Universal Battery Charger REF
7110-120-000
Stryker F1® Universal Battery Charger REF
7110-190-000
Instructions For Use
ENGLISH (EN) 2020-08
7110-120-700 Rev-AB
www.stryker.com
EN
Introduction This instructions for use manual contains information intended to ensure the safe, effective, and compliant use of your product. This manual is intended for in-service trainers, physicians, nurses, surgical technologists, and biomedical equipment technicians. Keep and consult this reference manual during the life of the product. The following conventions are used in this manual: • A WARNING highlights a safety-related issue. ALWAYS comply with this information to prevent patient and/or healthcare staff injury. • A CAUTION highlights a product reliability issue. ALWAYS comply with this information to prevent product damage. • A NOTE supplements and/or clarifies procedural information. For additional information, including safety information, in-service training, or current literature, contact your Stryker sales representative or call Stryker customer service at 1-269-323-7700 or 1-800-253-3210. Outside the US, contact your nearest Stryker subsidiary. NOTE: The user and/or patient should report any serious product-related incident to both the manufacturer and the Competent Authority of the European Member State where the user and/or patient is established. Trademarks not the property of Stryker Corporation are the property of their respective owners.
Indications For Use The Stryker Universal Battery Charger and Stryker F1 Universal Battery Charger (battery chargers) are four-station, modular battery chargers intended to charge Stryker handpiece battery packs only. The battery chargers and battery packs are specifically designed to work together so that the battery charger’s information screen will provide specific battery pack information.
7110-120-700 Rev-AB
Safety Directives WARNINGS: • Before using this equipment, or any component compatible with this equipment, read and understand the instructions for use. Pay particular attention to safety information. Become familiar with the equipment before use. • Only healthcare professionals trained and experienced in the use of this medical device should operate this equipment. • The Stryker Universal Battery Charger and Stryker F1 Universal Battery Charger are limited to professional use within a professional healthcare environment, excluding near high frequency (HF) surgical equipment and accessories. See the Electromagnetic Compatibility section. • Upon initial receipt and before each use, operate the equipment and inspect each component for damage. DO NOT use any equipment if damage is apparent. See the Inspection and Testing section for inspection criteria. • DO NOT use this equipment in areas in which flammable anesthetics or flammable agents are mixed with air, oxygen, or nitrous oxide. • Take special precautions regarding electromagnetic compatibility (EMC) when using medical electrical equipment. Place this equipment into service according to the EMC information contained in this manual. Portable and mobile radio frequency (RF) equipment can affect the function of this equipment. • ALWAYS use the appropriate battery charger to charge battery packs. • DO NOT operate the battery charger using a voltage inconsistent with the rating on the back of the unit. • DO NOT operate the battery charger with a damaged power cord or plug.
Contraindications None known.
2
www.stryker.com
7110-120-700 Rev-AB
EN
Accessories
CAUTION: ALWAYS select and install the appropriate charger module for the specific battery pack used. NOTES: • Four System 6™ Modules (REF 6110-625000) are supplied with the Universal Battery Charger. • Four Stryker F1 Charger Modules (REF 1900125-000) are supplied with the Stryker F1 Universal Battery Charger. • Alternative charger modules may be installed to accommodate specific types of battery packs as required. • For a complete list of accessories, contact your Stryker sales representative. Outside the US, contact your nearest Stryker subsidiary.
WARNINGS: • Use only Stryker-approved equipment, unless otherwise specified. • DO NOT modify any equipment without the authorization of the manufacturer. • Use only Stryker-approved electronic components and accessories. Failure to comply may result in degraded performance, increased electromagnetic emissions, or decreased electromagnetic immunity of the system. • DO NOT use any battery charger accessory that is not recommended or sold by Stryker. Failure to comply may result in fire, electric shock, or injury.
The following Stryker-approved accessories are sold separately: CHARGER MODULE
REF
FOR USE WITH BATTERY PACK REF
Charger Module*
6110-412-000
4112-000-000
System 5 Module
6110-415-000
4115-000-000
Charger Module
6110-422-000
4212-000-000 4215-000-000
Non-sterile Battery Charger Module
6110-426-000
4126-110-000 4222-110-000
System 6 Module
6110-625-000
6126-110-000 6212-000-000 6215-000-000 7126-110-000 7212-000-000 7215-000-000 7222-110-000 8212-000-000 8215-000-000
Stryker F1 Charger Module
1900-125-000
1900-012-000 1900-013-000
*This charger module is not design integrated. The battery charger will not display charge cycle count information for this charger module. POWER CORD
REF
Source Cord
0277-702-019
European Source Cord
0590-100-002
Power Cord (general)
0996-851-XXX series
www.stryker.com
3
EN
7110-120-700 Rev-AB
Features Battery Charger
A F
G
B H
I
C J
D E
A
Battery Pack – A variety of battery packs are sold separately.
