User Guide
32 Pages
Preview
Page 1
X8000 Light Source REF 220-200-000
User Guide English Deutsch Español Français, Italiano Português Nederlands Dansk Suomi Svenska Ελληνικά Polski 日本語 中文 한국어
2006/09
www.stryker.com
1000-400-885 D
Contents Warnings and Cautions ... 2 Symbol Definitions ... 4
Product Description and Intended Use ... 5 Setup and Assembly ... 8 Connecting the AC Power Cable ... 8 Connecting the Light Cable ... 8
System Operation ... 10 Powering the System On and Off ... 10 Selecting the Operation Mode ... 11 Adjusting the Brightness ... 12 Visual Display ... 12 Language Selection ... 13 Safety Shutoff ... 14 Checking the ESST Feature ... 15 Using the X8000 with a Voice-Controlled System Interface ... 16 Using the SFB Serial Interface ... 16 Troubleshooting ... 17
Cleaning and Maintenance ... 19 Cleaning the X8000 ... 19 Caring for the Bulb Module ... 19 Replacing the Bulb Module ... 20 Replacing the Fuses ... 21 Disposing of the X8000 ... 21
Technical Specifications ... 22 Electromagnetic Compatibility ... 23
Warranty ... 27 Service and Claims ... 28
1
Warnings and Cautions Please read this manual and follow its instructions carefully. The words warning, caution, and note carry special meanings and should be carefully reviewed: WARNING
THE PERSONAL SAFETY OF THE PATIENT OR USER MAY BE INVOLVED. DISREGARDING THIS INFORMATION COULD RESULT IN SERIOUS INJURY TO THE PATIENT OR USER.
Caution
Special procedures or precautions must be followed to avoid damaging the instrument.
Note
Special information to make maintenance easier or important information more clear. An exclamation mark within a triangle is intended to alert the user to the presence of important operating and maintenance instructions in the manual. A lightning bolt within a triangle is intended to warn of the presence of hazardous voltage. Refer all service to authorized personnel.
WARNING
IMPORTANT SAFETY NOTICE: BEFORE OPERATING THIS DEVICE, PLEASE READ THIS OPERATING MANUAL THOROUGHLY AND CAREFULLY. WHEN USING THIS DEVICE WITH A LIGHT SOURCE, FIRE AND/OR SEVERE INJURY MAY RESULT TO THE PATIENT, USER OR INANIMATE OBJECTS, IF THE INSTRUCTIONS IN THIS MANUAL ARE NOT FOLLOWED. ALL LIGHT SOURCES, INCLUDING THE X8000, CAN GENERATE SIGNIFICANT AMOUNTS OF HEAT AT THE SCOPE TIP, THE SCOPE LIGHT POST, THE LIGHT CABLE TIP, AND/OR NEAR THE LIGHT CABLE ADAPTER. HIGHER LEVELS OF BRIGHTNESS FROM THE LIGHT SOURCE RESULT IN HIGHER LEVELS OF HEAT. ALWAYS ADJUST THE BRIGHTNESS LEVEL OF THE CAMERA AND THE MONITOR, BEFORE ADJUSTING THE BRIGHTNESS LEVEL OF THE LIGHT SOURCE. ADJUST THE BRIGHTNESS LEVEL OF THE LIGHT SOURCE TO THE MINIMUM BRIGHTNESS NECESSARY TO ADEQUATELY ILLUMINATE THE SURGICAL SITE. IN ADDITION, ADJUST THE INTERNAL SHUTTER OF THE CAMERA HIGHER IN ORDER TO RUN THE LIGHT SOURCE AT A LOWER INTENSITY. AVOID TOUCHING THE SCOPE TIP OR THE LIGHT CABLE TIP TO THE PATIENT, AND NEVER PLACE THEM ON TOP OF THE PATIENT, AS DOING SO MAY RESULT IN BURNS TO THE PATIENT OR USER. IN ADDITION, NEVER PLACE THE SCOPE TIP, THE SCOPE LIGHT POST, THE LIGHT CABLE ADAPTER, OR THE LIGHT CABLE TIP ON THE SURGICAL DRAPES OR OTHER FLAMMABLE MATERIAL, AS DOING SO MAY RESULT IN FIRE.
