Stryker
System 8 Dual Trigger Rotary Handpiece Instructions for Use Rev AA Sept 2019
Instructions for Use
28 Pages
Preview
Page 1
System 8 Dual Trigger Rotary Handpiece REF
8205-000-000
Instructions For Use
ENGLISH (EN) 2019-09
8205-001-700 Rev-AA
www.stryker.com
EN
8205-001-700 Rev-AA
Contents
Definitions... 8
Introduction... 3
Instructions... 9
Audience... 3
To Install the Attachment and Accessory . . 9
Conventions... 3
To Install the Battery Pack... 10
Contact Information... 3
To Operate the Handpiece... 10
Indications For Use... 3
To Remove the Battery Pack... 12
Contraindications... 3
To Remove the Attachment and Accessory . 12
Safety Directives... 4
Care Instructions... 12
Accessories... 5
Troubleshooting... 13
Attachments... 5
Specifications... 17
Drill Bits... 5
Product Safety Certification... 18
Battery Packs... 6
Product Safety Compliance... 19
Insert Trays... 6
Electromagnetic Compatibility... 20
Features... 7 Handpiece... 7 Function Switch... 8 Shift Collar... 8
2
www.stryker.com
8205-001-700 Rev-AA
Introduction This instructions for use manual contains information intended to ensure the safe, effective, and compliant use of your product. Keep and consult this reference manual during the life of the product. NOTE: The user and/or patient should report any serious product-related incident to both the manufacturer and the Competent Authority of the European Member State where the user and/or patient is established.
Audience This manual is intended for in-service trainers, physicians, nurses, surgical technologists, and biomedical equipment technicians.
Conventions The following conventions are used in this manual: • A WARNING highlights a safety-related issue. ALWAYS comply with this information to prevent patient and/or healthcare staff injury.
EN
• A CAUTION highlights a product reliability issue. ALWAYS comply with this information to prevent product damage. • A NOTE supplements and/or clarifies procedural information.
Contact Information For additional information, including safety information, in-service training, or current literature, contact your Stryker sales representative or call Stryker customer service at 1-269-323-7700 or 1-800-253-3210. Outside the US, contact your nearest Stryker subsidiary.
Indications For Use The Stryker System 8 Dual Trigger Rotary Battery Powered Heavy Duty System is intended for use in the drilling, reaming, and decorticating of bone and other bone related tissue in a variety of surgical procedures. It is also usable in the placement of screws, wires, pins, and other fixation devices.
Contraindications None known.
www.stryker.com
3
EN
8205-001-700 Rev-AA
Safety Directives WARNINGS: • Before using this equipment, or any component compatible with this equipment, read and understand the instructions for use. Pay particular attention to safety information. Become familiar with the equipment before use. • Only healthcare professionals trained and experienced in the use of this medical device should operate this equipment. • The healthcare professional performing any procedure is responsible for determining the appropriateness of this equipment and the specific technique used for each patient. Stryker, as a manufacturer, does not recommend surgical procedure or technique.
• Upon initial receipt and before each use, operate the equipment and inspect each component for damage. DO NOT use any equipment if damage is apparent or the inspection criteria are not met. See the care instructions manual supplied with the handpiece. • Upon initial receipt and before each use, clean and sterilize the equipment as indicated. See the care instructions manual supplied with the handpiece. • DO NOT use this equipment in areas in which flammable anesthetics or flammable agents are mixed with air, oxygen, or nitrous oxide. • Take special precautions regarding electromagnetic compatibility (EMC) when using medical electrical equipment. Place this equipment into service according to the EMC information contained in this manual. Portable and mobile radio frequency (RF) communications equipment can affect the function of this equipment. • ALWAYS lock the handpiece triggers before installing or removing attachments or accessories.
4
www.stryker.com
8205-001-700 Rev-AA
Accessories WARNINGS: • Use only Stryker-approved electronic components and accessories. Failure to comply may result in increased electromagnetic emissions or decreased electromagnetic immunity of the system. • DO NOT modify any equipment without the authorization of the manufacturer. NOTES: • A variety of attachments are available for use with this handpiece. Each attachment has a specialized retainer for wires, pins, tools and/ or cutting accessories. See the instructions for use supplied with the attachment. • For a complete list of accessories, contact your Stryker sales representative or call Stryker customer service. Outside the US, contact your nearest Stryker subsidiary.
www.stryker.com
EN
The following Stryker-approved accessories are sold separately:
Attachments DESCRIPTION
REF (SERIES)
Heavy Duty Rotary Handpiece Attachments
8203-XXX-XXX 7203-XXX-XXX 6203-XXX-XXX
Drill Bits WARNINGS: • Upon initial receipt and before use, visually inspect the package for damage to confirm the integrity of the sterile barrier. Do not use the product if damage is apparent, the sterile barrier is compromised, or the package is unintentionally opened.