F
Power Cord Receptacle – Accepts the power cord.
B
Charger Module – An interchangeable component designed to accept a specific type of battery pack. A variety of charger modules are sold separately.
G
Power Cord – Connects the battery charger to a hospital-grade, facility power receptacle
H
Isolated Diagnostics Port – An Ethernet receptacle used to export device usage data from the battery charger to Stryker via Stryker’s Cloud, if desired.
I
Ethernet Cable (not supplied) – Connects the isolated diagnostics port to a facility Ethernet receptacle
J
MAC ID Label – Shows the Media Access Control Identifier (MAC Address) for the battery charger.
C
Indicator Lights – Provide status information for each charger module.
D
Information Screen – Provides status information and error messages for the battery charger, charger modules, and battery packs.
E
4
Input Buttons – Allow the user to interact with the battery charger. The information screen displays appropriate input button labels during the battery pack charging sequence.
www.stryker.com
7110-120-700 Rev-AB
EN
Information Screen
Indicator Lights
A
B
C
D
E
INDICATION
STATUS
GREEN (Steady State)
The battery pack is fully charged.
AMBER (Steady State)
The battery charger is either charging or discharging the battery pack.
AMBER (Slow Blinking)
The battery pack is at or near the end of its useful life. The battery charger or charger module failed diagnostic testing.
A
State-of-Health Area – Indicates charging activity, and the state-ofhealth of the battery pack.
AMBER (Fast Blinking)
B
Counter Area – Displays numeric information based on the state of the battery charger, charger modules, and/ or battery packs.
Definitions
C
Message Area – Displays status information and requests user action.
D
Symbol Area – Displays symbols based on the state of the battery charger, charger modules, and/or battery packs.
E
Input Button Label Area – Displays appropriate input button labels during the battery pack charging sequence.
The symbols located on the equipment and/ or labeling are defined in this section or in the Symbol Definition Chart. See the Symbol Definition Chart supplied with the equipment. SYMBOL
DEFINITION REPLACE – The battery pack failed the state-of-health test. Ethernet
MAC ID
Media Access Control Identifier Stryker F1 battery pack orientation
Refer to instruction manual/booklet General warning sign CSA Group certification mark for United States and Canada. These products were tested and meet medical electrical equipment certification requirements, including compliance with applicable 60601 series standards. For additional information, contact Stryker.
www.stryker.com
5
EN
7110-120-700 Rev-AB
Instructions To Connect the Battery Charger WARNINGS: • To avoid the risk of electric shock, ALWAYS connect this equipment to a hospital-grade, facility power receptacle with protective earth. • DO NOT modify the power cord. • ALWAYS position the equipment so that the power cord may be easily disconnected as required. • DO NOT place the battery charger within the sterile field. • DO NOT touch the battery charger and the patient simultaneously. Failure to comply may cause electrical shock. CAUTION: ALWAYS place the power cord where it will not be stepped on, tripped over, or otherwise subjected to damage or stress. 1. Use the power cord to connect the battery charger to a hospital-grade, facility power receptacle (figure 1). The power-up sequence will begin: - A diagnostic test will be performed to ensure the battery charger and each installed charger module is operational - The battery charger software and hardware revision numbers will appear (figures 2 and 3) - The charger module revision number will appear for each installed charger module (figure 4). NOTE: If an ERROR message appears during the power-up sequence, note the error code that appears in the counter area (figure 5). See the Troubleshooting section. 2. If desired, use an Ethernet cable (not supplied) to connect the isolated diagnostics port to a facility Ethernet receptacle. NOTES: • Confirm with a facility IT representative that the Ethernet receptacle has access to Stryker’s Cloud. • See the Network Requirements in the Specifications section. • For more information, contact your Stryker sales representative.