2
ALWAYS PLACE THE LIGHT SOURCE IN STANDBY MODE WHENEVER THE SCOPE IS REMOVED FROM THE LIGHT CABLE OR THE DEVICE IS UNATTENDED. THE SCOPE TIP, SCOPE LIGHT POST, LIGHT CABLE ADAPTER, AND LIGHT CABLE TIP WILL TAKE SEVERAL MINUTES TO COOL OFF AFTER BEING PLACED IN STANDBY MODE, AND THEREFORE MAY STILL RESULT IN FIRE OR BURNS TO THE PATIENT, USER, OR INANIMATE OBJECTS.
WARNING
TO HELP AVOID POTENTIAL SERIOUS INJURY TO THE USER AND THE PATIENT AND/OR DAMAGE TO THIS DEVICE, THE USER MUST:
1.
Read this operating manual thoroughly, especially the warnings, and be familiar with its contents prior to using this equipment.
2.
Carefully unpack the unit and check if any damage occurred during shipment. If damage is detected, please refer to the “Service and Claims” section in this manual.
3.
Be a qualified physician, having complete knowledge of the use of this equipment.
4.
Test this equipment prior to a surgical procedure. This unit was fully tested at the factory before shipment.
5.
Attempt no internal repairs or adjustments that are not specifically detailed in the Troubleshooting, Cleaning and Maintenance sections of this operating manual.
6.
Never sterilize any part of the X8000 console.
7.
Disconnect the X8000 from the electrical outlet when inspecting the fuses.
The warranty is void if any of these warnings or cautions contained in this manual are disregarded. The user must also be ensure that: • Readjustments, modifications, and/or repairs are carried out exclusively by Stryker Endoscopy. • The electrical installation of the relevant operating room complies with the applicable IEC, CEC, and NEC requirements. WARNING
FEDERAL LAW (UNITED STATES OF AMERICA) RESTRICTS THIS DEVICE TO USE BY, OR ON ORDER OF, A PHYSICIAN.
3
Symbol Definitions Type CF Applied Part Protective Earth Ground
Equipotentiality
Denotes compliance to CSA 22.2 No.601.1-M90 and UL60601-1.
This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact the manufacturer or other authorized disposal company to decommission your equipment. Caution: Surface may be hot
4
Product Description and Intended Use The Stryker X8000 Light Source is a light-generating unit designed to illuminate surgical sites during endoscopic applications. The X8000 uses a 300-watt xenon bulb to generate bright, crisp light, which it delivers to the surgical site via a fiberoptic light cable. The X8000 is compatible with all Stryker light cables, and, with the proper light cable and adapters, can connect to any flexible or rigid endoscope. The X8000 is equipped with Electronic Scope Sensing Technology (ESST), a special safety feature that helps prevent accidental burns caused by a light cable that is not connected to the scope. For more information, refer to the section “Checking the ESST Feature”. When operated with an ESST light cable, the X8000 senses when the scope and the light cable are separated and places the light source in Standby mode. In Standby mode, the X8000 will reduce light output to a minimum, preventing the light cable from generating excessive heat. WARNING
THE SCOPE TIP, SCOPE LIGHT POST, LIGHT CABLE ADAPTER, AND LIGHT CABLE TIP WILL TAKE SEVERAL MINUTES TO COOL OFF AFTER BEING PLACED IN STANDBY MODE AND THEREFORE MAY STILL RESULT IN FIRE AND/OR BURNS TO THE PATIENT, USER, OR AN INANIMATE OBJECT IF NOT USED PROPERLY. DO NOT PLACE THE SCOPE OR THE LIGHT CABLE ON THE PATIENT OR ON THE DRAPES OR OTHER FLAMMABLE MATERIAL, EVEN WHEN THE DEVICE IS IN STANDBY MODE.
The Stryker X8000 Light Source consists of one of each of the following: • light source console • power cord • xenon bulb module (spare or replacement bulb modules can be purchased separately as P/N 220-201-000)
5
The features of the X8000 console are described below (see Figures 1 and 2). 5
6
7
4 3 2
1
9
8
Figure 1 Front panel of the X8000 console
6
1.
Power Button: Powers the unit on and off.
2.
Standby LED: Illuminates when the unit is operating in the Standby mode. In Standby mode, light output is reduced to a minimum level.
3.
Mode Button: Selects either Standby or Run mode when pressed. Standby is the default mode when the light source is powered on.
4.
Run LED: Illuminates when the unit is operating in the Run mode. In Run mode, light output is determined by the up and down switches.
5.
LCD: Indicates the light intensity level (0-100%), bulb hours and language.
6.
Up Button: Increases light intensity in Run mode.
7.
