUTFLOW MODE)
Flow (INFLOW-ONLY MODE)
Set Pressure is the pump’s target pressure in the joint.
Suction refers to the flow rate the outflow pump will achieve through the outflow cannula, shaver, and RF tube.
Flow is the maximum flow rate the inflow pump will reach while attempting to achieve the Set Pressure.
NOTE: Suction will adjust in increments of 10%
NOTE: Flow will adjust in increments of 10%
NOTE: On the CrossFlow console, pressure is adjusted in increments of 5mmHg
3
Adjust Pressure:
NOTE: Actual Pressure indicates the pressure in the joint.
Adjust Suction/Flow: = Suction/Flow Rate Up = Suction/Flow Rate Down
NOTE: When used in Inflow-only, fluid usage will be read as “Flow”. When used in Inflow/Outflow mode it will be read as “Suction”.
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
FRONT Stryker CrossFlow Flash Card 01
Stryker Color Palette
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Wash Function
5 Wash Function:
How to access & adjust the Wash function.
1
2
Access Settings Menu:
Adjust Wash Settings:
3
= Pump Customization Menu
Start Pump to Perform Wash: = Wash Function
Understanding Wash:
= Wash Settings Screen
Wash Function
= Increase Pressure/Flow/Duration Settings
The Wash function allows for a customized, quick increase in pressure and flow to clear debris, control bleeding, and improve visualization in the joint. Can customize to last for a duration up to 120 seconds.
= Decrease Pressure/Flow/Duration Settings = Exit Setting Menu
EXAMPLE: 50% Pressure increase of 50 mmHg set pressure would become 75mmHg during Wash cycle.
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
BACK Stryker CrossFlow Flash Card 01
Stryker Color Palette
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Clear Function Clear Function: How to access & adjust the Clear function.
1
3
Access Settings Menu: = Pump Customization Menu
2
Adjust Clear Settings: = Clear Settings Screen
4
Start Pump to Perform Clear: = Clear can only be activated by footswitch
= Increase Pressure/Flow/Duration Settings
Adjust Clear Settings:
= Decrease Pressure/Flow/Duration Settings = Exit Setting Menu
Understanding Clear: Clear Function The Clear function is designed for quick dismissal and flushing of debris from the joint. It increases flow only and does not increase pressure.
= Wash/Clear Settings Screen
NOTE: Can customize to last for a duration up to 120 seconds.
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
FRONT Stryker CrossFlow Flash Card 01
Stryker Color Palette
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Wash Function
5 Wash Function:
How to access & adjust the Wash function.
1
2
Access Settings Menu:
Adjust Wash Settings:
3
= Pump Customization Menu
Start Pump to Perform Wash: = Wash Function
Understanding Wash:
= Wash Settings Screen
Wash Function
= Increase Pressure/Flow/Duration Settings
The Wash function allows for a customized, quick increase in pressure and flow to clear debris, control bleeding, and improve visualization in the joint. Can customize to last for a duration up to 120 seconds.
= Decrease Pressure/Flow/Duration Settings = Exit Setting Menu
EXAMPLE: 50% Pressure increase of 50 mmHg set pressure would become 75mmHg during Wash cycle.
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
BACK Stryker CrossFlow Flash Card 01
Stryker Color Palette
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Clear Function Clear Function: How to access & adjust the Clear function.
1
3
Access Settings Menu: = Pump Customization Menu
2
Adjust Clear Settings: = Clear Settings Screen
4
Start Pump to Perform Clear: = Clear can only be activated by footswitch
= Increase Pressure/Flow/Duration Settings
Adjust Clear Settings:
= Decrease Pressure/Flow/Duration Settings = Exit Setting Menu
Understanding Clear: Clear Function The Clear function is designed for quick dismissal and flushing of debris from the joint. It increases flow only and does not increase pressure.
= Wash/Clear Settings Screen
NOTE: Can customize to last for a duration up to 120 seconds.
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
FRONT Stryker CrossFlow Flash Card 02
Stryker Color Palette
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Hot Swap
6 Hot Swap Function:
How to adjust & toggle hardware and/or scope combinations with the Hot Swap feature.
1
2
Access Scope/Hardware Combination Selection
4
Choose Desired Option: = Number of Scope/Cannula combinations being customized
Press the top right section of the screen where the scope/hardware combination is shown.
= ONLY available during hip procedures
3
Perform Hot Swap: = Hot Swap Function
Use the Hot Swap function during your case by pressing the icon on the console or CrossFlow hand control.
Choose Combination: = Scroll Scope/Cannula Combinations = Assigned Scope/Cannula is Selected
NOTE: The “Hot Swap” function allows the user to switch the scope/hardware combination being used during the case, without recalibrating the pump. Users can add multiple scope/hardware options to toggle between during the case.
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
BACK Stryker CrossFlow Flash Card 02
Stryker Color Palette
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Hot Swap Hot Swap Function: If “Standard Settings” is chosen at the beginning of the case rather than a Surgeon Profile, these are the preselected scope/cannula combination settings for each joint:
Hardware Option
Joint
Default Combination
Knee
Stryker 5.8mm x 140mm Cannula with 4.0mm Scope
Shoulder
Stryker 5.8mm x 140mm Cannula with 4.0mm Scope
Hip
Stryker 5.0mm x 165mm Cannula (bridge) with 4.0mm Scope
Small
Stryker 4.0mm x 75mm Cannula with 2.7mm Scope
None
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
FRONT Stryker CrossFlow Flash Card 02
Stryker Color Palette
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Hot Swap
6 Hot Swap Function:
How to adjust & toggle hardware and/or scope combinations with the Hot Swap feature.