6
Figure 1 – Battery Charger Connections
Figure 2 – Software Revision Number
Figure 3 – Hardware Revision Number
Figure 4 – Charger Module Revision Number
Figure 5 – Error Code
www.stryker.com
7110-120-700 Rev-AB
To Charge a Battery Pack CAUTION: When battery packs require postsurgical sterilization, ALWAYS allow the battery packs to cool for one hour before charging. Excess heat buildup from sterilization and charging will damage the battery packs. NOTES: • To ensure maximum operating time, always charge battery packs before sterilization, even if the battery packs have not been used. • The charging sequence may take between 5 and 120 minutes, based on the type of battery pack and the existing level of charge in the battery pack. • On the Stryker F1 Universal Battery Charger, always install Stryker F1 SmartGrip® Battery Modules in the proper orientation (figure 6).
EN
After the battery pack is fully charged, a stateof-health test will be performed (figure 10): - Five to eight bars will illuminate in the state-of-health area. More illuminated bars indicates a better ability of the battery pack to store energy. - The counter area will update the charge cycle count (for design integrated components only). - A READY message will appear.
Figure 7 – No Battery Message
Figure 8 – Battery Pack Part Number
Figure 6 – Stryker F1 SmartGrip Battery Module Orientation 1. Install a clean, dry battery pack into a compatible charger module with a NO BATTERY message screen (figure 7). The information screen will momentarily display the battery pack part number (figure 8) and the charging sequence will begin. During the charging sequence, a CHARGING message will appear and the number of illuminated state-of-health bars will increase from left to right (figure 9).
Figure 9 – Charging Message and Bars
Figure 10 – Ready Message with State-ofHealth Bars and Charge Cycle Count
www.stryker.com
7
EN
7110-120-700 Rev-AB
NOTE: If the battery pack fails the state-of-health test, a REPLACE message and symbol will appear (figure 11). See the Troubleshooting section. 2. Store the battery pack on the battery charger to maintain a full charge before sterilization.
To Access Over Temperature Information NOTES: • Specific types of battery packs allow you to access information about battery pack exposure to over temperature conditions. • The information screen will return to its origin after displaying an over temperature message screen for five seconds. 1. From the READY, CHARGING, or REPLACE message screen, press the MORE button to access the # of OVER TEMP message screen. The counter will display the number of times the battery pack has been exposed to an over temperature condition (figure 12). 2. Press the MORE button again to access the TIME OVER TEMP message screen. The counter will display how much time, in minutes, the battery pack has been exposed to over temperature conditions (figure 13).
To Disconnect the Battery Charger CAUTION: To reduce the risk of damage to the power cord, ALWAYS pull the power cord plug, not the power cord when disconnecting the battery charger. Disconnect the battery charger from the facility power receptacle and Ethernet receptacle, as required.
To Replace a Charger Module See the instructions for use supplied with the replacement charger module.
8
Figure 11 – Replace Message and Symbol
Figure 12 – Total Number of Over Temperature Events
Figure 13 – Total Time Exposed to Over Temperature Conditions
Cleaning To Clean the Battery Charger and Charger Modules WARNING: Before cleaning, ALWAYS disconnect the power cord from the facility power receptacle and the battery charger to reduce the risk of electric shock. CAUTIONS: • DO NOT immerse the battery charger or charger modules in liquid. • DO NOT use solvents, lubricants, or other chemicals to clean the battery charger or charger modules unless otherwise directed.
www.stryker.com
7110-120-700 Rev-AB
• The use of abrasive disinfectants, such as Clorox bleach, can lead to build-up on the contacts which may affect the functionality of the charger module. • DO NOT allow liquid to collect in the charger modules or on top of the battery charger. • DO NOT allow liquid to enter the cooling vents on the back of the battery charger. • DO NOT spray or wipe liquid directly on the contacts of the charger modules. • DO NOT sterilize the battery charger or charger modules. 1. Disconnect the power cord from the facility power receptacle and the battery charger. 2. Disconnect the Ethernet cable, if used, from the isolated diagnostics port. 3. Gently wipe the external surfaces of the battery charger and power cord with a soft, lint-free cloth moistened with a prepared cleaning solution or non-abrasive, hospital disinfectant and water. 4. Immediately dry the battery charger with a soft, lint-free cloth or compressed air, < 140 kPa [< 20 psi]. 5. Inspect the battery charger and power cord for damage. See the Inspection and Testing section for inspection criteria.