5
EN
8205-001-700 Rev-AA
• DO NOT reuse, reprocess, or repackage a device that is intended for single use only. -- A single use device may not withstand chemical, chemical vapor, or high temperature sterilization reprocessing. -- Design features may make cleaning difficult. -- Reuse may create a contamination risk and compromise structural integrity resulting in operational failure. -- Critical product information may be lost during repackaging. Failure to comply may lead to infection or cross infection and result in patient and/or healthcare staff injury. NOTE: Sterile cutting accessories are sterilized by irradiation.
6
DESCRIPTION
REF
Drill Bits
0277-082-08X series
Battery Packs DESCRIPTION
REF
System 8 Battery Packs
8212-000-000 8215-000-000
SmartLife® Battery Packs
7212-000-000 7215-000-000
SmartLife Non-sterile Batteries
7126-110-000 7222-110-000
SmartLife Aseptic Housings
7126-120-000 7222-120-000
SmartLife Transfer Shields
7126-130-000 7222-130-000
System 6 Battery Packs
6212-000-000 6215-000-000
System 6 Aseptic Battery Kits
6126-000-000 6127-000-000
Insert Trays DESCRIPTION
REF
Insert Trays
7102-450-010 7102-452-010 7102-454-010 7102-458-010
www.stryker.com
8205-001-700 Rev-AA
EN
Features
A
NOTE: The Stryker System 8 Dual Trigger Rotary Handpiece (handpiece) is a component of the Stryker System 8 Battery Powered Heavy Duty System.
Function Switch – Locks one or both of the triggers. See the Function Switch section.
B
Shift Collar – Sets the speed and torque. See the Shift Collar section.
C
Attachment Collar – Retains the attachment in the handpiece.
D
Reverse Trigger – Controls the variable speed operation of the handpiece in a counterclockwise direction.
E
Forward Trigger – Controls the variable speed operation of the handpiece in a clockwise direction.
Handpiece The System 8 Dual Trigger Rotary Handpiece is battery powered and has two triggers and a function switch.
B C
NOTE: Press both triggers simultaneously to operate the handpiece in oscillate mode.
A
D E
www.stryker.com
7
EN
8205-001-700 Rev-AA
Function Switch Forward Mode – Only the forward trigger is functional. The reverse trigger is locked to prevent inadvertent operation of the handpiece in a counterclockwise direction.
Definitions The symbols located on the equipment and/ or labeling are defined in this section or in the Symbol Definition Chart. See the Symbol Definition Chart supplied with the equipment. SYMBOL
General warning sign
Forward/Reverse/Oscillate Mode – Both triggers are functional. Safe Mode – Both triggers are locked to prevent inadvertent operation of the handpiece.
DEFINITION
Drill Mode Ream Mode F/R 1 min / 4 min x 3 OSC 15s/15s x5
Duty Cycle – See the Specifications section.
Shift Collar Drill Mode – Fast speed, low torque operation
Ream Mode – Slow speed, high torque operation
8
www.stryker.com
8205-001-700 Rev-AA
Instructions To Install the Attachment and Accessory
EN
NOTE: If the attachment has notches, align one of the notches with a tab under the attachment collar before insertion.
1. Lock the handpiece triggers. 2. Insert the attachment into the handpiece until the attachment snaps into place.
3. Gently tug the attachment to make sure the attachment is securely locked in the handpiece. 4. Install a wire, pin, tool, or cutting accessory as required. See the instructions for use supplied with the attachment.
www.stryker.com
9
EN
To Install the Battery Pack NOTE: See the instructions for use supplied with the battery pack and/or battery charger for charging instructions and specifications. 1. Lock the handpiece triggers. 2. Slide a fully charged battery pack into the handpiece until the battery pack snaps into place.
8205-001-700 Rev-AA
To Operate the Handpiece WARNINGS: • ALWAYS lock the handpiece triggers when the handpiece is idle or when passing the handpiece to another person. • Before operating the handpiece, ALWAYS gently tug the attachment and accessory to make sure the attachment is securely locked in the handpiece and the accessory is securely locked in the attachment. • DO NOT grasp or touch any rotating component while the handpiece is operating. • DO NOT change the position of the function switch while the handpiece is operating. • DO NOT change the position of the shift collar while the handpiece is operating.
3. Gently tug the battery pack to make sure the battery pack is securely locked in the handpiece.
• DO NOT operate the handpiece in the drill mode position when a reamer attachment and/or accessory is installed in the handpiece.
