Stryker
System 7 Reciprocating Saw Instructions for Use Rev D Sept 2017
Instructions for Use
22 Pages
Preview
Page 1
NRWH: This portion of the document should not appear on produced Labels or IFUs
Dimensions:
Color/Material/Finish:
Label Stock:
6 inch (width) x 6 inch Booklet
Color Graphics on White Background 20# Bond or Equivalent
N/A Description/Type: Instructions For Use
Stryker Instruments (269) 323-7700 (800) 253-3210
DSGN-fm-50366 5ev. $ Effective: $SULO Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
6FRSH.=2,5(635 3URFHVV2ZQHU7HFK3XEV0DQDJHU
Part Number:
Rev.
7206-001-700
D
Page 1 of 1
System 7 Reciprocating Saw REF
7206-000-000
Instructions For Use
ENGLISH (EN) 2017-09
7206-001-700 Rev-D
Printed Date: Jul 03, 2018 22:49:23 GMT
0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
www.stryker.com
EN
Introduction This Instructions For Use manual is the most comprehensive source of information for the safe and effective use of your product. This manual may be used by in-service trainers, physicians, nurses, surgical technologists, and biomedical equipment technicians. Keep and consult this reference manual during the life of the product. The following conventions are used in this manual: • A WARNING highlights a safety-related issue. ALWAYS comply with this information to prevent patient and/or healthcare staff injury. • A CAUTION highlights a product reliability issue. ALWAYS comply with this information to prevent product damage. • A NOTE supplements and/or clarifies procedural information. If additional information or in-service training is required, contact your Stryker sales representative or call Stryker customer service. Outside the US, contact your nearest Stryker subsidiary.
Indications For Use The Stryker System 7 Battery Powered Heavy Duty System is intended for use in the cutting, drilling, decorticating, and smoothing of bone and other bone related tissue in a variety of surgical procedures. It is also usable in the placement of screws, wires, pins, and other fixation devices.
7206-001-700 Rev-D
NOTE: The Stryker System 7 Reciprocating Saw (handpiece) is a component of the Stryker System 7 Battery Powered Heavy Duty System.
Contraindications None known.
User/Patient Safety WARNINGS: • Before using any system component, or any component compatible with this system, read and understand the instructions. Pay particular attention to WARNING information. Become familiar with the system components prior to use. • Only trained and experienced healthcare professionals should use this equipment. • The healthcare professional performing any procedure is responsible for determining the appropriateness of this equipment and the specific technique used for each patient. Stryker, as a manufacturer, does not recommend surgical procedure or technique. • Upon initial receipt and before each use, clean and sterilize the equipment as indicated. See the Heavy Duty Care Instructions manual for processing instructions.
2
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
www.stryker.com
7206-001-700 Rev-D
• Upon initial receipt and before each use, operate the equipment and inspect each component for damage. DO NOT use any equipment if damage is apparent. See the Heavy Duty Care Instructions manual for inspection criteria. • DO NOT use this equipment in areas in which flammable anesthetics or flammable agents are mixed with air, oxygen, or nitrous oxide. • Take special precautions regarding electromagnetic compatibility (EMC) when using medical electrical equipment. Place this equipment into service according to the EMC information contained in this manual. Portable and mobile radio frequency (RF) communications equipment can affect the function of this equipment.
Accessories WARNINGS: • Use only Stryker-approved system components and accessories, unless otherwise specified. DO NOT modify any system component or accessory. • Using other electronic components and accessories may result in increased electromagnetic emissions or decreased electromagnetic immunity of the system.
EN
• DO NOT reuse, reprocess, or repackage a device that is intended for single use only. -- A single use device may not withstand chemical, chemical vapor, or high temperature sterilization reprocessing. -- Design features may make cleaning difficult. -- Reuse may create a contamination risk and compromise structural integrity resulting in operational failure. -- Critical product information may be lost during repackaging. Failure to comply may lead to infection or cross infection and result in patient and/or healthcare staff injury. • When using one of the following blades, be alert for blade whip when operating the handpiece below the maximum speed: • REF 0277-096-251 • REF 0277-096-275 • REF 0277-096-276 • REF 0277-096-277 • REF 0277-096-325 Blade whip increases the chance of blade fracture and may result in patient and/or healthcare staff injury. NOTE: For a complete list of accessories, contact your Stryker sales representative. Outside the US, contact your nearest Stryker subsidiary.