To Clean the Battery Pack For battery pack processing instructions, see the care instructions manual supplied with the battery pack and/or a compatible handpiece.
Inspection and Testing WARNINGS: • Only individuals trained and experienced in the maintenance of reusable medical devices should inspect and test this equipment. • Before attempting any inspection and testing, ALWAYS disconnect the power cord from the battery charger to reduce the risk of electric shock.
www.stryker.com
EN
• Perform recommended inspection and testing as indicated in these instructions. • DO NOT use any equipment if damage is apparent. • DO NOT use any system component if the inspection criteria are not met. • DO NOT disassemble or service this equipment. Failure to comply may result in electric shock or fire. NOTES: • If the equipment fails to meet the inspection and testing criteria, contact your Stryker sales representative or call Stryker customer service. Outside the US, contact your nearest Stryker subsidiary. • Maintenance documentation for this equipment is available upon request to Stryker-authorized service personnel only. INTERVAL
ACTIVITY
CRITERIA
Upon initial receipt and before each use
Inspect the equipment
No damage or signs of wear No loose or missing components No cuts in the power cord No bent pins or contacts No cracks in the housing
Storage and Handling CAUTION: ALWAYS store the equipment within the specified environmental condition values throughout its useful life. See the Specifications section. To ensure longevity, performance, and safety, use of the original packaging materials is recommended when storing or transporting this equipment.
9
EN
7110-120-700 Rev-AB
Disposal/Recycle WARNING: ALWAYS follow the current local recommendations and/or regulations governing environmental protection and the risks associated with recycling or disposing of the equipment at the end of its useful life. In accordance with European Directive 2012/19/EU on Waste Electrical and Electronic Equipment (WEEE) as amended, product should be collected separately for recycling. Do not dispose of as unsorted municipal waste. Contact local distributor for disposal information. Ensure infected equipment is decontaminated prior to recycling. NOTE: Contact Stryker for the recycling passport for this product.
China RoHS Standard SJ/T 11364 To comply with China RoHS Standard SJ/T 11364, this device has been marked with the environmentally-friendly use period (EFUP) number, measured in years. The device contains at least one of the listed hazardous substances above threshold.
China RoHS Disclosure Report REF 7110-120-000, REF 7110-190-000 Hazardous Substances Part Name
Electronic Circuit Boards
Lead (Pb)
Mercury (Hg)
Cadmium (Cd)
X
O
O
Hexavalent Polybrominated Polybrominated Chromium biphenyls diphenyl ethers (Cr (VI)) (PBB) (PBDE) O
O
O
This table is prepared in accordance with the provisions of SJ/T11364. O: Indicates that said hazardous substance contained in all of the homogenous materials for this part is below the limit requirement of GB/T26572. X: Indicates that said hazardous substance contained in at least one of the homogenous materials used for this part is above the limit requirement of GB/T26572.
10
www.stryker.com
7110-120-700 Rev-AB
EN
Troubleshooting WARNING: DO NOT disassemble or service this equipment. Failure to comply may result in electric shock or fire. NOTE: For service, contact your Stryker sales representative or call Stryker customer service. Outside the US, contact your nearest Stryker subsidiary. PROBLEM
COMPONENT
CAUSE
ACTION
Indicator lights do not illuminate.
Battery charger
No power to battery charger.
Reseat the cord connections on back of the battery charger and the facility power receptacle.
Battery charger is damaged.
Return the battery charger for repair.
Charger module is not properly connected.
Reseat the charger module connectors. See the instructions for use supplied with the charger module.
Internal problem.
Cycle the power on the charger.
NO MODULE appears in the message area of the information screen.
Charger module
Remove and reinstall the charger module. See the instructions for use supplied with the charger module. Charger module is damaged.
www.stryker.com
Replace the charger module.
11
EN
7110-120-700 Rev-AB
PROBLEM
COMPONENT
CAUSE
ACTION
NO BATTERY appears in the message area of the information screen when a battery is installed.
Battery pack
Battery pack is not properly seated in charger module.
Remove and reinstall the battery pack.
Stryker F1 battery pack release button is not allowing proper installation.
Push the release button during installation.
Stryker F1 battery pack is not installed in the correct orientation.
Reinstall battery pack in the correct orientation.
Contacts are dirty or corroded.
Clean the battery pack.
Battery pack is dead.
Replace the battery pack.
Charger module
Charger module is damaged.
Replace the charger module.
Battery pack
Battery pack has low voltage.