Cable Clamp: Grasps the light-source end of an inserted fiberoptic cable. (When no cable is inserted, the bulb automatically turns off to save bulb hours.)
8.
Jaw Handle: Opens the fiberoptic-cable holder.
9.
Down Button: Decreases light intensity in Run mode.
10
11
4USZLFS&OEPTDPQZ 0QUJDBM$PVSU 4BO+PTF $"64"
4*%/&
641BUFOUT 0UIFSQBUFOUTQFOEJOH
."%&*/64"
14
12, 13
Figure 2 Rear panel of the X8000 console
10. Voice-Control Port: Connects to Stryker voice-control systems. 11. SFB Series Connectors: Enables FireWire connection with Stryker FireWire devices. Provides connection for remote diagnoses and future software upgrades. 12. AC Inlet: Connects to the provided power cord for AC power supply. 13. Fuse Holder: Contains two 5A fuses. 14. Equipotentiality Plug: Provides a grounding post for common grounding with other equipment.
7
Setup and Assembly Note
Your local Stryker Endoscopy sales representative will perform inservice training at your convenience to instruct you and your staff on the safe and effective operation and maintenance of the X8000, and to help set up the equipment. Please contact your local Stryker Endoscopy sales representative to schedule an inservice after the equipment has arrived.
Note
When selecting a setup location for the X8000, consult the “Electromagnetic Compatibility” section included in this manual to determine the best location.
To set up the X8000, make the following connections: • Connect the AC power cable • Connect the light cable
Connecting the AC Power Cable 1.
Plug in the AC power cord to the AC Inlet on the rear console panel.
2.
Plug in the other end of the AC cord to a hospital-grade outlet.
WARNING
WHEN THE X8000 IS INTERCONNECTED WITH OTHER MEDICAL ELECTRICAL EQUIPMENT, LEAKAGE CURRENTS MAY BE ADDITIVE. TO MINIMIZE LEAKAGE CURRENT THAT MAY TRAVEL TO THE PATIENT OR USER, ANY TYPE CF APPLIED PART SHOULD BE USED ONLY WITH OTHER TYPE CF APPLIED PARTS. ENSURE ALL SYSTEMS ARE INSTALLED ACCORDING TO THE REQUIREMENTS OF
IEC 60601-1-1.
Connecting the Light Cable WARNING
USE ONLY NONCONDUCTIVE FIBEROPTIC CABLES WITH THE X8000 TO MAINTAIN ELECTRICAL ISOLATION.
Note
The X8000 Light Source is compatible with all Stryker Light Cables.
8
1.
Lock open the cable clamp by turning the jaw handle clockwise until it stops (see Figure 3).
Figure 3 Locking open the cable clamp
WARNING
KEEP FINGERS AWAY FROM THE CABLE CLAMP AS THE CLAMP MAY INADVERTENTLY DEPLOY AND CAUSE INJURY.
WARNING
DO NOT LOOK DIRECTLY INTO THE CABLE PORT. THE HIGHINTENSITY LIGHT MAY CAUSE DAMAGE TO THE EYES.
2.
Insert a clean, dry fiberoptic cable into the cable port until the jaw latch releases and the jaw clamps the cable in place (see Figure 4). Pull gently on the fiberoptic cable to test that it is securely seated in the cable port.
Figure 4 Inserting the light cable into the cable port
3.
Connect an endoscope to the opposite end of the fiberoptic cable.
4.
To remove the light cable, press the mode button to put the unit into standby. Then, turn the jaw handle clockwise until the latch opens.
9
WARNING
IF THE LIGHT SOURCE IS NOT IN STANDBY MODE PRIOR TO REMOVING THE CABLE, THE HIGH INTENSITY LIGHT WILL SHINE DIRECTLY OUT OF THE LIGHT SOURCE BRIEFLY BEFORE TURNING OFF, POSSIBLY CAUSING INJURY TO THE USER’S EYES.
Note
The Light Source will default to Standby mode when a light cable is inserted.
System Operation Note
Before operating the X8000, see the “Setup and Assembly” section of this manual.
Powering the System On and Off Caution
Do not power the system on and off in rapid succession. Allow the bulb to run for at least five minutes once it has been powered on, before turning the power off. Failure to do so can cause the bulb to rapidly darken and fail.
To power on the X8000: 1.
Confirm that the X8000 contains a properly installed bulb. The bulb handle must be turned all the way to the left, to ensure proper latching. If no bulb is installed, follow the instructions in the “Replacing the Bulb” section of this manual before proceeding.