1
2
Access Scope/Hardware Combination Selection
4
Choose Desired Option: = Number of Scope/Cannula combinations being customized
Press the top right section of the screen where the scope/hardware combination is shown.
= ONLY available during hip procedures
3
Perform Hot Swap: = Hot Swap Function
Use the Hot Swap function during your case by pressing the icon on the console or CrossFlow hand control.
Choose Combination: = Scroll Scope/Cannula Combinations = Assigned Scope/Cannula is Selected
NOTE: The “Hot Swap” function allows the user to switch the scope/hardware combination being used during the case, without recalibrating the pump. Users can add multiple scope/hardware options to toggle between during the case.
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
BACK Stryker CrossFlow Flash Card 02
Stryker Color Palette
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Hot Swap Hot Swap Function: If “Standard Settings” is chosen at the beginning of the case rather than a Surgeon Profile, these are the preselected scope/cannula combination settings for each joint:
Hardware Option
Joint
Default Combination
Knee
Stryker 5.8mm x 140mm Cannula with 4.0mm Scope
Shoulder
Stryker 5.8mm x 140mm Cannula with 4.0mm Scope
Hip
Stryker 5.0mm x 165mm Cannula (bridge) with 4.0mm Scope
Small
Stryker 4.0mm x 75mm Cannula with 2.7mm Scope
None
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
FRONT Stryker CrossFlow Flash Card 03
Stryker Color Palette
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Shaver & RF Settings
7 Shaver/RF Console Integration
How to access & adjust approved resection console integration settings.
1
3
Access Settings Menu:
4
Customize Shaver:
= RF Console Settings
= Shaver Console Settings
= Pump Customization Menu
PRESS:
2
Customize Footswitch:
PRESS: = When Using CrossFire Consoles
= When Using CrossFire Consoles
= When Using an Approved Shaver Console
= When Using an Approved RF Console
= When No Shaver Console is Connected
= When No RF Console is Connected
= Suction Up
5
= Suction Down
= Resection Integration Settings Screen
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
BACK Stryker CrossFlow Flash Card 03
Stryker Color Palette
Customize RF:
Return to Run Screen:
= Exit Menu Integration of Shaver and/or RF devices are now customized. NOTE: Suction must be checked if using RF devices with suction. Uncheck if surgeon will be using RF without suction.
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Footswitch/Hand Control
8 Footswitch/Hand Control Settings:
How to customize & program settings on compatible footswitches and the CrossFire 2 Hand Control.
1
2
Access Settings Menu: = Pump Customization Menu
Customize Footswitch: = Footswtich/Hand Control Settings
PRESS:
3
Assigning Button(s) Functionality to the Footswitch & Hand Control:
Select button(s) on the footswitch or hand control diagram (the selected button will outline in green). = Scroll Features Menu
= When Using CrossFlow Footswitch
Highlighted feature will assign to the selected button(s). = When Using CrossFire 2 Footswitch = When Using iSwitch Wireless Footswitch = When Using CrossFire 2 Hand Control Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
4
Return to Run Screen: = Exit Menu The hand control and footswitch are now customized.
NOTE: CrossFire 2 Hand Control and CrossFire 2 Footswitch can have only one button programmed with CrossFlow function.
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
FRONT Stryker CrossFlow Flash Card 03
Stryker Color Palette
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Shaver & RF Settings
7 Shaver/RF Console Integration
How to access & adjust approved resection console integration settings.
1
3
Access Settings Menu:
4
Customize Shaver:
= RF Console Settings
= Shaver Console Settings
= Pump Customization Menu
PRESS:
2
Customize Footswitch:
PRESS: = When Using CrossFire Consoles
= When Using CrossFire Consoles
= When Using an Approved Shaver Console
= When Using an Approved RF Console
= When No Shaver Console is Connected
= When No RF Console is Connected
= Suction Up
5
= Suction Down
= Resection Integration Settings Screen
Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
BACK Stryker CrossFlow Flash Card 03
Stryker Color Palette
Customize RF:
Return to Run Screen:
= Exit Menu Integration of Shaver and/or RF devices are now customized. NOTE: Suction must be checked if using RF devices with suction. Uncheck if surgeon will be using RF without suction.
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13
Stryker Color Palette
Prints: CMYK / Stock: Print Run: X,XXX
Footswitch/Hand Control
8 Footswitch/Hand Control Settings:
How to customize & program settings on compatible footswitches and the CrossFire 2 Hand Control.
1
2
Access Settings Menu: = Pump Customization Menu
Customize Footswitch: = Footswtich/Hand Control Settings
PRESS:
3
Assigning Button(s) Functionality to the Footswitch & Hand Control:
Select button(s) on the footswitch or hand control diagram (the selected button will outline in green). = Scroll Features Menu
= When Using CrossFlow Footswitch
Highlighted feature will assign to the selected button(s). = When Using CrossFire 2 Footswitch = When Using iSwitch Wireless Footswitch = When Using CrossFire 2 Hand Control Refer to the instructions for use for important safety and user advice. Do not operate the device without first reading and understanding the Instructions for Use. The IFU part number is: P17330 The CrossFlow Pump catalog number is: 0450-000-000
1.800.624.4422
www.stryker.com
4
Return to Run Screen: = Exit Menu The hand control and footswitch are now customized.
NOTE: CrossFire 2 Hand Control and CrossFire 2 Footswitch can have only one button programmed with CrossFlow function.
Stryker Endoscopy | 5900 Optical Court | San Jose, CA 95138 Literature Number: 1000900963 (Rev. B) | VID/FWG 0.0k 02/13