Leave the battery pack on the charger module for a minimum of five minutes. If the problem persists, check other causes.
Battery pack is dead.
Replace the battery pack.
Battery pack is wet.
Dry and reinstall the battery pack.
Battery pack has exceeded its operational life.
Replace the battery pack.
Charger module contact is wet.
Dry the contact and reinstall the battery pack.
Internal problem.
Reinsert the battery. If the REPLACE message does not occur again, then replace the charger module.
Message area of the information screen cycles between NO BATTERY and CHARGING.
REPLACE appears in the message area of the information screen.
Battery pack
Charger module
Battery pack does not fit into charger module.
Charger module
Charger module is intended for different battery pack.
Install the proper charger module. See the instructions for use supplied with the charger module.
Charger module is loose.
Charger module
Screw is not secure.
Tighten the screw.
12
www.stryker.com
7110-120-700 Rev-AB
EN
PROBLEM
COMPONENT
CAUSE
ACTION
Battery pack becomes unusually hot during use or while charging.
Charger module
Battery pack is intended for different charger module.
Install the battery pack in the correct charger module.
Battery pack
Internal problem.
Check battery pack status in the battery charger; replace the battery pack if indicated.
Battery charger
Internal problem.
Return the battery charger for repair.
Information screen does not display battery pack stateof-health information.
Battery charger
The battery charger will display the battery pack state-of-health at the end of the charge cycle.
Wait until the charge cycle is complete.
Information screen does not display charge cycle count information.
Battery pack
Battery pack is not a Stryker-approved battery pack.
Use Stryker battery packs only.
Battery pack/ Charger module
Battery pack (REF 4112000-000) is not a design integrated component.
No action required.
Amber indicator light flashes continuously.
Battery pack/ charger module
Battery pack and/or battery charger contacts may be dirty.
Clean the contacts.
ERROR code 005 appears in the counter area of the information screen.
Battery pack
Battery pack is wet.
Dry and reinstall the battery pack.
Charger module
Charger module contact is wet.
Dry the contact and reinstall the battery pack.
Charger module contact is corroded.
Clean the charger module contact. If problem persists, replace the charger module.
www.stryker.com
13
EN
7110-120-700 Rev-AB
PROBLEM
COMPONENT
CAUSE
ACTION
ERROR code 007 appears in the counter area of the information screen.
Battery pack
Battery pack contact is corroded.
Clean the battery pack contact.
Battery pack contact is damaged.
Replace the battery pack.
Battery pack ID is not present.
Replace the battery pack.
Battery pack has lost smart communication function.
Replace the battery pack.
Battery pack was repaired by an unauthorized provider and is not programmed.
Replace the battery pack.
Battery pack is not a Stryker-approved battery pack.
Use Stryker battery packs only.
Charger module contact is corroded.
Clean the charger module contact.
Charger module contact is damaged.
Replace the charger module.
Battery charger contact is corroded.
Clean the battery charger contact.
Battery charger contact is damaged.
Return the battery charger for repair.
Charger module
Battery charger
ERROR code 007 appears in the counter area of the information screen for a short duration, followed by normal charging operation.
Battery charger
Internal problem.
Replace the charger module.
Sporadic electrical interference is experienced.
Battery charger
Electrical noise is present.
Turn off all electrical equipment not in use. Relocate electrical equipment; increase spatial distance. Install electrical equipment into different facility power receptacles.
14
www.stryker.com
7110-120-700 Rev-AB
EN
Specifications Model:
Universal Battery Charger (REF 7110-120-000)
Stryker F1 Universal Battery Charger (REF 7110-190-000)
Dimensions:
130 mm [5.125 inch] height 257 mm [10.125 inch] width 394 mm [15.5 inch] length
193 mm [7.6 inch] height 257 mm [10.125 inch] width 394 mm [15.5 inch] length
Mass:
3.47 kg [7.65 lb]
3.47 kg [7.65 lb]
Material:
In accordance with the European REACH regulation and other environmental regulatory requirements, components within the charger contain Lead, CAS No. 7439-92-1; Diboron Trioxide, CAS No. 1303-86-2; Lead Monoxide, CAS No. 1317-36-8; and 1-Methyl-2-Pyrrolidone (NMP), CAS No. 872-50-4. This declaration is made in good faith and is either based on a technical evaluation, supplier data, and/or laboratory testing.