www.stryker.com
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
3
EN
7206-001-700 Rev-D
Accessories (continued)
Features
The following Stryker-approved accessories are sold separately:
Handpiece
DESCRIPTION
REF
Blades
0277-096-XXX series
System 6 Battery Pack, Large
6215-000-000
System 6 Aseptic Battery Kit, Large
6126-000-000
System 6 Aseptic Battery Kit, Small
6127-000-000
Stryker SmartLifeTM Battery Pack, Small
7212-000-000
Stryker SmartLife Battery Pack, Large
7215-000-000
Stryker SmartLife Non-sterile Battery, Large
7126-110-000
Stryker SmartLife Aseptic Housing, Large
7126-120-000
Stryker SmartLife Transfer Shield, Large
7126-130-000
Stryker SmartLife Non-sterile Battery, Small
7222-110-000
Stryker SmartLife Aseptic Housing, Small
7222-120-000
Stryker SmartLife Transfer Shield, Small
7222-130-000
4
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
B C A
D
A
Function Switch – Sets the speed or locks the trigger.
B
Applied Part – The distal end of the handpiece (as defined by the standards listed in the Specifications section under Product Safety Certification).
C
Blade Collar – Retains the blade in the handpiece.
D
Trigger – Controls the variable speed operation of the handpiece.
www.stryker.com
7206-001-700 Rev-D
EN
Features (continued)
Definitions
Function Switch
The symbols located on the equipment and/or labeling are defined in this section or in the Symbol Definition Chart. See the Symbol Definition Chart supplied with the equipment.
Fast Mode – The handpiece will operate at high speed when the trigger is depressed.
SYMBOL
DEFINITION Fast Mode
Standard Mode – The handpiece will operate at standard speed when the trigger is depressed. Safe Mode – The trigger is locked to prevent inadvertent operation of the handpiece.
Standard Mode
1 min / 4 min x 5
Duty Cycle – See the Specifications section. General warning sign To comply with European Community Waste Electrical and Electronic Equipment (WEEE) Directive 2012/19/EU, ALWAYS collect this product separately for recycling. DO NOT dispose of this product as unsorted municipal waste. Contact your local distributor for disposal information. Direct current
www.stryker.com
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
5
EN
7206-001-700 Rev-D
Instructions To Install the Blade
2. Release the blade collar. Make sure the blade collar springs back to its original position and the blade is securely locked in the blade collar.
WARNINGS: • ALWAYS slide the function switch to the safe mode position before installing the blade. • ALWAYS make sure the blade collar springs back to its original position and the blade is securely locked in the blade collar after installation. NOTE: The blade may be installed in any of four orientations.
CORRECT
INCORRECT
1. Rotate the spring-loaded blade collar to align the slots in the distal end of the collar, and then fully insert the blade into the blade collar. 3. Gently tug the blade to make sure the blade is securely locked in the blade collar.
6
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
www.stryker.com
7206-001-700 Rev-D
To Install the Battery Pack WARNING: ALWAYS slide the function switch to the safe mode position before installing the battery pack. NOTE: See the instructions for use supplied with the battery pack and/or battery charger for charging instructions and specifications. 1. Slide a fully charged battery pack into the handpiece until the battery pack snaps into place.
2. Gently tug the battery pack to make sure the battery pack is securely locked in the handpiece. 3. Test the operation of the handpiece by sliding the function switch to the fast or standard mode position, and then depressing the trigger.
EN
To Operate the Handpiece WARNINGS: • ALWAYS slide the function switch to the safe mode position when the handpiece is idle or when passing the handpiece to another person. • DO NOT change the position of the function switch while the handpiece is operating. • Before operating the handpiece, ALWAYS gently tug the blade to make sure the blade is securely locked in the blade collar. • ALWAYS operate the equipment within the specified environmental condition values. See the Specifications section. • ALWAYS follow the recommended duty cycle to prevent the equipment from overheating. See the Specifications section. • DO NOT apply excessive pressure, such as bending or prying, with the blade. Excessive pressure may bend or fracture the blade and result in tissue damage, loss of tactile control, and/or the ejection of blade fragments at a high velocity.
www.stryker.com
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
7
EN
7206-001-700 Rev-D
To Operate the Handpiece (continued)
To Remove the Battery Pack
CAUTIONS: • When operating the handpiece, let the blade do the cutting. DO NOT apply excessive pressure with the blade. Excessive pressure may bend the blade and reduce the cutting quality. • DO NOT stall the handpiece. Failure to comply may damage the electric motor and/or battery pack. If the handpiece jams, release the trigger immediately. Remove any obstructions before continuing to operate the handpiece. • If any power loss is experienced while using the handpiece, ALWAYS replace the battery pack with a fully charged battery pack. Failure to comply may result in a drained or damaged battery pack with a shortened life.
WARNING: ALWAYS slide the function switch to the safe mode position before removing the battery pack. Depress the battery latch and slide the battery pack out of the handpiece.
To Remove the Blade 1. Slide the function switch to the fast or standard position. 2. Depress the pressure-sensitive trigger for variable speed operation.