4. Test the operation of the handpiece by unlocking and then depressing the trigger.
10
www.stryker.com
8205-001-700 Rev-AA
• ALWAYS operate the equipment within the specified environmental condition values. See the Specifications section. • ALWAYS follow the recommended duty cycle to prevent the equipment from overheating. See the Specifications section and/or the instructions for use supplied with the attachment. • DO NOT apply excessive pressure, such as bending or prying, with the accessory. Excessive pressure may bend or fracture the accessory and result in tissue damage, loss of tactile control, and/or the ejection of accessory fragments at a high velocity. • A wobbling attachment and/or accessory may cause bone or tissue damage or inaccurate wire or pin placement. If wobbling occurs, take corrective action as indicated in the Troubleshooting section.
www.stryker.com
EN
CAUTIONS: • DO NOT stall the handpiece. Failure to comply may damage the electric motor and/or battery pack. If the handpiece jams, release the trigger immediately. Remove any obstructions before continuing to operate the handpiece. • If any power loss is experienced while using the handpiece, ALWAYS replace the battery pack with a fully charged battery pack. Failure to comply may result in a drained or damaged battery pack with a shortened life. NOTE: See the Features section for mode descriptions. 1. Rotate the shift collar to the drill mode or ream mode position. 2. Slide the function switch to the forward or forward/reverse/oscillate position. 3. Depress one of the pressure-sensitive triggers to operate the handpiece.
11
EN
To Remove the Battery Pack 1. Lock the handpiece triggers. 2. Depress the battery latch and slide the battery pack out of the handpiece.
8205-001-700 Rev-AA
To Remove the Attachment and Accessory 1. Lock the handpiece triggers. 2. Remove the accessory from the attachment. 3. Pull the attachment collar back and remove the attachment from the handpiece.
Care Instructions For processing instructions and disposal/recycle information, see the care instructions manual supplied with the equipment.
12
www.stryker.com
8205-001-700 Rev-AA
EN
Troubleshooting WARNING: DO NOT disassemble or service this equipment without the authorization of the manufacturer. NOTE: For service, contact your Stryker sales representative or call Stryker customer service. Outside the US, contact your nearest Stryker subsidiary. PROBLEM
CAUSE
The handpiece does not operate The battery pack is discharged. or operates at a reduced speed.
The handpiece operates but the cutting accessory does not move.
ACTION Use a Stryker battery charger to recharge the battery pack.
The battery pack is expended.
Replace the battery pack.
The handpiece trigger is locked.
Unlock the handpiece trigger. See the Features section.
The handpiece is damaged.
Return the equipment to Stryker for repair.
The attachment is not fully installed in the handpiece.
Remove and install the attachment. Make sure the attachment is securely locked in the handpiece.
The shift collar is positioned between the drill and ream mode positions.
Fully rotate the shift collar to either the drill mode or ream mode position.
Excessive pressure is being Release the pressure and allow applied to the cutting accessory. the cutting accessory to do the cutting. The handpiece is damaged.
www.stryker.com
Return the equipment to Stryker for repair.
13
EN
8205-001-700 Rev-AA
PROBLEM
CAUSE
ACTION
The handpiece continues to operate after the trigger is released.
The handpiece is damaged.
Depress the battery latch and slide the battery pack out of the handpiece. Return the equipment to Stryker for repair.
The equipment becomes unusually hot during use.
The duty cycle has been exceeded.
ALWAYS follow the recommended duty cycle to prevent the equipment from overheating. See the Specifications section and/or the instructions for use supplied with the attachment.
The handpiece is damaged.
Return the equipment to Stryker for repair.
The battery pack is damaged.
Use a Stryker battery charger to check the integrity of the battery pack. See the instructions for use supplied with the battery charger for more information. Replace the battery pack if required.
14
www.stryker.com
8205-001-700 Rev-AA
EN
PROBLEM
CAUSE
The attachment will not fit or cannot be secured in the handpiece.
The attachment and/or the distal See the care instructions end of the handpiece contains manual supplied with the debris. handpiece.
The pin accessory cannot be removed from the collet attachment.
www.stryker.com
ACTION
The attachment is damaged.
Return the equipment to Stryker for repair.
The handpiece is damaged.
Return the equipment to Stryker for repair.
The collet mechanism is jammed.
Gently tap the side of the pin to dislodge it from the collet. CAUTION: DO NOT attempt to force the pin back through the collet.
The collet is damaged.
Return the equipment to Stryker for repair.
15
EN
8205-001-700 Rev-AA
PROBLEM
CAUSE
ACTION
The attachment and/or accessory wobbles in the handpiece.
The accessory is damaged.
Inspect the accessory for damage and replace the accessory as required.
The accessory is the wrong size for the attachment.
Install a different accessory or attachment as required.
The accessory extends too far from the distal end of the attachment, or is not properly centered in the attachment.