2.
Press the power switch on the front panel. The Standby LED will illuminate, indicating the unit is in Standby mode.
Note
The bulb will not illuminate unless a light cable is installed in the cable port.
To power off the X8000: 1.
Put the light source unit into Standby mode.
2.
Disconnect the light cable from the X8000 console.
3.
Run the fan for at least one minute to cool the unit.
4.
Press the power switch on the front panel of the X8000.
WARNING
10
TO ALLOW FOR ADEQUATE COOLING, NEVER BLOCK THE REAR OR SIDE FAN VENTS. FAILURE TO FOLLOW THIS INSTRUCTION MAY RESULT IN DAMAGE TO THE X8000 OR A POSSIBLE FIRE.
Selecting the Operation Mode The X8000 has two operation modes, Run and Standby. • RUN mode: The Run mode is used during normal operation. It enables light output to be controlled by the brightness controls on the front console panel. • STANDBY mode: The Standby mode is used when the X8000 is powered on, but not in use. It reduces the light output to a minimum, thereby reducing the heat generated at the tip of the light cable or scope when the X8000 is not being used. To select either the Run or Standby mode, press the Mode Button. The LED for the selected mode will illuminate. WARNING
TO HELP PREVENT BURNS TO THE PATIENT, USER OR INANIMATE OBJECTS AND POSSIBLE FIRE, ALWAYS PUT THE X8000 IN STANDBY MODE WHENEVER THE ENDOSCOPE IS REMOVED FROM THE LIGHT CABLE. THE SCOPE TIP, SCOPE LIGHT POST, LIGHT CABLE ADAPTER, AND LIGHT CABLE TIP WILL TAKE SEVERAL MINUTES TO COOL OFF AFTER BEING PLACED IN STANDBY MODE AND THEREFORE MAY STILL RESULT IN FIRE AND/OR BURNS TO THE PATIENT, USER OR INANIMATE OBJECTS IF NOT USED PROPERLY. DO NOT PLACE THE SCOPE OR THE LIGHT CABLE ON THE PATIENT, ON THE DRAPES, OR OTHER FLAMMABLE MATERIAL, EVEN WHEN THE DEVICE IS IN STANDBY MODE.
11
Adjusting the Brightness The X8000 has up and down buttons for adjusting the bulb brightness. Press the up arrow to increase brightness, and press the down arrow to decrease brightness. The selection will appear on the LCD as a percentage between 0 and 100. When in Standby mode, the previous Run mode brightness level appears in the upper right-hand corner. The Run mode brightness level can be reset while still in Standby mode, by pressing the up and down buttons. WARNING
THE HIGHER THE BRIGHTNESS, THE MORE HEAT ENERGY THAT WILL BE GENERATED IN THE SCOPE AND THE TIP OF THE CABLE.
ALWAYS ADJUST THE BRIGHTNESS LEVEL OF THE CAMERA AND THE MONITOR BEFORE ADJUSTING THE BRIGHTNESS LEVEL OF THE LIGHT SOURCE. ADJUST THE BRIGHTNESS LEVEL OF THE LIGHT SOURCE TO THE MINIMUM BRIGHTNESS NECESSARY TO ILLUMINATE THE SURGICAL SITE. FAILURE TO FOLLOW THIS INSTRUCTION COULD RESULT IN FIRE OR BURNS TO THE PATIENT, USER OR AN INANIMATE OBJECT. IN ADDITION, ADJUST THE INTERNAL SHUTTER OF THE CAMERA HIGHER IN ORDER TO RUN THE LIGHT SOURCE AT A LOWER INTENSITY.
WARNING
THE SURFACE TEMPERATURE NEAR THE SCOPE ADAPTER AND AT THE TIP OF THE SCOPE MAY EXCEED 41ºC IF THE UNIT IS OPERATED AT HIGH LEVELS OF BRIGHTNESS FOR EXTENDED PERIODS OF TIME. THE HEATED SCOPE AND ADAPTER MAY CAUSE BURNS TO THE PATIENT, USER OR AN INANIMATE OBJECT.
Visual Display The X8000 provides feedback through the LCD. • Brightness: The LCD display shows the intensity level of the light as a percentage between 0 and 100. For example, if the LCD shows “70,” the light output to the fiberoptic cable is running at 70 percent of capacity. • Bulb Hours: The LCD display shows how many total hours the bulb has operated. For example, if the LCD shows “250,” the bulb has operated 250 total hours. The bulb hours will flash after 500 hours, and the LCD will display “Replace Bulb” after 1000 hours.