WARNING: ALWAYS slide the function switch to the safe mode position before removing the blade. Rotate the blade collar to align the slots in the distal end of the collar, and then remove the blade from the handpiece.
Care Instructions For processing instructions and disposal/recycle information, see the care instructions manual supplied with the equipment.
8
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
www.stryker.com
7206-001-700 Rev-D
EN
Troubleshooting WARNING: DO NOT disassemble or service this equipment. NOTE: For service, contact your Stryker sales representative or call Stryker customer service. Outside the US, contact your nearest Stryker subsidiary. PROBLEM
CAUSE
ACTION
The handpiece does not operate or operates at a reduced speed.
The battery pack is discharged.
Use a Stryker battery charger to recharge the battery pack.
The function switch is in the standard position.
Set the function switch to the fast position.
The battery pack is expended.
Replace the battery pack.
The function switch is in the safe mode position.
Set the function switch to the fast or standard position.
The drivetrain is malfunctioning.
Return the handpiece to Stryker for repair.
The handpiece operates but the blade does not move.
The drivetrain is malfunctioning.
Return the handpiece to Stryker for repair.
The handpiece continues to operate after the trigger is released.
The trigger is malfunctioning.
Depress the battery latch and slide the battery pack out of the handpiece. Return the handpiece to Stryker for repair.
The battery pack becomes unusually hot during use.
The circuitry is malfunctioning.
Use a Stryker battery charger to check the integrity of the battery pack. Replace the battery pack as required.
www.stryker.com
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
9
EN
7206-001-700 Rev-D
Troubleshooting (continued) PROBLEM
CAUSE
ACTION
The blade will not fit or cannot be secured in the blade collar.
The blade collar contains debris.
Use a small brush to clean the blade collar.
The blade is not a Stryker product.
Use a Stryker blade.
The blade collar is damaged.
Return the handpiece to Stryker for repair.
The blade collar does not spring back to its original position to securely lock the blade in the blade collar.
The blade collar contains debris.
Use a small brush to clean the blade collar. Actuate the collar several times to obtain smooth operation.
The handpiece is noisy and/ or vibrates.
The drivetrain is malfunctioning.
Return the handpiece to Stryker for repair.
The handpiece experiences Electrical noise is present. sporadic electrical interference.
Turn off all electrical equipment not in use in the operating room. Relocate electrical equipment and/ or increase spatial distance between electrical equipment. Plug operating room equipment into different operating room outlets.
10
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
www.stryker.com
7206-001-700 Rev-D
EN
Specifications WARNING: ALWAYS check any documentation that accompanies attachments, burs, pins, and/or blades for special duty cycle and usage instructions. CAUTION: ALWAYS store the equipment within the specified environmental condition values throughout its useful life. NOTE: Specifications are approximate and may vary between devices or as a result of power supply fluctuations. Model:
System 7 Reciprocating Saw (REF 7206-000-000)
Dimensions:
168 mm [6.6 inch] height 38 mm [1.5 inch] width 191 mm [7.5 inch] length
Mass:
0.94 kg [2.07 lb]
Speed:
13,000 cpm (fast) 10,000 cpm (standard)
Excursion:
3.9 mm [0.154 inch]
Mode of Operation:
Non-continuous Operation
Duty Cycle:
1 minute on/4 minutes off, 5 times
Rest Between Cycles:
3 hours
Maximum Temperature of Applied Part:
Less than 51 °C [124 °F] (Maximum surface temperature as tested to the standards listed under Product Safety Certification.)
Power Supply:
Internally Powered Refer to battery housing for voltage rating.