Remove and properly install the accessory. If wobble persists, return the handpiece and attachment to Stryker for repair.
A reamer attachment and/or accessory is installed in the handpiece and the shift collar is in the drill mode position.
Rotate the shift collar to the ream mode position.
The handpiece is noisy and/or vibrates.
The handpiece is damaged.
Return the equipment to Stryker for repair.
The handpiece experiences sporadic electrical interference.
Electrical noise is present.
Turn off all electrical equipment not in use in the operating room. Relocate electrical equipment and/or increase spatial distance between electrical equipment. Plug operating room equipment into different operating room outlets.
16
www.stryker.com
8205-001-700 Rev-AA
EN
Specifications WARNING: ALWAYS consult any documentation that accompanies attachments and/or accessories for product-specific duty cycles and instructions for use. CAUTION: ALWAYS store the equipment within the specified environmental condition values throughout its useful life. Model:
System 8 Dual Trigger Rotary Handpiece (REF 8205-000-000)
Dimensions:
150 mm [5.9 inch] height, 39 mm [1.5 inch] width, 154 mm [6.1 inch] length
Mass:
1.04 kg [2.3 lb]
Speed:
1200 rpm (drill mode), 270 rpm (ream mode) Forward and Reverse Modes
Oscillate Mode
Mode of Operation:
Non-continuous
Non-continuous
Duty Cycle:
1 minute on/4 minutes off, 3 times
15 seconds on/15 seconds off, 5 times
Rest Between Cycles:
3 hours
1.5 hours
Applied Part(s):
The distal end of the handpiece and the attachment as defined by the manufacturer
Maximum Temperature Less than 51 °C [124 °F] as tested to the Product Safety Certification of Applied Part(s): standards Power Supply:
Internally powered. Refer to battery housing for voltage rating.
Ingress Protection:
IPX9 during cleaning and sterilization
Equipment Type:
www.stryker.com
Type BF Applied Part
17
EN
8205-001-700 Rev-AA
Environmental Conditions:
Operation
Storage and Transportation
Temperature Limitation: 75 %
Humidity Limitation: 10 %
Atmospheric Pressure Limitation:
Product Safety Certification Canadian Standards Association (CSA) International Canadian Standards Association (CSA) CAN/CSA-C22.2 No. 60601-1:14, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance; (IEC 60601-1:2005+A1:2012, MOD) American National Standards Institute (ANSI)/Association for the Advancement of Medical Instrumentation (AAMI) ANSI/AAMI ES60601-1:2005/(R) 2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance; Consolidated Reprint (2009/(R) 2012); Amendment 2 (2010/(R) 2012); Amendment 1 (2012)
18
www.stryker.com
8205-001-700 Rev-AA
EN
Product Safety Compliance International Electrotechnical Commission (IEC) IEC 60601-1:2005, Ed: 3.1, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance; Corrigendum 1 (2006); Corrigendum 2 (2007); Amendment 1 (2012) IEC 60601-1-2:2014 Ed: 4, Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance – Electromagnetic Disturbances IEC 60601-1-2:2007 Ed: 3, Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance – Electromagnetic Compatibility IEC 60601-1-6:2010+ A1:2013 Ed. 3.1, Medical Electrical Equipment – Part 1-6: General Requirements for Basic Safety and Essential Performance – Usability IEC 62366-1:2007+ A1:2014 Ed 1.1, Medical Devices - Part 1: Application of Usability Engineering to Medical Devices European Committee for Electrotechnical Standardization (CENELEC) EN 60601-1:2006+A12:2014, Ed: 3.1, Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance; IEC Corrigendum 1 (2006); IEC Corrigendum 2 (2007); CENELEC Corrigendum (2010); CENELEC Amendment A11 (2011); IEC Amendment 1 (2013); IEC Corrigendum 3 (2014); CENELEC Amendment A12 (2014)
www.stryker.com
19
EN
8205-001-700 Rev-AA
Electromagnetic Compatibility Guidance and manufacturer’s declaration - electromagnetic emissions The System 8 Dual Trigger Rotary Handpiece (REF 8205-000-000) is intended for use in the electromagnetic environment specified below. The customer or the user of the System 8 Dual Trigger Rotary Handpiece (REF 8205-000-000) should assure that it is used in such an environment. Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
Group 1
The System 8 Dual Trigger Rotary Handpiece (REF 8205-000000) uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Class B
The System 8 Dual Trigger Rotary Handpiece (REF 8205-000000) is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
CISPR 11 RF emissions CISPR 11 Harmonic emissions
N/A
IEC 61000-3-2 Voltage fluctuations/flicker emissions
N/A
IEC 61000-3-3
20
www.stryker.com