12
• The LCD also displays warning and error codes. The table below lists and defines the warning and error codes displayed. Code
Definition
Recommended Action
E-1
All conditions are met for the bulb to illuminate, yet it remains off.
Return the X8000 for repair.
E-2
All conditions are not met for the bulb to illuminate, yet it remains on.
Return the X8000 for repair.
E-3
The bulb has higher than expected current or voltage applied to it.
Return the X8000 for repair.
E-4
The bulb is kept off because the ballast fan is not working properly.
Return the X8000 for repair.
E-5
The bulb is kept off because the bulb fan is not working properly.
Return the X8000 for repair.
Blinking 500
The bulb has exceeded its recommended lifetime of 500 hours.
Install a new Stryker bulb.
Replace Bulb
The bulb has reached 1000 hours and has exceeded its recommended lifetime of 500 hours.
Install a new Stryker bulb.
Language Selection The X8000 LCD has the capability of displaying text in the following languages: Danish Dutch English Finnish French
German Greek Italian Japanese Korean
Polish Portuguese Simplified Chinese Spanish Swedish
To select a particular language, perform the following steps: 1.
Hold down the Up and Down buttons until the current language identifier appears.
2.
Use Mode button to scroll through the available languages.
3.
Hold down the Up and Down buttons (until the light source resumes normal operation) to lock in the selected language.
13
Safety Shutoff The X8000 Light Source is equipped with a Safety Shutoff feature which will temporarily turn off the bulb in the event of excessive heat in the bulb assembly. WARNING
ONCE THE LIGHT SOURCE COOLS DOWN (AFTER 7-10 MINUTES), POWER WILL RESUME TO THE BULB AND THE UNIT WILL RESTART IN STANDBY MODE. TO PREVENT FIRES AND ACCIDENTAL BURNS TO THE PATIENT, USER OR INANIMATE OBJECTS, ALWAYS PLACE THE SCOPES AND/OR FIBEROPTIC CABLES IN A SAFE PLACE, AND NOT ON THE PATIENT, DRAPES, OR OTHER FLAMMABLE MATERIAL, TO ENSURE SAFE RESUMPTION OF LIGHT OUTPUT. IF THE X8000
LIGHT SOURCE EXPERIENCES A TEMPORARY SHUTDOWN, IT IS RECOMMENDED THAT THE DEVICE BE RETURNED FOR SERVICE.
Caution
14
Do not abruptly interrupt power to the unit. This will turn off the fan and may cause severe damage to the internal cooling system.
Checking the ESST Feature The X8000 is equipped with Electronic Scope Sensing Technology (ESST), a special safety feature that helps prevent accidental fires or burns to the patient or user caused by a light cable not connected to a scope. This feature will only work if the X8000 is used with an ESST light cable. When operated with an ESST light cable, the X8000 senses when the scope and light cable are separated, and places the light source in Standby mode. In Standby mode, the X8000 will reduce light output to a minimum, preventing the light cable from generating excessive heat. To verify the ESST feature is active, perform the following test before every surgical procedure: 1.
Set up the X8000 system with an ESST light cable and scope, and then power on the system.
2.
Place the X8000 in Run mode.
3.
Remove the light cable from the ESST scope adapter.
The X8000 should return to Standby mode, indicating that the ESST feature is functioning properly. WARNING
ALWAYS CHECK TO ENSURE THAT THE UNIT HAS SWITCHED INTO STANDBY MODE, BEFORE ASSUMING ESST SAFETY PROTECTION. IF THE UNIT FAILS TO RETURN TO STANDBY MODE, THERE MAY BE A FAULT WITH THE ESST FEATURE. IN THIS CASE, DO NOT ASSUME ESST SAFETY PROTECTION, AND RETURN THE UNIT FOR SERVICE.
WARNING
EVEN WITH ESST PROTECTION, OR WHEN THE X8000 IS IN STANDBY MODE, NEVER PLACE THE TIP OF THE LIGHT CABLE OR LIGHT CABLE ADAPTER DIRECTLY ON THE PATIENT, DRAPES OR OTHER FLAMMABLE MATERIAL, AS BURNS OR FIRE MAY RESULT. THE SCOPE TIP, SCOPE LIGHT POST, LIGHT CABLE ADAPTER, AND LIGHT CABLE TIP WILL TAKE SEVERAL MINUTES TO COOL OFF AFTER BEING PLACED IN STANDBY MODE, AND THEREFORE MAY STILL RESULT IN FIRE AND/OR BURNS TO THE PATIENT, USER, OR AN INANIMATE OBJECT IF NOT USED PROPERLY. DO NOT PLACE THE SCOPE OR THE LIGHT CABLE ON THE PATIENT, ON THE DRAPES, OR OTHER FLAMMABLE MATERIAL, EVEN WHEN THE DEVICE IS IN STANDBY MODE.