Power Cord (REF 0277-702-019 and REF 0590-100-002):
2 m [78.7 inch] length, fitted with NEMA 5-15 hospital-grade plug, or 2.5 m [98.4 inch] length, fitted with CEE 7/7 plug
Mode of Operation:
Continuous Operation
Electrical: Input:
100-230 V
Output:
22 V
Ingress Protection:
IPX0
Equipment Type:
Class I
Ground Type: Means of Isolation from Supply Mains:
50-60 Hz 2.2-1 A
Protective Earth Ground Disconnect the power cord from the battery charger power cord receptacle.
Network Requirements: Purpose of the Connection:
Device Usage Data
Required Network Characteristics:
Standard TCP/IP over Ethernet (IEEE 802.3 10/100 Mbs)
Required Network Configuration:
DHCP server, Mask and Gateway provided by site IT admin; Access to Stryker’s Cloud
Technical Specifications of the Connection:
Standard TCP/IP over Ethernet (IEEE 802.3 10/100 Mbs), utilized industry standard https (port 443) web services communication to Stryker’s Cloud.
Intended Information Flow:
Device Usage Data
Hazardous Situations Resulting from a Failure of the Network:
None known. CAUTION: Subsequent changes to the IT network could introduce new risks and require additional analysis. The hospital IT admin should identify, analyze, evaluate, and control these risks. Changes to the IT network include: changes in network configuration, connection of additional items, disconnection of items, update of equipment, upgrade of equipment.
www.stryker.com
15
EN
7110-120-700 Rev-AB
Power Cord Specifications (general) (REF 0996-851-XXX Series) Ratings: Current:
10 A
Voltage:
250 VAC minimum
Frequency:
50/60 Hz
Temperature:
0 °C to 70 °C minimum
Flammability:
UL 94 V-2 minimum, IEC 60332-1
Conductor: Size: Material:
3 X 1.00 mm2 ≤ size < 3 X 1.50 mm2 Copper
Connector Type:
IEC 60320 C13
Plug Type:
The facility power (mains) plug shall have a ground/earthing pin.
Cord: Type: Length:
SJT, H05VV-F, HVCTF, RVV or equivalent (unshielded) 3.0 m, 2.5 m
NOTE: The 2.5 m cord is not for use in Canada or the US.
Dielectric Withstand:
1500 VAC for 60 seconds between line and protective earth, and between neutral and protective earth
Certification:
All applicable in-country medical electrical requirements
Power Cord Requirements (specific) The Canadian and US power supply cord shall have a tag or label in English and French indicating that “GROUNDING RELIABILITY CAN ONLY BE ACHIEVED WHEN EQUIPMENT IS CONNECTED TO AN EQUIVALENT RECEPTACLE MARKED ‘HOSPITAL ONLY’ OR ‘HOSPITAL GRADE’ ” or equivalent wording. Agency Approval: CSA Certified for Canada and US or UL Recognized for Canada and US.
16
www.stryker.com
7110-120-700 Rev-AB
Environmental Conditions:
EN
Operation
Storage and Transportation
Temperature:
Humidity:
Atmospheric Pressure:
www.stryker.com
17
EN
7110-120-700 Rev-AB
Electromagnetic Compatibility WARNING: DO NOT stack or place equipment adjacent to the product. If such a configuration is necessary, observe the configuration to ensure that electromagnetic interference does not degrade performance. Guidance and manufacturer’s declaration - electromagnetic emissions The Universal Battery Charger (REF 7110-120-000) and Stryker F1 Universal Battery Charger (REF 7110-190-000) are intended for use in the electromagnetic environment specified below. The customer or the user of the Universal Battery Charger (REF 7110-120-000) and Stryker F1 Universal Battery Charger (REF 7110-190-000) should assure that they are used in such an environment.