Ingress Protection:
IPX0 Ordinary Equipment
Equipment Type: Type BF Applied Part
www.stryker.com
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
11
EN
7206-001-700 Rev-D
Specifications (continued) Product Safety Certification:
CSA International Canadian Standards Association (CSA) CAN/CSA-C22.2 No. 60601-1:14, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance; (IEC 60601-1:2005+A1:2012, MOD) American National Standards Institute (ANSI)/Association for the Advancement of Medical Instrumentation (AAMI) ANSI/AAMI ES60601-1:2005/(R) 2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance; Consolidated Reprint (2009/(R) 2012); Amendment 2 (2010/(R) 2012); Amendment 1 (2012)
Product Safety Compliance:
International Electrotechnical Commission (IEC) IEC 60601-1:2005, Ed: 3.1, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance; Corrigendum 1 (2006); Corrigendum 2 (2007); Amendment 1 (2012) European Committee for Electrotechnical Standardization (CENELEC) EN 60601-1:2006+A12:2014, Ed: 3.1, Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance; IEC Corrigendum 1 (2006); IEC Corrigendum 2 (2007); CENELEC Corrigendum (2010); CENELEC Amendment A11 (2011); IEC Amendment 1 (2013); IEC Corrigendum 3 (2014); CENELEC Amendment A12 (2014)
12
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
www.stryker.com
7206-001-700 Rev-D
EN
Specifications (continued) Environmental Conditions:
Operation
Storage and Transportation
Temperature Limitation:
Humidity Limitation: Atmospheric Pressure Limitation:
www.stryker.com
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
13
EN
7206-001-700 Rev-D
Specifications (continued) Guidance and manufacturer’s declaration - electromagnetic emissions The System 7 Reciprocating Saw (REF 7206-000-000) is intended for use in the electromagnetic environment specified below. The customer or the user of the System 7 Reciprocating Saw (REF 7206-000-000) should assure that it is used in such an environment. Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The System 7 Reciprocating Saw (REF 7206-000-000) uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
Harmonic emissions IEC 61000-3-2
N/A
The System 7 Reciprocating Saw (REF 7206-000-000) is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Voltage fluctuations/flicker emissions IEC 61000-3-3
N/A
14
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
www.stryker.com
7206-001-700 Rev-D
EN
Specifications (continued) Guidance and manufacturer’s declaration - electromagnetic immunity The System 7 Reciprocating Saw (REF 7206-000-000) is intended for use in the electromagnetic environment specified below. The customer or the user of the System 7 Reciprocating Saw (REF 7206-000-000) should assure that it is used in such an environment. Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment guidance
Electrostatic discharge (ESD) IEC 61000-4-2
±6 kV contact ±8 kV air
±6 kV contact ±8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/ burst IEC 61000-4-4
±2 kV for power supply lines ±1 kV for input/output lines
N/A
Surge IEC 61000-4-5
±1 kV line(s) to line(s) ±2 kV line(s) to earth
N/A N/A
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
<5% UT (>95% dip in UT) for 0,5 cycle 40% UT (60% dip in UT) for 5 cycles 70% UT (30% dip in UT) for 25 cycles <5% UT (>95% dip in UT) for 5 seconds
N/A
3 A/m
3 A/m
Power frequency (50/ 60 Hz) magnetic field IEC 61000-4-8
N/A
N/A
N/A
N/A N/A
N/A
N/A Power frequency magnetic fields should be at levels characteristics of a typical location in a typical commercial or hospital environment.
NOTE: UT is the alternating current mains voltage prior to application of the test level.
www.stryker.com
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
15
EN
7206-001-700 Rev-D
Specifications (continued) Guidance and manufacturer’s declaration - electromagnetic immunity The System 7 Reciprocating Saw (REF 7206-000-000) is intended for use in the electromagnetic environment specified below. The customer or the user of the System 7 Reciprocating Saw (REF 7206-000-000) should assure that it is used in such an environment. Immunity test
IEC 60601 test level
Compliance level
Conducted RF IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz
N/A
Radiated RF IEC 61000-4-3
3 V/m 80 MHz to 2.5 GHz
3 V/m
Electromagnetic environment - guidance Portable and mobile RF communications equipment should be used no closer to any part of the System 7 Reciprocating Saw (REF 7206-000-000), including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance: d=1.2√P 80 MHz to 800 MHz d=2.3√P 800 MHz to 2.5 GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol: (Non-ionizing electromagnetic radiation)
NOTE 1: At 80 MHz and 800MHz the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the System 7 Reciprocating Saw (REF 7206000-000) is used exceeds the applicable RF compliance level above, the System 7 Reciprocating Saw (REF 7206-000-000) should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating System 7 Reciprocating Saw (REF 7206-000-000). b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
16
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
www.stryker.com
7206-001-700 Rev-D
EN
Specifications (continued) Recommended separation distances between portable and mobile RF communications equipment and the System 7 Reciprocating Saw (REF 7206-000-000) The System 7 Reciprocating Saw (REF 7206-000-000) is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the System 7 Reciprocating Saw (REF 7206-000-000) can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the System 7 Reciprocating Saw (REF 7206-000000) as recommended below, according to the maximum output power of the communications equipment. Rated maximum output power of transmitter W
Separation distance according to frequency of transmitter m 150 kHz to 80 MHz N/A
80 MHz to 800 MHz d=1.2√P
800 MHz to 2.5 GHz d=2.3√P
0.01
N/A
0.12
0.23
0.1
N/A
0.38
0.73
1
N/A
1.2
2.3
10
N/A
3.8
7.3
100
N/A
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1: At 80 MHz and 800MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
www.stryker.com
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
17
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM
Printed Date: Jul 03, 2018 22:49:23 GMT 0000043646, Rev. D Effective Date: May 3, 2018 11:04:11 AM