15
Using the X8000 with a Voice-Controlled System Interface The X8000 can be used in conjunction with Stryker voice-control systems (SIDNE®). For more information about using the X8000 with Stryker voicecontrol systems, refer to the SIDNE® Operating and Maintenance Manual (P/N 1000-400-653).
Using the SFB Serial Interface The SFB serial connection on the rear panel of the X8000 enables FireWire connection to the Stryker Endoscopy Software Management site. Connecting to this site enables remote diagnostics and software updates. Note
This system feature is not necessary for regular light-source operation.
Note
This system feature requires an additional device (i.e., a computer) to connect to the Software Management site.
16
Troubleshooting Problem
Possible Solution
No light output
• Ensure the AC power cord is properly connected to a hospital-grade power outlet and the inlet on the rear console panel. • Ensure the power switch on the front panel is powered on. (It will illuminate when powered on.) • Ensure all fuses are operating. See the “Replacing the Fuses” section of this manual for further instructions. • Ensure the bulb is properly seated in the bulb housing. The bulb handle should be turned all the way to the left within the light source. • Ensure the bulb is in operating condition. Replace the bulb if necessary. • Ensure the light cable is correctly engaged with the cable port. As a safety feature, the X8000 will provide no light output unless a fiberoptic light cable is properly seated in the cable port. • Ensure the bulb access door is completely shut. • Check for error codes E-1, E-2, E-3, E-4 or E-5. See the “Visual Display” section of this manual for details. • Check that vents are not obstructed. • If the safety shutoff has been activated, please return the X8000 for service. Please see the section entitled “Safety Shutoff ” in this manual for additional information.
17
Too much or too little light output
• Ensure the light cable is correctly engaged with the cable port. • Ensure the bulb has adequate bulb life remaining. The bulb has a warranty of 500 hours. • Ensure the X8000 is in Run mode. (The Run LED should be illuminated.) If necessary, press the Mode button to switch from Standby to Run. If the unit remains in Standby: 1. Ensure the light cable is correctly engaged with the cable port. 2.
If an ESST cable is connected to the X8000, ensure the cable is attached to the scope using an ESST scope adapter. • Use the up/down buttons to adjust the brightness. For details, see the “Adjusting the Brightness” section in this manual. • Ensure the fiberoptic cable is transmitting light properly. Hold the light-source end of the cable up to an overhead room light and look into the scope end of the light cable. If the pattern contains any black spots, the light cable may be worn out and may require replacement. • Ensure the light cable is of an adequate size for the application. The cable diameter may be too small to provide adequate light transmission for the medical video camera in the endoscopic application. Excessive glare in the video
18
• Ensure the electronic shutter on the camera is operating properly to control the video signal brightness. If further light reduction is required, decrease the light source brightness with the down button.
Cleaning and Maintenance Cleaning the X8000 WARNING
UNPLUG THE X8000 BEFORE CLEANING THE UNIT.
1.
Clean the external surfaces of the X8000 using a cloth or sponge dampened with a mild detergent or disinfectant.
2.
Clean and maintain the light cable according to the manufacturer’s instructions.
Caution
Do not use any abrasive cleaners. Do not allow any liquid to drip into the unit.
Caution
Do not sterilize or immerse the X8000.
Caring for the Bulb Module The X8000 uses a Xenon bulb, which has a guaranteed life of 500 hours when used properly. Always follow these guidelines to ensure maximum bulb life: 1.
Clean the bulb face with alcohol and a cotton swab if it is touched. Dirt or oil on the bulb face will cause the bulb to heat unevenly and fail.
2.
Do not power on and off the bulb in rapid succession. Allow the bulb to run for at least five minutes once it has been powered on. Failure to do so can cause the bulb to rapidly darken and fail.
3.
The X8000 has been designed to start the bulb under most conditions, even when it is hot. However, if the bulb does not start within 10 seconds after the X8000 has powered on, turn the unit off and wait at least five minutes for the bulb to cool before restarting. Further attempts to start the bulb can damage the bulb and possibly the internal circuitry.
19