18
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The Universal Battery Charger (REF 7110-120-000) and Stryker F1 Universal Battery Charger (REF 7110-190-000) use RF energy only for their internal function. Therefore, their RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
Class A
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions IEC 61000-3-3
Complies
NOTE: The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment. The Universal Battery Charger (REF 7110-120-000) and Stryker F1 Universal Battery Charger (REF 7110-190-000) are suitable for use in all establishments other than domestic establishments and those directly connected to the public lowvoltage power supply network that supplies buildings used for domestic purposes.
www.stryker.com
7110-120-700 Rev-AB
EN
Guidance and manufacturer’s declaration - electromagnetic immunity The Universal Battery Charger (REF 7110-120-000) and Stryker F1 Universal Battery Charger (REF 7110-190-000) are intended for use in the electromagnetic environment specified below. The customer or the user of the Universal Battery Charger (REF 7110-120-000) and Stryker F1 Universal Battery Charger (REF 7110-190-000) should assure that they are used in such an environment. Electromagnetic environment - guidance
Immunity test
IEC 60601 test level
Compliance level
Electrostatic discharge (ESD) IEC 61000-4-2
±2 kV, ±4 kV, ±6 kV, ±8 kV contact
±2 kV, ±4 kV, ±6 kV, ±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transient/burst IEC 61000-4-4
±2 kV at 100 kHz repetition frequency for power supply lines ±1 kV at 100 kHz repetition frequency for input/output lines
±2 kV at 100 kHz repetition frequency for power supply lines ±1 kV at 100 kHz repetition frequency for input/output lines
Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5
±0.5, ±1 kV line(s) to line(s) ±0.5, ±1, ±2 kV line(s) to earth
±0.5, ±1 kV line(s) to line(s) ±0.5, ±1, ±2 kV line(s) to earth
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
<5% UT (>95% dip in UT) for 0.5 cycle 0% UT (100% dip in UT) for 0.5 cycle at 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315° 0% UT (100% dip in UT) for 1 cycle at 0° 40% UT (60% dip in UT) for 5 cycles 70% UT (30% dip in UT) for 25 & 30 cycles at 0° <5% UT (>95% dip in UT) for 5 s 0% UT (100% dip in UT) for 5 s
<5% UT (>95% dip in UT) for 0.5 cycle 0% UT (100% dip in UT) for 0.5 cycle at 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315° 0% UT (100% dip in UT) for 1 cycle at 0° 40% UT (60% dip in UT) for 5 cycles 70% UT (30% dip in UT) for 25 & 30 cycles at 0° <5% UT (>95% dip in UT) for 5 s 0% UT (100% dip in UT) for 5 s
Mains power quality should be that of a typical commercial or hospital environment. If the user of the Universal Battery Charger (REF 7110120-000) and Stryker F1 Universal Battery Charger (REF 7110-190-000) requires continued operation during power mains interruptions, it is recommended that the Universal Battery Charger (REF 7110-120-000) and Stryker F1 Universal Battery Charger (REF 7110-190000) be powered from an uninterruptible power supply or a battery.
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
30 A/m at 50 and 60 Hz
30 A/m at 50 and 60 Hz
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 20%.
NOTE: UT is the alternating current mains voltage prior to application of the test level.
www.stryker.com
19
EN
7110-120-700 Rev-AB
Guidance and manufacturer’s declaration - electromagnetic immunity The Universal Battery Charger (REF 7110-120-000) and Stryker F1 Universal Battery Charger (REF 7110-190-000) are intended for use in the electromagnetic environment specified below. The customer or the user of the Universal Battery Charger (REF 7110-120-000) and Stryker F1 Universal Battery Charger (REF 7110-190-000) should assure that they are used in such an environment. Immunity test
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
20
IEC 60601 test level
Compliance level
3 Vrms 150 kHz to 80 MHz outside ISM bands 80% AM at 1 kHz
3 Vrms 150 kHz to 80 MHz outside ISM bands 80% AM at 1 kHz
6 Vrms 150 kHz to 80 MHz in ISM bands 80% AM at 1 kHz
6 Vrms 150 kHz to 80 MHz in ISM bands 80% AM at 1 kHz
3 V/m 80 MHz to 2.7 GHz 80% AM at 1 kHz
3 V/m 80 MHz to 2.7 GHz 80% AM at 1 kHz
Electromagnetic environment guidance IEC 60601-1-2 3rd Edition: Portable and mobile RF equipment should be used no closer to any part of the Universal Battery Charger (REF 7110-120-000) and Stryker F1 Universal Battery Charger (REF 7110-190-000), including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance d=1.2√P d=1.2√P 80 MHz to 800 MHz d=2.3√P 800 MHz to 2.5 GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). IEC 60601-1-2 4th Edition: WARNING: Portable RF equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Universal Battery Charger (REF 7110-120-000) and Stryker F1 Universal Battery Charger (REF 7110-190-000), including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result. Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol:
www.stryker.com